Abstract

Fundamental Analytical Parameters of a Glucose Biosensor Based on TiO2 Nanotube Arrays and Chitosan as Immobilization Matrix †

by
Margalida Artigues-Cladera
,
Jordi Abella
and
Sergi Colominas
*
Electrochemical Methods Laboratory, Department of Analytical and Applied Chemistry, ETS Institut Químic de Sarrià, Universitat Ramon Llull, 08017 Barcelona, Spain
*
Author to whom correspondence should be addressed.
Presented at the 5th International Symposium on Sensor Science (I3S 2017), Barcelona, Spain, 27–29 September 2017.
Proceedings 2017, 1(8), 832; https://doi.org/10.3390/proceedings1080832
Published: 20 November 2017
Numerous analytical techniques are used for glucose determination, such as spectrophotometry, amperometry, HPLC, polarimetry and capillary electrophoresis. However, only electrochemical biosensors based on the use of glucose oxidase (GOx) have been able to combine the analytical power of electrochemical techniques with the specificity of biological recognition processes. Commonly, this combination of biological components with electrodes is used to produce low-cost, easy-to-use and compact devices for glucose quantification. Since the development of the first glucose biosensor, great efforts have been made in order to improve the response performances of these enzyme electrodes.
In the present work, we present the fundamental analytical parameters of an amperometric glucose biosensor based on GOx immobilization using a polymeric hydrogel (Chitosan) onto highly ordered titanium dioxide nanotube arrays (TiO2NTAs). The biosensor optimal working potential was evaluated and then fixed at −0.4 V. After that, the fundamental analytical parameters of the biosensor (linear range, limit of detection, sensitivity) were determined, as well as its storage stability. This biosensor showed a linear range from 0.3 mM to 1.5 mM, low limit of detection (0.07 mM) and high sensitivity (5.46 mA·mM−1). Furthermore, its lifetime was evaluated. After 30 days, the biosensor retained 85% of its initial current response.

Share and Cite

MDPI and ACS Style

Artigues-Cladera, M.; Abella, J.; Colominas, S. Fundamental Analytical Parameters of a Glucose Biosensor Based on TiO2 Nanotube Arrays and Chitosan as Immobilization Matrix. Proceedings 2017, 1, 832. https://doi.org/10.3390/proceedings1080832

AMA Style

Artigues-Cladera M, Abella J, Colominas S. Fundamental Analytical Parameters of a Glucose Biosensor Based on TiO2 Nanotube Arrays and Chitosan as Immobilization Matrix. Proceedings. 2017; 1(8):832. https://doi.org/10.3390/proceedings1080832

Chicago/Turabian Style

Artigues-Cladera, Margalida, Jordi Abella, and Sergi Colominas. 2017. "Fundamental Analytical Parameters of a Glucose Biosensor Based on TiO2 Nanotube Arrays and Chitosan as Immobilization Matrix" Proceedings 1, no. 8: 832. https://doi.org/10.3390/proceedings1080832

Article Metrics

Back to TopTop