Next Article in Journal
Experimental Investigation of an Incremental Contact Model for Hyperelastic Solids Using an In Situ Optical Interferometric Technique
Previous Article in Journal
Investigation of Cavitated Flow in Water-Lubricated Bearings Considering Surface Roughness, Thermal, and Elastic Effects
Previous Article in Special Issue
Green-Emitting Carbon Quantum Dots: Highly Sensitive Temperature Sensing Probe in Nanocomposite and Lubrication System
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Effect of Oil Acoustic Properties on Film Thickness Measurement by Ultrasound Using Spring and Resonance Models

by
Alvaro S. Piovesan
1,2,
Michele Schirru
3,
Fabio Tatzgern
3,
Jorge L. B. Medeiros
1 and
Henara L. Costa
1,2,*
1
School of Engineering, Universidade Federal do Rio Grande, Rio Grande 96203-900, RS, Brazil
2
Graduate Program on Materials Science and Engineering, Universidade Federal de Pelotas, Pelotas 96010-610, RS, Brazil
3
Austrian Center for Competence on Tribology (AC2T), 2700 Wiener Neustadt, Austria
*
Author to whom correspondence should be addressed.
Lubricants 2024, 12(4), 108; https://doi.org/10.3390/lubricants12040108
Submission received: 29 December 2023 / Revised: 8 March 2024 / Accepted: 20 March 2024 / Published: 25 March 2024

Abstract

:
The principle of reflection of ultrasonic waves at lubricated interfaces has been widely studied in recent years using different models. In this work, two different models (the spring model and the resonance model) were used to verify the influence of the acoustic properties of four different lubricating oils. A simple three-layer configuration was used, where carefully prepared, well-controlled gaps between stainless steel plates were established to accommodate a drop of oil. Optical measurements showed that the gaps formed were: gap 1 = 11 µm, gap 2 = 85 µm, gap 3 = 100 µm, and gap 4 = 170 µm. The smaller gap (11 µm) was found to be in the limit measurement range using the spring model for the sensor used in this work (14 MHz), whereas the resonance method was used for the thicker gaps. For the resonance model, the use of the phase spectra helped the identification of the resonance frequencies. The results showed good agreement between the measured thicknesses and the nominal gap values. There was little effect of the acoustic properties of the oils on the measured values, with the largest discrepancies found for the oil with the highest speed of sound (PAO4). This new way to characterize oil properties in a thin gap, where the material and geometry of the contact are fully characterized, enables us to compare different measurement methods and understand their sensitivity when testing similar materials of the same class of lubricants, as small deviations are crucial in real-life applications.

