Novel Portable Device for Post Mortem Interval Estimation Using Vitreous Humor Analysis

by
Louise Lijcklama à Nijeholt
1,2,3,*,†,
Michael Fleermann
3,†,
Micky Breukers
3,
Jaap Knotter
1,2,
Steven Staal
4 and
Brigitte Bruijns
1,2,5
1
Technologies for Criminal Investigations, Saxion University of Applied Sciences, M.H. Tromplaan 28, 7513 AB Enschede, The Netherlands
2
Police Academy, Arnhemseweg 348, 7334 AC Apeldoorn, The Netherlands
3
Faculty Life Science Engineering and Design (LED), Saxion University of Applied Sciences, M.H. Tromplaan 28, 7513 AB Enschede, The Netherlands
4
Fisic BV, High Tech Factory, De Veldmaat 17, 7522 NM Enschede, The Netherlands
5
CLHC, Amsterdam Center for Forensic Science and Medicine, University of Amsterdam, 1090 GD Amsterdam, The Netherlands
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Forensic Sci. 2025, 5(2), 18; https://doi.org/10.3390/forensicsci5020018
Submission received: 29 October 2024 / Revised: 20 February 2025 / Accepted: 13 March 2025 / Published: 23 April 2025

Abstract

Background: Accurately determining the post mortem interval (PMI) is critical in forensic investigations to reconstruct the sequence of events leading up to and following death. Traditional methods (e.g., observing livor mortis, rigor mortis, and using temperature-based models) have limitations due to their empirical nature and susceptibility to environmental variables. The objective of this study was to assess the potential of a novel portable device, the Fisic Medimate™ system, for estimating PMI through the analysis of the potassium/sodium (K+/Na+) ratio in vitreous humor. Method: Vitreous humor samples were collected from pigs at various intervals up to 44.5 hours post-mortem. These samples were analyzed using the Fisic Medimate™ system to determine the K+/Na+ ratio. The analysis was conducted at different time points to establish a relationship between the K+/Na+ ratio and PMI. Results: The results indicated a log-linear relationship between the K+/Na+ ratio and PMI for periods up to 31 hours, with normal residuals. This relationship was observed across all samples, suggesting that the method provides reliable and consistent data. Conclusions: This method offers a rapid and portable solution for PMI determination, providing valuable data within minutes directly at the crime scene. While these findings suggest potential for on-site forensic applications, further validation under real-world conditions is required to confirm its broader applicability.
Keywords: post mortem interval; forensic investigation; vitreous humor; potassium/sodium ratio; portable device; log-linear relationship post mortem interval; forensic investigation; vitreous humor; potassium/sodium ratio; portable device; log-linear relationship

Share and Cite

MDPI and ACS Style

Lijcklama à Nijeholt, L.; Fleermann, M.; Breukers, M.; Knotter, J.; Staal, S.; Bruijns, B. Novel Portable Device for Post Mortem Interval Estimation Using Vitreous Humor Analysis. Forensic Sci. 2025, 5, 18. https://doi.org/10.3390/forensicsci5020018

AMA Style

Lijcklama à Nijeholt L, Fleermann M, Breukers M, Knotter J, Staal S, Bruijns B. Novel Portable Device for Post Mortem Interval Estimation Using Vitreous Humor Analysis. Forensic Sciences. 2025; 5(2):18. https://doi.org/10.3390/forensicsci5020018

Chicago/Turabian Style

Lijcklama à Nijeholt, Louise, Michael Fleermann, Micky Breukers, Jaap Knotter, Steven Staal, and Brigitte Bruijns. 2025. "Novel Portable Device for Post Mortem Interval Estimation Using Vitreous Humor Analysis" Forensic Sciences 5, no. 2: 18. https://doi.org/10.3390/forensicsci5020018

APA Style

Lijcklama à Nijeholt, L., Fleermann, M., Breukers, M., Knotter, J., Staal, S., & Bruijns, B. (2025). Novel Portable Device for Post Mortem Interval Estimation Using Vitreous Humor Analysis. Forensic Sciences, 5(2), 18. https://doi.org/10.3390/forensicsci5020018

Article Metrics

Back to TopTop