Article

Urinary Catheters Coated with a Novel Biofilm Preventative Agent Inhibit Biofilm Development by Diverse Bacterial Uropathogens

1
Department of Immunology and Molecular Microbiology, Texas Tech University Health Sciences Center, Lubbock, TX 79430, USA
2
Next Science, Jacksonville, FL 32256, USA
3
Department of Medical Education, Texas Tech University Health Sciences Center, Lubbock, TX 79430, USA
4
Department of Urology, Texas Tech University Health Sciences Center, Lubbock, TX 79430, USA
5
Department of Surgery, Texas Tech University Health Sciences Center, Lubbock, TX 79430, USA
*
Author to whom correspondence should be addressed.
Current address: Hendrick Medical Center, Abilene, TX 79601, USA.
Antibiotics 2022, 11(11), 1514; https://doi.org/10.3390/antibiotics11111514
Submission received: 1 October 2022 / Revised: 22 October 2022 / Accepted: 25 October 2022 / Published: 30 October 2022
(This article belongs to the Section Novel Antimicrobial Agents)

Abstract

:
Despite the implementation of stringent guidelines for the prevention of catheter-associated (CA) urinary tract infection (UTI), CAUTI remains one of the most common health care-related infections. We previously showed that an antimicrobial/antibiofilm agent inhibited biofilm development by Gram-positive and Gram-negative bacterial pathogens isolated from human infections. In this study, we examined the ability of a novel biofilm preventative agent (BPA) coating on silicone urinary catheters to inhibit biofilm formation on the catheters by six different bacterial pathogens isolated from UTIs: three Escherichia coli strains, representative of the most common bacterium isolated from UTI; one Enterobacter cloacae, a multidrug-resistant isolate; one Pseudomonas aeruginosa, common among patients with long-term catheterization; and one isolate of methicillin-resistant Staphylococcus aureus, as both a Gram-positive and a resistant organism. First, we tested the ability of these strains to form biofilms on urinary catheters made of red rubber, polyvinyl chloride (PVC), and silicone using the microtiter plate biofilm assay. When grown in artificial urine medium, which closely mimics human urine, all tested isolates formed considerable biofilms on all three catheter materials. As the biofilm biomass formed on silicone catheters was 0.5 to 1.6 logs less than that formed on rubber or PVC, respectively, we then coated the silicone catheters with BPA (benzalkonium chloride, polyacrylic acid, and glutaraldehyde), and tested the ability of the coated catheters to further inhibit biofilm development by these uropathogens. Compared with the uncoated silicone catheters, BPA-coated catheters completely prevented biofilm development by all the uropathogens, except P. aeruginosa, which showed no reduction in biofilm biomass. To explore the reason for P. aeruginosa resistance to the BPA coating, we utilized two specific lipopolysaccharide (LPS) mutants. In contrast to their parent strain, the two mutants failed to form biofilms on the BPA-coated catheters, which suggests that the composition of P. aeruginosa LPS plays a role in the resistance of wild-type P. aeruginosa to the BPA coating. Together, our results suggest that, except for P. aeruginosa, BPA-coated silicone catheters may prevent biofilm formation by both Gram-negative and Gram-positive uropathogens.

1. Introduction

Complicated urinary tract infections (UTIs) involve conditions that compromise the urinary tract and/or its host defenses, including the presence of foreign bodies within the urinary tract, such as indwelling catheters and bladder or renal calculi, renal failure, and immunosuppression [1,2]. About 80% of complicated UTIs are related to indwelling catheters [3]. Indwelling urinary catheters, which are a closed system consisting of the catheter tubing, a small retention balloon, and a drainage collection bag, are inserted through the urethra or suprapubically to allow for bladder drainage [4]. Unfortunately, catheter usage, especially prolonged use, often leads to catheter-associated UTIs (CAUTIs) that arise when bacteria enter the urinary tract via insertion of the catheter, or during post-placement manipulation of the catheter or its drainage system [5,6,7]. Catheters also provide a surface to which bacteria can adhere and develop biofilms [7]. CAUTIs are among the most frequent types of hospital-acquired infections, currently ranking second [8,9]. It is estimated that, in the U.S., more than 30 million bladder catheters are annually inserted, with several hundred thousand CAUTIs associated with those insertions [10]. CAUTIs may lead to potentially serious consequences, including frequent febrile episodes, acute and chronic pyelonephritis, bacteremia with urosepsis and septic shock, catheter obstruction and renal and/or bladder stone formation, and even bladder cancer with prolonged use [11,12,13]. Placement of a urinary catheter led to a 39-fold increase in bacteremia in one study [11]. Of bloodstream infections noted 48 h or more after admission, 21% were from a urinary source and 71% of those were catheter-associated [12]. In long-term care facilities, CAUTIs have been estimated to be the source of 45–50% of bacteremia [14,15]. The most common pathogens associated with CAUTI include uropathogenic Escherichia coli, Klebsiella spp., Pseudomonas aeruginosa, Enterococcus faecalis, and other Enterobacterales [9,11,16].
Pathogens associated with CAUTIs are increasingly resistant to antimicrobial agents. In both acute and long-term care facilities, the urine of patients with indwelling catheters is the main source of resistant Gram-negative bacteria, including extended spectrum beta-lactamase (ESBL), producing Enterobacterales and carbapenem-resistant Enterobacterales (CRE) [17,18]. Uropathogens frequently form biofilms first on the surfaces (interior and/or exterior) of the indwelling catheter, and then on the uroepithelium and drainage bag [11,13]. These catheter-associated biofilms, which consist of the attached bacteria, their extracellular products, and host components, provide the uropathogens with many survival advantages, including resistance to the sheer force of the urine flow, resistance to phagocytosis, and resistance to antimicrobial agents [11,19,20,21]. Therefore, an ideal urinary catheter is one coated with a strong antimicrobial agent that prevents the uropathogens from colonizing the catheter and developing antibiotic-resistant biofilms. Multiple antimicrobial urinary tract catheter coating agents have been described [22] with variable effectiveness, including silver alloy [23], chlorhexidine [24,25], triclosan [24,26], antibiotics [23,27], antimicrobial peptides [28], and bacteriophages [29]. A meta-analysis of indwelling urethral catheters concluded that silver alloy-coated catheters were not associated with a statistically significant reduction in symptomatic CAUTI, and that while use of a nitrofurazone catheter reduced the risk of symptomatic CAUTI, the patients experienced more discomfort with these catheters [30]. It was further noted that several previously available antiseptic-coated and antimicrobial-coated catheters are no longer available [30]. Coatings that prevent bacterial adhesion to the catheter yet have no antimicrobial activity have been investigated, also with varying results: hydrogel- and polytetrafluoroethylene-coated urinary catheters are commercially available, while other compounds such as polyzwitterion polymers, polyethylene glycol, and enzymes (the quorum-sensing inhibitor acylase, and exopolysaccharide-specific glycoside hydrolase) have shown promising results in research [22,31]. The development of antimicrobial coatings for urinary catheters has been ongoing, with reports of successful killing of P. aeruginosa for up to 30 days by a catheter designed to release ciprofloxacin and azithromycin [32]; killing of Proteus mirabilis, Staphylococcus aureus, and E. coli for 7–12 weeks by slow release of rifampicin, sparfloxacin, and triclosan impregnated in a catheter [33]; and prevention of biofilm formation by S. aureus and E. coli on catheters coated with different acrylate compounds and the killing of these organisms by the release of antibiotics (rifampicin, sparfloxacin, and triclosan) impregnated within the coating [34].
We previously demonstrated the effectiveness of a biofilm dispersal formulation developed by the Next Science company (NS, when in conjunction with the different antimicrobial/antibiofilm agents tested) (Jacksonville, FL, USA), inhibiting the growth and biofilm development by different Gram-positive and Gram-negative pathogens isolated from wound and middle ear infections [35,36]. Additionally, we assessed the NS-formulated wound solution Bactisure in inhibiting biofilm development by an E. coli laboratory strain and three E. coli clinical isolates on three types of urinary catheters [37]. When added to the growth medium containing catheter pieces, NS Bactisure inhibited biofilm development by all four E. coli strains on each of the three catheter types [37]. However, for practical application, it is essential to coat urinary catheters with a specific NS formulation and examine the effectiveness of the coated catheter in preventing biofilm formation. In this study, we report an examination of biofilm development by eight bacterial strains isolated from UTIs on silicone urinary catheters coated internally and externally with a novel NS biofilm preventative agent (BPA). To provide additional relevance, biofilm development by the uropathogens was tested on BPA-coated catheters incubated in a medium that closely mimics human urine.

2. Results

2.1. E. coli Predominated among the UTI Isolates

Through a protocol approved by the Institutional Review Board at Texas Tech University Health Sciences Center (TTUHSC), we obtained 106 clinical isolates from urine samples of patients, many with long-term catheterization, presenting with UTI at the TTUHSC Urology Clinic. The most frequently isolated bacterial species were E. coli (32) followed by Klebsiella spp. (22), P. aeruginosa (8), and Enterobacter cloacae (7) (Figure 1). We then compared the distribution of our isolates to the most common agents of CAUTI, as reported by the National Healthcare Safety Network (NHSN) database for 2015–2017, containing 85,545 isolates from hospitalized patients [9], to determine how closely our outpatient isolates matched those of the hospitalized patients. By ranking, our first three, E. coli, Klebsiella spp., and P. aeruginosa, were the same as the national database [9]. The percentages of E. faecalis and Proteus spp. isolates among our collection were less (5.7% vs. 9.3% and 1.9% vs. 5.6%) while those of Enterobacter spp. and S. aureus were greater (6.6% vs. 4.9% and 4.7% vs 2.1%, respectively) (Figure 1) [9]. Additionally, our isolates included 4.7% Streptococcus agalactiae and 0.9% Candida albicans, while the inpatient database did not include either of those microorganisms among the top 15 species isolated (Figure 1) [9].

2.2. Several UIs Exhibited Multidrug Resistance

As antibiotic resistance has been increasing in general and among isolates from CAUTIs [17,18], we selected six isolates for which we obtained antibiotic susceptibility results for further examination: three E. coli isolates (UI-033-EsC, UI-038-EsC, and UI-061-EsC), representative of the most common agent of CAUTI; one E. cloacae (UI-095-EnC); one P. aeruginosa (UI-092-PA), a common cause of CAUTI in long-term catheterized patients; and one methicillin resistant S. aureus (UI-009-MRSA), to represent Gram-positive uropathogens. The three E. coli isolates and the E. cloacae isolate exhibited patterns of resistance that were quite similar (Table 1). Isolate UI-038-EsC was the most resistant among the E. coli, fully susceptible to only six of the twenty-three tested drugs (26%); UI-095-EnC was most resistant overall—susceptible to five of the twenty-one drugs tested (24%) (Table 1). Defining multidrug resistance (MDR) as an isolate that is resistant to three or more classes of antibiotics or that carries a key resistance marker (such as methicillin resistance for staphylococci or ESBL production for the Enterobacterales) [38,39], isolates UI-033-EsC, UI-038-EsC, UI-061-EsC, and UI-095-EnC were MDR and ESBL Enterobacterales, but none produced carbapenemase (Table S1 in Supplementary Materials). Of the ten drug classes tested, UI-038-EsC, UI-061-EsC, and UI-095-EnC were resistant to seven: cephalosporins, fluoroquinolones, monobactams, penicillins, sulfonamides, tetracyclines, and aminoglycosides (the E. coli isolates) or nitrofurans (the E. cloacae isolate) (Table S1 in Supplementary Materials). The only class of antibiotics for which no resistance was noted among any of the Enterobacterales isolates was the glycylcyclines (Table 1 and Table S1 in Supplementary Materials).
UI-092-PA was resistant to imipenem, but was susceptible to meropenem and the other beta-lactams tested (Table 1). Carbapenem resistance in such isolates is usually due to inactivation of the OprD porin, rather than production of carbapenemase [40,41]. The strain was not considered MDR as it showed intermediate susceptibility to the fluoroquinolones, and susceptibility to all of the other classes (Table S1 in Supplementary Materials). UI-009-MRSA was MDR due to the presence of methicillin resistance, even though it was resistant only to one other class of antibiotics—the macrolides, via inducible clindamycin resistance (Table 2 and Table S2 in Supplementary Materials) [38,39,42]. Of the three other S. aureus isolates for which susceptibilities were available, none were resistant to methicillin, but UI-073-SA and UI-078-SA were resistant to three classes of antibiotics and considered MDR, while UI-096-SA was susceptible to at least one drug in each class of antibiotics (Table S2 in Supplementary Materials). Analysis of eight Enterococcus spp. Isolates showed only two were MDR: E. faecalis isolate UI-065-Efc was resistant to three of eight classes of antibiotics tested and E. faecium isolate UI-048-Efm showed resistance to four classes (Table S3 in Supplementary Materials).

2.3. Artificial Urine Medium Supports Growth and Biofilm Development by the Uropathogens

To more closely mimic bacterial growth within the urinary bladder, we used artificial urine medium (AUM), which is similar to human urine [43]. This medium was previously shown to not only support the growth of Gram-positive and Gram-negative uropathogens, but to also promote crystal production by Staphylococcus epidermidis and P. mirabilis [43]. After 24 h of growth at 37 °C ± 0.5 °C, and compared to their growth in the rich medium TSB, the growth of five of the six uropathogens was significantly reduced in AUM (Figure 2). Only UI-033-EsC grew at a comparable level in both media. The observed reduction in growth in AUM was expected since the AUM is a chemically-defined medium with restricted nutrients. While growth was not as robust in AUM, the medium did support sufficient growth of the uropathogens for subsequent experiments (107 to 109 CFU/mL) (Figure 2).
The ability of the uropathogens to form biofilm on urinary catheters when grown in AUM was assessed using the microtiter plate biofilm assay. Urinary catheters are made of different materials, with the most common being silicone, rubber, and polyvinyl chloride (PVC). Urinary catheters made of these three materials were aseptically cut into 1.5-cm lengths and individual pieces were inoculated with the uropathogens in TSB or AUM. After 24 h at 37 °C ± 0.5 °C, all six uropathogens formed biofilms on each type of catheter when grown in TSB (Figure 3a). Except for UI-009-MRSA, which produced biofilms with significantly less biomass than the other strains on all three types of catheters, the biomass of the biofilms formed by the Gram-negative bacilli was similar: an average of 7.7 logs/catheter piece on the silicone catheter, 7.5 logs on the rubber catheter, and 7.5 logs on the PVC catheter (Figure 3a). In AUM, the amount of biofilm formed by all tested uropathogens on the three surfaces was reduced from that produced in TSB (Figure 3b); similar to the reduction in planktonic growth in AUM vs. TSB (Figure 2). The average biomass of all the biofilms formed in AUM on the silicone catheters was 5.3 logs/catheter piece, with 5.8 logs/catheter piece on the rubber catheters and 6.9 logs/catheter piece on the PVC catheters (Figure 3b). Again, the biomass of UI-009-MRSA was consistently and significantly less than that produced by the other isolates, regardless of surface (Figure 3b). The largest reduction in overall biomass, from 7.7 logs in TSB to 5.3 logs in AUM, was seen with the silicone catheters (Figure 3b), which substantiates reports that silicone catheters undergo less bacterial colonization in vivo [44,45]. Comparison of biofilm biomass formed by each isolate grown in AUM on the three types of catheters showed that the biomass of biofilms formed by UI-033-EsC, UI-038-EsC, and UI-009-MRSA on silicone was less than that formed on rubber or PVC (Figure 3b). The biomass of biofilms formed by UI-061-EsC and UI-095-EnC were less on silicone than on PVC, and that of UI-092-PA was less than that on rubber (Figure 3b). Direct comparison of biofilm biomass of each UI on each type of catheter material when grown in TSB versus AUM showed that the reduction in biomass was most consistent and most significant over all the Uis when silicone catheter material was tested (Figure S1 in Supplementary Material). Since our goal was to reduce bacterial colonization of the urinary catheter, we decided to continue with the silicone catheters.

2.4. BPA-Coated Catheters Inhibit Biofilm Development by the Uropathogens

It is estimated that 70%-80% of UTIs are associated with catheter use [3]. Thus, coating catheters with antimicrobial agents is an essential approach to prevent UTI. Multiple antimicrobial agents have been used to coat urinary catheters with varying degrees of reported success [46,47]. We previously reported successful inhibition of biofilm formation by several E. coli clinical isolates when urinary catheters were treated with NS Bactisure (0.13% benzalkonium chloride [BZK], sodium acetate, acetic acid, ethanol, and water; per list on package) [37]. Thus, for the next step, Next Science coated 1.5-cm lengths of silicone urinary catheters with BPA (0.175% BZK, 1.99% polyacrylic acid, 2.62% glutaraldehyde, and 95.215% sterile water for irrigation) and returned the coated pieces for determination of the ability of the BPA-coated catheters to prevent biofilm formation by uropathogens. Except for P. aeruginosa, BPA-coated catheters inhibited biofilm development by all tested uropathogens (Figure 4).
We confirmed the inhibition of biofilm development by UI-033-EsC by CLSM. Biofilms formed on the inner and outer surface of the catheter pieces were visualized by staining with LIVE/DEAD BacLight™ staining. UI-033-ESC formed biofilms on both sides of the catheter pieces, whether grown in TSB (Figure 5a,b) or AUM (Figure 5e,f). As expected from the more robust biofilm development by UI-033-EsC in TSB compared to AUM (Figure 3), the biofilms formed on both surfaces of the catheter pieces incubated in TSB appear to have more structure than those formed in AUM (Figure 5a,b,e,f). In contrast, we detected no attached bacteria on either surface of the BPA-coated catheter pieces following growth in TSB or AUM (Figure 5c,d,g,h). To confirm these results, we also visualized biofilms formed by UI-038-EsC; similar results were observed—structured biofilm development on outer and inner surfaces of the uncoated catheters when incubated in TSB and less robust biofilm formation in AUM while the BPA-coated catheters showed no biofilm present in TSB or AUM (Figure S2 in Supplementary Material). These results strongly suggest that the BPA coating is effective in preventing biofilm development by bacterial uropathogens on both surfaces of urinary catheters.

2.5. Lipopolysaccharide Contributes to the Tolerance of P. aeruginosa to the BPA-Coated Catheters

The BPA-coated catheters not only failed to inhibit UI-092-PA biofilm development, but the biofilm biomass was also not reduced (Figure 4). To investigate the possibility that this phenomenon is limited to this P. aeruginosa strain only, we examined the effectiveness of the BPA-coated catheter pieces in inhibiting biofilm development by two other P. aeruginosa uropathogens. Biofilm development by UI-040-PA and UI-051-PA on uncoated and BPA-coated catheter pieces was the same (Figure 6a). We then assessed the effect of the BPA-coated catheters on biofilm development by the virulent P. aeruginosa strain UCBPP-PA14 (PA14), which has been utilized extensively in various in vitro and in vivo studies [48,49,50,51,52]. Again, BPA-coated catheter pieces did not affect biofilm development by this strain (Figure 6a). This phenomenon was not limited to only P. aeruginosa; BPA-coated catheters also failed to prevent biofilm development by a P. putida clinical isolate (CF2945-PP) (Figure 6a). One component possibly contributing to Pseudomonas resistance to killing by the BPA-coated catheter (while the other Gram-negative bacilli were susceptible) could be its lipopolysaccharide (LPS). It has been previously shown that in contrast to E. coli, P. aeruginosa is more tolerant to the antibiotic colistin due to the release of large amounts of LPS [53]. To explore the possibility that LPS is a contributing factor, we examined two PA14 mutants deficient in the LPS O-antigen (PA14::wbpM and PA14::66100) [54]. Unlike the parent strain PA14, both mutants were sensitive to the BPA coating (Figure 6b), which suggests that P. aeruginosa tolerance to the BPA coating is due, at least in part, to its LPS.

3. Discussion

CAUTIs are caused by many different bacterial and fungal pathogens, whose prevalence varies according to their location of isolation. In this study, we categorized a pool of 106 Gram-positive and Gram-negative uropathogens from outpatients at the TTUHSC Urology Clinic. Similar to other reports, E. coli was the most common isolate at 30% (Figure 1) [9,55,56]. While the percentages and types of organisms isolated varied from the data reported by the NHSN [9], our data was quite similar to those reported by Khan et al., who analyzed a combination of inpatient and outpatient specimens: 30% E. coli, 21% Klebsiella spp., and 8% P. aeruginosa (Figure 1) vs. 36%, 21%, and 9%, respectively [55].
We were able to obtain the antibiotic resistance profiles for the five Gram-negative bacilli tested, four S. aureus isolates including UI-009-MRSA, and eight Enterococcus spp. (Table 1, Table 2 and Tables S1–S3 in Supplementary Materials). Nine of the seventeen strains met the definition of MDR pathogens. Of the six isolates selected for testing, all of the E. coli isolates, the E. cloacae isolate, and the MRSA isolate were MDR. Surprisingly, the P. aeruginosa isolate, which was resistant to the BPA coating, was not MDR. Despite their resistance to many different classes of antibiotics, the remaining uropathogens tested were sensitive to the BPA-coated silicone catheters (Figure 4), suggesting that BPA is a promising broad-spectrum antimicrobial coating agent. However, to support such an assumption, we need to test the sensitivity of the remaining 100 uropathogens.
Whether bacteria enter during insertion of the catheter or afterwards via the sheath of exudate associated with the catheter (extraluminally), or travel intraluminally along the tubing between the bladder and the collection bag [5,6,7], attachment of the bacteria to either surface (inner or outer) with subsequent biofilm development is a crucial component for initiation and establishment of CAUTIs [11,13,19,57]. Following attachment to the catheter, the uropathogens can produce thick condensed biofilms containing microbial polysaccharides, extracellular substances, and host products [58,59]. These biofilms, which can occlude the catheter, protect the bacteria from natural host defenses and antibiotic activities [19,59,60]. As a result, systemic antibiotics eradicate planktonic bacteria within the bladder but fail to eliminate the bacteria within the biofilm [59,60,61]. Due to the development of bacterial biofilms on the outer and inner surfaces of the catheter, the best approach to prevent CAUTIs is to coat both surfaces of the catheter with antibiofilm/antimicrobial agents [22]. The recovery of no bacteria from the BPA-coated catheters incubated with E. coli, E. cloacae, and S. aureus suggests that the BPA-coated catheters prevented biofilm formation on both surfaces (Figure 4). This was substantiated by the lack of visible biofilm in the CLSM examination of catheters exposed to two different E. coli strains (Figure 5 and Figure S1 in Supplementary Materials).
While approaches to preventing biofilm formation, targeting each stage of development from attachment to maturation and dispersion, have been described [62,63,64,65], the best approach for prevention of CAUTI seems to be preventing attachment in the first place, which is the most likely role of the BPA coating (0.175% BZK) used in this study. We previously showed successful inhibition of biofilm formation by different Gram-positive and Gram-negative pathogens in pre-market testing of NS BlastX wound gel (0.13% BZK, polyethylene glycol 400, polyethylene glycol 3500, sodium citrate, citric acid, and water; per listing on package) and NS Bactisure wound lavage (0.13% BZK) [35,36]. Although their formulations vary, the common active component among these three compounds is BZK, which is a quaternary ammonium compound (QAC) (in this study, PubChem CID 16372, https://pubchem.ncbi.nlm.nih.gov/compound/13762; accessed on 24 October 2022) [66]. QACs are cationic detergents that have been shown to be bactericidal when applied as coatings; essentially killing on contact [64,67,68]. Three mechanisms have been proposed for this contact killing. (1) Penetration of the cell membrane by long cationic polymers [68] or by shorter polymers at high density [69] causes the cell to break open. (2) Electrostatic interaction between the cationic QAC coating and ions on the net negative charge on the bacterial cytoplasmic membrane causes lysis and death [69,70]. (3) Selective adhesion of negatively charged phospholipids in the bacterial cell membrane to the cationic surface causes the leakage of cell contents (the phospholipid sponge effect) [70]. The exact mechanism by which BPA-coated catheters prevent biofilm formation by the different pathogens is not known, but the complete lack of organisms recovered from the biofilm inhibition experiments (Figure 4) suggests that bactericidal contact killing is taking place.
It is unlikely that the additional compounds used to produce the BPA coating made any contribution to the bactericidal contact killing. Polyacrylic acid used in the preparation of the BPA-coated catheters does not have antimicrobial effects (PubChem CID 6581, https://pubchem.ncbi.nlm.nih.gov/compound/6581) (accessed on 24 October 2022) [66], although polyacrylates have been shown to provide broad spectrum resistance to biofilm formation by several different uropathogens [34,71,72]. While the mechanism by which this occurs is not known, it is not due to killing, but is possibly due to the hydrophobicity of polyacrylate and its interaction with the bacterial lipophilic cell wall [71,72]. Polyacrylates can be used to form stable non-leaching coatings [69,73] or they can be used to produce coatings that release antimicrobials impregnated within the coating [34,74]. In contrast, glutaraldehyde, which is also used as chemical intermediate and a cross-linking agent, does have disinfectant properties and has been long used for cold sterilization of heat-sensitive equipment like endoscopes, as a biocide, and as an antimicrobial agent in a number of environments (PubChem CID 3485; https://pubchem.ncbi.nlm.nih.gov/compound/3485) (accessed on 24 October 2022) [66]. As there have been reports of toxicity associated with the use of glutaraldehyde as a disinfectant for endoscopes and other equipment [75,76], glutaraldehyde is unlikely to be an active ingredient.
It is probable that one or more of the three mechanisms described for contact killing (penetration of the cell membrane, electrostatic interaction between the cationic BZK and the negatively charged cell membrane, and selective adhesion of negatively-charged phospholipids to the cationic BZK, [68,69,70]) are involved in the observed prevention of biofilm formation by the MDR strains of E. coli, E. cloacae, and S. aureus. Results from our previous studies suggest that solutions containing BZK are more effective against Gram-positive pathogens than Gram-negative pathogens. Through titration experiments, we previously demonstrated that the minimum inhibitory concentration of such a NS solution (for prevention of biofilm on tympanostomy tubes) for the Gram-negative bacillus P. aeruginosa was 16-fold higher than that for the Gram-positive coccus S. aureus [35]. Similarly, we recently found that the minimum bactericidal concentrations (MBCs), as well as the minimum biofilm inhibition concentrations (MBfIC) of BZK in NS Bactisure, were about 32-fold less for UI-009-MRSA than those for UI-092-PA, and 8- to 16-fold less than those for UIs of E. coli and E. cloacae; additionally, the MBfIC for UI-092-PA was two to four times higher than that of E. coli or E. cloacae (Table S4 and Methods S1 and S2 in Supplementary Materials). Since their cell walls are composed of peptidoglycan and teichoic acid, which are slightly negatively charged, Gram-positive bacteria including S. aureus are more readily killed by BZK at lower concentrations than are Gram-negative ones [77].
However, the P. aeruginosa isolate UI-092-PA, which was the most susceptible to antibiotics, was completely resistant to the BPA coating (Figure 4; Table 2; and Table S1 in Supplementary Materials). The need for a higher concentration of BZK to kill Gram-negative bacteria is due to the presence of lipopolysaccharide (LPS) [78,79,80]. First, the tightly packed LPS molecules considerably slow diffusion of hydrophobic compounds; and second, the presence of divalent cations (calcium and magnesium) within the polyanionic lipid A and the inner core oligosaccharides allow bridging of the negative charges of the LPS molecules, linking them into a tight network [80,81]. Furthermore, the length of the LPS plays a role in susceptibility or resistance to ammonium-based ionic liquids, including QACs; wildtype E. coli and an outer core mutant were the least susceptible while inner core mutants showed elevated sensitivity [79]. The LPS of E. coli (EsC-LPS) and P. aeruginosa (PA-LPS) are structurally different and induce different reactions with the host defensive proteins: EsC-LPS complexed with sheep lung surfactant was more toxic than PA-LPS [82]; PA-LPS bound to human α-1-acid glycoprotein with the greatest affinity compared to EsC-LPS or LPS from Salmonella enterica or Serratia marcescens [83]; and rat cationic defensins effectively killed all the Gram-negative isolates tested, including EsC-LPS mutants, but failed to kill Gram-positive isolates [84]. Therefore, the LPS of Pseudomonas could be responsible for its resistance to BPA (Figure 6a). The observed resistance to BPA among our P. aeruginosa isolates was not a phenomenon of global resistance to BZK, as other BZK-containing NS antimicrobial agents effectively inhibited the growth and biofilm development of different P. aeruginosa clinical isolates and laboratory strains when applied as a solution or gel [35,36,37]. To confirm that the failure to inhibit biofilm development by Pseudomonas was specific to the BPA coating and not due to BZK resistance, we repeated the biofilm inhibition experiments on UI-092-PA and added NS Bactisure to one-half of the MBfIC of BZK to microtiter wells containing BPA-coated catheter pieces. The BPA-coated catheter pieces alone still failed to prevent biofilm formation, but the addition of one-half of the MBfIC of BZK inhibited UI-092-PA biofilm formation on the coated catheter (Figure S3 in Supplementary Materials). In further support of the possibility that the resistance to BPA is connected to the PA-LPS, we found that isogenic LPS mutants of PA14 lacking the O-antigen (a wbpM deletion deficient in O-antigen biosynthesis and a waaL deletion mutant lacking the ligase that links the O-antigen to the LPS core [85]) were completely susceptible to BPA (Figure 6b). It is possible that the length of the hydrophobic alkyl residues (paraffinic chains with eight to eighteen carbon atoms) were long enough to affect MRSA, E. coli, and E. cloacae, but too short to affect P. aeruginosa [69]. Another possibility is that the density of the BZK molecules on the BPA-coated catheters was sufficient to kill the enteric Gram-negative bacteria (E. coli and E. cloacae), but not P. aeruginosa [69]. The synergy obtained when NS Bactisure was added to the BPA-coated catheter supports this possibility (Figure S3 and Table S4 in Supplementary Materials).

4. Materials and Methods

4.1. Bacterial Strains

All urinary isolates (UI) were obtained from patients seen at the TTUHSC Urology Clinic. E. coli (UI-033-EsC, UI-038-EsC, UI-061-EsC), E. cloacae (UI-095-EnC), P. aeruginosa (UI-092-PA, UI-040-PA, UI-051-PA), and methicillin-resistant S. aureus (MRSA) (UI-009-MRSA) were used for the majority of the experiments. To analyze the effect of Pseudomonas LPS on its resistance to the BPA coating, two additional P. aeruginosa UIs (UI-040-PA and UI-051-PA), the P. aeruginosa laboratory strain UCBPP-PA14 (PA14) originally isolated from a burn wound [52], the PA14 LPS mutants PA14/MrT7::PA14_23470 (PA14::wbpM) and PA14/MrT7::PA14_66100 (PA14::66100) (http://pa14.mgh.harvard.edu/cgi-bin/pa14/home.cgi) (accessed on 24 October 2022) [54], and the P. putida strain (CF2954-PP) isolated from sputum of a cystic fibrosis patient (JA Colmer-Hamood, TTUHSC, Lubbock, TX, USA) were used. Gentamicin (50 μg/mL) was added to cultures of PA14::wbpM and PA14::66100. Antibiotic susceptibility profiles were analyzed for the first six strains listed above, three additional S. aureus isolates (UI-073-SA, UI-078-SA, UI-096-SA), six Enterococcus faecalis isolates (UI-031-Efl, UI-065-Efl, UI-080-Efl, UI-090-Efl, UI-099-Efl, UI-100-Efl), and two E. faecium isolates (UI-020-Efm, UI-048-Efm).

4.2. Media, and Growth Conditions

Frozen stock cultures were inoculated into tryptic soy broth (TSB) and incubated overnight at 37 °C ± 0.5 °C with gentle shaking. Aliquots of the overnight cultures were inoculated into fresh TSB and artificial urine media (AUM) to an OD600 of 0.02 for subsequent experiments. The AUM was prepared as described by Brooks and Keevil, 1997 [43]. The diluted cultures (OD600 0.02) were further diluted tenfold in TSB or AUM and 10-μL aliquots were spotted in triplicate on TSB agar plates for quantification of colony forming units (CFU). The CFU/mL were calculated by the following formula:
CFU/mL = average number of colonies in 3 spots × dilution factor × 100
to determine a standard inoculum of ~104 CFU/mL for each strain.

4.2.1. Planktonic Growth in TSB and AUM

An amount of 1-mL aliquots of the diluted suspensions calculated to contain an inoculum of ~104 CFU/mL were pipetted into the wells of a 24-well microtiter plate (Costar untreated, Corning, Durham, NC, USA) and the plate was incubated for 24 h with gentle shaking at 37 °C ± 0.5 °C. After 24 h, the medium in each well was serially diluted tenfold and 10-μL aliquots were spotted in triplicate on Luria-Bertani (LB) agar plates to determine the numbers of CFU/mL.

4.2.2. Biofilm Development on Urinary Catheters

Magic3 intermittent catheters (100% silicone) (Bard, Murray Hill, NJ, USA), Dover red rubber urethral catheters (Cardinal Health, Dublin, OH, USA), and Clean Cath polyvinyl chloride (PVC) urethral catheters (Bard) were used for testing of biofilm development. The sterile silicone, rubber, and PVC catheters were aseptically cut into 1.5-cm pieces for the assays. Biofilm development on the surface of the catheter pieces was conducted using the previously described microtiter plate assay [86]. Experiments were done in triplicate. Catheter pieces were placed individually in wells of a 24-well microtiter plate. The inoculum for each strain was prepared as described above in TSB and AUM to yield 105 CFU/mL, and 1 mL was added to wells containing the catheter pieces. The plates were incubated for 24 h with gentle shaking at 37 °C ± 0.5 °C. After 24 h, the catheter pieces were removed to 1.5 mL microcentrifuge tubes and gently washed twice in 1 mL of phosphate-buffered saline (PBS). Catheters were placed into fresh tubes containing 1 mL of PBS, and the tubes were vortexed for 2 min at 3000 rpm to dislodge the biofilms from the catheter pieces. The suspensions were serially diluted tenfold and 10-μL aliquots were spotted in triplicate on LB agar plates to determine the CFU/catheter piece.

4.3. Coating of Silicone Urinary Catheters with the Biofilm Preventative Agent (BPA)

Sterile 1.5-cm lengths of the silicone catheters were washed for two days in sterile PBS. After washing, the catheter pieces were rinsed in sterile water and dried at room temperature. The washed catheter pieces were aseptically transferred to a Petri dish and the dish was sealed and mailed to Next Science (Jacksonville, FL, USA) for coating. The novel biofilm preventative agent BPA, which contains 0.175% BZK (63349-41-2), 1.99% polyacrylic acid (PAA) (9003-01-4), 2.62% glutaraldehyde (10.47% of 25% glutaraldehyde solution in water) (110-30-8) (all from MilliporeSigma, St. Louis, MO, USA), and 95.215% sterile water for irrigation USP (7732-18-5) (B. Braun Medical, Bethlehem, PA, USA) (personal communication from Ethan Sherman, Next Science), was applied on both the inner and outer surface of the catheter using the simple dipping method. The PAA was mixed with water in a 100 mL flask using a stir bar for 20 min until the PAA was solubilized. Following PAA solubilization, glutaraldehyde solution was added and mixed for 5 min until combined. Finally, the BZK was added to the solution and mixed until the solution cleared. For catheter coating, individual 1.5-cm catheter pieces were dipped into the solution and allowed to soak for 5 min. The catheters were then removed from the solution and strung up to air dry at room temperature for 24 h. Following drying, the samples were re-dipped into the solution for 5 min and air dried again for 24 h (2× coating). After coating, the pieces were returned to our laboratory for testing. Efficacy of the BPA-coating in preventing biofilm formation on the catheter pieces was evaluated using the same method used to develop biofilms on untreated urinary catheters (above).

4.4. Confocal Scanning Laser Microscopy

Biofilms formed on catheter pieces by UI-033-EsC and UI-038-EsC in TSB and AUM were stained using a LIVE/DEAD™ BacLight™ Bacterial Viability kit (Invitrogen, Thermo Fisher Scientific, Waltham, MA, USA). The kit contained two fluorescent dyes, SYTO 9 that stains cells with intact membranes green, and propidium iodide that stains cells with damaged membranes red. Catheter pieces were removed from the microtiter plate and placed in microcentrifuge tubes containing 1 mL of 0.85% NaCl and gently washed twice with an additional 0.85% NaCl. After washing, catheters were placed in 1 mL of 0.85% NaCl, 1.5 μL of SYTO 9, and 2 μL of propidium iodide were added, and the catheters were incubated for 20 min at room temperature in the dark. The catheter pieces were removed from the staining solution and washed twice in 0.85% NaCl, gently blotted on Kimwipes (Kimtech Science, Roswell, GA, USA) to remove excess liquid, and placed in a Petri dish to be visualized. Biofilms on the catheter pieces were visualized using a Nikon Ti-E microscope with A1 confocal and STORM super-resolution modules. (Nikon Instruments, Melville, NY, USA). Images were obtained through a 10X objective with 525/488 and 595/562 filters for SYTO 9 and propidium iodide, respectively. Image stacks were acquired at 2 μm intervals for 200 steps and three-dimensional image reconstructions were carried out using NIS-Elements 5.21.03 (Nikon Instruments). All instrument settings were consistent for each imaging session.

4.5. Statistical Analysis

GraphPad Prism version 9.4.0 (673) (GraphPad Software, San Diego, CA, USA) was used for all statistical analyses. The CFU data were routinely log transformed prior to graphing and statistical analysis. One-way ANOVA was used for comparison of multiple isolates with each other, the three types of catheter material with each other, and the effect of catheter material type on biofilm formation by individual isolates. Tukey’s multiple comparisons posttest was applied to all pairs of pairs or to selected relevant pairs. Unpaired t tests were used to compare individual pairs. Data represent the means ± SEM of three independent experiments for each group (n = 3). Statistical significance is shown as *, p < 0.05; **, p < 0.01; ***, p < 0.001; ****, p < 0.0001; ns, no significant difference.

5. Conclusions

In this study, we have described BPA as a novel broad spectrum biofilm preventative catheter coating. BPA-coated catheters prevented biofilm development by E. coli, E. cloacae, and MRSA, suggesting that this coating provides a broad-spectrum approach to the prevention of CAUTI. Further studies will be required to determine the specific mechanisms by which P. aeruginosa resisted killing by the BPA coating, and to discover a way to prevent this loss of efficacy during the coating process.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/antibiotics11111514/s1, Figure S1. Biofilm biomass formed by the UIs was consistently and most significantly reduced when silicone catheter pieces were incubated in AUM compared to TSB. Figure S2. BPA-coating inhibits biofilm formation by UI-038-ESC on the inner and outer surfaces of the coated catheter. Table S1. Antibiotic resistance patterns of the Gram-negative uropathogens. Table S2. Antibiotic resistance patterns of S. aureus urinary isolates. Table S3. Antibiotic resistance patterns of Enterococcus spp. urinary isolates. Table S4. MBC and MBfIC of BZK in NS Bactisure. Figure S3. NS Bactisure solution and BPA-coated catheters synergize to prevent UI-092-PA biofilm formation. Methods S1. Minimum Bactericidal Concentration (MBC) of Benzalkonium Chloride (BZK) in NS Bactisure. Methods S2. Minimum Biofilm Inhibition Concentration (MBfIC) of Benzalkonium Chloride in NS Bactisure. References [87,88,89,90] are only cited in Supplementary Materials.

Author Contributions

Conceptualization, A.N.H. and T.N.; Formal analysis, S.N., A.N.H. and J.A.C.-H.; Funding acquisition, M.M.; Investigation, S.N.; Methodology, S.N., E.S. and M.M.; Resources, E.S., T.N. and M.M.; Supervision, A.N.H.; Writing—original draft, S.N. and A.N.H.; Writing—review and editing, J.A.C.-H. All authors have read and agreed to the published version of the manuscript.

Funding

This study was supported in part by Next Science (Jacksonville, FL USA). Additional support was provided by the Texas Tech University Health Sciences Center (TTUHSC) Graduate School of Biomedical Sciences, Lubbock, TX, USA, for S.N.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review Board of TTUHSC for collection of urine samples for culture and susceptibility testing from patients presenting at the TTUHSC Urology Clinic. Patients were consented by the staff of the Clinical Research Institute of TTUHSC, and the samples were submitted to the University Medical Center Microbiology Department for testing. Deidentified isolates and their susceptibility results (if applicable) were retrieved from the laboratory by S.N and brought to the research laboratory for the study. No animals were used in this study.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Acknowledgments

The authors thank the Clinical Research Institute at TTUHSC for their help in preparing the IRB protocol and obtaining patient consent. We also thank Joanna E. Swickard for critical reading of the manuscript.

Conflicts of Interest

M.M. is the Chief Technology Officer and E.S. is Director of Research and Development for Next Science LLC (Jacksonville, FL, USA). The Next Science company designed the tested compounds and coated the silicone catheters with BPA. The company had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Levison, M.E.; Kaye, D. Treatment of complicated urinary tract infections with an emphasis on drug-resistant gram-negative uropathogens. Curr. Infect. Dis. Rep. 2013, 15, 109–115. [Google Scholar] [CrossRef] [PubMed]
  2. Lichtenberger, P.; Hooton, T.M. Complicated urinary tract infections. Curr. Infect. Dis. Rep. 2008, 10, 499–504. [Google Scholar] [CrossRef] [PubMed]
  3. Foxman, B. The epidemiology of urinary tract infection. Nat. Rev. Urol. 2010, 7, 653–660. [Google Scholar] [CrossRef]
  4. Newman, D.K.; Cumbee, R.P.; Rovner, E.S. Indwelling (Transurethral and Suprapubic) Catheters. In Clinical Application of Urologic Catheters, Devices and Products; Newman, D.K., Rovner, E.S., Wein, A.J., Eds.; Springer International Publishing: Cham, Switzerland, 2018; pp. 1–45. [Google Scholar] [CrossRef]
  5. Feneley, R.C.; Kunin, C.M.; Stickler, D.J. An indwelling urinary catheter for the 21st century. BJU Int. 2012, 109, 1746–1749. [Google Scholar] [CrossRef] [PubMed]
  6. Newman, D.K.; Wyman, J.F.; Welch, V.W. Core Curriculum for Urologic Nursing; Jannetti Publications: Pitman, NJ, USA, 2016. [Google Scholar]
  7. Vergidis, P.; Patel, R. Novel approaches to the diagnosis, prevention, and treatment of medical device-associated infections. Infect. Dis. Clin. 2012, 26, 173–186. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. Magill, S.S.; O’Leary, E.; Janelle, S.J.; Thompson, D.L.; Dumyati, G.; Nadle, J.; Wilson, L.E.; Kainer, M.A.; Lynfield, R.; Greissman, S.; et al. Changes in prevalence of health care-associated infections in U.S. hospitals. N. Engl. J. Med. 2018, 379, 1732–1744. [Google Scholar] [CrossRef]
  9. Weiner-Lastinger, L.M.; Abner, S.; Edwards, J.R.; Kallen, A.J.; Karlsson, M.; Magill, S.S.; Pollock, D.; See, I.; Soe, M.M.; Walters, M.S.; et al. Antimicrobial-resistant pathogens associated with adult healthcare-associated infections: Summary of data reported to the National Healthcare Safety Network, 2015-2017. Infect. Control. Hosp. Epidemiol. 2020, 41, 1–18. [Google Scholar] [CrossRef] [Green Version]
  10. Tambyah, P.A. Catheter-associated urinary tract infections: Diagnosis and prophylaxis. Int. J. Antimicrob. Agents 2004, 24 (Suppl. 1), S44–S48. [Google Scholar] [CrossRef]
  11. Hooton, T.M.; Bradley, S.F.; Cardenas, D.D.; Colgan, R.; Geerlings, S.E.; Rice, J.C.; Saint, S.; Schaeffer, A.J.; Tambayh, P.A.; Tenke, P.; et al. Diagnosis, prevention, and treatment of catheter-associated urinary tract infection in adults: 2009 International Clinical Practice Guidelines from the Infectious Diseases Society of America. Clin. Infect. Dis. 2010, 50, 625–663. [Google Scholar] [CrossRef] [Green Version]
  12. Lo, E.; Nicolle, L.E.; Coffin, S.E.; Gould, C.; Maragakis, L.L.; Meddings, J.; Pegues, D.A.; Pettis, A.M.; Saint, S.; Yokoe, D.S. Strategies to prevent catheter-associated urinary tract infections in acute care hospitals: 2014 update. Infect. Control. Hosp. Epidemiol. 2014, 35, 464–479. [Google Scholar] [CrossRef] [Green Version]
  13. Warren, J.W. Catheter-associated urinary tract infections. Int. J. Antimicrob. Agents 2001, 17, 299–303. [Google Scholar] [CrossRef]
  14. Hollingsworth, J.M.; Rogers, M.A.; Krein, S.L.; Hickner, A.; Kuhn, L.; Cheng, A.; Chang, R.; Saint, S. Determining the noninfectious complications of indwelling urethral catheters: A systematic review and meta-analysis. Ann. Intern. Med. 2013, 159, 401–410. [Google Scholar] [CrossRef] [PubMed]
  15. Mylotte, J.M. Nursing home-acquired bloodstream infection. Infect. Control. Hosp. Epidemiol. 2005, 26, 833–837. [Google Scholar] [CrossRef] [PubMed]
  16. Flores-Mireles, A.; Hreha, T.N.; Hunstad, D.A. Pathophysiology, treatment, and prevention of catheter-associated urinary tract infection. Top. Spinal Cord Inj. Rehabil. 2019, 25, 228–240. [Google Scholar] [CrossRef] [PubMed]
  17. Arnoldo, L.; Migliavacca, R.; Regattin, L.; Raglio, A.; Pagani, L.; Nucleo, E.; Spalla, M.; Vailati, F.; Agodi, A.; Mosca, A.; et al. Prevalence of urinary colonization by extended spectrum-beta-lactamase Enterobacteriaceae among catheterised inpatients in Italian long term care facilities. BMC Infect. Dis. 2013, 13, 124. [Google Scholar] [CrossRef]
  18. Brennan, B.M.; Coyle, J.R.; Marchaim, D.; Pogue, J.M.; Boehme, M.; Finks, J.; Malani, A.N.; VerLee, K.E.; Buckley, B.O.; Mollon, N.; et al. Statewide surveillance of carbapenem-resistant Enterobacteriaceae in Michigan. Infect. Control Hosp. Epidemiol. 2014, 35, 342–349. [Google Scholar] [CrossRef]
  19. Donlan, R.M.; Costerton, J.W. Biofilms: Survival mechanisms of clinically relevant microorganisms. Clin. Microbiol. Rev. 2002, 15, 167–193. [Google Scholar] [CrossRef] [Green Version]
  20. Hall-Stoodley, L.; Costerton, J.W.; Stoodley, P. Bacterial biofilms: From the natural environment to infectious diseases. Nat. Rev. Microbiol. 2004, 2, 95–108. [Google Scholar] [CrossRef]
  21. Trautner, B.W.; Darouiche, R.O. Role of biofilm in catheter-associated urinary tract infection. Am. J. Infect. Control 2004, 32, 177–183. [Google Scholar] [CrossRef]
  22. Tran, P.L.; Hamood, A.N.; Reid, T.W. Antimicrobial coatings to prevent biofilm formation on medical devices. In Antibiofilm Agents; Rumbaugh, K.P., Ahmad, I., Eds.; Springer: Berlin/Heidelberg, Germany, 2014; Volume 8, pp. 175–204. [Google Scholar]
  23. Pickard, R.; Lam, T.; MacLennan, G.; Starr, K.; Kilonzo, M.; McPherson, G.; Gillies, K.; McDonald, A.; Walton, K.; Buckley, B.; et al. Antimicrobial catheters for reduction of symptomatic urinary tract infection in adults requiring short-term catheterisation in hospital: A multicentre randomised controlled trial. Lancet 2012, 380, 1927–1935. [Google Scholar] [CrossRef] [Green Version]
  24. Gaonkar, T.A.; Caraos, L.; Modak, S. Efficacy of a silicone urinary catheter impregnated with chlorhexidine and triclosan against colonization with Proteus mirabilis and other uropathogens. Infect. Control. Hosp. Epidemiol. 2007, 28, 596–598. [Google Scholar] [CrossRef] [PubMed]
  25. Segev, G.; Bankirer, T.; Steinberg, D.; Duvdevani, M.; Shapur, N.K.; Friedman, M.; Lavy, E. Evaluation of urinary catheters coated with sustained-release varnish of chlorhexidine in mitigating biofilm formation on urinary catheters in dogs. J. Vet. Intern. Med. 2013, 27, 39–46. [Google Scholar] [CrossRef] [PubMed]
  26. Darouiche, R.O.; Mansouri, M.D.; Gawande, P.V.; Madhyastha, S. Antimicrobial and antibiofilm efficacy of triclosan and DispersinB combination. J. Antimicrob. Chemother. 2009, 64, 88–93. [Google Scholar] [CrossRef] [PubMed]
  27. Park, J.H.; Cho, Y.W.; Cho, Y.H.; Choi, J.M.; Shin, H.J.; Bae, Y.H.; Chung, H.; Jeong, S.Y.; Kwon, I.C. Norfloxacin-releasing urethral catheter for long-term catheterization. J. Biomater. Sci. Polym. Ed. 2003, 14, 951–962. [Google Scholar] [CrossRef] [PubMed]
  28. Lim, K.; Chua, R.R.; Bow, H.; Tambyah, P.A.; Hadinoto, K.; Leong, S.S. Development of a catheter functionalized by a polydopamine peptide coating with antimicrobial and antibiofilm properties. Acta Biomater. 2015, 15, 127–138. [Google Scholar] [CrossRef] [PubMed]
  29. Lehman, S.M.; Donlan, R.M. Bacteriophage-mediated control of a two-species biofilm formed by microorganisms causing catheter-associated urinary tract infections in an in vitro urinary catheter model. Antimicrob. Agents Chemother. 2015, 59, 1127–1137. [Google Scholar] [CrossRef] [Green Version]
  30. Lam, T.B.; Omar, M.I.; Fisher, E.; Gillies, K.; MacLennan, S. Types of indwelling urethral catheters for short-term catheterisation in hospitalised adults. Cochrane Database. Syst. Rev. 2014, CD004013. [Google Scholar] [CrossRef]
  31. Andersen, M.J.; Flores-Mireles, A.L. Urinary catheter coating modifications: The race against catheter-associated infections. Coatings 2020, 19, 23. [Google Scholar] [CrossRef] [Green Version]
  32. Saini, H.; Chhibber, S.; Harjai, K. Antimicrobial and antifouling efficacy of urinary catheters impregnated with a combination of macrolide and fluoroquinolone antibiotics against Pseudomonas aeruginosa. Biofouling 2016, 32, 511–522. [Google Scholar] [CrossRef]
  33. Fisher, L.E.; Hook, A.L.; Ashraf, W.; Yousef, A.; Barrett, D.A.; Scurr, D.J.; Chen, X.; Smith, E.F.; Fay, M.; Parmenter, C.D.; et al. Biomaterial modification of urinary catheters with antimicrobials to give long-term broadspectrum antibiofilm activity. J. Control. Release 2015, 202, 57–64. [Google Scholar] [CrossRef]
  34. Burroughs, L.; Ashraf, W.; Singh, S.; Martinez-Pomares, L.; Bayston, R.; Hook, A.L. Development of dual anti-biofilm and anti-bacterial medical devices. Biomater. Sci. 2020, 8, 3926–3934. [Google Scholar] [CrossRef] [PubMed]
  35. Banerjee, D.; Tran, P.L.; Colmer-Hamood, J.A.; Wang, J.C.; Myntti, M.; Cordero, J.; Hamood, A.N. The antimicrobial agent, Next-Science, inhibits the development of Staphylococcus aureus and Pseudomonas aeruginosa biofilms on tympanostomy tubes. Int. J. Pediatr. Otorhinolaryngol. 2015, 79, 1909–1914. [Google Scholar] [CrossRef] [PubMed]
  36. Miller, K.G.; Tran, P.L.; Haley, C.L.; Kruzek, C.; Colmer-Hamood, J.A.; Myntti, M.; Hamood, A.N. Next Science wound gel technology, a novel agent that inhibits biofilm development by gram-positive and gram-negative wound pathogens. Antimicrob. Agents Chemother. 2014, 58, 3060–3072. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  37. Siddiqui, S.Z.; Kruczek, C.; Colmer-Hamood, J.A.; Myntti, M.; Hamood, A.N. Next-Science: A novel antimicrobial agent that inhibits biofilm development by Escherichia coli clinical isolates on urinary tract catheters. J. Med. Microbiol. Diagn 2017, 6, 1–3. [Google Scholar] [CrossRef]
  38. Siegel, J.D.; Rhinehart, E.; Jackson, M.; Chiarello, L. Multidrug-Resistant Organisms (MDRO) Management. Available online: https://www.cdc.gov/infectioncontrol/guidelines/mdro/index.htm (accessed on 29 July 2022).
  39. Waters, A.E.; Contente-Cuomo, T.; Buchhagen, J.; Liu, C.M.; Watson, L.; Pearce, K.; Foster, J.T.; Bowers, J.; Driebe, E.M.; Engelthaler, D.M.; et al. Multidrug-resistant Staphylococcus aureus in US meat and poultry. Clin. Infect. Dis. 2011, 52, 1227–1230. [Google Scholar] [CrossRef]
  40. Horcajada, J.P.; Montero, M.; Oliver, A.; Sorli, L.; Luque, S.; Gomez-Zorrilla, S.; Benito, N.; Grau, S. Epidemiology and treatment of multidrug-resistant and extensively drug-resistant Pseudomonas aeruginosa infections. Clin. Microbiol. Rev. 2019, 32, e00031-19. [Google Scholar] [CrossRef]
  41. Pragasam, A.K.; Raghanivedha, M.; Anandan, S.; Veeraraghavan, B. Characterization of Pseudomonas aeruginosa with discrepant carbapenem susceptibility profile. Ann. Clin. Microbiol. Antimicrob. 2016, 15, 12. [Google Scholar] [CrossRef] [Green Version]
  42. Timsina, R.; Shrestha, U.; Singh, A.; Timalsina, B. Inducible clindamycin resistance and erm genes in Staphylococcus aureus in school children in Kathmandu, Nepal. Future Sci. OA 2020, 7, FSO361. [Google Scholar] [CrossRef]
  43. Brooks, T.; Keevil, C.W. A simple artificial urine for the growth of urinary pathogens. Lett. Appl. Microbiol. 1997, 24, 203–206. [Google Scholar] [CrossRef]
  44. Huang, W.Y.; Wei, L.P.; Ji, Y.G.; Xu, D.X.; Mo, J.K. [Effect of silicon and latex urinary catheters: A comparative study]. Di Yi Jun Yi Da Xue Xue Bao 2005, 25, 1026–1028. [Google Scholar]
  45. Verma, A.; Bhani, D.; Tomar, V.; Bachhiwal, R.; Yadav, S. Differences in bacterial colonization and biofilm formation property of uropathogens between the two most commonly used indwelling urinary catheters. J. Clin. Diagn Res. 2016, 10, PC01–PC03. [Google Scholar] [CrossRef] [PubMed]
  46. Johnson, J.R.; Johnston, B.; Kuskowski, M.A. In vitro comparison of nitrofurazone- and silver alloy-coated foley catheters for contact-dependent and diffusible inhibition of urinary tract infection-associated microorganisms. Antimicrob. Agents Chemother. 2012, 56, 4969–4972. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  47. Singha, P.; Locklin, J.; Handa, H. A review of the recent advances in antimicrobial coatings for urinary catheters. Acta Biomater. 2017, 50, 20–40. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  48. Anderson, G.G.; Moreau-Marquis, S.; Stanton, B.A.; O’Toole, G.A. In vitro analysis of tobramycin-treated Pseudomonas aeruginosa biofilms on cystic fibrosis-derived airway epithelial cells. Infect. Immun. 2008, 76, 1423–1433. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  49. Dzvova, N.; Colmer-Hamood, J.A.; Griswold, J.A.; Hamood, A.N. Heparinase is essential for Pseudomonas aeruginosa virulence during thermal injury and infection. Infect. Immun. 2018, 86, e00755-17. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  50. Friedman, L.; Kolter, R. Genes involved in matrix formation in Pseudomonas aeruginosa PA14 biofilms. Mol. Microbiol. 2004, 51, 675–690. [Google Scholar] [CrossRef] [PubMed]
  51. Korgaonkar, A.; Trivedi, U.; Rumbaugh, K.P.; Whiteley, M. Community surveillance enhances Pseudomonas aeruginosa virulence during polymicrobial infection. Proc. Natl. Acad. Sci. USA 2013, 110, 1059–1064. [Google Scholar] [CrossRef] [Green Version]
  52. Rahme, L.G.; Stevens, E.J.; Wolfort, S.F.; Shao, J.; Tompkins, R.G.; Ausubel, F.M. Common virulence factors for bacterial pathogenicity in plants and animals. Science 1995, 268, 1899–1902. [Google Scholar] [CrossRef] [Green Version]
  53. Yokota, S.; Hakamada, H.; Yamamoto, S.; Sato, T.; Shiraishi, T.; Shinagawa, M.; Takahashi, S. Release of large amounts of lipopolysaccharides from Pseudomonas aeruginosa cells reduces their susceptibility to colistin. Int. J. Antimicrob. Ag. 2018, 51, 888–896. [Google Scholar] [CrossRef]
  54. Liberati, N.T.; Urbach, J.M.; Miyata, S.; Lee, D.G.; Drenkard, E.; Wu, G.; Villanueva, J.; Wei, T.; Ausubel, F.M. An ordered, nonredundant library of Pseudomonas aeruginosa strain PA14 transposon insertion mutants. Proc. Natl. Acad. Sci. USA 2006, 103, 2833–2838. [Google Scholar] [CrossRef]
  55. Khan, M.I.; Xu, S.; Ali, M.M.; Ali, R.; Kazmi, A.; Akhtar, N.; Bilal, M.; Hu, Y.; Li, F. Assessment of multidrug resistance in bacterial isolates from urinary tract-infected patients. J. Radiat. Res. Appl. Sci. 2020, 13, 267–275. [Google Scholar] [CrossRef] [Green Version]
  56. Perez Heras, I.; Sanchez-Gomez, J.C.; Beneyto-Martin, P.; Ruano-de-Pablo, L.; Losada-Pinedo, B. Community-onset extended-spectrum beta-lactamase producing Escherichia coli in urinary tract infections in children from 2015 to 2016: Prevalence, risk factors, and resistances. Medicine 2017, 96, e8571. [Google Scholar] [CrossRef] [PubMed]
  57. Ong, C.L.; Ulett, G.C.; Mabbett, A.N.; Beatson, S.A.; Webb, R.I.; Monaghan, W.; Nimmo, G.R.; Looke, D.F.; McEwan, A.G.; Schembri, M.A. Identification of type 3 fimbriae in uropathogenic Escherichia coli reveals a role in biofilm formation. J. Bacteriol. 2008, 190, 1054–1063. [Google Scholar] [CrossRef] [Green Version]
  58. Flemming, H.C.; Neu, T.R.; Wozniak, D.J. The EPS matrix: The “house of biofilm cells”. J. Bacteriol. 2007, 189, 7945–7947. [Google Scholar] [CrossRef] [Green Version]
  59. Pelling, H.; Nzakizwanayo, J.; Milo, S.; Denham, E.L.; MacFarlane, W.M.; Bock, L.J.; Sutton, J.M.; Jones, B.V. Bacterial biofilm formation on indwelling urethral catheters. Lett. Appl. Microbiol. 2019, 68, 277–293. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  60. Fux, C.A.; Costerton, J.W.; Stewart, P.S.; Stoodley, P. Survival strategies of infectious biofilms. Trends Microbiol. 2005, 13, 34–40. [Google Scholar] [CrossRef]
  61. Gu, H.; Lee, S.W.; Carnicelli, J.; Jiang, Z.; Ren, D. Antibiotic susceptibility of Escherichia coli cells during early-stage biofilm formation. J. Bacteriol 2019, 201, e00034-19. [Google Scholar] [CrossRef] [Green Version]
  62. Ghosh, A.; Jayaraman, N.; Chatterji, D. Small-molecule inhibition of bacterial biofilm. ACS Omega 2020, 5, 3108–3115. [Google Scholar] [CrossRef]
  63. Koo, H.; Allan, R.N.; Howlin, R.P.; Stoodley, P.; Hall-Stoodley, L. Targeting microbial biofilms: Current and prospective therapeutic strategies. Nat. Rev. Microbiol. 2017, 15, 740–755. [Google Scholar] [CrossRef]
  64. Muhammad, M.H.; Idris, A.L.; Fan, X.; Guo, Y.; Yu, Y.; Jin, X.; Qiu, J.; Guan, X.; Huang, T. Beyond risk: Bacterial biofilms and their regulating approaches. Front. Microbiol. 2020, 11, 928. [Google Scholar] [CrossRef]
  65. Roy, R.; Tiwari, M.; Donelli, G.; Tiwari, V. Strategies for combating bacterial biofilms: A focus on anti-biofilm agents and their mechanisms of action. Virulence 2018, 9, 522–554. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  66. Kim, S.; Chen, J.; Cheng, T.; Gindulyte, A.; He, J.; He, S.; Li, Q.; Shoemaker, B.A.; Thiessen, P.A.; Yu, B.; et al. PubChem in 2021: New data content and improved web interfaces. Nucleic Acids Res. 2021, 49, D1388–D1395. [Google Scholar] [CrossRef] [PubMed]
  67. Hasan, J.; Crawford, R.J.; Ivanova, E.P. Antibacterial surfaces: The quest for a new generation of biomaterials. Trends Biotechnol. 2013, 31, 295–304. [Google Scholar] [CrossRef] [PubMed]
  68. Tiller, J.C.; Liao, C.J.; Lewis, K.; Klibanov, A.M. Designing surfaces that kill bacteria on contact. Proc. Natl. Acad. Sci. USA 2001, 98, 5981–5985. [Google Scholar] [CrossRef] [Green Version]
  69. Murata, H.; Koepsel, R.R.; Matyjaszewski, K.; Russell, A.J. Permanent, non-leaching antibacterial surface—2: How high density cationic surfaces kill bacterial cells. Biomaterials 2007, 28, 4870–4879. [Google Scholar] [CrossRef]
  70. Achinas, S.; Charalampogiannis, N.; Euverlink, G.J.W. A brief recap of microbial adhesion and biofilms. Appl. Sci. 2019, 9, 2801. [Google Scholar] [CrossRef] [Green Version]
  71. Dundas, A.A.; Sanni, O.; Dubern, J.F.; Dimitrakis, G.; Hook, A.L.; Irvine, D.J.; Williams, P.; Alexander, M.R. Validating a predictive structure-property relationship by discovery of novel polymers which reduce bacterial biofilm formation. Adv. Mater. 2019, 31, e1903513. [Google Scholar] [CrossRef] [Green Version]
  72. Sanni, O.; Chang, C.Y.; Anderson, D.G.; Langer, R.; Davies, M.C.; Williams, P.M.; Williams, P.; Alexander, M.R.; Hook, A.L. Bacterial attachment to polymeric materials correlates with molecular flexibility and hydrophilicity. Adv. Healthc. Mater. 2015, 4, 695–701. [Google Scholar] [CrossRef] [Green Version]
  73. Tran, P.; Hamood, A.; Mosley, T.; Gray, T.; Jarvis, C.; Webster, D.; Amaechi, B.; Enos, T.; Reid, T. Organo-selenium-containing dental sealant inhibits bacterial biofilm. J. Dent. Res. 2013, 92, 461–466. [Google Scholar] [CrossRef] [Green Version]
  74. Cho, Y.W.; Park, J.H.; Kim, S.H.; Cho, Y.H.; Choi, J.M.; Shin, H.J.; Bae, Y.H.; Chung, H.; Jeong, S.Y.; Kwon, I.C. Gentamicin-releasing urethral catheter for short-term catheterization. J. Biomater. Sci. Polym. Ed. 2003, 14, 963–972. [Google Scholar] [CrossRef]
  75. Hsu, C.W.; Lin, C.H.; Wang, J.H.; Wang, H.T.; Ou, W.C.; King, T.M. Acute rectocolitis following endoscopy in health check-up patients--glutaraldehyde colitis or ischemic colitis? Int. J. Colorectal. Dis. 2009, 24, 1193–1200. [Google Scholar] [CrossRef] [PubMed]
  76. Unal, M.; Yucel, I.; Akar, Y.; Oner, A.; Altin, M. Outbreak of toxic anterior segment syndrome associated with glutaraldehyde after cataract surgery. J. Cataract. Refract. Surg. 2006, 32, 1696–1701. [Google Scholar] [CrossRef] [PubMed]
  77. Sandle, T. Disinfectants. In Encyclopedia of Infection and Immunity; Rezaei, N., Ed.; Elsevier: Amsterdam, The Netherlands, 2022; Volume 4, pp. 630–639. [Google Scholar]
  78. Fox, L.J.; Kelly, P.P.; Humphreys, G.J.; Waigh, T.A.; Lu, J.R.; McBain, A.J. Assessing the risk of resistance to cationic biocides incorporating realism-based and biophysical approaches. J. Ind. Microbiol. Biotechnol. 2022, 49, kuab074. [Google Scholar] [CrossRef] [PubMed]
  79. Gundolf, T.; Rauch, B.; Kalb, R.; Rossmanith, P.; Mester, P. Influence of bacterial lipopolysaccharide modifications on the efficacy of antimicrobial ionic liquids. J. Mol. Liq. 2018, 271, 220–227. [Google Scholar] [CrossRef]
  80. Paracini, N.; Schneck, E.; Imberty, A.; Micciulla, S. Lipopolysaccharides at solid and liquid interfaces: Models for biophysical studies of the Gram-negative bacterial outer membrane. Adv. Colloid. Interface Sci. 2022, 301, 102603. [Google Scholar] [CrossRef]
  81. Ebbensgaard, A.; Mordhorst, H.; Aarestrup, F.M.; Hansen, E.B. The role of outer membrane proteins and lipopolysaccharides for the sensitivity of Escherichia coli to antimicrobial peptides. Front. Microbiol. 2018, 9, 2153. [Google Scholar] [CrossRef] [Green Version]
  82. Brogden, K.A.; Cutlip, R.C.; Lehmkuhl, H.D. Complexing of bacterial lipopolysaccharide with lung surfactant. Infect. Immun. 1986, 52, 644–649. [Google Scholar] [CrossRef] [Green Version]
  83. Huang, J.X.; Azad, M.A.; Yuriev, E.; Baker, M.A.; Nation, R.L.; Li, J.; Cooper, M.A.; Velkov, T. Molecular characterization of lipopolysaccharide binding to human alpha-1-acid glycoprotein. J. Lipids 2012, 2012, 475153. [Google Scholar] [CrossRef] [Green Version]
  84. Kohashi, O.; Ono, T.; Ohki, K.; Soejima, T.; Moriya, T.; Umeda, A.; Meno, Y.; Amako, K.; Funakosi, S.; Masuda, M.; et al. Bactericidal activities of rat defensins and synthetic rabbit defensins on staphylococci, Klebsiella pneumoniae (Chedid, 277, and 8N3), Pseudomonas aeruginosa (mucoid and nonmucoid strains), Salmonella Typhimurium (Ra, Rc, Rd, and Re of LPS mutants) and Escherichia coli. Microbiol. Immunol. 1992, 36, 369–380. [Google Scholar] [CrossRef]
  85. Lam, J.S.; Taylor, V.L.; Islam, S.T.; Hao, Y.; Kocincova, D. Genetic and functional diversity of Pseudomonas aeruginosa lipopolysaccharide. Front. Microbiol. 2011, 2, 118. [Google Scholar] [CrossRef]
  86. Haley, C.L.; Colmer-Hamood, J.A.; Hamood, A.N. Characterization of biofilm-like structures formed by Pseudomonas aeruginosa in a synthetic mucus medium. BMC Microbiol. 2012, 12, 181–201. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  87. Andrews, J.M. Determination of minimum inhibitory concentrations. J. Antimicrob. Chemother. 2001, 48 (Suppl. 1), 5–16. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  88. Balouiri, M.; Sadiki, M.; Ibnsouda, S.K. Methods for in vitro evaluating antimicrobial activity: A review. J. Pharm. Anal. 2016, 6, 71–79. [Google Scholar] [CrossRef] [Green Version]
  89. Ceri, H.; Olson, M.E.; Stremick, C.; Read, R.R.; Morck, D.; Buret, A. The Calgary Biofilm Device: New technology for rapid determination of antibiotic susceptibilities of bacterial biofilms. J. Clin. Microbiol. 1999, 37, 1771–1776. [Google Scholar] [CrossRef] [Green Version]
  90. Wiegand, I.; Hilpert, K.; Hancock, R.E. Agar and broth dilution methods to determine the minimal inhibitory concentration (MIC) of antimicrobial substances. Nat. Protoc. 2008, 3, 163–175. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Distribution of the 106 uropathogenic isolates. Numbers in parentheses indicate numbers of isolates. Klebsiella spp. includes K. pneumoniae (18), K. oxytoca (2), and K. aerogenes (2); other Enterobacterales includes Serratia marcescens (3), Citrobacter freundii (3), C. koseri (2), Proteus mirabilis (2), and Hafnia alvei (1). Enterococcus spp. includes E. faecalis (6) and E. faecium (2). Other GPC isolates include Streptococcus agalactiae (5), S. anginosus (1), Staphylococcus epidermidis (2), Aerococcus urinae (2); GPB isolates are Lactobacillus sp. (1) and Actinomyces neuii (1). GPC, Gram-positive cocci; GPB, Gram-positive bacilli.
Figure 1. Distribution of the 106 uropathogenic isolates. Numbers in parentheses indicate numbers of isolates. Klebsiella spp. includes K. pneumoniae (18), K. oxytoca (2), and K. aerogenes (2); other Enterobacterales includes Serratia marcescens (3), Citrobacter freundii (3), C. koseri (2), Proteus mirabilis (2), and Hafnia alvei (1). Enterococcus spp. includes E. faecalis (6) and E. faecium (2). Other GPC isolates include Streptococcus agalactiae (5), S. anginosus (1), Staphylococcus epidermidis (2), Aerococcus urinae (2); GPB isolates are Lactobacillus sp. (1) and Actinomyces neuii (1). GPC, Gram-positive cocci; GPB, Gram-positive bacilli.
Antibiotics 11 01514 g001
Figure 2. Artificial urine medium supports the growth of clinical uropathogens. Clinical isolates were grown in the microtiter plate assay to assess planktonic growth in tryptic soy broth (TSB) or artificial urine medium (AUM). Colony-forming units (CFU) were determined after 24 h at 37°C. Data were transformed to log10 before graphing. Bars represent the average of three independent experiments ± SEM; ns, no significance; *** p < 0.001, **** p < 0.0001.
Figure 2. Artificial urine medium supports the growth of clinical uropathogens. Clinical isolates were grown in the microtiter plate assay to assess planktonic growth in tryptic soy broth (TSB) or artificial urine medium (AUM). Colony-forming units (CFU) were determined after 24 h at 37°C. Data were transformed to log10 before graphing. Bars represent the average of three independent experiments ± SEM; ns, no significance; *** p < 0.001, **** p < 0.0001.
Antibiotics 11 01514 g002
Figure 3. Artificial urine medium supports the growth of clinical uropathogens and their formation of biofilm on urinary catheters. Biofilms were allowed to develop for 24 h on 1.5-cm lengths of urinary catheters made of silicone (S), rubber (R), or PVC (P) incubated in TSB (a) or AUM (b). The amounts of biofilm formed by each isolate on each type of catheter material were compared. For five of the six isolates, biofilm biomass on silicone is significantly less than that formed on PVC; for four of the six isolates, biofilm biomass on silicone is less than that formed on rubber. Each dot represents an independent experiment; ns, no significance; * p < 0.05; ** p < 0.01; *** p < 0.001, **** p < 0.0001. Black asterisks indicate significant differences for each isolate and the different catheter materials; magenta asterisks indicate significant differences in biofilm biomass of UI-009-MRSA on each type of catheter material compared to the other bacteria.
Figure 3. Artificial urine medium supports the growth of clinical uropathogens and their formation of biofilm on urinary catheters. Biofilms were allowed to develop for 24 h on 1.5-cm lengths of urinary catheters made of silicone (S), rubber (R), or PVC (P) incubated in TSB (a) or AUM (b). The amounts of biofilm formed by each isolate on each type of catheter material were compared. For five of the six isolates, biofilm biomass on silicone is significantly less than that formed on PVC; for four of the six isolates, biofilm biomass on silicone is less than that formed on rubber. Each dot represents an independent experiment; ns, no significance; * p < 0.05; ** p < 0.01; *** p < 0.001, **** p < 0.0001. Black asterisks indicate significant differences for each isolate and the different catheter materials; magenta asterisks indicate significant differences in biofilm biomass of UI-009-MRSA on each type of catheter material compared to the other bacteria.
Antibiotics 11 01514 g003
Figure 4. BPA-coating on silicone catheters inhibits biofilm formation by most of the uropathogens. The BPA coating was not effective in preventing biofilm formation by UI-092-PA. Values represent the average of three independent experiments ± SEM; ns, no significance; **** p < 0.0001.
Figure 4. BPA-coating on silicone catheters inhibits biofilm formation by most of the uropathogens. The BPA coating was not effective in preventing biofilm formation by UI-092-PA. Values represent the average of three independent experiments ± SEM; ns, no significance; **** p < 0.0001.
Antibiotics 11 01514 g004
Figure 5. BPA-coating inhibits biofilm formation by UI-033-ESC on the inner and outer surfaces of the coated catheter. Biofilm development was visualized using CSLM following staining with the LIVE/DEAD BacLight™ Bacterial Viability kit; live bacteria are stained green and dead bacteria are stained red. Outer (a) and inner (b) surfaces of uncoated catheters and outer (c) and inner (d) surfaces of BPA-coated catheters incubated in TSB. Outer (e) and inner (f) surfaces of uncoated catheters and outer (g) and inner (h) surfaces of BPA-coated catheters incubated in AUM. Yellow bars = 100 μm; images are representative of three independent experiments.
Figure 5. BPA-coating inhibits biofilm formation by UI-033-ESC on the inner and outer surfaces of the coated catheter. Biofilm development was visualized using CSLM following staining with the LIVE/DEAD BacLight™ Bacterial Viability kit; live bacteria are stained green and dead bacteria are stained red. Outer (a) and inner (b) surfaces of uncoated catheters and outer (c) and inner (d) surfaces of BPA-coated catheters incubated in TSB. Outer (e) and inner (f) surfaces of uncoated catheters and outer (g) and inner (h) surfaces of BPA-coated catheters incubated in AUM. Yellow bars = 100 μm; images are representative of three independent experiments.
Antibiotics 11 01514 g005
Figure 6. Pseudomonas species are tolerant to BPA-coated catheters. (a) Uncoated and BPA-coated catheters were placed in AUM inoculated with two additional P. aeruginosa UTI isolates (UI-040-PA, UI-051-PA), the virulent P. aeruginosa laboratory strain (PA14), and a respiratory isolate of P. putida (CF2954-PP). (b) Isogenic LPS mutants of PA14 are sensitive to the BPA-coated catheters. Uncoated and BPA-coated catheters were placed in AUM inoculated with PA14 (parent strain) and its isogenic mutants PA14::wbpM and PA14::66100. Bars represent the average of three independent experiments ± SEM; ns, no significance; *** p < 0.001; **** p < 0.0001.
Figure 6. Pseudomonas species are tolerant to BPA-coated catheters. (a) Uncoated and BPA-coated catheters were placed in AUM inoculated with two additional P. aeruginosa UTI isolates (UI-040-PA, UI-051-PA), the virulent P. aeruginosa laboratory strain (PA14), and a respiratory isolate of P. putida (CF2954-PP). (b) Isogenic LPS mutants of PA14 are sensitive to the BPA-coated catheters. Uncoated and BPA-coated catheters were placed in AUM inoculated with PA14 (parent strain) and its isogenic mutants PA14::wbpM and PA14::66100. Bars represent the average of three independent experiments ± SEM; ns, no significance; *** p < 0.001; **** p < 0.0001.
Antibiotics 11 01514 g006aAntibiotics 11 01514 g006b
Table 1. Antibiotic susceptibility patterns of the five Gram-negative uropathogens.
Table 1. Antibiotic susceptibility patterns of the five Gram-negative uropathogens.
AntibioticMIC in μg/mL *
E. coliE. cloacaeP. aeruginosa
UI-033-EsCUI-038-EsCUI-061-EsCUI-095-EnCUI-092-PA
Amikacin≤4, S *>32, R *≤4, S≤4, S≤4, S
Amoxicillin/clavulanate16/8, R8/4, S8/4, S>16/8, R
Ampicillin>16, R>16, R>16, R>16, R
Aztreonam>16, R>16, R>16, R>16, R≤1, S
Cefazolin>32, R>32, R>32, R>32, R
Cefepime>16, R16, R16, R4, I *1, S
Cefoxitin16, I16, I≤4, S>16, R
Ceftazidime>16, R>16, R8, R>16, R1, S
Ceftriaxone>32, R>32, R>32, R>32, R
Cefuroxime>16, R>16, R>16, R>16, R
Ciprofloxacin>2, R>2, R>2, R>2, R2, I
Gentamicin>8, R>8, R≤1, S1, S≤0.5, S
Imipenem≤0.25, S≤0.25, S≤0.25, S8, R
Levofloxacin>4, R>4, R>4, R>4, R4, I
Meropenem≤0.125, S≤0.125, S≤0.125, S≤0.125, S1, S
Moxifloxacin>4, R>4, R>4, R>4, R
Nalidixic Acid>32, R>32, R>32, R§§
Nitrofurantoin≤16, S≤16, S≤16, S>64, R
Piperacillin/tazobactam8/4, S4/4, S8/4, S>64/4, R≤2/4, S
Tetracycline>8, R>8, R>8, R>8, R
Tigecycline≤0.5, S≤0.5, S≤0.5, S2, S
Tobramycin>8, R>8, R>8, R1, S≤0.5, S
Trimethoprim/sulfamethoxazole≤0.5/9.5, S>2/38, R>2/38, R>2/38, R
* MIC: the minimum inhibitory concentration of each antibiotic was determined by the University Medical Center Clinical Microbiology Laboratory; S, susceptible; I, intermediate (light orange); R, resistant (orange); † Antibiotic not tested as P. aeruginosa is innately resistant; ‡ Antibiotic not tested by laboratory; § Antibiotic not tested as relevant only for E. coli.
Table 2. Antibiotic susceptibility pattern of UI-009-MRSA.
Table 2. Antibiotic susceptibility pattern of UI-009-MRSA.
AntibioticMIC in μg/mL *
ClindamyciniCR †
Daptomycin≤1, S
ErythromyciniCR
Gentamicin≤1, S
Levofloxacin≤1, S
Linezolid2, S
Nitrofurantoin≤16, S
Oxacillin>2, R
Penicillin>1, R
Quinupristin/dalfopristin≤0.5, S
Rifampin≤0.5, S
Tetracycline≤0.5, S
Trimethoprim/sulfamethoxazole≤0.5/9.5, S
Vancomycin1, S
* MIC: the minimum inhibitory concentration of each antibiotic was determined by the University Medical Center Clinical Microbiology Laboratory; S, susceptible; R, resistant (orange); † iCR, inducible clindamycin resistance [42].
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Navarro, S.; Sherman, E.; Colmer-Hamood, J.A.; Nelius, T.; Myntti, M.; Hamood, A.N. Urinary Catheters Coated with a Novel Biofilm Preventative Agent Inhibit Biofilm Development by Diverse Bacterial Uropathogens. Antibiotics 2022, 11, 1514. https://doi.org/10.3390/antibiotics11111514

AMA Style

Navarro S, Sherman E, Colmer-Hamood JA, Nelius T, Myntti M, Hamood AN. Urinary Catheters Coated with a Novel Biofilm Preventative Agent Inhibit Biofilm Development by Diverse Bacterial Uropathogens. Antibiotics. 2022; 11(11):1514. https://doi.org/10.3390/antibiotics11111514

Chicago/Turabian Style

Navarro, Stephany, Ethan Sherman, Jane A. Colmer-Hamood, Thomas Nelius, Matthew Myntti, and Abdul N. Hamood. 2022. "Urinary Catheters Coated with a Novel Biofilm Preventative Agent Inhibit Biofilm Development by Diverse Bacterial Uropathogens" Antibiotics 11, no. 11: 1514. https://doi.org/10.3390/antibiotics11111514

APA Style

Navarro, S., Sherman, E., Colmer-Hamood, J. A., Nelius, T., Myntti, M., & Hamood, A. N. (2022). Urinary Catheters Coated with a Novel Biofilm Preventative Agent Inhibit Biofilm Development by Diverse Bacterial Uropathogens. Antibiotics, 11(11), 1514. https://doi.org/10.3390/antibiotics11111514

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop