1. Introduction
Since the early days of hemodialysis, which is based on the diffusion of uremic toxins through the membrane, the counter-current flow of blood in fiber lumen versus dialysate in the shell side has been employed in clinical practice to treat End-Stage Renal Disease (ESRD) patients. However, in recent decades, it has been demonstrated (e.g., [
1,
2]) that other employed hemocatharsis modes (i.e., high-flux hemodialysis or expanded hemodialysis—HDX and hemodiafiltration—HDF), involving significant fluid/plasma convection to dialysate, improve the removal of uremic toxins from ESRD patients; nevertheless, they introduce complications regarding the determination of hemocatharsis performance parameters, which are briefly outlined as follows.
For convenience, the generic acronyms HF (hemofilter) and HC (hemocatharsis) will be used to designate, respectively, the membrane module and the various treatment modes. Key standardized parameters, employed to characterize HF performance, include the ultrafiltration coefficient K
UF and the sieving coefficient SC, defined [
3] as follows:
where Q
UF and TMP are the ultrafiltrate rate (mL/min) and the effective trans-membrane pressure, accounting for osmotic pressure difference; C
UF, C
fo, and C
f1, which designate the specific species concentration in the ultrafiltrate, fiber/lumen inlet, and fiber/lumen outlet streams, respectively. For the determination of both K
UF and SC, in vitro, standard methods (i.e., ISO, 8637-1 [
3]) are commonly employed, involving a “closed system” [
3], i.e., without a dialysate feed, only an ultrafiltrate outlet, and a blood-side inlet/outlet flow, using either blood or plasma. Under such conditions, there is a unidirectional trans-membrane flow (i.e., ultrafiltration) from the blood to the dialysate side. Typical K
UF and SC
single values are commonly reported for commercial HF (e.g., [
4]), which are obtained under the specific blood-side feed flow rate Q
bin. It should be also noted that K
UF is proportional to the effective hydraulic permeability K
P, as follows:
for the HF membrane surface area S, m
2. Additionally, this K
P definition holds only if there is a unidirectional trans-membrane flow from the blood to the dialysate side.
However, for the following two reasons, the aforementioned standardized/reported single K
UF and SC numerical values are of dubious usefulness for the characterization of the HF performance under the actually prevailing clinical conditions: (i) As is well known (e.g., [
1,
2]), with the currently favored high-flux membranes, particularly in the expanded hemodialysis (HDX) mode, the externally controlled ultrafiltration rate Q
UF is not unidirectional. In fact, Q
UF is the net value of the “forward filtration” (blood- to dialysate-side) rate in the front part, minus the “back-filtration” rate (from the dialysate to the blood side) in the rear part of the HF module [
1,
2]. This bidirectionality of the flow, which is different from the unidirectional flow mode of the ISO standard [
3], employed for determining both the K
UF and SC values, creates problems in clinical data treatment and interpretation (e.g., [
5,
6,
7,
8]). Moreover, under such bidirectional flow conditions, the K
UF values have no direct quantitative relation with the effective permeability K
P. (ii) The HF performance is characterized by significant temporal and spatial variability, depending on the implemented HC mode and the imposed flow rate and pressure conditions. For fixed feed flow rates, the temporal HF performance variability (reflected in the varying K
UF and SC) is quite strong, particularly in the early stage of the HC treatment/session. Therefore, the commonly reported single K
UF and SC “standard” values are merely indicative and of unclear physical significance, as is also evident from recent studies [
5,
6]. Furthermore, it is well known (e.g., [
9,
10]) that HF membrane “fouling”, mainly by proteins, is responsible for the gradual reduction in the effective HF membrane permeability and the aforementioned temporal variability. The need to address and resolve the above issues is crucial for improving the HF performance evaluation and motivates this work.
Of particular interest to the present study are results and observations regarding the early stage of membrane–plasma interaction and its impact on membrane permeation and species rejection characteristics. Based on early in vivo studies, Rockel et al. [
9] reported that the sieving coefficient of several low-molecular-weight proteins decreases (rather sharply) on membrane exposure to blood, with a tendency to stabilize after approx. 20 min. The authors attributed this tendency to the initial/transient period of formation of a protein layer (or “secondary membrane”) on the HF membrane surfaces. Langsdorf and Zydney [
10], working with flat-sheet Cuprophan and PAN membranes, proposed a “two-layer membrane model” involving a fouling layer (comprised of adsorbed plasma proteins) acting as an additional resistance to permeation in series with that of the membrane itself. Boschetti-de-Fierro et al. [
11] experimented with dextran sieving to characterize some high-molecular-weight cut-off membranes (before and after exposure to blood). They observed that the sieving coefficients after blood contact exhibited a significant reduction compared to the initial values obtained with the pristine membranes over a seemingly transient period of ~20 min, thus essentially corroborating the earlier results. It should be added that the temporal variability of the HF parameters has not been quantified or modeled for predictive purposes.
In summary, a significant research priority emerges from the above brief literature review, i.e., the need to determine and correlate, as best as possible, the effective HF membrane permeability K
P and related process parameters under sufficiently uniform local TMP and unidirectional (blood- to dialysate-side) flow, focusing on the early stage of the hemocatharsis process. Co-current flow (
Figure 1) is employed to implement this approach. As a next step, using the results from such a study, one could proceed to model and predict the HF module performance under various conditions for clinical applications. It should be stressed that the case implemented in this study of co-current (blood- and dialysate-side) flow with significant convection/ultrafiltration has not been dealt with in the HC literature so far. Only very limited work has been reported [
12] using conventional hemodialysis, where convection is absent and only species diffusion occurs.
Experiments were performed with a commercial hemofilter with
co-currently flowing fluids at the lumen and shell sides. Under such conditions, the local/axial variability of the TMP, the fluid permeation rate, foulant deposition, and species rejection tends to be reduced. Therefore, by determining the (inlet and outlet) flow rates and pressures at the two end-surfaces of the cylindrical/active HF section (
Figure 1), one can obtain the
mean parameter values fairly representative of the entire HF. Two types of fluids were employed in this work at the blood side (i.e., human plasma and an aqueous BSA solution as a reference). A theoretical solution to the co-current flow mode of the HF operation, for Newtonian fluids, is also presented in support of the experimental work.
3. Experimental Part
3.1. Materials and Equipment
The commercially available module
Elisio 19H (Nipro Medical Corporation, Osaka, Japan), comprising the Polynephron™ (polyethersulfone) high-flux hollow fiber membranes, with a 1.9 m
2 membrane area [
4], was employed in all of the HC simulation tests using BSA solutions and human plasma. Its main characteristics are listed in
Table S1 (Supplement).
Bovine serum albumin (BSA, Sigma-Aldrich, Darmstadt, Germany) at a concentration of 3 g/L was used as the feed solution at the blood side in some of the tests. The main experiments were conducted using frozen human plasma, kindly provided by the “George Papanikolaou” General Hospital of Thessaloniki. The plasma was received in appropriate packaging after testing for several viruses, such as hepatitis B virus (HBsAg), based on the official guidelines for experimental purposes. All of the experiments were carried out using deionized water at the dialysate side.
The experimental set-up, detailed in previous publications [
13,
14], was equipped with two feed vessels with a capacity of 5.0 L and 5.5 L for the blood- and dialysate-side fluids, respectively. Two magnetic drive gear pumps (with a flow rate range of 0–1000 mL/min, type MS204, Fluid-O-Tech, Milan, Italy) were employed for both the blood and dialysate feed solutions; they were placed at the top of the module for both solutions (
Figure 1). Additionally, four precision pressure transducers (with a range of 0–15 psi, type A-10, Wika, Klingenberg, Germany) were installed at the inlet and outlet at each side of the module. The blood inlet, dialysate inlet, and outlet flow rates were monitored using three flowmeters (101-Flo-Sequate data, McMillan Co., San Francisco, USA). Furthermore, the experimental set-up was equipped with a Programmable Learning Controller, PLC (CMT Series, Weintek, Taipei, Taiwan), for the continuous adjustment, monitoring, and recording of all operating parameters, including pressures and flow rates. Data were continuously monitored and recorded every 30 s.
3.2. Experimental Conditions—Test Protocols
The following three types of experiments are reported here, all performed under co-current flow: (1) a few tests with water for the preliminary assessment of the co-current mode; (2) experiments using an aqueous BSA solution at the blood side in recycling mode; (3) experiments using human plasma at the blood side in the “once-through“ mode. After some type (2) and (3) tests, hemofilter cleaning/washing took place, followed by the measurement of clean water permeability.
In the type (2) tests, 3 L of BSA solution (i.e., 3 g/L in deionized water) was pumped into the hollow fibers at the blood-side inlet (
Figure 1), while deionized water was fed co-currently into the shell/dialysate sides. It is noted that this aqueous BSA solution is Newtonian. Two sets of flow rates were tested in this operating mode, i.e., 300/120 mL/min (denoted as experiment “A2”) and 300/360 mL/min (experiment “A4”) for blood/dialysate flow rates, respectively. Intermediate HF cleaning was performed as noted above.
In the type (3) experiments, human plasma was fed at the blood side in the co-current flow mode. However, in these experiments, plasma was pumped “once-through” in the hemodialyzer to maintain a constant feed quality and to closely simulate hemocatharsis conditions. In all of these tests, the flow rates were 250 and 200 mL/min for the blood and dialysate sides, respectively, and three new/clean HF modules (designated as E1, F1, and F2) were employed. It should be stressed that the “once-through” flow mode with the plasma was selected because the preliminary tests in the plasma-recirculating mode exhibited a significant TMP pressure increase. The latter was attributed to the increasing protein concentration in the feed, possible coagulation effects, as well as excessive/unrepresentative membrane fouling. It should be also noted that the plasma samples in tests E1, F1 were used ‘as received’; however, the feed to test F2 was pre-filtered to remove some visible small clots of undetermined origin.
The previously described experiments began after removing air from the system and the module by circulating deionized water in both lumen and shell sides. This step was followed by a co-current clean water test (to determine the clean water permeability KP) using deionized water at both sides, lasting for 10 min at the same flow rates as the main experiment. It is noted that (at the set flow rates) the data recording (i.e., designated as time zero) started at the moment the lumen side was filled with the BSA solution (or human plasma).
In each test, the samples were collected from the blood-side feed solution and from both the blood- and dialysate-side outlets. Determination of the albumin and total protein concentration in those samples was carried out in Thessaloniki Gen. Hospital “G. Papanikolaou”. Immuno-turbidimetry was used to determine the albumin in the plasma and dialysate samples by employing ALBU2 (Tina-quant Albumin Gen.2, Roche Diagnostics GmbH, Germany). For the total protein concentration in the plasma and dialysate samples, colorimetry (TP2, cobas c, Roche Diagnostics) and nephelometry (TPUC3, cobas c) were employed, respectively.
Two flow modes were employed for cleaning/washing the used HF modules. In the first cleaning mode, deionized water was pumped exclusively into the lumen side of the dialyzer (entry #1,
Figure 1), with the two valves at the dialysate side closed. The aim was to dislodge the proteins and other foulants (i.e., the “gel” layer) covering the membrane/lumen inner surface. In the following cleaning mode, deionized water was fed at one dialysate-side inlet (entry #3,
Figure 1) and exited from the blood-side outlet (exit #2,
Figure 1), whereas the valves at the blood-side inlet and the other dialysate-side outlet (exit #4) were closed. Cleaning with this mode aimed to remove foulants adsorbed in the membrane pores by the pressurized water, permeating from the shell side into the fiber lumen.
5. Discussion
This study has demonstrated that the co-current flow of blood/plasma and dialysate leads to relatively uniform (axially) local TMP, thus ensuring a unidirectional ultrafiltration flow along the entire HF under conditions representative of those prevailing in clinical practice. Since the local TMP is a driving force for fluid/plasma permeation and protein deposition (e.g., [
15,
16]), such TMP spatial uniformity tends to reduce the axial variation in membrane fouling, which similarly affects the effective permeability K
P of the HF, thus benefiting its overall performance. It should be stressed that the observed
temporal TMP variability is also rather small in typical tests with plasma (
Figure 5a). These advantages of the co-current flow compared to the presently practiced counter-current flow facilitate the study of the evolution of key HC parameters and the acquisition of accurate data, particularly under the presently favored high-convection HC modes [
1,
2]. Therefore, co-current flow merits particular attention in future studies and the further investigation of its application in clinical practice.
To interpret the present data and relate the HF effective permeability K
P to membrane fouling, the following general form of the Darcy law is invoked, applicable to
unidirectional transmembrane flow:Here, the quantity J = Q
UF/S is the mean fluid permeation flux (in mL/min/m
2), S (m
2) is the HF effective membrane area, μ is the fluid viscosity, and [ΔP − Δπ] is the mean/effective trans-membrane pressure accounting for the osmotic pressure difference. The total resistance to permeation (R
m + R
c) is comprised of the membrane resistance R
m and an additional resistance R
c, often called “
secondary” or cake/gel resistance [
9,
10] due to fouling/deposit formation. However, the clean/unused membrane resistance R
m is also increased due to the mechanisms of
pore constriction and
pore blockage (e.g., [
19,
20]) by tightly adsorbed organic species, i.e., by proteins in the case of hemocatharsis [
18,
21]. It should be noted that these resistances are related to the directly measured quantities in this study, K
p and K
UF, as follows:
The observed in tests with human plasma (
Figure 6) showing a rather strong decline in the effective permeability K
P (during the initial ~30 min) is most likely due to the mechanisms of
pore constriction and
pore blockage, which tends to increase the membrane resistance R
m, as also demonstrated in previous studies [
15,
19,
20]. A protein ‘gel’/cake-layer formation (represented by R
c) possibly follows, caused by larger proteins and possible agglomerates depositing on the initially adsorbed proteins (e.g., [
22]). The observed rather small decline of total resistance (R
m + R
c), beyond the initial (~30 min) period, may indicate that pore constriction/blockage is no more effective and that the gel/cake layer is relatively porous. In support of this interpretation is also the small reduction in the resistance (R
m + R
c) beyond an initial period, exhibited by the BSA solution data (
Figure 4), in particular with a reused membrane (test A4) that has apparently suffered irreversible fouling due to pore constriction and blockage. However, additional detailed, well-focused experiments are required to clarify and quantify the effect of the deposited protein mass on the HF permeability and to facilitate modeling.
The proposed here, for the first time, key parameter M
m (permeation mass flux of particular species, e.g., albumin) is considered very useful for direct/facile computation of the total/cumulative mass [M] of species leaking in the dialysate (e.g., albumin [
23]) as well as for future modeling studies. A typical realistic example is provided here using the data on M
m depicted in
Figure 7. For instance, the data on albumin permeation from test E1 show that the temporal M
m variation is very well represented by a power function, i.e., M
m = 45.897t
−1.03. Making the fair assumption that this function holds for the entire 4-h period of an HC session, one can readily predict the cumulative (or total) albumin loss to permeate [M] at various times using a simple integration, as shown in
Figure 9. It should be noted that the trend as well as the magnitude of these albumin loss projections are in general accord with similar projections by Zawada et al. [
24]. However, in that study [
24], the less-sensitive sieving coefficient (SC) data were employed, as well as other approximations commonly made for such predictions, e.g., [
25].
Regarding the SC data, which are widely used as a performance index (with significant uncertainties in clinical practice, e.g., [
7]), the following remarks can be made in comparison with the respective mass flux M
m values:
- i.
An inherent weakness in the definition of the SC is due to the quite small magnitude of the numerator (the dialysate concentration) compared to the denominator (blood concentrations). This leads to the greatly reduced sensitivity of the SC (and increased error margins), thus rendering the SC inappropriate for predictive purposes.
- ii.
The SC can be reliably determined in vivo
only under unidirectional ultrafiltration flow from blood to dialysate. Therefore, in HC modes, such as expanded hemodialysis [
1], involving both ultrafiltration and back-filtration, the true value of the ultrafiltrate concentration C
UF (and therefore SC) cannot be determined.
The new data (i.e.,
Figure 6 and
Figure 7) suggest that the parameter M
m, particularly for the middle MW species (such as albumin and some other proteins) depends on the effective permeability K
P, exhibiting a qualitatively similar temporal evolution. Considering that it is relatively easy to quantify the temporal variability of K
P, a correlation of M
m with K
p would be useful. Indeed, as shown in
Figure 10a,b, the data from tests E1 and F1 with plasma under co-current flow exhibit a rather strong (exponential-type) dependence of M
m on K
p for both albumin and total proteins. However, although the trend is qualitatively similar between the two tests (E1 and F1) performed under the same conditions, there is some quantitative difference to be clarified in future studies. It should be also added that there is a somewhat weaker dependence of the total protein mass flux M
m on M
p (compared to that for albumin), as shown in
Figure 11, which is possibly due to the fact that the fouled membrane is more permeable by the smaller-size/MW proteins of the human plasma than by the albumin molecules.
6. Conclusions
This study, involving realistic in vitro HC experiments with human plasma (supported by tests with BSA solution), demonstrates that the co-current flow direction of blood/plasma and dialysate leads to an axially fairly uniform TMP, particularly under the presently favored high-convection HC modes. In turn, the TMP axial uniformity favors spatially uniform membrane fouling, as well as relatively uniform effective hemofilter permeability KP and overall performance. These conditions facilitate the study of the temporal evolution of key HF parameters, as summarized below. Therefore, the co-current blood and dialysate flow mode clearly deserves particular attention and additional testing towards clinical applications.
Accurate/representative data on the temporal evolution of key HC parameters were obtained, focusing on the HF effective permeability Kp and the mass flux Km of permeating albumin and total proteins into the dialysate. Kp can be readily determined in co-current flow and exhibits significant temporal variability, particularly during the initial HC period of ~30 min.
The parameter Km introduced here (for the first time for the HC) affords the following clear advantages compared to other similar indices, such as the widely used sieving coefficient (SC): (i) it can be easily determined in vivo under unidirectional trans-membrane flow/ultrafiltration; (ii) it is physically sound (also accounting for the membrane surface area) and is more sensitive than the SC; (iii) it facilitates the accurate determination of the cumulative permeating mass [M] of specific species/solutes, unlike the presently used approximate methods; (iv) it can be readily correlated with other key parameters, particularly the effective Kp. Such correlations are expected to facilitate modeling and further research towards improved HC clinical applications.