Influence of Different Surface Modifications on the Reverse Torque Values of Abutment Screws: An In Vitro Study

by
Merve Dede
1,
Ozgun Yusuf Ozyilmaz
1,*,
Ozge Doganay Ozyilmaz
2 and
Gamze Hanci Abay
3
1
Department of Prosthodontics, Faculty of Dentistry, Istanbul Galata University, 34430 Istanbul, Turkey
2
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Bezmialem Vakif University, 34093 Istanbul, Turkey
3
Private Practice, 34570 Istanbul, Turkey
*
Author to whom correspondence should be addressed.
J. Funct. Biomater. 2026, 17(9), 475; https://doi.org/10.3390/jfb17090475 (registering DOI)
Submission received: 20 July 2026 / Revised: 6 September 2026 / Accepted: 16 September 2026 / Published: 18 September 2026

Abstract

Evidence regarding tantalum nitride (TaN)-coated gold abutment screws is lacking, and studies on diamond-like carbon (DLC)-coated and anaerobic adhesive-coated abutment screws remain limited. This study evaluated the reverse torque values (RTVs) of abutment screws with different surface modifications. Sixty abutment screws from the same implant system were divided into five groups (n = 12): gold, TaN-coated gold, titanium, DLC-coated titanium, and anaerobic adhesive-coated titanium. Abutments were tightened according to the manufacturer’s recommended torque. Specimens were then subjected to cyclic loading (50 N, 1 Hz, 480,000 cycles) in a dual-axis chewing simulator, after which RTVs were evaluated. ANOVA showed a significant difference only for the uncoated gold screw group (p < 0.05). The lowest RTVs were observed in the gold screws. Although the differences among the modified screw groups were not statistically significant, surface-treated screws demonstrated higher and more stable RTVs than their uncoated counterparts. TaN-, DLC-, and anaerobic adhesive-treated screws exhibited higher RTVs following cyclic loading, suggesting that these surface modifications may enhance screw stability and reduce the risk of abutment screw loosening.
Keywords: implant; abutment screw; torque; reverse torque; abutment screw surface implant; abutment screw; torque; reverse torque; abutment screw surface

Share and Cite

MDPI and ACS Style

Dede, M.; Ozyilmaz, O.Y.; Doganay Ozyilmaz, O.; Hanci Abay, G. Influence of Different Surface Modifications on the Reverse Torque Values of Abutment Screws: An In Vitro Study. J. Funct. Biomater. 2026, 17, 475. https://doi.org/10.3390/jfb17090475

AMA Style

Dede M, Ozyilmaz OY, Doganay Ozyilmaz O, Hanci Abay G. Influence of Different Surface Modifications on the Reverse Torque Values of Abutment Screws: An In Vitro Study. Journal of Functional Biomaterials. 2026; 17(9):475. https://doi.org/10.3390/jfb17090475

Chicago/Turabian Style

Dede, Merve, Ozgun Yusuf Ozyilmaz, Ozge Doganay Ozyilmaz, and Gamze Hanci Abay. 2026. "Influence of Different Surface Modifications on the Reverse Torque Values of Abutment Screws: An In Vitro Study" Journal of Functional Biomaterials 17, no. 9: 475. https://doi.org/10.3390/jfb17090475

APA Style

Dede, M., Ozyilmaz, O. Y., Doganay Ozyilmaz, O., & Hanci Abay, G. (2026). Influence of Different Surface Modifications on the Reverse Torque Values of Abutment Screws: An In Vitro Study. Journal of Functional Biomaterials, 17(9), 475. https://doi.org/10.3390/jfb17090475

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop