Next Article in Journal
One-Class Drift Compensation for an Electronic Nose
Next Article in Special Issue
Current Progress of Magnetoresistance Sensors
Previous Article in Journal
Comparison of Optical Ammonia-Sensing Properties of Conducting Polymer Complexes with Polysulfonic Acids
Previous Article in Special Issue
Electronic Surveillance and Security Applications of Magnetic Microwires
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

MEMS-Based Cantilever Sensor for Simultaneous Measurement of Mass and Magnetic Moment of Magnetic Particles

by
Wilson Ombati Nyang’au
1,2,3,
Tamara Kahmann
2,4,
Thilo Viereck
2,4 and
Erwin Peiner
1,2,*
1
Institute of Semiconductor Technology (IHT), Technische Universität Braunschweig, D38106 Braunschweig, Germany
2
Laboratory for Emerging Nanometrology (LENA), Technische Universität Braunschweig, D38106 Braunschweig, Germany
3
Department of Metrology, Kenya Bureau of Standards (KEBS), Nairobi 00200, Kenya
4
Institute of Electrical Measurement Science and Fundamental Electrical Engineering (EMG), Technische Universität Braunschweig, D38106 Braunschweig, Germany
*
Author to whom correspondence should be addressed.
Chemosensors 2021, 9(8), 207; https://doi.org/10.3390/chemosensors9080207
Submission received: 1 July 2021 / Revised: 30 July 2021 / Accepted: 2 August 2021 / Published: 4 August 2021
(This article belongs to the Special Issue Advances in Magnetic Sensors with Nanocomponents)

Abstract

:
This study presents a measurement approach suitable for the simultaneous determination of both the mass mp and magnetic moment µp of magnetic particles deposited on a micro electro mechanical system (MEMS) resonant cantilever balance, which is operated in parallel to an external magnetic field-induced force gradient F′(z). Magnetic induction B(z) and its second spatial derivative δ2Bz2 is realized, beforehand, through the finite element method magnetics (FEMM) simulation with a pair of neodymium permanent magnets configured in a face-to-face arrangement. Typically, the magnets are mounted in a magnet holder assembly designed and fabricated in-house. The resulting F′ lowers the calibrated intrinsic stiffness k0 of the cantilever to k0-F′, which can, thus, be obtained from a measured resonance frequency shift of the cantilever. The magnetic moment µp per deposited particle is determined by dividing F′ by δ2Bz2 and the number of the attached monodisperse particles given by the mass-induced frequency shift of the cantilever. For the plain iron oxide particles (250 nm) and the magnetic polystyrene particles (2 µm), we yield µp of 0.8 to 1.5 fA m2 and 11 to 19 fA m2 compared to 2 fA m2 and 33 fA m2 nominal values, respectively.

Graphical Abstract

1. Introduction

The merit of magnetic particles (MPs) and their diverse and widespread applications is an ever-present reality that continues to trigger endless possibilities. These particle types have, for instance, greatly revolutionized many fields of science and technology, including biomedicine and environmental monitoring. MPs often serve as magnetic separation tools in effluent processing for contaminants removal, remediation and water treatment [1,2]. They also primarily serve as diagnostic tools [3,4], e.g., as contrast enhancement gents and as tracers for magnetic resonance imaging (MRI) and magnetic particle imaging (MPI) processes, respectively. Their appreciable use is quite evident in the targeted delivery of therapeutic drugs, gene and radionuclides, magnetic hyperthermia (for the treatment of tumors), treatment of iron deficiency, magnetic separation of biomolecules such as proteins, cells, antibodies and bacteria and as magnetic markers (for biosensing) [1,2,5,6,7,8,9,10]. For a successful and versatile application, knowledge of the magnetic properties of these particles (e.g., particle size, the chemical nature of magnetic material, magnetic content, saturation magnetization) is essential and influential. Magnetic particles are easily manipulated by applying a sufficiently high external magnetic field, which tends to align the particle’s magnetic moment(s) with the magnetizing field. In addition, if the latter is turned off, the magnetic dipole moments begin to randomize, thereby reducing the bulk magnetization [7].
The detection and characterization of magnetic particles deploy several techniques. Notwithstanding, various methods are utilized for the colloidal and physicochemical characterization of MPs. This consists of electrical conductivity measurements, a thermogravimetric analysis (TGA), scanning electron microscopy (SEM), transmission electron microscopy (TEM), X-ray diffraction (XRD) technique, Fourier transform infrared spectroscopy (FTIR) and atomic absorption spectroscopy (AAS). Thermogravimetric analysis has, for instance, been described as an effective chemical composition measurement method for quantitative determination of the mass content and magnetic content percentages of MPs [11,12,13]. Other magnetic particle characterization approaches include optical methods (e.g., dynamic light scattering), the AC susceptibility method, nuclear magnetic resonance (NMR) spectroscopy, magnetic resonance imaging (MRI) and magnetic particle force microscopy (MPFM) [14,15], magnetic particle spectroscopy (MPS) and magnetorelaxometry (MRX) [16,17,18,19].
Some of the utilized magnetic sensing devices also include induction coils; solid-state magnetic sensors such as fluxgate magnetometers, magnetoresistive magnetometers, Hall effect magnetometers, magnetodiodes, magnetotransitors, magneto-optical magnetometers and optically pumped magnetometers, and superconducting quantum interference device (SQUID) magnetometers [12,20,21,22]. Whilst induction coils usually suffer from low sensitivity, SQUIDs offer an exceptional sensitivity but are rather bulky and consume high power. Solid-state magnetic sensors are usually designed to detect the stray field of magnetized MPs, either in parallel or perpendicular to the plane [12]. Moreover, the determination of the magnetic moment of magnetic particles using Hall-effect sensors necessitates a prior knowledge of particle geometry.
On the other hand, micro-/nanoelectromechanical systems (M/NEMS)-based cantilever sensors have also been utilized to detect and characterize MPs, and have remained afloat over the years due to their miniaturizability and versatility. In this particular study, the effect of externally induced magnetic fields on the resonant frequency f0 of a cantilever mass sensor (with and without adsorbed magnetic particles) is investigated. We optimized the magnetic field gradient by tuning the separation between a pair of identical permanent neodymium magnets. Furthermore, the magnetic moment of the MPs is also determined as a function of the resonant frequency of a vibrating cantilever. For comparison, the magnetic moment of each type of MPs was also experimentally characterized using a SQUID magnetometer. Ultimately, our method provides an inexpensive mass-magnetometry method for characterizing the MPs.

2. Experimental

2.1. Materials Description and Characterization

2.1.1. Magnetic Particles Samples

In this study, two types of commercial monodisperse magnetic particle samples were used: 250 nm iron oxide nanoparticles (Fe3O4 NPs, product code: 45-00-252) and 2 µm magnetic polystyrene particles (micromer®-M, product code: 08-03-203), both from micromod Partikeltechnologie GmbH, Rostock, Germany. Particle concentration Cp,v of the stock solution was 50 mg/mL (with 5.5 × 109 MPs per mL, micromer®-M) and 25 mg/mL (with 5.7 × 1011 MPs per mL, Fe3O4 NPs). The iron oxide particles and magnetic polystyrene beads were designed with plain surface and hydrophilic alkyl-OH coating, respectively [23,24]. The magnetic polystyrene particles and iron oxide NPs consist of a large number of crystallites with an average size of about 6.62 ± 0.23 nm (embedded in a polymer matrix) and 10 nm (cluster-type), respectively. Prior to sampling, the particles in the test solution were homogeneously dispersed by ultrasonic dispersion.
The properties of the used MPs are listed in Table 1. Here, the particle volume Vp was calculated from the nominal particle diameter dp, whereas particle mass mp was computed from the product of the particle volume and effective particle density. Consequently, the magnetic moment per particle is determined from:
μ p = m p M m ,
where Mm is the mass magnetization (in A m2/kg particles). The particle mass is calculated using:
m p = V p ρ eff
ρ eff = ( 1 φ Fe 3 O 4 ) ρ PS + φ Fe 3 O 4 ρ Fe 3 O 4
where ρeff is the effective particle mass density; ρPS and ρFe3O4 are the mass densities of polystyrene (PS) and magnetite (Fe3O4), respectively, and φFe3O4 denotes the volume fraction magnetite in each particle. Magnetic polystyrene particles have a low magnetic content, ~15% (w/w) [25]. Consequently, the volume ratio of magnetite VFe3O4 to the total particle volume Vp is φFe3O4 ≈ 3.41%.

2.1.2. MEMS Cantilevers Sensors

Here, three main types of silicon-based micro electro mechanical system (MEMS) cantilever sensors were used: CAN50-2-5, CAN30-1-2 and CAN15-3-2, all of which were commercially obtained from CiS Forschungsinstitut für Mikrosensorik GmbH, Erfurt, Germany. The sensing mechanism of these sensors was piezoresistive, while their excitation was externally induced using an extensional piezoelectric stack actuator (PL055.30, from PI Ceramic GmbH, Lederhose, Germany). The main attributes of the deployed cantilever sensors are listed in Table 2. These include: cantilever’s mass m0, resonant frequency f0 (the fundamental mode), stiffness k0, mechanical quality factor Q and frequency stability σ (Allan deviation). From these, we could calculate the minimum detectable frequency Δfmin, mass Δmmin and magnetic moment μmin.
For a cantilever with length L, width w, thickness h and material density ρ (silicon: 2.33 g/cm3), its mass m0 = ρLwh. Given knowledge of material’s Young’s modulus E (for a (100) silicon, along the [011] direction of cantilever axis, E = 169 GPa [27]), one can calculate the cantilever’s geometric spring constant k0 (based on Euler–Bernoulli beam theory) using:
k 0 = 3 E I L 3 ,
where the parameter I = wh3/12 denotes the geometric moment of inertia for an out-of-plane excited rectangular cantilever. It should be noted that k0 can also be determined using Equation (5) from the measured resonant frequency f0 and the calculated cantilever mass m0. The k0 values given in Table 2 were determined based on Equations (4) and (5).
k 0 π 2 f 0 2 m 0
A magnetic dipole in a magnetic field B is subjected to a magnetic force according to [22]:
F = ( μ · B ) .
We considered the particles as sufficiently small that the magnetic field was spatially uniform, and both µ and B were aligned in the z direction. For the magnetic force gradient along z we, thus, obtained:
F = δ F δ z = μ δ 2 B δ z 2
With the experimentally measured frequency noise of the sensor in the time domain (σ, Allan deviation), the minimum detectable values of mass (Δmmin) and magnetic moment μmin were estimated using Equations (8) and (9), respectively [28,29]:
Δ m min = 2 m 0 σ
μ min = 2 k 0 δ 2 B δ z 2 σ
where δ2Βz2 is the second derivative of a magnetic flux density (in T/m2) along the direction z, in which the magnetized cantilever is oscillating (magnetization and cantilever deflection are parallel). Here, µ is the total magnetic moment (in A m2) of the magnetized cantilever:
μ = N p μ p ,
where, Np is the number of particles (given by their volume Vp and the total mass of the attached particles) and µp is the magnetic moment per magnetic particle. We found minimum detectable values of the magnetic moment (corresponding to resolution) for the considered piezoresistive cantilevers of μmin = 6 pA m2 to 169 pA m2, which was slightly larger than 0.65 pA m2 of a recently reported micro-ring force sensor-based magnetometer [30].
Since a clear knowledge of k0 is necessary for precise determination of F′ (using Equation (16)), stiffness calibration was exemplarily performed with some selected cantilevers. This was conducted using the force/balance method [31]. Here, a nano-force measuring setup comprising of a one dimensional (1D) translational positioning stage (P-726 PIFOC, from Physik Instrumente (PI) GmbH & Co. KG, Karlsruhe, Germany: resolution 1 nm, uncertainty 10 nm) was used to precisely move the cantilever, as well as a commercially available compensation balance (Sartorius SE-2 micro-balance: resolution 1 nN, reproducibility 2.5 nN, linearity deviation 9 nN, uncertainty 0.1 µN) (see Figure 1). The cantilever chip comprising a conical tip at its free end was mounted on the 1D stage, with the tip oriented downwards. A 2-micron-sized sphere (composed of polished ruby) was mounted on the balance pan and then utilized as a probing area.
During stiffness calibration measurement, the cantilever was equidistantly and incrementally moved down in defined steps (~0.1 µm) until it touched the ruby sphere and some micrometers further in order to measure the probing curve. Force deflection curves were then measured up to a maximum force of 10.2 µN. The position of the cantilever and the mass of the compensation balance were then simultaneously recorded. In this case, five measurement cycles were performed, each cycle comprising of 20 measurements. The averages and standard deviations from each measurement cycle were then pooled together to determine the mean value and the repeatability (standard uncertainty). If we take into consideration other uncertainty contributions such as temperature drift, humidity, balance/mass comparator and the tip (orientation), an overall (maximum expanded) uncertainty in spring constant calibration of ± 4% (corresponding to about 95% confidence level) was estimated.
Another important pre-calibration process that was performed was to clean up the cantilevers in acetone to remove dust and any other contaminants that would otherwise be on the surface. Moreover, to guarantee and render the sensors re-usable after adsorbing magnetic particles thereon, each cantilever was cleaned by sonicating in acetone for about 3–5 min. Before sonication, a cantilever was mounted on a specially designed metallic holder to mitigate the risk of breakage [32]. Thus, repetitive use of the cantilever sensor was possible.
Individual sensor chips were glued onto in-house carrier chips, and their contact pads were electrically connected via wire-bonding (Figure 2a). The prepared chip was then mounted onto a cantilever sensor PCB (fixed on a workbench). Consequently, the contact pads of the carrier chip were contacted through metallic shafts (fixed onto the PCB, see Figure 2b) and then fastened/screwed.
To supply (or obtain) voltage to (or from) the cantilever sensor (Wheatstone bridge), SMA coaxial cables were connected to dedicated SMA connectors (ports) on the cantilever sensor holder PCB. Here, a lock-in amplifier (MFLI, from Zurich Instruments AG, Zurich, Switzerland) was used to supply voltage to both the cantilever sensor (up-to 5 Vdc) and piezo actuator (up-to 10 Vpk). Each cantilever was placed on the piezo actuator (within the cantilever sensor holder PCB), as depicted in Figure 2b.
Subsequently, magnetic particles solution was deposited onto the cantilever sensors. Deposition of MPs solution on the sensor’s surface was performed based on droplet dispensing approach, as discussed in a previous work [33]. In general, a small liquid volume was deposited and solvent vaporized under ambient conditions, leaving behind (on the beam) an adsorbate of MPs. Localized deposition of the MPs was conducted at a point towards the free-end of the sensing beam, aided by a digital camera. The added particle mass Δm proportionately changed the unloaded resonant frequency of the cantilever by Δf as follows [32]:
Δ m = 2 m 0 f 0 1 U   2 ( x Δ m ) Δ f
where U(xΔm) is the dimensionless parameter denoting the mode shape function of the cantilever. The number of particles added on the cantilever sensor was then determined from the ratio of total measured Δm to the mass per particle, mp.
To realize magnetization of the particles into the direction perpendicular to the cantilever surface prior to their immobilization after solvent evaporation, we dispensed the droplet on the cantilever after positioning it in the field of a permanent magnet as described below. It should be noted that whenever the particles were immobilized after solvent evaporation, Brownian relaxation was suppressed and a constant/remanent magnetization prevailed.

2.1.3. Magnets and Magnet Assembly

A magnetic field, created by a pair of Neodymium (NdFeB) permanent disk magnets (from Webcraft GmbH, Gottmadingen, Germany) was used to align the magnetic moment of the magnetic particles and to induce a magnetic force gradient in the same direction. The radius rm and length lm of these axially magnetized identical NdFeB magnets (type N42) were 5 mm and 3 mm, respectively. Their average residual magnetism and coercive field strength was ~1.31 T and 908 kA/m [34], respectively. These magnets were mounted in the magnets holder assemblies, whose designs and fabricated samples are shown in Figure 3.

2.2. Finite Element Modeling of Second-Derivative Magnetic Field

The magnetic force gradient was determined by finite element modeling using the Finite Element Method Magnetics, FEMM v. 4.2 software [35]. To tune the magnetic field B and its second spatial derivative along z2Bz2), two parameters were considered: the distance between a pair of magnets d12 and the arrangement of the permanent magnets. Here, as depicted in Figure 4a, the two NdFeB permanent magnets were configured with a face-to-face arrangement. These magnets possess axisymmetric magnetization (i.e., with the magnetization direction through the axis of the magnet). The spatial magnetic field distribution generated by the two repelling (opposing) faces of magnets (i.e., north-to-north pole, N–N, or south-to-south pole, S–S) along the z-axis and r-axis is for instance depicted in Figure 4b,c, in which magnets had a spacing of d12 ≈ 2 mm.
In Figure 5, we show the maximum magnetic induction field (for two identical NdFeB disk magnets in repelling arrangement) vs. magnets–pole spacing, along the axisymmetric (z) and radial (r) axes. With the latter, the indicated maximum induction field was observed nearly at the edge of the magnet (i.e., at r ≈ 2.5 mm). With the former (i.e., Figure 4b) as expected, the zero B-field was observed at the mid-point between the two permanent magnets with opposing faces (N-N or S-S). Usually, B-values would be negative to one side (field lines point in opposite directions) and then the graph was almost a straight line over the entire field of view. However, as depicted in Figure 4a, an absolute value of B (i.e., |B|) was used, leading to a sharp change of slope at z = 0. This was in contrast to Figure 5, where with increasing r the B-field approached a maximum at r ≈ 2.5 mm under a steadily decreasing slope.
For a 2 mm magnets spacing, we found a maximum second derivative of magnetic flux density δ2Bz2|max ≈ 1.57 T/mm2 along z-axis (cf. red solid line, Figure 4b), whereas along the radial axis the maximum gradient δ2Br2|max ≈ 0.30 T/mm2 was much lower (i.e., at point P2, cf. Figure 4a; red solid line, Figure 4c). Therefore, the former condition was adopted for the experiments as subsequently described. Here, the second derivative fields were determined by finite difference approximation of the spatial field distribution along the z and r axis, respectively.

3. Results and Discussion

3.1. Magnetic Force Gradient Sensing

To generate an interaction of the magnetic force gradient induced by the second derivative of the B field with the magnetic particles on the cantilever, pre-magnetization was performed. Hence, the particle-laden droplet (on the cantilever) was deposited and immediately subjected to the magnetizing field prior to complete solvent evaporation. We assumed that through the Brownian relaxation mechanism the particles easy axes rotated in the z-direction (i.e., along the applied field) [36]. To realize this, we used the magnet holder in Figure 3e. This holder assembly gave the possibility to deposit the particles on the cantilever in the form of a water suspension (as a droplet) and then simultaneously align them into the direction perpendicular to the cantilever surface. In essence, this happened prior to the particle immobilization after the solvent evaporation. Under the influence of an external magnetic field (induced by a permanent magnet), it suffices to mention that the dispersion of magnetic particles within the droplet could be biased towards the droplet base (contact line). In this work though, this was inconsequentially assumed since the liquid was evaporated before conducting the measurement. Upon complete evaporation, the particles were adsorbed on the sensing surface (cf. Figure 2c,d).
To detect the cantilever-stiffness-changing effect of the magnetic force gradient, the setup depicted in Figure 6 was employed. This mainly comprised the sensor (cantilever plus attaching particles magnetized in z direction) and a magnetic force gradient (induced also in z direction by two permanent magnets spaced d12 apart). From Figure 5, δ2Bz2|max and δ2Br2|max varied with respect to the distance d12 between the poles of the two magnets. Thus, as deducible from Equation (5), to generate a sufficiently large magnetic force gradient F′, d12 had to be set as small as possible. Moreover, by subjecting the immobilized particles to a magnetizing field (in z direction) during the cantilever frequency response measurement, a further alignment (through the Néel relaxation mechanism, i.e., the rotation/alignment of the particle’s magnetic moment inside the magnetic core with the applied field against an energy barrier without interacting with the particle’s surrounding [16]) could be assumed. This was favorably due to the fact that both magnetic particles were composed of small crystallites (less than 10 nm in size).
In detecting the magnetic force gradient F′(z0) = ∂F/∂z|z0, a cantilever should be sinusoidally driven around its equilibrium position by a force F(z0) along the z axis (Figure 6) in accordance to the equation of motion for a lumped element oscillator [37]:
m 0 2 z t 2 + γ z t + ( k 0 F ( z 0 ) ) z = F ( z 0 )
where γ denote the damping coefficient, and the term k0F′(z0) is equivalent to the effective spring constant of the cantilever. Based on Equation (12), it is evidently clear that the magnetic force gradient F’ acted to soften (reduce) or harden (increase) the spring constant. Consequently, this served to modulate and shift the cantilever resonance frequency Δf’. Utilizing Equations (5) and (12) yielded Equation (13). If we assume F′/k0 to be << 1, then using the first order Taylor series expansion, Equation (13) reduces to Equation (14). By substituting the angular frequency ω with the ordinary frequency f, we obtained the resonant frequency shift Δf’ due to the influence of the magnetic force gradient F′ as delineated in Equation (15).
ω = k 0 F m = ω 0 1 F k 0
ω   ω 0 ( 1 F 2 k 0 )
Δ f f 0 2 k 0 F
Thus, by rearranging Equation (15) we obtained:
F 2 k 0 f 0 Δ f
Under ideal measurement conditions (e.g., free of dust and contaminants), the total mass of the cantilever (i.e., m0 + ∆m) does not change upon exposure to the external magnetic force field gradient. Thus, the attached magnetic particles mass ∆m was assumed to be invariant.
In Equations (15) and (16), Δf is the effective typical resonant frequency difference between a cantilever (containing magnetic particles), within and without external magnetic fields exposure. To determine it, blank measurements were performed beforehand, as depicted with the black line graph in Figure 7. Here, a change in the resonant frequency of an unloaded clean cantilever was unexpectedly observed upon exposure to an external magnetic force field. It should be noted that the silicon-based cantilever contains electrical/metallic connection lines which were not shielded from the external magnetic field. Thus, if a current was induced in the electrical conductors it would produce thermal energy (Joule heating). Upon a rise in temperature, the Young’s modulus of silicon and stiffness k0 will decrease. Furthermore, we suppose that the cantilever may experience stress, e.g., by a bimorph effect due the stack of materials of different thermal expansion coefficients built in the cantilever, that consequently alter the fundamental resonant frequency of the vibrating system.
In Figure 7, Bon and Boff represent the manual switch on and off of the magnetic force field, respectively. It should be noted that the average frequency difference between the Boff states corresponded to the resonant frequency shift Δf due to the added mass Δm = 8.22 ± 0.54 ng (Np = 2050 ± 135). Furthermore, it is worth noting that there was no systematic phase shift between the unloaded and loaded cantilever excitation conditions. Thus, to determine the change in resonant frequency (Δf’) due to the externally induced magnetic field on the magnetic moment of particles, the average resonant frequency due to Bon state was subtracted from that of Boff.
To determine F′ using Equation (16), a known (calibrated) value of k0 and the frequency shift Δf’ were required. For instance, in Figure 7, Δf’ ≈ −22 ± 12 mHz (i.e., the effective frequency difference obtained from a cantilever (CAN30-1-2) with and without magnetic polystyrene particles under a magnetic force field condition comprising of two permanent magnets placed 2 mm apart). This cantilever’s calibrated stiffness k0 was 4.46 ± 0.09 N/m. By substituting k0 and Δf’ into Equation (16), we obtained a magnetic force gradient F′ ≈ 44 ± 24 µN/m, which was considered significant since it was much larger than F′min = 10 ± 2 µN/m.
A much clearer trend was observed upon the attaching iron oxide NPs on the cantilever (→ frequency shift by added particles mass, Figure 8a) and subjecting it to the magnetic force gradient under the same conditions as above (Figure 8b). With these particles, Δf and F′ were about −4.1 ± 0.2 Hz and 4.53 ± 0.94 mN/m, respectively. Since the stiffness of this type of cantilever (k0 ≈ 12.5 ± 0.8 N/m, CAN15-3-2, Equation (5)) was larger than that of CAN30-1-2 (cf. Figure 7), its sensitivity to the external magnetic force field was lower. However, as depicted in Figure 8b, the induced frequency shift was much larger compared to Figure 7. This was mainly attributed to the much larger number of attached particles (Np = (254 ± 21) × 103 vs. 2050 ± 135 in case of the magnetic polystyrene beads). Consequently, this yielded an almost one-order of magnitude higher total magnetic moment of the plain iron oxide particles compared to magnetic polystyrene beads (even though their magnetic moment per particle was more than one order of magnitude lower, cf. Table 1). Nonetheless, a slight continuous change in frequency was visible in the Bon state on Figure 8b which is supposedly due to temperature change effect on the magnetic particle magnetization or the cantilever (whose increase led to a decrease in cantilever stiffness).
To determine the frequency response of a cantilever sensor (with 230 ± 20 ng of 250 nm-sized iron oxide particles, S84/CAN30-1-2) under different magnetic force gradients, the distance between a pair of magnets was sequentially reduced in steps (see Figure 9a). In this condition, the position of the cantilever was fixed at a point nearly at the midpoint of the magnets, while the magnets’ spacing (d12) was varied from about 9 mm (green line) to 4 mm (blue line), 2 mm (red line) and 1 mm. As expected, in the magnetic force gradient environment, the resonant frequency changes on the bare cantilever tend to zero (nearly negligible, open red circles in Figure 9b), while for the cantilever loaded with magnetic particles it increased with decrease in the distance between the two identical magnets (see full red circles in Figure 9b). Consequently, in Figure 9c we show the magnetic force gradient (derived from the effective frequency shifts in Figure 9b using Equation (10)) against the corresponding second-derivative magnetic-induction gradient. According to Equation (7), the resultant slope of the linear fitting in Figure 9c was equivalent to the total magnetic moment (μ = 3.8 ± 0.5 nA m2) in the test sample.
In Table 3, the derived magnetic force gradient observed with a varying attached particles mass (numbers) on various assorted piezoresistive cantilevers is presented. For sensors with nearly the same k0 and comparable f0, an increase in the particle mass resulted in a corresponding increment in the magnetic force gradient. It is also evidently clear that the force gradient varied with the particles type on the beam. In which case, for cantilevers with comparable masses (such as on sensors S97 and S89), the frequency response due to Fe3O4 NPs was found to be much greater compared to the microbeads (magnetic polystyrene particles), which was attributed to the much larger particle numbers. Moreover, taking the S80/CAN30-1-2 cantilever (Table 3) into consideration, the calibrated cantilever stiffness was found to differ with the calculated values by about 45% (Equation (4)) and 28% (Equation (5)), which is generally expected mainly due to an uncertainty of the cantilever thickness related to the fabrication tolerances. Replication of these variances was obviously expected on their respective calculated magnetic force gradients. Consequently, the calibrated stiffness has a lower uncertainty compared to calculated values. With calibration, it should be noted that the measurements were performed under stable ambient conditions, and the repeatability (the major source of uncertainty) was small, ~6 mN/m.

3.2. Magnetic Moment Determination

The nominal magnetic moment calculated for each magnetic particle sample (Table 1) assumed that all the particles were 100% polarized along the easy axis. For a known particle mass, the total nominal magnetic moment μ on the cantilever could be determined based on Equation (10). Considering a mass-induced frequency change Δf = −3.72 ± 0.02 Hz due to the added magnetic polystyrene particles (Δm = 8.22 ± 0.54 ng; cf. Table 3) yielded a total nominal magnetic moment of 28 ± 15 pA m2. For iron oxide NPs on a CAN15-3-2 cantilever with Δf = −131.67 ± 0.13 Hz (→Np = (254 ± 21) × 103), μ = 2.9 ± 0.6 nA m2 (Equation (7)).
Using the magnet arrangement with d12 ≈ 2 mm (→ δ2Bz2|max ≈ 1.57 T/mm2), the corresponding magnetic moment per particle was determined from our measurements using Equation (7) as indicated in Table 4. In case of the micromer beads, the average resultant magnetic moment per particle was about 15 ± 10 fA m2 (Table 4) compared to the 33 fA m2 nominal value (see Table 1). For Fe3O4 particles, an average magnetic moment per particle of about 1.2 ± 0.8 fA m2 was achieved (from S89 and S84), which was also slightly lower than the nominal magnetic moment (2 fA m2). Unlike the other sensors, S88 yielded a much larger magnetic moment, which we consider as an outlier. Of all the used cantilevers, S88/CAN15-3-2 was evidently the stiffest. Nevertheless, the observed large deviation from nominal magnetic moment could probably arise from sensor positioning error within the second derivative B field, and variance in particle size distribution.
Furthermore, it was also assumed that the magnetic moments in the MP were 100% aligned perpendicular to the cantilever surface (along the cantilever deflection) along the easy axis. This was essentially important for the quantitative determination of magnetic moment per particle µp. At the center position between the magnets, the magnetizing field was usually zero; thus, we could not expect a supporting (dynamic) alignment and only the pre-aligned moments contributed to the responses.
To compare and validate the above results, static magnetic moments of the magnetic particles were exemplarily measured using a SQUID-based magnetometer (MPMS-3, from Quantum Design GmbH, Darmstadt, Germany [38,39]), as depicted in Figure 10. To perform these measurements, a 15 µL stock solution of each magnetic particles sample was placed on an in-house designed liquid sample holder. It should be noted that the concentration of the stock solutions was 5.5 × 109 MPs per mL (micromer) and 5.7 × 1011 MPs per mL (iron oxide particles) [23,24]. Thus, a test solution of 0.015 mL corresponds to ~8.25 × 107 MPs (micromer) and 8.55 × 109 MPs (Fe3O4). The volume of magnetite for each magnetite polystyrene and iron oxide particles was determined to be 0.15 µm3 and 0.008 µm3, respectively (Table 1). Therefore, the corresponding magnetite total volume (and mass) in the 15 µL test sample for micromer beads and Fe3O4 NPs was 0.01 µL (65 µg) and 0.07 µL (373 µg), respectively.
At zero field, the magnetic moment of a particle was nearly zero, or tends thereto (see Figure 10). Moreover, as expected, by increasing the magnitude of the magnetizing field, the magnetic moment concomitantly increased and eventually saturated. It should be noted that in our SQUID measurements, particle suspensions were used and the observed phenomenon (Figure 10) would connotate the superparamagnetic behavior of these particular type of particles. Thus, at the saturating magnetizing field, we could use a corresponding magnetic moment and the total volume of magnetite to determine saturation magnetization Ms (i.e., M s = μ p V Fe 3 O 4 ). Consequently, it is possible to determine the mass magnetization Mm,s from the ratio of Ms to the density of magnetite ρFe3O4. For the microbeads (micromer MPs), the magnetic moment saturated at ~9.97 ± 0.03 µA m2 (at a magnetizing field H = 1592 kA/m). With Fe3O4 NPs, a total magnetic moment of ~21.98 ± 0.11 µA m2 was obtained (at H = 1592 kA/m). From Figure 10, the corresponding total magnetic moment at H = 800 kA/m was 20.34 ± 0.01 μA m2 (iron oxide plain NPs) and 9.62 ± 0.04 μA m2 (magnetic polystyrene particles).
To obtain magnetic moment per particle from SQUID measurements, it’s necessary to divide the total magnetic moment with the total number of magnetic particles in the test sample. Microbead (micromer) and nanobead (Fe3O4) yielded 116.6 ± 0.5 fA m2 and 2.379 ± 0.001 fA m2, respectively, per particle. Obviously, the obtained moment for the microbeads was much greater than both the nominal value (by factor of ~3.5) and the cantilever-based determined moment (Table 4). It should be noted that with SQUID-based measurements, the number of particles in the test samples was estimated from the volume fraction. The observed discrepancies would probably arise from insufficient knowledge of the particle number in the test sample, e.g., homogeneity in particle concentration as well as pipetting errors (volume sampling). In a previous work [39], the possibility of interaction was also observed to affect the effective magnetic moment of the multi-core particle. We also further suppose that a change in polarization over time might affect the measured values. Unlike cantilever-based samples where diluted particle solutions were employed, SQUID-based samples were taken from stock solutions, which is highly concentrated; hence, particles agglomeration was much more probable here.
Unlike SQUID, the determination of the particle number (from the added mass) in the cantilever-based resonance balance approach was much more precise. Moreover, a much smaller liquid sample (pico-sized droplet) could be deposited on the cantilever sensor. In principle, magnetic nanoparticles dynamically reacted to a change in external magnetic field. In order to show their superparamagnetic behavior, here, magnetization measurements could also be performed as the solvent evaporates. Since the evaporation process was rather fast for a small droplet, faster measurements were highly desirable. To effectively facilitate this, a replacement of the manual switching on and off of the field with a much more precise control mechanism involving an electromagnets setup was intended. Moreover, since this measurement system/approach is quite promising, it can also be flexibly adapted for localized particle detection in mapping (magnetic particle mapping [40]).

4. Conclusions

Here, we have presented a resonant-based measurement approach for determining both magnetic particles’ mass m and the magnetic moment µ. With mass determination, an external magnetic force field was negated. Thus, MEMS cantilever-based cantilever sensors containing particles of interest were excited and the induced proportionate changes in resonant frequency (in fundamental mode) were determined. Given the knowledge of the added mass, the particle number concentration of the test samples on the cantilever was evaluated. To obtain µ, a clear knowledge of magnetic force gradient F’ and second derivative magnetic field δ2Bz2 was necessary. The latter was determined from the linear field gradient produced from a magnetic arrangement between two opposing faces of permanent magnets (N–N, or S–S) through FEMM simulation, whereas the former was established from the measured resonant frequency shift Δf’ under an externally induced magnetic gradient field. Using plain iron oxide nanoparticles (250 nm) and magnetic polystyrene beads (2 µm), we obtained a magnetic moment per particle of ~ 1.2 ± 0.8 fA m2 and 15 ± 10 fA m2 compared to 2 fA m2 and 33 fA m2 nominal values, respectively. The range of SNR (µ/µmin) obtainable from the present sensing system was nearly 0.8 to 57. To enhance sensitivity, an ultrathin cantilever design (L × w × h = 2000 × 50 × 5 µm3) based on silicon-on-insulator (SOI)-based MEMS technology was envisaged. This sensor had a much smaller stiffness k0 ≈ 0.025 N/m and could yield µmin = 36 fA m2 and mmin = 0.62 pg. An array of cantilevers with different spring constants k0 (or functionalization) could also be considered to further improve the effective sensing response of the cantilever sensor to the magnetic moment. To mitigate the influence of ambient temperature on the results, measurements could also be conducted under controlled environmental conditions.

Author Contributions

Conceptualization, W.O.N.; methodology, W.O.N.; validation, W.O.N., T.K. and E.P.; formal analysis, W.O.N.; investigation, W.O.N. and T.K.; resources, T.V. and E.P.; writing—original draft preparation, W.O.N.; writing—review and editing, W.O.N., T.K., T.V. and E.P.; supervision, E.P.; project administration, T.V. and E.P.; funding acquisition, E.P. All authors have read and agreed to the published version of the manuscript.

Funding

This project has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme under no. 17IND05 MicroProbes, and Niedersächsisches Vorab through the “Quantum- and Nano-Metrology (QUANOMET)” initiative within the project NP-2.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available upon request from the corresponding author.

Acknowledgments

W.O.N. is grateful to the German Federal Ministry for Economic Cooperation and Development (BMZ) for a doctoral scholarship within the Braunschweig International Graduate School of Metrology (B-IGSM). T.K. gratefully acknowledges the support of the B-IGSM and the DFG Research Training Group 1952 Metrology for Complex Nanosystems. Moreover, we gratefully thank Radovan Popadic (PTB), Michael Fahrbach and Agus Budi Dharmawan for their valuable assistance and fruitful discussions on cantilever stiffness calibration measurements, wire bonding of the sensors and on 3D-printing of the magnet holder assemblies, respectively. The technical support given by Juliane Breitfelder, Angelika Schmidt, Aileen Michalski, Diana Herz, Andreas Heidemann and Karl-Heinz Lachmund is also gratefully acknowledged.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Mohammed, L.; Gomaa, H.G.; Ragab, D.; Zhu, J. Magnetic nanoparticles for environmental and biomedical applications: A review. Particuology 2017, 30, 1–14. [Google Scholar] [CrossRef]
  2. Horák, D.; Babic, M.; Macková, H.; Benes, M.J. Preparation and properties of magnetic nano- and microsized particles for biological and environmental separations. J. Sep. Sci. 2007, 30, 1751–1772. [Google Scholar] [CrossRef]
  3. Schaller, V.; Kräling, U.; Rusu, C.; Petersson, K.; Wipenmyr, J.; Krozer, A.; Wahnström, G.; Sanz-Velasco, A.; Enoksson, P.; Johansson, C. Motion of nanometer sized magnetic particles in a magnetic field gradient. J. Appl. Phys. 2008, 104, 93918. [Google Scholar] [CrossRef]
  4. Ruffert, C. Magnetic Bead-Magic Bullet. Micromachines 2016, 7, 21. [Google Scholar] [CrossRef] [Green Version]
  5. Wang, Z.; Wang, L.; Brown, S.I.; Frank, T.G.; Cuschieri, A. Ferromagnetization of target tissues by interstitial injection of ferrofluid: Formulation and evidence of efficacy for magnetic retraction. IEEE Trans. Biomed. Eng. 2009, 56, 2244–2252. [Google Scholar] [CrossRef]
  6. Pankhurst, Q.A.; Connolly, J.; Jones, S.K.; Dobson, J. Applications of magnetic nanoparticles in biomedicine. J. Phys. D Appl. Phys. 2003, 36, R167–R181. [Google Scholar] [CrossRef] [Green Version]
  7. Philippova, O.; Barabanova, A.; Molchanov, V.; Khokhlov, A. Magnetic polymer beads: Recent trends and developments in synthetic design and applications. Eur. Polym. J. 2011, 47, 542–559. [Google Scholar] [CrossRef] [Green Version]
  8. Yoon, T.-J.; Kim, J.S.; Kim, B.G.; Yu, K.N.; Cho, M.-H.; Lee, J.-K. Multifunctional nanoparticles possessing a “magnetic motor effect” for drug or gene delivery. Angew. Chem. Int. Engl. 2005, 44, 1068–1071. [Google Scholar] [CrossRef] [PubMed]
  9. Williams, H.M. The application of magnetic nanoparticles in the treatment and monitoring of cancer and infectious diseases. Biosci. Horiz. Int. J. Stud. Res. 2017, 10, 257. [Google Scholar] [CrossRef] [Green Version]
  10. Alexiou, C.; Diehl, D.; Henninger, P.; Iro, H.; Rockelein, R.; Schmidt, W.; Weber, H. A High Field Gradient Magnet for Magnetic Drug Targeting. IEEE Trans. Appl. Supercond. 2006, 16, 1527–1530. [Google Scholar] [CrossRef]
  11. Tenório-Neto, E.T.; Jamshaid, T.; Eissa, M.; Kunita, M.H.; Zine, N.; Agusti, G.; Fessi, H.; El-Salhi, A.E.; Elaissari, A. TGA and magnetization measurements for determination of composition and polymer conversion of magnetic hybrid particles. Polym. Adv. Technol. 2015, 26, 1199–1208. [Google Scholar] [CrossRef]
  12. Liu, C.; Stakenborg, T.; Peeters, S.; Lagae, L. Cell manipulation with magnetic particles toward microfluidic cytometry. J. Appl. Phys. 2009, 105, 102014. [Google Scholar] [CrossRef]
  13. Aziz, T.; Masum, S.; Qadir, M.; Gafur, A.; Huq, D. Physicochemical Characterization of Iron Oxide Nanoparticle Coated with Chitosan for Biomedical Application. IRJPAC 2016, 11, 1–9. [Google Scholar] [CrossRef]
  14. Moreland, J.; Nakashima, Y.; Alldredge, J.W.; Zabow, G. Novel methods for in situ characterization of individual micro- and nanoscale magnetic particles. MRS Bull. 2013, 38, 933–937. [Google Scholar] [CrossRef] [Green Version]
  15. Alldredge, J.W.; Moreland, J. Magnetic particle imaging with a cantilever detector. J. Appl. Phys. 2012, 112, 23905. [Google Scholar] [CrossRef]
  16. Draack, S.; Viereck, T.; Nording, F.; Janssen, K.-J.; Schilling, M.; Ludwig, F. Determination of dominating relaxation mechanisms from temperature-dependent Magnetic Particle Spectroscopy measurements. J. Magn. Magn. Mater. 2019, 474, 570–573. [Google Scholar] [CrossRef]
  17. Draack, S.; Lucht, N.; Remmer, H.; Martens, M.; Fischer, B.; Schilling, M.; Ludwig, F.; Viereck, T. Multiparametric Magnetic Particle Spectroscopy of CoFe 2 O 4 Nanoparticles in Viscous Media. J. Phys. Chem. C 2019, 123, 6787–6801. [Google Scholar] [CrossRef]
  18. van de Loosdrecht, M.M.; Draack, S.; Waanders, S.; Schlief, J.G.L.; Krooshoop, H.J.G.; Viereck, T.; Ludwig, F.; Ten Haken, B. A novel characterization technique for superparamagnetic iron oxide nanoparticles: The superparamagnetic quantifier, compared with magnetic particle spectroscopy. Rev. Sci. Instrum. 2019, 90, 24101. [Google Scholar] [CrossRef] [Green Version]
  19. Viereck, T.; Draack, S.; Schilling, M.; Ludwig, F. Multi-spectral Magnetic Particle Spectroscopy for the investigation of particle mixtures. J. Magn. Magn. Mater. 2019, 475, 647–651. [Google Scholar] [CrossRef]
  20. Llandro, J.; Palfreyman, J.J.; Ionescu, A.; Barnes, C.H.W. Magnetic biosensor technologies for medical applications: A review. Med. Biol. Eng. Comput. 2010, 48, 977–998. [Google Scholar] [CrossRef]
  21. Ejsing, L.; Hansen, M.F.; Menon, A.K.; Ferreira, H.A.; Graham, D.L.; Freitas, P.P. Magnetic microbead detection using the planar Hall effect. J. Magn. Magn. Mater. 2005, 293, 677–684. [Google Scholar] [CrossRef]
  22. Moreland, J. Micromechanical instruments for ferromagnetic measurements. J. Phys. D: Appl. Phys. 2003, 36, R39. [Google Scholar] [CrossRef]
  23. Micromod GmbH. Magnetic Polystyrene Particles. Available online: https://www.micromod.de/en/produkte-11-magnetic_psm.html (accessed on 29 June 2021).
  24. Micromod GmbH. Iron Oxide Particles. Available online: https://www.micromod.de/pdf/generate_pdf.php?article_number=45-00-252&tds_en=true (accessed on 29 June 2021).
  25. Graham, D.L.; Ferreira, H.; Bernardo, J.; Freitas, P.P.; Cabral, J.M.S. Single magnetic microsphere placement and detection on-chip using current line designs with integrated spin valve sensors: Biotechnological applications. J. Appl. Phys. 2002, 91, 7786. [Google Scholar] [CrossRef]
  26. Cardarelli, F. Materials Handbook; Springer: Cham, Germany, 2018; ISBN 978-3-319-38923-3. [Google Scholar]
  27. Hopcroft, M.A.; Nix, W.D.; Kenny, T.W. What is the Young′s Modulus of Silicon? J. Microelectromech. Syst. 2010, 19, 229–238. [Google Scholar] [CrossRef] [Green Version]
  28. Stipe, B.C.; Mamin, H.J.; Stowe, T.D.; Kenny, T.W.; Rugar, D. Magnetic dissipation and fluctuations in individual nanomagnets measured by ultrasensitive cantilever magnetometry. Phys. Rev. Lett. 2001, 86, 2874–2877. [Google Scholar] [CrossRef]
  29. Ono, T.; Esashi, M. Magnetic force and optical force sensing with ultrathin silicon resonator. Appl. Phys. Lett. 2003, 74, 5141–5146. [Google Scholar] [CrossRef]
  30. Punyabrahma, P.; Jayanth, G.R. A magnetometer for estimating the magnetic moment of magnetic micro-particles. Rev. Sci. Instrum. 2017, 88, 15008. [Google Scholar] [CrossRef]
  31. Behrens, I.; Doering, L.; Peiner, E. Piezoresistive cantilever as portable micro force calibration standard. J. Micromech. Microeng. 2003, 13, S171–S177. [Google Scholar] [CrossRef]
  32. Nyang’au, W.O.; Setiono, A.; Schmidt, A.; Bosse, H.; Peiner, E. Sampling and Mass Detection of a Countable Number of Microparticles Using on-Cantilever Imprinting. Sensors 2020, 20, 2508. [Google Scholar] [CrossRef]
  33. Nyang’au, W.O.; Setiono, A.; Bertke, M.; Bosse, H.; Peiner, E. Cantilever-Droplet-Based Sensing of Magnetic Particle Concentrations in Liquids. Sensors 2019, 19, 4758. [Google Scholar] [CrossRef] [Green Version]
  34. Webcraft GmbH. Data Sheet Article S-05-03-N. Available online: https://www.supermagnete.de/eng/disc-magnets-neodymium/disc-magnet-5mm-3mm_S-05-03-N (accessed on 30 June 2021).
  35. Meeker, D. Finite Element Method Magnetics, Version 4.2 (21 April 2019 Build); 2015. Available online: https://www.femm.info/wiki/Files/files.xml?action=download&file=femm42bin_x64_21Apr2019.exe (accessed on 30 June 2021).
  36. Elrefai, A.L.; Enpuku, K.; Yoshida, T. Effect of easy axis alignment on dynamic magnetization of immobilized and suspended magnetic nanoparticles. J. Appl. Phys. 2021, 129, 93905. [Google Scholar] [CrossRef]
  37. Schmid, S.; Villanueva, L.G.; Roukes, M.L. Fundamentals of Nanomechanical Resonators; Springer International Publishing: Cham, Switzerland, 2016; ISBN 978-3-319-28691-4. [Google Scholar]
  38. Quantum Design GmbH. MPMS3 SQUID Magnetometer. Available online: https://qd-europe.com/de/en/product/mpms3-squid-magnetometer/ (accessed on 14 December 2020).
  39. Kahmann, T.; Ludwig, F. Magnetic field dependence of the effective magnetic moment of multi-core nanoparticles. J. Appl. Phys. 2020, 127, 233901. [Google Scholar] [CrossRef]
  40. Friedrich, R.-M.; Zabel, S.; Galka, A.; Lukat, N.; Wagner, J.-M.; Kirchhof, C.; Quandt, E.; McCord, J.; Selhuber-Unkel, C.; Siniatchkin, M.; et al. Magnetic particle mapping using magnetoelectric sensors as an imaging modality. Sci. Rep. 2019, 9, 2086. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Experimental setup for cantilever stiffness calibration: (a) schematics and (b) photograph.
Figure 1. Experimental setup for cantilever stiffness calibration: (a) schematics and (b) photograph.
Chemosensors 09 00207 g001
Figure 2. (a) Commercial cantilever chip (CAN30-1-2, from CiS Forschungsinstitut für Mikrosensorik GmbH, Erfurt, Germany) glued onto an in-house fabricated carrier chip (with contact pads labelling 1–10) and then wire bonded. (b) The prepared cantilever sensor was mounted onto the cantilever holder PCB and contacted using metallic pins (fixed into the PCB). (c,d) are the SEM images of the adsorbate (deposited magnetic polystyrene particles).
Figure 2. (a) Commercial cantilever chip (CAN30-1-2, from CiS Forschungsinstitut für Mikrosensorik GmbH, Erfurt, Germany) glued onto an in-house fabricated carrier chip (with contact pads labelling 1–10) and then wire bonded. (b) The prepared cantilever sensor was mounted onto the cantilever holder PCB and contacted using metallic pins (fixed into the PCB). (c,d) are the SEM images of the adsorbate (deposited magnetic polystyrene particles).
Chemosensors 09 00207 g002
Figure 3. Magnet holder assembly. Typical designs(ac) and fabricated samples (d,e).
Figure 3. Magnet holder assembly. Typical designs(ac) and fabricated samples (d,e).
Chemosensors 09 00207 g003
Figure 4. (a) FEMM simulation contour plot of the magnetic field distribution of two identical repelling permanent disk magnets (with radius rm ≈ 2.5 mm and length lm ≈ 3 mm). The axisymmetric (central) axis of each magnet was along its length lm—through the z-axis of the FEMM coordinate system. The magnetic fields also distributed radially outwards (along the radial axis r). The origin of the coordinate system was at point zero “0”, and it corresponded to the position of near-zero field. (b,c) are the magnetic induction B (black line) and its second derivatives δ2Βz2 and δ2Br2 (red line) as a function of the distance from the magnets’ pole face along the symmetric axis and the radial axis, respectively. The identical magnets used here were 2 mm apart.
Figure 4. (a) FEMM simulation contour plot of the magnetic field distribution of two identical repelling permanent disk magnets (with radius rm ≈ 2.5 mm and length lm ≈ 3 mm). The axisymmetric (central) axis of each magnet was along its length lm—through the z-axis of the FEMM coordinate system. The magnetic fields also distributed radially outwards (along the radial axis r). The origin of the coordinate system was at point zero “0”, and it corresponded to the position of near-zero field. (b,c) are the magnetic induction B (black line) and its second derivatives δ2Βz2 and δ2Br2 (red line) as a function of the distance from the magnets’ pole face along the symmetric axis and the radial axis, respectively. The identical magnets used here were 2 mm apart.
Chemosensors 09 00207 g004
Figure 5. A plot of maximum second derivative magnetic induction field obtained from FEMM simulation with respect to magnet poles spacing (d12). Here, two identical disk NdFeB magnets were repulsively arranged face-to-face (2 mm apart) and the δ2Βz2|max and δ2Βr2|max were determined along symmetric (z) and the radial (r) axes, respectively.
Figure 5. A plot of maximum second derivative magnetic induction field obtained from FEMM simulation with respect to magnet poles spacing (d12). Here, two identical disk NdFeB magnets were repulsively arranged face-to-face (2 mm apart) and the δ2Βz2|max and δ2Βr2|max were determined along symmetric (z) and the radial (r) axes, respectively.
Chemosensors 09 00207 g005
Figure 6. Magnetic force gradient detection setup, mainly comprising of a cantilever sensor (plus aligned magnetic particles), lock-in amplifier and pair of neodymium permanent magnets.
Figure 6. Magnetic force gradient detection setup, mainly comprising of a cantilever sensor (plus aligned magnetic particles), lock-in amplifier and pair of neodymium permanent magnets.
Chemosensors 09 00207 g006
Figure 7. Frequency response of a bare (black line) and magnetic particle-laden (red line) cantilever (S80/CAN30-1-2) under an external magnetic force field gradient induced by a pair of NdFeB permanent magnets (2 mm apart). The observed frequencies were monitored over time under phase-locked loop (PLL) condition when the field was switched on (Bon) and off (Boff). The test sample consisted of 2 μm-sized magnetic polystyrene particles. The cantilever sensor was excited in an out-of-plane mode (along the z-axis) and the magnetizing field was induced parallel with the excitation direction.
Figure 7. Frequency response of a bare (black line) and magnetic particle-laden (red line) cantilever (S80/CAN30-1-2) under an external magnetic force field gradient induced by a pair of NdFeB permanent magnets (2 mm apart). The observed frequencies were monitored over time under phase-locked loop (PLL) condition when the field was switched on (Bon) and off (Boff). The test sample consisted of 2 μm-sized magnetic polystyrene particles. The cantilever sensor was excited in an out-of-plane mode (along the z-axis) and the magnetizing field was induced parallel with the excitation direction.
Chemosensors 09 00207 g007
Figure 8. (a) Frequency shift response of a cantilever sensor (S88/CAN15-3-2, f0 = 22.46 kHz) containing 11.1 ± 0.7 ng of 250 nm-sized iron oxide particles. Here, the baseline differences between a blank and particle-laden cantilever was supposedly due to variance in ambient temperature. (b) Observed frequency time response under PLL condition after exposure to external magnetic field gradient, in an on-state (Bon) and off-state (Boff). In this case, a pair of two permanent magnets was placed 2 mm apart in a face-to-face arrangement as shown in Figure 6. The visible further decrease in f0 in the Bon state was assigned to a temperature rise by induced currents in the cantilever metallization. After switching-off the field, temperature and; consequently, f0 tended to level out to the initial condition at Boff (marked by the horizontal blue dashed line, see inset).
Figure 8. (a) Frequency shift response of a cantilever sensor (S88/CAN15-3-2, f0 = 22.46 kHz) containing 11.1 ± 0.7 ng of 250 nm-sized iron oxide particles. Here, the baseline differences between a blank and particle-laden cantilever was supposedly due to variance in ambient temperature. (b) Observed frequency time response under PLL condition after exposure to external magnetic field gradient, in an on-state (Bon) and off-state (Boff). In this case, a pair of two permanent magnets was placed 2 mm apart in a face-to-face arrangement as shown in Figure 6. The visible further decrease in f0 in the Bon state was assigned to a temperature rise by induced currents in the cantilever metallization. After switching-off the field, temperature and; consequently, f0 tended to level out to the initial condition at Boff (marked by the horizontal blue dashed line, see inset).
Chemosensors 09 00207 g008
Figure 9. (a) Observed frequency responses under various magnetic force gradient conditions (magnets spacing) for a cantilever (S84/CAN30-1-2) attached with magnetic particles (Fe3O4 NPs). The used cantilever had k0 ≈ 2.85 ± 0.19 N/m and f0 ≈ 4.148 kHz, and it was excited in an out-of-plane mode (same direction as the induced external magnetic field and second its derivative). (b) Average shift in resonant frequency with magnets spacing (d12) for a blank cantilever (open red circles) and effective resonant frequency shifts (filled red circles). (c) Magnetic force gradient F (Equation (16)) versus the maximum second derivative magnetic field gradient (determined by FEMM). The linear fitting in (c) was performed for d12 = 2 mm to 9 mm; its slope was equivalent to the particles’ magnetic moment (→ Equation (7)). The values of the polarizing field and its second derivative depended strongly on the positioning accuracy of the cantilever, which had to be performed against the risk of breaking the cantilever. Consequently, since positioning at δ2Bz2|max was much more difficult at decreased spacing, a deviation from linearity was observed.
Figure 9. (a) Observed frequency responses under various magnetic force gradient conditions (magnets spacing) for a cantilever (S84/CAN30-1-2) attached with magnetic particles (Fe3O4 NPs). The used cantilever had k0 ≈ 2.85 ± 0.19 N/m and f0 ≈ 4.148 kHz, and it was excited in an out-of-plane mode (same direction as the induced external magnetic field and second its derivative). (b) Average shift in resonant frequency with magnets spacing (d12) for a blank cantilever (open red circles) and effective resonant frequency shifts (filled red circles). (c) Magnetic force gradient F (Equation (16)) versus the maximum second derivative magnetic field gradient (determined by FEMM). The linear fitting in (c) was performed for d12 = 2 mm to 9 mm; its slope was equivalent to the particles’ magnetic moment (→ Equation (7)). The values of the polarizing field and its second derivative depended strongly on the positioning accuracy of the cantilever, which had to be performed against the risk of breaking the cantilever. Consequently, since positioning at δ2Bz2|max was much more difficult at decreased spacing, a deviation from linearity was observed.
Chemosensors 09 00207 g009
Figure 10. Total magnetic moment versus applied magnetic induction field (B = µ0H, spanning from –7 T to 7 T) for: (a) Magnetic polystyrene beads (2 µm in size), and (b) iron oxide NPs (250 nm in size). The sinusoidal magnetic induction field was applied under a stable temperature environment (300 K).
Figure 10. Total magnetic moment versus applied magnetic induction field (B = µ0H, spanning from –7 T to 7 T) for: (a) Magnetic polystyrene beads (2 µm in size), and (b) iron oxide NPs (250 nm in size). The sinusoidal magnetic induction field was applied under a stable temperature environment (300 K).
Chemosensors 09 00207 g010
Table 1. Properties of magnetic particle samples [23,24].
Table 1. Properties of magnetic particle samples [23,24].
Property/ParameterMagnetic Polystyrene ParticlesIron Oxide NPs
Particle size (nominal diameter), dp2 µm250 nm
Particle volume, Vp (π/6 × dp3)4.19 × 10−18 m38.18 × 10−21 m3
Density of polystyrene, ρPS1070 kg/m3 [26]-
Density of magnetic material (magnetite), ρFe3O45350 kg/m3 5350 kg/m3
Effective particle density, ρeff, Equation (3) 1100 kg/m3 5350 g/m3
Volume of magnetite per particle, Vp,Fe3O4 1.43 × 10−19 m38.18 × 10−21 m3
Mass per particle, mp, Equation (2) 5.1 × 10−15 kg4.4 × 10−17 kg
Mass of magnetite per particle, mp,Fe3O47.6 × 10−16 kg4.4 × 10−17 kg
Particles mass concentration, Cp,m2.2 × 1014 MPs per kg2.3 × 1016 MPs per kg
Saturation mass magnetization, Mm,s (at H = 800 kA/m) 6.5 A m2/kg 51 A m2/kg
Magnetic moment per particle, µp (Equation (1), Mm = Mm,s)33 fA m22 fA m2
  M m = M m , s * C f where Mm,s* = 71 A m2/kg iron [24], and Cf is the conversion factor from iron to magnetite: C f = Molar   mass   of   magnetite 3   ×   Molar   mass   of   iron = 231.533   g / mol 3   ×   55.845   g / mol = 1.39 .
Table 2. Resonant MEMS cantilever sensor characteristics.
Table 2. Resonant MEMS cantilever sensor characteristics.
ParameterCantilever Type
CAN50-2-5CAN30-1-2CAN15-3-2
m0 (µg)116.5 ± 2.317.48 ± 0.072.62 ± 0.17
f0 (Hz)2400.0 ± 0.34377.0 ± 0.222,462.4 ± 0.1
k0 (N/m), Equation (4)8.5 ± 0.32.4 ± 0.210.1 ± 0.8
k0 (N/m), Equation (5)6.4 ± 0.13.2 ± 0.112.5 ± 0.8
Q379 ± 3432 ± 12768 ± 50
σ (ppm)20.81.54.1
fmin (mHz)50 ± 106.7 ± 1.093 ± 1
mmin (pg), Equation (8)1164 ± 2412.8 ± 0.55.2 ± 0.4
μmin (pA m2), Equation (9)169 ± 346 ± 166 ± 4
Table 3. Magnetic force gradient under 2 mm magnets spacing (with δ2Bz2|max ≈ 1.57 T/mm2) from assorted cantilever sensors with different resonance attributes and attached particle masses.
Table 3. Magnetic force gradient under 2 mm magnets spacing (with δ2Bz2|max ≈ 1.57 T/mm2) from assorted cantilever sensors with different resonance attributes and attached particle masses.
Cantilever ID/Type (MPs)f0
(Hz)
Δf0
(Hz)
Δm
× 10−9 (g)
k0
(N/m)
Δf0
(Hz)
F’(z)
(mN/m)
S97/CAN30-1-2 (Micromer)4667.75 ± 0.02−1.87 ± 0.043.79 ± 0.263.61 ± 0.24−0.01 ± 0.010.02 ± 0.01
S80/CAN30-1-2 (Micromer)4377.08 ± 0.01−3.72 ± 0.028.22 ± 0.544.46 ± 0.09 −0.02 ± 0.010.04 ± 0.02
3.17 ± 0.130.03 ± 0.02
2.45 ± 0.20 0.02 ± 0.01
S95/CAN30-1-2 (Micromer)4611.90 ± 0.05−10.51 ± 0.0523.2 ± 1.53.52 ± 0.23−0.11 ± 0.020.17 ± 0.04
S89/CAN15-3-2 (Fe3O4)21,448.63 ± 0.06−18.50 ± 0.061.63 ± 0.1111.43 ± 0.75−0.08 ± 0.030.09 ± 0.03
S88/CAN15-3-2 (Fe3O4)22,465.52 ± 0.12−131.67 ± 0.1311.06 ± 0.7312.53 ± 0.82−4.06 ± 0.204.53 ± 0.94
S84/CAN30-1-2 (Fe3O4)4148.50 ± 0.01−84.90 ± 0.01230 ± 202.85 ± 0.19−3.89 ± 0.125.34 ± 0.39
from load-deflection calibration curves. from Equation (4).
Table 4. Magnetic force gradient (under 2 mm magnets spacing) and derived magnetic moment per magnetic particle.
Table 4. Magnetic force gradient (under 2 mm magnets spacing) and derived magnetic moment per magnetic particle.
Cantilever ID/Type (MPs)F’(z0)
(mN/m)
Σµ
(pA m2)
Npµ per MP (fA m2)
by CantileverNominalby SQUID
S97/CAN30-1-2 (Micromer)0.02 ± 0.0110 ± 6945 ± 6511 ± 733116.6 ± 0.5
120.8 ± 0.4
S80/CAN30-1-2 (Micromer)0.04 ± 0.0228 ± 152050 ± 13514 ± 8
S95/CAN30-1-2 (Micromer)0.17 ± 0.04110 ± 215792 ± 38219 ± 4
S88/CAN15-3-2 (Fe3O4)4.53 ± 0.942887 ± 602(254 ± 21) × 10311 ± 322.379 ± 0.001
2.571 ± 0.013
S89/CAN15-3-2 (Fe3O4)0.09 ± 0.0355 ± 21(37 ± 3) × 1031.5 ± 0.8
S84/CAN30-1-2 (Fe3O4)5.34 ± 0.393402 ± 247(4506 ± 372) × 1030.75 ± 0.08
3761 ± 526 *0.83 ± 0.1
* From the slope of F’ vs. δ2Bz2 in Figure 9c, based on Equation (7); In determining stiffness k0, the full cantilever length L was used here. However, since particles were not concentrated at the free end of the cantilever a smaller length (corresponding to the average position of the particles along L) would yield a larger stiffness keff. Consequently, a larger magnetic moment which ultimately depicts a better agreement with the nominal value would be expected; at 800 kA/m (= 1 T); at 1592 kA/m (= 2 T).
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Nyang’au, W.O.; Kahmann, T.; Viereck, T.; Peiner, E. MEMS-Based Cantilever Sensor for Simultaneous Measurement of Mass and Magnetic Moment of Magnetic Particles. Chemosensors 2021, 9, 207. https://doi.org/10.3390/chemosensors9080207

AMA Style

Nyang’au WO, Kahmann T, Viereck T, Peiner E. MEMS-Based Cantilever Sensor for Simultaneous Measurement of Mass and Magnetic Moment of Magnetic Particles. Chemosensors. 2021; 9(8):207. https://doi.org/10.3390/chemosensors9080207

Chicago/Turabian Style

Nyang’au, Wilson Ombati, Tamara Kahmann, Thilo Viereck, and Erwin Peiner. 2021. "MEMS-Based Cantilever Sensor for Simultaneous Measurement of Mass and Magnetic Moment of Magnetic Particles" Chemosensors 9, no. 8: 207. https://doi.org/10.3390/chemosensors9080207

APA Style

Nyang’au, W. O., Kahmann, T., Viereck, T., & Peiner, E. (2021). MEMS-Based Cantilever Sensor for Simultaneous Measurement of Mass and Magnetic Moment of Magnetic Particles. Chemosensors, 9(8), 207. https://doi.org/10.3390/chemosensors9080207

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop