Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide
Abstract
1. Introduction
2. Materials and Methods
2.1. Chemicals and Solutions
2.2. Polymer Microelectrode Arrays
2.3. Scanning Electron Microscopy and Elemental Composition Characterization
2.4. X-Ray Diffraction Characterization
2.5. Electrochemical Instrumentation
2.6. Microelectrode Calibration
2.7. Electrochemical Impedance Spectroscopy Measurements
2.8. Selectivity Measurements
2.9. Operational Stability and Biofouling Assessment
2.10. Data Analysis
3. Results and Discussion
3.1. Morphology and Chemical Analysis
3.2. Electrochemical Behavior in Acidic Electrolyte and in Neutral PBS
3.3. Electrochemical Active Surface Area
3.4. Electrochemical Impedance Spectroscopy
3.5. Oxygen Reduction Reaction at the Pt Surface
3.6. Hydrogen Peroxide Oxidation Reaction at the Pt Surface
3.7. Selectivity Evaluation
3.8. Operational Stability and Biofouling Effects
4. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Kress, G.J.; Mennerick, S. Action Potential Initiation and Propagation: Upstream Influences on Neurotransmission. Neuroscience 2009, 158, 211–222. [Google Scholar] [CrossRef] [PubMed]
- Wu, F.; Yu, P.; Mao, L. New Opportunities of Electrochemistry for Monitoring, Modulating, and Mimicking the Brain Signals. JACS Au 2023, 3, 2062–2072. [Google Scholar] [CrossRef] [PubMed]
- Liu, Y.; Liu, Z.; Tian, Y. Real-Time Tracking of Electrical Signals and an Accurate Quantification of Chemical Signals with Long-Term Stability in the Live Brain. Acc. Chem. Res. 2022, 55, 2821–2832. [Google Scholar] [CrossRef] [PubMed]
- Li, H.; Wang, J.; Fang, Y. Recent Developments in Multifunctional Neural Probes for Simultaneous Neural Recording and Modulation. Microsyst. Nanoeng. 2023, 9, 4. [Google Scholar] [CrossRef] [PubMed]
- Fernandes, E.; Ledo, A.; Gerhardt, G.A.; Barbosa, R.M. Amperometric Bio-Sensing of Lactate and Oxygen Concurrently with Local Field Potentials during Status Epilepticus. Talanta 2024, 268, 125302. [Google Scholar] [CrossRef] [PubMed]
- Ledo, A.; Lourenço, C.F.; Laranjinha, J.; Gerhardt, G.A.; Barbosa, R.M. Concurrent Measurements of Neurochemical and Electrophysiological Activity with Microelectrode Arrays: New Perspectives for Constant Potential Amperometry. Curr. Opin. Electrochem. 2018, 12, 129–140. [Google Scholar] [CrossRef]
- Siwakoti, U.; Jones, S.A.; Kumbhare, D.; Cui, X.T.; Castagnola, E. Recent Progress in Flexible Microelectrode Arrays for Combined Electrophysiological and Electrochemical Sensing. Biosensors 2025, 15, 100. [Google Scholar] [CrossRef] [PubMed]
- Ledo, A.; Lourenço, C.F.; Laranjinha, J.; Gerhardt, G.A.; Barbosa, R.M. Combined in Vivo Amperometric Oximetry and Electrophysiology in a Single Sensor: A Tool for Epilepsy Research. Anal. Chem. 2017, 89, 12383–12390. [Google Scholar] [CrossRef] [PubMed]
- Malekoshoaraie, M.H.; Wu, B.; Krahe, D.D.; Ahmed, Z.; Pupa, S.; Jain, V.; Cui, X.T.; Chamanzar, M. Fully Flexible Implantable Neural Probes for Electrophysiology Recording and Controlled Neurochemical Modulation. Microsyst. Nanoeng. 2024, 10, 91. [Google Scholar] [CrossRef] [PubMed]
- Vetter, R.J.; Williams, J.C.; Hetke, J.F.; Nunamaker, E.A.; Kipke, D.R. Chronic Neural Recording Using Silicon-Substrate Microelectrode Arrays Implanted in Cerebral Cortex. IEEE Trans. Biomed. Eng. 2004, 51, 896–904. [Google Scholar] [CrossRef] [PubMed]
- Burmeister, J.J.; Gerhardt, G.A. Self-Referencing Ceramic-Based Multisite Microelectrodes for the Detection and Elimination of Interferences from the Measurement of L-Glutamate and Other Analytes. Anal. Chem. 2001, 73, 1037–1042. [Google Scholar] [CrossRef] [PubMed]
- Yi, D.; Yao, Y.; Wang, Y.; Chen, L. Design, Fabrication, and Implantation of Invasive Microelectrode Arrays as in Vivo Brain Machine Interfaces: A Comprehensive Review. J. Manuf. Process. 2024, 126, 185–207. [Google Scholar] [CrossRef] [PubMed]
- Ledo, A.; Lourenço, C.F.; Laranjinha, J.; Brett, C.M.A.; Gerhardt, G.A.; Barbosa, R.M. Ceramic-Based Multisite Platinum Microelectrode Arrays: Morphological Characteristics and Electrochemical Performance for Extracellular Oxygen Measurements in Brain Tissue. Anal. Chem. 2017, 89, 1674–1683. [Google Scholar] [CrossRef] [PubMed]
- Xu, M.; Zhao, Y.; Xu, G.; Zhang, Y.; Sun, S.; Sun, Y.; Wang, J.; Pei, R. Recent Development of Neural Microelectrodes with Dual-Mode Detection. Biosensors 2023, 13, 59. [Google Scholar] [CrossRef] [PubMed]
- Choi, J.R.; Kim, S.M.; Ryu, R.H.; Kim, S.P.; Sohn, J.W. Implantable Neural Probes for Brain-Machine Interfaces-Current Developments and Future Prospects. Exp. Neurobiol. 2018, 27, 453–471. [Google Scholar] [CrossRef] [PubMed]
- Lecomte, A.; Descamps, E.; Bergaud, C. A Review on Mechanical Considerations for Chronically-Implanted Neural Probes. J. Neural Eng. 2018, 15, 031001. [Google Scholar] [CrossRef] [PubMed]
- Fu, T.M.; Hong, G.; Zhou, T.; Schuhmann, T.G.; Viveros, R.D.; Lieber, C.M. Stable Long-Term Chronic Brain Mapping at the Single-Neuron Level. Nat. Methods 2016, 13, 875–882. [Google Scholar] [CrossRef] [PubMed]
- Fuchsberger, K.; Goff, A.L.; Gambazzi, L.; Toma, F.M.; Goldoni, A.; Giugliano, M.; Stelzle, M.; Prato, M. Multiwalled Carbon-Nanotube-Functionalized Microelectrode Arrays Fabricated by Microcontact Printing: Platform for Studying Chemical and Electrical Neuronal Signaling. Small 2011, 7, 524–530. [Google Scholar] [CrossRef] [PubMed]
- Xu, T.; Ji, W.; Wang, X.; Zhang, Y.; Zeng, H.; Mao, L.; Zhang, M. Support-Free PEDOT:PSS Fibers as Multifunctional Microelectrodes for In Vivo Neural Recording and Modulation. Angew. Chem.-Int. Ed. 2022, 61, e202115074. [Google Scholar] [CrossRef] [PubMed]
- Steins, H.; Mierzejewski, M.; Brauns, L.; Stumpf, A.; Kohler, A.; Heusel, G.; Corna, A.; Herrmann, T.; Jones, P.D.; Zeck, G.; et al. A Flexible Protruding Microelectrode Array for Neural Interfacing in Bioelectronic Medicine. Microsyst. Nanoeng. 2022, 8, 131. [Google Scholar] [CrossRef] [PubMed]
- Wang, X.; Xu, M.; Yang, H.; Jiang, W.; Jiang, J.; Zou, D.; Zhu, Z.; Tao, C.; Ni, S.; Zhou, Z.; et al. Ultraflexible Neural Electrodes Enabled Synchronized Long-Term Dopamine Detection and Wideband Chronic Recording Deep in Brain. ACS Nano 2024, 18, 34272–34287. [Google Scholar] [CrossRef] [PubMed]
- Zhao, H.; Liu, R.; Zhang, H.; Cao, P.; Liu, Z.; Li, Y. Research Progress on the Flexibility of an Implantable Neural Microelectrode. Micromachines 2022, 13, 386. [Google Scholar] [CrossRef] [PubMed]
- Metallo, C.; White, R.D.; Trimmer, B.A. Flexible Parylene-Based Microelectrode Arrays for High Resolution EMG Recordings in Freely Moving Small Animals. J. Neurosci. Methods 2011, 195, 176–184. [Google Scholar] [CrossRef] [PubMed]
- De La Oliva, N.; Mueller, M.; Stieglitz, T.; Navarro, X.; Del Valle, J. On the Use of Parylene C Polymer as Substrate for Peripheral Nerve Electrodes. Sci. Rep. 2018, 8, 5965. [Google Scholar] [CrossRef] [PubMed]
- Zhang, Y.; Zhang, J.; Le, S.; Niu, L.; Tao, J.; Liang, J.; Zhang, L.; Kang, X. Parylene C as an Insulating Polymer for Implantable Neural Interfaces: Acute Electrochemical Impedance Behaviors in Saline and Pig Brain In Vitro. Polymers 2022, 14, 3033. [Google Scholar] [CrossRef] [PubMed]
- Scholten, K.; Xu, H.; Lu, Z.; Jiang, W.; Jin, Z.; Ortigoza-Diaz, J.; Petrossians, A.; Orler, S.; Gallonio, R.; Liu, X.; et al. A Comprehensive Research Dissemination Model for Polymer-Based Neural Interfaces. IEEE Trans. Biomed. Eng. 2026, 73, 934–944. [Google Scholar] [CrossRef] [PubMed]
- Gatti, I.; Eisawi, K.; Siwakoti, U.; Jadwin, G.; Sigdel, A.; Chiappalone, M.; Kumbhare, D.; Naguib, M.; Castagnola, E. MXene/PEDOT Functional Coatings on Flexible Microelectrode Arrays for Multianalyte In Vivo Neurochemical Sensing and Electrophysiology. Adv. Funct. Mater. 2026, 36, e75402. [Google Scholar] [CrossRef]
- Cortés-Salazar, F.; Deng, H.; Peljo, P.; Pereira, C.M.; Kontturi, K.; Girault, H.H. Parylene C Coated Microelectrodes for Scanning Electrochemical Microscopy. Electrochim. Acta 2013, 110, 22–29. [Google Scholar] [CrossRef]
- Sander, R. Compilation of Henry’s Law Constants (Version 4.0) for Water as Solvent. Atmos. Chem. Phys. 2015, 15, 4399–4981. [Google Scholar] [CrossRef]
- Daubinger, P.; Kieninger, J.; Unmüssig, T.; Urban, G.A. Electrochemical Characteristics of Nanostructured Platinum Electrodes-A Cyclic Voltammetry Study. Phys. Chem. Chem. Phys. 2014, 16, 8392–8399. [Google Scholar] [CrossRef] [PubMed]
- Talauliker, P.M.; Price, D.A.; Burmeister, J.J.; Nagari, S.; Quintero, J.E.; Pomerleau, F.; Huettl, P.; Hastings, J.T.; Gerhardt, G.A. Ceramic-Based Microelectrode Arrays: Recording Surface Characteristics and Topographical Analysis. J. Neurosci. Methods 2011, 198, 222–229. [Google Scholar] [CrossRef] [PubMed]
- Scholten, K.; Xu, H.; Song, D.; Meng, E. A shared resource for building polymer-based microelectrode arrays as neural interfaces. In Proceedings of the 11th International IEEE/EMBS Conference on Neural Engineering (NER 2023), Baltimore, MD, USA, 24–27 April 2023; IEEE Computer Society: Washington, DC, USA, 2023. [Google Scholar] [CrossRef]
- Pandey, A.; Dalal, S.; Dutta, S.; Dixit, A. Structural Characterization of Polycrystalline Thin Films by X-Ray Diffraction Techniques. J. Mater. Sci. Mater. Electron. 2021, 32, 1341–1368. [Google Scholar] [CrossRef]
- Marszalek, T.; Gazicki-Lipman, M.; Ulanski, J. Parylene C as a Versatile Dielectric Material for Organic Field-Effect Transistors. Beilstein J. Nanotechnol. 2017, 8, 1532–1545. [Google Scholar] [CrossRef] [PubMed]
- Saksena, A.; Chien, Y.-C.; Chang, K.; Kümmerl, P.; Hans, M.; Völker, B.; Schneider, J.M. Metastable Phase Formation of Pt-X (X = Ir, Au) Thin Films. Sci. Rep. 2018, 8, 10198. [Google Scholar] [CrossRef] [PubMed]
- Cozma, S.; Vlǎdoiu, R.; Mandes, A.; Dinca, V.; Prodan, G.; Buršíková, V. Characterization of Platinum-Based Thin Films Deposited by Thermionic Vacuum Arc (TVA) Method. Materials 2020, 13, 1796. [Google Scholar] [CrossRef] [PubMed]
- Bulowski, W.; Skibińska, K.; Żabiński, P.; Wojnicki, M. Optimization of Gold Thin Films by DC Magnetron Sputtering: Structure, Morphology, and Conductivity. Coatings 2025, 15, 1240. [Google Scholar] [CrossRef]
- Sheng, W.; Gasteiger, H.A.; Shao-Horn, Y. Hydrogen Oxidation and Evolution Reaction Kinetics on Platinum: Acid vs Alkaline Electrolytes. J. Electrochem. Soc. 2010, 157, B1529. [Google Scholar] [CrossRef]
- Weltin, A.; Kieninger, J. Electrochemical Methods for Neural Interface Electrodes. J. Neural Eng. 2021, 18, 052001. [Google Scholar] [CrossRef] [PubMed]
- Rinaldo, S.G.; Lee, W.; Stumper, J.; Eikerling, M. Mechanistic Principles of Platinum Oxide Formation and Reduction. Electrocatalysis 2014, 5, 262–272. [Google Scholar] [CrossRef]
- Lai, B.-C.; Wu, J.-G.; Luo, S.-C. Revisiting Background Signals and the Electrochemical Windows of Au, Pt, and GC Electrodes in Biological Buffers. ACS Appl. Energy Mater. 2019, 2, 6808–6816. [Google Scholar] [CrossRef]
- Ivanovskaya, A.N.; Belle, A.M.; Yorita, A.M.; Qian, F.; Chen, S.; Tooker, A.; Lozada, R.G.; Dahlquist, D.; Tolosa, V. Electrochemical Roughening of Thin-Film Platinum for Neural Probe Arrays and Biosensing Applications. J. Electrochem. Soc. 2018, 165, G3125–G3132. [Google Scholar] [CrossRef]
- Johnson, A. Electrochemical Surface Area (ECSA) Evaluation in Electrocatalysis: Principles, Measurement Techniques, and Future Perspectives. J. Eng. Ind. Res. 2025, 6, 212–222. [Google Scholar] [CrossRef]
- Yohannes, W.; Belenov, S.V.; Guterman, V.E.; Skibina, L.M.; Volotchaev, V.A.; Lyanguzov, N.V. Effect of Ethylene Glycol on Electrochemical and Morphological Features of Platinum Electrodeposits from Chloroplatinic Acid. J. Appl. Electrochem. 2015, 45, 623–633. [Google Scholar] [CrossRef]
- Barsoukov, E.; Macdonald, J.R. (Eds.) Impedance Spectroscopy Theory, Experiment, and Applications, 2nd ed.; Wiley-Interscience: Hoboken, NJ, USA, 2005. [Google Scholar]
- Hirschorn, B.; Orazem, M.E.; Tribollet, B.; Vivier, V.; Frateur, I.; Musiani, M. Determination of Effective Capacitance and Film Thickness from Constant-Phase-Element Parameters. Electrochim. Acta 2010, 55, 6218–6227. [Google Scholar] [CrossRef]
- Harris, A.R.; Molino, P.J.; Kapsa, R.M.I.; Clark, G.M.; Paolini, A.G.; Wallace, G.G. Correlation of the Impedance and Effective Electrode Area of Doped PEDOT Modified Electrodes for Brain-Machine Interfaces. Analyst 2015, 140, 3164–3174. [Google Scholar] [CrossRef] [PubMed]
- Zeng, Q.; Xia, K.; Zhang, Y.; Wu, T. Well Controlled 3D Iridium Oxide/Platinum Nanocomposites with Greatly Enhanced Electrochemical Performances. Adv. Mater. Interfaces 2019, 6, 1900356. [Google Scholar] [CrossRef]
- Huang, L.; Zaman, S.; Tian, X.; Wang, Z.; Fang, W.; Xia, B.Y. Advanced Platinum-Based Oxygen Reduction Electrocatalysts for Fuel Cells. Acc. Chem. Res. 2021, 54, 311–322. [Google Scholar] [CrossRef] [PubMed]
- Ahsan, M.A.; He, T.; Eid, K.; Abdullah, A.M.; Sanad, M.F.; Aldalbahi, A.; Alvarado-Tenorio, B.; Du, A.; Puente Santiago, A.R.; Noveron, J.C. Controlling the Interfacial Charge Polarization of MOF-Derived 0D–2D vdW Architectures as a Unique Strategy for Bifunctional Oxygen Electrocatalysis. ACS Appl. Mater. Interfaces 2022, 14, 3919–3929. [Google Scholar] [CrossRef] [PubMed]
- Keeley, T.P.; Mann, G.E. Defining Physiological Normoxia for Improved Translation of Cell Physiology to Animal Models and Humans. Physiol. Rev. 2019, 99, 161–234. [Google Scholar] [CrossRef] [PubMed]
- Tedjo, W.; Obeidat, Y.; Catandi, G.; Carnevale, E.; Chen, T. Real-Time Analysis of Oxygen Gradient in Oocyte Respiration Using a High-Density Microelectrode Array. Biosensors 2021, 11, 256. [Google Scholar] [CrossRef] [PubMed]
- Xiang, L.; Yu, P.; Zhang, M.; Hao, J.; Wang, Y.; Zhu, L.; Dai, L.; Mao, L. Platinized Aligned Carbon Nanotube-Sheathed Carbon Fiber Microelectrodes for in Vivo Amperometric Monitoring of Oxygen. Anal. Chem. 2014, 86, 5017–5023. [Google Scholar] [CrossRef] [PubMed]
- Bazzu, G.; Puggioni, G.G.M.; Dedola, S.; Calia, G.; Rocchitta, G.; Migheli, R.; Desole, M.S.; Lowry, J.P.; O’Neill, R.D.; Serra, P.A. Real-Time Monitoring of Brain Tissue Oxygen Using a Miniaturized Biotelemetric Device Implanted in Freely Moving Rats. Anal. Chem. 2009, 81, 2235–2241. [Google Scholar] [CrossRef] [PubMed]
- Bolger, F.B.; Bennett, R.; Lowry, J.P. An in Vitro Characterisation Comparing Carbon Paste and Pt Microelectrodes for Real-Time Detection of Brain Tissue Oxygen. Analyst 2011, 136, 4028–4035. [Google Scholar] [CrossRef] [PubMed]
- Sosna, M.; Denuault, G.; Pascal, R.W.; Prien, R.D.; Mowlem, M. Development of a Reliable Microelectrode Dissolved Oxygen Sensor. Sens. Actuators B Chem. 2007, 123, 344–351. [Google Scholar] [CrossRef]
- Chen, Y.P.; Liu, S.Y.; Fang, F.; Li, S.H.; Liu, G.; Tian, Y.C.; Xiong, Y.; Yu, H.Q. Simultaneous Determination of Nitrate and Dissolved Oxygen under Neutral Conditions Using a Novel Silver-Deposited Gold Microelectrode. Environ. Sci. Technol. 2008, 42, 8465–8470. [Google Scholar] [CrossRef] [PubMed]
- Pennings, F.A.; Schuurman, P.R.; Van Den Munckhof, P.; Bouma, G.J. Brain Tissue Oxygen Pressure Monitoring in Awake Patients during Functional Neurosurgery: The Assessment of Normal Values. Proc. J. Neurotrauma 2008, 25, 1173–1177. [Google Scholar] [CrossRef] [PubMed]
- Liu, S.-Y.; Chen, Y.-P.; Fang, F.; Xu, J.; Sheng, G.-P.; Yu, H.-Q.; Liu, G.; Tian, Y.-C. Measurement of Dissolved Oxygen and Its Diffusivity in Aerobic Granules Using a Lithographically-Fabricated Microelectrode Array. Environ. Sci. Technol. 2009, 43, 1160–1165. [Google Scholar] [CrossRef] [PubMed]
- Xiao, T.; Li, X.; Wei, H.; Ji, W.; Yue, Q.; Yu, P.; Mao, L. In Vivo Monitoring of Oxygen Fluctuation Simultaneously at Multiple Sites of Rat Cortex during Spreading Depression. Anal. Chem. 2018, 90, 13783–13789. [Google Scholar] [CrossRef] [PubMed]
- Lourenço, C.F.; Caetano, M.; Ledo, A.; Barbosa, R.M. Platinized Carbon Fiber-Based Glucose Microbiosensor Designed for Metabolic Studies in Brain Slices. Bioelectrochemistry 2019, 130, 107325. [Google Scholar] [CrossRef] [PubMed]
- Yu, H.; Yu, J.; Li, L.; Zhang, Y.; Xin, S.; Ni, X.; Sun, Y.; Song, K. Recent Progress of the Practical Applications of the Platinum Nanoparticle-Based Electrochemistry Biosensors. Front. Chem. 2021, 9, 677876. [Google Scholar] [CrossRef] [PubMed]
- Katsounaros, I.; Schneider, W.B.; Meier, J.C.; Benedikt, U.; Biedermann, P.U.; Auer, A.A.; Mayrhofer, K.J.J. Hydrogen Peroxide Electrochemistry on Platinum: Towards Understanding the Oxygen Reduction Reaction Mechanism. Phys. Chem. Chem. Phys. 2012, 14, 7384–7391. [Google Scholar] [CrossRef] [PubMed]
- Li, Y.; Sella, C.; Lemaître, F.; GuilleCollignon, M.; Thouin, L.; Amatore, C. Highly Sensitive Platinum-Black Coated Platinum Electrodes for Electrochemical Detection of Hydrogen Peroxide and Nitrite in Microchannel. Electroanalysis 2013, 25, 895–902. [Google Scholar] [CrossRef]
- Chatard, C.; Sabac, A.; Moreno-Velasquez, L.; Meiller, A.; Marinesco, S. Minimally Invasive Microelectrode Biosensors Based on Platinized Carbon Fibers for in Vivo Brain Monitoring. ACS Cent. Sci. 2018, 4, 1751–1760. [Google Scholar] [CrossRef] [PubMed]
- Wang, B.; Wen, X.; Chiou, P.Y.; Maidment, N.T. Pt Nanoparticle-Modified Carbon Fiber Microelectrode for Selective Electrochemical Sensing of Hydrogen Peroxide. Electroanalysis 2019, 31, 1641–1645. [Google Scholar] [CrossRef]
- Yao, Y.; Bai, X.; Shiu, K.K. Spontaneous Deposition of Prussian Blue on Multi-Walled Carbon Nanotubes and the Application in an Amperometric Biosensor. Nanomaterials 2012, 2, 428–444. [Google Scholar] [CrossRef] [PubMed]
- Chang, S.H.; Yeh, M.H.; Rick, J.; Su, W.N.; Liu, D.G.; Lee, J.F.; Liu, C.C.; Hwang, B.J. Bimetallic Catalyst of PtIr Nanoparticles with High Electrocatalytic Ability for Hydrogen Peroxide Oxidation. Sens. Actuators B Chem. 2014, 190, 55–60. [Google Scholar] [CrossRef]
- Sellen, S.C.; Siwakoti, U.; Sigdel, A.; Jaiswal, B.; Zivanovic, S.; Castagnola, E. Advancing Glassy Carbon Microelectrode Arrays for Neurochemical Sensing: Impact of Double Pyrolysis on Structure and Function. Front. Bioeng. Biotechnol. 2025, 13, 1642063. [Google Scholar] [CrossRef] [PubMed]
- Giaretta, J.E.; Duan, H.; Oveissi, F.; Farajikhah, S.; Dehghani, F.; Naficy, S. Flexible Sensors for Hydrogen Peroxide Detection: A Critical Review. ACS Appl. Mater. Interfaces 2022, 14, 20491–20505. [Google Scholar] [CrossRef] [PubMed]
- Spanos, M.; Gras-Najjar, J.; Letchworth, J.M.; Sanford, A.L.; Toups, J.V.; Sombers, L.A. Quantitation of Hydrogen Peroxide Fluctuations and Their Modulation of Dopamine Dynamics in the Rat Dorsal Striatum Using Fast-Scan Cyclic Voltammetry. ACS Chem. Neurosci. 2013, 4, 782–789. [Google Scholar] [CrossRef] [PubMed]
- Wilson, L.R.; Panda, S.; Schmidt, A.C.; Sombers, L.A. Selective and Mechanically Robust Sensors for Electrochemical Measurements of Real-Time Hydrogen Peroxide Dynamics in Vivo. Anal. Chem. 2018, 90, 888–895. [Google Scholar] [CrossRef] [PubMed]
- Burmeister, J.J.; Moxon, K.; Gerhardt, G.A. Ceramic-Based Multisite Microelectrodes for Electrochemical Recordings. Anal. Chem. 2000, 72, 187–192. [Google Scholar] [CrossRef] [PubMed]
- Carriere, V.M.; Rodrigues, J.P.; Tan, C.; Arumugam, P.; Poh, S. In Vitro Electrochemical Detection of Hydrogen Peroxide in Activated Macrophages via a Platinum Microelectrode Array. Sensors 2021, 21, 5607. [Google Scholar] [CrossRef] [PubMed]
- Ben-Amor, S.; Vanhove, E.; Sékli Belaïdi, F.; Charlot, S.; Colin, D.; Rigoulet, M.; Devin, A.; Sojic, N.; Launay, J.; Temple-Boyer, P.; et al. Enhanced Detection of Hydrogen Peroxide with Platinized Microelectrode Arrays for Analyses of Mitochondria Activities. Electrochim. Acta 2014, 126, 171–178. [Google Scholar] [CrossRef]
- Pang, H.; Yin, B.; Liu, X.; Jiang, T.; Wu, S. Simultaneous Electrochemical Detection of Dopamine and Hydrogen Peroxide Based on Bifunctional Microelectrode Array. J. Mater. Sci. Mater. Electron. 2023, 34, 883. [Google Scholar] [CrossRef]
- Killoran, S.J.; O’Neill, R.D. Characterization of Permselective Coatings Electrosynthesized on Pt-Ir from the Three Phenylenediamine Isomers for Biosensor Applications. Electrochim. Acta 2008, 53, 7303–7312. [Google Scholar] [CrossRef]
- Soldatkina, O.V.; Kucherenko, I.S.; Pyeshkova, V.M.; Alekseev, S.A.; Soldatkin, O.O.; Dzyadevych, S.V. Improvement of Amperometric Transducer Selectivity Using Nanosized Phenylenediamine Films. Nanoscale Res. Lett. 2017, 12, 594. [Google Scholar] [CrossRef] [PubMed]
- Adediji, M.Y.; Billa, S.; Siddiqui, S.; Arumugam, P.U. Enhancing the Sensitivity of Gamma-Aminobutyric Acid and Glutamate Biosensors by Electrochemically Roughening Platinum Microelectrodes. Front. Neurosci. 2025, 19, 1679591. [Google Scholar] [CrossRef] [PubMed]
- Miele, M.; Fillenz, M. In Vivo Determination of Extracellular Brain Ascorbate. J. Neurosci. Methods 1996, 70, 15–19. [Google Scholar] [CrossRef] [PubMed]
- Lakard, S.; Pavel, I.-A.; Lakard, B. Electrochemical Biosensing of Dopamine Neurotransmitter: A Review. Biosensors 2021, 11, 179. [Google Scholar] [CrossRef] [PubMed]
- Keteklahijani, Y.Z.; Sharif, F.; Roberts, E.P.L.; Sundararaj, U. Enhanced Sensitivity of Dopamine Biosensors: An Electrochemical Approach Based on Nanocomposite Electrodes Comprising Polyaniline, Nitrogen-Doped Graphene, and DNA-Functionalized Carbon Nanotubes. J. Electrochem. Soc. 2019, 166, B1415–B1425. [Google Scholar] [CrossRef]
- Castagnola, E.; Lee, K.H.; Ricci, D. Editorial: Carbon Microelectrodes for Neurochemical Sensing. Front. Bioeng. Biotechnol. 2026, 14, 1799334. [Google Scholar] [CrossRef] [PubMed]
- Liljeström, T.; Kontturi, K.S.; Durairaj, V.; Wester, N.; Tammelin, T.; Laurila, T.; Koskinen, J. Protein Adsorption and Its Effects on Electroanalytical Performance of Nanocellulose/Carbon Nanotube Composite Electrodes. Biomacromolecules 2023, 24, 3806–3818. [Google Scholar] [CrossRef] [PubMed]







| Rs (kΩ) | Re | CPE | n |
|---|---|---|---|
| 2.48 | 1.3 GΩ | 13.0 nF sn−1 | 0.84 |
| 45.76 [a] kΩ cm2 | 369.3 [a] µF cm−2 sn−1 |
| Microelectrode Type | Electrode Material (Dimensions, μm) | Sensitivity (nA μM−1) | Sensitivity/Unit Area (mA mM−1 cm−2) | LOD (nM) | |Z| at 1 kHz/Unit Area (Ω cm2) | Response Time (s) | Ref. |
|---|---|---|---|---|---|---|---|
| Flexible (Parylene-C) | Pt (30 ⌀) | −0.25 ± 0.04 | −11.54 ± 4.75 | 5.4 ± 1.4 | 1.17 | 5.7 ± 0.6 | This work |
| Flexible (SU-8) | Pt (20 × 20) | −0.00426 b | −1.065 b | nd | nd | <15 | [59] |
| Ceramic | Pt (15 × 333) | −0.16 ± 0.02 | −3.2 ± 0.5 | 330 ± 200 | 10.8 | na | [13] |
| Silicon | Pt (30 ⌀) | nd | nd | nd | nd | nd | [60] |
| Silicon | Pt (17.5 × 17.5) | −0.0082 | −2.68 | 20,090 | nd | nd | [52] |
| Microelectrode Type | Electrode Material (Dimensions, μm) | Sensitivity (nA μM−1) | Sensitivity/Unit Area (mA mM−1 cm−2) | LOD (nM) | |Z| at 1 kHz/Unit Area (Ω cm2) | Response Time (s) | Ref. |
|---|---|---|---|---|---|---|---|
| Flexible (Parylene-C) | Pt (30 ⌀) | 88.13 ± 7.61 | 2.56 ± 0.48 | 41.9 ± 5.6 | 1.17 | 8.5 ± 0.2 | This work |
| Ceramic | Pt (50 × 50) | 0.0122 | 0.488 b | 133 | nd | nd | [73] |
| Ceramic | Pt (100 × 50) | 0.006743 | 0.135 b | nd | nd | nd | [74] |
| Silicon | Pt Black (20 ⌀) | nd | nd | 10 | nd | <1 | [75] |
| Glass | Pt Black (20 ⌀) | 0.0172 b | 5.48 | nd | nd | nd | [76] |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Fernandes, E.; Ledo, A.; Scholten, K.; Meng, E.; Gerhardt, G.A.; Barbosa, R.M. Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide. Sensors 2026, 26, 4929. https://doi.org/10.3390/s26154929
Fernandes E, Ledo A, Scholten K, Meng E, Gerhardt GA, Barbosa RM. Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide. Sensors. 2026; 26(15):4929. https://doi.org/10.3390/s26154929
Chicago/Turabian StyleFernandes, Eliana, Ana Ledo, Kee Scholten, Ellis Meng, Greg A. Gerhardt, and Rui M. Barbosa. 2026. "Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide" Sensors 26, no. 15: 4929. https://doi.org/10.3390/s26154929
APA StyleFernandes, E., Ledo, A., Scholten, K., Meng, E., Gerhardt, G. A., & Barbosa, R. M. (2026). Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide. Sensors, 26(15), 4929. https://doi.org/10.3390/s26154929