Intra-Rater and Inter-Rater Reliability of the Kinvent Hand-Held Dynamometer in Young Adults †
Abstract
:1. Introduction
2. Materials and Methods
2.1. Study Design
2.2. Participants and Setting
2.3. Procedure
2.4. Statistics
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Chamorro, C.; Armijo-Olivo, S.; De la Fuente, C.; Fuentes, J.; Chirosa, L.J. Absolute reliability and concurrent validity of hand held dynamometry and isokinetic dynamometry in the hip, knee and ankle joint: Systematic review and meta-analysis. Open Med. 2017, 12, 359–375. [Google Scholar] [CrossRef] [PubMed]
- Stark, T.; Walker, B.; Phillips, J.K.; Fejer, R.; Beck, R. Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: A systematic review. PMR 2011, 3, 472–479. [Google Scholar] [CrossRef] [PubMed]
- Clarke, M.; Ni Mhuircheartaigh, D.; Walsh, G.; Walsh, J.; Meldrum, D. Intra-tester and inter-tester reliability of the MicroFET 3 hand-held dynamometer. Physiother. Pract. Res. 2011, 32, 13–18. [Google Scholar] [CrossRef]
- Contreras-Díaz, G.; Chirosa-Ríos, L.J.; Chirosa-Ríos, I.; Intelangelo, L.; Jerez-Mayorga, D.; Martinez-Garcia, D. Reliability of Isokinetic Hip Flexor and Extensor Strength Measurements in Healthy Subjects and Athletes: A Systematic Review and Meta-Analysis. Int. J. Environ. Res. Public Health 2021, 18, 11326. [Google Scholar] [CrossRef] [PubMed]
- Sørensen, L.; Oestergaard, L.G.; van Tulder, M.; Petersen, A.K. Measurement Properties of Isokinetic Dynamometry for Assessment of Shoulder Muscle Strength: A Systematic Review. Arch. Phys. Med. Rehabil. 2021, 102, 510–520. [Google Scholar] [CrossRef] [PubMed]
- Hirano, M.; Katoh, M.; Gomi, M.; Arai, S. Validity and reliability of isometric knee extension muscle strength measurements using a belt-stabilized hand-held dynamometer: A comparison with the measurement using an isokinetic dynamometer in a sitting posture. J. Phys. Ther. Sci. 2020, 32, 120–124. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Olds, M.; McLaine, S.; Magni, N. Validity and Reliability of the Kinvent Handheld Dynamometer in the Athletic Shoulder Test. J. Sport Rehabil. 2023, 1, 1–9. [Google Scholar] [CrossRef] [PubMed]
- Mentiplay, B.F.; Perraton, L.G.; Bower, K.J.; Adair, B.; Pua, Y.-H.; Williams, G.P.; McGaw, R.; Clark, R.A. Assessment of lower limb muscle strength and power using hand-held and fixed dynamometry: A reliability and validity study. PLoS ONE 2015, 10, e0140822. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Movement (kgf) | Rater 1—Day1 (Mean ± SD) | Rater 1—Day2 (Mean ± SD) | Intra-Rater Reliability | Rater 2—Day2 (Mean ± SD) | Inter-Rater Reliability | ||
---|---|---|---|---|---|---|---|
ICC (95% CI) | p | ICC (95% CI) | p | ||||
Upper limbs | |||||||
Shoulder Flex. | 19.13 ± 4.58 | 16.93 ± 5.90 | 0.707(0.273–0.905) | 0.003 | 16.00 ± 6.13 | 0.922 (0.273–0.905) | <0.001 |
Shoulder Abd. | 18.80 ± 4.86 | 16.98 ± 6.42 | 0.850 (0.575–0.945) | <0.001 | 16.14 ± 5.65 | 0.917 (0.739–0.975) | <0.001 |
Elbow Flex. | 26.69 ± 8.98 | 25.38 ± 10.03 | 0.951 (0.854–0.985) | <0.001 | 25.30 ± 9.68 | 0.985 (0.948–0.996) | <0.001 |
Elbow Ext. | 17.77 ± 6.60 | 16.49 ± 6.69 | 0.912 (0.734–0.973) | <0.001 | 16.39 ± 6.69 | 0.961 (0.872–0.989) | <0.001 |
Wrist Flex. | 17.88 ± 6.28 | 17.51 ± 7.17 | 0.854 (0.586–0.955) | <0.001 | 17.59 ± 6.46 | 0.894 (0.675–0.968) | <0.001 |
Wrist Ext. | 15.73 ± 5.24 | 14.84 ± 5.55 | 0.874 (0.636–0.962) | <0.001 | 15.65 ± 6.70 | 0.929 (0.773–0.979) | <0.001 |
Lower limbs | |||||||
Hip Flex. | 32.27 ± 9.68 | 31.94 ± 7.44 | 0.796 (0.453–0.936) | <0.001 | 32.08 ± 8.97 | 0.900 (0.692–0.970) | <0.001 |
Hip Abd. | 35.15 ± 10.26 | 34.00 ± 8.75 | 0.840 (0.552–0.951) | <0.001 | 33.70 ± 9.36 | 0.916 (0.734–0.975) | <0.001 |
Knee Flex. | 18.90 ± 5.19 | 18.50 ± 4.95 | 0.939 (0.811–0.982) | <0.001 | 18.19 ± 4.87 | 0.863 (0.594–0.959) | <0.001 |
Knee Ext. | 57.74 ± 17.68 | 59.06 ± 15.67 | 0.909 (0.726–0.972) | <0.001 | 54.02 ± 14.78 | 0.876 (0.604–0.965) | <0.001 |
Ankle Dorsiflex. | 26.20 ± 7.08 | 23.86 ± 5.76 | 0.707 (0.392–0.926) | <0.001 | 22.85 ± 3.97 | 0.722 (0.281–0.911) | 0.03 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
de Almeida, M.B.; Oliveira, C.; Ornelas, G.; Soares, T.; Souto, J.; Póvoa, A.R.; Ferreira, L.M.A.; Ricci-Vitor, A.L. Intra-Rater and Inter-Rater Reliability of the Kinvent Hand-Held Dynamometer in Young Adults. Med. Sci. Forum 2023, 22, 12. https://doi.org/10.3390/msf2023022012
de Almeida MB, Oliveira C, Ornelas G, Soares T, Souto J, Póvoa AR, Ferreira LMA, Ricci-Vitor AL. Intra-Rater and Inter-Rater Reliability of the Kinvent Hand-Held Dynamometer in Young Adults. Medical Sciences Forum. 2023; 22(1):12. https://doi.org/10.3390/msf2023022012
Chicago/Turabian Stylede Almeida, Manuel Barbosa, Carlos Oliveira, Guilherme Ornelas, Tiago Soares, Jean Souto, Ana Rita Póvoa, Luciano Maia Alves Ferreira, and Ana Laura Ricci-Vitor. 2023. "Intra-Rater and Inter-Rater Reliability of the Kinvent Hand-Held Dynamometer in Young Adults" Medical Sciences Forum 22, no. 1: 12. https://doi.org/10.3390/msf2023022012
APA Stylede Almeida, M. B., Oliveira, C., Ornelas, G., Soares, T., Souto, J., Póvoa, A. R., Ferreira, L. M. A., & Ricci-Vitor, A. L. (2023). Intra-Rater and Inter-Rater Reliability of the Kinvent Hand-Held Dynamometer in Young Adults. Medical Sciences Forum, 22(1), 12. https://doi.org/10.3390/msf2023022012