1. Introduction

Ultrasonic waves can travel through different media due to the excitation and displacement of particles in the medium. Generally, this excitation is carried out using piezoelectric sensors, which are elements capable of producing sound waves in response to stimuli from a voltage pulse. When this piezoelectric element is coupled to a solid material, a mechanical displacement propagates through this material [1].
The principle of measuring different characteristics of tribological contacts using ultrasonic reflection was first established at the University of Sheffield around two decades ago [2]. Since then, it has been used for measuring lubricant film thickness [3,4,5], lubricant viscosity [6], thermoelastic compliance [7], contact forces [8], friction [9,10], contact dimensions [11], contact surface roughness [2], among other relevant information about tribological contacts. In addition to a number of works under laboratory conditions, this technique has been adapted to a range of real applications, varying from typical engineering applications such as the piston/liner system in engines [3,7,12,13], sliding bearings [14,15,16,17], roller bearings [18,19], and metal-forming applications [11,20], to a much broader realm of applications, such as measuring the thickness of leaves in tree canopies [21].
In lubricated systems, monitoring the thickness of the lubricating film layer is fundamental to guaranteeing the correct functioning of mechanical components and equipment [22,23]. One of the main difficulties for real-time monitoring of the lubricating film layer is the complexity involved in developing non-invasive systems that can be directly applied to mechanical components [24]. Lubricating film thickness measurement techniques based on optical [25] and electrical [26] principles have been widely discussed by several authors [27]. Although both methods prove to be effective, they have limitations. In the case of electrical methods, there may be a need to electrically insulate the contacting elements. As for the use of optical methods, it is necessary to use transparent contacting elements, which makes application outside the laboratory environment difficult. The ultrasound film thickness measurement technique does not have such limitations. Furthermore, it has proven to be a very promising technique, mainly due to its versatility and measurement capacity [13].
In a lubricated contact, the ultrasonic wave will have part of its energy reflected and part transmitted when it encounters a thin layer of fluid. The amount of this reflection is then associated with the acoustic impedance of the materials creating the interfaces. The greater the disparity of this acoustic property of the materials, the greater the proportion of the reflected wave. The part of the wave transmitted through the solid-oil interface will be reflected again upon reaching a new interface. These successive reflections of the waves tend to overlap into very thin film layers (in the order of the pulse wavelength) [28,29].
Several reliable models for measuring film thickness using ultrasonic sensors have been developed, allowing the investigation of a wide range of film thicknesses. In fact, the different methods complement each other, suppressing the limitations of each one [5,29,30,31]. Some criteria, such as the magnitude of the film thickness, the geometry of the lubricated contact, and the characteristics of the ultrasonic pulse, need to be considered to guarantee the accuracy of the chosen model [30].
The information regarding the incident and reflected waves has been used in the different models proposed in the literature to compute lubricant film thickness. The spring model considers the reflection coefficient (R) of the reflected waves to extract information about the lubricant layer [19]. When the spring model is used to obtain film thicknesses, it is considered that the behavior of the ultrasonic waves is governed by the stiffness of the lubricant layer (K), which depends on the mass modulus (B) and the thickness of the film layer (h). This is due to the fact that the film layers investigated by the spring model are very thin (below 10 µm), causing the interaction of sound waves with this thin liquid layer to present a very high first resonance frequency. The spring model uses the information obtained in the amplitude of the returned signal together with the acoustic properties of both the fluid and the materials of the contacting surfaces [1,13,22,32,33]. A simplified relationship between the stiffness of the film layer (K) and the reflection coefficient (R) was presented by Reddyhoff, Dwyer-Joyce, Zhang, and Drinkwater [22] to define the film thickness between two surfaces:
h = ρ c 2 π f z ( R 2 1 R 2 )
where ρ is the density of the liquid, c is the speed of sound in the lubricant, z′ is the acoustic impedance of the materials on each side of the fluid (in this case the same material was considered for both sides), f is the center frequency of the ultrasonic sensor, and R is the reflection coefficient. The signals referring to the pulses returned both from the reference interface and from the lubricant layer, first obtained in the time domain, are converted to the frequency domain through a fast Fourier transform (FFT). The amplitude of the signal reflected by the oil film layer (Ameas) and a reference amplitude (Aref) obtained without the presence of lubricant are used to obtain the reflection coefficient. To perform the film thickness calculations, the reflection coefficient (R) can be calculated by Equation (2) [17]:
R = A m e a s A r e f R r e f
As the lubricant film gets thicker, the spring model no longer results in accurate values; thus another model known as the resonance model can be used to determine subsequent thickness [14,15]. Similarly to the spring model, the resonance model approach also considers that the behavior of the lubricating film layer is governed by the reflection coefficient. However, this model relates the thickness of the lubricating film layer to the resonant frequency of the incident wave along with its velocity in the liquid medium. The resonance frequencies can be defined in the regions where the reflection coefficient, R, presents periodic minima. For the calculation of the film thickness via the resonance model, the minimum coefficient of reflection is considered when the resonance frequencies occur within the bandwidth region of the ultrasonic sensor. The relationship between film thickness and resonance frequency range is given by Equation (3) [34,35]:
h = c 2 f m
where c is the speed of sound in the lubricant and ∆fm corresponds to the interval between two minimum frequencies. Sometimes the periodicity between these minima is not evidenced, and only a resonance is perceived in the spectrum of the reflection coefficient in the frequency domain. When this occurs, the frequency corresponding to this minimum is considered as the ∆fm interval itself. Thus, very similar to Equation (3), the thickness of the film layer can be calculated by [36]:
h = c m 2 f m
Despite the increasing use of ultrasound to monitor tribological contacts, including works that have used the technique very successfully in real mechanical components under industrial conditions, it is important to point out that many factors can affect the measurements. For example, the type of oil and temperature affect the acoustic properties of the lubricants, but it is common that the models use room temperature values, and sometimes the values for the speed of sound in the lubricant are not measured but taken as average values for a similar lubricant. Moreover, the effects of such effects may vary depending on the model used, since the film thickness values may vary either linearly or with the square of the speed of sound for the different models. Although the models used in this work have been widely used in the literature, in this work we utilized a calibrated gap between the steel plates so that the real geometry of the lubricant film thickness was accurately known. Therefore, it was possible to compare and identify the limitations of the models for oils with different acoustic properties. Since the steel-to-steel gap was completely characterized in terms of geometry and materials, this test setup can be proposed as a new way to characterize oil properties in a thin gap. Moreover, although not addressed in the literature regarding acoustic measurements of tribological contacts, when air (or other gaseous matter) is trapped in the gap, the measurement could be affected, so that an incorrect measurement could be obtained. This setup avoids this problem by overflowing the gap; since the gap is not closed, air should not be trapped, thus ensuring that only the lubricant’s properties will be addressed.
The spring model and the resonance model were used to verify the relationship between the acoustic properties of the measured film layer in terms of its stiffness, K, and its ability to reflect incident ultrasonic waves. For the spring model, a nominal gap geometry around 10 µm was chosen because it is around the limit recommended in the literature for using the spring model [34]. The other three gaps with a thickness greater than 10 µm were employed for the resonance model. The phase variation of the incident and reflected waves was used to assist in the identification of resonance frequencies. Four oils, with different properties, were allocated between stainless steel plates so that they were kept static and with a previously defined spacing. A sensor with a central frequency of 14 MHz, coupled to the opposite face of one of the plates, was used to emit and receive ultrasonic signals. The aim was to identify the limitations of the models in terms of the acoustic properties of the fluid for a well-controlled gap geometry.

2. Materials and Methods

Figure 1 shows a scheme for the experimental setup used to measure lubricant film thickness by ultrasound. Two stainless steel plates separated by a lubricant layer created a three-layer system (steel–oil–steel). A piezoelectric sensor with a central frequency of 14 MHz was fixed onto the upper plate. The excitation of the sensor was performed using a function generator to create arbitrary functions of 100 MHz and a sampling rate of 250 megapoints/s, which provided a burst of sine waves in the range of the central frequency of the sensors. The reception and acquisition of signals emitted and returned by the sensors were performed using a 100 MHz oscilloscope with a sampling rate of 2 gigapoints/s. A dedicated Labview (Version 18.0) interface was created for data acquisition and processing.
The plates were manufactured from the same block of martensitic stainless steel 1.4313—X3CrNiMo13-4, both with identical dimensions (90 × 90 × 14 mm). During the execution of the tests, the piezoelectric sensor, fixed on the upper plate, operated in pulse-echo mode. The bottom plate is responsible for delimiting the amount of lubricant between the two plates. For this, two standard steel gauges with the same thickness were fixed at the ends of the face of the lower plate. The steel gauges used on the lower plate limit the approximation between the two plates, thus creating a channel with a height corresponding to the thickness of the steel gauges.
In this work, the surface finish of the plates used was brought to its best condition according to the needs and characteristics of the film thickness measurement models adopted. The plates were carefully ground to obtain the lowest achievable roughness. Attempts were made to achieve even lower roughness levels via lapping, but that led to a deterioration of the surface flatness. Even if lapping can confer an even better roughness condition, the surface profile tends to vary in flatness. In Figure 2, it is possible to notice the variation in the flatness of the bottom plate when a very refined polishing was applied. This indicates that there really is a limit to which the decrease in roughness does not interfere with the surface profile. For the film thickness measurement settings and conditions adopted in this work, it is essential that the surfaces that limit the lubricating oil are parallel to each other in order to ensure the accuracy of the results.
The 3D topographic maps presented in Figure 3 represent the best possible surface finish for the lapping process used so that the surface profile was maintained. The plate profile was measured using an Alicona 3D optical microscope. The quantification of the surface roughness of the plates was carried out via the 2D roughness parameters presented in Table 1.
After the plates were polished to their best condition of surface finish, standard steel gauges were glued at the ends of the lower plate. Figure 4 shows the 3D topography map of the lower plate after gluing the gauges that formed the channel intended to restrict the oil sample between the plates for different gaps. To facilitate the visualization of the channel heights created by the standard steel gauges on the lower steel plate, cross-sections with topographic profiles are shown in the left column. The measurements for the smaller gap (nominal thickness = 10 µm) were carried out using an Alicona 3D optical microscope. The channels formed in the bottom plate presented heights close to those specified by standard steel gauges that were fixed in the lower plate, around 11–12 µm. The peak in Figure 4b (close to 78 mm) does not concern the height of the channel and can be considered as a reflection measurement error. For the thicker gaps, the measurements used a confocal microscope model Leica, TCS SP8. Since the measured areas that could be measured by the confocal microscope were substantially smaller, only one side of the gap could be evaluated, but no significant differences were found when measurements were carried out for both sides of the channel were compared.
To use the spring model, a reference signal is necessary in addition to the signals reflected at the steel–liquid interface. The reference signals are obtained when the upper plate does not contact with any oil sample, that is, it considers a steel–air interface. An example of how the signals are presented can be seen in Figure 5, showing both the reference and the measured signals in the presence of a drop of oil between the two plates. The signals are shown in the time domain (Figure 5a) and in the frequency domain after the fast Fourier transform (FFT) (Figure 5b). The determination of a reference signal is carried out using only the top plate of the measuring system. When an ultrasonic wave reaches the boundary of the plate’s interface with the air, it will be reflected almost in its entirety, presenting a large amount of information about the incident wave. The spring model was used only for the smaller gap of 11–12 µm, which is around the limit for the piezoelectric sensor used in this work, with a central frequency of 14 MHz.
For the resonance model, Equations (3) and (4) were used to identify the lubricant film thickness present between the metallic plates for the gaps of 85, 100, and 170 µm. The precision of the measurements using this model is linked to the extraction of resonant frequencies that occur within the useful frequency region of the sensor used; hence, the importance of correctly identifying the resonance frequencies in the reflection coefficient spectrum.
Four different oils were used in the measurements. The choice of lubricants took into account the need for testing oils with different viscosities, as well as choosing both a base oil (PAO 4) and a wide range of formulated lubricants. The time-of-flight method was also used, but simply to determine the speed of sound propagation (c) in each of the oils and stainless-steel plates. For this, the response time for ultrasonic waves imposed on lubricant samples when they were limited to a container with known dimensions was recorded. From the values of the speed at which sound propagates in a medium and having information on the density of the material (ρ), it is possible to determine its acoustic impedance (z = cρ). The acoustic properties of the four lubricating oils used in this study, together with the acoustic characteristics of the stainless-steel plates, are presented in Table 2, where the speed of sound for the steel and each lubricant was measured (three repetitions) at a temperature of 25 °C using a commercial ultrasound system model Phasor XS from manufacturer GE, Boston, MA, USA.
The measurements were performed by depositing a drop of oil on the lower steel plate and then positioning the upper plate, containing the piezoelectric sensor, on the metal shims. Each measurement was repeated six times. The test temperature was kept constant at around 25 °C, which is the same temperature for the measurement of the material´s acoustic properties. This is an important point, as a few degrees of difference between measurements will have a considerable impact on the results. Both the reference signal and the other signals measured during the tests were acquired in the time domain; therefore, they needed to be converted to the frequency domain through a fast Fourier transform. This conversion was performed using a VI (virtual instrument), created in the LabView (version 18.0) interface, simultaneously with the data collection.

3. Results and Discussion

First, the spring model was used to verify the agreement between the measured film thickness values and the real geometry of the thinner gap (10 µm). The spectra of the signals reflected in the oil layer for each of the lubricants used are presented in Figure 6 and Figure 7, together with the spectra of the reference signal. The signals obtained in the time domain, Figure 6, were converted to the frequency domain through the fast Fourier transform, Figure 7. With this, the maximum amplitudes were used to define the reflection coefficient (R) of each layer of oil measured.
During the measurements, it was possible to notice that the central frequency of the received signals was slightly below the central frequency of the piezoelectric sensor, around 13.6 MHz. This difference between the incident and received frequencies can be related to the attenuation of the waves when crossing the upper plate [37]. For this central frequency of the received signals, the bandwidth (−6 dB) was defined between 12.6 MHz and 14.6 MHz, as shown in Figure 7. The amplitude spectra show agreement in the behavior of the different measured layer spectra. This was already expected since the gap size and measurement conditions were kept identical for the four samples tested. The magnification in the region of the peaks of the spectra shows that for the PAO4 oil sample, there was a smaller attenuation of the reflected signal. This behavior may be related to its lower acoustic impedance compared to other oils.
From the spectra of the reference signal and the signals measured in the lubricant film layer in the frequency domain, the values for the reflection coefficient were obtained using Equation (2). From Figure 6, the reflection coefficient values referring to the central frequency can be seen in Table 3, corresponding to averages from six repetitions for each oil. When the standard deviation of the measurements is taken into account, no significant differences are observed between the average reflection coefficients for the different oils.
In Figure 8, it is possible to notice the stability of the reflection coefficient curves within the bandwidth (BW) region of the FFT spectrum. This is one of the indications that an adequate choice of measurement model was made for the investigated film thickness range. The reflection coefficient values were very close to one (≈0.98). This value is in line with the results presented by Dwyer-Joyce, Harper, and Drinkwater [33], where reflection coefficients above 0.9 were obtained when using 3 MHz sensors to determine films in the 10 µm range via the spring model. For sensors with a central frequency of 14 MHz, such as those used in this study, the reflection coefficient values are very close to one, favoring the increase in measurement uncertainty.
Figure 9 presents the thickness values within the bandwidth (BW) for the four oils measured using a 10 µm gap. The film thickness values are in good agreement with the true gap height shown in Figure 4b, confirming the suitability of the ultrasonic model for this thickness range. and the adopted configuration are suitable for the purpose of this work.
In the measurements carried out, there is a limit for the values of the reflection coefficient where measured thicknesses present acceptable values. Below 0.975 and above 0.986, these measurements showed significant variations in relation to the nominal gap (given by the gauge thickness). As the gap used in this work is close to the measurement limits for the spring model, the sensitivity of the measurement system can be easily affected by factors inherent to the measurement process. As with the reflection coefficient, the thickness of the measured lubricant film layer is governed by its stiffness, K.
Having the thickness limitation for the spring model confirmed, the same experimental configuration was used to measure the three thicker gaps using the resonance model. The presence of minimum amplitude in the reflected signal spectrum is one of the main indications that the choice of the resonance model may be the most appropriate to determine the film thickness. The minimums representing the resonance frequencies for the gaps used in the experiments are shown in Figure 10.
For the first reference thickness used (85 µm), these minima were not evidenced in the amplitude spectrum of the reflected signal within the bandwidth, but a minimum occurred slightly below the beginning of the bandwidth region. For the other two reference thicknesses, 100 and 170 µm, more than one resonance frequency was observed, and at least one of these occurred within the bandwidth. For the thicker gaps (100 µm and 170 µm), it was thus possible to identify the minimum amplitude in the spectrum of the measured signal, as can be seen in Figure 11. By identifying the minimum frequency, the thickness of each of the tested oils can be defined by Equation (4).
The amplitude minima identified in Figure 11 within the bandwidth show a behavior that goes according to what is expected by Equation (3), with a displacement in the occurrence of valleys as the thickness of the film increases. Comparing the curves in Figure 12a,b, it is noted that with the increase in thickness, the minima tend to occur at lower frequencies. Figure 12 shows the behavior of the phase angle that undergoes a sudden change when approaching resonance frequencies, crossing the zero line. From the identification of the crossing point, it is also possible to apply Equation (4) to extract film thickness values.
Comparing the spectra of the phase reflection coefficient (Figure 12) with the spectra of R (Figure 10), it is possible to observe that the points where the phase angle inversion occurs coincide with the minimum frequency that characterizes the resonances in the film layer. From the identification of these frequencies, the thicknesses were calculated by Equations (3) and (4) for the three gaps used (85 µm, 100 µm, and 170 µm), as shown in Table 4.
To evaluate the statistical significance of the differences between the oils, an analysis of variance (ANOVA) was carried out for a confidence interval of 95%. Table 5 summarizes the p-values obtained from the ANOVA, showing that at least one of the oils had a value that was statistically different from the others in the group (p-values lower than 0.05).
However, the results in Table 5 do not discriminate which oils showed statistically different values. For that, a Tukey test was carried out, which is summarized in Table 6. In this table, the oils that were different appear highlighted. For the spring model, the only significant difference occurs between the oils ISO VG6 and PAO4. For the analysis made for the resonance model, there is a significant difference almost in all cases, not only in the measurements of gap 4 (170 µm) between the oils ISO VG46 and SAE-30. Therefore, this statistical analysis showed that the acoustic properties of the oil (speed of sound) affect the measurements, since the largest difference in speed of sound was found for the PAO4 oil, which, in turn, always showed different thickness values in relation to the other oils.
The results in Table 4 showed a reasonably good agreement with the gap heights measured by optical methods. In order to compare the effects of the oil acoustic properties using the two models, Table 7 summarizes the errors calculated in relation to the nominal gap height measured by optical methods.
The larger discrepancy was observed for the smaller gap of 85 µm measured with the resonance model (gap 2), probably because the resonance occurred outside the bandwidth. Despite this, the use of the phase spectrum to confirm the resonance frequencies can be helpful when no minima for the reflection coefficient are found within the sensor bandwidth. For the larger gaps (3 and 4), the errors were up to around 6% for all the oils. When using the spring model, the errors were slightly higher, in particular for the PAO oil, which showed the highest speed of sound. For the spring model, the lubricant film thickness varies with the square of the speed of sound of the oil, so that small errors in the measurement of the speed of sound can have a larger impact on the calculated thickness.

4. Conclusions

This work measured the lubricant film thickness in a three-layer system with well-controlled gaps for oils with different acoustic properties. The results showed that:
  • When very well controlled gaps are used, the effects of the properties of the fluids can be more robustly investigated.
  • With the spring model, for all the oils, a good agreement was found between the true gap height (11–12 µm) and the measured values. For the oil with the lowest speed of sound (PAO4), the accuracy of the measurement is reduced due to the lower attenuation of the reflected signal.
  • The resonance model showed little effect of the lubricant’s acoustic properties on the measured thickness. The larger discrepancy was again found for the oil with the lowest speed of sound (PAO4).
  • When the resonance model is used for thinner films, the phase reflection coefficient spectrum can help identify the resonance frequency.

Author Contributions

Conceptualization, A.S.P., M.S. and H.L.C.; methodology, A.S.P., M.S. and J.L.B.M.; formal analysis, A.S.P., M.S. and H.L.C.; resources, M.S., J.L.B.M., F.T. and H.L.C.; writing—original draft preparation, A.S.P. and H.L.C.; writing—review and editing, M.S., F.T. and H.L.C.; supervision, H.L.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the CNPq/Brazil via the INCT CT Trib (grant number 406654/2022-0). ASP received a scholarship from Capes/Brazil.

Data Availability Statement

All relevant data are included in this manuscript.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Dwyer-Joyce, R.S. The application of ultrasonic NDT techniques in tribology. Proc. Inst. Mech. Eng. Part J J. Eng. Tribol. 2005, 219, 347–366. [Google Scholar] [CrossRef]
  2. Dwyer-Joyce, R.S.; Drinkwater, B.W.; Quinn, A.M. The use of ultrasound in the investigation of rough surface interfaces. J. Tribol. 2001, 123, 8–16. [Google Scholar] [CrossRef]
  3. Harper, P.; Dwyer-Joyce, R.S.; Sjödin, U.; Olofsson, U. Evaluation of an Ultrasonic Method for Measurement of Oil Film Thickness in a Hydraulic Motor Piston Ring. In Tribology and Interface Engineering Series; Elsevier: Amsterdam, The Netherlands, 2005; Volume 48, pp. 305–312. [Google Scholar]
  4. Kaeseler, R.L.; Johansen, P. Adaptive ultrasound reflectometry for lubrication film thickness measurements. Meas. Sci. Technol. 2020, 31, 025108. [Google Scholar] [CrossRef]
  5. Dou, P.; Jia, Y.; Wu, T.; Peng, Z.; Yu, M.; Reddyhoff, T. High-accuracy incident signal reconstruction for in-situ ultrasonic measurement of oil film thickness. Mech. Syst. Signal Process. 2021, 156, 107669. [Google Scholar] [CrossRef]
  6. Schirru, M.; Li, X.; Cadeddu, M.; Dwyer-Joyce, R.S. Development of a shear ultrasonic spectroscopy technique for the evaluation of viscoelastic fluid properties: Theory and experimental validation. Ultrasonics 2019, 94, 364–375. [Google Scholar] [CrossRef]
  7. Littlefair, B.; De la Cruz, M.; Theodossiades, S.; Mills, R.; Howell-Smith, S.; Rahnejat, H.; Dwyer-Joyce, R.S. Transient Tribo-Dynamics of Thermo-Elastic Compliant High-Performance Piston Skirts. Tribol. Lett. 2014, 53, 51–70. [Google Scholar] [CrossRef]
  8. Liu, Y.; Liu, E.; Chen, Y.; Wang, X.; Sun, C.; Tan, J. Measurement of fastening force using dry-coupled ultrasonic waves. Ultrasonics 2020, 108, 106178. [Google Scholar] [CrossRef] [PubMed]
  9. Li, X.; Dwyer-Joyce, R.S. Measuring friction at an interface using ultrasonic response. Proc. R. Soc. A Math. Phys. Eng. Sci. 2020, 476, 20200283. [Google Scholar] [CrossRef]
  10. Blanloeuil, P.; Croxford, A.J.; Meziane, A. Numerical and experimental study of the nonlinear interaction between a shear wave and a frictional interface. J. Acoust. Soc. Am. 2014, 135, 1709–1716. [Google Scholar] [CrossRef]
  11. Carretta, Y.; Hunter, A.; Boman, R.; Ponthot, J.P.; Legrand, N.; Laugier, M.; Dwyer-Joyce, R. Ultrasonic roll bite measurements in cold rolling: Contact length and strip thickness. Proc. Inst. Mech. Eng. Part J J. Eng. Tribol. 2018, 232, 179–192. [Google Scholar] [CrossRef]
  12. Avan, E.Y.; Mills, R.; Dwyer-Joyce, R. Ultrasonic imaging of the piston ring oil film during operation in a motored engine-Towards oil film thickness measurement. SAE Tech. Pap. 2010, 3, 786–793. [Google Scholar] [CrossRef]
  13. Dwyer-Joyce, R.S. An ultrasonic approach for the measurement of oil films in the piston zone. In Tribology and Dynamics of Engine and Powertrain: Fundamentals, Applications and Future Trends; Woodhead Publishing: Sawston, UK, 2010; pp. 426–457. [Google Scholar] [CrossRef]
  14. Beamish, S.; Li, X.; Brunskill, H.; Hunter, A.; Dwyer-Joyce, R. Circumferential film thickness measurement in journal bearings via the ultrasonic technique. Tribol. Int. 2020, 148, 106295. [Google Scholar] [CrossRef]
  15. Dwyer-Joyce, R.S.; Harper, P.; Pritchard, J.; Drinkwater, B.W. Oil film measurement in polytetrafluoroethylene-faced thrust pad bearings for hydrogenerator applications. Proc. Inst. Mech. Eng. Part A J. Power Energy 2006, 220, 619–628. [Google Scholar] [CrossRef]
  16. Geng, T.; Meng, Q.; Zhang, K.; Yuan, X.; Jia, Q. Ultrasonic measurement of lubricant film thickness in sliding bearings with thin liners. Meas. Sci. Technol. 2015, 26, 025002. [Google Scholar] [CrossRef]
  17. Zhang, K.; Meng, Q.; Geng, T.; Wang, N. Ultrasonic measurement of lubricant film thickness in sliding Bearings with overlapped echoes. Tribol. Int. 2015, 88, 89–94. [Google Scholar] [CrossRef]
  18. Dwyer-Joyce, R.; Reddyhoff, T.; Drinkwater, B. Operating limits for acoustic measurement of rolling bearing oil film thickness. Tribol. Trans. 2004, 47, 366–375. [Google Scholar] [CrossRef]
  19. Zhang, K.; Meng, Q.; Chen, W.; Li, J.; Harper, P. Ultrasonic measurement of oil film thickness between the roller and the inner raceway in a roller bearing. Ind. Lubr. Tribol. 2015, 67, 531–537. [Google Scholar] [CrossRef]
  20. Adeyemi, G.J.; Dwyer-Joyce, R.S.; Stephen, J.T.; Adebayo, A. Measurement of metal–roll interface during metal rolling using normal and oblique ultrasonic reflection. Tribol. Mater. Surf. Interfaces 2020, 14, 79–91. [Google Scholar] [CrossRef]
  21. Zhou, H.; Jia, W.; Li, Y.; Ou, M. Method for Estimating Canopy Thickness Using Ultrasonic Sensor Technology. Agriculture 2021, 11, 1011. [Google Scholar] [CrossRef]
  22. Reddyhoff, T.; Dwyer-Joyce, R.S.; Zhang, J.; Drinkwater, B.W. Auto-calibration of ultrasonic lubricant-film thickness measurements. Meas. Sci. Technol. 2008, 19, 045402. [Google Scholar] [CrossRef]
  23. de Mello, J.D.B.; Costa, H.L. Brazilian Tribology: Origin, status quo and future perspectives. Surf. Topogr. Metrol. Prop. 2023, 11, 030201. [Google Scholar] [CrossRef]
  24. Spikes, H.A. In situ methods for tribology research. Tribol. Lett. 2003, 14, 8883. [Google Scholar] [CrossRef]
  25. Guangteng, G.; Cann, P.M.; Olver, A.V.; Spikes, H.A. Lubricant film thickness in rough surface, mixed elastohydrodynamic contact. J. Tribol. 2000, 122, 65–76. [Google Scholar] [CrossRef]
  26. Costa, H.L.; Hutchings, I.M. Hydrodynamic lubrication of textured steel surfaces under reciprocating sliding conditions. Tribol. Int. 2007, 40, 1227–1238. [Google Scholar] [CrossRef]
  27. Sherrington, I. 11-Measurement Techniques for Piston-Ring Tribology; Elsevier Ltd.: Amsterdam, The Netherlands, 2010; pp. 387–425. [Google Scholar]
  28. Wang, J.; He, Y.; Zhang, C.; Gu, L.; Li, P.; Zheng, D.; Wang, L.; Mao, Y. A unified model for large-scale thickness measurement of lubricating film based on ultrasonic lag phase slope. Ultrasonics 2023, 131, 106956. [Google Scholar] [CrossRef]
  29. Wang, M.; Zheng, D.; Xu, Y. A new method for liquid film thickness measurement based on ultrasonic echo resonance technique in gas-liquid flow. Meas. J. Int. Meas. Confed. 2019, 146, 447–457. [Google Scholar] [CrossRef]
  30. Dou, P.; Jia, Y.; Zheng, P.; Wu, T.; Yu, M.; Reddyhoff, T.; Peng, Z. Review of ultrasonic-based technology for oil film thickness measurement in lubrication. Tribol. Int. 2022, 165, 107290. [Google Scholar] [CrossRef]
  31. Dou, P.; Wu, T.; Luo, Z. Wide range measurement of lubricant film thickness based on ultrasonic reflection coefficient phase spectrum. J. Tribol. 2019, 141, 031702. [Google Scholar] [CrossRef]
  32. Dwyer-Joyce, R.S.; Drinkwater, B.W.; Donohoe, C.J. The measurement of lubricant-film thickness using ultrasound. Proc. R. Soc. A Math. Phys. Eng. Sci. 2003, 459, 957–976. [Google Scholar] [CrossRef]
  33. Dwyer-Joyce, R.S.; Harper, P.; Drinkwater, B.W. A method for the measurement of hydrodynamic oil films using ultrasonic reflection. Tribol. Lett. 2004, 17, 337–348. [Google Scholar] [CrossRef]
  34. Hunter, A.; Dwyer-Joyce, R.; Harper, P. Calibration and validation of ultrasonic reflection methods for thin-film measurement in tribology. Meas. Sci. Technol. 2012, 23, 105605. [Google Scholar] [CrossRef]
  35. Ouyang, W.; Zhou, Z.; Jin, Y.; Yan, X.; Liu, Y. Ultrasonic measurement of lubricant film thickness distribution of journal bearing. Rev. Sci. Instrum. 2020, 91, 065111. [Google Scholar] [CrossRef] [PubMed]
  36. Hunter, A. Ultrasonic Measurements of the Strip Thickness, Lubricant Film Thickness, Roll Deflection and Roll Stress in the Roll Bite in the Cold Rolling of Steel. Ph.D. Thesis, University of Sheffield, Sheffield, UK, 2018. [Google Scholar]
  37. Geng, T.; Meng, Q.; Xu, X.; Wang, P.; Yuan, X. An extended ultrasonic time-of-flight method for measuring lubricant film thickness. Proc. Inst. Mech. Eng. Part J J. Eng. Tribol. 2015, 229, 861–869. [Google Scholar] [CrossRef]
Figure 1. Diagram of the system used for the emission, reception, and processing of ultrasonic pulses.
Figure 1. Diagram of the system used for the emission, reception, and processing of ultrasonic pulses.
Lubricants 12 00108 g001
Figure 2. Variation in surface flatness when exposed to excessive polishing. The elevations on the profile indicated in 1 and 2 concern the region where the standard steel gauges were glued to the lower plate.
Figure 2. Variation in surface flatness when exposed to excessive polishing. The elevations on the profile indicated in 1 and 2 concern the region where the standard steel gauges were glued to the lower plate.
Lubricants 12 00108 g002
Figure 3. Topography maps of the (a) top plate and (b) bottom plate.
Figure 3. Topography maps of the (a) top plate and (b) bottom plate.
Lubricants 12 00108 g003
Figure 4. Optical measurements of the gaps used in this work; 3D topographic maps (left) and 2D profiles (right): (a,b) gap 1 = 11–12 µm; (c,d) gap 2 = 85 µm; (e,f) gap 3 = 100 µm; and (g,h) gap 4 = 170 µm.
Figure 4. Optical measurements of the gaps used in this work; 3D topographic maps (left) and 2D profiles (right): (a,b) gap 1 = 11–12 µm; (c,d) gap 2 = 85 µm; (e,f) gap 3 = 100 µm; and (g,h) gap 4 = 170 µm.
Lubricants 12 00108 g004
Figure 5. Reference signals reflected from the steel–air layer and pulses reflected from the oil layer: (a) time domain and (b) frequency domain.
Figure 5. Reference signals reflected from the steel–air layer and pulses reflected from the oil layer: (a) time domain and (b) frequency domain.
Lubricants 12 00108 g005
Figure 6. Time domain spectra for the four measured oils, spring model, gap = 11–12 µm.
Figure 6. Time domain spectra for the four measured oils, spring model, gap = 11–12 µm.
Lubricants 12 00108 g006
Figure 7. FFT of the signals obtained for each of the four oils and the reference signal, spring model, gap = 11–12 µm. In detail, the region of maximum amplitude has been enlarged.
Figure 7. FFT of the signals obtained for each of the four oils and the reference signal, spring model, gap = 11–12 µm. In detail, the region of maximum amplitude has been enlarged.
Lubricants 12 00108 g007
Figure 8. Reflection coefficient spectra for each oil.
Figure 8. Reflection coefficient spectra for each oil.
Lubricants 12 00108 g008
Figure 9. Variation of film thickness within the BW for the four oils used.
Figure 9. Variation of film thickness within the BW for the four oils used.
Lubricants 12 00108 g009
Figure 10. Minimum reflection identification of the resonance frequency for (a) 80 µm; (b) 100 µm; and (c) 170 µm gaps.
Figure 10. Minimum reflection identification of the resonance frequency for (a) 80 µm; (b) 100 µm; and (c) 170 µm gaps.
Lubricants 12 00108 g010
Figure 11. Minimum reference signal amplitude for (a) 100 µm and (b) 170 µm gaps.
Figure 11. Minimum reference signal amplitude for (a) 100 µm and (b) 170 µm gaps.
Lubricants 12 00108 g011
Figure 12. Phase spectrum for (a) 85 µm; (b) 100 µm; and (c) 170 µm gaps.
Figure 12. Phase spectrum for (a) 85 µm; (b) 100 µm; and (c) 170 µm gaps.
Lubricants 12 00108 g012
Table 1. Roughness parameters for the surfaces of steel plates, no form filter, λs Gauss filter: 2.500 µm, λc Gauss Filter: 0.8000 mm.
Table 1. Roughness parameters for the surfaces of steel plates, no form filter, λs Gauss filter: 2.500 µm, λc Gauss Filter: 0.8000 mm.
Top PlateBottom Plate
Amplitude ParametersAmplitude Parameters
Rp1.383 µmRp1.363 µm
Rv1.025 µmRv0.991 µm
Rz2.408 µmRz2.354 µm
Rc1.380 µmRc1.302 µm
Rt8.349 µmRt6.919 µm
Ra0.3260 µmRa0.319µm
Rq0.4246 µmRq0.418 µm
Rsk0.306Rsk0.443
Rku3.987Rku4.172
Load Bearing Ratio ParametersLoad Bearing Ratio Parameters
Rmr0.0116% Rmr0.04253%
Rdc0.615 µmRdc0.6025 µm
Table 2. Acoustic properties of the materials; SD = standard deviation.
Table 2. Acoustic properties of the materials; SD = standard deviation.
Speed of Sound,
c (m/s) *
Acoustic Impedance,
z (MRayl)
Density,
ρ (kg.m3)
PAO 4 Oil1450 1.19820
15W40 Oil14131.25882
SAE-30 Oil14091.24881
ISO VG46 Oil14051.22868
Stainless Steel Plates588345.47700
* Longitudinal waves.
Table 3. Values of the reflection coefficient, R.
Table 3. Values of the reflection coefficient, R.
Reflection Coefficient, RStandard Deviation
PAO 4 Oil0.98140.0022
15W40 Oil0.97980.0004
SAE-30 Oil0.98020.0015
ISO VG46 Oil0.97960.0010
Table 4. Average thickness values (±standard deviation) obtained using the resonance model.
Table 4. Average thickness values (±standard deviation) obtained using the resonance model.
Gap 2 (85 µm)Gap 3 (100 µm)Gap 4 (170 µm)
15W40 Oil65 ± 0.059 µm98 ± 0.121 µm161 ± 0.071 µm
ISO VG46 Oil64 ± 0.057 µm95 ± 0.180 µm160 ± 0.381 µm
PAO4 Oil67 ± 0.200 µm104 ± 0.375 µm173 ± 0.560 µm
SAE-30 Oil63 ± 0.074 µm95 ± 0.162 µm160 ± 0.250 µm
Table 5. Analysis of variance for two measurement models.
Table 5. Analysis of variance for two measurement models.
Reference Gapp-Value
Spring Model11 μm4.92 × 10−2
Resonance Model85 μm4.28 × 10−24
100 μm2.33 × 10−24
170 μm4.27 × 10−24
Table 6. Results of the Tukey tests comparing the oils in each test group.
Table 6. Results of the Tukey tests comparing the oils in each test group.
Gap 1 = 11 µm (Spring Model)
15W40ISO VG46PAO 4SAE 30
15W40 0.370.520.99
ISO VG462.36 0.030.32
PAO 41.964.31 0.59
SAE 300.152.511.80
Gap 2 = 85 µm (Resonance model)
15W40ISO VG46PAO 4SAE 30
15W40 2.05 × 10−101.92 × 10−141.95 × 10−14
ISO VG4618.30 1.92 × 10−146.89 × 10−11
PAO 456.3174.61 1.92 × 10−14
SAE 3037.7319.4494.05
Gap 3 = 100 µm (Resonance model)
15W40ISO VG46PAO 4SAE 30
15W40 3.55 × 10−141.92 × 10−143.53 × 10−13
ISO VG4630.26 1.92 × 10−140.02
PAO 460.5390.79 1.92 × 10−14
SAE 3025.804.4686.34
Gap 4 = 170 µm (Resonance model)
15W40ISO VG46PAO 4SAE 30
15W40 1.17 × 10−111.92 × 10−143.80 × 10−10
ISO VG4621.44 1.92 × 10−140.07
PAO 465.8687.31 1.92 × 10−14
SAE 3017.673.76883.54
Table 7. Relative error values calculated for the investigated film thicknesses.
Table 7. Relative error values calculated for the investigated film thicknesses.
Relative Error (%)
Spring ModelResonance Model
Gap 1 (11 µm)Gap 2 (85 µm)Gap 3 (100 µm)Gap 4 (170 µm)
15W407.623.92.34.3
ISO VG460.224.95.16.2
PAO 414.220.83.41.5
SAE 308.125.94.75.8
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Piovesan, A.S.; Schirru, M.; Tatzgern, F.; Medeiros, J.L.B.; Costa, H.L. Effect of Oil Acoustic Properties on Film Thickness Measurement by Ultrasound Using Spring and Resonance Models. Lubricants 2024, 12, 108. https://doi.org/10.3390/lubricants12040108

AMA Style

Piovesan AS, Schirru M, Tatzgern F, Medeiros JLB, Costa HL. Effect of Oil Acoustic Properties on Film Thickness Measurement by Ultrasound Using Spring and Resonance Models. Lubricants. 2024; 12(4):108. https://doi.org/10.3390/lubricants12040108

Chicago/Turabian Style

Piovesan, Alvaro S., Michele Schirru, Fabio Tatzgern, Jorge L. B. Medeiros, and Henara L. Costa. 2024. "Effect of Oil Acoustic Properties on Film Thickness Measurement by Ultrasound Using Spring and Resonance Models" Lubricants 12, no. 4: 108. https://doi.org/10.3390/lubricants12040108

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop