Next Article in Journal
Atherosclerotic Cardiovascular Disease: Risk Assessment, Prevention and Treatment Strategies
Next Article in Special Issue
Comparative Analysis of Temperature Rise between Convective Heat Transfer Method and Computational Fluid Dynamics Method in an Anatomy-Based Left Atrium Model during Pulsed Field Ablation: A Computational Study
Previous Article in Journal
Novel Hybrid Treatment for Pulmonary Arterial Hypertension with or without Eisenmenger Syndrome: Double Lung Transplantation with Simultaneous Endovascular or Classic Surgical Closure of the Patent Ductus Arteriosus (PDA)
Previous Article in Special Issue
Healthcare Utilization and Outcomes in Atrial Fibrillation Patients Treated by Drug Therapy versus a Catheter Ablation Strategy—A Middle European Propensity Score Matched Cohort Study
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation

1
School of Engineering, University of Galway, H91 TK33 Galway, Ireland
2
Translational Medical Device Lab, University of Galway, H91 YR71 Galway, Ireland
3
BioMIT, Department of Electronic Engineering, Universitat Politècnica de València, 46022 Valencia, Spain
4
AtriAN Medical Limited, Unit 204, University of Galway Business Innovation Centre, Upper Newcastle, H91 W60E Galway, Ireland
*
Author to whom correspondence should be addressed.
J. Cardiovasc. Dev. Dis. 2022, 9(12), 458; https://doi.org/10.3390/jcdd9120458
Submission received: 14 November 2022 / Revised: 7 December 2022 / Accepted: 12 December 2022 / Published: 14 December 2022
(This article belongs to the Special Issue Catheter Ablation of Cardiac Arrhythmias: Practices and Outcomes)

Abstract

Background: Pulsed Electric Field (PEF) ablation has been recently proposed to ablate cardiac ganglionic plexi (GP) aimed to treat atrial fibrillation. The effect of metal intracoronary stents in the vicinity of the ablation electrode has not been yet assessed. Methods: A 2D numerical model was developed accounting for the different tissues involved in PEF ablation with an irrigated ablation device. A coronary artery (with and without a metal intracoronary stent) was considered near the ablation source (0.25 and 1 mm separation). The 1000 V/cm threshold was used to estimate the ‘PEF-zone’. Results: The presence of the coronary artery (with or without stent) distorts the E-field distribution, creating hot spots (higher E-field values) in the front and rear of the artery, and cold spots (lower E-field values) on the sides of the artery. The value of the E-field inside the coronary artery is very low (~200 V/cm), and almost zero with a metal stent. Despite this distortion, the PEF-zone contour is almost identical with and without artery/stent, remaining almost completely confined within the fat layer in any case. The mentioned hot spots of E-field translate into a moderate temperature increase (<48 °C) in the area between the artery and electrode. These thermal side effects are similar for pulse intervals of 10 and 100 μs. Conclusions: The presence of a metal intracoronary stent near the ablation device during PEF ablation simply ‘amplifies’ the E-field distortion already caused by the presence of the vessel. This distortion may involve moderate heating (<48 °C) in the tissue between the artery and ablation electrode without associated thermal damage.
Keywords: cardiac arrhythmia; computer modelling; coronary artery; epicardial ablation; metal stent; pulsed field ablation cardiac arrhythmia; computer modelling; coronary artery; epicardial ablation; metal stent; pulsed field ablation

Share and Cite

MDPI and ACS Style

González-Suárez, A.; Pérez, J.J.; O’Brien, B.; Elahi, A. In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation. J. Cardiovasc. Dev. Dis. 2022, 9, 458. https://doi.org/10.3390/jcdd9120458

AMA Style

González-Suárez A, Pérez JJ, O’Brien B, Elahi A. In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation. Journal of Cardiovascular Development and Disease. 2022; 9(12):458. https://doi.org/10.3390/jcdd9120458

Chicago/Turabian Style

González-Suárez, Ana, Juan J. Pérez, Barry O’Brien, and Adnan Elahi. 2022. "In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation" Journal of Cardiovascular Development and Disease 9, no. 12: 458. https://doi.org/10.3390/jcdd9120458

APA Style

González-Suárez, A., Pérez, J. J., O’Brien, B., & Elahi, A. (2022). In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation. Journal of Cardiovascular Development and Disease, 9(12), 458. https://doi.org/10.3390/jcdd9120458

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop