Next Article in Journal
W-Band 4th Order Waveguide Filter Based on Double Layer SU8 Microfabrication
Next Article in Special Issue
Design of Under-Actuated Soft Adhesion Actuators for Climbing Robots
Previous Article in Journal
Virtual Reality Game for Physical and Emotional Rehabilitation of Landmine Victims
Previous Article in Special Issue
Development of Wrist Interface Based on Fully Actuated Coaxial Spherical Parallel Mechanism for Force Interaction
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Electrical Optimization Method Based on a Novel Arrangement of the Magnetic Navigation System with Gradient and Uniform Saddle Coils

Department of Robotics, Kwangwoon University, Seoul 01897, Korea
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(15), 5603; https://doi.org/10.3390/s22155603
Submission received: 30 June 2022 / Revised: 21 July 2022 / Accepted: 23 July 2022 / Published: 27 July 2022

Abstract

:
The magnetic navigation system (MNS) with gradient and uniform saddle coils is an effective system for manipulating various medical magnetic robots because of its compact structure and the uniformity of its magnetic field and field gradient. Since each coil of the MNS was geometrically optimized to generate strong uniform magnetic field or field gradient, it is considered that no special optimization is required for the MNS. However, its electrical characteristics can be still optimized to utilize the maximum power of a power supply unit with improved operating time and a stronger time-varying magnetic field. Furthermore, the conventional arrangement of the coils limits the maximum three-dimensional (3D) rotating magnetic field. In this paper, we propose an electrical optimization method based on a novel arrangement of the MNS. We introduce the objective functions, constraints, and design variables of the MNS considering electrical characteristics such as resistance, current density, and inductance. Then, we design an MNS using an optimization algorithm and compare it with the conventional MNS; the proposed MNS generates a magnetic field or field gradient 22% stronger on average than that of the conventional MNS with a sevenfold longer operating time limit, and the maximum three-dimensional rotating magnetic field is improved by 42%. We also demonstrate that the unclogging performance of the helical robot improves by 54% with the constructed MNS.

1. Introduction

Magnetic robots and tools have attracted considerable attention as alternatives to conventional medical devices such as catheters, endoscopes, grippers, and forceps [1,2,3,4,5,6,7,8]. Remote magnetic actuation not only provides an unlimited operation time, but also provides the advantage of being able to miniaturize the robot because it does not require batteries. As a part of this miniaturization, magnetic particles also have been studied for various medical purpose such as targeted drug delivery, embolization, and imaging technology [9,10,11,12,13]. These advantages render magnetic robots suitable for minimally invasive medical devices that must operate for a long time inside small and narrow human organs. However, magnetic robots require a magnetic navigation system (MNS), which generates an external magnetic field to the magnetic robots. MNSs can be classified into two types, with and without a magnetic core. The MNSs with a core are particularly advantageous in generating a strong magnetic field to manipulate small magnetic nanoparticles or microrobots [14,15,16,17] because the core amplifies the generated magnetic field. However, the coreless MNSs allow a magnetic robot to be controlled without localization because they generate uniform magnetic field or field gradient over a large workspace. Magnetic resonance imaging (MRI) or magnetic particle imaging (MPI) devices can also be used for manipulation, but they are not satisfactory for various medical magnetic robots because their functions are focused on imaging [8,9,10].
A simple coreless MNS combines several pairs of Helmholtz coils (HCs) and Maxwell coils (MCs). These circular coils are structurally simple; hence, many researchers have used this combination to actuate magnetic robots [18,19,20,21,22,23,24,25]. Three pairs of HCs are the basic combination for generating a three-dimensional (3D) magnetic field [8,9,10]. Several pairs of MCs can also be integrated with the HCs to generate both a magnetic field and a field gradient [21,22,23]. However, the combination of multiple pairs of circular coils is geometrically inefficient. These combinations produce large empty spaces between pairs of coils. To address this issue, rotatable MNSs were developed [24,25]; instead of increasing the number of coils, the coils can be rotated to ensure the required degree of freedom (DOF). However, this method makes the MNS complex because it requires additional electromechanical systems. In addition, uniform saddle coils (USCs) and gradient saddle coils (GSCs) have been developed [26]. These coils can surround circular coils to form a compact cylindrical structure without requiring extra space. The cylindrical shape is appropriate for accommodating the human body and is used in other similar medical devices, such as magnetic resonance imaging (MRI) machines and computerized tomography (CT) scanners. There are various combinations of circular and saddle coils [26,27,28,29,30]. In particular, the MNS composed of five pairs of coils (two USCs, a GSC, an HC, and an MC) can generate a 3D rotation and two-dimensional (2D) translation of magnetic robots [27]. Since each coil of the MNS was geometrically optimized to generate strong uniform magnetic field or field gradient, it is considered that no special optimization is required for the MNS. However, even if coils have the same geometry, electronical characteristics, such as resistance, current density, and inductance, can be different depending on the number of turns and thickness of wires. These electronic characteristics affect the performance of the MNS. For example, if the resistance of a coil is too large or too small, a power supply unit cannot provide maximum output power owing to its current or voltage limit. If a coil has high current density, it reduces the operating time limit owing to the rapid temperature rise of the coil. If a coil has large inductance, we obtain a small time-varying magnetic field owing to large inductance effect. Furthermore, the arrangement of the coils can affect the performance of the MNS for a 3D rotating magnetic field, but this has not been considered.
In this study, an electrical optimization method is proposed on the basis of a novel arrangement of the MNS with five pairs of coils. First, two pipes for installing the five pairs of coils were determined so as to have the required inner space. Then, we propose a novel arrangement of the five pairs of coils that enables the MNS to generate similar three-axis magnetic fields per current because the available amplitude of a 3D rotating magnetic field is restricted by the coil that generates the smallest magnetic field. Next, the objective functions, design variables, and constraints were defined for optimization, with some assumptions to simplify the calculation. The constraints include the inductance effect and current density of the coils such that the MNS can generate a strong time-varying magnetic field under a given temperature limit. The MNS was optimized using an optimization algorithm with the given constraints. Finally, the optimized MNS was constructed and verified. Then, the performance was compared with the conventional MNS. We confirmed that the optimized MNS could generate a stronger magnetic field in both static and dynamic conditions. The improved magnetic field can be used to enhance the quasi-static and dynamic motion of magnetic robots. As one example, we demonstrated that the optimized MNS can effectively improve the unclogging performance of a helical robot.

