A Comparative Analysis of Morphology and Dimensions of Functional Blebs following PRESERFLO-Microshunt and XEN-Gel-Stent, a Study Using Anterior Segment OCT

Department of Ophthalmology, Jena University Hospital, 07747 Jena, Germany
*
Author to whom correspondence should be addressed.
Diagnostics 2023, 13(14), 2318; https://doi.org/10.3390/diagnostics13142318
Submission received: 4 June 2023 / Revised: 22 June 2023 / Accepted: 6 July 2023 / Published: 9 July 2023
(This article belongs to the Special Issue Optical Coherence Tomography (OCT) Imaging in Ophthalmology)

Abstract

Evaluation of bleb morphology is a vital part of successful filtration glaucoma surgery. The PRESERFLO-MicroShunt (PRESERFLO) and XEN-Gel-Stent (XEN) are drainage devices implanted using different surgical approaches (ab externo and ab interno, respectively), potentially resulting in distinct bleb morphology. Understanding these morphological differences is essential for postoperative care. In this study, we retrospectively examined functioning blebs following PRESERFLO and XEN implantation using high-resolution anterior segment OCT imaging. Qualitative assessment utilizing the Jenaer Bleb Grading System and quantitative assessment measuring 12 parameters representing bleb dimensions were conducted, and the results were compared between the two groups. A total of 80 eyes from 80 patients were included (41 after PRESERFLO, 39 after XEN). Functioning blebs following PRESERFLO exhibited a higher frequency of hyperreflective tenon changes compared to XEN (31.7% vs. 10.3%, respectively, p = 0.02) and a lower frequency of cavernous changes (17.1% vs. 35.9%, p = 0.05). Additionally, PRESERFLO blebs showed a higher frequency of visible episcleral lakes (92.7% vs. 30.8%, p < 0.001). Furthermore, PRESERFLO blebs demonstrated larger height (2.13 ± 0.5 vs. 1.85 ± 0.6 mm, p = 0.03), width (10.31 ± 2.3 vs. 9.1 ± 2.3 mm, p = 0.02), length (9.13 ± 1.8 vs. 8.24 ± 1.9 mm, p = 0.04), posterior location relative to the limbus (6.21 ± 1.2 vs. 5.21 ± 1.8 mm, p = 0.005), and a thicker bleb wall (1.60 ± 0.5 vs. 1.1 ± 0.4 mm, p = 0.004). Functioning blebs following PRESERFLO and XEN displayed morphological distinctions, likely attributed to variations in surgical techniques (ab externo vs. ab interno) and stent dimensions. These morphological differences should be taken into consideration when evaluating blebs, as they could impact assessments of bleb functionality and influence decisions regarding postoperative interventions.
Keywords: glaucoma filtering surgery; minimally invasive glaucoma surgery; bleb surgery; optical coherence tomography; bleb morphology; PRESERFLO Microshunt; XEN Gel Stent glaucoma filtering surgery; minimally invasive glaucoma surgery; bleb surgery; optical coherence tomography; bleb morphology; PRESERFLO Microshunt; XEN Gel Stent

Share and Cite

MDPI and ACS Style

Hasan, S.M.; Theilig, T.; Papadimitriou, M.; Meller, D. A Comparative Analysis of Morphology and Dimensions of Functional Blebs following PRESERFLO-Microshunt and XEN-Gel-Stent, a Study Using Anterior Segment OCT. Diagnostics 2023, 13, 2318. https://doi.org/10.3390/diagnostics13142318

AMA Style

Hasan SM, Theilig T, Papadimitriou M, Meller D. A Comparative Analysis of Morphology and Dimensions of Functional Blebs following PRESERFLO-Microshunt and XEN-Gel-Stent, a Study Using Anterior Segment OCT. Diagnostics. 2023; 13(14):2318. https://doi.org/10.3390/diagnostics13142318

Chicago/Turabian Style

Hasan, Somar M., Theresa Theilig, Menelaos Papadimitriou, and Daniel Meller. 2023. "A Comparative Analysis of Morphology and Dimensions of Functional Blebs following PRESERFLO-Microshunt and XEN-Gel-Stent, a Study Using Anterior Segment OCT" Diagnostics 13, no. 14: 2318. https://doi.org/10.3390/diagnostics13142318

APA Style

Hasan, S. M., Theilig, T., Papadimitriou, M., & Meller, D. (2023). A Comparative Analysis of Morphology and Dimensions of Functional Blebs following PRESERFLO-Microshunt and XEN-Gel-Stent, a Study Using Anterior Segment OCT. Diagnostics, 13(14), 2318. https://doi.org/10.3390/diagnostics13142318

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop