Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre–Post Motion Analysis Study
Abstract
1. Introduction
2. Materials and Methods
2.1. Design and Setting
2.2. Participants
2.3. Study Procedure
2.4. Balance Tasks and Experimental Setup
2.5. Data Processing and Outcome Variables
2.6. Statistical Analysis
3. Results
3.1. Sit-to-Stand Task
3.2. Standing with Eyes Open and Closed
3.3. Tandem Stance and Feet-Together Tasks
4. Discussion
Limitations
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
| ESAR | Energy-Storing-and-Return |
| ML | Mediolateral |
| BBS | Berg Balance Scale |
| CoP | Center of Pressure |
| CoM | Center of Mass |
| AP | Anteroposterior |
| MDIST | Mean Distance |
| RDIST | Root Mean Square Distance |
| ROM | Range of Motion |
| TSB | Total Surface Bearing |
| PTB | Patellar Tendon Bearing |
| EMG | Electromyography |
References
- Petersen, B.A.; Sparto, P.J.; Fisher, L.E. Clinical measures of balance and gait cannot differentiate somatosensory impairments in people with lower-limb amputation. Gait Posture 2023, 99, 104–110. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Moisan, G.; Miramand, L.; Younesian, H.; Turcot, K. Balance control deficits in individuals with a transtibial amputation with and without visual input. Prosthet. Orthot. Int. 2022, 46, 134–139. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gaffney, B.M.; Murray, A.M.; Christiansen, C.L.; Davidson, B.S. Identification of trunk and pelvis movement compensations in patients with transtibial amputation using angular momentum separation. Gait Posture 2016, 45, 151–156. [Google Scholar] [CrossRef] [Scilit] [PubMed][Green Version]
- Miramand, L.; Moisan, G.; Richard, V.; McFadyen, B.J.; Turcot, K. Whole body movement strategies during sit-to-stand and stair ascent in individuals with a lower limb amputation: A systematic review. Clin. Biomech. 2022, 100, 105811. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kolářová, B.; Janura, M.; Svoboda, Z.; Kolář, P.; Tečová, D.; Elfmark, M. Postural Control Strategies and Balance-Related Factors in Individuals with Traumatic Transtibial Amputations. Sensors 2021, 21, 7284. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Poonsiri, J.; Dijkstra, P.U.; Geertzen, J.H.B. Fitting transtibial and transfemoral prostheses in persons with a severe flexion contracture: Problems and solutions—A systematic review. Disabil. Rehabil. 2022, 44, 3749–3759. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hunter, S.W.; Batchelor, F.; Hill, K.D.; Hill, A.M.; Mackintosh, S.; Payne, M. Risk Factors for Falls in People with a Lower Limb Amputation: A Systematic Review. PM&R 2017, 9, 170–180.e171. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sawers, A.; McDonald, C.L.; Hafner, B.J. A survey for characterizing details of fall events experienced by lower limb prosthesis users. PLoS ONE 2022, 17, e0272082. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Seth, M.; Horne, J.R.; Pohlig, R.T.; Sions, J.M. Pain, Balance-Confidence, Functional Mobility, and Reach Are Associated with Risk of Recurrent Falls Among Adults with Lower-Limb Amputation. Arch. Rehabil. Res. Clin. Transl. 2023, 5, 100309. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fleming, A.; Liu, W.; Huang, H.H. Neural prosthesis control restores near-normative neuromechanics in standing postural control. Sci. Robot. 2023, 8, eadf5758. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Vaca, M.; Stine, R.; Hammond, P., 2nd; Cavanaugh, M.; Major, M.J.; Gard, S.A. The Effect of Prosthetic Ankle Dorsiflexion Stiffness on Standing Balance and Gait Biomechanics in Individuals with Unilateral Transtibial Amputation. J. Prosthet. Orthot. 2022, 36, 110–123. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Koehler-McNicholas, S.R.; Savvas Slater, B.C.; Koester, K.; Nickel, E.A.; Ferguson, J.E.; Hansen, A.H. Bimodal ankle-foot prosthesis for enhanced standing stability. PLoS ONE 2018, 13, e0204512. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Olenšek, A.; Zadravec, M.; Burger, H.; Matjačić, Z. Dynamic balancing responses in unilateral transtibial amputees following outward-directed perturbations during slow treadmill walking differ considerably for amputated and non-amputated side. J. Neuroeng. Rehabil. 2021, 18, 123. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Murray, A.M.; Gaffney, B.M.; Davidson, B.S.; Christiansen, C.L. Biomechanical compensations of the trunk and lower extremities during stepping tasks after unilateral transtibial amputation. Clin. Biomech. 2017, 49, 64–71. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sawers, A.; Fatone, S. The relationship of hip strength to walking and balance performance in unilateral lower limb prosthesis users differs by amputation level. PM&R 2025, 17, 147–158. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sahoo, S.; Mohanty, R.K.; Mohapatra, A.K. A systematic review of energy storing dynamic response foot for prosthetic rehabilitation. Proc. Inst. Mech. Eng. Part H J. Eng. Med. 2024, 238, 1069–1090. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Major, M.J.; Kenney, L.P.; Twiste, M.; Howard, D. Stance phase mechanical characterization of transtibial prostheses distal to the socket: A review. J. Rehabil. Res. Dev. 2012, 49, 815–829. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Shell, C.E.; Segal, A.D.; Klute, G.K.; Neptune, R.R. The effects of prosthetic foot stiffness on transtibial amputee walking mechanics and balance control during turning. Clin. Biomech. 2017, 49, 56–63. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ernst, M.; Altenburg, B.; Schmalz, T. Characterizing adaptations of prosthetic feet in the frontal plane. Prosthet. Orthot. Int. 2020, 44, 225–233. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wernke, M.M.; Ficanha, E.M.; Thomas, Z.; Maitland, M.E.; Allyn, K.J.; Albury, A.; Colvin, J. Mechanical testing of frontal plane adaptability of commercially available prosthetic feet. J. Rehabil. Assist. Technol. Eng. 2022, 9, 20556683221123330. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hsieh, T.H.; Song, H.; Shu, T.; Qiao, J.; Yeon, S.H.; Carney, M.; Mooney, L.; Duval, J.F.; Herr, H. Design, Characterization, and Preliminary Assessment of a Two-Degree-of-Freedom Powered Ankle-Foot Prosthesis. Biomimetics 2024, 9, 76. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Maitland, M.E.; Imsdahl, S.I.; Fogelberg, D.J.; Allyn, K.J.; Cain, K.C.; Humbert, A.T.; Albury, A.; Ficanha, E.M.; Colvin, J.M.; Wernke, M.M. Motion Analysis of a Frontal Plane Adaptable Prosthetic Foot. J. Prosthet. Orthot. 2024, 36, 255–262. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Maitland, M.E.; Allyn, K.J.; Ficanha, E.M.; Colvin, J.M.; Wernke, M.M. Finite element simulation of frontal plane adaptation using full-foot, split-toe and cam-linkage designs in prosthetic feet. J. Prosthet. Orthot. 2022, 34, 14–21. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Major, M.J.; Fatone, S.; Roth, E.J. Validity and reliability of the Berg Balance Scale for community-dwelling persons with lower-limb amputation. Arch. Phys. Med. Rehabil. 2013, 94, 2194–2202. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hashimoto, H.; Kobayashi, T.; Gao, F.; Kataoka, M. A proper sequence of dynamic alignment in transtibial prosthesis: Insight through socket reaction moments. Sci. Rep. 2023, 13, 458. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Morgan, S.J.; McDonald, C.L.; Halsne, E.G.; Cheever, S.M.; Salem, R.; Kramer, P.A.; Hafner, B.J. Laboratory- and community-based health outcomes in people with transtibial amputation using crossover and energy-storing prosthetic feet: A randomized crossover trial. PLoS ONE 2018, 13, e0189652. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ray, S.F.; Wurdeman, S.R.; Takahashi, K.Z. Prosthetic energy return during walking increases after 3 weeks of adaptation to a new device. J. Neuroeng. Rehabil. 2018, 15, 6. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Berg, K.O.; Maki, B.E.; Williams, J.I.; Holliday, P.J.; Wood-Dauphinee, S.L. Clinical and laboratory measures of postural balance in an elderly population. Arch. Phys. Med. Rehabil. 1992, 73, 1073–1080. [Google Scholar] [PubMed]
- Chorin, F.; Rahmani, A.; Beaune, B.; Cornu, C. Determination of reliable force platform parameters and number of trial to evaluate sit-to-stand movement. Aging Clin. Exp. Res. 2015, 27, 473–482. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Fiedler, G.; Slavens, B.; Smith, R.O.; Briggs, D.; Hafner, B.J. Criterion and construct validity of prosthesis-integrated measurement of joint moment data in persons with transtibial amputation. J. Appl. Biomech. 2014, 30, 431–438. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Agrawal, V.; Gailey, R.; Gaunaurd, I.; Gailey, R., 3rd; O’Toole, C. Weight distribution symmetry during the sit-to-stand movement of unilateral transtibial amputees. Ergonomics 2011, 54, 656–664. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Teater, R.H.; Wolf, D.N.; McDonald, K.A.; Zelik, K.E. Unilateral transtibial prosthesis users load their intact limb more than their prosthetic limb during sit-to-stand, squatting, and lifting. Clin. Biomech. 2023, 108, 106041. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gailey, R.; Allen, K.; Castles, J.; Kucharik, J.; Roeder, M. Review of secondary physical conditions associated with lower-limb amputation and long-term prosthesis use. J. Rehabil. Res. Dev. 2008, 45, 15–29. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Palazzo, F.; Nardi, A.; Lamouchideli, N.; Caronti, A.; Alashram, A.; Padua, E.; Annino, G. The effect of age, sex and a firm-textured surface on postural control. Exp. Brain Res. 2021, 239, 2181–2191. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tisserand, R.; Plard, J.; Robert, T. Relative contributions of postural balance mechanisms reveal studying the CoP displacement alone may be incomplete for analysis of challenging standing postures. Gait Posture 2023, 101, 134–137. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Curtze, C.; Hof, A.L.; Postema, K.; Otten, B. The relative contributions of the prosthetic and sound limb to balance control in unilateral transtibial amputees. Gait Posture 2012, 36, 276–281. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Runciman, P.; Cockcroft, J.; Derman, W. A novel pivot ankle/foot prosthesis reduces sound side loading and risk for osteoarthritis: A pragmatic randomized controlled trial. Prosthet. Orthot. Int. 2022, 46, 258–266. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yang, J.R.; Yang, H.S.; Ahn, D.H.; Ahn, D.Y.; Sim, W.S.; Yang, H.E. Differences in Gait Patterns of Unilateral Transtibial Amputees With Two Types of Energy Storing Prosthetic Feet. Ann. Rehabil. Med. 2018, 42, 609–616. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Baharuddin, M.H.; Abdullah, N.N.A.A.; Kori, M.I.; Gan, H.-S.; Abdul Kadir, M.R.; Ramlee, M.H. Prosthetic leg socket design: New insight on different tibia length and distal bevelled angle using finite element analysis. Injury 2026, 57, 113472. [Google Scholar] [CrossRef] [Scilit] [PubMed]






| Participants | Height (cm) | Mass (kg) | Age (yrs.) | Sex | Feet | Socket | RTL (cm) | TSA (yrs.) | K-Level | Suspension | Side |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 179.3 | 83.1 | 46 | M | C-walk a | TSB | 9.1 | 27 | 3 | Suction | Rt |
| 2 | 171.2 | 85.5 | 61 | M | Vari-flex b | TSB | 10.1 | 11 | 3 | Pin | Lt |
| 3 | 172.7 | 74.6 | 74 | M | 1D35 a | TSB | 20.1 | 51 | 3 | Pin | Lt |
| 4 | 173.1 | 92.8 | 54 | M | HeraFT c | TSB | 16.2 | 7 | 3 | Suction | Lt |
| 5 | 167.5 | 62.7 | 74 | M | Trias a | TSB | 17.9 | 51 | 2 | Suction | Lt |
| 6 | 171.4 | 70.8 | 75 | M | C-walk a | TSB | 13.8 | 51 | 3 | Pin | Rt |
| 7 | 158.8 | 56.4 | 75 | M | C-walk a | TSB | 13.9 | 3 | 3 | Pin | Rt |
| 8 | 167.3 | 60.35 | 76 | M | 1H38 a | PTB | 19.6 | 53 | 3 | Cuff | Lt |
| 9 | 178.8 | 73.9 | 62 | M | Pro-flex b | TSB | 14.1 | 55 | 3 | Suction | Lt |
| 10 | 166.9 | 61.3 | 62 | M | 1D35 a | TSB | 16.7 | 39 | 2 | Suction | Rt |
| Mean | 170.7 | 72.15 | 65.9 | 15.15 | 34.8 |
| Variables | Sit-to-Stand | ||||
|---|---|---|---|---|---|
| CPF | IF | p | g (95% CI) | ||
| Range of CoP | AP (mm) | 7.51 ± 5.77 | 6.22 ± 4.35 | 0.445 | 0.31 (−0.28 to 0.88) |
| ML (mm) | 10.79 ± 6.68 | 9.65 ± 3.83 | 0.386 | 0.23 (−0.35 to 0.80) | |
| MDIST of CoP | AP (mm) | 8.83 ± 2.57 | 8.56 ± 1.75 | 0.575 | 0.08 (−0.49 to 0.64) |
| ML (mm) | 8.45 ± 2.79 | 6.87 ± 2.93 | 0.059 | 0.57 (−0.06 to 1.18) | |
| Resultant (mm) | 13.72 ± 2.68 | 12.17 ± 2.52 | 0.139 | 0.46 (−0.15 to 1.06) | |
| RDIST of CoP | AP (mm) | 10.45 ± 3.10 | 10.13 ± 2.38 | 0.878 | 0.07 (−0.50 to 0.64) |
| ML (mm) | 10.20 ± 3.59 | 8.29 ± 3.32 | 0.047 * | 0.68 (0.02 to 1.31) | |
| Resultant (mm) | 15.10 ± 3.09 | 13.39 ± 2.98 | 0.241 | 0.45 (−0.17 to 1.04) | |
| Range of CoM | Axis X (mm) | 345.51 ± 31.72 | 326.45 ± 47.40 | 0.008 ** | 0.77 (0.05 to 1.44) |
| Axis Y (mm) | 28.21 ± 9.07 | 24.44 ± 9.58 | 0.374 | 0.21 (−0.40 to 0.80) | |
| Axis Z (mm) | 302.89 ± 34.12 | 303.26 ± 33.05 | 0.260 | 0.43 (−0.21 to 1.04) | |
| Amputee ankle ROM | Axis X (°) | 4.88 ± 2.03 | 4.22 ± 1.71 | 0.285 | 0.31 (−0.28 to 0.88) |
| Axis Y (°) | 1.93 ± 1.01 | 2.23 ± 1.00 | 0.445 | −0.23 (−0.80 to 0.36) | |
| Sound ankle ROM | Axis X (°) | 11.67 ± 4.03 | 12.24 ± 3.28 | 0.241 | −0.20 (−0.77 to 0.38) |
| Axis Y (°) | 3.67 ± 0.79 | 3.87 ± 1.05 | 0.508 | −0.16 (−0.73 to 0.41) | |
| Variables | Eyes Open | Eyes Closed | |||||||
|---|---|---|---|---|---|---|---|---|---|
| CPF | IF | p | g (95% CI) | CPF | IF | p | g (95% CI) | ||
| Range of CoP | AP (mm) | 2.20 ± 0.76 | 2.22 ± 0.94 | 0.959 | −0.02 (−0.59, 0.55) | 4.01 ± 1.70 | 3.54 ± 1.43 | 0.799 | 0.17 (−0.40, 0.74) |
| ML (mm) | 1.17 ± 0.45 | 1.41 ± 0.54 | 0.169 | −0.53 (−1.13, 0.10) | 1.99 ± 0.65 | 2.17 ± 0.67 | 0.333 | −0.27 (−0.84, 0.32) | |
| MDIST of CoP | AP (mm) | 0.54 ± 0.23 | 0.62 ± 0.36 | 0.374 | −0.18 (−0.74, 0.40) | 1.01 ± 0.45 | 0.91 ± 0.49 | 0.575 | 0.11 (−0.46, 0.68) |
| ML (mm) | 0.31 ± 0.14 | 0.39 ± 0.16 | 0.038 * | −0.66 (−1.29, −0.01) | 0.49 ± 0.14 | 0.58 ± 0.18 | 0.139 | −0.45 (−1.04, 0.16) | |
| Resultant (mm) | 0.67 ± 0.27 | 0.80 ± 0.33 | 0.214 | −0.36 (−0.94, 0.24) | 1.22 ± 0.46 | 1.18 ± 0.48 | 0.646 | 0.05 (−0.52, 0.61) | |
| RDIST of CoP | AP (mm) | 0.64 ± 0.25 | 0.70 ± 0.38 | 0.575 | −0.13 (−0.69, 0.45) | 1.21 ± 0.54 | 1.05 ± 0.50 | 0.508 | 0.16 (−0.41, 0.73) |
| ML (mm) | 0.36 ± 0.16 | 0.44 ± 0.18 | 0.059 | −0.65 (−1.27, 0.004) | 0.59 ± 0.18 | 0.68 ± 0.21 | 0.169 | −0.42 (−1.01, 0.18) | |
| Resultant (mm) | 0.74 ± 0.28 | 0.87 ± 0.35 | 0.285 | −0.30 (−0.88, 0.29) | 1.37 ± 0.54 | 1.29 ± 0.49 | 0.575 | 0.08 (−0.49, 0.65) | |
| Range of CoM | Axis X (mm) | 4.41 ± 1.09 | 4.19 ± 1.37 | 0.374 | 0.24 (−0.37, 0.83) | 3.02 ± 1.56 | 3.44 ± 2.34 | 0.859 | −0.14 (−0.73, 0.45) |
| Axis Y (mm) | 1.84 ± 0.77 | 1.74 ± 0.62 | 0.594 | 0.17 (−0.43, 0.76) | 0.95 ± 0.33 | 1.27 ± 0.67 | 0.139 | −0.53 (−1.16, 0.12) | |
| Axis Z (mm) | 0.46 ± 0.17 | 0.41 ± 0.12 | 0.173 | 0.39 (−0.24, 1.00) | 0.45 ± 0.23 | 0.41 ± 0.13 | 0.314 | 0.27 (−0.35, 0.86) | |
| Variables | Tandem Stance | Feet Together | |||||||
|---|---|---|---|---|---|---|---|---|---|
| CPF | IF | p | g (95% CI) | CPF | IF | p | g (95% CI) | ||
| Range of CoP | AP (mm) | 63.54 ± 46.61 | 107.14 ± 66.35 | 0.114 | −0.52 (−1.12, 0.11) | 3.10 ± 0.82 | 2.60 ± 0.98 | 0.594 | 0.38 (−0.24, 0.99) |
| ML (mm) | 19.18 ± 10.42 | 27.39 ± 10.69 | 0.074 | −0.71 (−1.35, −0.04) | 2.65 ± 1.62 | 2.32 ± 1.19 | 0.515 | 0.21 (−0.39, 0.80) | |
| MDIST of CoP | AP (mm) | 12.27 ± 8.98 | 12.93 ± 6.31 | 0.646 | −0.07 (−0.63, 0.50) | 0.82 ± 0.29 | 0.68 ± 0.31 | 0.594 | 0.24 (−0.37, 0.84) |
| ML (mm) | 4.54 ± 2.03 | 5.66 ± 1.88 | 0.169 | −0.43 (−1.02, 0.18) | 0.67 ± 0.42 | 0.64 ± 0.36 | 0.767 | 0.11 (−0.49, 0.70) | |
| Resultant (mm) | 14.02 ± 9.12 | 15.13 ± 6.13 | 0.445 | −0.12 (−0.68, 0.46) | 1.20 ± 0.36 | 1.03 ± 0.46 | 0.214 | 0.37 (−0.26, 0.97) | |
| RDIST of CoP | AP (mm) | 14.83 ± 10.47 | 18.50 ± 9.34 | 0.285 | −0.27 (−0.84, 0.32) | 0.95 ± 0.31 | 0.79 ± 0.34 | 0.678 | 0.28 (−0.33, 0.88) |
| ML (mm) | 5.34 ± 2.47 | 6.80 ± 2.15 | 0.074 | −0.50 (−1.10, 0.12) | 0.78 ± 0.48 | 0.74 ± 0.40 | 0.767 | 0.11 (−0.48, 0.70) | |
| Resultant (mm) | 16.16 ± 10.39 | 20.05 ± 9.08 | 0.285 | −0.29 (−0.87, 0.30) | 1.33 ± 0.39 | 1.12 ± 0.48 | 0.214 | 0.37 (−0.25, 0.98) | |
| Range of CoM | Axis X (mm) | 29.20 ± 43.68 | 53.45 ± 51.51 | 0.086 | −0.49 (−1.12, 0.16) | 3.58 ± 1.99 | 2.93 ± 1.02 | 0.374 | 0.34 (−0.28, 0.94) |
| Axis Y (mm) | 23.94 ± 18.77 | 55.99 ± 30.90 | 0.021 * | −0.78 (−1.47, −0.07) | 3.61 ± 2.68 | 3.36 ± 2.18 | 0.859 | 0.11 (−0.49, 0.70) | |
| Axis Z (mm) | 4.14 ± 4.18 | 9.45 ± 7.72 | 0.038 * | −0.73 (−1.40, −0.03) | 0.48 ± 0.16 | 0.41 ± 0.15 | 0.214 | 0.41 (−0.23, 1.02) | |
| Amputee ankle ROM | Axis X (°) | 0.47 ± 0.26 | 0.66 ± 0.53 | 0.646 | −0.26 (−0.83, 0.32) | - | - | - | - |
| Axis Y (°) | 0.27 ± 0.13 | 0.43 ± 0.34 | 0.169 | −0.41 (−1.00, 0.19) | - | - | - | - | |
| Sound ankle ROM | Axis X (°) | 1.54 ± 1.43 | 1.70 ± 0.91 | 0.508 | −0.08 (−0.64, 0.49) | - | - | - | - |
| Axis Y (°) | 1.44 ± 1.20 | 2.48 ± 1.30 | 0.022 * | −0.82 (−1.48, −0.13) | - | - | - | - | |
| Maintenance time (s) | 3.57 ± 3.13 | 7.27 ± 7.00 | 0.093 | −0.50 (−1.10, 0.12) | - | - | - | - | |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Jeong, H.-Y.; Yang, H.S.; Lee, Y.-E.; Choi, P.-h.; Jeong, C.-h.; Choi, H.-W.; Lee, Y. Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre–Post Motion Analysis Study. Bioengineering 2026, 13, 949. https://doi.org/10.3390/bioengineering13080949
Jeong H-Y, Yang HS, Lee Y-E, Choi P-h, Jeong C-h, Choi H-W, Lee Y. Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre–Post Motion Analysis Study. Bioengineering. 2026; 13(8):949. https://doi.org/10.3390/bioengineering13080949
Chicago/Turabian StyleJeong, Ho-Yong, Hee Seung Yang, Yea-Eun Lee, Pyoung-hwa Choi, Chan-hyeok Jeong, Hui-Woo Choi, and Young Lee. 2026. "Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre–Post Motion Analysis Study" Bioengineering 13, no. 8: 949. https://doi.org/10.3390/bioengineering13080949
APA StyleJeong, H.-Y., Yang, H. S., Lee, Y.-E., Choi, P.-h., Jeong, C.-h., Choi, H.-W., & Lee, Y. (2026). Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre–Post Motion Analysis Study. Bioengineering, 13(8), 949. https://doi.org/10.3390/bioengineering13080949