2. Novel Arrangement of the MNS

2.1. Magnetic Field and Field Gradient Generated by Each Pair of Coils

The magnetic field and field gradient from the MNS produce a magnetic torque and force for a magnetic robot to generate rotation and translation motion [28]. The torque and force can be expressed as
F = ( m ) B ,
T = ( m × B ) ,
respectively, where m is the magnetic moment of the robot and B is the external magnetic field. This external magnetic field can be generated by five pairs of coils, as shown in Figure 1, and the three components of the magnetic field vector can be expressed as follows [26]:
B M N S = [ B h + ( G g + G m ) x B u y + ( 2.4398 G g 0.5 G m ) y B u z + ( 1.4398 G g 0.5 G m ) z ] ,
B h = ( 4 / 5 ) 3 / 2 N h I h μ 0 / r h ,   B u y = 0.6004 N u y I u y μ 0 / r u y ,   B u z = 0.6004 N u z I u z μ 0 / r u z , G m = ( 16 / 3 ) ( 3 / 7 ) 5 / 2 N m I m μ 0 / r m 2 ,   G g = 0.3286 N g I g μ 0 / r g 2 ,
where N k , I k , and r k are the number of turns, current, and radius of the k-th coil, respectively, the subscripts h, m, g, uy, and uz represent the HC, MC, GSC, and y- and z-directional USCs, respectively, and μ 0 is the magnetic permeability of the free space. In Equation (3), x, y, and z are assumed to be negligible around the center of the MNS, where the workspace is located. Then, the magnetic torque is generated by three components ( B h , B u y , and B u z ), whereas the magnetic force is generated by two components ( G g and G m ).

2.2. Frame of the MNS

The frame of the MNS can be made of pipes owing to its cylindrical structure of the MNS. Figure 2a shows the configuration of the MNS using two pipes. Because the coils can be attached to both sides of the pipes, the two pipes are sufficient to install all coils. The diameter of each pipe was determined by considering the volume and magnetic field of the MNS. If the diameter is too small, the MNS cannot provide sufficient inner space for a magnetic robot. In contrast, if the diameter is too large, the coils cannot generate a sufficiently strong magnetic field to actuate the magnetic robot. According to this idea, the inner and outer diameters of the pipes ( d i n and d o u t ) were determined to be 31 and 46 cm, respectively. The thickness of the pipes ( t p ) was 5 mm, and the gap between the pipes ( d g a p ) was 7 cm. The pipes had four square windows to observe the inner space of the MNS during the experiment.

2.3. Arrangement of the MNS

Various mechanisms of magnetic robots are based on a 3D rotating magnetic field; however, the available amplitude of a 3D rotating magnetic field is restricted by the coil that generates the smallest magnetic field. Thus, the three components of the magnetic field ( B h , B u y , and B u z ) should have the same values. In Equation (3), the HC can generate a larger magnetic field than the USC under the same conditions ( N k , I k , and r k ). Therefore, the HC should be placed outside the USC. Then, B h can be equal to B u y or B u z with a larger radius. However, because USCy and USCz are of the same type of coil, if the same N k and I k are given, the magnetic field generated by the outer coil is inevitably smaller. To solve this problem, we divided USCz and placed USCy between them; USCz1 and USCz2 were connected in series. Then, by adjusting the ratio of USCz1 to USCz2, USCy and USCz could generate the same magnetic field ( B u y and B u z ).
Figure 2 shows this novel arrangement of the divided USCz. For two pipes, this arrangement was the only possible case. The MC and GSC were then placed in the empty spaces. The MC was placed on the outermost side of the MNS because it could be overlapped with the HC. In contrast, the GSC could be overlapped with USCy or USCz. The GSC could generate the strongest magnetic field gradient if it is placed with USCz; however, in this case, the GSC reduced the inner space of the MNS because the GSC was thicker than the divided coil (USCz2). Thus, we placed the GSC with USCy.

3. Electrical Optimization of the MNS

3.1. Objective Function and Design Variables of Each Coil

The MNS has three magnetic field components ( B h , B u y , and B u z ) and two field gradient components ( G g and G m ). We assume that the maximum output of the power supply unit ( P o u t ) is utilized for each coil. Then, the five components of the MNS can be rewritten as
B k , m a x = n k N k μ 0 r k P o u t R k   ,
G k , m a x = n k N k μ 0 r k 2 P o u t R k   ,
where n k and R k are the coefficient (in Equation (3)) and resistance of the k-th coil, respectively. These five components become the objective functions of the coils, and our goal is to maximize these values with several constraints, which are described in Section 3.2. In this study, we defined the design variables as the number of turns ( N k ) and the thickness of the wire ( t k ). We could then reorganize the objective functions using these design variables. In Equations (4) and (5), the resistance ( R k ) of the k-th coil can be expressed as
R k = l k A k ρ c o i l ,
where l k , A k , and ρ k are the total length, cross-sectional area, and resistivity of the wound wire with N k turns, respectively. In Equation (6), the cross-sectional area ( A k ) can be calculated assuming a circular wire, and the total length ( l k ) can be calculated assuming that the wire is wound only at the center of the coil. Consequently, l k and A k can be calculated as follows:
l k   = c k r k N k ,
c h   = c m = 4 π ,   c u y = c u z = 22.3788 , c g = 13.3704 ,
A k = π t k 2 / 4 ,
where c k is the geometric coefficient of the k-th coil. The coefficients of c h and c m can be calculated considering the circular shape of the HC and MC, and the coefficients of c u y , c u z , and c g can be calculated considering the saddle shapes of the USC and GSC. Furthermore, r k can be calculated assuming that the bundle of wires forms a square cross-section, as shown in Figure 3. Although the actual cross-section is close to a circle, this assumption is valid because the radius of the bundle is very small compared to the radius of the coil. Then, r k can be expressed as
r k = 0.5 d k ± 0.5 t k N k ,
d h , m = d o u t ,   d u z 1 = d o u t 2 t p ,   d u y ,   g = d i n ,   d u z 2 = d i n 2 t p ,
where d k is the outer or inner diameter of the pipe to which the k-th coil is attached. In Equation (9), the sign is determined on the basis of the location of the coil attached to the pipe. If the coil is attached to the outside, it has a positive sign (+); if the coil is attached to the inside, it has a negative sign (−). Finally, the objective function of the k-th coil can be rewritten using the two design variables ( N k and t k ) by substituting Equations (6)–(9) into Equations (4) and (5).
B k , m a x = μ 0 n k t k ( 0.5 d k ± 0.5 t k N k ) 3 / 2 π N k P o u t 4 c k ρ c o i l ,
G k , m a x = μ 0 n k t k ( 0.5 d k ± 0.5 t k N k ) 5 / 2 π N k P o u t 4 c k ρ c o i l .
In Equations (10) and (11), the resistivity ( ρ k ) can be changed depending on a temperature of the coil, but we assumed it as a constant value because we could not control the temperature without an additional cooling system.

3.2. Constraints of Each Coil

We considered several constraints to optimize the MNS. The first constraint was that the maximum values of the three components of the magnetic field are equal.
B h , m a x = B u y , m a x = B u z , m a x .
This constraint was described in Section 2.3. Using this constraint, the 3D rotating magnetic field can be maximized. We also considered the temperature rise of the coils because an insulated wire has an allowable temperature for safe use, which limits the available operating time of each coil. In addition, this temperature rise can deteriorate the performance of the MNS because of the increase in resistivity. Because the temperature of a coil is proportional to its current density, enlarging the cross-sectional area of each coil ( T k 2 ) can be the solution to increase the operating time limit. However, USCy, USCz1, and the GSC share a limited space between the pipes, as shown in Figure 2c. To enlarge the cross-sectional area without interference from the coils, we introduce the following constraints:
T u y + T u z 1 < d g a p ,
T g + T u z 1 < d g a p ,
0.5 d g a p < T k .
In Equation (15), T u z 1 and T u z 2 are not considered separately because they are connected in series. Instead, their summation ( T u z 1 + T u z 2 ) is considered as T u z . We also consider the inductance of the coils ( L k ). The inductance effect attenuates the time-varying current of the coils. Thus, a small inductance is advantageous for generating a strong time-varying magnetic field. Because the inductance is proportional to N k , we can obtain a small inductance by reducing N k . However, this also reduces R k in Equation (6) by increasing A k and decreasing l k to keep T k constant. Thus, we can minimize the inductance by minimizing the resistance. However, we should consider the range of R k to utilize the maximum power of the power supply unit. Figure 4 shows the output range of the power supply unit (3001iX by California Instruments); the resistance must be in the range of 8.62 Ω ≤ R k ≤ 19.05 Ω to utilize the maximum power. Then, the available minimum resistance becomes 8.62 Ω to minimize inductance of each coil, and we obtain the following resistance constraint:
R k = 8.62   Ω .
We also constrained t k because a wire that is too thick is difficult to wind. We experientially introduce the following constraint:
t k < 2   mm .

3.3. Algorithm to Optimize the MNS

We developed an optimization algorithm using Equations (10) and (11) and the given constraints, shown in Figure 5. We set the range of the two design variables as 0.001 t k 2 and 1 N k < 1420 . The range of t k was determined using Equation (17), and its minimum incremental value was 0.001 mm. The maximum number of turns of wires between the pipes was 1420 when t k had a minimum value of 0.001 mm. Thus, the maximum limit of N k was 1420. Using this algorithm, we obtained the optimized MNS as shown in Table 1. Because the mathematic functions in Equations (10) and (11) were utilized for the algorithm, the calculation time for each step was very short. Thus, the results of the optimization could be obtained in a few seconds.

4. Experiments

4.1. Construction of the MNS

The optimized MNS was constructed, as shown in Figure 6, and the major variables were measured, as shown in Table 2. Each coil was connected to a power supply unit (3001iX, California Instruments). A magnetic robot inside the MNS was tracked using a real-time camera and was controlled by a joystick controller. The two pipes for the coils were made of fiber-reinforced plastic that could withstand the weight and heat of the coils. Plastic is a nonmetallic material; therefore, the pipes have no iron loss that attenuates the magnetic field. In contrast, the structure used to support the pipes was made of metallic aluminum. Although aluminum can cause iron loss, it does not affect the magnetic field inside the MNS because the structure is located outside the coils.
The constructed MNS had considerable geometrical errors because there was no available winding machine for the coils, owing to their unusual size and shape. In particular, the circular HC and MC were wound with the larger radius ( r k ) than the designed value. This is because they were attached to the outside of the circular pipe. If the radius ( r k ) is smaller than the designed value, they cannot be assembled with the pipe. Thus, a margin was added to manufacture the HC and MC. As a result, these errors can be reflected by the correction factors, which are the ratios of the measured and designed values. Table 3 lists the correction factors; the correct B k or G k can be obtained by multiplying the correction factor and each designed B k or G k , respectively.
We also measured the magnetic field near the center of the MNS to verify the spatial homogeneity of the magnetic field. During the experiment, a current of 1 A was applied to each coil, and the magnetic field was measured using a Gauss meter (Model 8030 by F. W. Bell). We calculated the magnetic field of each coil considering the correction factor, and Figure 7 shows the comparison between the calculated and measured magnetic field. We confirmed that the magnetic fields of the HC and MC matched well with the calculated values. However, the saddle coils had relatively large error at the outside of the center because they could not be ideally constructed due to their complex geometry. In particular, the maximum error of 2.9% was measured at the USCy which had the maximum thickness of wire.

4.2. Heating Effect

The electrical insulation system for the wires was divided into different classes by temperature; we used an F-class wire for each coil. The F-class wire has an allowable temperature of 155 °C; this must be a temperature limit for safe use. The temperature rise due to the current of the coils was measured using an infrared thermometer, as shown in Figure 8, while the maximum output power was utilized. To obtain an average value, the temperature was measured at six points on each coil. Only USCz2 demonstrated an unlimited operating time because its temperature converged at 96 °C, while the temperatures of the other coils rose over 155 °C. In the figure, the USCz1 had the minimum operating time limit of 16 min, which can be considered the operating time limit of the MNS. However, when a magnetic robot is actuated, each coil operates discontinuously below the maximum output power. Thus, the practical operating time limit of the MNS would be longer than 16 min. For example, if we generate a 2D rotating magnetic field of 15 mT in the xy-plane, the operating time limit would be much longer than 40 min because we discontinuously utilize the HC and USCy below the maximum output power.

4.3. Inductance Effect

Each coil was designed to have a minimum inductance within the maximum output range of the power supply unit. The current drop due to the inductance effect can be expressed as follows [21]:
I k = V k R k 2 + ( 2 π L k f k ) 2   ,
where V k and f k are the input voltage and frequency of the power supply unit for the k-th coil, respectively. Because the magnetic field is proportional to the current in Equation (3), the calculated maximum magnetic field and field gradient in Table 1 cannot be obtained with this current drop. Figure 9 shows the measured maximum magnetic field and field gradient with a variation in the frequency. In particular, the amplitude of each maximum magnetic field decreased to 50% or less at 20 Hz. Thus, this inductance effect should be considered if we generate a time-varying magnetic field.

4.4. Comparison to the Conventional System

To verify the optimized MNS, we compared it with the conventional MNS in [21,22], as shown in Table 4 and Table 5. We selected this conventional MNS because it had an identical maximum output power and similar outer and inner diameters to the optimized MNS. As shown in the tables, each coil of the optimized MNS generated a larger magnetic field than that of the conventional MNS. In particular, the maximum 3D rotating magnetic field increased by 42%. However, the magnetic field gradient of the optimized MC decreased. Instead, it had half the inductance as before. This is because the conventional MC was designed to have a relatively large number of turns. Although several inductances of the optimized MNS increased, the maximum inductance of the MNS was reduced by 53% from 859.9 mH to 405.1 mH, which allowed the optimized MNS to generate a larger time-varying magnetic field and field gradient. We also observed an increase in the operating time limit owing to the enlarged cross-sectional area of each coil. If we consider the extreme condition in which the five coils are simultaneously utilized with the maximum output power, the optimized MNS can be operated for seven times longer than the conventional MNS.

4.5. Performance Test Using a Rotating Magnetic Field

The maximum 3D rotating magnetic field of the optimized MNS was 42% larger than that of the conventional MNS in the static state. This improvement of static magnetic field can enhance the quasi-static motions such as steering motion of a magnetic catheter and sampling motion of a magnetic capsule [2,16,31]. In addition, the optimized MNS can generate a stronger time-varying magnetic field with a smaller inductance effect, because the inductance was improved by 53%. This improvement can also enhance the dynamic motion of magnetic robots. As one example, we compared the unclogging ability of a helical robot in each MNS, as shown in Figure 10. First, the step-out frequencies of the helical robot were measured using the maximum 3D rotating magnetic fields of the two MNSs. The measured step-out frequency of the optimized MNS was 18 Hz, which is 38% higher than the 13 Hz of the conventional MNS. These frequency differences may affect the drilling ability of the robots. As shown in Figure 10b, the helical robot was actuated in front of the clogged area using agar. Considering the frictional energy consumption during the unclogging motion, slightly lower frequencies (17 Hz and 12 Hz) than the step-out frequencies (18 Hz and 13 Hz) were used for each experiment. As a result, the unclogging time improved by 54% from 96 s to 44 s. Because the propulsive force of a helical robot is proportional to the square of the robot dimension [32], larger robots would exhibit a better improvement. In particular, for the magnetic robots actuated in a fluidic flow such as blood flow in a vessel, their drag forces are proportional to the square of the fluidic velocity [24]. Thus, this improvement of the 3D rotating magnetic field would significantly help the magnetic robots to accomplish their mission.

5. Conclusions

In this paper, we proposed an electrical optimization method based on a novel arrangement of the MNS with saddle coils. The MNS was optimized to generate a greater magnetic field and field gradient than a conventional MNS with the same output power and similar size. As a result, each coil could generate an average of 22% stronger magnetic field or field gradient, and the maximum 3D rotating magnetic field was improved by 42%. The proposed optimization method also minimized the current density and inductance of the coils so that the MNS could generate a stronger time-varying magnetic field, with at least sevenfold longer operating time. The optimized MNS could effectively enhance the magnetic robots. As one example, we demonstrated that the unclogging performance of a helical robot was improved by 54%.
Through this study, we verified that, even if the coil is geometrically optimized, the performance of the coil can be limited in the case that the electrical characteristics are not properly designed. In contrast to coreless MNS, many researchers have studied the optimization method of MNSs with a core [15,16,17], because the performance of the MNSs with a core can significantly vary depending on various factors of the core such as the ratio between the core and coil, material property of the core, and shape of the core tip. However, they did not focus on the electrical characteristics of the MNSs. We believe that this electrical optimization can be applied to MNSs both with and without cores.
The proposed optimization method increases the cross-sectional area of the coils to suppress the temperature rise of the coils by reducing the current density. However, this may increase the material cost because copper coil is quite expensive. Thus, an excessively large cross-sectional area of the coil is not recommended in terms of cost. Although the thickness of the wire was limited, it was still too thick to manufacture complex coils without manufacturing errors; therefore, the thickness of wires should be carefully determined considering the shapes and volumes of the coils.

Author Contributions

Project administration and funding acquisition, J.N.; investigation, S.K. and M.C.; methodology, S.K. and J.N.; validation, S.K., M.C., S.I., J.Y. and J.N.; visualization, S.K.; writing—original draft, S.K.; writing—review and editing, S.K. and J.N. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by a National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (No. 2021R1C1C1014661) and a research grant from Kwangwoon University in 2020.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Nam, J.; Lai, Y.P.; Gauthier, L.; Jang, G.; Diller, E. Resonance-Based Design of Wireless Magnetic Capsule for Effective Sampling of Microbiome in Gastrointestinal Tract. Sens. Actuators A 2022, 342, 113654. [Google Scholar] [CrossRef]
  2. Shokrollahi, P.; Lai, Y.P.; Rash-Ahmadi, S.; Stewart, V.; Mohammadigheisar, M.; Huber, L.-A.; Matsuura, N.; Zavodni, A.E.H.; Parkinson, J.; Diller, E. Blindly Controlled Magnetically Actuated Capsule for Noninvasive Sampling of the Gastrointestinal Microbiome. IEEE/ASME Trans. Mechatron. 2021, 26, 2616–2628. [Google Scholar] [CrossRef]
  3. Forbrigger, C.; Lim, A.; Onaizah, O.; Salmanipour, S.; Looi, T.; Drake, J.; Diller, E.D. Cable-Less, Magnetically Driven Forceps for Minimally Invasive Surgery. IEEE Robot. Autom. Lett. 2019, 4, 1202–1207. [Google Scholar] [CrossRef]
  4. Bernat, J.; Gajewski, P.; Kapela, R.; Marcinkowska, A.; Superczyńska, P. Design, Fabrication and Analysis of Magnetorheological Soft Gripper. Sensors 2022, 22, 2757. [Google Scholar] [CrossRef] [PubMed]
  5. Lee, W.; Nam, J.; Kim, J.; Jung, E.; Kim, N.; Jang, G. Steering, Tunneling, and Stent Delivery of a Multifunctional Magnetic Catheter Robot to Treat Occlusive Vascular Disease. IEEE Trans. Ind. Electron. 2020, 68, 391–400. [Google Scholar] [CrossRef]
  6. Kim, M.C.; Kim, E.S.; Park, J.O.; Choi, E.; Kim, C.S. Robotic Localization Based on Planar Cable Robot and Hall Sensor Array Applied to Magnetic Capsule Endoscope. Sensors 2020, 20, 5728. [Google Scholar] [CrossRef] [PubMed]
  7. Ji, D.M.; Jung, W.S.; Kim, S.H. Wireless Manipulation Mechanism and Analysis for Actively Assistive Pinch Movements. Sensors 2021, 21, 6216. [Google Scholar] [CrossRef]
  8. Mathieu, J.B.; Beaudoin, G.; Martel, S. Method of Propulsion of a Ferromagnetic Core in the Cardiovascular System through Magnetic Gradients Generated by an MRI System. IEEE Trans. Biomed. Eng. 2006, 53, 292–299. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  9. Li, N.; Jiang, Y.; Plantefève, R.; Michaud, F.; Nosrati, Z.; Tremblay, C.; Saatchi, K.; Häfeli, U.O.; Kadoury, S.; Moran, G.; et al. Magnetic Resonance Navigation for Targeted Embolization in a Two-Level Bifurcation Phantom. Ann. Biomed. Eng. 2019, 47, 2402–2415. [Google Scholar] [CrossRef]
  10. Griese, F.; Knopp, T.; Gruettner, C.; Thieben, F.; Müller, K.; Loges, S.; Ludewig, P.; Gdaniec, N. Simultaneous Magnetic Particle Imaging and Navigation of Large Superparamagnetic Nanoparticles in Bifurcation Flow Experiments. J. Magn. Magn. Mater. 2020, 498, 166206. [Google Scholar] [CrossRef] [Green Version]
  11. Karvelas, E.G.; Lampropoulos, N.K.; Karakasidis, T.E.; Sarris, I.E. Computational Study of the Optimum Gradient Magnetic Field for the Navigation of the Spherical Particles in the Process of Cleaning the Water from Heavy Metals. Procedia Eng. 2016, 162, 77–82. [Google Scholar] [CrossRef] [Green Version]
  12. Zhang, X.; Le, T.-A.; Yoon, J. Development of a Real Time Imaging-Based Guidance System of Magnetic Nanoparticles for Targeted Drug Delivery. J. Magn. Magn. Mater. 2017, 427, 345–351. [Google Scholar] [CrossRef]
  13. Hoshiar, A.K.; Le, T.-A.; Valdastri, P.; Yoon, J. Swarm of Magnetic Nanoparticles Steering in Multi-Bifurcation Vessels under Fluid Flow. J. Micro-Bio Robot. 2020, 16, 137–145. [Google Scholar] [CrossRef]
  14. Kummer, M.P.; Abbott, J.J.; Kratochvil, B.E.; Borer, R.; Sengul, A.; Nelson, B.J. OctoMag: An Electromagnetic System for 5-DOF Wireless Micromanipulation. IEEE Trans. Robot. 2010, 26, 1006–1017. [Google Scholar] [CrossRef]
  15. Diller, E.; Giltinan, J.; Lum, G.Z.; Ye, Z.; Sitti, M. Six-Degree-of-Freedom Magnetic Actuation for Wireless Microrobotics. Int. J. Robot. Res. 2016, 35, 114–128. [Google Scholar] [CrossRef]
  16. Nam, J.; Lee, W.; Jung, E.; Jang, G. Magnetic Navigation System Utilizing a Closed Magnetic Circuit to Maximize Magnetic Field and a Mapping Method to Precisely Control Magnetic Field in Real Time. IEEE Trans. Ind. Electron. 2017, 65, 5673–5681. [Google Scholar] [CrossRef]
  17. Erni, S.; Schürle, S.; Fakhraee, A.; Kratochvil, B.E.; Nelson, B.J. Comparison, Optimization, and Limitations of Magnetic Manipulation Systems. J. Micro-Bio Robot. 2013, 8, 107–120. [Google Scholar] [CrossRef]
  18. Mahoney, A.W.; Sarrazin, J.C.; Bamberg, E.; Abbott, J.J. Velocity Control with Gravity Compensation for Magnetic Helical Microswimmers. Adv. Robot. 2011, 25, 1007–1028. [Google Scholar] [CrossRef] [Green Version]
  19. Jeong, S.; Choi, H.; Cha, K.; Li, J.; Park, J.; Park, S. Enhanced Locomotive and Drilling Microrobot Using Precessional and Gradient Magnetic Field. Sens. Actuators A 2011, 171, 429–435. [Google Scholar] [CrossRef]
  20. Bell, D.J.; Leutenegger, S.; Hammar, K.M.; Dong, L.X.; Nelson, B.J. Flagella-Like Propulsion for Microrobots Using a Nanocoil and a Rotating Rlectromagnetic Field. In Proceedings of the 2007 IEEE International Conference on Robotics and Automation, Rome, Italy, 10–14 April 2007; pp. 1128–1133. [Google Scholar] [CrossRef]
  21. Choi, H.; Cha, K.; Jeong, S.; Park, J.-O.; Park, S. 3-D Locomotive and Drilling Microrobot Using Novel Stationary EMA System. IEEE/ASME Trans. Mechatron. 2013, 18, 1221–1225. [Google Scholar] [CrossRef]
  22. Choi, H.; Choi, J.; Jeong, S.; Yu, C.; Park, J.; Park, S. Two-Dimensional Locomotion of a Microrobot with a Novel Stationary Electromagnetic Actuation System. Smart Mater. Struct. 2009, 18, 115017. [Google Scholar] [CrossRef]
  23. Arcese, L.; Fruchard, M.; Ferreira, A. Adaptive Controller and Observer for a Magnetic Microrobot. IEEE Trans. Robot. 2013, 29, 1060–1067. [Google Scholar] [CrossRef] [Green Version]
  24. Yesin, K.B.; Vollmers, K.; Nelson, B.J. Modeling and Control of Untethered Biomicrorobots in a Fluidic Environment Using Electromagnetic Fields. Int. J. Robot. Res. 2006, 25, 527–536. [Google Scholar] [CrossRef]
  25. Yu, C.; Kim, J.; Choi, H.; Choi, J.; Jeong, S.; Cha, K.; Park, J.; Park, S. Novel Electromagnetic Actuation System for Three-Dimensional Locomotion and Drilling of Intravascular Microrobot. Sens. Actuators A 2010, 161, 297–304. [Google Scholar] [CrossRef]
  26. Jeon, S.; Jang, G.; Choi, H.; Park, S. Magnetic Navigation System with Gradient and Uniform Saddle Coils for the Wireless Manipulation of Micro-robots in Human Blood Vessels. IEEE Trans. Magn. 2010, 46, 1943–1946. [Google Scholar] [CrossRef]
  27. Jeon, S.M.; Jang, G.H.; Choi, H.C.; Park, S.H.; Park, J.O. Magnetic Navigation System for the Precise Helical and Translational Motions of a Microrobot in Human Blood Vessels. J. Appl. Phys. 2012, 111, 07E702. [Google Scholar] [CrossRef] [Green Version]
  28. Jeon, S.M.; Jang, G.H.; Choi, H.C.; Park, S.H.; Park, J.O. Utilization of Magnetic Gradients in a Magnetic Navigation System for the Translational Motion of a Micro-robot in Human Blood Vessels. IEEE Trans. Magn. 2011, 47, 2403–2406. [Google Scholar] [CrossRef]
  29. Choi, H.; Cha, K.; Choi, J.; Jeong, S.; Jeon, S.; Jang, G.; Park, J.; Park, S. EMA System with Gradient and Uniform Saddle Coils for 3D Locomotion of Microrobot. Sens. Actuators A 2010, 163, 410–417. [Google Scholar] [CrossRef]
  30. Go, G.; Choi, H.; Jeong, S.; Lee, C.; Ko, S.Y.; Park, J.-O.; Park, S. Electromagnetic Navigation System Using Simple Coil Structure (4 Coils) for 3-D Locomotive Microrobot. IEEE Trans. Magn. 2015, 51, 1–4. [Google Scholar] [CrossRef]
  31. Son, D.; Gilbert, H.; Sitti, M. Magnetically Actuated Soft Capsule Endoscope for Fine-Needle Biopsy. Soft Robot. 2019, 7, 10–21. [Google Scholar] [CrossRef]
  32. Abbott, J.J.; Peyer, K.E.; Lagomarsino, M.C.; Zhang, L.; Dong, L.; Kaliakatsos, I.K.; Nelson, B.J. How Should Microrobots Swim? Int. J. Robot. Res. 2009, 28, 1434–1447. [Google Scholar] [CrossRef]
Figure 1. The proposed novel arrangement of five pairs of coils to enable the MNS to generate the maximum 3D rotating magnetic field: (a) three pairs of coils (HC, USCy, and USCz) generate the uniform magnetic field; (b) two pairs of coils (MC and GSC) generate the uniform magnetic field gradient.
Figure 1. The proposed novel arrangement of five pairs of coils to enable the MNS to generate the maximum 3D rotating magnetic field: (a) three pairs of coils (HC, USCy, and USCz) generate the uniform magnetic field; (b) two pairs of coils (MC and GSC) generate the uniform magnetic field gradient.
Sensors 22 05603 g001
Figure 2. The proposed arrangement of the pipes (frame) and five pairs of coils: (a) HC and MC are installed outside the outer pipe, and USCz1 is installed inside the outer pipe; (b) USCy and GSC are installed outside the inner pipe, and USCz2 is installed inside the inner pipe; (c) the five pairs of coils integrated with the two pipes.
Figure 2. The proposed arrangement of the pipes (frame) and five pairs of coils: (a) HC and MC are installed outside the outer pipe, and USCz1 is installed inside the outer pipe; (b) USCy and GSC are installed outside the inner pipe, and USCz2 is installed inside the inner pipe; (c) the five pairs of coils integrated with the two pipes.
Sensors 22 05603 g002
Figure 3. The design variables and parameters of the k-th coil. This cross-section shows the case in which the coil is attached to the outside of the pipe.
Figure 3. The design variables and parameters of the k-th coil. This cross-section shows the case in which the coil is attached to the outside of the pipe.
Sensors 22 05603 g003
Figure 4. Output range of the power supply unit (3001iX by California Instruments).
Figure 4. Output range of the power supply unit (3001iX by California Instruments).
Sensors 22 05603 g004
Figure 5. Optimization algorithm for the MNS with the given constraints.
Figure 5. Optimization algorithm for the MNS with the given constraints.
Sensors 22 05603 g005
Figure 6. The constructed MNS and experimental setup to actuate a magnetic robot.
Figure 6. The constructed MNS and experimental setup to actuate a magnetic robot.
Sensors 22 05603 g006
Figure 7. Calculated and measured magnetic field distribution along the axis of each coil when the current of 1 A was applied.
Figure 7. Calculated and measured magnetic field distribution along the axis of each coil when the current of 1 A was applied.
Sensors 22 05603 g007
Figure 8. Temperature rise of coils for elapsed time.
Figure 8. Temperature rise of coils for elapsed time.
Sensors 22 05603 g008
Figure 9. Calculated and measured maximum magnetic field or field gradient of each coil with a variation of frequency.
Figure 9. Calculated and measured maximum magnetic field or field gradient of each coil with a variation of frequency.
Sensors 22 05603 g009
Figure 10. Performance test using a rotating magnetic field: (a) helical robot made of a diametrically magnetized cylindrical magnet (N52 grade) for the demonstration; (b) unclogging motion of the helical robot inside the optimized and conventional MNS. The optimized MNS and conventional MNS generated a 3D rotating magnetic field at 17 Hz and 12 Hz, respectively.
Figure 10. Performance test using a rotating magnetic field: (a) helical robot made of a diametrically magnetized cylindrical magnet (N52 grade) for the demonstration; (b) unclogging motion of the helical robot inside the optimized and conventional MNS. The optimized MNS and conventional MNS generated a 3D rotating magnetic field at 17 Hz and 12 Hz, respectively.
Sensors 22 05603 g010
Table 1. Major variables of the optimized MNS.
Table 1. Major variables of the optimized MNS.
VariablesHCUSCyUSCz1USCz2MCGSC
Radius   of   the   coil   ( r k ) (mm)249.8173.1220.0137.0248.2173.2
Turns   of   the   wire   ( N k ) (turns)44837040269413479
Thickness   of   the   wire   ( t k ) (mm)1.861.891.591.791.65
Resistance   of   the   coil   ( R k ) (Ω)8.628.628.648.628.63
Max .   magnetic   field   ( B k , m a x )   ( mT )   or   field   gradient   ( G k , m a x ) (mT/m)25.1625.1525.2684.26102.99
Table 2. Measured major values of the MNS.
Table 2. Measured major values of the MNS.
VariablesHCUSCyUSCz1USCz2MCGSC
Radius   of   the   coil   ( r k ) (mm)262.3175.0220.0137.1260.7170.0
Resistance   of   the   coil   ( R k ) (Ω)10.29.89.010.19.0
Inductance   of   the   coil   ( L k ) (mH)405.1364.6170.2303.0248.5
Max .   magnetic   field   ( B k , m a x )   ( mT )   or   field   gradient   ( G k , m a x ) (mT/m)20.0023.1125.0270.04106.67
Table 3. Correction coefficients of the MNS.
Table 3. Correction coefficients of the MNS.
HCUSCyUSCzMCGSC
Correction coefficients0.90450.95401.03130.86231.0739
Table 4. Comparison of major values between the optimized and conventional MNS.
Table 4. Comparison of major values between the optimized and conventional MNS.
VariablesConventional MNSOptimized MNSDifferences
[%]
Max magnetic
field ( B k , m a x ) (mT)
HC14.1820.0041
USCy21.6923.117
USCz14.0425.0278
Max magnetic field
gradient ( G k , m a x ) (mT/m)
MC121.370.04−42
GSC83.70106.6727
Inductance ( L k ) (mH)HC344.5405.118
USCy201.3364.181
USCz394.3170.2−57
MC859.9303.0−65
GSC84.60248.5194
Operating time limit
with max. power (min)
HC760757
USCy1540167
USCz101660
MC1435150
GSC217750
Diameter of an MNS (mm)Outer47052612
Inner2352402
Max power for each coil (W)21002100-
Table 5. Comparison of practical values between the optimized and conventional MNS.
Table 5. Comparison of practical values between the optimized and conventional MNS.
VariablesConventional MNSOptimized MNSDifferences
(%)
Max. 3D rotating magnetic field (mT)14.0420.0042
Min. magnetic field gradient (mT/m)83.7070.04−16
Max. inductance (mH)859.9405.1−53
Min. operating time limit
of the coils with max. power (min.)
216700
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Kim, S.; Cho, M.; Im, S.; Yun, J.; Nam, J. Electrical Optimization Method Based on a Novel Arrangement of the Magnetic Navigation System with Gradient and Uniform Saddle Coils. Sensors 2022, 22, 5603. https://doi.org/10.3390/s22155603

AMA Style

Kim S, Cho M, Im S, Yun J, Nam J. Electrical Optimization Method Based on a Novel Arrangement of the Magnetic Navigation System with Gradient and Uniform Saddle Coils. Sensors. 2022; 22(15):5603. https://doi.org/10.3390/s22155603

Chicago/Turabian Style

Kim, Sungjun, Mingyu Cho, Seyeong Im, Joongho Yun, and Jaekwang Nam. 2022. "Electrical Optimization Method Based on a Novel Arrangement of the Magnetic Navigation System with Gradient and Uniform Saddle Coils" Sensors 22, no. 15: 5603. https://doi.org/10.3390/s22155603

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop