Comparison of Multivariate Regression Models Based on Water- and Carbohydrate-Related Spectral Regions in the Near-Infrared for Aqueous Solutions of Glucose

Institute of Analytical Chemistry and Radiochemistry, CCB-Center for Chemistry and Biomedicine, Innrain 80/82, 6020 Innsbruck, Austria
*
Author to whom correspondence should be addressed.
Molecules 2019, 24(20), 3696; https://doi.org/10.3390/molecules24203696
Received: 3 September 2019 / Revised: 7 October 2019 / Accepted: 10 October 2019 / Published: 15 October 2019
The predictive power of the two major water bands centered at 6900 cm - 1 and 5200 cm - 1 in the near-infrared (NIR) region was compared to carbohydrate-related spectral areas located in the first overtone (around 6000 cm - 1 ) and combination (around 4500 cm - 1 ) region using glucose in aqueous solutions as a model substance. For the purpose of optimal coverage of stronger as well as weaker absorbing NIR regions, cells with three different declared optical pathlengths were employed. The sample set consisted of multiple separately prepared batches in the range of 50–200 mmol/L. Moreover, the samples were divided into a calibration set for the construction of the partial least squares regression (PLS-R) models and a test set for the validation process with independent samples. The first overtone and combination region showed relative prediction errors between 0.4–1.6% with only one PLS-R factor required. On the other hand, the errors for the water bands were found between 1.6–8.3% and up to three PLS-R factors required. The best PLS-R models resulted from the cell with 1 mm optical pathlength. In general, the results suggested that the carbohydrate-related regions in the first overtone and combination region should be preferred over the regions of the two dominant water bands. View Full-Text
Keywords: FT-NIR spectroscopy; PLS-R; water; glucose; test set validation; RMSEP FT-NIR spectroscopy; PLS-R; water; glucose; test set validation; RMSEP
Show Figures

Figure 1

MDPI and ACS Style

Beganović, A.; Moll, V.; Huck, C.W. Comparison of Multivariate Regression Models Based on Water- and Carbohydrate-Related Spectral Regions in the Near-Infrared for Aqueous Solutions of Glucose. Molecules 2019, 24, 3696.

Show more citation formats Show less citations formats
Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Article Access Map by Country/Region

1
Search more from Scilit
 
Search
Back to TopTop