Beyond Couch Correction in Head and Neck Radiotherapy: A Narrative Review of Action-Oriented Positioning Management

1
Department of Radiological Technology, National Cancer Center Hospital East, 6-5-1 Kashiwanoha, Kashiwa 277-8577, Japan
2
Section of Radiation Safety and Quality Assurance, National Cancer Center Hospital East, 6-5-1 Kashiwanoha, Kashiwa 277-8577, Japan
3
Department of Radiation Oncology, National Cancer Center Hospital East, 6-5-1 Kashiwanoha, Kashiwa 277-8577, Japan
*
Author to whom correspondence should be addressed.
Cancers 2026, 18(14), 2249; https://doi.org/10.3390/cancers18142249
Submission received: 5 June 2026 / Revised: 11 July 2026 / Accepted: 11 July 2026 / Published: 14 July 2026
(This article belongs to the Section Cancer Therapy)

Simple Summary

Head and neck radiotherapy requires highly reproducible patient positioning because treatment fields are close to critical organs such as the spinal cord, brainstem, salivary glands, and swallowing-related structures. Modern image guidance and six-degree couch correction can improve setup accuracy, but they cannot fully correct non-rigid posture differences, regional mismatch, mask-fit deterioration, or anatomical changes during treatment. This review explains why the main clinical challenge is no longer simply detecting mismatch, but deciding what action should follow. We summarize current evidence and future technologies that may help treatment teams determine when couch correction is sufficient and when repositioning, re-immobilization, repeat simulation, or adaptive replanning is needed.

Abstract

Accurate positioning remains a central challenge in head and neck radiotherapy because highly conformal dose distributions are delivered near multiple critical organs, and clinically relevant discrepancies may not be resolved by rigid image-guided correction alone. This narrative review synthesizes evidence on positioning management in head and neck radiotherapy, with particular emphasis on the transition from couch correction to action-oriented clinical decision-making. Peer-reviewed studies, guidelines, and key reviews published between January 2005 and April 2026 were identified using PubMed/MEDLINE and reference screening. Positioning uncertainty is often region-specific and dynamic, reflecting lower-neck and shoulder mismatch, mandibular variation, mask-fit deterioration, functional motion, and progressive anatomical changes during treatment. Thermoplastic immobilization, customized headrests, bite blocks, cone-beam computed tomography, six-degree-of-freedom correction, and adaptive workflows have improved reproducibility and verification; however, these approaches do not eliminate mismatches caused by non-rigid posture, immobilization failure, or evolving anatomy. In such situations, the key clinical task is to distinguish rigid setup deviations suitable for online correction from discrepancies requiring repositioning, re-immobilization, repeat simulation, or adaptive replanning. Emerging technologies, including artificial intelligence, dose-of-the-day estimation, corrected CBCT, synthetic CT, rapid replanning, and workflow-integrated decision support, may help close the gap between mismatch detection and intervention selection. Positioning management should therefore evolve from a correction-centered model to an action-oriented framework that links observed geometric or anatomical change to the most appropriate clinical response.

Graphical Abstract

1. Introduction

Accurate patient positioning is a central requirement in head and neck intensity-modulated radiotherapy (IMRT), in which highly conformal dose distributions are delivered near multiple critical organs, including the spinal cord, brainstem, salivary glands, and swallowing-related structures. Residual geometric uncertainty is usually managed using planning margins, with clinical target volumes commonly expanded by 3–5 mm to generate planning target volumes and serial organs, such as the spinal cord or brainstem, which are often assigned planning organ-at-risk volume margins of 2 mm. These margins indicate that in head and neck IMRT, millimeter-scale deviations are not negligible, particularly in regions with steep dose gradients. However, margin-based compensation cannot fully resolve the complex regional mismatches caused by lower neck and shoulder displacement, mandibular rotation, swallowing-related laryngeal motion, weight loss, or progressive anatomical changes during treatment. Therefore, even in the era of advanced immobilization, image guidance, six-degrees-of-freedom correction, and adaptive radiotherapy, positioning uncertainty remains a clinically relevant limitation that requires a systematic evaluation that goes beyond couch-based correction [1,2].
Over the past two decades, the management of positioning uncertainty in head and neck radiotherapy has evolved. Contemporary practice now incorporates improved immobilization approaches, volumetric image guidance, rotational correction with six-degree-of-freedom couches, adaptive radiotherapy workflows, and advanced delivery techniques, such as stereotactic treatment, MR-guided radiotherapy, and non-coplanar or trajectory-based delivery. Simultaneously, open-face immobilization, surface-guided workflows, and patient-specific oral positioning devices have been used to expand the concept of positioning beyond simple rigid fixation. These developments have undoubtedly improved precision; however, the underlying problem has not been eliminated. Rather, they have clarified that positioning uncertainty in the head and neck region is anatomically heterogeneous, dynamically evolving during treatment, and only partly addressable with rigid image-based correction alone [3,4].
This issue is important because head and neck radiotherapy is affected by lower neck and shoulder mismatch, mask fit deterioration, weight loss, tumor regression, and other treatment-related anatomical changes. Thus, the practical challenge now extends to determining when immobilization, repositioning or re-immobilization, and adaptive replanning are necessary. The field is moving from a correction-centered model toward a broader framework of positioning management [4,5,6,7,8,9,10].
This narrative review therefore reframes head and neck positioning management around an unresolved clinical question: how should treatment teams act when the observed mismatch cannot be adequately resolved by couch correction alone? We integrate evidence on sources of positioning uncertainty, immobilization, daily setup, image guidance, technology-specific positioning requirements, repositioning, re-immobilization, adaptive response, and emerging decision-support technologies. The objective is not only to summarize current practice, but also to organize the evidence around an action-oriented framework that links mismatch detection to clinically meaningful interventions.

2. Search Strategy and Selection Criteria, and Terminology

This narrative review was structured with reference to published quality-assessment principles for narrative review articles [11]. We searched PubMed/MEDLINE for peer-reviewed articles published from January 2005 to April 2026 using combinations of terms related to head and neck radiotherapy, positioning, setup error, immobilization, image-guided radiotherapy, cone-beam computed tomography, six-degree-of-freedom correction, adaptive radiotherapy, surface-guided radiotherapy, and artificial intelligence. We also screened reference lists of relevant guidelines, reviews, and original studies.
We selected articles that were clinically or technically relevant to positioning management in head and neck radiotherapy, particularly studies addressing regional mismatch, immobilization failure, image-guided correction, adaptive response, or emerging decision-support technologies. Articles were excluded when they were unrelated to head and neck radiotherapy, did not address positioning or image-guidance issues, or were outside the scope of the review. As this was a narrative review, no formal meta-analysis or quantitative risk-of-bias assessment was performed. For terminology, positioning uncertainty was used as a broad concept that includes measurable setup errors and anatomical mismatches arising from regional deformation, immobilization variation, functional motion, or progressive anatomical change. The term mismatch was used particularly for regional, non-rigid, or anatomically discordant findings that may not be fully resolved by couch correction.

3. Sources of Positioning Uncertainty

Positioning uncertainty is significant in head and neck IMRT because millimeter-scale geometric deviations may remain relevant even when conventional planning margins are applied. Dosimetric studies have shown the measurable effects of setup uncertainty on target coverage and normal tissue doses [1,2].
Importantly, the positioning uncertainty in the head and neck region cannot be understood solely as a rigid translational problem. Setup deviations vary across anatomical subregions, with postures involving the neck, especially the lower neck, being less reproducible than those involving the cranial regions. Giske et al. demonstrated that local deformations persisted even with dedicated immobilization and were not adequately represented by a single global registration. Using megavoltage cone-beam computed tomography (CBCT), Motegi et al., reported that most deviations were approximately ±8 mm in the neck region and ±5 mm in the head region. Djordjevic et al. further reported that the systematic and random components of subregional deviations were 0.6–2.3 mm and 1.0–1.7 mm, respectively, despite daily image guidance. These findings indicate that setup deviations differ by anatomical sub-region and should not be interpreted as whole-volume rigid displacements [7,12,13].
Clinically relevant uncertainties are associated with the lower neck and shoulder regions. Neubauer et al. reported an average interfractional shoulder motion of 2–6 mm and showed that superior shoulder displacement reduced lower neck target coverage [6]. Tachibana et al. further showed that shoulder deformations greater than 6 mm could affect volumetric modulated arc therapy (VMAT) dose distribution [9]. Region-dependent effects have also been reported in the lower neck and shoulders. Hirotaki et al. found that regional positioning deviations involving the vertebrae, clivus, and mandible were associated with dose-distribution deterioration and that mandibular rotation is correlated with parotid dose indices [10]. Thus, clinically meaningful mismatches can arise across multiple anatomical subregions [14,15].
Progressive anatomical changes that occur during treatment, such as weight loss, tumor shrinkage, nodal regression, and soft-tissue deformation are major sources of uncertainty, as they may alter both the immobilization fit and the geometric relationship between the targets and organs at risk. Adaptive radiotherapy reviews have identified such changes as the major reasons for replanning for head and neck cancer [4]. Contesini et al. suggested that positioning deterioration is multifactorial and may be influenced by treatment time, patient habitus, and immobilization-related factors [16].
The clinical relevance of setup error differs according to the treated subsite and target extent. For localized nasal cavity or sinonasal tumors, strict alignment of cranial bony structures is often required, whereas structures inferior to the mandible may be outside the treatment volume. In contrast, localized laryngeal radiotherapy is more affected by swallowing-related motion, and daily image guidance may require particular attention to the larynx and hyoid bone. For definitive treatment of oropharyngeal, hypopharyngeal, or nasopharyngeal cancers, elective nodal volumes are often extensive and are located close to multiple organs at risk, including the oral cavity, parotid glands, pharyngeal constrictor muscles, spinal cord, and brainstem. In these settings, reproducibility of neck curvature, mandibular position, head rotation, and shoulder position becomes clinically important. Nasopharyngeal cancer is particularly challenging because precise cranial alignment is required near the skull base and brainstem, while elective coverage may extend to the lower neck; therefore, cranially focused six-degree-of-freedom matching may leave clinically relevant residual mismatch in the lower neck.
The representative regional and functional mismatch patterns that may not be fully corrected by image-guided couch correction are summarized in Table 1. These data suggest that positioning uncertainty in head and neck IMRT is dynamic and anatomy-dependent rather than a simple rigid-body displacement. This review cannot define universal action-level thresholds for repositioning, re-immobilization, or adaptive replanning. Nevertheless, the values summarized in Table 1 may provide practical reference data for developing institution-specific action levels.
The table includes lower neck, shoulder, mandibular, and swallowing-related factors; measurement sites; and key quantitative findings. The studies differed in design, imaging methods, and endpoints; thus, the values should be interpreted as representative rather than directly comparable. Abbreviations: AP, anteroposterior; CC, craniocaudal; CTV, clinical target volume; IMPT, intensity-modulated proton therapy; LR, left–right; VMAT, volumetric modulated arc therapy.

4. Immobilization and Daily Treatment Setup

Immobilization is a key component of head and neck radiotherapy because a reproducible treatment geometry must be established before image guidance and couch-based correction are applied. Current guidance for radiation therapists places immobilization, patient positioning, and position verification within a unified workflow, emphasizing that an accurate daily treatment setup depends on the quality of simulation positioning and immobilization [3].
Although thermoplastic masks remain the standard for immobilization, mask fixation alone may not ensure adequate reproducibility across all anatomical regions. Compared with standard headrests, customized head and neck support improves setup performance, and cone-beam CT-based studies have suggested improved consistency in regions such as the chin, cervical spine, and shoulder line. These findings indicate that the headrest design is an important determinant of setup reproducibility [19,20,21]. Representative studies on immobilization approaches and the reported setup accuracies are summarized in Table 2.
Bite blocks and mouthpieces also contribute to reproducible head and neck positioning. In addition to separating target volumes from adjacent oral structures, these devices improve setup reproducibility by stabilizing mandibular, tongue, and oral cavity position [22,23,24].
Immobilization of the lower neck and shoulder region requires special attention, particularly because its reproducibility is not always linked to that of cranial alignment. Even when skull–base matching is acceptable, variations in shoulder position, neck posture, body habitus, and mask fit may produce a caudal–neck mismatch. Therefore, careful shoulder positioning, patient-specific support, and a consistent neck posture are important prerequisites for effective image-guided correction [6,9,16].
Open face masks and surface guided radiation therapy (SGRT)-assisted workflows may improve patient comfort and enable surface-based monitoring during setup and treatment. These approaches can maintain acceptable positional stability in selected patients; however, surface information cannot fully replace internal anatomical verification in detecting regional mismatches in head and neck radiotherapy [25,26,27,28].
Table 2. Representative studies of immobilization and setup accuracy in head and neck radiotherapy.
Table 2. Representative studies of immobilization and setup accuracy in head and neck radiotherapy.
StudynTechniqueCorrectionImmobilizationKey Finding
Anderson et al., 2022 [21]20VMAT6DoF + daily CBCTMask + standard/custom headrestCustom headrest reduced translational setup errors versus standard headrest: 1.2–1.5 vs. 1.6–2.9 mm; rotational errors were 0.51–0.91° vs. 0.75–0.93°.
Rodrigues et al., 2019 [29]30VMAT3/6DoF + daily CBCTMask + standard/custom headrestCustom support + 6DoF reduced interfraction translation from 1.4 ± 0.7 to 0.8 ± 0.3 mm; intrafraction translation was 0.7–1.0 mm.
Wang et al., 2016 [30]21SRT6DoF + ExacTracMask + custom headrest + bite-blockSetup error (3D vector) decreased from 2.7 ± 1.4 to 0.4 ± 0.3 mm after correction; intrafraction motion was 0.9 ± 0.6 mm.
Kang et al., 2020 [31]41SRTCyberKnife + XsightMask + headrest ± vacuum bagIntrafraction motion was 0.92 ± 0.58 mm without vacuum bag and 1.12 ± 0.88 mm with vacuum bag.
Mail et al., 2024 [19]189VMAT6DoF + daily CBCTMask + standard/custom headrest + mouth biteTranslational setup errors: 1.21/1.00/1.07 mm vs. 0.83/1.05/0.85 mm (CC/AP/LR).
Malone et al., 2025 [27]56VMAT3/6DoF + CBCT/SGRTClosed- vs. open-face maskOpen-face masks showed comparable intrafraction stability to closed-face masks; CBCT mean translation/rotation deviations were <0.2 mm/<0.1°.
Essers et al., 2024 [28]19VMAT6DoF + CBCT/SGRTMaskless SGRT vs. maskIntrafraction motion > 2 mm occurred in 19% of maskless fractions and 9% of masked fractions.
Studies were organized according to the number of patients, treatment technique, guidance or correction method, immobilization approach, and key quantitative findings. Mask refers to a thermoplastic immobilization mask, unless otherwise specified. The studies differed in design, imaging method, correction strategy, and endpoint; thus, the values should be interpreted as representative rather than directly comparable. Abbreviations: AP, anteroposterior; CBCT, cone-beam computed tomography; CC, craniocaudal; CTV, clinical target volume; IMRT, intensity-modulated radiotherapy; LR, left–right; SRT, stereotactic radiotherapy; SGRT, surface-guided radiotherapy; VMAT, volumetric modulated arc therapy; 3/6DoF, three- or six-degrees-of-freedom correction.

5. Positioning Requirements Across Treatment Technologies

Technological advances in head and neck radiotherapy have increased the importance of identifying geometric deviations that can be corrected onsite and those that require repositioning, re-immobilization, or adaptive replanning. In conventional fractionated IMRT and VMAT, highly conformal dose distributions and steep dose gradients make the treatment sensitive to both translational and rotational mismatches. Six-degree-of-freedom correction can improve setup reproducibility, particularly when combined with individualized head support; however, its effectiveness depends on reproducible simulation positioning and stable immobilization throughout the treatment [29,32].
This principle is more critical in stereotactic radiotherapy (SRT), where high dose per fraction and limited geometric tolerance leave little room for intrafraction motion. Although customized immobilization and bite-block-based approaches improve setup accuracy and motion control, the primary challenge lies in identifying when residual motion or postural shifts exceed the safely manageable range using image-guided correction alone [30,31,33].
Magnetic resonance-guided radiotherapy and adaptive workflows shift positioning management from image matching to daily anatomical evaluation. Improved soft tissue visualization may help detect tumor regression, weight loss, organ displacement, or mask-fit deterioration; however, adaptive intervention still requires a reproducible posture, compatible immobilization, and efficient workflow [34,35,36,37,38].
Non-coplanar and trajectory-based delivery techniques illustrate how advanced treatment delivery may require a setup-specific robustness evaluation. Although these approaches can improve organ-at-risk sparing by expanding the beam-angle freedom, their clinical value depends on whether such benefits are preserved under setup and delivery uncertainties [39]. Loebner et al. demonstrated that dynamic trajectory radiotherapy maintained lower average normal tissue complication probabilities for xerostomia and dysphagia than VMAT in most scenarios of uncertainty; however, the dose variation differed according to scenario and anatomical structure [40]. Similarly, Hirotaki et al. reported that in a robustness evaluation of biaxially rotational dynamic radiation therapy with robust optimization, parotid and oral cavity sparing was achieved, whereas some target-coverage metrics were more affected by setup uncertainty [41,42,43]. These findings suggest that non-coplanar delivery may require trajectory-specific consideration of setup accuracy, rotational mismatch, image guidance, and delivery reproducibility. Particle- and neutron-based modalities add another layer of sensitivity because setup errors may interact with range uncertainty, beam geometry, or biologically weighted dose distributions [44,45,46,47,48,49]. Particle therapy is particularly sensitive to setup error and anatomical change because it exploits characteristic physical properties, including the finite range and Bragg peak. In particle therapy, the therapeutic dose is concentrated near the end of the beam range; therefore, body contour change, tumor shrinkage, or setup error can have a greater dosimetric impact than in VMAT. Unlike the more robust dose distribution of VMAT, such changes in particle therapy may reduce target coverage. Therefore, strict positioning management is particularly important in head and neck particle therapy, where residual bony mismatch that cannot be fully corrected by six-degree-of-freedom matching may occur. BNCT also requires careful positioning management, because neutron flux decreases with distance from the irradiation aperture and patient setup can affect dose error in sitting-position head and neck treatments [48,49]. Therefore, across advanced delivery platforms, setup uncertainty should be evaluated not only as a geometric displacement but also as a treatment-specific factor that may affect target coverage and organ-at-risk sparing.
Among treatment technologies, some advanced modalities require more stringent positioning than conventional IMRT. In these settings, image guidance and six-degree-of-freedom correction are insufficient; clinicians must determine whether the observed bony or regional anatomical mismatch can be corrected by repositioning or whether it reflects a persistent anatomical change that should trigger adaptive replanning.

6. Image Guidance and Online Correction Strategies

Image guidance in head and neck radiotherapy has shifted from planar bony verification to volumetric imaging and structure-specific assessment. This suggests that external setup reproduction or a single global bony match alone does not ensure accurate treatment geometry [50,51,52]. CBCT has therefore become central to daily or protocol-based verification because it can identify regional mismatch and anatomical change that may not be adequately appreciated on planar images [52,53,54,55,56].
However, the clinical value of image guidance depends on the image interpretation. Whole-volume registration may improve overall alignment; however, residual mismatch may still be observed in clinically relevant subregions, such as the lower neck, shoulders, mandible, or laryngeal region [14,15,57,58]. Similarly, six-degree-of-freedom correction can reduce rigid translational and rotational errors but cannot fully resolve discrepancies caused by posture change, regional deformation, mask-fit deterioration, or progressive anatomical changes [29,32,51,59].
Thus, online image guidance should be used not only to calculate couch corrections but also to identify mismatches that may require repositioning, re-immobilization, or adaptive replanning. Automatic registration may improve workflow efficiency and consistency, but its reliability depends on the image quality, registration target, and the type of discrepancy being assessed [60]. Human oversight remains essential when image guidance is used to evaluate whether standard corrections suffice or if further intervention is required [50,51].

7. When Couch Correction Is Insufficient: Repositioning and Adaptive Response

Image guidance and couch-based correction are essential in modern head and neck radiotherapy; however, they do not solve all clinically relevant positioning problems. In daily practice, the skull base or upper cervical spine may be well aligned after image matching, while residual mismatch remains in the lower neck, shoulders, mandible, or soft tissues. Such findings suggest a region-specific difference between the treatment and simulation geometries rather than a simple rigid setup error [7,13,14,15,57,61]. Figure 1 summarizes a practical interpretation pathway for escalating from couch correction to repositioning, re-immobilization, or replanning when positioning mismatch is not clinically acceptable after daily image guidance.
In such situations, additional couch shifts are not always appropriate. If the mismatch reflects neck flexion or extension, shoulder displacement, mandibular rotation, mask loosening, or poor patient posture, the first intervention should be repeat setup or repositioning. The aim is to restore the planned treatment posture before irradiation, rather than to optimize one anatomical region at the expense of another. This distinction is particularly important in head and neck IMRT, in which different anatomical levels may have different dosimetric priorities [6,7,9,10,13,61].
Re-immobilization should be considered when the original immobilization device no longer reproduces the planned geometry. This may occur because of weight loss, tumor regression, edema reduction, patient discomfort, or progressive loosening of the mask. In such cases, repeated daily corrections may temporarily hide the underlying problem; however, they do not restore the relationship among the patient, mask, and planned dose distribution. A new mask, additional support, or repeat simulation may be required when reproducibility cannot be achieved by repositioning alone [4,16].
Adaptive replanning becomes relevant when the issue is no longer about setup but an anatomical change that alters the delivered dose. Examples include target shrinkage, nodal regression, parotid medial shift, body contour change, or persistent separation between the planning CT and the current treatment anatomy. In these situations, the clinical question is not only whether alignment can be achieved but also whether the original plan remains dosimetrically appropriate. Trigger-based adaptive radiotherapy provides a practical framework for this decision, because it aims to identify patients with large anatomical or positional changes that require repeat imaging, replanning, or both [4,5,62,63,64,65,66].
However, the central issue is not simply which action is selected, but how the need for action is judged. In this context, IGRT should not be regarded merely as an imaging or registration technique. In head and neck radiotherapy, IGRT should also be understood as a clinical judgment process performed at the treatment unit, in which radiation therapists determine whether the daily patient geometry is acceptable for the specific treatment intent and dose distribution. Perfect reproduction of all anatomical regions is desirable but not always achievable, because the target is a living human body with region-dependent motion, deformation, and anatomical change. Therefore, the key technical skill is not simply to minimize the global registration error, but to understand which anatomical region must be prioritized for each patient.
For example, in nasopharyngeal cancer with skull-base or clival invasion, the high-dose CTV may be located immediately adjacent to the brainstem or intracranial structures. In such cases, the treatment plan often relies on a steep dose gradient to maintain target coverage while respecting the brainstem dose constraint. The radiation therapist performing daily IGRT must therefore understand that even a small residual mismatch around the skull base may be clinically more important than a comparable mismatch in a lower-neck elective nodal region. In contrast, a small residual mismatch in the lower neck may be managed differently when cranial target coverage and brainstem sparing are the dominant priorities. Thus, action-oriented positioning management requires patient-specific prioritization of image matching based on target extent, organs at risk, dose-gradient location, and treatment intent, rather than uniform alignment of the entire head and neck volume.
Therefore, beyond couch correction, positioning management should be understood as a stepwise and multidisciplinary clinical process. First, the treatment team determines whether the observed discrepancy can be adequately managed using couch-based translational or rotational correction. If the discrepancy reflects altered neck posture, shoulder position, mandibular rotation, or regional anatomical mismatch, repositioning is performed to restore the planned treatment posture. If reproducibility remains poor, re-immobilization or repeat simulation should be considered. If anatomical changes compromise target coverage or organ-at-risk sparing, adaptive replanning should be considered and, when appropriate, implemented. Importantly, radiation therapists must perform this process with an understanding of the treatment intent, target extent, planning margins, dose-gradient location, and the priorities embedded in the treatment plan. When residual mismatch cannot be fully corrected, positioning should be guided by patient-specific priorities regarding which anatomical structures must be matched most strictly and which regions may tolerate limited residual uncertainty. This requires a multidisciplinary understanding of the radiation oncologist’s clinical intent and the planner’s dose-distribution strategy. Thus, action-oriented positioning management is not simply a technical procedure for image matching, but a patient-specific positioning skill that integrates daily imaging, repositioning, dose-distribution awareness, and clinical judgment.

8. Clinical Impact of Positioning Quality

The clinical importance of positioning quality in head and neck radiotherapy lies in its effect on the delivered dose distribution. Deviations of only a few millimeters can alter the target coverage and organ-at-risk dose, including parotid sparing. Therefore, positioning quality should be regarded not only as a geometric endpoint but also as a determinant of dosimetric treatment quality [1,2,67].
The dosimetric effect of the positioning uncertainty is not uniform. Depending on the anatomical region involved and the structure of the treatment plan, a mismatch may reduce target coverage and/or increase the organ-at-risk dose. Therefore, regional deviations involving the lower neck, shoulders, mandible, or soft tissues may have different clinical consequences, even when the overall setup appears acceptable [10,57,67].
Adaptive radiotherapy provides indirect evidence suggesting that these changes are clinically significant. Published studies and reviews have shown that adaptive replanning can improve target coverage and reduce the dose to at-risk organs, such as the parotid glands and spinal cord, in selected patients. These findings support the concept that deterioration in positioning reproducibility or anatomical fidelity during treatment may justify intervention when it compromises the planned dose distribution [4,64,68,69,70].
The most relevant downstream clinical endpoints are toxicity and tumor control, which should be distinguished from dosimetric endpoints such as target coverage and organ-at-risk dose. Current evidence is strongest for dosimetric effects, more limited for toxicity reduction, and least mature for improved tumor control. Margin reduction combined with daily CBCT-guided VMAT has been associated with reduced radiation-related toxicity, and adaptive radiotherapy reviews have suggested potential benefits for quality of life. However, evidence for improved tumor control remains less mature and should be interpreted cautiously [64,68,71].
Improving positioning quality also incurs practical costs. Frequent volumetric imaging, repeated verification, and adaptive reviews can increase treatment complexity and reduce throughput. Reported online ART workflows have required approximately 20 min per fraction in a head and neck series and 34.5 ± 11.4 min per adaptive session in a broader kV-CBCT implementation study. Repeated CBCT imaging also adds protocol-dependent imaging dose, with reported head and neck kV-CBCT organ doses of 0.03–3.43 mGy per scan [72,73,74]. CBCT-based dose reconstruction and dose-of-the-day estimation may support adaptive decision-making, but these approaches require additional resources and should be balanced against workflow burden and imaging dose [75,76,77].

9. Emerging Technologies and Future Directions

Emerging technologies are shifting positioning management in head and neck radiotherapy from a largely reactive process to a more predictive and decision-support-oriented framework. Recent advances in artificial intelligence (AI), accelerated image reconstruction, synthetic imaging, and automated workflow design suggest that future positioning systems may not only detect mismatches, but also help classify their causes, estimate their dosimetric relevance, and guide the next appropriate action. Some enabling technologies, including online adaptive workflows, synthetic imaging, corrected or high-quality CBCT, and rapid replanning, have already entered clinical use in selected institutions or commercial platforms. In contrast, AI-based mismatch classification and automated action selection remain largely investigational, and their clinical value has not yet been established through robust prospective validation [78,79,80,81].
Figure 2 summarizes the central unmet need and the major technology-enabled directions for moving beyond couch correction in head and neck radiotherapy. In this framework, emerging tools are not viewed as isolated technical advances; rather, they are organized according to the clinical gap they aim to address: reproducing posture, standardizing setup, estimating dose impact, selecting escalation, guiding repositioning, and enabling rapid adaptation.
One of the most promising developments in this field is AI-based mismatch detection. Machine learning can be used to identify setup-related abnormalities and distinguish them from other causes of treatment deviations. In this context, the significance of AI is not merely that it can automate pattern recognition, but that it may eventually support a more consistent identification of clinically meaningful positioning failures. Related work on in vivo dosimetry tolerance failures also suggests that AI can classify the underlying cause of an observed discrepancy, thereby moving closer to a clinically useful framework of error attribution than to simple error flagging. Nevertheless, most available studies are retrospective and have been evaluated in relatively constrained technical settings rather than in full routine clinical workflows [65,66,78,81,82].
Another step beyond detection is AI-assisted correction suggestion. The central future question is whether a system can move from stating that a mismatch exists to indicating how it should be managed. Current evidence suggests that this transition has started but remains incomplete. Current systems are more likely to generate geometric correction estimates, reconstructed images, or classification outputs than to provide human-readable and clinically actionable repositioning advice, such as whether the patient should be re-setup, whether the chin or neck posture is incorrect, or whether the shoulder position is driving the discrepancy. This constitutes an important knowledge gap. For head and neck radiotherapy, the most clinically valuable future systems may not be those that simply automate correction, but those that can distinguish between rigid error, postural change, immobilization failure, and evolving anatomy, and then recommend appropriate intervention [65,66,78,81]. To make such systems clinically reliable, future studies should use multi-institutional datasets that include variations in immobilization devices, imaging protocols, registration strategies, treatment platforms, and institutional action levels. External validation should evaluate not only geometric prediction accuracy, but also whether AI-assisted recommendations improve action selection, using post-repositioning verification images and subsequent clinical decisions as practical feedback.
Another key frontier is the development of 2D-only or low-information-image-based high-precision positioning. This direction is attractive because it can reduce the imaging burden and improve throughput while preserving geometric accuracy. Although fully mature clinical systems remain limited, recent work on corrected CBCT, synthetic CT generation, and AI-assisted adaptive workflows indicate that clinically useful inference from reduced or degraded imaging information is becoming more feasible. In head and neck radiotherapy, this line of development is particularly relevant because high-frequency volumetric imaging may improve confidence; however, improvement will be at the cost of time, dose, and workflow burden. Future systems capable of extracting actionable geometric or dosimetric information from fewer projections, lower-dose scans, or limited-angle imaging may therefore become highly valuable [65,66,83,84,85,86].
Low-dose and fast CBCT is a highly practical area of development. If image guidance evolves into an adaptive decision platform rather than a simple verification step, repeated volumetric imaging will become more efficient and scalable. Recent studies on CBCT correction algorithms, synthetic CT generation from CBCT, and online adaptive radiotherapy workflows suggest that image quality limitations can increasingly be mitigated computationally. This raises the possibility that lower-dose or faster acquisitions may become sufficient not only for setup verification, but also for dose monitoring and adaptive responses. However, this potential depends on the reliability of the correction algorithms under real-world variations, and is essential to validate whether such reconstructed or enhanced images are robust enough for clinical decision-making across different scanners, institutions, and patient anatomies [83,84,85].
Beyond image reconstruction and enhancement, other emerging technologies may improve the earlier phases of treatment preparation, including initial setup support. Although the head- and neck-specific literature remains less mature in this area than in segmentation or adaptive planning, AI-driven and interactive workflows are increasingly entering the radiotherapy pipeline. These developments suggest that future positioning systems may extend upstream, standardizing the initial posture, reducing operator dependence, and identifying problematic setup conditions before treatment imaging is acquired. Such approaches could be especially valuable at a disease site where reproducibility depends not only on image matching but also on subtle, difficult-to-standardize features, such as neck flexion, mandibular position, shoulder posture, tongue position, swallowing-related motion, and patient cooperation. Since swallowing and tongue motion can produce transient intrafraction displacement even after an apparently acceptable setup and verification, future positioning support may need to incorporate not only static alignment but also motion-aware detection, prediction, or monitoring strategies [17,18,75,86,87,88,89].
However, there are substantial barriers to clinical implementation. Reviews of AI in head and neck oncology have consistently identified limited dataset size, single-institution development, class imbalance, domain shift, insufficient external validation, and poor generalizability as major limitations. These concerns are particularly relevant in positioning-related applications because anatomical heterogeneity, immobilization practices, scanner differences, and patient posture variability may influence model performance. Moreover, AI systems trained on relatively controlled datasets may perform unpredictably when confronted with outlier situations that matter the most in clinical positioning, such as unusual shoulder deformation, mask looseness, substantial weight loss, or atypical neck tilt. Recent evidence also indicates that deep learning performance in related head and neck imaging tasks can be degraded by positional variations such as neck tilt, underscoring the importance of data diversity and anatomical realism in model development [79,80].
Ethical and governance issues should also be considered. A positioning support system that recommends an intervention, alters the clinical workflow, or implicitly prioritizes one registration strategy over another is not simply a technical tool; it becomes a part of the clinical decision-making process. Explainability, accountability, bias control, and prospective oversight are essential. Data sharing and federated learning strategies may help overcome the limitations of small institutional datasets; however, they also introduce challenges in standardization, privacy, and governance. Thus, the development of AI-based positioning support will require not only technical progress but also careful attention to validation design, regulatory expectations, and ethical deployment [76,78].
What matters most is not simply building smarter detection tools but ensuring that these systems can translate observed changes into reliable, clinically actionable guidance. In this field, the most valuable emerging technologies are likely to help the treatment team decide when to reposition and adapt, and when the original plan will no longer be sufficient.

10. Conclusions

Positioning in head and neck radiotherapy has evolved substantially with advances in immobilization, image guidance, rotational correction, adaptive workflows, and emerging AI-enabled technologies. Nevertheless, positioning uncertainty remains clinically meaningful because a proportion of relevant mismatches cannot be fully resolved by six-degree-of-freedom couch correction alone. These mismatches may reflect non-rigid posture differences, region-specific deformation, immobilization failure, functional motion, or progressive anatomical change.
Accordingly, the current challenge is no longer limited to detecting setup errors. The more important question is whether the observed mismatch should be managed by online correction, repeat setup, modification of immobilization, repeat simulation, adaptive replanning, or continued treatment with careful monitoring. Although image guidance has become increasingly sophisticated, standardized and actionable support for these decisions remains limited. In this setting, beyond-couch positioning management should be understood as a patient-specific judgment process rather than a purely technical registration task. Radiation therapists must interpret daily images in relation to the planned dose distribution, treatment intent, target extent, organs at risk, planning margins, and dose-gradient location, and determine which anatomical structures require the strictest alignment when perfect anatomical reproduction is not achievable. Thus, action-oriented positioning management is not simply a technical procedure for image matching, but a patient-specific positioning skill that integrates daily imaging, repositioning, dose-distribution awareness, multidisciplinary understanding, and clinical judgment.
Future progress should focus on closing the gap between detection and action. Promising directions include AI-based standardized mismatch detection, AI-assisted repositioning support, dose-impact estimation from IGRT, and workflow-integrated decision systems that distinguish rigid setup errors from posture-related changes, immobilization failures, and evolving anatomy. In parallel, faster CBCT acquisition, automated contouring, rapid replanning, and ultrafast online adaptive radiotherapy may reduce the workflow barriers that currently limit adaptive response. In this context, the most valuable innovation may not simply automate positioning, but help the treatment team choose the next clinically appropriate action.

Author Contributions

K.H.: Conceptualization, literature review, evidence synthesis, and writing—original draft. H.O.: Expert interpretation of head and neck radiotherapy positioning, supervision of dose-distribution and positioning-related content, literature review, and writing—review and editing. K.M.: Expert interpretation of head and neck radiotherapy positioning, critical advice, detailed content review, and writing—review and editing. A.M.: Clinical expertise as a radiation oncologist specializing in head and neck cancer, clinical interpretation, and writing—review and editing. M.I.: Supervision and writing—review and editing. S.Z.: Clinical expertise as a radiation oncologist specializing in head and neck cancer, supervision, and writing—review and editing. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by joint research funding from Hitachi High-Tech Corporation [contract number: A2025-129; funding period: 1 April 2025 to 31 March 2028]. Funding was provided for research-related support, including the preparation of the manuscript. The funder had no role in the conception of the review, literature interpretation, writing of the manuscript, or decision to submit the article for publication.

Institutional Review Board Statement

Not applicable. This manuscript is a narrative review and did not involve new experiments on human participants or animals.

Informed Consent Statement

Not applicable. This manuscript is a narrative review and does not report identifiable patient information, individual patient data, or patient-specific case details.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

The authors are grateful to the late Takaki Ariji, formerly of the Department of Radiation Oncology at National Cancer Center Hospital East, for his extensive guidance and mentorship in head and neck radiotherapy. His expert instructions on dose distribution generation, anatomical interpretation, patient positioning, and image-guided radiotherapy provided an important foundation for this review. During the preparation of this manuscript, the authors used ChatGPT, version GPT-5.5 (OpenAI), to assist with English translation, grammar checking, and English-language editing. The authors reviewed and edited all AI-assisted outputs and take full responsibility for the content of the final manuscript.

Conflicts of Interest

The authors declare no conflicts of interest. This work was supported by joint research funding from Hitachi High-Tech Corporation, as disclosed in the Funding statement. The funder had no role in the conception of the review, literature interpretation, writing of the manuscript, or decision to submit the article for publication.

References

  1. Prabhakar, R.; Laviraj, M.A.; Haresh, K.P.; Julka, P.K.; Rath, G.K. Impact of patient setup error in the treatment of head and neck cancer with intensity modulated radiation therapy. Phys. Med. 2010, 26, 26–33. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  2. Kaur, I.; Rawat, S.; Ahlawat, P.; Kakria, A.; Gupta, G.; Saxena, U.; Mishra, M. Dosimetric impact of setup errors in head and neck cancer patients treated by image-guided radiotherapy. J. Med. Phys. 2016, 41, 144–148. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  3. Leech, M.; Coffey, M.; Mast, M.; Moura, F.; Osztavics, A.; Pasini, D.; Vaandering, A. ESTRO ACROP guidelines for positioning, immobilisation and position verification of head and neck patients for radiation therapists. Tech. Innov. Patient Support Radiat. Oncol. 2017, 1, 1–7. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Castadot, P.; Lee, J.A.; Geets, X.; Grégoire, V. Adaptive radiotherapy of head and neck cancer. Semin. Radiat. Oncol. 2010, 20, 84–93. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  5. Figen, M.; Çolpan Öksüz, D.; Duman, E.; Prestwich, R.; Dyker, K.; Cardale, K.; Ramasamy, S.; Murray, P.; Şen, M. Radiotherapy for head and neck cancer: Evaluation of triggered adaptive replanning in routine practice. Front. Oncol. 2020, 10, 579917. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  6. Neubauer, E.; Dong, L.; Followill, D.S.; Garden, A.S.; Court, L.E.; White, R.A.; Kry, S.F. Assessment of shoulder position variation and its impact on IMRT and VMAT doses for head and neck cancer. Radiat. Oncol. 2012, 7, 19. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  7. Djordjevic, M.; Sjöholm, E.; Tullgren, O.; Sorcini, B. Assessment of residual setup errors for anatomical sub-structures in image-guided head-and-neck cancer radiotherapy. Acta Oncol. 2014, 53, 646–653. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  8. Moore, S.J.; Herst, P.M.; Louwe, R.J.W. Review of the patient positioning reproducibility in head-and-neck radiotherapy using statistical process control. Radiother. Oncol. 2018, 127, 183–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  9. Tachibana, H.; Motegi, K.; Moriya, S. Impact of shoulder deformation on volumetric modulated arc therapy doses for head and neck cancer. Phys. Med. 2018, 53, 118–128. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Hirotaki, K.; Tomizawa, K.; Moriya, S.; Ito, M.; Sakae, T. Impact of anatomical position errors on dose distribution in head and neck radiotherapy and robust image registration against anatomical changes. Anticancer Res. 2023, 43, 1827–1834. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  11. Baethge, C.; Goldbeck-Wood, S.; Mertens, S. SANRA-a scale for the quality assessment of narrative review articles. Res. Integr. Peer Rev. 2019, 4, 5. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  12. Motegi, K.; Kohno, R.; Ueda, T.; Shibuya, T.; Ariji, T.; Kawashima, M.; Akimoto, T. Evaluating positional accuracy using megavoltage cone-beam computed tomography for IMRT with head-and-neck cancer. J. Radiat. Res. 2014, 55, 568–574. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  13. Giske, K.; Stoiber, E.M.; Schwarz, M.; Stoll, A.; Muenter, M.W.; Timke, C.; Roeder, F.; Debus, J.; Huber, P.E.; Thieke, C.; et al. Local setup errors in image-guided radiotherapy for head and neck cancer patients immobilized with a custom-made device. Int. J. Radiat. Oncol.* Biol.* Phys. 2011, 80, 582–589. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. van Kranen, S.; van Beek, S.; Rasch, C.; van Herk, M.; Sonke, J.J. Setup uncertainties of anatomical sub-regions in head-and-neck cancer patients after offline CBCT guidance. Int. J. Radiat. Oncol.* Biol.* Phys. 2009, 73, 1566–1573. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  15. van Beek, S.; van Kranen, S.; Mencarelli, A.; Remeijer, P.; Rasch, C.; van Herk, M.; Sonke, J.J. First clinical experience with a multiple region of interest registration and correction method in radiotherapy of head-and-neck cancer patients. Radiother. Oncol. 2010, 94, 213–217. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  16. Contesini, M.; Guberti, M.; Saccani, R.; Braglia, L.; Iotti, C.; Botti, A.; Abbati, E.; Iemmi, M. Setup errors in patients with head-neck cancer (HNC), treated using the intensity modulated radiation therapy (IMRT) technique: How it influences the customised immobilisation systems, patient’s pain and anxiety. Radiat. Oncol. 2017, 12, 72. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  17. Hamlet, S.; Ezzell, G.; Aref, A. Larynx motion associated with swallowing during radiation therapy. Int. J. Radiat. Oncol.* Biol.* Phys. 1994, 28, 467–470. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. van Asselen, B.; Raaijmakers, C.P.J.; Lagendijk, J.J.W.; Terhaard, C.H.J. Intrafraction motions of the larynx during radiotherapy. Int. J. Radiat. Oncol.* Biol.* Phys. 2003, 56, 384–390. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Mail, N.; Alshamrani, K.M.; Lodhi, R.N.; Khawandanh, E.; Saleem, A.; Khan, B.; Alghamdi, M.; Nadershah, M.; Althaqafy, M.S.; Subahi, A.; et al. Evaluation of positioning accuracy in head-and-neck cancer treatment: A cone beam computed tomography assessment of three immobilization devices with volumetric modulated arc therapy. J. Biol. Methods 2024, 11, e99010025. [Google Scholar] [CrossRef] [PubMed]
  20. Howlin, C.; O’Shea, E.; Dunne, M.; Mullaney, L.; McGarry, M.; Clayton-Lea, A.; Finn, M.; Carter, P.; Garret, B.; Thirion, P. A randomized controlled trial comparing customized versus standard headrests for head and neck radiotherapy immobilization in terms of set-up errors, patient comfort and staff satisfaction (ICORG 08-09). Radiography 2015, 21, 74–83. [Google Scholar] [CrossRef] [Scilit]
  21. Anderson, E.; Pilling, K.; Iqbal, S.; Pearson, R.L. A retrospective analysis comparing set up errors from standard versus customised headrests for head and neck radiotherapy. Radiography 2022, 28, 746–750. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  22. Feng, Z.; Wang, P.; Gong, L.; Xu, L.; Zhang, J.; Zheng, J.; Zhang, D.; Tian, T.; Wang, P. Construction and clinical evaluation of a new customized bite block used in radiotherapy of head and neck cancer. Cancer/Radiothérapie 2019, 23, 125–131. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  23. Suzuka, M.; Tomita, M.; Nakata, A.; Takeuchi, A.; Minowa, K. Evaluation of fixation accuracy according to differences in mouthpiece configuration used in radiotherapy for head and neck cancer. Hokkaido J. Dent. Sci. 2024, 45, 22–30. (In Japanese) [Google Scholar]
  24. Kitamori, H.; Sumida, I.; Tsujimoto, T.; Shimamoto, H.; Murakami, S.; Ohki, M. Evaluation of mouthpiece fixation devices for head and neck radiotherapy patients fabricated in PolyJet photopolymer by a 3D printer. Phys. Med. 2019, 58, 90–98. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  25. Sharkey, M.; Bridge, P. A systematic review of open face versus mask less surface guided radiotherapy for patients undergoing radiotherapy for head, neck and brain tumours. J. Radiother. Pract. 2025, 24, e15. [Google Scholar] [CrossRef] [Scilit]
  26. Lastrucci, A.; Morelli, I.; Votta, C.; Maran, I.; Iosca, N.; Monaco, I.P.; Salvestrini, V.; Desideri, I.; Marrazzo, L.; Wandael, Y.; et al. Open-face masks in radiotherapy: Enhancing therapeutic strategies for head and neck and brain cancer patients-a comprehensive scoping review. Cancers 2024, 16, 2899. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Malone, C.; Ryan, S.; Nicholson, J.; O’Maolalai, R.; O’Donovan, R.; McArdle, O.; Duane, F.; Armstrong, J.; Keane, L.; Glynn, A.; et al. Intrafraction motion stability of open vs. closed facemasks in head and neck radiotherapy: Insights from the OPEN phase III trial. Radiother. Oncol. 2025, 209, 110941. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  28. Essers, M.; Mesch, L.; Beugeling, M.; Dekker, J.; de Kruijf, W. Setup and intra-fractional motion measurements using surface scanning in head and neck cancer radiotherapy: A feasibility study. Phys. Imaging Radiat. Oncol. 2024, 29, 100563. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Rodrigues, M.F.; Veen, S.; van Egmond, J.; van Hameren, M.; van Oorschot, T.; de Vet, S.; van Santvoort, J.P.C.; Wiggenraad, R.G.J.; Mast, M.E. The influence of a six degrees of freedom couch and an individual head support in patient positioning in radiotherapy of head and neck cancer. Phys. Imaging Radiat. Oncol. 2019, 11, 30–33. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  30. Wang, H.; Wang, C.; Tung, S.; Dimmitt, A.W.; Wong, P.F.; Edson, M.A.; Garden, A.S.; Rosenthal, D.I.; Fuller, C.D.; Gunn, G.B.; et al. Improved setup and positioning accuracy using a three-point customized cushion/mask/bite-block immobilization system for stereotactic reirradiation of head and neck cancer. J. Appl. Clin. Med. Phys. 2016, 17, 180–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  31. Kang, C.L.; Lee, T.F.; Chan, S.H.; Liu, S.C.; Wang, J.C.; Tsai, C.H.; Liao, K.C.; Fang, F.M.; Chang, L.; Huang, C.C. Comparison of intrafractional motion in head and neck cancer between two immobilization methods during stereotactic ablative radiation therapy by CyberKnife. Cancer Manag. Res. 2020, 12, 13599–13606. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  32. Kung, J.S.; Tran, W.T.; Poon, I.; Atenafu, E.G.; Courneyea, L.; Higgins, K.; Enepekides, D.; Sahgal, A.; Chin, L.; Karam, I. Evaluation of the efficacy of rotational corrections for standard-fractionation head and neck image-guided radiotherapy. Technol. Cancer Res. Treat. 2019, 18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  33. Mesko, S.; Wang, H.; Tung, S.; Wang, C.; Pasalic, D.; Chapman, B.V.; Moreno, A.C.; Reddy, J.P.; Garden, A.S.; Rosenthal, D.I.; et al. Estimating PTV margins in head and neck stereotactic ablative radiation therapy (SABR) through target site analysis of positioning and intrafractional accuracy. Int. J. Radiat. Oncol.* Biol.* Phys. 2020, 106, 185–193. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  34. Boeke, S.; Mönnich, D.; van Timmeren, J.E.; Balermpas, P. MR-Guided radiotherapy for head and neck cancer: Current developments, perspectives, and challenges. Front. Oncol. 2021, 11, 616156. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  35. McDonald, B.A.; Dal Bello, R.; Fuller, C.D.; Balermpas, P. The use of MR-guided radiation therapy for head and neck cancer and recommended reporting guidance. Semin. Radiat. Oncol. 2024, 34, 69–83. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  36. McDonald, B.A.; Vedam, S.; Yang, J.; Wang, J.; Castillo, P.; Lee, B.; Sobremonte, A.; Ahmed, S.; Ding, Y.; Mohamed, A.S.R.; et al. Initial feasibility and clinical implementation of daily MR-guided adaptive head and neck cancer radiation therapy on a 1.5T MR-linac system: Prospective R-IDEAL 2a/2b systematic clinical evaluation of technical innovation. Int. J. Radiat. Oncol.* Biol.* Phys. 2021, 109, 1606–1618. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  37. Zhou, Y.; Yuan, J.; Wong, O.L.; Fung, W.W.K.; Cheng, K.F.; Cheung, K.Y.; Yu, S.K. Assessment of positional reproducibility in the head and neck on a 1.5-T MR simulator for an offline MR-guided radiotherapy solution. Quant. Imaging Med. Surg. 2018, 8, 925–935. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  38. Zhou, Y.; Wong, O.L.; Cheung, K.Y.; Yu, S.K.; Yuan, J. A pilot study of highly accelerated 3D MRI in the head and neck position verification for MR-guided radiotherapy. Quant. Imaging Med. Surg. 2019, 9, 1255–1269. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  39. Smyth, G.; Evans, P.M.; Bamber, J.C.; Bedford, J.L. Recent developments in non-coplanar radiotherapy. Br. J. Radiol. 2019, 92, 20180908. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  40. Loebner, H.A.; Bertholet, J.; Mackeprang, P.H.; Volken, W.; Elicin, O.; Mueller, S.; Guyer, G.; Aebersold, D.M.; Stampanoni, M.F.M.; Fix, M.K.; et al. Robustness analysis of dynamic trajectory radiotherapy and volumetric modulated arc therapy plans for head and neck cancer. Phys. Imaging Radiat. Oncol. 2024, 30, 100586. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  41. Hirotaki, K.; Tomizawa, K.; Kitou, S.; Jinno, S.; Moriya, S.; Fujisawa, T.; Zenda, S.; Sakae, T.; Ito, M. Dosimetric comparison of noncoplanar VMAT without rotating the patient couch versus conventional coplanar/noncoplanar VMAT for head and neck cancer: First report of Dynamic Swing Arc. Adv. Radiat. Oncol. 2024, 10, 101706. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  42. Hirotaki, K.; Motegi, K.; Moriya, S.; Wakabayashi, M.; Sakae, T.; Ito, M.; Zenda, S. Dosimetric and robustness evaluation of biaxially rotational dynamic radiation therapy with robust optimization. J. Appl. Clin. Med. Phys. 2025, 26, e70194. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  43. Hiraoka, S.; Hirashima, H.; Nakamura, M.; Tanaka, F.; Adachi, H.; Ono, Y.; Ono, T.; Nakashima, R.; Nakajima, A.; Mizowaki, T. Integration test of biaxially rotational dynamic-radiation therapy for nasopharyngeal carcinoma: Efficacy evaluation and dosimetric analysis. Radiother. Oncol. 2025, 207, 110879. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  44. Malyapa, R.; Lowe, M.; Bolsi, A.; Lomax, A.J.; Weber, D.C.; Albertini, F. Evaluation of robustness to setup and range uncertainties for head and neck patients treated with pencil beam scanning proton therapy. Int. J. Radiat. Oncol.* Biol.* Phys. 2016, 95, 154–162. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  45. Chang, L.; Shaaban, S.G.; Gogineni, E.; Page, B.; Quon, H.; Li, H.; Ger, R. Daily head and neck treatment assessment for optimal proton therapy planning robustness. Cancers 2023, 15, 3719. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  46. Jensen, A.D.; Winter, M.; Kuhn, S.P.; Debus, J.; Nairz, O.; Münter, M.W. Robotic-based carbon ion therapy and patient positioning in 6 degrees of freedom: Setup accuracy of two standard immobilization devices used in carbon ion therapy and IMRT. Radiat. Oncol. 2012, 7, 51. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  47. Varnava, M.; Kawashima, M.; Matsumura, A.; Oohashi, Y.; Miyazawa, M.; Koya, J.; Yamaguchi, N.; Ogano, T.; Tashiro, M.; Ohno, T. Assessment of upright immobilization methods for abdominal and head-and-neck cancer treatments in a carbon ion radiotherapy setting. Tech. Innov. Patient Support Radiat. Oncol. 2025, 36, 100356. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  48. Komori, S.; Hirose, K.; Sato, M.; Yamazaki, Y.; Takeuchi, A.; Kato, R.; Motoyanagi, T.; Narita, Y.; Kato, T.; Takai, Y. Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer. Phys. Med. 2024, 126, 104818. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  49. Kumada, H.; Takada, K.; Terunuma, T.; Aihara, T.; Matsumura, A.; Sakurai, H.; Sakae, T. Monitoring patient movement with boron neutron capture therapy and motion capture technology. Appl. Radiat. Isot. 2020, 163, 109208. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  50. Kearney, M.; Coffey, M.; Leong, A. A review of Image Guided Radiation Therapy in head and neck cancer from 2009–2019—Best Practice Recommendations for RTTs in the Clinic. Tech. Innov. Patient Support Radiat. Oncol. 2020, 14, 43–50. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  51. Radiotherapy Board. On Target 2: Updated Guidance for Image-Guided Radiotherapy; The Royal College of Radiologists: London, UK, 2021. [Google Scholar]
  52. Li, H.; Zhu, X.R.; Zhang, L.; Dong, L.; Tung, S.; Ahamad, A.; Chao, K.S.C.; Morrison, W.H.; Rosenthal, D.I.; Schwartz, D.L.; et al. Comparison of 2D radiographic images and 3D cone beam computed tomography for positioning head-and-neck radiotherapy patients. Int. J. Radiat. Oncol.* Biol.* Phys. 2008, 71, 916–925. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  53. Den, R.B.; Doemer, A.; Kubicek, G.; Bednarz, G.; Galvin, J.M.; Keane, W.M.; Xiao, Y. Daily image guidance with cone-beam computed tomography for head-and-neck cancer intensity-modulated radiotherapy: A prospective study. Int. J. Radiat. Oncol.* Biol.* Phys. 2010, 76, 1353–1359. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  54. Delishaj, D.; Ursino, S.; Pasqualetti, F.; Matteucci, F.; Cristaudo, A.; Soatti, C.P.; Barcellini, A.; Paiar, F. Set-up errors in head and neck cancer treated with IMRT technique assessed by cone-beam computed tomography: A feasible protocol. Radiat. Oncol. J. 2018, 36, 54–62. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  55. Nyarambi, I.; Chamunyonga, C.; Pearce, A. CBCT image guidance in head and neck irradiation: The impact of daily and weekly imaging protocols. J. Radiother. Pract. 2015, 14, 362–369. [Google Scholar] [CrossRef] [Scilit]
  56. Bhatia, J.V.; Kothavade, V.D.; Ramachandran, V.K. Impact of different image guidance schedules in head and neck irradiation: A retrospective analysis. Precis. Radiat. Oncol. 2021, 5, 34–42. [Google Scholar] [CrossRef] [Scilit]
  57. Graff, P.; Kirby, N.; Weinberg, V.; Chen, J.; Yom, S.S.; Lambert, L.; Pouliot, J. The residual setup errors of different IGRT alignment procedures for head and neck IMRT and the resulting dosimetric impact. Int. J. Radiat. Oncol.* Biol.* Phys. 2013, 86, 170–176. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  58. Lin, W.; Xiao, L.; Chen, J.; Lin, X.; Li, Z.; Fang, J.; Fang, Y. Comparison of six-degree-of-freedom registration errors across different anatomical registration regions in head and neck radiotherapy under identical CBCT conditions. BMC Cancer 2026, 26, 477. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  59. Pisani, L.; Lockman, D.; Jaffray, D.; Yan, D.; Martinez, A.; Wong, J. Setup error in radiotherapy: On-line correction using electronic kilovoltage and megavoltage radiographs. Int. J. Radiat. Oncol.* Biol.* Phys. 2000, 47, 825–839. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  60. Kanakavelu, N.; Samuel, E.J.J. Accuracy in automatic image registration between MV cone beam computed tomography and planning kV computed tomography in image guided radiotherapy. Rep. Pract. Oncol. Radiother. 2016, 21, 487–494. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  61. Xu, J.; Wang, T.; Luo, Y.; Shang, L.; Mai, X.; Ruan, J.; Pan, X.; Chi, F. Set-up errors of the neck are underestimated using the overall registration frame of head and neck in IMRT for NPC. J. X-Ray Sci. Technol. 2023, 31, 1067–1077. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  62. Bak, B.; Skrobala, A.; Adamska, A.; Kazmierska, J.; Jozefacka, N.; Piotrowski, T.; Malicki, J. Criteria for verification and replanning based on the adaptive radiotherapy protocol “Best for Adaptive Radiotherapy” in head and neck cancer. Life 2022, 12, 722. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  63. Gupta, T.; Maheshwari, G.; Joshi, K.; Sawant, P.; Mishra, A.; Khairnar, S.; Patel, P.; Sinha, S.; Swain, M.; Budrukkar, A.; et al. Image-guidance triggered adaptive radiation therapy in head and neck squamous cell carcinoma: Single-institution experience and implications for clinical practice. J. Med. Imaging Radiat. Sci. 2023, 54, 88–96. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  64. Nuyts, S.; Bollen, H.; Eisbruch, A.; Strojan, P.; Mendenhall, W.M.; Ng, S.P.; Ferlito, A. Adaptive radiotherapy for head and neck cancer: Pitfalls and possibilities from the radiation oncologist’s point of view. Cancer Med. 2024, 13, e7192. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  65. Hirotaki, K.; Moriya, S.; Tachibana, H.; Sakae, T. Detection of anatomical changes using two-dimensional x-ray images for head and neck adaptive radiotherapy. Med. Phys. 2022, 49, 3288–3297. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  66. Hirotaki, K.; Moriya, S.; Tomizawa, K.; Wakabayashi, M.; Motegi, A.; Ito, M.; Sakae, T. Prediction and monitoring of adaptive radiation therapy timing using two-dimensional x-ray image-based water equivalent thickness. Anticancer Res. 2025, 45, 387–397. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  67. Manning, M.A.; Wu, Q.; Cardinale, R.M.; Mohan, R.; Lauve, A.D.; Kavanagh, B.D.; Morris, M.M.; Schmidt-Ullrich, R.K. The effect of setup uncertainty on normal tissue sparing with IMRT for head-and-neck cancer. Int. J. Radiat. Oncol.* Biol.* Phys. 2001, 51, 1400–1409. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  68. Castelli, J.; Simon, A.; Lafond, C.; Perichon, N.; Rigaud, B.; Chajon, E.; De Bari, B.; Ozsahin, M.; Bourhis, J.; de Crevoisier, R. Adaptive radiotherapy for head and neck cancer. Acta Oncol. 2018, 57, 1284–1292. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  69. Maheshwari, G.; Dhanawat, A.; Kumar, H.S.; Sharma, N.; Jakhar, S.L. Clinical and dosimetric impact of adaptive intensity-modulated radiotherapy in locally advanced head-and-neck cancer. J. Cancer Res. Ther. 2020, 16, 600–604. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  70. Ghosh, A.; Gupta, S.; Johny, D.; Bhosale, V.V.; Negi, M.P.S. A study to assess the dosimetric impact of the anatomical changes occurring in the parotid glands and tumour volume during intensity modulated radiotherapy using simultaneous integrated boost (IMRT-SIB) in head and neck squamous cell cancers. Cancer Med. 2021, 10, 5175–5190. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  71. Navran, A.; Heemsbergen, W.; Janssen, T.; Hamming-Vrieze, O.; Jonker, M.; Zuur, C.; Verheij, M.; Remeijer, P.; Sonke, J.J.; van den Brekel, M.; et al. The impact of margin reduction on outcome and toxicity in head and neck cancer patients treated with image-guided volumetric modulated arc therapy (VMAT). Radiother. Oncol. 2019, 130, 25–31. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  72. Blumenfeld, P.; Arbit, E.; Den, R.; Salhab, A.; Michaeli, T.F.; Wygoda, M.; Hillman, Y.; Pfeffer, R.M.; Fang, M.; Misrati, Y.; et al. Real world clinical experience using daily intelligence-assisted online adaptive radiotherapy for head and neck cancer. Radiat. Oncol. 2024, 19, 43. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  73. Stanley, D.N.; Harms, J.; Pogue, J.A.; Belliveau, J.G.; Marcrom, S.R.; McDonald, A.M.; Dobelbower, M.C.; Boggs, D.H.; Soike, M.H.; Fiveash, J.A.; et al. A roadmap for implementation of kV-CBCT online adaptive radiation therapy and initial first year experiences. J. Appl. Clin. Med. Phys. 2023, 24, e13961. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  74. Özseven, A.; Dirican, B. Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy. Rep. Pract. Oncol. Radiother. 2021, 26, 251–258. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  75. Shinde, P.; Jadhav, A.; Shankar, V.; Gupta, K.K.; Dhoble, N.S.; Dhoble, S.J. Evaluation of kV-CBCT based 3D dose calculation accuracy and its validation using delivery fluence derived dose metrics in head and neck cancer. Phys. Med. 2022, 96, 32–45. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  76. Washio, H.; Ohira, S.; Funama, Y.; Ueda, Y.; Isono, M.; Inui, S.; Miyazaki, M.; Teshima, T. Accuracy of dose calculation on iterative CBCT for head and neck radiotherapy. Phys. Med. 2021, 86, 106–112. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  77. Trivedi, G.; Singh, P.P.; Oinam, A.S.; Singh, R. Cone-beam computed tomography (CBCT) dose optimization technique and image quality assessment scoring. J. Cancer Res. Ther. 2024, 20, 71–78. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  78. Muhammed, A.; Hassan, H.; Zarb, F.; McEntee, M.F. The potential use of deep learning in performing autocorrection of setup errors in patients receiving radiotherapy. Radiography 2025, 31, 102881. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  79. Rao, K.N.; Fernandez-Alvarez, V.; Guntinas-Lichius, O.; Sreeram, M.P.; de Bree, R.; Kowalski, L.P.; Forastiere, A.; Pace-Asciak, P.; Rodrigo, J.P.; Saba, N.F.; et al. The limitations of artificial intelligence in head and neck oncology. Adv. Ther. 2025, 42, 3042–3055. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  80. Brooks, J.; Harmsen, W.; Routman, D.; Tryggestad, E.; Moseley, D. The impact of neck tilt on the accuracy of deep learning generated contours for CT images of the head and neck. J. Appl. Clin. Med. Phys. 2025, 26, e70316. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  81. Butterworth, V.; Young, T.; Drake, H.; Palmer, I.; Avgoulea, T.; Ivy, E.; Andriolo, J.; Creppy, C.; Routledge, C.; Adjogatse, D.; et al. Data-centric artificial intelligence and cancer research: Construction of a real-world head and neck treatment data repository. ESMO Real World Data Digit. Oncol. 2025, 9, 100162. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  82. Saiyo, N.; Kojima, H.; Noto, K.; Isomura, N.; Tsukamoto, K.; Yamaguchi, S.; Segawa, Y.; Kohigashi, J.; Takemura, A. Detection of the failed-tolerance causes of electronic-portal-imaging-device-based in vivo dosimetry using machine learning for volumetric-modulated arc therapy: A feasibility study. Phys. Imaging Radiat. Oncol. 2025, 34, 100785. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  83. Gambetta, V.; Stützer, K.; Richter, C. Current status and upcoming developments for online adaptive proton therapy enabling a closed feedback loop for near real-time adaptation. Front. Oncol. 2025, 15, 1660605. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  84. Kaushik, S.; Vatterodt, N.; Ödén, J.; Fredriksson, A.; Korreman, S.S.; Toma-Dasu, I. Synthetic computed tomography techniques for adaptive proton therapy in head and neck cancers. Phys. Imaging Radiat. Oncol. 2025, 36, 100847. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  85. Sawyer, C.; Koo, J.; Naghavi, A.; Caudell, J.J.; Qayyum, M.; Redler, G.; Feygelman, V.; Mueller, J.; Johansson, W.; Latifi, K. Evaluating a CBCT correction algorithm for adaptive radiotherapy. Technol. Cancer Res. Treat. 2026, 25. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  86. Wei, Z.; Ren, J.; Eriksen, J.G.; Jensen, K.; Mortensen, H.R.; Korreman, S.S.; Nijkamp, J. An interactive deep-learning workflow for head and neck gross tumour volume segmentation. Phys. Imaging Radiat. Oncol. 2025, 35, 100820. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  87. Heilemann, G.; Zimmermann, L.; Nyholm, T.; Simko, A.; Widder, J.; Goldner, G.; Georg, D.; Kuess, P. Ultra-fast, one-click radiotherapy treatment planning outside a treatment planning system. Phys. Imaging Radiat. Oncol. 2025, 33, 100724. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  88. Zhao, B.; Park, Y.K.; Gu, X.; Reynolds, R.; Timmerman, R.; Sher, D.J. Surface guided motion management in glottic larynx stereotactic body radiation therapy. Radiother. Oncol. 2020, 153, 236–242. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  89. Jung, S.; Kim, B.; Lee, S.Y.; Chang, W.I.; Son, J.; Park, J.M.; Choi, C.H.; Lee, J.H.; Wu, H.G.; Kim, J.; et al. Novel tongue-positioning device to reduce tongue motion during radiation therapy for head and neck cancer: Geometric and dosimetric evaluations. PLoS ONE 2023, 18, e0291712. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Figure 1. Practical interpretation pathway for positioning mismatch in head and neck radiotherapy. After daily 2DkV or CBCT assessment, discrepancies that are clinically acceptable after couch correction may proceed to couch correction and treatment. If residual mismatch remains unacceptable, regional or posture-related mismatch should prompt repositioning and repeat imaging. Persistent mismatch, anatomical change, or mask-fit failure should prompt further evaluation. Depending on dosimetric concern and clinical feasibility, subsequent actions may include adaptive replanning, re-simulation CT and replanning, or re-immobilization or replanning. This figure was created by the authors and was not reproduced or adapted from previously published material.
Figure 1. Practical interpretation pathway for positioning mismatch in head and neck radiotherapy. After daily 2DkV or CBCT assessment, discrepancies that are clinically acceptable after couch correction may proceed to couch correction and treatment. If residual mismatch remains unacceptable, regional or posture-related mismatch should prompt repositioning and repeat imaging. Persistent mismatch, anatomical change, or mask-fit failure should prompt further evaluation. Depending on dosimetric concern and clinical feasibility, subsequent actions may include adaptive replanning, re-simulation CT and replanning, or re-immobilization or replanning. This figure was created by the authors and was not reproduced or adapted from previously published material.
Cancers 18 02249 g001
Figure 2. Conceptual framework for action-oriented positioning management beyond couch correction. Six-degree couch correction alone cannot resolve head and neck-specific positioning mismatch. Future technologies should support detection of mismatch, interpretation of its cause, estimation of dose impact, selection of appropriate action, and repositioning or adaptation. This figure was created by the authors and was not reproduced or adapted from previously published material.
Figure 2. Conceptual framework for action-oriented positioning management beyond couch correction. Six-degree couch correction alone cannot resolve head and neck-specific positioning mismatch. Future technologies should support detection of mismatch, interpretation of its cause, estimation of dose impact, selection of appropriate action, and repositioning or adaptation. This figure was created by the authors and was not reproduced or adapted from previously published material.
Cancers 18 02249 g002
Table 1. Representative regional and functional sources of positioning uncertainty in head and neck radiotherapy.
Table 1. Representative regional and functional sources of positioning uncertainty in head and neck radiotherapy.
StudynSiteMeasurementKey Quantitative Finding
Motegi et al., 2014 [12]67Lower neckC51 SD was 1.9 mm (LR), 1.7 mm (CC), and 2.3 mm (AP).
van Kranen et al., 2009 [14]38Lower neckC5–C7Systematic/random errors were 1.9/1.7 mm (LR), 1.1/1.9 mm (CC), and 2.2/1.6 mm (AP).
Hirotaki et al., 2023 [10]20Lower neck/shoulder/mandibleT1, shoulders, lower jawboneT1 vertical deviation increased from approximately 3.0 to 4.0–4.2 mm; mandibular roll correlated with parotid D50 in both VMAT and IMPT (r > 0.7).
Neubauer et al., 2012 [6]10ShoulderHumeral headsDisplacement: 2–6 mm; CTV D99 reduction: up to 1.01 Gy.
Hamlet et al., 1994 [17]NRLarynx/
Swallowing
LarynxLaryngeal motion during swallowing: ~20 mm superiorly and <10 mm anteriorly; estimated dose impact: ~0.5%.
van Asselen et al., 2003 [18]10Larynx/
Swallowing
Epiglottis tipSwallowing occupied 0.45% of irradiation time; epiglottis motion was within 7.1 mm for 95% of irradiation time.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Hirotaki, K.; Oyoshi, H.; Motegi, K.; Motegi, A.; Ito, M.; Zenda, S. Beyond Couch Correction in Head and Neck Radiotherapy: A Narrative Review of Action-Oriented Positioning Management. Cancers 2026, 18, 2249. https://doi.org/10.3390/cancers18142249

AMA Style

Hirotaki K, Oyoshi H, Motegi K, Motegi A, Ito M, Zenda S. Beyond Couch Correction in Head and Neck Radiotherapy: A Narrative Review of Action-Oriented Positioning Management. Cancers. 2026; 18(14):2249. https://doi.org/10.3390/cancers18142249

Chicago/Turabian Style

Hirotaki, Kouta, Hajime Oyoshi, Kana Motegi, Atsushi Motegi, Masashi Ito, and Sadamoto Zenda. 2026. "Beyond Couch Correction in Head and Neck Radiotherapy: A Narrative Review of Action-Oriented Positioning Management" Cancers 18, no. 14: 2249. https://doi.org/10.3390/cancers18142249

APA Style

Hirotaki, K., Oyoshi, H., Motegi, K., Motegi, A., Ito, M., & Zenda, S. (2026). Beyond Couch Correction in Head and Neck Radiotherapy: A Narrative Review of Action-Oriented Positioning Management. Cancers, 18(14), 2249. https://doi.org/10.3390/cancers18142249

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop