Previous Article in Journal
Flexipede: A Bio-Inspired, Modular Myriapod Robot for Rough-Terrain Traversal
Previous Article in Special Issue
Gynecological Tendon-Driven Continuum Robots: Design and Experimentation
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Experimental Validation of ASSIST-FEEv3 Elbow Assisting Device with Physiotherapy Considerations

by
Cuauhtémoc Morales-Cruz
1,
Fortunato Frisina
2,
Francesco Scerbo
2,
Rocco Mazzotta
2 and
Marco Ceccarelli
1,*
1
LARM2: Laboratory of Robot Mechatronics, University of Rome Tor Vergata, 00133 Rome, Italy
2
Department of Biomedicine and Prevention, University of Rome Tor Vergata, 00133 Rome, Italy
*
Author to whom correspondence should be addressed.
Robotics 2026, 15(7), 130; https://doi.org/10.3390/robotics15070130
Submission received: 28 May 2026 / Revised: 1 July 2026 / Accepted: 2 July 2026 / Published: 3 July 2026

Abstract

Upper-limb rehabilitation and elderly exercise programs require lightweight, reliable, and physiotherapy-oriented assistive technologies capable of supporting controlled joint motion while enabling objective performance assessment. This paper presents experimental validation of ASSIST-FEEv3, a cable-driven elbow assisting device that is designed for flexion–extension exercises with emphasis on usability, portability, and physiotherapy integration. The device employs a dual-cable antagonistic mechanism that is actuated by servomotors housed in a compact module, allowing guided motion in the arm sagittal plane with minimal wearable load mass. A testing campaign was conducted with 25 healthy volunteers under the supervision of physiotherapy experts following a properly designed protocol for three sessions of ten repetitions each. Joint kinematics was acquired through integrated sensing, and performance metrics including maximum flexion, maximum extension, and range of motion (ROM) were analyzed to assess repeatability, motion smoothness, and user-specific variability. The results demonstrate consistent motion assistance across repeated cycles, variability between sessions, and comparable ROM distributions between sexes. Observed deviations were considered due to individual temporary conditions rather than device-related limitations. The device operated with low energy consumption as required in home-based applications. Test findings validate both the mechanical reliability and the physiotherapy-oriented operational framework of the ASSIST-FEEv3 device.

1. Introduction

The global demographic transition is causing a rapid increase in the proportion of elderly individuals within the overall population. This structural change presents significant challenges to healthcare systems, particularly in maintaining autonomy, mobility, and quality of life among aging citizens [1,2]. Preserving the functional independence of elderly people requires structured physical activity programs that are designed to sustain muscle strength, joint mobility, and neuromuscular coordination. Evidence in the large literature highlights the effectiveness of resistance-based and functional training interventions in mitigating age-related decline and reducing the risk of frailty, as pointed out, for example, in [3,4]. Practical guidelines also emphasize the importance of implementing accessible exercise routines in both community and home settings to ensure continuity of care and adherence [5].
Frailty is a multidimensional clinical condition that is characterized by increased vulnerability to adverse outcomes, such as falls, hospitalization, and loss of independence [6]. Identifying and quantifying frailty indicators are essential for customizing motion exercise, rehabilitation interventions, and monitoring progress over time. Therefore, technological solutions that assist movement while enabling measurable and controlled exercise execution are becoming increasingly important, especially for elderly populations who need safe, lightweight, low-cost, and user-friendly devices. The increasing clinical demand for effective upper-limb rehabilitation and motion assistance has driven extensive research into wearable robotic systems, particularly exoskeletons and exosuits [7,8,9,10]. Over the past decade, a substantial amount of work has explored design methodologies, actuation strategies, control paradigms, and human–robot interaction approaches specifically for upper-limb rehabilitation devices. Comprehensive review studies provide valuable insights into existing developments and the key challenges that continue to shape this field, as discussed in [7,8,9,10]. Recent reviews have highlighted the diversity of mechanical architectures and interactive design strategies that are employed in upper-limb exoskeletons. Dhatrak et al. [7] present a comprehensive analysis of more than thirty existing upper-limb exoskeleton prototypes, focusing on interactive design principles, human-in-the-loop control, and challenges related to ergonomics, adaptability, and usability. Their work underscores the multidisciplinary nature of exoskeleton development and emphasizes the need for systems that better integrate physiotherapy requirements with user-centered design. In addition, increasing attention has been devoted to soft robotic wearable devices that can be designed to overcome the limitations associated with rigid exoskeletons. A systematic review by Bardi et al. [8] examines upper-limb exosuits that are based on cable-driven and pneumatic actuation, covering more than fifty systems. The authors highlight the advantages of soft, lightweight solutions in terms of comfort, portability, and preservation of natural joint motion, while also identifying challenges that are related to force transmission, control accuracy, and repeatability in rehabilitation contexts. Focusing specifically on the elbow joint, Supriyono et al. [9] provide a comprehensive review of elbow exoskeletons, classifying devices based on mechanical structure, actuation methods, and sensing technologies. Their analysis reveals recurring trade-offs among mechanical simplicity, assistance capability, and user comfort, emphasizing the importance of aligning device kinematics with physiological joint motion. Complementing this, Gonçalves et al. [10] review robotic interfaces developed for post-stroke upper-limb rehabilitation, examining assistance strategies, actuation methods, and control mechanisms. Their review identifies a clear trend toward lightweight, modular, and portable solutions, while also highlighting persistent barriers to clinical translation and widespread adoption. Collectively, these reviews demonstrate strong and sustained research interest in upper-limb and elbow assisting devices while also highlighting unresolved challenges related to portability, adaptability, and integration within physiotherapy-oriented rehabilitation frameworks, although applications specifically targeting elderly healthcare have not yet been thoroughly considered [2]. These gaps underscore the need for continued development and experimental validation of compact, lightweight, and user-centered assisting devices capable of supporting functional rehabilitation beyond traditional clinical settings. Moreover, recent biomechanical works suggest that upper-limb assistive and prosthetic systems may influence whole-body movement patterns beyond the assisted limb itself. For instance, the use of customized upper-limb prostheses has been shown to modify gait symmetry, plantar pressure distribution, and compensatory loading strategies, highlighting the complex interactions between upper-limb assistance and overall motor function [11]. These findings further motivate the need for comprehensive experimental evaluations of assistive devices within rehabilitation-oriented frameworks.
In response to specific needs, extensive research has focused on developing upper-limb rehabilitation robots and exoskeletons. Early clinically oriented systems, such as ARMin, introduced patient-cooperative arm therapy through semi-exoskeleton kinematics combined with multimodal feedback, demonstrating the potential of robot-assisted intensive training [12]. Highly articulated rigid exoskeletons, such as ANYexo 2.0, offer full joint actuation and enable joint-oriented, task-based training across various rehabilitation stages [13]. Self-aligning shoulder mechanisms have been proposed to enhance kinematic compatibility and to reduce the effects of joint misalignment [14], while clinically evaluated systems like AGREE have demonstrated therapeutic effectiveness through randomized controlled trials [15]. Lightweight anthropomorphic exoskeletons that integrate series elastic actuators and Bowden cable transmissions, such as CURER [16], aim to improve compliance and comfort while maintaining torque accuracy. Additionally, hybrid drive solutions combining cable and linkage mechanisms have been introduced to enhance force transmission and decoupling in cable-driven architectures like the one in [17].
In addition to rigid-link systems, cable-driven and wire-based devices have been explored to reduce inertia and enhance intrinsic safety. CAREX demonstrated the early feasibility of cable-based force-field generation for neural rehabilitation [18], while Bowdendriven soft exoskeletons focused on minimizing interaction forces through optimized structural configurations [19]. Earlier wire-driven robots, such as NeReBot and its evolution MariBot, emphasized lightweight construction and an enlarged workspace while maintaining safe human–robot interaction [20]. More recent developments include soft and hybrid exosuits that reduce muscle activation during assisted tasks [21,22,23], underactuated systems for adaptive arm de-weighting [24], and accessible open-source actuator solutions such as OpenSEA [25]. Affordable, elbow-focused devices like ARDEL further demonstrate the interest in compact and low-cost active-assist solutions [26]. Current accessible solutions also include compact and low-cost elbow rehabilitation tools fabricated through 3D printing technologies, such as the ERT device, which integrates basic actuation and motion analysis functionalities for structured exercise programs [27]. Collectively, these contributions reveal sustained efforts toward balancing biomechanical performance, portability, compliance, and user comfort. Recent studies have also emphasized the importance of objective motion assessment and sensor integration in upper-limb assistive technologies. For example, inertial-magnetic measurement units have been employed to quantify movement quality and intersegmental coordination during activities of daily living, revealing compensatory strategies and reduced movement smoothness in prosthesis users [28]. Similarly, sensorized and individualized upper-limb prostheses fabricated through additive manufacturing have demonstrated the value of integrating force and motion measurements for functional assessment and iterative design improvements [29].
Within this context, the team at the LARM2 Laboratory of Robotics and Mechatronics has worked out activities to develop new cable-driven elbow assisting devices. The CADEL mechanism introduced a portable, low-cost elbow assistance concept that is based on a cable-driven parallel architecture, whose performance was validated through both simulation and experimental testing [30]. Subsequent refinements resulted in the L-CADEL series, featuring structural redesign, lightweight construction, and digital interfacing to better support elderly-oriented home exercise applications [31]. Laboratory evaluations of intermediate prototypes demonstrated satisfactory efficiency while highlighting areas needing ergonomic and functional improvements [32]. A systematic analysis of requirements for elderly motion-assisting systems further clarified critical factors such as modularity, ease of wearing, adaptability, and safe interaction [2,33]. Related developments within the same research framework include the AlmatyExoElbow device, a lightweight antagonistic cable-driven exoskeleton integrating sensor-based motion tracking and experimental validation, demonstrating the feasibility of compact and modular elbow assistance architectures [34].
These findings and further challenges motivated the structured research activities of the ASSIST project [35], aimed to develop modular, portable cable-driven exoskeletons both for elderly exercise and rehabilitation in home use that were validated through significant experimental campaigns. Within this framework, the ASSIST-FEE device family was designed for elbow flexion–extension assistance and evolved through three successive prototypes [36]. The progression from versions v1 and v2 to ASSIST-FEEv3 involved refining cable routing patterns, upgrading actuators, and iteratively repositioning the extension actuation point, resulting in improved torque transmission and smoother operation. A distinctive feature of ASSIST-FEEv3 is its extremely low wearable mass of approximately 100 g, significantly lighter than earlier cable-driven systems [18,19] and substantially lighter than other multi-DOF rigid exoskeletons [13]. The ASSIST-FEEv3 device has undergone a structured testing campaign with healthy volunteers under physiotherapist supervision [37,38], extending preliminary investigations and providing a broader assessment of usability and assisted motion performance. The originality of the solution is further recognized through patent registration [39].
Despite extensive research in upper-limb rehabilitation robotics, many studies primarily focus on mechanical design, actuation capabilities, or control strategies, while structured experimental validation under physiotherapy-guided protocols remains relatively limited. Specifically, lightweight cable-driven elbow devices often lack comprehensive evaluations that incorporate therapist-supervised exercise execution, usability within structured rehabilitation sessions, and preliminary assessments of user interaction. Consequently, there is a need for studies that integrate engineering validation with physiotherapy considerations to evaluate feasibility, operational reliability, and suitability for both structured rehabilitation and home-based exercise scenarios.
This work presents the experimental validation of the ASSIST-FEEv3 elbow assisting device, incorporating explicit physiotherapy considerations. It aims to provide systematic evidence of the device’s performance, usability, and potential applicability in supervised rehabilitation settings and elderly home use.

2. Materials and Methods

2.1. Physiotherapy Requirements for Assisted Exercise

Figure 1 shows a conceptual framework outlining the fundamental design and operational requirements for developing a cable-actuated elbow assisting device intended for exercise and rehabilitation applications. These requirements are organized into two complementary categories: design features and operational features. Design requirements define the technical and human-centered principles guiding the mechanical structure, actuation mechanism, sensing integration, and intelligent control features of the hardware. Operational requirements specify the practical conditions necessary to ensure safe, efficient, and user-friendly performance, particularly in home environments where users may operate the device autonomously. Successful home use depends on key aspects related to motion, force, and sensing during exercise execution, including the provision of an appropriate range of motion (ROM) with both active and passive modes measurable through a goniometer, the assessment of muscle strength according to standardized scales such as the Medical Research Council (MRC) scale [40], and the evaluation of motion performance through customized initial measurements of movement speed and accuracy. In addition, anatomical and physiological characteristics should be carefully considered when programming rehabilitation exercises, allowing parameter settings to be tailored according to medical reference criteria. Altogether, these interconnected elements establish a comprehensive framework that integrates biomechanical compatibility, user adaptability, objective assessment, and intelligent interaction to enable effective and comfortable motion assistance during rehabilitation-oriented and physical training.
Within the design requirements, biomechanics concepts focus on the physical compatibility between the device and the human body. It includes concepts such as range of motion, which defines the angular displacement the mechanism can support; payload capacity, referring to the maximum torque or resistance it can safely handle; adjustability, which ensures adaptability to different users and anatomical variations; and compactness, which provides portability and ease of storage.
The user-oriented design features address aspects related to comfort and usability, including ergonomics for natural anatomical attachment, comfort to reduce fatigue, customization for personalized settings, and noise reduction to improve the overall user experience. The sensing and monitoring concepts encompass the device’s capacity to perceive and evaluate user performance through sensors that measure motion and power consumption; precision that ensures accurate data capture; data storage for tracking performance over time; and real-time feedback that provides immediate corrective or adaptive responses. Finally, intelligent integration focuses on incorporating adaptive and connected technologies, including self-calibration for automatic adjustment of parameters; connectivity for communication with external systems or cloud platforms; data elaboration for analysis and interpretation of information; and alerts or notifications to enhance safety and performance.
The operation requirements complement design aspects by defining the essential conditions for practical, efficient application. Safety issues ensure protection against mechanical and electrical risks; portability facilitates transportation and setup; low cost provides accessibility and scalability; wearability guarantees easy donning and removal; sanitation or ease of cleaning ensures proper hygiene and maintenance; and intensity checker allows adjustment of assistance levels to match user needs and therapeutic goals.
Together, these design and operation considerations form a conceptual foundation for developing a functional, adaptive, and user-centered cable-driven motion-assisting device.

2.2. Design and Prototype of ASSIST-FEEv3

Figure 2 presents a conceptual design of the ASSIST-FEEv3 device, along with a practical solution highlighting all components that are required to assist elbow motion. The schemes illustrate the overall device architecture and show how each component interacts to provide motion support. A laptop communicates with the onboard electronics, while a battery powers the ASSIST-FEEv3 prototype. A current sensor (CS) monitors electrical consumption, and a microcontroller (µC) manages data processing and controls the various operating modes. Dedicated motors for flexion and extension (Mf and Me) drive the corresponding pulleys (Pf and Pe), which pull the respective flexion and extension cables. These cables are routed through guiding tubes (Gf and Ge) toward the actuation points (APf and APe), which are strategically positioned at the shoulder support (Ss) and leg support (Ls) to assist elbow motion (E). Attachment points secure the cables to the arm structure, and an inertial measurement unit (IMU), positioned near the wrist, records real-time motion for performance monitoring.
The proposed device may have potential applications in several rehabilitation and exercise contexts, including upper-limb functional recovery after orthopedic injuries, neurological rehabilitation, and general conditioning programs. The possibility of integrating sensor-based feedback into therapeutic exercises could enhance both patient engagement and objective clinical assessment.

2.3. Testing Layout and Modes

The experimental setup with the ASSIST-FEEv3 worn by a volunteer is shown in Figure 3a, while Figure 3b presents the corresponding conceptual layout of the testing configuration. The setup is designed to evaluate both the functional performance of the assisting device and its interaction with a user. It enables synchronized data acquisition from the volunteer and from the integrated sensing system, including signals collected by the control unit from the wrist-mounted inertial measurement unit (IMU). According to the defined testing protocol, the procedure begins with a medical supervisor selecting and configuring the exercise parameters. These settings determine the operating mode of the device. At the current stage of development, no adaptive control algorithm is implemented, but the device operates in a predefined repetition cycle of 28 s. Adaptation to individual user characteristics is therefore managed during the initial fitting phase, when the device is worn, and the exercise parameters are programmed. Once the session starts, the control unit continuously acquires and processes data from both the assisting mechanism actuation and the sensors. The recorded signals are simultaneously transmitted to a laptop/PC via a USB connection, allowing the medical supervisor to monitor the assisted movement in real time and to perform immediate assessments when necessary.
Figure 4 shows the sequence of steps for the experimental procedure, ensuring repeatability across different volunteers in accordance with a specifically designed protocol in accordance with the Declaration of Helsinki and approved by the Institutional Ethics Committee of Policlinico di Tor Vergata, Rome, with protocol code RS. 197.22 on 15 November 2022. This protocol has been designed to validate and assess the functionality of motion devices, particularly in supporting arm and elbow movement within the Italian PNR project ASSIST [35]. The initial two steps are fundamental to the success of the exercise session, especially for this kind of cable-driven system. These first two steps are device preparation and adjustment to a volunteer. Once the guided movement begins, the medical supervisor must monitor real-time data and correlate them with the corresponding arm motion, particularly during the initial stage. As the device operation progresses through the subsequent steps, the need for intervention from a medical supervisor can be reduced, ultimately requiring only a final data review, which can even be conducted remotely.

3. Results

From a clinical rehabilitation perspective, the possibility of objectively quantifying acceleration and power output during exercise may represent a valuable tool for physiotherapists, providing quantitative metrics for monitoring patient performance, exercise adherence, and progression throughout rehabilitation programs.
Figure 5a–c present snapshots of a laboratory test with the ASSIST-FEEv3 device worn by a healthy volunteer. Constructed using commercially available components, as reported in [36,37], the device effectively assists elbow motion. Its performance is experimentally evaluated through the motion, acceleration, and power-related results whose illustrative representative outcomes are presented in Figure 6, Figure 7, Figure 8 and Figure 9. The testing campaign has been planned as per a proof-of-concept validation with 25 healthy student volunteers to check the feasibility of the device and to characterize its operation from the physiotherapist’s perspective. The limited number of volunteers and tests can still be considered suitable for the above targets, as referred to in the physiotherapist testing practice [41].
According to physiotherapy recommendations, the duration for one complete flexion–extension repetition was established at approximately 28 s to ensure a controlled and safe exercise-oriented movement. During the experiments, the volunteer performed the experiments in a standing position, keeping their back close to the wall and following the physiotherapist’s instructions. Throughout the experiments, the arm remained passive, while the ASSIST-FEEv3 device was solely responsible for moving it up and down. This posture was selected to guarantee a proper body configuration for all the volunteers during the exercise and to minimize undesired compensatory movements while using the ASSIST-FEEv3 device.
Figure 5a shows the initial configuration corresponding to the full flexion position, just before the assisting motion starts. The figure also illustrates the IMU reference frame adopted in this work, where the roll angle is measured about the x-axis, indicated in red color, and the pitch angle is measured about the y-axis, indicated in green color. This reference frame configuration is maintained as a reference for testing to acquire different data, as in Figure 6, Figure 7 and Figure 8.
Figure 6 shows the time-domain data of the elbow motion during two extension-flexion repetitions, corresponding to a section of a test duration of approximately 1 min. The measurements were obtained from the IMU mounted on the ASSIST-FEEv3 device, using the reference frame illustrated in Figure 5a, and are presented for a selected volunteer as representative experimental evidence of the obtained motion data. The results describe the range of motion (ROM) that is achieved during the assisted motion exercise, from full flexion to full extension of the elbow joint. In particular, the roll angle exhibits the main variation associated with the flexion–extension motion, ranging approximately from −24.34° to 48°, showing a smooth and continuous assisted motion throughout the exercise repetitions. Although the pitch angle presents comparatively small variations, its value is not negligible. This effect indicates that the elbow joint is allowed to rotate freely during supination–pronation movements while the assistance is provided. Such kinematic adaptability is important because it enables the ASSIST-FEEv3 device to accommodate natural arm motion according to the specific movement needs of the volunteer during the rehabilitation task.
Figure 7 shows the measured linear acceleration components at the wrist together with the corresponding acceleration magnitude as a function of time during the assisted motion. The acceleration signals were obtained according to the IMU reference frame shown in Figure 5a. The reported data further confirm the smooth and continuous motion behavior recorded with data in Figure 6, as indicated by the relatively low acceleration levels throughout the two exercise cycles.
A more representative assessment of the motion quality can be obtained from the free-of-gravity acceleration magnitude, denoted as a f , which is also reported in Figure 8. This parameter allows an evaluation of the dynamic contribution that is associated only with the arm motion by compensating for the gravitational component. The quantity a f is computed as
a f = a x 2 + a y 2 + a z 2 g
where a x , a y , and a z are the measured acceleration components along the IMU axes, and g is the gravitational acceleration. The relatively low values of a f indicate that the assisted motion is performed without abrupt changes in acceleration, which is desirable for physiotherapy-oriented rehabilitation exercises and contributes to user comfort and safe interaction with the ASSIST-FEEv3 device.
The acceleration trends shown in Figure 8 correlate with the changes in motion direction occurring during the flexion and extension phases that are shown in Figure 5, providing additional evidence of the smoothness and stability of the assisted movement. The free-of-gravity acceleration does not reach zero values because the arm motion is continuously maintained throughout the assisting task.
These acceleration values correspond to instantaneous measurements acquired by the inertial sensor and cannot be interpreted as a constant acceleration acting over the entire experiment duration, giving motion displacement, but an evaluation of the motion variability and smoothness. Most of the observed acceleration peaks occur at the transition instants where the movement changes direction from flexion to extension, or vice versa. These variations can also be associated with the concentric and eccentric phases of the upper-limb motion, reflecting the subject’s ability to maintain coordinated and rhythmic motion control during repetitive exercise. Between these transition phases, the acceleration behavior remains quasi-stationary with relatively low values, which can be associated with proper assisted motion performance and adequate recovery of muscle strength while maintaining smooth arm motion.
Figure 9 shows the computed power consumption that is calculated from the measured current supplied to the two servomotors actuating the cable transmission system. The observed power peaks correspond to the inversion points between flexion and extension, where the largest torque is required. The small and acceptable level of noise visible in the acquired signals is very likely attributable to minor physiological tremors of the arm, sensor resolution limits, and mechanical backlash within the transmission mechanism.
Overall, test results from a testing campaign with 25 volunteers demonstrate that the device is capable of assisting a wide range of elbow motion while maintaining low power consumption, with an average current demand of approximately 0.26 A. Such a low current requirement allows operation using lightweight commercial batteries, supporting system portability. These characteristics confirm that ASSIST-FEEv3 is a suitable solution for elbow exercise applications in both home-based and clinical environments.

4. Discussion

Figure 10 shows statistical results from the testing campaign using the ASSIST-FEEv3 device, showing the distribution of flexion–extension repetitions across all participants, grouped by sex. Each panel illustrates the variability and spread of the measurements collected during the study. Each box represents the distribution of repetitions recorded for participants of the corresponding sex, as elaborated from test results with 283 repetitions for female participants (F) and 436 repetitions for male participants (M). The boxes are colored according to participant sex, with red representing females and blue representing males, and the y-axis indicates the corresponding measurement in degrees. Figure 10a shows the maximum flexion values, Figure 10b shows the maximum extension values, and Figure 10c shows the range of motion (ROM), which is calculated as the difference between maximum flexion and extension. The data indicate that female participants exhibited smaller variability when compared with results by males, which is clearly observed in the smaller interquartile ranges (IQRs) across all three plots. The Shapiro–Wilk test indicated that the assumption of normality was not satisfied for at least one of the groups, and non-parametric comparisons between sexes were performed using the Mann–Whitney U test. The analyses were conducted using R version 4.6.0. The results show no statistically significant differences between female and male participants for maximum flexion (p = 0.14), maximum extension (p = 0.071), or ROM (p = 0.35). Therefore, despite the large variability observed among male participants, the central tendency of the kinematic variables remained statistically similar between female and male tested volunteers, suggesting similar functional elbow mobility across sexes. Therefore, the results do not show statistically significant differences between female and male participants. This study, conducted with healthy young adults, serves to validate and characterize the data acquisition and analysis procedures before using the device in elderly populations in future clinical testing campaigns.
Figure 11 shows the distribution of maximum elbow flexion values for each participant, as a function of individual ID and color-coded sex, red for females and blue for males. Each box represents the spread of flexion–extension repetitions recorded for a single participant. According to the testing protocol defined together with the collaborating physiotherapists, each participant was expected to perform a total of 30 repetitions, derived from three separate sessions of 10 repetitions each. This figure provides a visual summary of the variability in maximum flexion performance across all volunteers. The height of each distribution reflects the maximum values achieved, where large distributions indicate better performance, since they correspond to large elbow flexion angles. The plot shows participant-specific differences in both performance and variability, which may reflect individual physical characteristics, engagement levels, or temporary conditions during testing.
Notably, participant ID 06 exhibits a substantially larger interquartile range (IQR) when compared with other participants, indicating larger variability and a different behavior pattern during the flexion measurements. Overall, Figure 11 not only shows individual performance in terms of flexion but also highlights the presence of inter-subject variability within a standardized test campaign.
Figure 12 shows the distribution of maximum elbow extension values for each participant, as a function of individual ID and sex, red for female and blue for male participants. Each box represents the distribution of flexion–extension repetitions recorded for a single individual. As defined in the experimental protocol developed in collaboration with physiotherapists, each participant was expected to complete 30 repetitions, derived from three sessions of 10 repetitions each. This figure shows the variability in extension performance across all volunteers. Smaller distributions are indicative of better performance, as they reflect a greater ability to fully extend the elbow joint, closer to its anatomical limit. The figure highlights individual differences in extension range and consistency, with some participants demonstrating tightly clustered values and others showing greater variability. As observed previously in Figure 11, participant ID 06 again exhibits a notably wider distribution and reduced extension values when compared with other participants, indicating larger variability in the measured performance. Nevertheless, in Figure 13, the ROM values for this participant remain within the overall tendency observed across the study population, suggesting that the atypical flexion and extension distributions did not substantially alter the overall functional range of motion.
Figure 13 shows the distribution of elbow joint range of motion (ROM) values for each participant as a function of individual ID and sex, red for female and blue for male participants. Each boxplot reflects the variability in ROM across 30 flexion–extension repetitions, following the standardized protocol of three sessions with 10 repetitions each, as established in collaboration with physiotherapists. The ROM was calculated as the difference between maximum flexion and maximum extension values for each repetition, representing the total movement achieved by the elbow joint.
The collected evidence suggests a significant potential clinical impact of the device. The combination of an extremely low-burden wearable system, smooth and repeatable assistance, and the ability to acquire objective movement parameters (ROM, maximum flexion/extension, and acceleration profiles) enables the integration of ASSIST-FEEv3 into standard rehabilitation protocols without increasing the physical or cognitive load on the patient.
These characteristics are particularly relevant for older adults, post-surgical patients, and individuals with reduced muscular endurance, for whom safety, portability, and the possibility of performing controlled exercises outside the hospital environment are key factors for therapeutic adherence. Moreover, the availability of objective and repeatable movement measurements can support physiotherapists in monitoring clinical progression, tailoring exercise intensity, and evaluating treatment effectiveness over time.

5. Conclusions

The functionality of the ASSIST-FEEv3 elbow assisting device has been demonstrated through characterization tests that were conducted within a medical physiotherapy context, with results confirming both its technically feasible performance and user-centered design. The device employs a cable-driven mechanism to guide the forearm through controlled elbow flexion–extension exercises in the sagittal plane. This motion is generated by two coordinated cables that are actuated by servomotors housed within a lightweight module positioned on the user’s body. Experimental results indicate that the ASSIST-FEEv3 device reliably supports the full range of elbow motion during repeated cyclic exercises while maintaining low energy consumption, high operational comfort, and positive acceptance among physiotherapy practitioners.
The proposed device also enables objective monitoring of elbow kinematics while ensuring safe, comfortable, and repeatable exercise execution. These characteristics support its potential application in supervised rehabilitation, elderly exercise programs, and home-based therapeutic scenarios. In particular, the obtained results confirm that ASSISTFEEv3 is a valid and reliable platform for collecting and analyzing biomechanical data related to elbow motion. Tests conducted with healthy volunteers validated the proposed operational protocols, demonstrating consistency and repeatability in the measured data. Minor deviations observed during the experiments were mainly attributed to transient individual variations rather than intrinsic limitations of the device.
From physiotherapeutic perspectives, the developed system represents a significant step toward more objective, sensitive, and reproducible assessment methods for upper-limb motion assistance. Its capability to continuously monitor joint kinematics and motion-related dynamic parameters provides valuable support for evaluating functional recovery, elderly exercise performance, and the adaptation of rehabilitation programs according to patient progress. This aspect is particularly relevant for elderly users and individuals with neuromotor impairments, for whom systematic monitoring and the possibility of supervised home-based rehabilitation may contribute to improved therapeutic outcomes and exercise adherence.
From an operational perspective, patient safety, comfort, and adaptability remain fundamental requirements for practical clinical implementation. Features such as physiotherapy-friendly setup, portability, and rapid session configuration enhance the suitability of ASSISTFEEv3 for outdoor patient clinics, home-based care, and low-intensity rehabilitation facilities. Furthermore, the integration of non-invasive sensing technology and compatibility with clinical reporting and monitoring tools increase its applicability within existing rehabilitation workflows, supporting efficient, data-driven, and patient-centered physiotherapeutic practices.

6. Patents

Ceccarelli, Marco.; Morales-Cruz, Cuauhtemoc. Device for motion assistance of the elbow. Patent, 2025. Number: IT102025000006393, 25 March 2025.

Author Contributions

Conceptualization, M.C. and C.M.-C.; methodology, C.M.-C., F.F., F.S. and R.M.; software, C.M.-C.; validation, F.F., F.S. and R.M.; formal analysis, C.M.-C. and M.C.; investigation, M.C. and F.F.; resources, M.C.; data curation, C.M.-C.; writing—original draft preparation, C.M.-C.; writing—review and editing, M.C.; visualization, C.M.-C.; supervision, M.C.; project administration, M.C.; funding acquisition, M.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Italian 2022 PRIN-PNRR funding program with grant number P2022A4ELB.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Ethics Committee of Policlinico di Tor Vergata, Rome, with protocol code RS. 197.22 on 15 November 2022.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Data available upon request.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Feigin, V.L.; Norrving, B.; Mensah, G.A. Global Burden of Stroke. Circ. Res. 2017, 120, 439–448. [Google Scholar] [CrossRef] [PubMed]
  2. Ceccarelli, M. Challenges in service robot devices for elderly motion assistance. Robotica 2024, 42, 4186–4199. [Google Scholar] [CrossRef]
  3. Izquierdo, M.; Merchant, R.A.; Morley, J.E.; Anker, S.D.; Aprahamian, I.; Arai, H.; Aubertin-Leheudre, M.; Bernabei, R.; Cadore, E.L.; Cesari, M.; et al. International Exercise Recommendations in Older Adults (ICFSR): Expert Consensus Guidelines. J. Nutr. Health Aging 2021, 25, 824–853. [Google Scholar] [CrossRef] [PubMed]
  4. Schmitter-Edgecombe, M.; Luna, C.; Cook, D.J. Technologies for Health Assessment, Promotion, and Intervention: Focus on Aging and Functional Health. In Positive Neuropsychology: Evidence-Based Perspectives on Promoting Brain and Cognitive Health; Randolph, J.J., Ed.; Springer: Cham, Switzerland, 2022; pp. 111–138. [Google Scholar] [CrossRef]
  5. Merchant, R.A.; Morley, J.E.; Izquierdo, M. Exercise, Aging and Frailty: Guidelines for Increasing Function. J. Nutr. Health Aging 2021, 25, 405–409. [Google Scholar] [CrossRef] [PubMed]
  6. Rockwood, K.; Song, X.; MacKnight, C.; Bergman, H.; Hogan, D.B.; McDowell, I.; Mitnitski, A. A global clinical measure of fitness and frailty in elderly people. CMAJ 2005, 173, 489–495. [Google Scholar] [CrossRef] [PubMed]
  7. Dhatrak, P.; Durge, J.; Dwivedi, R.K.; Pradhan, H.K.; Kolke, S. Interactive design and challenges on exoskeleton performance for upper-limb rehabilitation: A comprehensive review. Int. J. Interact. Des. Manuf. (IJIDeM) 2025, 19, 4565–4587. [Google Scholar] [CrossRef]
  8. Bardi, E.; Gandolla, M.; Braghin, F.; Resta, F.; Pedrocchi, A.L.G.; Ambrosini, E. Upper limb soft robotic wearable devices: A systematic review. J. Neuroeng. Rehabil. 2022, 19, 87. [Google Scholar] [CrossRef] [PubMed]
  9. Supriyono, C.S.A.; Dragusanu, M.; Malvezzi, M. A Comprehensive Review of Elbow Exoskeletons: Classification by Structure, Actuation, and Sensing Technologies. Sensors 2025, 25, 4263. [Google Scholar] [CrossRef] [PubMed]
  10. Gonçalves, A.; Silva, M.F.; Mendonça, H.; Rocha, C.D. A Review of Robotic Interfaces for Post-Stroke Upper-Limb Rehabilitation: Assistance Types, Actuation Methods, and Control Mechanisms. Robotics 2025, 14, 141. [Google Scholar] [CrossRef]
  11. Glowinski, S.; Pecolt, S.; Błażejewski, A.; Maciejewski, I.; Królikowski, T. Gait Analysis with an Upper Limb Prosthesis in a Child with Thrombocytopenia–Absent Radius Syndrome. J. Clin. Med. 2025, 14, 2245. [Google Scholar] [CrossRef] [PubMed]
  12. Nef, T.; Mihelj, M.; Riener, R. ARMin: A robot for patient-cooperative arm therapy. Med. Biol. Eng. Comput. 2007, 45, 887–900. [Google Scholar] [CrossRef] [PubMed]
  13. Zimmermann, Y.; Sommerhalder, M.; Wolf, P.; Riener, R.; Hutter, M. ANYexo 2.0: A Fully Actuated Upper-Limb Exoskeleton for Manipulation and Joint-Oriented Training in All Stages of Rehabilitation. IEEE Trans. Robot. 2023, 39, 2131–2150. [Google Scholar] [CrossRef]
  14. Pan, J.; Astarita, D.; Baldoni, A.; Dell’Agnello, F.; Crea, S.; Vitiello, N.; Trigili, E. A Self-Aligning Upper-Limb Exoskeleton Preserving Natural Shoulder Movements: Kinematic Compatibility Analysis. IEEE Trans. Neural Syst. Rehabil. Eng. 2023, 31, 4954–4964. [Google Scholar] [CrossRef] [PubMed]
  15. Gandolla, M.; Luciani, B.; Longatelli, V.; Tropea, P.; Seregni, A.; Corbo, M.; Braghin, F.; Pedrocchi, A. AGREE: An upper limb motorized exoskeleton for restoring arm functions: A single-blinded randomized controlled trial. J. Neuroeng. Rehabil. 2025, 22, 134. [Google Scholar] [CrossRef] [PubMed]
  16. Qian, W.; Liao, J.; Lu, L.; Ai, L.; Li, M.; Xiao, X.; Guo, Z. CURER: A Lightweight Cable-Driven Compliant Upper Limb Rehabilitation Exoskeleton Robot. IEEE/ASME Trans. Mechatron. 2023, 28, 1730–1741. [Google Scholar] [CrossRef]
  17. Cui, Z.; Zhao, Z.; Su, C.; Meng, J.; Sang, H.; Zhao, X.; You, T. Mechanism design and drive performance analysis of hybrid drive upper limb rehabilitation robot. J. Mech. Sci. Technol. 2025, 39, 7775–7789. [Google Scholar] [CrossRef]
  18. Mao, Y.; Agrawal, S.K. Design of a Cable-Driven Arm Exoskeleton (CAREX) for Neural Rehabilitation. IEEE Trans. Robot. 2012, 28, 922–931. [Google Scholar] [CrossRef]
  19. Wei, W.; Qu, Z.; Wang, W.; Zhang, P.; Hao, F. Design on the Bowden Cable-Driven Upper Limb Soft Exoskeleton. Appl. Bionics Biomech. 2018, 2018, 1925694. [Google Scholar] [CrossRef]
  20. Rosati, G.; Gallina, P.; Masiero, S.; Rossi, A. Design of a new 5 d.o.f. wire-based robot for rehabilitation. In Proceedings of the 9th International Conference on Rehabilitation Robotics, 2005, ICORR 2005, Chicago, IL, USA, 28 June–1 July 2005; pp. 430–433. [Google Scholar] [CrossRef]
  21. Xiloyannis, M.; Chiaradia, D.; Frisoli, A.; Masia, L. Physiological and kinematic effects of a soft exosuit on arm movements. J. Neuroeng. Rehabil. 2019, 16, 29. [Google Scholar] [CrossRef] [PubMed]
  22. KhalilianMotamed Bonab, A.; Chiaradia, D.; Frisoli, A.; Leonardis, D. A Framework for Modeling, Optimization, and Musculoskeletal Simulation of an Elbow–Wrist Exosuit. Robotics 2024, 13, 60. [Google Scholar] [CrossRef]
  23. Bonab, A.K.; Camardella, C.; Frisoli, A.; Chiaradia, D. Hybrid Soft-Rigid Elbow Exosuit: Theory, Mechatronic Design, and Experimental Assessment. IEEE Trans. Robot. 2026, 42, 912–930. [Google Scholar] [CrossRef]
  24. Jeyabalan, P.K.M.; Thiagu, S.; Nehrujee, A.; Sujatha, S.; Balasubramanian, S. Motorized Arm Support (MARS)—An Underactuated Robot for Adaptive Arm De-weighting in Neurorehabilitation. IEEE Access 2026, 14, 23244–23256. [Google Scholar] [CrossRef]
  25. Jenks, B.; Levan, H.; Stefanovic, F. OpenSEA: A 3D printed planetary gear series elastic actuator for a compliant elbow joint exoskeleton. Front. Robot. AI 2025, 12, 1528266. [Google Scholar] [CrossRef] [PubMed]
  26. Desai, S.; Piovesan, D.; Dagli, C. Design and Development of ARDEL (Active-Assist Rehabilitation Device for Elbow). In ASME International Mechanical Engineering Congress and Exposition; Volume 5: Biomedical and Biotechnology; ASME: New Orleans, LA, USA, 2023. [Google Scholar] [CrossRef]
  27. Jameel, H.F.; Alazawi, A.; Mahmood, A.I. Design and Investigation of a low-cost elbow rehabilitation tool. Ann. 3D Print. Med. 2024, 15, 100167. [Google Scholar] [CrossRef]
  28. Franzke, A.W.; Kristoffersen, M.B.; Farina, D.; van der Sluis, C.K.; Bongers, R.M.; Murgia, A. Testing the Use of Advanced Upper Limb Prostheses: Towards quantifying the Movement Quality with Inertial-Magnetic Measurement Units. Prosthesis 2023, 5, 264–281. [Google Scholar] [CrossRef]
  29. Górski, F.; Rybarczyk, D.; Wichniarek, R.; Wierzbicka, N.; Kuczko, W.; Żukowska, M.; Regulski, R.; Pacurar, R.; Comsa, D.S.; Baila, D.I.; et al. Development and Testing of an Individualized Sensorised 3D Printed Upper Limb Bicycle Prosthesis for Adult Patients. Appl. Sci. 2023, 13, 12918. [Google Scholar] [CrossRef]
  30. Zuccon, G.; Bottin, M.; Ceccarelli, M.; Rosati, G. Design and Performance of an Elbow Assisting Mechanism. Machines 2020, 8, 68. [Google Scholar] [CrossRef]
  31. Kotov, S.; Ceccarelli, M. Design and Prototype of L-CADEL.v5 Elbow Assisting Device. Designs 2025, 9, 126. [Google Scholar] [CrossRef]
  32. Ceccarelli, M.; Kotov, S.; Russo, M. Results and Problems from Lab Testing with L-CADEL.V3. In Proceedings of the Mechanism Design for Robotics; Lovasz, E.C., Ceccarelli, M., Ciupe, V., Eds.; Springer: Cham, Switzerland, 2024; pp. 193–201. [Google Scholar] [CrossRef]
  33. Ceccarelli, M.; Russo, M.; Boschetti, G.; Bottin, M. Problems and requirements for motion-assisting devices for elderly people. In New Trends in Mechanism and Machine Science (EuCoMeS 2024); Rosati, G., Gasparetto, A., Ceccarelli, M., Eds.; Mechanisms and Machine Science; Springer: Cham, Switzerland, 2024; Volume 165, pp. 20–27. [Google Scholar] [CrossRef]
  34. Bizhanov, D.; Ceccarelli, M.; Ozhikenov, K.; Zhetenbayev, N. Design Evolution and Experimental Validation of the AlmatyExoElbow Assisting Device. Robotics 2026, 15, 12. [Google Scholar] [CrossRef]
  35. Ceccarelli, M. ASSIST Project: Plans and Expected Results. In Advances in Motion Assisting Devices for Limbs. ASSIST 2025; Ceccarelli, M., Boschetti, G., Eds.; Mechanisms and Machine Science; Springer: Cham, Switzerland, 2025; Volume 200, pp. 3–20. [Google Scholar] [CrossRef]
  36. Morales-Cruz, C. Advances in Motion Assisting Devices for Limbs. ASSIST 2025; Ceccarelli, M., Boschetti, G., Eds.; Mechanisms and Machine Science; Springer: Cham, Switzerland, 2025; Volume 200, pp. 68–85. [Google Scholar] [CrossRef]
  37. Frisina, F.; Morales-Cruz, C.; Scerbo, F.; Mazzotta, R.; Ceccarelli, M. Physiotherapy Considerations from Test Results with ASSIST-FEE.v3 Elbow Assisting Device. In New Trends in Medical and Service Robotics; Laribi, M.A., Carbone, G., Pisla, D., Zeghloul, S., Eds.; Mechanisms and Machine Science; Springer: Cham, Switzerland, 2025; Volume 186, pp. 196–206. [Google Scholar] [CrossRef]
  38. Frisina, F.; Francesco, S.; Mazzotta, R.; Ceccarelli, M.; Morales-Cruz, C.; Ofonaike, E.U.; Russo, M. Physiotherapy perspectives on robotic assistance of elbow motion. In Proceedings of the Engineering Methodologies for Medicine and Sports: Proceedings of EMMS 2026, Rome, Italy, 7–9 February 2026. [Google Scholar]
  39. Ceccarelli, M.; Morales-Cruz, C. Device for Motion Assistance of Elbow. Patent Number: IT102025000006393, 25 March 2025. [Google Scholar]
  40. James, M.A. Use of the Medical Research Council muscle strength grading system in the upper extremity. J. Hand Surg. 2007, 32, 154–156. [Google Scholar] [CrossRef] [PubMed]
  41. Eldridge, S.M.; Chan, C.L.; Campbell, M.J.; Bond, C.M.; Hopewell, S.; Thabane, L.; Lancaster, G.A. CONSORT 2010 statement: Extension to randomised pilot and feasibility trials. BMJ 2016, 355, i5239. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Design and operation requirements for ASSIST-FEEv3 elbow assisting device for arm motion assistance.
Figure 1. Design and operation requirements for ASSIST-FEEv3 elbow assisting device for arm motion assistance.
Robotics 15 00130 g001
Figure 2. Conceptual design of elbow assisting device ASSIST-FEEv3 for flexion and extension of the elbow joint: (a) block components; (b) a wearable configuration.
Figure 2. Conceptual design of elbow assisting device ASSIST-FEEv3 for flexion and extension of the elbow joint: (a) block components; (b) a wearable configuration.
Robotics 15 00130 g002
Figure 3. Testing layout of the worn ASSIST-FEEv3 prototype: (a) experimental testing setup on a healthy volunteer; (b) conceptual layout.
Figure 3. Testing layout of the worn ASSIST-FEEv3 prototype: (a) experimental testing setup on a healthy volunteer; (b) conceptual layout.
Robotics 15 00130 g003
Figure 4. The designed physiotherapy testing procedure using the ASSISTFEEv3 device presented in Figure 2.
Figure 4. The designed physiotherapy testing procedure using the ASSISTFEEv3 device presented in Figure 2.
Robotics 15 00130 g004
Figure 5. Snapshots of a laboratory test with the ASSIST-FEEv3 device for elbow flexion and extension: (a) starting configuration showing the IMU reference frame; (b) middle phase; (c) final flexed configuration.
Figure 5. Snapshots of a laboratory test with the ASSIST-FEEv3 device for elbow flexion and extension: (a) starting configuration showing the IMU reference frame; (b) middle phase; (c) final flexed configuration.
Robotics 15 00130 g005
Figure 6. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of elbow angle for two cycles.
Figure 6. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of elbow angle for two cycles.
Robotics 15 00130 g006
Figure 7. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of wrist acceleration for two cycles.
Figure 7. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of wrist acceleration for two cycles.
Robotics 15 00130 g007
Figure 8. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of acceleration free-of-gravity magnitude in Figure 7.
Figure 8. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of acceleration free-of-gravity magnitude in Figure 7.
Robotics 15 00130 g008
Figure 9. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of computed power consumption for two cycles.
Figure 9. Test results during a test with ASSIST-FEEv3, like in Figure 5, in terms of computed power consumption for two cycles.
Robotics 15 00130 g009
Figure 10. Statistical elaboration of results from the testing campaign using the ASSIST-FEEv3 device in terms of distribution of flexion–extension repetitions: (a) maximum flexion values; (b) maximum extension values; and (c) range of motion (ROM).
Figure 10. Statistical elaboration of results from the testing campaign using the ASSIST-FEEv3 device in terms of distribution of flexion–extension repetitions: (a) maximum flexion values; (b) maximum extension values; and (c) range of motion (ROM).
Robotics 15 00130 g010
Figure 11. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of maximum flexion values by participant ID and by sex.
Figure 11. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of maximum flexion values by participant ID and by sex.
Robotics 15 00130 g011
Figure 12. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of maximum extension values by participant ID and by sex.
Figure 12. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of maximum extension values by participant ID and by sex.
Robotics 15 00130 g012
Figure 13. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of range of motion (ROM) values by participant ID and grouped by sex.
Figure 13. Results from the testing campaign using the ASSIST-FEEv3 device in terms of the distribution of range of motion (ROM) values by participant ID and grouped by sex.
Robotics 15 00130 g013
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Morales-Cruz, C.; Frisina, F.; Scerbo, F.; Mazzotta, R.; Ceccarelli, M. Experimental Validation of ASSIST-FEEv3 Elbow Assisting Device with Physiotherapy Considerations. Robotics 2026, 15, 130. https://doi.org/10.3390/robotics15070130

AMA Style

Morales-Cruz C, Frisina F, Scerbo F, Mazzotta R, Ceccarelli M. Experimental Validation of ASSIST-FEEv3 Elbow Assisting Device with Physiotherapy Considerations. Robotics. 2026; 15(7):130. https://doi.org/10.3390/robotics15070130

Chicago/Turabian Style

Morales-Cruz, Cuauhtémoc, Fortunato Frisina, Francesco Scerbo, Rocco Mazzotta, and Marco Ceccarelli. 2026. "Experimental Validation of ASSIST-FEEv3 Elbow Assisting Device with Physiotherapy Considerations" Robotics 15, no. 7: 130. https://doi.org/10.3390/robotics15070130

APA Style

Morales-Cruz, C., Frisina, F., Scerbo, F., Mazzotta, R., & Ceccarelli, M. (2026). Experimental Validation of ASSIST-FEEv3 Elbow Assisting Device with Physiotherapy Considerations. Robotics, 15(7), 130. https://doi.org/10.3390/robotics15070130

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop