Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power
Highlights
- The distinct evolutionary trends of sensitivity, bandwidth, and dynamic range were systematically quantified as a function of tunable laser power in a miniaturized SERF magnetometer.
- An optimized comprehensive performance was achieved within a compact 35 × 22 × 15 mm3 probe, featuring 16 fT/√Hz sensitivity, 230 Hz bandwidth, and a ±5.4 nT dynamic range.
- Active modulation of laser power enables a flexible transition between ultra-high sensitivity and high-bandwidth operating modes without the need for hardware reconfiguration.
- The established power–performance framework enhances the environmental adaptability and operational robustness of portable biomagnetic sensing systems in diverse application scenarios.
Abstract
1. Introduction
2. Principles and Experimental Configuration
2.1. Operating Principles
2.2. Experimental Configuration
3. Experimental Results and Analysis
3.1. Characterization of VCSEL Performance
3.2. Characterization of Sensor Sensitivity
3.3. Sensor Bandwidth and Dynamic Range
4. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
| SERF | Spin-exchange relaxation-free |
| VCSEL | vertical-cavity surface-emitting laser |
| OPMs | Optically Pumped Magnetometers |
| SQUIDs | Superconducting Quantum Interference Devices |
| MCG | magnetocardiography |
| MEMS | Micro-Electro-Mechanical Systems |
| NIST | National Institute of Standards and Technology |
| MEG | magnetoencephalography |
| DFB | distributed feedback |
| VCSELs | vertical-cavity surface-emitting lasers |
| PD | photodetector |
| FPC | flexible printed circuit |
| TA | transimpedance amplifier |
| LIA | lock-in amplifier |
| DAQ | data acquisition |
References
- Ripka, P.; Janošek, M. Advances in magnetic field sensors. IEEE Sens. J. 2010, 10, 1108–1116. [Google Scholar]
- Ghemes, C.; Dragos-Pinzaru, O.G.; Tibu, M.; Lostun, M.; Lupu, N.; Chiriac, H. Tunnel magnetoresistance-based sensor for biomedical application: Proof-of-concept. Coatings 2023, 13, 227. [Google Scholar] [CrossRef] [Scilit]
- Khan, M.A.; Sun, J.; Li, B.; Przybysz, A.; Kosel, J. Magnetic sensors-A review and recent technologies. Eng. Res. 2021, 3, 022005. [Google Scholar] [CrossRef] [Scilit]
- Dang, H.B.; Maloof, A.C.; Romalis, M.V. Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer. Appl. Phys. Lett. 2010, 97, 151110. [Google Scholar] [CrossRef] [Scilit]
- Kominis, I.K.; Kornack, T.W.; Allred, J.C.; Romalis, M.V. A subfemtotesla multichannel atomic magnetometer. Nature 2003, 422, 596–599. [Google Scholar] [CrossRef] [Scilit]
- Budker, D.; Romalis, M. Optical magnetometry. Nat. Phys. 2007, 3, 227–234. [Google Scholar] [CrossRef] [Scilit]
- Boto, E.; Holmes, N.; Leggett, J.; Roberts, G.; Shah, V.; Meyer, S.S.; Muñoz, L.D.; Mullinger, K.J.; Tierney, T.M.; Bestmann, S.; et al. Moving magnetoencephalography towards real-world applications with a wearable system. Nature 2018, 555, 657–661. [Google Scholar] [CrossRef] [Scilit]
- Holmes, N.; Rea, M.; Hill, R.M.; Boto, E.; Leggett, J.; Edwards, L.J.; Rhodes, N.; Shah, V.; Osborne, J.; Fromhold, T.M.; et al. Naturalistic hyperscanning with wearable magnetoencephalography. Sensors 2023, 23, 5454. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sun, Y.; Xixi, W.; Zhou, Y.; Qin, J.; Bai, D.; Wang, Y.; Zhou, Z. A single-axis atomic magnetometer-based mouse magnetocardiography measurement system. IEEE Trans. Instrum. Meas. 2023, 72, 9505510. [Google Scholar] [CrossRef] [Scilit]
- Shah, V.K.; Wakai, R.T. A compact, high performance atomic magnetometer for biomedical applications. Phys. Med. Biol. 2013, 58, 8153–8161. [Google Scholar] [CrossRef] [Scilit]
- Schwindt, P.D.D.; Knappe, S.; Shah, V.; Hollberg, L.; Kitching, J.; Liew, L.A.; Moreland, J. Chip-scale atomic magnetometer. Appl. Phys. Lett. 2004, 85, 6409–6411. [Google Scholar] [CrossRef] [Scilit]
- Johnson, C.N.; Schwindt, P.D.D.; Weisend, M. Multi-sensor magnetoencephalography with atomic magnetometers. Phys. Med. Biol. 2013, 58, 6065–6077. [Google Scholar] [CrossRef] [Scilit]
- Zhou, P.; Quan, W.; Wei, K.; Liang, Z.; Hu, J.; Liu, L.; Hu, G.; Wang, A.; Ye, M. Application of VCSEL in bio-sensing atomic magnetometers. Biosensors 2022, 12, 1098. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Liu, X.; Bi, W.; Li, Y.; Hu, Y.; Chang, M.; Zhang, X. A review of fiber-coupled atomic magnetometer. Microw. Opt. Technol. Lett. 2023, 65, 1516–1524. [Google Scholar] [CrossRef] [Scilit]
- Song, M.; Liu, Y.; Lu, F.; Cao, Q.; Zhai, Y. A CNN-GS Hybrid Algorithm for Generating Pump Light Fields in Atomic Magnetometers. Photonics 2025, 12, 796. [Google Scholar] [CrossRef] [Scilit]
- Wu, T.; Xiao, W.; Peng, X.; Wu, T.; Guo, H. Compact high-bandwidth single-beam optically-pumped magnetometer for biomagnetic measurement. Biomed. Opt. Express 2024, 16, 235–244. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Li, X.; Guo, Z.; Yang, R.; Feng, Y. Single-beam double-pass miniaturized atomic magnetometer for biomagnetic imaging systems. IEEE Sens. J. 2023, 23, 12433–12440. [Google Scholar] [CrossRef] [Scilit]
- Savukov, I.M.; Romalis, M.V. Effects of spin-exchange collisions in a high-density alkali-metal vapor in low magnetic fields. Phys. Rev. A. 2005, 71, 023405. [Google Scholar] [CrossRef] [Scilit]
- Allred, J.C.; Lyman, R.N.; Kornack, T.W.; Romalis, M.V. High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation. Phys. Rev. Lett. 2002, 89, 130801. [Google Scholar] [CrossRef] [Scilit]
- Happer, W.; Tang, H. Spin-exchange shift and narrowing of magnetic resonance lines in optically pumped alkali vapors. Phys. Rev. Lett. 1973, 31, 273–276. [Google Scholar] [CrossRef] [Scilit]
- Savukov, I.M.; Seltzer, S.J.; Romalis, M.V. Detection of NMR signals with a radio-frequency atomic magnetometer. J. Magn. Reson. 2007, 185, 214–220. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Auzinsh, M.; Budker, D.; Kimball, D.F.; Rochester, S.M.; Stalnaker, J.E.; Sushkov, A.O.; Yashchuk, V.V. Can a Quantum Nondemolition Measurement Improve the Sensitivity of an Atomic Magnetometer? Phys. Rev. Lett. 2004, 93, 173002. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tierney, T.M.; Holmes, N.; Mellor, S.; López, J.D.; Roberts, G.; Hill, R.M.; Boto, E.; Leggett, J.; Shah, V.; Brookes, M.J.; et al. Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalography. NeuroImage 2019, 199, 598–608. [Google Scholar] [CrossRef] [Scilit] [PubMed]






Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Shi, P.; Zuo, C.; Li, Q.; Zhang, S. Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power. Sensors 2026, 26, 2000. https://doi.org/10.3390/s26062000
Shi P, Zuo C, Li Q, Zhang S. Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power. Sensors. 2026; 26(6):2000. https://doi.org/10.3390/s26062000
Chicago/Turabian StyleShi, Peng, Chen Zuo, Qisong Li, and Shulin Zhang. 2026. "Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power" Sensors 26, no. 6: 2000. https://doi.org/10.3390/s26062000
APA StyleShi, P., Zuo, C., Li, Q., & Zhang, S. (2026). Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power. Sensors, 26(6), 2000. https://doi.org/10.3390/s26062000