The Impact of a New “Inverted Arch” Prosthetic Annuloplasty Ring on the Mitral Valve’s 3-D Motion: An Experimental Ex-Vivo Study

1
Department of Medical Direction of University Hospital of Novara, Corso Mazzini 18, 28100 Novara, Italy
2
Department of Cardiac Surgery, “Attikon” University Hospital, 12462 Athens, Greece
3
Department of Cardiac Surgery, University Hospital of Novara, 28100 Novara, Italy
4
Department of Experimental Surgery, University Hospital of Novara, 28100 Novara, Italy
5
Department of Physiology, University of Eastern Piedmont “Amedeo Avogadro”, 28100 Novara, Italy
*
Author to whom correspondence should be addressed.
Bioengineering 2019, 6(2), 31; https://doi.org/10.3390/bioengineering6020031
Received: 2 February 2019 / Revised: 31 March 2019 / Accepted: 4 April 2019 / Published: 8 April 2019
(This article belongs to the Special Issue Implantable Medical Devices)
  |  
PDF [2151 KB, uploaded 10 April 2019]
  |  

Abstract

This experimental study aimed to evaluate the ex-vivo three-dimensional (3-D) motion of the Inverted Arch Ring (IAR), an innovative new design concept for a flexible incomplete annuloplasty prosthesis with an incorporated stabilizing rigid arch that can be used in correcting mitral valve regurgitation. Twenty explanted porcine hearts were placed in a circulation simulation system. Ultrasonometry transducers implanted in the mitral annulus were used to measure the 3-D valvular motion during a simulated cardiac cycle. Annular distance measurements were recorded and compared in each heart before and after the implantation of the IAR prosthesis at pressures corresponding to mid-systole and mid-diastole. Distances measured in mid-systole and mid-diastole demonstrated no significant differences in annular motion or in valve areas either prior to or after IAR implantation. Therefore, the results of this study confirm the minimal effects exerted by the IAR prosthesis on the mitral valve’s 3-D motion during a simulated cardiac cycle. View Full-Text
Keywords: mitral valve annuloplasty; mitral valve repair; prosthetic mitral ring mitral valve annuloplasty; mitral valve repair; prosthetic mitral ring
Figures

Figure 1

This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited (CC BY 4.0).
SciFeed

Share & Cite This Article

MDPI and ACS Style

Caimmi, P.; Kapetanakis, E.; Beggino, C.; Vacca, G.; Grossini, E.; Stratica, F.; Sacco, R.; Capponi, A. The Impact of a New “Inverted Arch” Prosthetic Annuloplasty Ring on the Mitral Valve’s 3-D Motion: An Experimental Ex-Vivo Study. Bioengineering 2019, 6, 31.

Show more citation formats Show less citations formats

Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Related Articles

Article Metrics

Article Access Statistics

1

Comments

[Return to top]
Bioengineering EISSN 2306-5354 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert
Back to Top