Development and Comparative Evaluation of Two Enzyme-Based Amperometric Biosensor Designs for Alanine Aminotransferase Determination in Biological Fluids

by
Daryna Mruga
1,*,
Yevhen Vakhovskyi
1,2,
Veronika Bakhmat
1,
Viktoriya Pyeshkova
1,3,*,
Sergii Dzyadevych
1,4 and
Oleksandr Soldatkin
1,5
1
Institute of Molecular Biology and Genetics, National Academy of Sciences of Ukraine, 150 Zabolotnoho Str., 03680 Kyiv, Ukraine
2
ESC “Institute of Biology and Medicine”, Taras Shevchenko National University of Kyiv, 64 Volodymyrska Str., 01003 Kyiv, Ukraine
3
School of Engineering, University of Warwick, Coventry CV4 7AL, UK
4
Educational Scientific Institute of High Technologies, Taras Shevchenko National University of Kyiv, 64 Volodymyrska Str., 01003 Kyiv, Ukraine
5
Faculty of Biomedical Engineering, National Technical University of Ukraine “Igor Sikorsky Kyiv Polytechnic Institute”, 37 Beresteysky Avenue, 03056 Kyiv, Ukraine
*
Authors to whom correspondence should be addressed.
Micromachines 2025, 16(10), 1168; https://doi.org/10.3390/mi16101168
Submission received: 30 August 2025 / Revised: 25 September 2025 / Accepted: 10 October 2025 / Published: 15 October 2025

Abstract

Alanine aminotransferase (ALT) is a key biomarker of liver function. Compared with conventional assays for ALT detection—which are expensive, time-consuming, labor-intensive, and require experienced personnel—biosensors represent a promising alternative, but it remains unclear which biorecognitive enzymatic configuration offers the best analytical performance for ALT detection. This study presents the development and comparative evaluation of two amperometric biosensors based on oxidase biorecognition elements: pyruvate oxidase (POx) and glutamate oxidase (GlOx). Enzymes were immobilized onto platinum electrodes under optimized conditions using entrapment for POx (pH 7.4, enzyme loading 1.62 U/µL, PVA-SbQ concentration 13.2%) and covalent crosslinking for GlOx (pH 6.5, enzyme loading 2.67%, glutaraldehyde concentration 0.3%). Analytical parameters were systematically assessed, including linear range (1–500 U/L for POx vs. 5–500 U/L for GlOx), limit of detection (1 U/L for both), and sensitivity (0.75 vs. 0.49 nA/min at 100 U/L). The POx-based biosensor demonstrated higher sensitivity and lower detection limits, whereas the GlOx-based biosensor exhibited greater stability in complex solutions and reduced assay costs due to a simpler working solution. Moreover, while the POx-based system is uniquely suited for ALT determination, the GlOx-based sensor can be affected by AST activity in samples but may also be adapted for targeted AST detection. Overall, the study highlights a trade-off between sensitivity, robustness, and versatility in ALT biosensor design, providing guidance for the rational development of clinically relevant devices.
Keywords: alanine aminotransferase; pyruvate oxidase; glutamate oxidase; amperometry; biosensor; enzyme; ALT activity alanine aminotransferase; pyruvate oxidase; glutamate oxidase; amperometry; biosensor; enzyme; ALT activity

Share and Cite

MDPI and ACS Style

Mruga, D.; Vakhovskyi, Y.; Bakhmat, V.; Pyeshkova, V.; Dzyadevych, S.; Soldatkin, O. Development and Comparative Evaluation of Two Enzyme-Based Amperometric Biosensor Designs for Alanine Aminotransferase Determination in Biological Fluids. Micromachines 2025, 16, 1168. https://doi.org/10.3390/mi16101168

AMA Style

Mruga D, Vakhovskyi Y, Bakhmat V, Pyeshkova V, Dzyadevych S, Soldatkin O. Development and Comparative Evaluation of Two Enzyme-Based Amperometric Biosensor Designs for Alanine Aminotransferase Determination in Biological Fluids. Micromachines. 2025; 16(10):1168. https://doi.org/10.3390/mi16101168

Chicago/Turabian Style

Mruga, Daryna, Yevhen Vakhovskyi, Veronika Bakhmat, Viktoriya Pyeshkova, Sergii Dzyadevych, and Oleksandr Soldatkin. 2025. "Development and Comparative Evaluation of Two Enzyme-Based Amperometric Biosensor Designs for Alanine Aminotransferase Determination in Biological Fluids" Micromachines 16, no. 10: 1168. https://doi.org/10.3390/mi16101168

APA Style

Mruga, D., Vakhovskyi, Y., Bakhmat, V., Pyeshkova, V., Dzyadevych, S., & Soldatkin, O. (2025). Development and Comparative Evaluation of Two Enzyme-Based Amperometric Biosensor Designs for Alanine Aminotransferase Determination in Biological Fluids. Micromachines, 16(10), 1168. https://doi.org/10.3390/mi16101168

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop