Pitfalls of Using NIR-Based Clinical Instruments to Test Eyes Implanted with Diffractive Intraocular Lenses

1
Departament d’Òptica i Optometria, Universitat Politècnica de Catalunya, Barcelonatech, ViolinistaVellsolà 37, 08222 Terrassa, Spain
2
Physics Center of Minho and Porto Universities (CF-UM-UP), University of Minho, 4710-057 Braga, Portugal
*
Author to whom correspondence should be addressed.
Diagnostics 2023, 13(7), 1259; https://doi.org/10.3390/diagnostics13071259
Submission received: 30 January 2023 / Revised: 13 March 2023 / Accepted: 20 March 2023 / Published: 27 March 2023
(This article belongs to the Special Issue Refining Diagnosis in Ophthalmology through Laboratory Research)

Abstract

The strong wavelength dependency of diffractive elements casts reasonable doubts on the reliability of near-infrared- (NIR)-based clinical instruments, such as aberrometers and double-pass systems, for assessing, post-surgery, the visual quality of eyes implanted with diffractive multifocal intraocular lenses (DMIOLs). The results obtained for such patients when using NIR light can be misleading. Ordinary compensation for the refractive error bound to chromatic aberration is not enough because it only considers the best focus shift but does not take into account the distribution of light energy among the foci which strongly depends on the wavelength-dependent energy efficiency of the diffractive orders used in the DMIOL design. In this paper, we consider three commercial DMIOL designs with the far focus falling within the range of (−1, 0, +1)-diffractive orders. We prove theoretically the differences existing in the physical performance of the studied lenses when using either the design wavelength in the visible spectrum or a NIR wavelength (780 to 850 nm). Based on numerical simulation and on-bench experimental results, we show that such differences cannot be neglected and may affect all the foci of a DMIOL, including the far focus.
Keywords: intraocular lens; diffractive lens; retinal image quality; ocular aberrations; near infrared; through focus analysis; presbyopia; cataract surgery intraocular lens; diffractive lens; retinal image quality; ocular aberrations; near infrared; through focus analysis; presbyopia; cataract surgery

Share and Cite

MDPI and ACS Style

Vega, F.; Faria-Ribeiro, M.; Armengol, J.; Millán, M.S. Pitfalls of Using NIR-Based Clinical Instruments to Test Eyes Implanted with Diffractive Intraocular Lenses. Diagnostics 2023, 13, 1259. https://doi.org/10.3390/diagnostics13071259

AMA Style

Vega F, Faria-Ribeiro M, Armengol J, Millán MS. Pitfalls of Using NIR-Based Clinical Instruments to Test Eyes Implanted with Diffractive Intraocular Lenses. Diagnostics. 2023; 13(7):1259. https://doi.org/10.3390/diagnostics13071259

Chicago/Turabian Style

Vega, Fidel, Miguel Faria-Ribeiro, Jesús Armengol, and María S. Millán. 2023. "Pitfalls of Using NIR-Based Clinical Instruments to Test Eyes Implanted with Diffractive Intraocular Lenses" Diagnostics 13, no. 7: 1259. https://doi.org/10.3390/diagnostics13071259

APA Style

Vega, F., Faria-Ribeiro, M., Armengol, J., & Millán, M. S. (2023). Pitfalls of Using NIR-Based Clinical Instruments to Test Eyes Implanted with Diffractive Intraocular Lenses. Diagnostics, 13(7), 1259. https://doi.org/10.3390/diagnostics13071259

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop