Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study
Abstract
1. Introduction
2. Materials and Methods
2.1. Study Design
2.2. Study Population and Setting
2.3. Description of the Earsensor
2.4. Description of the Bedside Routine Standard
2.5. Signal Analysis
2.6. Outcomes and Statistical Analysis
3. Results
3.1. Oxygen Saturation
3.2. Pulse Rate
3.3. Respiratory Rate
| Number of Measurement Pairs | Bias | Lower LoA | Upper LoA | ARMS | SD | |
|---|---|---|---|---|---|---|
| SpO2 | 4646 | 1.7 | −1.8 | 5.1 | 2.4 | 1.8 |
| Pulse rate | 9354 | 1.2 | −6.1 | 8.4 | 3.9 | 3.7 |
| Respiratory rate | 2131 | 0.3 | −8.4 | 8.9 | 4.4 | 4.4 |



3.4. Technical Performance
3.5. Patient Experiences
4. Discussion
Limitations
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
Appendix A
How comfortable was it to wear the sensor?![]() Very uncomfortable Very comfortable |
Did the sensor irritate your ear?![]() Always irritating Not irritating |
Did the sensor interfere with your sleep?![]() Always Not at all |
Were you aware of wearing the sensor?![]() Not aware Always aware |
Did you mind continuously wearing the sensor?![]() Always Not at all |
Did the sensor cause pain?![]() Painful Not painful |
Did the sensor have the correct size?![]() Not at all Perfect fit |
References
- Schein, R.M.; Hazday, N.; Pena, M.; Ruben, B.H.; Sprung, C.L. Clinical antecedents to in-hospital cardiopulmonary arrest. Chest 1990, 98, 1388–1392. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Jones, D.; Mitchell, I.; Hillman, K.; Story, D. Defining clinical deterioration. Resuscitation 2013, 84, 1029–1034. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Goldhill, D.R.; McNarry, A.F.; Mandersloot, G.; McGinley, A. A physiologically-based early warning score for ward patients: The association between score and outcome. Anaesthesia 2005, 60, 547–553. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ludikhuize, J.; Smorenburg, S.M.; de Rooij, S.E.; de Jonge, E. Identification of deteriorating patients on general wards; measurement of vital parameters and potential effectiveness of the Modified Early Warning Score. J. Crit. Care 2012, 27, 424.e7–424. e13. [Google Scholar] [CrossRef] [Scilit]
- Van Galen, L.S.; Struik, P.W.; Driesen, B.E.J.M.; Merten, H.; Ludikhuize, J.; Van Der Spoel, J.I.; Kramer, M.H.H.; Nanayakkara, P.W.B. Delayed Recognition of Deterioration of Patients in General Wards Is Mostly Caused by Human Related Monitoring Failures: A Root Cause Analysis of Unplanned ICU Admissions. PLoS ONE 2016, 11, e0161393. [Google Scholar] [CrossRef] [Scilit]
- Eddahchouri, Y.; Koeneman, M.; Plokker, M.; Brouwer, E.; van de Belt, T.H.; van Goor, H.; Bredie, S.J. Low compliance to a vital sign safety protocol on general hospital wards: A retrospective cohort study. Int. J. Nurs. Stud. 2021, 115, 103849. [Google Scholar] [CrossRef] [Scilit]
- Fuhrmann, L.; Lippert, A.; Perner, A.; Østergaard, D. Incidence, staff awareness and mortality of patients at risk on general wards. Resuscitation 2008, 77, 325–330. [Google Scholar] [CrossRef] [Scilit]
- Beckett, D.; Gordon, C.; Paterson, R.; Chalkley, S.; Macleod, D.; Bell, D. Assessment of clinical risk in the out of hours hospital prior to the introduction of Hospital at Night. Acute Med. 2009, 8, 33–38. [Google Scholar] [CrossRef] [Scilit]
- Calzavacca, P.; Licari, E.; Tee, A.; Egi, M.; Downey, A.; Quach, J.; Haase-Fielitz, A.; Haase, M.; Bellomo, R. The impact of Rapid Response System on delayed emergency team activation patient characteristics and outcomes—A follow-up study. Resuscitation 2010, 81, 31–35. [Google Scholar] [CrossRef] [Scilit]
- Kause, J.; Smith, G.; Prytherch, D.; Parr, M.; Flabouris, A.; Hillman, K. A comparison of Antecedents to Cardiac Arrests, Deaths and EMergency Intensive care Admissions in Australia and New Zealand, and the United Kingdom—The ACADEMIA study. Resuscitation 2004, 62, 275–282. [Google Scholar] [CrossRef] [Scilit]
- Watkinson, P.J.; Pimentel, M.A.; Clifton, L.; Clifton, D.A.; Vollam, S.; Young, D.; Tarassenko, L. Early detection of physiological deterioration in post-surgical patients using wearable technology combined with an integrated monitoring system: A pre-and post-interventional study. medRxiv 2020. [Google Scholar] [CrossRef] [Scilit]
- Downey, C.; Randell, R.; Brown, J.; Jayne, D.G. Continuous Versus Intermittent Vital Signs Monitoring Using a Wearable, Wireless Patch in Patients Admitted to Surgical Wards: Pilot Cluster Randomized Controlled Trial. J. Med. Internet Res. 2018, 20, e10802. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Barker, S.J.; Shah, N.K. The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication). Anesthesiology 1997, 86, 101–108. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Allen, J. Photoplethysmography and its application in clinical physiological measurement. Physiol. Meas. 2007, 28, R1–R39. [Google Scholar] [CrossRef] [Scilit]
- Almarshad, M.A.; Islam, M.S.; Al-Ahmadi, S.; BaHammam, A.S. Diagnostic Features and Potential Applications of PPG Signal in Healthcare: A Systematic Review. Healthcare 2022, 10, 547. [Google Scholar] [CrossRef] [Scilit]
- Saab, R.; Wu, B.P.; Rivas, E.; Chiu, A.; Lozovoskiy, S.; Ma, C.; Yang, D.; Turan, A.; Sessler, D.I. Failure to detect ward hypoxaemia and hypotension: Contributions of insufficient assessment frequency and patient arousal during nursing assessments. Br. J. Anaesth. 2021, 127, 760–768. [Google Scholar] [CrossRef] [Scilit]
- Charlton, P.H.; Bonnici, T.; Tarassenko, L.; Clifton, D.A.; Beale, R.; Watkinson, P.J. An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram. Physiol. Meas. 2016, 37, 610–626. [Google Scholar] [CrossRef] [Scilit]
- van Melzen, R.; Haveman, M.E.; Schuurmann, R.C.; van Amsterdam, K.; El Moumni, M.; Tabak, M.; Struys, M.M.R.F.; de Vries, J.-P.P.M. Validity and Reliability of Wearable Sensors for Continuous Postoperative Vital Signs Monitoring in Patients Recovering from Trauma Surgery. Sensors 2024, 24, 6379. [Google Scholar] [CrossRef] [Scilit]
- Monnink, S.H.J.; Vliet Mvan Kuiper, M.J.; Constandse, J.C.; Hoftijzer, D.; Muller, M.; Ronner, E. Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate. J. Clin. Monit. Comput. 2024, 39, 451–457. [Google Scholar] [CrossRef] [Scilit]
- Harskamp, R.E.; Bekker, L.; Himmelreich, J.C.L.; Clercq, L.D.; Karregat, E.P.M.; Sleeswijk, M.E.; Lucassen, W.A.M. Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: A cross-sectional validation study in intensive care patients. BMJ Open Respir. Res. 2021, 8, e000939. [Google Scholar] [CrossRef] [Scilit]
- Poorzargar, K.; Pham, C.; Ariaratnam, J.; Lee, K.; Parotto, M.; Englesakis, M.; Chung, F.; Nagappa, M. Accuracy of pulse oximeters in measuring oxygen saturation in patients with poor peripheral perfusion: A systematic review. J. Clin. Monit. Comput. 2022, 36, 961–973. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Saugel, B.; Hoppe, P.; Khanna, A.K. Automated Continuous Noninvasive Ward Monitoring: Validation of Measurement Systems Is the Real Challenge. Anesthesiology 2020, 132, 407–410. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- ISO 80601-2-61:2017; Medical Electrical Equipment—Part 2–61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment. International Organization for Standardization: Geneva, Switzerland, 2017.
- Haahr-Raunkjaer, C.; Skovbye, M.; Rasmussen, S.M.; Elvekjaer, M.; Sørensen, H.B.D.; Meyhoff, C.S.; Aasvang, E.K. Agreement between standard and continuous wireless vital sign measurements after major abdominal surgery: A clinical comparison study. Physiol. Meas. 2022, 43, 115007. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Breteler, M.J.M.; Leigard, E.; Hartung, L.C.; Welch, J.R.; Brealey, D.A.; Fritsch, S.J.; Konrad, D.; Hertzberg, D.; Bell, M.; Rienstra, H.; et al. Reliability of an all-in-one wearable sensor for continuous vital signs monitoring in high-risk patients: The NIGHTINGALE clinical validation study. J. Clin. Monit. Comput. 2025, 39, 1087–1100. [Google Scholar] [CrossRef] [Scilit]
- Bland, J.M.; Altman, D.G. Agreement between methods of measurement with multiple observations per individual. J. Biopharm. Stat. 2007, 17, 571–582. [Google Scholar] [CrossRef] [Scilit]
- Myles, P.S.; Cui, J. Using the Bland-Altman method to measure agreement with repeated measures. Br. J. Anaesth. 2007, 99, 309–311. [Google Scholar] [CrossRef] [Scilit]
- Parker, R.A.; Weir, C.J.; Rubio, N.; Rabinovich, R.; Pinnock, H.; Hanley, J.; McCloughan, L.; Drost, E.M.; Mantoani, L.C.; MacNee, W.; et al. Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients. PLoS ONE 2016, 11, e0168321. [Google Scholar] [CrossRef] [Scilit]
- Reijmers, N.; van Kootwijk, A.; de Waal, E.E.C. Accuracy of vital sign monitoring using a photoplethysmography upper arm wearable device in postoperative non-cardiac surgery patients: A prospective observational clinical validation study. J. Clin. Monit. Comput. 2025, 1–9. [Google Scholar] [CrossRef] [Scilit]
- Breteler, M.J.; KleinJan, E.J.; Dohmen, D.A.; Leenen, L.P.; van Hillegersberg, R.; Ruurda, J.P.; van Loon, K.; Blokhuis, T.J.; Kalkman, C.J. Vital Signs Monitoring with Wearable Sensors in High-risk Surgical Patients: A Clinical Validation Study. Anesthesiology 2020, 132, 424–439. [Google Scholar] [CrossRef] [Scilit]
- van der Stam, J.A.; Mestrom, E.H.; Scheerhoorn, J.; Jacobs, F.E.; Nienhuijs, S.; Boer, A.K.; Riel, N.A.W.V.; de Morree, H.M.; Bonomi, A.G.; Scharnhorst, V.; et al. The Accuracy of Wrist-Worn Photoplethysmogram-Measured Heart and Respiratory Rates in Abdominal Surgery Patients: Observational Prospective Clinical Validation Study. JMIR Perioper. Med. 2023, 6, e40474. [Google Scholar] [CrossRef] [Scilit]
- Leenen, J.P.L.; Leerentveld, C.; van Dijk, J.D.; van Westreenen, H.L.; Schoonhoven, L.; Patijn, G.A. Current Evidence for Continuous Vital Signs Monitoring by Wearable Wireless Devices in Hospitalized Adults: Systematic Review. J. Med. Internet Res. 2020, 22, e18636. [Google Scholar] [CrossRef] [Scilit]
- Leenen, J.P.L.; Rasing, H.J.M.; Kalkman, C.J.; Schoonhoven, L.; Patijn, G.A. Process Evaluation of a Wireless Wearable Continuous Vital Signs Monitoring Intervention in 2 General Hospital Wards: Mixed Methods Study. JMIR Nurs. 2023, 6, e44061. [Google Scholar] [CrossRef] [Scilit]






| Characteristic | Value | |
|---|---|---|
| Gender, n (%) | Female | 9 (43) |
| Male | 12 (57) | |
| Age, years (median [IQR]) | 53 [40–66] | |
| Type of surgery, n (%) | Neurosurgery | 18 (86%) |
| Vascular surgery | 2 (9%) | |
| Urology | 1 (5%) | |
| Average duration of monitoring hours, median (range) | 14 (0.5–17) | |
| Total duration of monitoring, hours | 264 | |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
van den Berge, P.; van Loon, K.; Zevenbergen, L.; van den Heuvel, P.A.; Breteler, M.J.M. Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study. Sensors 2026, 26, 1201. https://doi.org/10.3390/s26041201
van den Berge P, van Loon K, Zevenbergen L, van den Heuvel PA, Breteler MJM. Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study. Sensors. 2026; 26(4):1201. https://doi.org/10.3390/s26041201
Chicago/Turabian Stylevan den Berge, Patrick, Kim van Loon, Lianne Zevenbergen, Pascalle A. van den Heuvel, and Martine J. M. Breteler. 2026. "Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study" Sensors 26, no. 4: 1201. https://doi.org/10.3390/s26041201
APA Stylevan den Berge, P., van Loon, K., Zevenbergen, L., van den Heuvel, P. A., & Breteler, M. J. M. (2026). Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study. Sensors, 26(4), 1201. https://doi.org/10.3390/s26041201