3.1. Spherical for Small Pixel Values, 2d = 3λF/#
We investigate the relationship between the
OCE and the
EOD for the increase in spherical aberration from
a40 = 0 to
a40 =
1.25λ. The results for small pixel size
2d =
3λF/# are presented as the
OCE vs.
EOD graph in
Figure 1. The
EOD values are arranged in the ascending order from about
0.3 to about
0.9, corresponding to the decrease in the amount of the spherical aberration from bottom left (
a40 =
1.25λ) to top right (
a40 =
0) of the graph. As the amount of spherical aberration increases, the energy in the central peak of the
psf-s is pushed out symmetrically to the side lobes in the radial direction. The addition of spherical aberration results in the reduction of the
EOD.
When there is no aberration, the OCE achieves a value of 0.88 for the maximum energy on the detector enclosed by the square pixel, and an EOD value of 0.88 for pixel size 2d = 3λF/#. The OCE first smoothly decreases with increasing amount of aberration, arriving at a local minimum of 0.82 for the EOD of 0.58. Then OCE climbs to a sharp peak, nearly equal to the case of zero (0) aberration around the aberration value a40 = 1λ. When the aberration further increases, the lowest value is achieved for the OCE, equal to 0.80 for an EOD of 0.25.
Here we note that the OCE is proportional to the EOD for about half of the EOD interval and inversely proportional over the other half. We may simply observe that they are not correlated. We use the description that two quantities are correlated when they either increase or decrease under similar circumstances. Two quantities are not correlated when their changes occur under different circumstances.
For a small pixel size of 2d = 3λF/#, the OCE first rapidly increases with increasing values of EOD, reaching a maximum of 0.872 when the EOD is equal to 0.31. The low value of the EOD is caused by the fact that the energy moves from the central spot to the higher-order rings for a high amount of aspherical aberration. When the optical axis relative to the pixel position moves around in the OCE determination, the position of the image centroid has no deteriorating consequence because the amount of energy is about the same all over the pixel surface. The OCE value is relatively high because the normalization with the EOD is included in its definition.
The OCE peak around 1λ aberration is followed by a sharp decrease in the OCE values when the EOD is further increased, until a broad minimum is attained at EOD = 0.53, with the still-high value of 0.82 for the OCE. In this region, the amount of spherical aberration has decreased, so the spot is becoming better defined, resulting in increased energy collection over the pixel surface. A more compact spot results in smaller average energy on detector for a displaced pixel center and increased energy on detector for the centered image. Both effects combine to decrease the OCE for small pixel size 2d = 3λF/#. Here, the displacement of the spot position over the pixel area is very sensitive. Only with decreasing aberration values are the OCE and the EOD correlated as expected when the spot over the pixel surface becomes compact under conditions of no or a small amount of aberration.
Figure 2 presents cross-sections of the instrument point response function
prf as a function of a radial coordinate, normalized to pixel size for the points labeled in
Figure 1, that is,
A,
B,
C,
D,
E, and
F.
First, we study the three points near the narrow and sharp peak, that is, Points
A,
B, and
C. The corresponding
prf-graphs are plotted in
Figure 2a–c. At first glance, these three graphs, lined in the first row, appear to be quite like the ones in the second row.
The most important difference to note is the relatively large difference in the prf-values between the center (0) and edge of the detector pixel (0.5). Point B has the highest difference between these two values; therefore, it is located near the local peak for the lower EOD regime. The prf-value at the pixel center (0.) is equal to 0.30. The value of the prf at the pixel edge (0.5) is 0.17. The difference in the prf-values at the center and the edge of the detector pixel is 0.13, for a small pixel value of 2d =3λF/#.
The point A is located at nearly the lowest value of the EOD, 0.279. The prf-value at the pixel center is equal to 0.279. The value of the prf at the pixel edge (0.5) is 0.22. The difference in the prf-values at the center and the edge of the detector pixel is 0.123, a value smaller than that at Point B; therefore, the OCE is also smaller for Point A than for Point B. Point C is located at the value of the EOD higher than that for the highest point (B), at 0.345. The prf-value at the pixel center is equal to 0.345. The value of the prf at the pixel edge (0.5) is 0.24. The difference in the prf-values at the center and the edge of the detector pixel is 0.105, a value smaller than that of Point B; therefore, the OCE is also smaller for Point C than for Point B.
Next, we investigate the
OCE vs.
EOD relationship around the minimum.
Figure 2d–f display the three
prf plots for Points
D,
E, and
F around the minimum. Point
F, featured on
Figure 2f, exhibits a peak of the
prf graph, that is, the
EOD, of
0.66. We see that the value of the
prf at the pixel edge (
0.5) is 0.36. The difference in the
prf at the center, or the
EOD, and that the edge of the pixel (
0.5) is
0.30.
We examine the minimum next, approaching it from the large EOD values. Point
E, at the very minimum of the
OCE vs.
EOD curve is presented in
Figure 2e. It features an
EOD value of
0.53, a value smaller than that of Point
F. We see that the value of the
prf at the pixel edge (
0.5) is
0.33. This
prf graph exhibits a relatively small difference in the
prf values between the center and edge of the pixel of
0.20. Therefore, the value of the
OCE is the lowest there.
Figure 2d features the
prf for point
D. The peak of the
prf, the
EOD, is low among the triplets, at
0.446. The value of the
prf at the pixel edge (
0.5) is
0.32. This is a relatively smaller difference in the
prf-values between the center and edge of the pixel of 0.13. Because of the relatively low peak value of the prf at the center and the relatively high value of the
prf at the pixel edge, Point
D has a higher
OCE value than the neighboring Point
E.
3.2. Spherical with Large Pixel Values, 2d = 7λF/#
The next interesting result is observed when the detector size is increased from
2d =
3λF/# to
2d =
7λF/#.
Figure 3 presents the
OCE vs. the
EOD for a lens with the spherical aberration (
a40 -- term) for the pixel size
2d =
7λF/#. For the larger pixel size discussed here, we start graphing the
OCE with the value of
0.4 for the
EOD with spherical aberration equal to
a40 =
1.25λ. Traveling along the curve, the spherical aberration decreases in the upper right corner to zero.
General features of the
OCE vs.
EOD curve in
Figure 3 for large pixel
2d =
7λF/# are quite similar to those for the small pixel in
Figure 1. The
OCE values increase for very small values of the
EOD and for large values of the
EOD with the increasing amount of aberration. A peak and a valley arise between these two regions, like the shape in
Figure 1. The amount of spherical aberration decreases from the lower left corner with the value of
a40 =
1.25λ along the curve to the upper right corner with
a40 =
0.
In the absence of aberration, the maximum energy on the detector enclosed by the square pixel is 0.95. When the amount of spherical aberration increases from 0 to a40 = 1.25λ, the EOD decreases from about 0.95 to 0.4 when using a pixel size of 2d = 7λF/#. When there is no aberration, the OCE achieves the value of 0.89. The OCE first rapidly decreases with increasing amount of aberration, achieving the valley bottom of the OCE = 0.78 at the EOD = 0.80. The OCE then, with a slope of about (minus) 45 degrees, approaches and forms a rounded peak, nearly equal to the case of no aberration around the aberration value of a40 = 1λ. When the aberration further increases, about an average value is achieved for the OCE, equal to 0.84 for the EOD of 0.4. Here we note that the OCE is proportional to the EOD in about one half of the EOD interval and inversely proportional in the other half. They are not correlated.
Next, we endeavor to explain the first one-third of the curve for the small
EOD values, featuring a narrow, sharp
OCE peak. The
prf-s corresponding to the three points around the local peak (Points
A,
B, and
C) are plotted in
Figure 4. Point response functions
prf-s for Points
A,
B, and
C have a similar profile across the pixel surface. For these points, the
EOD increases from
0.42 through
0.47 to
0.50. The
prf(
0) and
prf(
0.5) for Point
A are
0.421 and
0.25, resulting in their difference of
0.17. The
prf(
0) and
prf(
0.5) for Point
B are
0.47 and 0.31, resulting in their difference of 0.16. Even so, Point
A has a lower
EOD value than Point
B. This leads to a relative higher variation of the
EOD across the pixel surface and therefore a lower
OCE value for Point
A.
The prf-curve corresponding to Point C has the highest EOD value (that is, prf(0)), and a more narrowly formed cone at the center than the other two. Additionally, it also features a relatively larger difference in the prf-values between the center and edge of the detector pixel than that of Point B. The edge value is 0.32, and the center value is 0.5, resulting in the difference from the center to the edge of 0.18. Thus, Point C presents a larger variation in the prf-values across the pixel surface. Therefore, Point C ends up having a somewhat lower OCE value than Point B.
In the presence of 3rd order spherical aberration in an otherwise ideal lens-optical detector combination, the
EOD decreases in the presence of aberration because the energy in the central spot is being pushed out to the outer rings. The amount of energy that is incident on a given pixel and the average
prf across the detector determine the relationship between
EOD and
OCE. For a different size pixel, with its value somewhere between
2d =
3λF/# and
2d =
7λF/#, the initial points and the end points would be lying between the corresponding points on
Figure 1 and
Figure 3. The
OCE-vs-
EOD curve would exhibit a shape similar to the two limiting cases presented here and lying in-between them.