Review

The Versatility of Biological Field-Effect Transistor-Based Biosensors (BioFETs) in Point-of-Care Diagnostics: Applications and Future Directions for Peritoneal Dialysis Monitoring

1
Department of Biomedical Engineering, Faculty of Engineering, The Hong Kong Polytechnic University, Hong Kong
2
Department of Medicine, Alice Ho Miu Ling Nethersole Hospital, Hong Kong
3
School of Medicine and Pharmacy, Ocean University of China, Qingdao 266003, China
4
Research Institute for Smart Ageing, The Hong Kong Polytechnic University, Hong Kong
*
Authors to whom correspondence should be addressed.
Biosensors 2025, 15(3), 193; https://doi.org/10.3390/bios15030193
Submission received: 17 January 2025 / Revised: 10 March 2025 / Accepted: 16 March 2025 / Published: 18 March 2025
(This article belongs to the Special Issue Microelectronics and MEMS-Based Biosensors for Healthcare Application)

Abstract

:
Peritoneal dialysis (PD) is a vital treatment for end-stage renal disease patients, but its efficacy is often compromised by complications such as infections and peritoneal fibrosis. Biological field-effect transistors (BioFETs) present a promising solution for rapid, sensitive, and non-invasive detection of indicators and biomarkers associated with these complications, potentially enabling early intervention. However, BioFETs are yet to be adopted for PD monitoring. This review presents a forward-looking analysis of the capacity and potential integration of BioFETs into PD management systems, highlighting their capacity to monitor both routine indicators of dialysis efficiency and metabolic status, as well as specific biomarkers for complications such as inflammation and fibrosis. We examine the challenges in adapting BioFETs for PD applications, focusing on key areas for improvement, including sensitivity, specificity, stability, reusability, and clinical integration. Furthermore, we discuss various approaches to address these challenges, which are crucial for developing point-of-care (PoC) and multiplexed wearable devices. These advancements could facilitate continuous, precise, and user-friendly monitoring, potentially revolutionizing PD complication management and enhancing patient care.

1. Introduction

Peritoneal dialysis (PD) is a home-based therapy for kidney failure [1], which is primarily used for patients with chronic kidney disease (CKD) to help preserve the residual renal function, particularly in the late stages—end-stage renal disease (ESRD)—when kidney function drops to 10–15% of its normal capacity [2]. CKD affects approximately 700 million people worldwide [3], representing about 10% of the population [4]. PD accounts for around 11% of global dialysis therapies [5] and PD patients often report better health-related quality of life [6]. This is largely due to the greater autonomy and flexibility PD offers [7], allowing patients to perform the treatment at home, especially during the initial years of treatment [8].
Despite these advantages, PD is primarily challenged by infection and peritoneal fibrosis (PF). Infection-related problems, including peritonitis and catheter infection, occur in 20% to 40% of cases depending on the region [9]. Patients undergoing prolonged PD treatment are at risk of developing PF, with incidence rates ranging from 0.5% to 19.4% depending on the duration of therapy. As PF progresses, it leads to a decline in dialysis efficiency, and affected patients may no longer be able to use PD. These patients are then forced to return to hemodialysis (HD), which is more cumbersome and costly [10,11]. Additionally, patient-specific factors, including reduced peritoneal membrane longevity, declining ultrafiltration capacity [12], overhydration [13], and hypokalemia [6], further contribute to the eventual treatment failure. Therefore, real-time monitoring and early detection of complications are crucial for delivering targeted interventions and minimizing the risk of PD failure [14].
To meet these challenges, PD patients undergo regular prescriptions and follow-ups, including assessments of peritoneal transport status, solute clearance, and screening for complications [15]. Peritoneal biopsy requires surgical procedures to obtain peritoneal tissue, making it an invasive procedure [16,17]. The peritoneal equilibration test (PET) is commonly used to evaluate the peritoneal solute transfer rate (PSTR) and optimize solute clearance according to the kidney status [18]. However, this is a time-consuming process, as it involves instilling dialysate fluid into the peritoneal cavity and collecting samples at specific intervals over a 4 h period to monitor the transfer of solutes across the peritoneal membrane [19]. Non-invasive and efficient alternatives focus on the analysis of peritoneal dialysis effluent (PDE), which serves as a rich source of indicators and biomarkers, such as glucose, urea, creatinine [16], cancer antigen 125 (CA125) [17], and lnterleukin-6 (IL-6) [20], which can help predict clinical PD outcomes [21].
The International Society of Peritoneal Dialysis (ISPD) recommends screening PDE through cell culture, Gram stain, white blood cell count, and clinical feature observation [22]. However, these methods are time-consuming and complex to perform, leading to the use of empirical antibiotics as an initial treatment before the infection source is confirmed [23]. This delay poses a challenge for timely intervention, as the risk of complications and death increases by 5.5% for each additional hour of delay in treatment [23]. So researchers have been exploring proteomics and metabolomics approaches for PDE analysis [24], including nuclear magnetic resonance (NMR)-based identification of metabolites [25], as well as proteomics techniques such as Western blot [26], ELISA, and two-dimensional differentiation gel electrophoresis [10]. Other novel technologies, such as polymerase chain reaction/electrospray ionization mass spectrometry (PCR/ESI-MS) and 16S rRNA gene sequencing, have been developed to quickly detect pathogens in PDE as well [27]. However, PDE metabolomics may differ significantly depending on patient peritoneal type [25]. Proteomics are often interfered by high-abundance proteins, necessitating sample pretreatment, which slows clinical translation [10]. PCR/ESI-MS lacks accuracy [28], and gene sequencing struggles to differentiate pathogens with high genetic similarity [29]. Furthermore, all these techniques require sophisticated instruments and laboratory conditions, which makes them less accessible and incapable of being used for point-of-care (PoC) diagnostics for PD patients. This limitation highlights the need for efficient, non-invasive, and label-free detection platforms that can enable PoC testing, thereby minimizing the risk of death or complications caused by delays in infection treatment [30].
In response to the limitations of current diagnostic methods, transistor-based biosensors emerged as a promising avenue for advancing PD monitoring. These transistors primarily fall into two categories: bipolar junction transistors (BJTs) and biological field-effect transistors (BioFETs). BioFETs consist of a semiconductor channel connecting source and drain electrodes, with a gate electrode that modulates the channel’s conductivity via an applied voltage [31]. BJTs, on the other hand, are three-terminal devices composed of two p-n junctions, where a small current injected into the base controls a larger current between the collector and emitter. While BioFETs rely on voltage-controlled modulation of channel conductivity, BJTs offer current-controlled amplification [32]. Recent research suggests that BJTs can surpass BioFETs in certain performance aspects, including achieving simpler calibration curves independent of applied voltage, lower signal-to-noise ratios, and higher sensitivity. For example, researchers demonstrated a heterojunction MoTe2/GeSe/MoTe2 BJT capable of highly sensitive detection of diverse proteins and DNA, achieving a limit of detection (LOD) of 5 pM for streptavidin within 10 s—a performance exceeding many existing FET-based detection methods [33]. Similarly, BJTs have been used to detect C-reactive protein (CRP) with ultra-low detection limits (1 pmol/L), surpassing the sensitivity of many FET-based CRP sensors [34,35]. Despite these advantages, the application of BJTs in the medical field, and particularly in biosensing, remains less prevalent than that of FETs, with current BJT applications primarily focused on areas such as X-ray radiation therapy [36], pressure sensing [37], and protein detection [33]. Although BJTs offer potential for point-of-care (PoC) diagnostics and demonstrate superior performance in some areas, including the possibility of atomic-layer thickness [33], a critical challenge lies in scaling down BJT fabrication to below 100 nanometers in an easy method for mass production and widespread commercialization [38]. Given that achieving ultra-low detection limits often depends on maximizing sensor surface area, highly miniaturized BioFETs may offer a more practical approach in this specific context. Therefore, this review centers on BioFETs, exploring their potential for addressing the unique challenges of PoC PD monitoring while acknowledging the exciting possibilities offered by BJT technology. While PET is valuable for assessing long-term peritoneal membrane function, BioFETs could provide real-time monitoring of key biomarkers, alerting clinicians to acute changes that might be missed by infrequent PET testing. BioFETs could offer a bacterial detection limit over 10,000 times lower than electrochemical impedance spectroscopy (EIS) and matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-ToF) using the same modified surfaces [39]. For PD patients, BioFETs-based monitoring is non-invasive, while the lower manufacturing costs make them an affordable option [40]. The miniaturization capabilities of BioFETs enable PoC applications [41], allowing patients to perform home-based self-testing and see the results directly from wireless communication modules such as an app [42,43]. These advantages support personalized treatment plans, as regular monitoring of residual renal function can be tailored to meet the unique needs of individuals [44].
BioFETs have been successfully applied in various fields [45], demonstrating their effective capabilities for integration with existing PD monitoring. These applications mainly include antigen detection for communicable diseases [46], particularly COVID-19 [47] and influenza [48], as well as biomarker detection for non-communicable diseases [49], mainly cancers [50], cardiovascular conditions [51,52], and neurodegenerative disorders [53]. Additionally, BioFETs are utilized in drug screening through monitoring cell membrane potential changes [54]. Despite their success in other areas, their direct use in PD monitoring remains limited and underexplored [55,56]. Since PD monitoring requires long-term follow-up [20], the integration of BioFETs faces several challenges, including the stringent demands on stability, sensitivity, specificity, and the need for consistent performance. Many BioFETs remain at the laboratory stage [57], and the complex composition of PDE effluent and bodily fluid [58], along with issues related to long-term device performance [59], manufacturing costs [31], and system integration [60], further complicates their application in routine PD monitoring. Several reviews explored various aspects of BioFETs, including their configuration [61], transducing materials, probes [62], and applications in wearable healthcare technologies [52]. However, the potential of BioFETs specifically for PD monitoring remains largely unexplored.
The primary objective of this review is to critically evaluate the adaptability of BioFETs for PD monitoring, focusing on three key areas: (1) the design and optimization of BioFETs components (probes, transducing materials) for PD-specific biomarker detection; (2) the technical and clinical barriers to integrating BioFETs into PD workflows; and (3) emerging strategies (e.g., multiplexing, PoC) to address these challenges. By synthesizing advances in BioFETs engineering with unmet clinical demands in PD, this work delineates a roadmap for translating laboratory innovations into practical, patient-centric diagnostic tools. The successful integration of BioFETs into PD management could revolutionize personalized medicine, offering tailored interventions for individual patients and greatly improving the quality of life and long-term outcomes for those with ESRD undergoing PD. To provide a conceptual overview, Figure 1 illustrates the interplay between the core components of BioFETs and their applications in PD monitoring.
The outline of this review is as follows: The next section delves into the core components of BioFETs, focusing on biosensing elements and transducing materials, which are crucial for improving device performance. Section 3 and Section 4 explore the functions of BioFETs applicable to PD monitoring and assess the technical and clinical barriers hindering their integration into PD management. Lastly, this work explores solutions to overcome these challenges for the improvement of usability and clinical applicability, including optimized device design, multiplexing capabilities, wearable BioFETs for portable PoC testing, and advances in fabrication techniques. These innovations demonstrate how BioFETs can be effectively integrated with other approaches to enhance PD monitoring.

2. Overview of BioFETs

2.1. Introduction of BioFETs

BioFETs are fundamentally voltage-controlled amplifiers [63] widely employed in biosensor design for analyte detection, with applications spanning environmental monitoring, agriculture, the food industry, and particularly the medical field [61]. In typical BioFETs, the source and drain electrodes are connected via semiconductor channels, with a gate electrode controlling the flow of charge carriers by applying a bias potential [64]. Biosensing elements are immobilized on the semiconductor channels of BioFETs to specifically capture the target molecules [62]. When a target binds to a probe, this interaction induces a change in the charge carrier density within the channel, altering its conductivity. This shift in conductivity, driven by the binding reaction, can be easily detected and measured, allowing for the sensitive detection of specific biomolecular interactions [49,61,65]. Figure 2 illustrates the operational workflow of BioFETs, encompassing their structure, functionalization, signal output, and user interface.
BioFETs enable label-free, real-time electrical detection of biomolecules, offering advantages in direct signal transduction without requiring fluorescent labeling or optical instrumentation [66]. While other systems, such as fluorescence-based systems, also excel in single-molecule sensitivity, BioFETs are particularly suited for applications demanding miniaturization, continuous monitoring, and integration with portable electronic systems [31,42,67]. Taking these advantages, BioFETs have already been applied in various healthcare areas, including precision medicine and the early detection of biomarkers such as urea [55], glucose [68], and creatinine [69]. However, the commercial deployment of BioFETs remains limited due to issues such as the development of disposable devices, difficulties in large-scale production [57], low biocompatibility [70], and especially Debye screening effect [71]. BioFETs rely on detecting the electrostatic potential of charged biomolecules, but in physiological solutions, mobile ions rapidly screen these charges over a characteristic distance called the Debye length (κ−1). If the target analyte is located beyond this length—typically <1 nm in physiological conditions—its signal is effectively shielded, limiting the sensor’s ability to detect larger biomolecules and reducing sensitivity in clinical applications [72,73]. So, physiological samples analyzed using BioFETs usually require dilution or pretreating, but recent advancements proposed solutions to streamline this, as is later detailed in the following sections.
The performance of BioFETs is also determined by several other factors. Depending on the biometric event and sensor operation parameters, BioFETs sensitivity formulation can be expressed as [74]:
d I D I D = d C d σ d C · d V E G d σ · d I D / d V E G I D
where dC is the change in the analyte concentration, is the change in the sensor conductivity due to the charge transfer process, and dVEG refers to gate voltage change, which results in the variation in drain current ID. The first half of the formula represents a change in the sensor conductivity caused by a biological reaction, expressing the response to the presence of the analyte. The second half represents a change in conductivity during signal transduction that changes the gate threshold voltage, resulting in a detectable change in leakage current ID. Thus, biosensing elements and transducing materials can be spotted as two main factors when designing BioFETs [62], which determine their sensitivity, selectivity, limit of detection (LOD), stability, and reusability [75]. Structural parameters including channel length, gate insulator thickness, electrode metal type, channel material thickness, and gating structure contribute to the device performance as well (Figure 3) [76]. While general designs of BioFETs are covered in other literature [74], our emphasis here is on the biosensing elements and transducing materials employed in BioFETs for the PD-related biomarkers detection, demonstrating their suitability for PD diagnostics.

2.2. Biosensing Elements

BioFETs are highly sensitive to the surrounding environment but have low selectivity. To address this, biosensing elements which have high specificity for the target are usually immobilized on the BioFET surface through appropriate connectors to capture the targets [70]. Based on different recognition elements and binding principles, they can be divided into enzyme-based, DNA-based, aptamer-based, and immune-based FET biosensors (Figure 4). They have been used for the biomarker detection according to different mechanisms, for instance, enzyme-based BioFETs are used to detect glucose and urea through enzyme-catalyzed reactions; DNA probe-based BioFETs can identify pathogen genes during PD peritonitis through base complementarity; aptamer-based BioFETs can overcome the Debye screening length in PDE with a smaller size; immune reaction-based BioFETs can recognize cytokines or bacteria through specific antigen–antibody binding. This section selects the four most common probe types as the basis to illustrate how they can be applied for PD monitoring (Figure 5).

2.2.1. Enzyme-Based BioFETs

Enzyme-based BioFETs (EnFETs) have been broadly adapted in the detection of PD-related biomarkers, especially glucose [81,82]. EnFETs typically operate based on enzyme-catalyzed reactions, where enzymes immobilized on the channel surface facilitate the conversion of substrates into products. This reaction occurs within the enzyme membrane, inducing alterations in the accumulation of charged carriers at the gate surface, along with the generation or consumption of protons in the channel, leading to pH changes. These variations can then be used to quantify the target analyte [43]. In glucose detection, glucose oxidase (GOD) fixed on the EnFETs catalyzes the oxidation of glucose, generating positive charges. The accumulation of these positive charges on the gate of the n-type FET enhances the carrier concentration in the channel, leading to an increase in drain current, which enables the detection of glucose molecules [81]. Real-time urea detection in dialysis follows a similar principle. Urease was immobilized on a graphene-based EnFET using the EDC/NHA method to specifically recognize urea. In this system, urease catalyzes the hydrolysis of urea into carbon dioxide and ammonia, increasing the local pH and triggering a potential change within minutes [55]. This rapid response has shown significant potential for improving the efficiency of indirect urea clearance estimation for clinical PD monitoring.
These enzymes are popular because of their high specificity, availability [81], and even reversible reactions, which facilitate the reusability of BioFETs [83]. However, they are prone to deactivation when exposed to fluctuations in pH, temperature, and humidity [84]. This situation makes the immobilization of enzymes on BioFETs quite challenging, especially in biofluids such as PDE [85]. While the immobilization method and uniformity of immobilization can have a huge effect on the sensitivity fluctuations [81], new strategies have been conducted to deal with these problems. For examples, nanocomposites such as potassium-doped carbon nanotubes (CNT) have been used as the enzyme loading matrix [86], while Au nanoparticles (AuNPs) and polyelectrolytes have been employed for multilayer enzyme encapsulation to enhance system stability [87].
To improve the biocompatibility of the system, biomimetic materials are utilized as substitutes for traditional coatings. Silk fibroin (SF) hydrogel had been integrated with graphene BioFETs (GFETs) for glucose detection in physiological solution [84]. The enzymatically cross-linked SF serves as a carrier for GOD, effectively encapsulating the GFET while prolonging the lifetime of GOD and reducing non-specific adhesion of interferents. This device demonstrates high application potential with a LOD as low as 200 nM and a broad detection range from 1µM to 10 mM for glucose sensing, thereby fully encompassing the clinical diagnostic range for glucose [84]. Using a same-packaging approach, GOD was incorporated into an APTMS: PEG complex to create a porous biopolymer layer on the surface of silicon nanowire (SiNW) BioFETs. This “hybrid catalytic layer” maintains a sufficient density and activity of GOD on the BioFET surface while also increasing the Debye screening length, facilitating direct glucose detection in high-ionic-strength solutions. The device exhibits a dynamic glucose detection range of 10 nM~10 mM and a response time of less than 8 s, indicating high practical value, particularly in solutions with elevated ionic concentrations [77].

2.2.2. DNA-Based BioFETs

In typical DNA probe-based BioFETs, single-stranded DNA (ssDNA) probes are usually immobilized on the BioFETs channel to hybridize the target DNA through complementary base pairing to alter the electron mobility and generate an electricity signal [70]. This approach overcomes the limitations of an optical-based real-time polymerase chain reaction (RT-PCR), which often requires toxic fluorescent dyes and unstable light source signals [51]. More importantly, it offers the potential for direct and label-free detection of DNA/RNA molecules without the need for time-consuming amplification processes [88,89], even down to the level of individual viral long RNA [90].
These properties make DNA-based BioFETs valuable for PD monitoring, including the detection of DNA or 16S rRNA for screening pathogens associated with prolonged catheter placement [91], mitochondrial DNA detection as indicators of peritoneal injury [92], and the assessment of elevated levels of pathological microRNA (miRNA) [93]. Notably, DNA-based BioFETs have been designed to detect Gram-negative bacterial (GNB) E. coli O157:H7, which is known to cause peritonitis. In this case, an ssDNA probe based on the E. coli intimin gene was covalently bound to the gold gate surface through a reduction reaction, enabling specific binding to the complementary DNA sequence (cDNA) in the target bacteria. The binding of negatively charged complementary DNA leads to the redistribution of the double-layer capacitance in the n-type BioFETs, resulting in a decrease in gate voltage and drain current, which enables the detection of the target [94].
Furthermore, studies have shown that it is possible to distinguish sequences that differ by a single base, which facilitates the identification of bacterial variants induced by peritoneal dialysis fluid (PDF), which are often underestimated [95]. In this context, ultra-thin film indium oxide has been utilized as the semiconductor channel material, with thiolated ssDNA probe fixed on the indium oxide channel via an amine–thiol linker to selectively capture the target sequence. The detection capabilities of BioFETs were then assessed by exposing them to non-complementary, fully complementary, and mismatched sequences, respectively. The results demonstrate that hybridization with fully complementary sequences led to a significant increase in the stable conductivity of the BioFETs, while the exposure to non-complementary and mismatched sequences resulted in minimal changes in conductivity following stabilization [96]. This differential signal response effectively enables the distinction among these three types of sequences and may facilitate the identification of pathogen variants during PD infection.
Though ssDNA probe has a high affinity with the cDNA, its large size usually increases the distance between the sensing surface and the target–probe reaction [97], which may lower the sensitivity due to the charge screening effect. Shorter probes, such as peptide nucleic acid (PNA), are introduced to replace ssDNA while it offers higher sensitivity. They are also named as peptide-based BioFETs to help the probe and target remaining within the Debye screening length [70], which should be approximately 0.7 nm to 2.2 nm in typical physiological samples [98]. The sensing performance of BioFETs modified with PNA and DNA probes, respectively, has been investigated to verify the enhanced sensitivity brought by smaller probe size. Additionally, the GFET modified with PNA probes demonstrated lower background noise and relatively higher selectivity than the latter. Its dynamic range and LOD were three orders of magnitude higher than those of the DNA-based GFET [99]. To integrate the benefits of this technology and lab-on-chip to develop PoC devices, it was further combined with the commercially manufactured printed circuit board (PCB) techniques. PCB electrodes can be directly used as the source, drain, and gate of FET to exempt the extra deposition process. Then, graphene ink can be drop-casted to form the FET channel, and PNA probes can then be fixed on it to recognize the target cDNA. This device has been applied for the label-free DNA detection of cell-free nucleic acid biomarkers and exhibited nearly twice the sensitivity compared with EGFET based on traditional silicon substrates [79].

2.2.3. Aptamer-Based BioFETs

Aptamers, artificially synthesized ssDNA or RNA, can be chemically modified in a sequence-specific manner to improve their binding affinity to targets [100]. Notably, they can combine certain toxic substances that antibodies cannot detect [101]. In comparation to DNA probes, aptamers offer several advantages, including lower cost, higher stability, reduced susceptibility to interference from charged particles, and a smaller size [78,102]. These characteristics allow aptamers to retain their binding affinity even in high-ion concentration matrices [103], such as PDE. Within BioFETs, when an aptamer binds to its target, it typically transitions from a less structured two-dimensional (2D) form into a more organized and complex three-dimensional (3D) structure. This conformational change brings the target closer to the channel surface, improving the system conductivity [78].
For PD-related applications, aptamer-based BioFETs can be utilized to detect various biomarkers to understand the patient inflammation status; for instance, exosome surface transmembrane proteins differentiation 9 (CD9) protein, which reflects the dialysis efficiency [104], or C-reactive protein (CRP), which may indicate the membrane characteristics [105]. In the case of CRP detection, aptamers were immobilized on multi-wall carbon nanotube (MWCNT) BioFETs for rapid sensing. The BioFETs were easily functionalized by soaking overnight in an aptamer, where the aptamers initially adopted a stable 2D structure [78]. Upon the introduction of CRP to the sensing area, the negatively charged CRP induced positive charges in the MWCNT, increasing the number of free electron carriers and leading to a drain current change. Despite interference from PD-associated biomarkers such as IL-6 and TNF-α, the device rapidly detected changes in CRP concentration within minutes. It achieved a LOD as low as 0.017 mg/L, with a detection range that covers the CRP levels in the human body [78].
However, for practical applications, aptamers often require further optimization to enhance their binding affinity and stability, as they can be affected by factors such as ionic strength and interference [106]. To address this, additional selection processes, such as the systematic evaluation by exponential enrichment technique (SELEX) [107], are typically employed. In studies targeting the nucleic acid sequence of the CD9 protein, SELEX identified three sequences with high reproducibility and low ΔG values (Gibbs free energy) as the most effective candidates. Among these, CD9-26, which had the lowest dissociation constant (Kd) and thus the highest affinity for the target, was selected as the optimal tool for capturing the target sequence. It was then covalently immobilized on MXene (Ti3C2Tx) via the EDC/NHS method. MXene, with its numerous functional groups including hydroxyl and carboxyl groups, played a crucial role in facilitating the attachment and enhancing the overall performance of the sensing platform. The BioFETs exhibited a LOD of 6.41 × 102 exosomes/mL and a detection range of 1 × 103 to 1 × 107 exosomes/mL in human serum, which significantly outperformed antibody-based methods [101]. Additionally, in comparation to traditional qRT-PCR [108], the proposed platform eliminates the need for exosome purification, separation, and gene extraction, offering a more economical and rapid alternative for clinical exosome detection.

2.2.4. Immuno-Based BioFETs

In immune BioFETs, antibodies are usually used as probes and immobilized on FET channels. Changes in drain leakage current after antigen–antibody binding will generate detectable electrical signals since they are both charged molecules [70]. For PD-related applications, immune BioFETs demonstrated potential in bacterial screening [109] and biomarkers detection [80] as well. A BioFET-based sensing platform was proposed for detecting exosome membrane protein CD63 by integrating CD63 antibodies. The interaction between the antibodies and negatively charged exosomes induced the accumulation of positive charges in the graphene channel, leading to a shift in the Fermi level [80]. The method bypasses traditional exosome extraction and analysis techniques such as PCR and Western blot, avoiding the associated complexity and time consumption [110]. In experiments conducted by Kwong Hong Tsang et al. [80], this device achieved a LOD as low as 0.1 μg/mL. Albumin, another important PD biomarker, is closely linked to PD peritonitis and patient mortality [111]. Based on the same antigen–antibody reaction principle, albumin antibodies were immobilized on BioFETs for albumin detection [112]. In this study, the thickness of the transducing material was precisely controlled by using spray gun technology to obtain a single-layer film, which improved system sensitivity. This approach increased the surface area of transducing material and enhanced covalent binding during the biorecognition process [112].
However, the size of antibody is generally larger than the Debye length, which results in reduced sensitivity [43], so many studies redesigned antibodies by retaining antibody-specific fragments. Single-chain variable fragment antibodies (scFv) have been used in the design of GFETs for the detection of pathogen antigens. They are obtained through the connection of the variable heavy chain and variable light chain domains of the parent IgG through short peptides, which have a smaller size and higher binding density. The result of binding experiments showed that the use of this probe increased the LOD by nearly 4000 times compared to traditional IgG GFETs [113]. Nanobodies have been incorporated into BioFETs as well. As single-domain antibodies that retain the specificity of the homologous antigen [114], they are usually less than 3 nm, much smaller than antibodies (~150 kDa, 15 nm), antibody fragments (Fab) (~50 kDa, 7–8 nm), and scFv (~27 kDa) (Figure 6), and have a more stable and simpler structure [43]. The LOD of nanobody-based BioFETs has been proved to be sub-picomolar, which is at least five orders of magnitude lower than the previous reports of BioFETs based on aptamer [115].
Through comparison and analysis (Table 1), it is evident that the sensitivity of detection can vary significantly depending on the binding principle employed. These variations are further influenced by the methods used to immobilize the probes on the BioFET surface and the choice of transducing materials, highlighting the complex interplay between these factors. This observation underscores the role of transducing materials in dictating the overall performance of BioFETs. The following section will delve into the impact of transducing materials, discussing their properties, functionalization strategies, and implications for enhancing sensitivity and selectivity.

2.3. Transducing Materials

The transducing material plays a key role in BioFET performance, as it directly influences sensitivity, selectivity, and stability, which is critical for optimizing probe immobilization, preventing nonspecific adsorption, and reducing noise [46]. It is expected to drive significant improvements in device performance if designed properly [88]. Given that PD is a long-term therapy, the monitoring devices must prioritize long-term accuracy, efficiency, stability, and reusability to ensure reliable, ongoing patient care.
Transducing materials can be broadly categorized based on their dimensional structure (Figure 7); 3D bulk materials, such as silicon (Si) and germanium (Ge), are commonly used [134], as well as nanomaterials, such as metal oxide semiconductors (MOS), including zinc oxide (ZnO) and indium gallium zinc oxide (IGZO) (Figure 8) [135]. One-dimensional (1D) nanostructures, such as silicon wires (SiNWs) [50] and carbon nanotubes (CNTs) [86], along with two-dimensional (2D) nanomaterials, such as graphene [79], black phosphorus [136], or transition metal dichalcogenides (TMDCs), such as MoS2 [137], are also frequently employed (Figure 9). Additionally, metal nanoparticles are often used to modify the surfaces of these materials to amplify detection signals [138].

2.3.1. Bulk Materials

Conventional FETs use 3D semiconductors such as Si/Ge as channel materials, typically with a thickness of 102–104 nm [143]. However, these materials are often seen as limiting in terms of signal conversion efficiency due to poor modulation of channels near the substrate [143]. While there have been efforts to enhance the sensitivity of BioFETs based on 3D materials by adjusting oxide thickness [134] or tuning the mole fraction in Si/Ge [144], they still present inherent limitations. Their bulkier structure results in weaker electrostatics and [145], although the increased surface area in 3D nanostructured electrodes can strengthen signal output, it also tends to introduce more noise [75,146]. In contrast, nanomaterials are far more appealing due to their high surface-to-volume ratio, superior electron transport properties, and sizes that are inherently compatible with biomolecules, etc. [147]. These attributes allow for the immobilization of a higher density of bioreceptors and provide more active binding sites, which can significantly improve the sensitivity and performance of BioFETs [138].

2.3.2. Metal-Oxide Semiconductors

Compared to other materials, the main advantages of MOS are its ease of fabrication and compatibility with flexible substrates, especially in the field of wearable devices [148]. MOS, such as In2O3, ZnO, and SnO2, have wide bandgap characteristics, and their electrical properties are less affected by sensing environmental factors [149], which makes them competitive as the candidates of transducing materials in PD monitoring devices. They have been widely incorporated into BioFETs to develop PoC devices [149]. ZnO, which has a large and direct band gap (3.37 eV), can maintain a low noise level even under high pressure up to 200 kPa [150,151]. However, to enhance the stability of the system, high-k dielectric layers are frequently used to modify the MOS surface. The same passivation strategy works in other kinds of MOS as well, and the spin-coated Al2O3 layer on the surface of In2O3 BioFETs improved its insufficient stability in high voltage and liquid. The results show that the addition of the Al2O3 dielectric layer not only suppressed the damage of the In2O3 layer by leakage current and ion diffusion, but also exhibited long-term continuous monitoring performance with low signal drift [152].
Although MOS-based BioFETs exhibit high electrical performance, issues with mechanical flexibility usually limit their development in the field of wearable devices [153]. To fix this, it can be integrated with soft substrate to enhance flexibility. Studies showed that fabricating an array of MOS nanoribbon BioFETs on a polyethylene terephthalate (PET) substrate can maintain stable performance and high electron mobility even over 100 bending cycles, while it has a wide serotonin detection range spanning eight orders of magnitude [148]. A flexible printed circuit board (FPCB), which is more accessible, has been applied as well. In2O3 BioFETs were passivated with Al2O3 and the SU-8 layer and then incorporated into a FPCB to achieve multiplexing and simultaneous measurement of H+, Na+, and K+ concentrations. The FPCB-BioFETs devices exhibited strong stability and maintain ion monitoring for more than 90 days [153].

2.3.3. One-Dimensional Nanomaterials

The 1D materials including SiNWs and CNTs offer a high surface area-to-volume ratio, with most of their atoms exposed to the surface. This feature brings them unprecedented sensitivity to the environmental changes [138], enabling detection at the single molecule level [154] or even down to single amino acid bases [155]. For instance, when CNTs are positioned near biomolecules, target molecules are attracted to the nanotube surface through covalent bonds or π–π interactions. These interactions cause the target molecules to fold or bind to the probes, altering the electrostatic potential (ESP) on the CNT surface [154]. The conductivity of these materials can be further enhanced through polymer coatings or improved deposition techniques. For example, loading polyethyleneimine (PEI) onto the CNT surface can create stronger binding [69], while optimizing deposition methods can improve the purity and yield of semiconductors [35], which in turn facilitates more efficient charge transfer. These strategies significantly enhance the performance of CNT-based BioFETs.
However, variations in the manufacturing of sensors can lead to inconsistencies in their electrical properties, making recalibration cumbersome and often unreliable, which hinders their commercialization and broader applicability [156]. One of the key advantages of 1D materials is their compatibility with the CMOS process, which allows for precise control over parameters such as length, width, density, and substrate doping level [126]. This compatibility enables large-scale manufacturing and integration [50], making 1D materials promising for future BioFETs applications. For instance, Zhao et al. [50] fabricated SiNWs BioFETs using a standard 8-inch CMOS processing platform, with nanowire dimensions of 45 nm in width and 10 μm in length. They used the APTES/glutaraldehyde method to immobilize CD63 antibodies on the surface of SiNWs for specific capture of exosomes by utilizing the properties of aldehyde and amino groups. They immobilized CD63 antibodies onto the SiNWs using the APTES/glutaraldehyde method, leveraging the aldehyde and amino group properties to specifically capture exosomes. The repeatability of the device was verified with a relative standard deviation (RSD) of 2.56% across three tests in exosome solution, indicating consistent device performance. Additionally, reusability was demonstrated by successfully separating captured exosomes using IgG elution buffer and reintroducing the exosome solution, with a threshold voltage change of less than 1.5. The standardized CMOS process not only minimized performance variation between devices, but also ensured good stability and reusability. These advantages make BioFETs more affordable, reliable, and suitable for long-term monitoring in PD patients.
The 1D materials show compatibility with smaller probes such as monoclonal antibody (Fab) fragments as well. These probes reduce the overall thickness of BioFET functionalization while it retains the binding sites and enhances the selectivity compared to antibodies, which makes the FET more resilient to Debye screening in high-ionic solution [126]. Studies have shown that Fab-based SiNWs BioFETs can achieve highly sensitive detection of CRP [126]. To further improve sensitivity, the surface of nanowires is optimized to have smooth edges, which increases mechanical stability and reduces electron scattering [157]. Techniques such as wet etching based on tetramethylammonium hydroxide (TMAH) and isopropyl alcohol (IPA) are employed to maintain the mask pattern and smooth edge defects [158]. However, research also demonstrated that increasing the surface roughness of SiNWs, by adjusting etching molecule concentration, can boost sensitivity. A rougher surface increases the surface-to-volume ratio, which enhances the system’s ability to capture target molecules more effectively, ultimately improving the BioFETs’ performance [159].

2.3.4. Two-Dimensional Nanomaterials

Similar to 1D materials, 2D materials such as graphene and TDMC offer a high surface area and exceptional electron transport properties, while also being compatible with existing manufacturing processes. However, the cost of large-scale production for 1D nanomaterials is relatively high compared to 2D materials [57]. BioFETs based on 2D nanomaterials benefit from better control over their geometric properties, including minimizing impurities and crystal defects [160], as well as fine-tuning the band gap [147]. These advantages allow for more precise sensor performance and adaptability in various applications. Additionally, a 2D-based material channel can cause significant changes in Fermi level with its higher quantum capacitance [70]. This is important when detecting biomarkers in physiological fluids such as saliva, where concentrations are typically lower than in serum [161]. Achieving high sensitivity and LOD is crucial in these scenarios. As a result, BioFETs based on 2D materials are increasingly being widely utilized for detecting PD-related biomarkers and pathogens, offering improved performance in low-concentration environments [162].
Graphene, a single-atom-thick, tightly bound sp2-hybridized carbon layer [147], was the first discovered and utilized 2D nanomaterial [163]. Its unique electronic property—where the valence and conduction bands meet at a single point—makes graphene highly sensitive to external stimuli, a feature that is key to its use in biosensors. This sensitivity stems from its “fragile” band gap, allowing graphene to detect minute changes in its environment [147]. Furthermore, it remains the only 2D material with potential for large-scale commercial applications [163], largely due to its compatibility with chemical vapor deposition (CVD) technology [65]. This technique ensures the high-quality, cost-effective production of graphene with controllable size and thickness, making it suitable for mass production [164]. Monolayer graphene, in particular, preserves optimal carrier mobility, which is critical for the performance of graphene-based BioFETs [165]. The quality of graphene can be validated using Raman spectroscopy, where a 2.5:1 ratio of the G-band to the 2D-band and the absence of the D-band confirm its monolayer purity [166]. Compared to traditional silicon-based BioFETs, the operating voltage (VDS) required for these carbon-based materials is extremely low, reducing the risk of denaturing or degrading sensitive biological probes due to high voltage [138]. Because graphene’s almost nonexistent band gap and exceptional conductivity allow carriers to be easily generated and moved under an electric field [31]. This low VDS is beneficial for the development of portable and wearable BioFETs in the PD applications, as it extends battery life and makes these devices more user-friendly in real-world settings.
In addition to graphene, its derivatives—graphene oxide (GO) and reduced graphene oxide (rGO)—offer distinct advantages due to their high carrier concentration, light transmittance, chemical inertness, and biocompatibility [167]. The introduction of oxygen-containing functional groups in GO and rGO enhances their ability to form conductive films and facilitates molecular immobilization, which is crucial for sensor design [168]. Additionally, it can be used as a quencher for the fluorescent group to develop multiple sensing mechanism BioFETs. The oxygen-rich groups and excellent light transmittance of GO quench fluorescence from labeled aptamers, generating a measurable signal [169]. Compared to BioFETs employing a single sensing approach, dual-sensing BioFETs that integrate both electrical and optical detection methods are thought to offer improved reliability and detection accuracy [169]. However, while pristine graphene layers exhibit no D band, indicating a defect-free structure as shown by Kong, et al. [166], the presence of a D band in Zhang, et al. [169] suggested surface defects introduced during the transfer process. Interestingly, Kwon, et al. [170] demonstrated that intentional introduction of edge defects in graphene can enhance its sensing performance. These defects create stronger chemical adsorption sites, improving charge transfer at the probe–graphene interface, thereby boosting the sensitivity of the BioFETs.
While graphene’s zero bandgap provides high sensitivity, it also results in higher leakage currents, reducing the sensor’s dynamic range [160]. This issue arises because electrons can easily pass through the barrier in graphene, leading to increased leakage current and degraded sensor performance due to a higher subthreshold swing. In contrast, 2D materials such as MoS2, which have appropriate bandgaps, offer better alternatives for improving BioFET performance [147]. MoS2 is the most widely studied TMDC, which has a simpler structure than 1D materials [171]. Its thin atomic layer structures can reduce the short channel effect of BioFETs while maintaining high carrier mobility to provide excellent electrostatic effects [65]. Meanwhile, the planar architecture and exposed overhanging surfaces of MoS2 simplify its patterning as a semiconductor channel for BioFETs [147]. This makes it more suitable for high-precision applications, especially with nanoporous MoS2, which shows sensitivity up to 10⁹ times higher than pristine MoS2 in BioFETs sensing [172]. The bioreceptors can be coupled on the edge of MoS2 nanopores for specific signal generation. The edges of MoS2 nanopores can be functionalized with bioreceptors, which generate specific signals, and this edge-functionalization approach has been shown to induce larger conductivity changes in the device [173].
However, similar to graphene, MoS2 is susceptible to surface degradation, which can compromise the electrical reliability of BioFETs. To address this, MoS2 is often passivated with grafted layers such as Al2O3, TiO2, or gold nanoparticles. For example, depositing an ultrathin Al2O3 dielectric layer on nanoporous MoS2 helps prevent surface degradation when exposed to air or liquids, while also providing dangling bonds for the chemical adsorption of biomolecules [174]. However, the conflict is that this passivation increases the physical distance between the FET channel and the charged biomolecules, potentially reducing sensitivity. Therefore, careful control of the thickness of the passivation layer is critical to maintaining the balance between protection and sensitivity [172].
In sum, each material offers distinct advantages for BioFETs in detecting PD biomarkers, making it essential to consider the specific requirements of the application. Factors such as the electrical and mechanical stability of the measurement environment (i.e., PDE), sensitivity to light and temperature, material availability for scalable production (especially for PD PoC devices), and compatibility with flexible substrates [43] are crucial in selecting the optimal material. For example, 2D materials can provide higher sensitivity in some cases due to their larger surface area, improved electronic uniformity, and stability, making them well-suited for integration into flexible and wearable electronics, while 1D materials may excel in applications requiring high specificity and selectivity. Examples and omparison of these different transducing materials are summarized in Table 2.

2.4. Configuration

The configuration of BioFETs plays a crucial role in influencing the sensor’s performance, even though the underlying principle remains the same across different types [182]. For example, in graphene-based FETs (GFETs), detection of bioreceptor–analyte interactions depends heavily on the electrostatic gating effects of graphene [163]. Based on how gate voltage is applied, BioFETs can be categorized into several configurations, including back-gated, dual-gate, electrolyte-gated (EGFETs), and extended-gate designs (Figure 10).
Each configuration offers distinct advantages and trade-offs for biosensing applications: Back-gated BioFETs typically involve a gate placed beneath the substrate, allowing basic electrostatic control, but may limit sensitivity due to the separation between the gate and the channel [183]. An example is the use of capacitive sensing applications, using Al as the back gate electrode and AlOx formed by natural oxidation of Al as the insulating layer. The capacitance of the device proved to be well modulated by the gate voltage under vacuum conditions due to the quantum capacitance effect [184]. Dual-gate BioFETs, on the other hand, provide better sensitivity and control by using two gates to modulate the channel’s charge carrier concentration more effectively. For instance, Meng et al. [142] demonstrated a dual-gate single-molecule GFET with HfO2/Al2O3 as the dielectric layer, with nanocracked graphene electrodes and a single dinuclear ruthenium diene (Ru-DAE) complex as the channel, achieving a maximum on/off ratio of greater than three orders of magnitude. Extended-gate configurations decouple the sensing area from the active region of BioFETs, which helps reduce interference from direct exposure to the electrolyte, making it suitable for long-term and more durable sensor applications [101]. The limitation of EnFETs in combining several different active layers in one device also indicates the importance of extended-gate configuration [185]. For example, extending the gate through an externally connected platinum electrode functionalized with a graphene sheet decorated with bioreceptors can overcome the Debye screening limitation [186].
Compared to these gate configurations, electrolyte gate is the most common structure of GFETs [187]. Studies showed that electrolyte-gated configurations can be two orders of magnitude more efficient than traditional back gates [79] because the formation of a thin electrical double layer (EDL) between the electrolyte and the graphene interface improving device conductivity while ensuring low leakage current, and also the charge density in these devices depends more on the thickness of the EDL rather than the distance between the gate and channel [188]. In electrolyte gate GFETs, the sample solution is directly applied as the electrolyte or proxy of physical fluid and bioreceptors are fixed on the graphene surface to capture the targets [189]. However, such devices often need an additional passivation layer to avoid contact between the source and drain electrodes through the electrolyte solution to ensure the stable contact with the analyte [167]. Overall, each of these configurations affects the overall sensitivity, stability, and selectivity of the BioFETs, and the choice of configuration depends on the specific application requirements.

3. Functions of BioFETs Applicable to Peritoneal Dialysis Monitoring

Continuous monitoring of key substances such as urea, creatinine, and electrolytes (H+, K+, and Na+) is essential for assessing dialysis status (Table 1) [190]. They can serve as critical risk assessment tools for identifying patients at the highest risk of developing PD-related complications, facilitating personalized treatment interventions [191]. For instance, biomarkers that indicate peritoneal inflammation can help identify patients with progressive loss of peritoneal function, while those reflecting weakened immunity can identify PD patients with increased susceptibility. Shifts in cytokine patterns may provide insights into the nature of an infection [191]. However, most of the current methods of monitoring PD patients are limited to delivering doses and measuring the transport status of the peritoneal membranes, and the medical requirements for biomarkers are still unmet [10,191]. The integration of BioFETs into PD monitoring systems has the potential to enable sensitive, real-time and PoC analysis of PDE composition, helping to identify complications including peritonitis, membrane failure, or fibrosis [192] at an early stage. In this section, we explore the potential application scenarios of BioFETs applicable to PD monitoring by dividing their functions into routine monitoring and biomarker detection for early diagnosis and intervention, highlighting their significance in improving BioFET management.

3.1. Routine Monitoring of Key Indicators

3.1.1. Dynamic Monitoring of Glucose Imbalance

In PD, the glucose concentration in the dialysate is typically 10 to 50 times higher than physiological serum levels, creating an osmotic gradient essential for effective fluid removal [193]. However, elevated glucose levels contribute to complications such as hypertriglyceridemia, hyperglycemia [15], structural and functional changes in the peritoneum [20], and, eventually, fibrosis, either directly or indirectly, through the formation and adsorption of glucose degradation products (GDPs) and advanced glycation end-products (AGEs) [194]. Although various alternative therapies, such as stem cells [195] and peptides or glucose polymers [196], and macromolecular solutions such as L-carnitine and xylitol [193], have been proposed, none gained widespread acceptance. Consequently, glucose remains the predominant osmotic agent due to its effectiveness, low cost, and favorable safety profile [58]. So, the latest Kidney Disease Improving Global Outcomes (KDIGO) guidelines recommend continuous glucose monitoring (CGM) for ESRD patients [197]. However, recent CGM studies focus on hemodialysis, with relatively little attention to blood glucose self-monitoring (SMBG) [197]. It is, therefore, important to offer PD patients accessible methods for self-monitoring glucose levels to enable personalized treatment adjustment, especially since absorption varies depending on individual peritoneal characteristics and transporter status [198].
BioFETs-based glucose monitoring operates through an enzymatic reaction between glucose oxidase (GOD) and glucose [167]. In this process, glucose is catalyzed by GOD to produce gluconate and H2O2, where the H2O2 goes on to be dissociated and produces H+, which can be detected by the BioFETs [199]. More specifically, H2O2 undergoes electrocatalytic oxidation at a microelectrode to generate H3O+, and the subsequent diffusion-mediated transport of hydronium ions creates localized pH gradients detectable by BioFETs [200]. However, as mentioned above, enzyme-based BioFETs are intolerant to challenges such as long-term use and storage. The immobilization efficiency of enzymes on BioFETs can be compromised by factors such as ambient dissolved oxygen levels, temperature, and humidity fluctuations [68], resulting in lower dissociation constants for the gluconic acid and a limited dynamic range in practical applications [200].
To eradicate the limitation of enzyme-based BioFETs, research shifted toward developing non-enzymatic BioFETs. One approach involves depositing CuO NPs in the channel of an EGFET using a low-cost inkjet printing technique. The catalytic properties of CuO NPs influence the glucose oxidation process, with the oxidation peak current increasing as glucose concentration rises, enabling quantitative glucose analysis [85]. However, these BioFETs suffer from poor selectivity toward different sugars. Recent studies demonstrated that transition metal oxide-based BioFETs offered improved electrocatalytic activity and significantly enhanced glucose oxidation, particularly in compound in the form of MWO4 (M could be Ni, Cu, etc.). Shaping these materials into porous microspheres further increased their surface-to-volume ratio, enhancing sensitivity. Notably, NiWO4-based BioFETs retained good selectivity for glucose in the presence of interfering substances such as ascorbic acid, uric acid, lactose, fructose, and dopamine [68]. These advancements present promising potential for non-invasive glucose detection in next-generation biosensors.
Efforts to introduce more stable glucose-sensing components and improved immobilization strategies have also been explored to enhance FET-based glucose detection. For instance, modifying the rGO surface with AuNPs can significantly boost the electro-oxidation of glucose, resulting in a more stable structure with a wider linear detection range compared to EnFETs. Notably, this configuration achieves a LOD for glucose that is nearly ten times lower than traditional electrochemical sensors [201]. Optimizing the GOD immobilization further contributes to better FET performance. Compounds such as glutaraldehyde and Nafion have been employed to strengthen the adhesion of GOD to the channel. Additionally, incorporating a TiO2 film into the system enhances sensitivity at extremely low glucose concentrations [202]. TiO2’s photoelectrochemical properties make it particularly promising for glucose detection, with ongoing research aimed at increasing its light absorption efficiency. This is achieved through heterojunctions with other carbon or gold nanomaterials, as they need to be sensitive enough to detect trace glucose levels in, for example, saliva and sweat [203]. This heterojunction method not only increases the signal strength, but also improves the stability of the sensor, which may facilitate glucose detection in other human fluids for PD patients.
However, the biocompatibility of BioFETs material needs to be considered to hinder adverse clinical reactions [204]. For example, polymeric ion-sensitive membranes (ISMs), which are often conjugated into BioFETs, may involve plasticizers with possible cytotoxicity [75]. To address these concerns, more biocompatible materials, such as vinylphenylphenylboronic acid (VPBA), have been explored as glucose-responsive monomers. Combined with 2-hydroxyethyl methacrylate [160], these can form a functionalized hydrogel coating on the BioFETs electrodes for glucose monitoring [205]. The hydrogel-based BioFETs offer high sensitivity and biocompatibility, making them suitable for both in vivo glucose monitoring and in vitro applications, including wearable devices such as glucose-sensing contact lenses [205]. The development of these devices aligns with the needs of PD for daily glucose monitoring, particularly as they can be adapted into PoC systems equipped with wireless capabilities, enabling non-invasive and wearable biomarker detection for both patents and clinicians [75].
In a similar vein, hydrogels can encapsulate other glucose-sensitive parts, such as boronic acid-containing peptides, in BioFETs. These peptides can locally self-assemble into a hydrogel nanofiber network through non-covalent interactions, where phenylpropionic acid interacts reversibly with glucose, producing borate ions that influence the conductivity of the semiconductor [59]. These hydrogel-gated BioFETs demonstrate potential for wearable applications, allowing real-time monitoring of glucose in complex environments. It also proves that enclosing interfacial hydrogels on BioFETs may be a general strategy to overcome the stability issues of wearable BioFETs. Furthermore, enhancing the reusability of non-enzymatic BioFETs is possible by using sensing parts such as N-dimethylaminopropyl acrylamide, which undergoes reversible reactions with the glucose. Studies have shown that such modified BioFETs exhibited a glucose LOD of 1.9 μM in human urine, and they can regain the detection ability after hydrochloric acid treatment [178].

3.1.2. Urea and Creatinine Clearance

Routine monitoring of urea clearance (Kt/V) and creatinine clearance (CCr) is a clinical standard for assessing residual renal function [206,207]. Both urea and creatinine are filtered by the glomerulus, but urea will be reabsorbed while creatinine will be excreted. Therefore, evaluating urea or creatinine clearance alone may misjudge the residual renal function [44]. This makes the urea-to-creatinine ratio (UCR) particularly valuable in clinical practice, as an elevated UCR typically signals declining renal function [208], while their concentrations in PDE are commonly used to assess the clearance of toxic molecules during dialysis [209]. According to the International Society for Peritoneal Dialysis (ISPD), current dialysis prescriptions are adjusted based on the proportion of urea removed from the body [210]. Urea concentration during PD should be calculated through the weekly total kt/V [211]. The dimensionless parameter Kt/V represents the dialysis level, where K is the urea clearance, t is the duration of the treatment and V is the body urea distribution volume. It is usually measured before and after treatment, or indirectly by the conductivity of the dialysate [55].
Spectrophotometry has long been a standard laboratory technique for urea detection [212]. However, its high cost and infrastructure requirements make it unsuitable for real-time monitoring. Moreover, to obtain an accurate Kt/V, samples ideally need to be taken 30 min after treatment to account for the urea rebound effect, which cannot be captured without continuous monitoring, leading to potential inaccuracies [213]. While urease-based urea sensors were once popular due to the enzyme’s low cost, availability, and stability [214], they faced significant drawbacks. These sensors rely on the hydrolysis of urea by urease, producing hydroxyl ions after being catalyzed by urease, leading to a change in pH so that the urea concentration can be obtained by measuring the change in hydroxyl ion concentration [215]. However, challenges such as low selectivity, frequent calibration, and maintenance issues hindered their widespread adoption [213]. To avoid unnecessary delays during the treatment, urea concentration in PDE should be better measured under flow conditions [216] which poses high pressure on the sensitivity of sensors.
BioFETs, by contrast, offer significant potential for real-time urea detection in PD systems. EnFETs can detect pH changes or dissociated NH4+ and HCO3− generated by urease-catalyzed reactions, converting these into electronic signals proportional to urea concentration [217]. Early designs of BioFETs for urea detection utilized pH-sensitive ion-selective BioFETs (ISFETs), where urease was immobilized on a hydrated silicon nitride surface using glutaraldehyde [218]. However, the small sensing area of traditional BioFETs limited the enzyme density, resulting in weaker output signals [139]. To allow for a larger sensing area, EGFETs were developed, where the gate electrode is positioned away from the BioFETs, preventing direct contact with the target analyte [101]. This design not only expanded the sensing area, but also improved immobilization efficiency and sensitivity. In the case of urea detection, urease was immobilized on membranes and applied to the extended gates of ISFETs to create PoC devices, enhancing sensitivity by up to 50 times compared to earlier urea detection systems [219]. The setup can be cost-effective by directly fixing urease on commercially available fluorine-doped thin oxides (SnO2:F), yielding stable results within one minute using these simple assembly BioFETs [214]. Another approach involves embedding urease in magnetic alginate microcubes, which are immobilized on the EGFET surface via an external magnet. This straightforward packaging technique is also applicable to other substances, such as antibodies. By integrating different functionalized magnetic beads into a microfluidic chip, multiple biomarkers, such as urea, glucose, and proteins, can be monitored simultaneously in real time [215].
However, EGFETs have limitations as mentioned above, such as the inherent interface between the gate and the membrane, which introduces additional parasitic capacitance and resistance, negatively impacting conductivity and reproducibility [139]. To maintain the sensitivity and usability of BioFETs in real human fluids, urease was directly fixed on the Ag gate of the EGFET. This modification allowed for higher receptor densities, significantly enhancing the output signal. The device exhibits very low extraction power consumption compared to other conventional BioFETs, making it more applicable for practical applications [139].
Creatinine is also one of the most widely used biomarkers for assessing kidney function, though it should be accompanied with other biomarkers to optimize diagnosis or indicate the kidney status [220]. The conventional method for detecting creatinine in clinical settings is the Jaffé method [221], but this approach lacks specificity and a timely response, especially in the presence of interfering substances found in biological fluids [222]. Low accuracy, toxicity, and non-specific adhesion of ammonium ions used in this system limited this method as well [223]. To address these limitations, modern electrochemical biosensors have been developed to offer a more sensitive and specific interface for creatinine detection, utilizing biological receptors such as enzymes or novel synthetic responsive materials [224]. However, enzymatic systems are the most reported electrochemical creatinine sensors, prized for their high selectivity [225], and enzyme layers have been thoroughly investigated and applied in BioFET development [226].
EnFETs have been employed as potentiometric methods for creatinine detection, with creatinine deiminase [227] immobilized on the carboxyl-functionalized multi-walled carbon nanotube (COOH-MWCNT) films through crosslinking techniques. This setup, achieved through a chemical solution method, does not require complex instruments [228]. The strong bonding between COOH-MWCNT and CD over a large area improves the conductivity of the biosensitive membrane. However, while the crosslinking technique enhances stability, it is thought to reduce sensitivity and analytical range [224]. Moreover, despite improved assay efficiency [223], these enzymes still suffer from stability issues and are costly to apply in practice, as previously mentioned [229].
To improve the reliability of EnFETs for creatinine detection, special consideration should be given to issues related to enzyme handling, storage, or aging. Researchers explored various strategies to address these shortcomings, including immobilizing enzymes with nanoparticles to increase surface area and facilitate higher charge transfer [225], as well as improving CD adsorption [230]. One approach involved immobilizing CD on silicate particles, which were then coated onto the surface of a pH-sensitive BioFET for creatinine detection in physiological solutions. This design demonstrated high signal reproducibility and stability, with the device functions remaining for over a year in storage. Compared with traditional enzyme covalent crosslinking through glutaraldehyde vapor, this technique provided two- to three-fold increases in creatinine sensitivity, three- to four-fold reductions in response and recovery times, and significantly lower assay thresholds [231]. Adsorption of CD onto zeolites has also proven to be a non-toxic and stable method to retain enzyme activity. The effect of different modified zeolites on the BioFETs-based detection response has been investigated, with the zeolite modified with AuNPs (BEA-Gold) showing the highest sensitivity for creatinine detection. The incorporation of gold nanoparticles not only helped prevent enzyme denaturation, but also increased the surface area [230]. This aligns with early studies, since the nitrogen on the aromatic ring of creatinine exhibits a strong affinity for metal ions, making metal nanomaterials especially effective for creatinine detection [225]. In fact, several reviews discussed creatinine detection methods using different metal-centered nanomaterials [232], further emphasizing their suitability for this application.
Despite these improvements, enzymes remain costly and susceptible to denaturation over time [225]. Although studies on non-enzymatic BioFETs are less extensive than their enzymatic counterparts, these devices can be tailored to meet specific sensing requirements and are simpler to fabricate [232]. One of the most studied non-enzymatic approaches involves the use of molecularly imprinted polymers (MIPs), which are often used to replace the gate ends in ion-sensitive BioFETs, creating a sensor that responds specifically to creatinine-sensitive ions [223]. MIPs are artificially synthesized receptors that mimic the mechanism of antibody–antigen formation with high affinity interactions [232]. In this method, templates and cavities are created in a high-affinity polymer matrix [223], and MIPs are imprinted onto elongated BioFET electrodes, reporting creatinine concentrations through integration with a digital readout circuit ring oscillator [233]. To improve the sensitivity of these systems, optimizing MIP-related parameters is crucial for improving the conductivity of BioFETs, a topic that will be explored in further sections.

3.2. Detection of Potential Biomarkers

In addition to daily monitoring functions for glucose, creatinine, and urea, BioFETs have shown significant potential for detecting a wide range of biomarkers. These include PD-relevant biomarkers such as cystatin C (CysC), beta-2 microglobulin, albumin [234], cell-free nucleic acids, microRNA [235], inflammation markers such as interleukin-6, tumor necrosis factor-alpha (TNF-α) [236], and exosomes [237], among others (Table 3). PDE is a rich source of these potential biomarkers [238], containing solutes diffusing from the circulation along with peptides and proteins released locally from the peritoneal tissues. These components can provide insights into mesothelial cell mass, peritoneal fibrosis, and local inflammation, to name a few, during PD [239]. Additionally, other biological fluids, such as saliva, tears, sweat, serum, plasma, and urine, have also been explored as sources for PD biomarkers [240,241,242]. However, despite promising findings, the integration of new PDE biomarkers into routine PD monitoring remains limited while there is no proof that these biomarkers directly contribute to the PD outcomes [58]. The only two identified biomarkers so far are CA125 and IL-6, both of which can be readily measured in unconcentrated PDE [58,237].

3.2.1. IL-6

IL-6 is the most thoroughly studied marker of inflammation in PD patients, and its systemic levels of IL-6 and its soluble receptor are always elevated in ESRD patients while inducing hepatic acute phase protein synthesis [191,239]. Additionally, it is a key factor in increasing solute transport associated with inflammation, which predisposes the peritoneum to fibrosis [254]. BioFETs-based IL-6 detection primarily leverages the highly specific interaction between IL-6 and its antibody, IL-6R [255]. When IL-6R is immobilized on transducing materials such as SWCNTs, it captures IL-6, causing a measurable drain current change in the BioFETs due to the antigen–antibody interaction. Simultaneously, inter-tube contact between the SWCNTs and the FET reduces resistance, further improving the detection sensitivity [255].
Efforts have been made to transform this system into a wearable PoC platform. For instance, IL-6 antibody was immobilized on the rGO layer and transferred to a flexible polycarbonate substrate to develop BioFETs, facilitating the ultra-sensitive IL-6 detection even in tears [256]. The IL-6 detection has been expanded to saliva as a non-invasive and accessible pathway as well. Aptamer-functionalized GFETs were integrated onto PCBs to form a portable PoC device for real-time monitoring of IL-6 in saliva. This system can respond to the changes in IL-6 concentration within several minutes and display results via an online app, making it a promising tool for the self-monitoring of PD patient users [187].
To increase the detection range of IL-6, a novel metal carbide nanocomposite MXene film with an accordion multilayer structure was applied to amplify the electrical signals. An IL-6 aptamer was also decorated with thiolate group at its 3’ end to improve the efficacy. Due to the combination of the optimized BioFET structure with the multi-helix structure, the BioFETs demonstrated an IL-6 detection range nearly two orders of magnitude higher than the previous one with better selectivity, reproducibility, and stability [257]. Despite the advancements in detecting IL-6 using BioFETs, its role as a definitive marker for peritoneal inflammation remains debated. IL-6 has been shown to downregulate other inflammatory molecules by activating IL-1 and TNF receptor antagonists, suggesting it also possesses anti-inflammatory properties. This dual functionality complicates its use as a straightforward indicator of peritoneal inflammation, as IL-6 may contribute to preserving the peritoneum rather than solely indicating inflammation [258].

3.2.2. CA125

CA125, another identified substance in the PDE regular monitoring, is produced by cells lining the peritoneum and so is often used to reflect the quality of mesothelial cells of PD patients as an alternative parameter for determining the status of the peritoneum [259]. A sudden drop in CA125 levels may signal severe mesothelial cell damage [239] or a heightened risk of encapsulating peritoneal sclerosis [133] development [260]. However, elevated CA125 levels in the PDE may also occur during PD-related peritonitis [259], making CA125 monitoring essential for accurately diagnosing and evaluating the overall status of PD patients.
Relatively few studies explored CA125 detection using BioFETs compared to IL-6. A previous study immobilized CA125 ssDNA [261] on the surface of carboxylated MWCNT-BioFETs via amide bonding and then stacked it on the graphene surface of rGO-based BioFETs via π–π interactions. Through specific interactions with CA125 antigens, the BioFETs were capable of detecting CA125 with high specificity and a LOD of 5.0 × 10−10 U/mL, possessing the potential for application in biological fluids and clinical samples. To ensure BioFETs have high electrical performance and degradation stability in physical solutions for practical CA125 application, they were integrated with microfluidic channels using the liquid channel as a passivation layer, which effectively improves the electrical stability of BioFETs [262].
However, the routine measurement of peritoneal CA125 is not widely recommended [263] due to ongoing debates about its reliability as a biomarker, similar to IL-6. Additionally, CA125 levels are positively correlated with IL-6 [258]. Moreover, some CA125 in the PDE may be partially derived from systemic circulation [263], while the ideal PDE biomarker should be released or produced locally in the peritoneal cavity or be involved in the pathology of the peritoneum [239]. Furthermore, studies have shown no significant impact of peritoneal urea clearance or residual renal function on serum CA125 levels [264]. Consequently, CA125 is generally regarded as an auxiliary biomarker, often used in conjunction with other markers, such as Chemokine CCL2, to provide a more comprehensive understanding of the inflammatory process [265].

3.2.3. Other Potential Biomarkers and Pathogens

The controversies surrounding IL-6 and CA125 have driven the development of new biomarkers, including various cytokines and growth factors such as IL-8 and IL-17 [58], many of which are soluble, making them readily accessible in human fluids [237]. PoC diagnosis devices have been developed for TNF-α detection in tears, where the TNF-α aptamer was directly immobilized on GFETs. A nonionic surfactant was used to minimize nonspecific binding, subsequently enhancing the specificity [266]. Similarly, CNT-based BioFETs designed for CRP detection had been reported, where CRP antibodies were immobilized, enabling the detection of CRP levels from normal to inflammatory ranges, showing promise for clinical use [35]. To further improve the sensitivity, CRP Fab fragments were used in place of full antibodies, addressing the Debye screening effect and reducing the overall thickness of the capture layer. These Fab fragments were incorporated into SiNW array BioFETs, which provided a larger surface area and more binding sites to accommodate more analytes, thereby expanding the detectable concentration range [250]. However, it is important to note that the possibility of systemic production of these biomarkers cannot be entirely ruled out. For example, VEGF and TNF-α detected in the PDE may also originate from systemic circulation [258].
In addition to the biomarkers mentioned, germ infections are common during PD due to prolonged catheterization [15], yet relatively few studies directly applied BioFETs to detect pathogens in the context of PD diagnosis [56]. While some research summarized the use of various FETs for pathogen detection, these studies largely focused on applications related to food safety and environment monitoring [267]. In the context of PD, BioFETs-based detection mainly targeted bacterial biofilms [56], cell wall proteins [268], and genetic material such as DNA or RNA [94]. For instance, fibronectin, a protein that binds specifically to certain bacterial components, has been fixed on Au-EGFETs to adhere to the Staphylococcus epidermidis (S. epidermidis) biofilm, leading to changes in conductivity. In the same study, ssDNA probes were used to target the 16S rRNA of Staphylococcus Aureus (S. aureus) and S. epidermidis, resulting in pathogen-specific detection [56]. Another noble study employed Au nanoporous membranes in combination with an extended-gate FET to hybridize with S. aureus 16S rRNA, demonstrating high specificity even in the presence of non-specific DNA from other pathogens. The nanoporous gold membrane, by providing a large surface area on both its surface and walls to attach probes, enhanced the hybridization signal. This enables better control over DNA orientation and immobilization, making further boost detection accurate [269]. Additionally, innovative research utilized DNA origami techniques to design RNA-cleaving deoxyribonuclease (RCD) specific to S. aureus, capable of detecting concentrations ranging from 1 to 105 CFU/mL, with a remarkable lower LOD of 1 CFU/mL [270]. While the treatment approaches for Gram-positive (GPB) and Gram-negative bacteria (GNB) infections in PD are distinct [23], distinguishing bacterial types is crucial for effective initial antibiotic therapy. Recent advancements led to the development of BioFETs-based platforms designed specifically for this purpose. In these systems, antibiotics such as vancomycin, targeting GPB, and magainin I, targeting GNB, were immobilized on GFETs. These GFETs were subsequently integrated into a dual-channel microfluidic chip, enabling the simultaneous detection of multiple pathogens during PD treatment [268].
The ongoing development of these BioFETs is significant, as it enhances the capability to diagnose early infections in PD patients, thereby facilitating timely and appropriate treatment. Current examples and applications of using BioFETs to monitor these potential biomarkers related to PD are summarized in Table 4.

4. Challenges and Strategies for Implementation of BioFETs in PD Monitoring

4.1. Technical Barriers and Clinical Applicability

In practical applications, the complex composition of the PDE usually biases the detection and quantification, which poses a challenge for daily monitoring. The plasma-like structure of PDE, for instance, contains high-abundance proteins that interfere with proteomic analyses, masking signals from target biomarkers [238]. For other potential diagnostic samples, such as saliva, sweat, and tears, the concentration of target biomarkers is often considerably lower, posing additional hurdles for sensitive and accurate detection [281]. Moreover, to ensure reliable detection of analytes such as urea, measurement often requires ambulatory conditions [216]. These challenges are further exacerbated by the high-ionic concentrations in human bodily fluids, which induce Debye screening effects. This phenomenon involves an ionic cloud surrounding the target charge, diminishing its influence on the surface potential of the BioFETs gate. Consequently, the target charge can be significantly screened, leading to a substantial reduction in signal strength, potentially rendering the detection signal undetectable [71]. This interplay of complex fluid composition and stringent sensitivity and specificity demands underscores the significant technical barriers BioFETs face in PD monitoring applications.
In clinical applications, PDE samples could significantly heighten the risk of degradation for the transducing materials of BioFETs. These samples can vary widely in appearance and composition due to patient-specific conditions—for instance, red from peritoneal bleeding, milky due to triglyceride or chylomicron accumulation, yellow-green from bile leakage, or brown-black from hemolysis [282]. Such variations will make BioFETs channel prone to fouling, oxidation, and delamination over time [174]. Therefore, stability is paramount to ensure the reliable performance of BioFETs during prolonged monitoring in clinical settings, yet achieving this stability amidst the diverse and dynamic composition of clinical PDE samples remains a significant challenge.
Translating BioFETs from laboratory prototypes to commercially viable clinical tools poses significant scalability challenges. While current nanofabrication techniques are effective for small-scale production, they frequently fall short in providing the precision and reproducibility required for large-scale manufacturing processes. For example, most BioFETs built with 2D materials are only performed on a laboratory scale and not allowed for reproducible high-throughput manufacturing processes [57], because the non-uniform morphology of the substrate and the scattering effect can make the output of each device unreliable [147], but fortunately, advances such as standardized CMOS processes [50,126], compatibility with large-scale CVD [65], and lower-cost fabrication techniques including spray gun coating [112], inkjet printing [85], and reversible reaction designs [178] covered above, offer promising pathways to address these issues and promote commercial potential. However, integrating multiple probes into a single BioFET chip further increases manufacturing complexity, requiring highly specialized processes that are not yet widely available.
In addition to the consideration for BioFETs themselves, for BioFETs being adopted in clinical practice, they must integrate seamlessly into existing PD monitoring frameworks. This requires compatibility with standard clinical tools and protocols to ensure that BioFETs can be effectively used alongside current diagnostic and monitoring platforms. For instance, wearable BioFETs continuously monitoring inflammatory markers could trigger an alert if levels exceed a pre-defined threshold. This would prompt the clinicians to collect PDE samples for the further identification of peritonitis and initiation of targeted antibiotic therapy. This integrated approach could significantly reduce the time to diagnosis compared to relying solely on intermittent PDE analysis. Alternatively, a BioFETs-based patch could be connected to the catheter for the regular monitoring of solute clearance to personalize dialysis prescriptions, supplementing the information obtained from PET scans. By tracking trends in solute clearance over time, clinicians could identify patients who are experiencing a decline in peritoneal membrane function and adjust their dialysis regimen accordingly. Lastly, the data generated by these BioFETs can be transmitted via IoT devices (e.g., smartphones, tablets) to a secure cloud platform (Figure 2). This enables integration with existing patient records and facilitates clinical decision making. New algorithms or decision support tools may be required to effectively interpret the BioFETs data in conjunction with other clinical parameters, a topic that will be discussed in more detail subsequently. This integrated approach allows for a shift from traditional, clinician-centered monitoring towards a more patient-centered model, empowering individuals to actively participate in their PD management.

4.2. Solutions to Overcome Technical Barriers

4.2.1. Enhancing Sensitivity and Selectivity

Efforts to enhance the sensitivity and selectivity of BioFETs for practical applications face challenges due to the high ionic strength of physiological environments and the complexity of sample compositions, and in particular, efforts to overcome the latter impairment are still very limited [62]. In recent years, pairwise efforts focused on new BioFETs designed to address these challenges, including the introduction of new materials [49], development of controlled immobilization methods, and optimization of probe designs (Figure 11) [62]. Additional innovations in fabrication architectures aim to reduce internal electrical noise [283] and extend the Debye screening length in solutions [284], advancing the integration potential of BioFETs within PD monitoring systems.
Appropriate immobilization strategies are crucial for achieving the good performance of BioFETs. For example, in urea detection, the immobilization efficiency of urease would have a large effect on the sensitivity and response time of the urea biosensor by increasing the hydrolysis rate of urea [287]. The immobilization step was thoroughly investigated to improve its stability and reduce signal interference, including enzyme encapsulation in alginate or hydrogel beads, gate surface modification with gold or magnetic nanoparticles [87], incorporation of the enzyme into polyelectrolyte multilayers such as polyaniline or Nafion [232] or silicalite [69], and nano-spotting technique [288]. These methods do not usually aid the enzymes directly, but rather serve as an enzyme loading matrix [86]. Alginate microbeads carrying enzymes can be fixed to the BioFET surface by an external magnetic field so that the analyte sample can be injected into the microchannel and react with the enzymes contained in the alginate beads. The potential change in the BioFET surface caused by the release of H+ during the reaction can indicate the glucose, urea, and creatinine levels, respectively [289]. A major advantage of this encapsulation method is that a single sensor chip can detect multiple analytes simultaneously and the enzyme depletion can be countered by easily replacing the alginate microbeads with a new one [290]. However, a comparison with the results of the commercial assay kits revealed that the sensitivity of this kind of device was not substantially increased compared to the sensitivity previously obtained using the conventional bulk alginate enzyme encapsulation method [290]. The differences between the two sets of results for the glucose, urea and, creatinine metrics were 5.17%, 6.22%, and 13.53%, respectively. This result might be attributed to the partial depletion of the enzyme in the microbeads and absorption of the biological sample by the surface of the microchannels, resulting in a slight decrease in the intensity of the response during the measurement [289]. This could be avoided if the enzyme was properly stabilized by selecting a more suitable substrate and better immobilization methods [291]. Despite the excellent sensitivity, other researchers argued that the use of such biomaterials is problematic due to the absence of enzymes or antibodies for each target biomarker, as well as for some other reasons, such as poor long-term stability, costly and time-consuming production, and difficulty in controlling the quality of the production, and that therefore there is a need to look for other alternatives with high sensitivity [75].
Au alone is usually considered to have no specificity and selectivity when used as an electrode film coating [75], but AuNPs are widely used for doping modification of carbon materials due to their high catalytic efficiency, good biocompatibility and chemical stability [30]. In creatinine detection, creatinine can be physically adsorbed on the active sites of various metal centers to form metal–creatinine complexes, and then the current signal changes [232]. The same principle has been applied to the detection of CRP. AuNPs acted as a defect marker for the MoS2 layer of FET and non-covalently binding the defected MoS2 edges. The formation of strong S-Au bonds improves the charge transfer in FET while the immobilization of the CRP antibody enabled the detection of CRP [137]. Recently, AuNPs were employed for the detection of cystatin C in urine samples for the CKD patients, which is likewise a promising biomarker for early detection of renal injury [223]. In this system, AuNPs not only served as a carrier for papain, which can selectively bind to cystatin C to form a complex, but also improved the surface chemical inertness of laser-induced graphene, eliminating the need for an additional surface activation step [30]. Despite offering so many benefits, it is not suitable for large-scale manufacturing and practical application due to the high cost of raw materials [292], while clinically more cost-effective tools are preferred.
Artificial and functional membranes are quite promising, serving as platforms for small biomarker molecular recognition sites [72]. Molecularly imprinted polymers (MIPs) can not only overcome the above-mentioned biomaterial problems, but also enhance the binding constants (Ka) and LOD [75]. The response principle of MIPs to the target substance is that the chemical reaction between the template and the molecule generates charge, or the template molecule is embedded in the template, resulting in a weakened charge transfer to the electrode surface [275]. These polymer films can mimic the specific interaction between antibody and antigen, allowing the monomer to trap the template molecule during the polymerization process and to form three-dimensional imprinted sites for the specific recognition of the target molecule [292]. Due to these properties, they are widely designed into various molecular-specific templates while offering higher sensitivity. In the cases of urea detection, poly (methyl methacrylate) PMMA and urea were used as functional polymers and molecular templates, respectively, to fabricate the MIPs on the ion-sensitive BioFETs. The response voltage level is higher than that of conventional Si3N4-based BioFETs, which implies a great improvement in the sensitivity and selectivity for urea detection [275]. MIPs work in the detection of unamplified samples as well. The direct binding of the MIP to the GFETs resulted in the generation of additional electronic labels, allowing the device to observe characteristic bumps in the noise spectrum only in the presence of the target molecule. This property enables the MIP-based GFETs to have a serotonin LOD two orders of magnitude higher than the existing reports [132].
MIP-based BioFET performance can be analyzed and optimized through Langmuir adsorption isotherm equation. This formula has been used to evaluate parameters such as the binding affinity between small biomolecules and MIP cavities, and even to show the adsorption mechanism of small biomolecules on specific MIP templates [286], which might benefit the pathological studies of PD metabolism. Glucose detection based on MIP-BioFETs has been fully investigated before. MIP for glucose formed by a glucose template (GluMIP), phenylboronic acids (PBA), and glucose molecules was developed for the detection of low concentrations of glucose in an enzyme-free manner in tears [72]. As diol compounds, the glucose molecules used as templates can be removed under acidic pH conditions upon binding to PBA. Subsequently, BioFETs can detect the density changes of the negative charge generated by the PBA-diol compound [75]. The detected output voltage changes at different sugar concentrations can then be put into the Langmuir adsorption isotherm equation to calculate the LOD. Notably, MIP-based BioFETs can detect glucose about 15–20 times more sensitively than other types of sugars [286].
Based on previous studies and derivations [75], the Langmuir isotherm equation can be described as [286]
A = N [ c ] 1 + K a [ c ] .
A is the signal observed at equilibrium for the MIP-bound template, while [c] refers to the free concentration of the template at equilibrium, N is the number of active centers available per unit volume of MIP, and Ka is the binding constant. It can be easily seen that the potential difference at the interface of the electrolyte solution varies with the ion concentration, and in this experiment Ka is the binding constant for glucose and PBA in the MIP solution, and the value of Ka for MIP is an important parameter in controlling the LOD and imprinting factor (IF) [75].
K a = [ P B A S u g a r ] c h a r g e d [ P B A ] n o n c h a r g e d [ S u g a r ]
where [PBA-Sugar]charged indicates the concentration of charged PBA-sugar conjugate [72]. Based on this equation, the Ka of PBA with glucose was found to be several hundred times higher than values obtained through non-electrochemical detection methods [72]. This enhanced affinity is attributed to the role of MIPs, which function as hydrophilic interfaces with immobilized receptors. By positioning the target molecule closer to the surface, MIPs mitigate the ionic shielding effect, thereby reducing screening and facilitating more accurate measurements [62].
In addition to MIP, the combination of other polymer membranes and BioFETs frees it from the reliance on enzyme-specific detection and transcends ionic shielding length limitations, allowing for continued research as a non-invasive real-time monitoring method in PD systems. Bioreceptors can be directly blended with poly (styrene-co-methacrylic acid) (PSMA) to make a sensing membrane and then be spun-coated onto the extended gate of BioFETs. This method can reduce the ionic shielding length within 0.2 nm [293]. However, when it comes into direct contact with the human body, the design of such chemically synthesized interfacial materials should consider biocompatibility [75]. Meanwhile, several factors of MIPs or other polymer membranes still need to be considered when designing, including the cumbersome polymerization process, complete removal of the template, and deeply embedded template molecules leading to high background signals and narrower detection range, among other issues [292].
In addition to the optimization of immobilization, enhancing the sensitivity can be achieved through reducing Debye screening effect and noise signals caused by non-specific adsorption of interfering species in the solution of biological samples [75]. Most of the current protocols focus on shortening the distance between the BioFET surface and detected biomolecule analytes by either diluting the concentration of the analyte or using a small molecule as a bioreceptor [62,75,294]. However, dilution of the solution may affect the biomolecular conformation, stability or reduced affinity for capture-target molecule interactions [71,98]. More focus can be put on the development of a new interface material structures to realize the efficient signal conduction [45].
The solution to Debye screening can be translated into the filtration of non-specific electrical signals [102]. Theoretically, coating with hydrophilic polymers can inhibit non-specific adsorption of large interfering species such as gate proteins, while preventing small interfering species from approaching the gate surface can be achieved by filter membranes such as polymer nanofilter to improve the sensitivity and specificity of FETs [75]. The polymer nanofilter usually consists of an anchor layer and a filter layer in which the components we added can trap the interference substances. The anchor layer is in contact with the surface of the FET electrode, and target can pass through the filter layer to reach the anchor layer producing a detectable potential change. The polymer nanofilter applied effectively prevents the interaction between protein macromolecules and Au electrodes, thereby reducing non-specific electrical signals. This kind of nanofilter-based FET has been used for the potentiometric detection of small biomolecules [295]. It can further design to be an aptamer-based nanofilter interface, where the filter layer consists of DNA, RNA, or peptide aptamer, which can be used to capture dopamine (DA) molecules. The targets can reach the anchor layer to contact the electrode surface to generate an electrical signal [102]. Devices coated with nanofilter exhibit satisfactory filtration effect and suppress non-specific signals.
These methods enable BioFETs to conduct assays in solutions with high ionic strengths. However, for clinical applications, several factors must be carefully considered to optimize performance. These include the thickness of the polymer layer and the density of immobilized aptamer molecules, which significantly impact assay sensitivity. Additionally, controlling the initial potential of the Au-gated electrodes is crucial to prevent non-specific adsorption of interfering species that could oxidize at the electrodes, among other considerations. To obtain a uniform distribution of the electric field, the materials thickness should be controlled or narrowed down [296], since their oxide thickness, length, orientation and radius play important roles in improving BioFETs sensitivity [297]. In a previous study, Au nanoporous membranes with different sizes, thicknesses, and highly ordered and homogeneous structures were combined to the EGFETs for the hybridization tests for 16S rRNA of Staphylococcus aureus, respectively. The results reveal that the thicker Au membrane layer with smaller diameters promotes a larger hybridization signal [269].
These membranes are then developed to the nanoporous structure since the electrostatic shielding effect is weaker near concave surfaces and stronger near convex surfaces [298]. That is, molecular charges bound to convex portal surfaces (e.g., nanopillars) are more easily screened by counterions in the sample solution and will not induce any electrical signals in FET biosensors [75]. On the other hand, porous membranes (concave surfaces) that allow filtration of biomolecules are more of potential [75]. Bioreceptors embedded in the nanopores of EGFET were verified to enable higher molecular throughput and signal-to-noise ratio [299]. This kind of novel nanosensor combines the advantages of nanopore platforms and BioFETs through the use of nanopipette-based polypyridinium ion FETs [163], which allow real-time adjustment of the nanopore dimensions to fit the size of the target molecule, and even efficient control of molecular transport at the single-molecule level by controlling the gate voltage.
Conventional transducing materials have also been developed to various structures and scientists have been attempting to combine BioFETs with these new nanostructures to enhance the sensitivity of detection because of their larger surface area ratio [62]. For instance, in addition to the SiNWs, SEM analysis indicated that silicon nanoribbon has a width of 120 nm and a height of 25 nm which is larger than the functional surface area of conventional nanowires [296]. SiNW BioFETs can detect protein adsorption and desorption events at the single-molecule level while multiple Si-nanochannels can improve device performance by reducing biochemical noise and minimizing discrete dopant fluctuations [300]. A multiplexed SiNW tunnel FET can independently detect several distinct biomolecules with almost equivalent sensing current levels, regardless of the gold layer position, and have higher sensitivity than traditional BioFETs [301]. Even 120 SiNWs were fixed on the BioFETs for the parallel detection of ctDNA. Due to the high dimensional coherence of 120 SiNWs, they enable signal superposition compared to individual SiNWs, resulting in stronger response signals [302]. Vertically stacked nanosheet gate-all-around BioFETs were developed as well, which showed superior sensitivity compared with nanowire BioFETs or vertically stacked nanowire BioFETs, due to its multi-channel structure [303].
To sum up, methods to enhance the sensitivity and performance of BioFETs-based detection encompass a variety of strategies. These include optimizing probe immobilization techniques, such as enzyme encapsulation, attachment to gold nanoparticles or magnetic beads, and employing MIPs, which offer controlled recognition capabilities and improved detection limits for a wide range of substances, including small molecules. Additional approaches focus on mitigating the Debye screening effect using nanofiltration layers, polymer protective coatings, or passivation layers. Lastly, advancements in synthesis techniques and stringent quality control of transducing materials are crucial for ensuring consistent and high-performance FET-based sensors as well.

4.2.2. Stability and Durability Improvement

The long-term stability and durability of BioFETs are crucial for obtaining accurate and reliable detection results, given that PD is a long-cycle treatment modality. However, stability testing of biosensors in is usually performed prior to or separately from the target receptor-bound biosensing, and the neglect of temporal effects and signal drift, may mask actual biomarker detection and adversely affect device performance [59]. When BioFETs are immersed in physiological environmental solutions for prolonged periods of time, the surface of the SiO2 insulating layer may be hydrolyzed by cations, and reactive magazines such as airborne water and oxygen are more likely to be absorbed at the interface, affecting charge injection and transfer, thus affecting the final performance of the device [304]. One way to improve stability and reusability is to separate the sensing area from the electrode area, allowing connections to be made between the two parts, such as extended-gate BioFETs did [57]. Since the sensing area functionalized with bioreceptors is the only part of the device exposed to the environment, this configuration maintains excellent stability and reproducibility. Studies also explored integrating extended-gate BioFETs with electrical sensors to establish redundant validation systems, enhancing reliability in complex environments. The IFN-γ aptamer was immobilized on the extended gate of the BioFETs, which simultaneously serves as the sensing region of the BioFETs and the working electrode for electrochemical detection in the electrolyte solution. Upon encountering the target IFN-γ, its specific binding to the electrode induces changes in the redox peak current of ruthenium hexamine chloride ([Ru(NH3)6]CI3), triggering signal variations in both detection systems. The dual-mode sensor demonstrated higher sensitivity and a broader detection range, effectively mitigating BioFETs response saturation when exposed to high-concentration samples [305].
Apart from the configuration, there have been many cases showing that the precise control of nanoscale thickness through chemical deposition technology helps to stabilize the channel materials. Atomic layer deposition (ALD) is a novel technique for the evaporation of chemical substances, which has been widely utilized in BioFETs design. In current research, ALD was used to control the number of cycles to optimize the thickness and morphology of the generated MoS2-SWCNT film for the miRNA detection. This adjustment benefits the charge transfer between SWCNT and MoS2 and improves the on/off ratio and mobility of BioFETs. Raman spectra and thermogravimetric analysis proved that ALD enables the growth of MoS2 on SWCNT with almost no change and loss of the intrinsic properties of SWCNT, and even helps to remove the polymer PCz that wraps the CNT [306].
In addition to deposition technologies, device surface passivation is more common for enhancing stability and sensor lifetime. For example, even for continuous monitoring under conditions with numerous interferences such as in real biological fluids, BioFETs coated with rGO layer has been proved to be useful [307]. Deposition of an ultrathin layer of Al2O3 on the SiNW-BioFETs also enabled the device to remain stable in physiological solution for at least four months [308]. Combing the forementioned ALD, another research passivated the entire device with an insulating high-κ dielectric HfO2 which greatly improved the sensitivity and long-term stability of BioFETs. In this system, ALD allows thickness control on the nanometer scale and provides excellent uniformity and very low defect density [309]. However, the surface passivation strategy for BioFETs is not without side effects. A current study tried to passivate the metal source and drain of CNT-BioFETs with a patterned epoxy layer (SU-8) and then passivated the entire chip with a thin dielectric. They hoped that the charge-blocking effect created by this double passivation strategy would mitigate the device’s leakage currents and greatly improve the long-term stability of the device, only to find that as the device exhibited more stable detection performance, it usually came with a thicker layer of HfO2 (30 nm) and the cost of reduced device sensitivity [59]. Table 5 was proposed to summarize the different solutions mentioned above.

4.3. Strategies for Enhancing Clinical Usability

Enhancing the clinical usability of BioFETs is not merely about convenience; it is a critical factor in ensuring their seamless integration into existing PD monitoring workflows and maximizing their impact on patient care. This integration allows for the continuous monitoring of biomarkers and enables adjustments to the frequency and composition of dialysate exchanges based on real-time data, thereby optimizing fluid balance. This is particularly crucial for PD patients, as insufficient fluid removal can result in hypertension and symptoms such as subcutaneous and pulmonary edema, while excessive removal can lead to hypotension and diminished residual renal function [264]. The removal of excess solutes during PD is not selective and unintentional removal of useful substances may occur [10], such as the loss of vitamin D-binding proteins leading to symptoms of malnutrition [310]. Moreover, due to the diversity of factors such as age or diabetes status of different patients, personalized approaches should be explored to address the heterogeneity among different populations and meet the clinical requirements [8].
BioFETs could be miniaturized to provide wearable PoC devices, providing patients with immediate feedback. This is an effective way for managing their health and performing PD-related behavior [311]. These PoC systems demonstrated improvements in health management, as well as albumin and hemoglobin levels [312]. Notably, BioFETs can detect biological targets across a range of human biofluids [70], including not only PDE but also saliva, sweat and tears, making them less invasive compared to traditional plasma or serum tests [228]. No sample preparation are required during BioFETs-based monitoring which is also a unique attribute that adds to its commercial appeal [313]. To develop BioFETs-based devices for PD patients, several aspects must be considered: the multiplexed detection capability, data processing, and the requirements for PoC applications, including wearable design, biocompatibility during use (to minimize adverse immune responses), and device reusability.

4.3.1. Multiplexed Detection Capability

The composition of real clinical samples is so complex that the detection of a single analyte is far from sufficient for PD early diagnosis. Developing an efficient method for the simultaneous detection of multiple markers is particularly crucial, especially when these potential biomarkers must be analyzed collectively to assess PD patients’ condition [314]. Integrated biosensing is thought to have the highest accuracy and sensitivity [315], while it can directly convert interactions between biological target analytes and receptors into charge signals without any other labeling process [45]. A unique feature of BioFETs is their miniaturization ability to integrate semiconductor circuits or be in conjunction with microfluidic chips to measure multiple samples at the same time (Figure 12) [75,316]. Constructing a BioFETs array with multiple sensing windows, where each module independently detects a specific biomarker without cross-interference, has been demonstrated to be a feasible approach while maintaining high sensitivity and precision [314]. To create multiple sensing windows, polymer layers such as PMMA were frequently used to spin-coat the array to passivate the source/drain electrodes, leaving a sensing area in the channel using electron beam lithography. Different sensing windows were then functionalized separately with specific antibodies to enable independent biomarker detection [314].
Compared to sensor array, integrating microfluidics and immunoassays into lab-on-a-chip [319] devices is more effective and standardized. Microfluidic technology is frequently used to integrate multiple BioFETs into one system and guide samples to the channel positions of different BioFETs to achieve specific binding and differentiation of targets [320]. It helps to detect biomarkers in shorter response time and with lower reagent consumption [45], which contributes to the prototype development of a miniaturized rapid PoC platform for PD routine monitoring. The fabrication of microfluidic systems typically involves coating a Si wafer with photoresist, patterning the microfluidic structures using a photomask, sealing them with PDMS, and bonding the system to BioFETs via O2 plasma treatment [268]. Alternatively, many studies employed 3D printing to directly fabricate polycarbonate as the top and bottom layers of the chip, using a polyimide film as the channel and sensing layer. The source, drain, and gate electrodes of the BioFETs are then patterned on the PI film through photolithography and thermal deposition [321]. Through this technology, a study integrated two different antibiotics-functionalized BioFETs into a dual-channel microfluidic chip to distinguish between GPB and GNB. The portable BioFETs exhibited unprecedented high sensitivity and real-time target specificity with a LOD of 100 CFU/mL, which would facilitate physicians to quickly make a judgement on the type of bacterial infections during PD, reducing the likelihood of other complications as well as the overdose of antibiotic usage [268]. Another study developed a dual-mode diagnostic platform based on a microfluidic chip using 3D printing, integrating electrochemiluminescence (ECL) sensors and BioFETs within the microfluidic device. The microfluidic channels guide the test samples through the electro-optical sensor units, and similar to the previous setup, the gate electrode of the FET is also used as the working electrode for the ECL sensor. Due to the different signal transduction mechanisms, the combined ECL and FET sensors do not interfere with each other in practical tests. To further reduce uncertainties in complex testing environments, the dual-mode output is refined using machine learning (ML) algorithms to denoise the incoming signals and construct a three-dimensional hyperplane that facilitates multi-class diagnosis. Finally, the platform demonstrated diagnostic accuracy of 99% for 501 clinical samples of tuberculosis (TB), human rhinovirus (HRV), and Group B Streptococcus (GBS) from serum and nasopharyngeal/rectal swabs, whereas existing PoC technologies achieved diagnostic accuracy of 77–93% at 100% statistical power [321].
To improve the reliability of sensor arrays and precisely differentiate signals from multiple analytes, various algorithms were implemented to optimize both the detection and analysis processes, including those refine signal clarity by processing interference [322] and compensate for device variations caused by imperfections in material synthesis and inconsistencies in device fabrication [318]. While combined systems provide two categories of clinical data (positive and negative results), in practice, there are ambiguous samples that cannot be identified by the combined system, thus transforming the problem into a multi-class classification task [321]. For example, in one previous study, a dual-channel GFET was designed to simultaneously detect dopamine (DA) and glucose. The sensor utilized 11-mercapto-1-undecanoic acid-gold nanoclusters (MUA-AuNCs) and PBA to enhance the selectivity and sensitivity of the FET for DA and glucose detection. To differentiate the signals from the two analytes, a unique data fusion algorithm based on a least squares method was employed. This approach utilized precise DA readings from one channel to correct for interference and calculate the accurate glucose concentration in the mixed solution, thereby ensuring high detection accuracy [322].
Building on the previous example of data fusion algorithms, another study employed a different approach to enhance the performance of GFET-based sensors. Three distinct ion-selective membranes (ISMs) were integrated onto the GFET to selectively recognize potassium, sodium, and calcium ions in urine and sweat. The GFET was composed of over 200 integrated sensing units (16 × 16) and divided into three separate regions, which made it challenging to accurately quantify the concentration of three ions. To address this issue, the team utilized a source measure unit (SMU) to assess the electrical performance of the fabricated BioFETs and confirmed their typical output and transfer curves. Subsequently, various algorithms were employed to establish regression models, including linear regression (LR), support vector regression (SVR), decision tree regression (DTR), and random forest regression [323], with the goal of analyzing their respective feature curves. By evaluating metrics such as mean square error [115], root mean square error (RMSE), mean absolute error, and R2, it was found that the RFR model exhibited the best performance in estimating Ca2+ concentration without the need for calibration, with the predicted concentrations closely aligning with the actual concentrations (y = x). The integration of these algorithms not only provided a calibration method to reduce device-to-device variability, but also enhanced the accuracy of ion classification, further improving the performance of the GFET-based sensor system [317]. Subsequent studies employed RFR to extract additional parameters reflecting the changes in graphene films following deposition. Beyond the Dirac point [324], other features derived from the I-V curve, such as the maximum and minimum transconductance (for both electrons and holes), were analyzed to study the physical property variations of graphene. Compared to other algorithms that are also compatible with multiplexed sensor chip data analysis, such as deep neural networks (DNN) and support vector machines (SVM), RFR, owing to its tree-based structure, is more interpretable and requires less data. This makes RFR particularly suitable for analyzing data sets from sensor chips with fewer variables or simpler data structures [318].

4.3.2. Development of PoC Devices

With further research into PD, additional physiological fluids—such as saliva—emerged as potential alternatives for monitoring metabolic by-products associated with renal failure. Moreover, tears and sweat could also be considered physiological solutions for assessing creatinine or glucose levels during PD [207]. BioFETs hold significant potential for developing wearable devices tailored to PD patient monitoring through these various fluids, enabling personalized adjustments to treatment regimens. They have been incorporated with different flexible materials and designs to ensure wearability (Figure 13).
For instance, the extended gate of the BioFETs was designed as a microneedle to penetrate the skin and measure Na+ in the skin interstitial fluid when worn [325]. In their study, researchers first created solid polystyrene microneedles using PDMS molds. These microneedles then served as the base for depositing gold electrodes, while silver was applied to another microneedle electrode to function as a reference electrode. The stretchable design was achieved by patterning AgNWs into elastomeric SIS substrate. Thanks to the interconnected network structure and high conductivity of AgNW, the electrodes continue to perform reliably even when bent, twisted, or stretched. Mechanical tests using SIS films with a thickness gradient loaded with these rigid microneedles demonstrated that, even when the patch is stretched to 100% strain, the central region remains intact and flexible. During the detection process, the BioFETs convert the concentration input into an electrical signal, which can be sent directly to doctors and clinics via Internet-of-Things-enabled smartphones or computers. ISFET was integrated with microfluidics to make a wearable sweat analyzer to track the Na+ and K+ concentrations changes in sweat in real time. An SU-8 passivation layer was deposited on the top of the fabricated ISFET and patterned to create an integrated passive microfluidic interface. This setup leverages capillary action to direct biological fluids into an array of functionalized ISFETs for detection and transport to the outlet area [326]. Furthermore, nanoFETs were incorporated into injectable electronics or direct implantation into specific tissues for vivo monitoring or intracellular measurements, such as the development of SiNW-BioFETs arrays embedded in engineered tissue patches [294].
Figure 13. Wearable BioFETs for the detection of biomarkers in different human fluids. (A) Schematic of a wearable BioFET patch for assessment of Na+ concentration in sweat, where a plasticizer-free ISM based on fluoropolysiloxane (FPS) improves biocompatibility [327]. (B) Schematic of smart contact lens modified with antibody for the cortisol detection [328]. (C) Schematic of an ultra-flexible GFET for the cytokine detection in biofluids. The flexibility and stretchability of GFET allows it to attach onto the human wrist and crooked fingers [279]. (D) Top: schematic of a smart watch functionalized by aptamer-based FET for the cortisol detection in saliva and sweat samples. Bottom: Schematic of FET structure. Right: top view and side view of In2O3-based FET with a flexible polyimide substrate. ADC, analog-digital converter [129] (images used and edited under Creative Common licenses).
Figure 13. Wearable BioFETs for the detection of biomarkers in different human fluids. (A) Schematic of a wearable BioFET patch for assessment of Na+ concentration in sweat, where a plasticizer-free ISM based on fluoropolysiloxane (FPS) improves biocompatibility [327]. (B) Schematic of smart contact lens modified with antibody for the cortisol detection [328]. (C) Schematic of an ultra-flexible GFET for the cytokine detection in biofluids. The flexibility and stretchability of GFET allows it to attach onto the human wrist and crooked fingers [279]. (D) Top: schematic of a smart watch functionalized by aptamer-based FET for the cortisol detection in saliva and sweat samples. Bottom: Schematic of FET structure. Right: top view and side view of In2O3-based FET with a flexible polyimide substrate. ADC, analog-digital converter [129] (images used and edited under Creative Common licenses).
Biosensors 15 00193 g013
The biocompatibility of these BioFETs should be highlighted to prevent adverse reactions. For instance, while higher sodium-to-creatinine ratios are often associated with the progression of end-stage renal failure [329], ISMs are frequently combined with wearable BioFETs to detect Na+ in sweat, in which ISMs are drop-cast on top of the sensing dielectric. The ionophores in the ISM interacted with the target ions with high selectivity to reduce the non-specific adhesion [330]. However, this system usually needs a reference electrode [331], so CNT was then incorporated with ISFETs. Ag/AgCl ink and polyvinyl butyral membrane were used to fabricate an integrated reference electrode, and the CNT surface was functionalized with a polyvinyl chloride (PVC)-based ISM. The BioFETs can detect sodium ions in the range of 0.1 to 100 mM with high sensitivity. This structure solved the problem of large size of the traditional ISFETs reference electrode, and it can provide a stable potential comparable to that of the commercial reference electrode [332]. However, PVC-based ISMs usually include a plasticizer, and if ISM-coated sensors are used to monitor health parameters in daily life, prolonged immersion of the plasticizer in solution may lead to cytotoxicity and shorter sensor lifetime [333]. New strategies have been developed using Ta2O5 oxide membranes and fluoropolycrystalline silicon (FPS) to fabricate the plasticizer-free ISMs, and the device exhibited higher biocompatibility as well as retained sufficient sensitivity and selectivity to Na+ [327]. The combination of biocompatible and flexible coating materials such as hydrogels or liquid metals with these systems has been proved to minimize immune reactions and potential tissue damage as well [334]. Additionally, these coatings contribute to enhance the sensitivity of BioFETs as mentioned above. Spin-coated hydrogels synthesized by free radical polymerization were deposited on SiNW BioFETs serving as dielectric and functional layers. Compared to conventional SiNW BioFETs, the ones modified with hydrogel coatings have higher biocompatibility and time response [335].
Building on the advancements in biocompatibility and flexible designs, reusability is another essential aspect of PoC devices that ensure long-term functionality and cost-efficiency for patients and clinicians. The primary requirement for reusable sensors is the regenerability of probe–analyte interactions. Methods such as acid base rinsing, heating, and voltage biasing are commonly employed to achieve in situ dissociation of probe–analyte complexes [336]. For instance, one approach leverages the reversible interaction between PBA and glucose. In acidic conditions, the bonds between glucose and PBA are disrupted, allowing glucose molecules to dissociate from boronate ester complexes and regenerating the device surface for subsequent cycle [130].
Several other innovative strategies further enhance the reusability of FET-based biosensors. In the urea detection, back-gated BioFETs were placed within a sealed configuration to measure the urea. Urea molecules are enzymatically dissociated into ammonia gas, enabling reversible adsorption-desorption cycles that facilitate multiple uses without significant performance loss [274]. In another study, they immobilized the cortisol antibody onto the magnetic beads. Following analyte detection, the magnetic nanoparticles can be removed using an external magnetic field, effectively resetting the sensor surface while preserving the functional integrity of the device [136]. Lastly, a novel approach has been demonstrated with highly durable and reusable drug molecule-based OECTs (DM-OECTs). In this system, protonated gefitinib is electrostatically adsorbed onto the PEDOT:PSS layer. The introduction of competitive gefitinib-EGFR binding induces additional surface gating and de-gating of the active layer, creating a dynamic and refreshable sensing interface [336]. These strategies exemplify the potential of innovative reusability solutions to extend the operational lifespan and functionality of PoC devices, furthering their applicability in real-world diagnostics.

5. Concluding Remarks

BioFETs offer promising advancements for the field of PD by enabling precise and real-time monitoring of critical biomarkers, which is essential for personalized patient care. By addressing the sensitivity and selectivity challenges posed by the complex physiological environment, BioFETs can be optimized to deliver accurate biomarkers detection such as urea, creatinine, inflammatory markers, and bacterial infections, enhancing their clinical relevance for daily monitoring. However, the heterogeneous composition of PDE and the high-ionic environment pose considerable challenges to the performance of BioFETs, underscoring the need for stable and specific transducing materials that can maintain functionality over extended clinical use.
The scalability of BioFET fabrication and integration into existing PD frameworks are equally vital for transitioning from laboratory prototypes to clinically viable devices. Advances in nanomaterials, such as MoS2 and carbon-based technologies, alongside cost-effective manufacturing processes, including CMOS and CVD compatibility, will be instrumental in making BioFETs accessible at a commercial level. Additionally, the development of wearable, PoC BioFETs offers a novel approach to continuous and minimally invasive monitoring across various biofluids, facilitating frequent and accessible assessments for PD patients.
Future work should prioritize embedding BioFETs into closed-loop PD systems, where real-time biomarker data from BioFETs automatically adjust dialysate composition via feedback-controlled pumps. Such integration would minimize human error and optimize solute clearance. The long-term stability, biocompatibility, and reusability of BioFETs under variable physiological conditions, as well as the integration of multiplexing capabilities to allow concurrent monitoring of multiple biomarkers should be further studies as well. With these improvements, BioFETs could become a cornerstone of patient-centered care in PD, delivering actionable insights and facilitating personalized treatment strategies that ultimately enhance patient outcomes and quality of life. Despite the challenges, the forward momentum in research suggests a transformative future for PD management.

Author Contributions

Conceptualization, Y.-L.C., D.S.-H.W., D.W.-C.W. and J.C.-W.C.; methodology, D.W.-C.W. and J.C.-W.C.; validation, Y.-L.C. and D.S.-H.W.; formal analysis, Q.W., Z.-A.Z. and K.-Y.Y.; investigation, Q.W., Z.-A.Z. and K.-Y.Y.; data curation, Q.W., Z.-A.Z. and K.-Y.Y.; writing—original draft preparation, Q.W., Z.-A.Z. and K.-Y.Y.; writing—review and editing, D.S.-H.W., D.W.-C.W. and J.C.-W.C. visualization, Q.W., Z.-A.Z. and K.-Y.Y.; supervision, D.W.-C.W. and J.C.-W.C.; project administration, D.W.-C.W. and J.C.-W.C.; funding acquisition, J.C.-W.C. All authors have read and agreed to the published version of the manuscript.

Funding

The study was supported by the collaborative fund with Vincent RenalCan Medical Technology Company Limited (reference number: P0051331). The funder had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Bello, A.K.; Okpechi, I.G.; Osman, M.A.; Cho, Y.; Cullis, B.; Htay, H.; Jha, V.; Makusidi, M.A.; McCulloch, M.; Shah, N. Epidemiology of peritoneal dialysis outcomes. Nat. Rev. Nephrol. 2022, 18, 779–793. [Google Scholar] [CrossRef] [PubMed]
  2. Chen, T.K.; Knicely, D.H.; Grams, M.E. Chronic kidney disease diagnosis and management: A review. JAMA 2019, 322, 1294–1304. [Google Scholar] [CrossRef]
  3. Francis, A.; Harhay, M.N.; Ong, A.; Tummalapalli, S.L.; Ortiz, A.; Fogo, A.B.; Fliser, D.; Roy-Chaudhury, P.; Fontana, M.; Nangaku, M. Chronic kidney disease and the global public health agenda: An international consensus. Nat. Rev. Nephrol. 2024, 20, 473–486. [Google Scholar] [CrossRef]
  4. Jadoul, M.; Aoun, M.; Imani, M.M. The major global burden of chronic kidney disease. Lancet Glob. Health 2024, 12, e342–e343. [Google Scholar] [CrossRef]
  5. Teitelbaum, I. Peritoneal dialysis. N. Engl. J. Med. 2021, 385, 1786–1795. [Google Scholar] [CrossRef]
  6. Chuasuwan, A.; Pooripussarakul, S.; Thakkinstian, A.; Ingsathit, A.; Pattanaprateep, O. Comparisons of quality of life between patients underwent peritoneal dialysis and hemodialysis: A systematic review and meta-analysis. Health Qual. Life Outcomes 2020, 18, 191. [Google Scholar] [CrossRef]
  7. Kalantar-Zadeh, K.; Jafar, T.H.; Nitsch, D.; Neuen, B.L.; Perkovic, V. Chronic kidney disease. Lancet 2021, 398, 786–802. [Google Scholar] [CrossRef]
  8. Stepanova, N. SGLT2 inhibitors in peritoneal dialysis: A promising frontier toward improved patient outcomes. Ren. Replace. Ther. 2024, 10, 5. [Google Scholar] [CrossRef]
  9. Perl, J.; Fuller, D.S.; Bieber, B.A.; Boudville, N.; Kanjanabuch, T.; Ito, Y.; Nessim, S.J.; Piraino, B.M.; Pisoni, R.L.; Robinson, B.M. Peritoneal dialysis–related infection rates and outcomes: Results from the Peritoneal Dialysis Outcomes and Practice Patterns Study (PDOPPS). Am. J. Kidney Dis. 2020, 76, 42–53. [Google Scholar] [CrossRef]
  10. Bonomini, M.; Borras, F.E.; Troya-Saborido, M.; Carreras-Planella, L.; Di Liberato, L.; Arduini, A. Proteomic research in peritoneal dialysis. Int. J. Mol. Sci. 2020, 21, 5489. [Google Scholar] [CrossRef]
  11. Suryantoro, S.D.; Thaha, M.; Sutanto, H.; Firdausa, S. Current insights into cellular determinants of peritoneal fibrosis in peritoneal dialysis: A narrative review. J. Clin. Med. 2023, 12, 4401. [Google Scholar] [CrossRef] [PubMed]
  12. Davies, S.J. Peritoneal dialysis—Current status and future challenges. Nat. Rev. Nephrol. 2013, 9, 399–408. [Google Scholar] [CrossRef] [PubMed]
  13. Alexandrou, M.-E.; Balafa, O.; Sarafidis, P. Assessment of hydration status in peritoneal dialysis patients: Validity, prognostic value, strengths, and limitations of available techniques. Am. J. Nephrol. 2020, 51, 589–612. [Google Scholar] [CrossRef] [PubMed]
  14. Fung, W.W.-S.; Li, P.K.-T. Recent advances in novel diagnostic testing for peritoneal dialysis-related peritonitis. Kidney Res. Clin. Pract. 2022, 41, 156. [Google Scholar] [CrossRef]
  15. Andreoli, M.C.C.; Totoli, C. Peritoneal dialysis. Rev. Assoc. Médica Bras. 2020, 66, s37–s44. [Google Scholar] [CrossRef]
  16. Mizumasa, T.; Honda, K.; Aoki, S.; Hamada, C.; Miyazaki, M.; Ito, Y.; Tanno, Y.; Nakano, T.; Nakayama, M.; Peritoneal Biopsy Study Group of the Japanese Society for Peritoneal Dialysis. Proposal of peritoneal biopsy procedures for patients undergoing peritoneal dialysis. Ren. Replace. Ther. 2020, 6, 8. [Google Scholar] [CrossRef]
  17. Yang, J.; Cai, M.; Wan, J.; Wang, L.; Luo, J.; Li, X.; Gong, W.; He, Y.; Chen, J. Effluent decoy receptor 2 as a novel biomarker of peritoneal fibrosis in peritoneal dialysis patients. Perit. Dial. Int. 2022, 42, 631–639. [Google Scholar] [CrossRef]
  18. Morelle, J.; Stachowska-Pietka, J.; Öberg, C.; Gadola, L.; La Milia, V.; Yu, Z.; Lambie, M.; Mehrotra, R.; de Arteaga, J.; Davies, S. ISPD recommendations for the evaluation of peritoneal membrane dysfunction in adults: Classification, measurement, interpretation and rationale for intervention. Perit. Dial. Int. 2021, 41, 352–372. [Google Scholar] [CrossRef]
  19. Hao, N.; Chiou, T.T.-Y.; Wu, C.-H.; Lei, Y.-Y.; Liang, P.-L.; Chao, M.-C.; Yang, H.; Chen, J.-B. Longitudinal Changes of PAI-1, MMP-2, and VEGF in Peritoneal Effluents and Their Associations with Peritoneal Small-Solute Transfer Rate in New Peritoneal Dialysis Patients. BioMed Res. Int. 2019, 2019, 2152584. [Google Scholar] [CrossRef]
  20. Krediet, R.T.; Struijk, D.G. Peritoneal changes in patients on long-term peritoneal dialysis. Nat. Rev. Nephrol. 2013, 9, 419–429. [Google Scholar] [CrossRef]
  21. Wiesenhofer, F.M.; Herzog, R.; Boehm, M.; Wagner, A.; Unterwurzacher, M.; Kasper, D.C.; Alper, S.L.; Vychytil, A.; Aufricht, C.; Kratochwill, K. Targeted metabolomic profiling of peritoneal dialysis effluents shows anti-oxidative capacity of alanyl-glutamine. Front. Physiol. 2019, 9, 1961. [Google Scholar] [CrossRef]
  22. Szeto, C.-C.; Li, P.K.-T. Peritoneal dialysis–associated peritonitis. Clin. J. Am. Soc. Nephrol. 2019, 14, 1100–1105. [Google Scholar] [CrossRef] [PubMed]
  23. Li, P.K.-T.; Chow, K.M.; Cho, Y.; Fan, S.; Figueiredo, A.E.; Harris, T.; Kanjanabuch, T.; Kim, Y.-L.; Madero, M.; Malyszko, J. ISPD peritonitis guideline recommendations: 2022 update on prevention and treatment. Perit. Dial. Int. 2022, 42, 110–153. [Google Scholar] [CrossRef]
  24. Kim, H.J.; Seong, E.Y.; Lee, W.; Kim, S.; Ahn, H.-S.; Yeom, J.; Kim, K.; Kwon, C.H.; Song, S.H. Comparative analysis of therapeutic effects between medium cut-off and high flux dialyzers using metabolomics and proteomics: Exploratory, prospective study in hemodialysis. Sci. Rep. 2021, 11, 17335. [Google Scholar] [CrossRef]
  25. Kim, H.J.; Choo, M.; Kwon, H.N.; Yoo, K.D.; Kim, Y.; Tsogbadrakh, B.; Kang, E.; Park, S.; Oh, K.-H. Metabolomic profiling of overnight peritoneal dialysis effluents predicts the peritoneal equilibration test type. Sci. Rep. 2023, 13, 3803. [Google Scholar] [CrossRef] [PubMed]
  26. Bowen, T.; Meran, S.; Williams, A.P.; Newbury, L.J.; Sauter, M.; Sitter, T. Regulation of synthesis and roles of hyaluronan in peritoneal dialysis. BioMed Res. Int. 2015, 2015, 427038. [Google Scholar] [CrossRef] [PubMed]
  27. Simoes-Silva, L.; Araujo, R.; Pestana, M.; Soares-Silva, I.; Sampaio-Maia, B. Peritoneal microbiome in end-stage renal disease patients and the impact of peritoneal dialysis therapy. Microorganisms 2020, 8, 173. [Google Scholar] [CrossRef]
  28. Szeto, C.C.; Ng, J.K.-C.; Fung, W.W.-S.; Lai, K.-B.; Chow, K.-M.; Li, P.K.-T.; Massiah, A.; Alcolea-Medina, A.; Wilks, M.; Fan, S.L. Polymerase chain reaction/electrospray ionization–mass spectrometry (PCR/ESI-MS) is not suitable for rapid bacterial identification in peritoneal dialysis effluent. Perit. Dial. Int. 2021, 41, 96–100. [Google Scholar] [CrossRef]
  29. van Hougenhouck-Tulleken, W.G.; Lebre, P.H.; Said, M.; Cowan, D.A. Bacterial pathogens in peritoneal dialysis peritonitis: Insights from next-generation sequencing. Perit. Dial. Int. 2020, 40, 581–586. [Google Scholar] [CrossRef]
  30. Chen, X.; Liang, Y.; Tang, N.; Li, C.; Zhang, Y.; Xu, F.; Shi, G.; Zhang, M. Ultrasensitive sensing urinary cystatin C via an interface-engineered graphene extended-gate field-effect transistor for non-invasive diagnosis of chronic kidney disease. Biosens. Bioelectron. 2024, 249, 116016. [Google Scholar] [CrossRef]
  31. Singh, J.; Thakur, P.; Jayram, S.; Singh, G. Impacts of graphene mass-manufacturing processes on bio-FET performance. In Proceedings of the 2021 International Conference on Electrical, Computer and Energy Technologies (ICECET), Cape Town, South Africa, 9–10 December 2021; pp. 1–5. [Google Scholar]
  32. Zafar, S.; Lu, M.; Jagtiani, A. Comparison between field effect transistors and bipolar junction transistors as transducers in electrochemical sensors. Sci. Rep. 2017, 7, 41430. [Google Scholar] [CrossRef] [PubMed]
  33. Dastgeer, G.; Shahzad, Z.M.; Chae, H.; Kim, Y.H.; Ko, B.M.; Eom, J. Bipolar junction transistor exhibiting excellent output characteristics with a prompt response against the selective protein. Adv. Funct. Mater. 2022, 32, 2204781. [Google Scholar] [CrossRef]
  34. Yuan, H.; Wang, B.; Yeom, S.-H.; Kang, B.-H.; Kim, K.-J.; Lee, J.-H.; Kang, S.-W.; Kwon, D.-H. Novel Biosensor Based on MOSFET-BJT Hybrid Mode of Gated Lateral Bipolar Junction Transistor for C-reactive Protein Detection. In Proceedings of the 2012 Third International Conference on Intelligent Systems Modelling and Simulation, Kota Kinabalu, Malaysia, 8–10 February 2012; pp. 756–759. [Google Scholar]
  35. Rabbani, G.; Khan, M.E.; Ahmad, E.; Khan, M.V.; Ahmad, A.; Khan, A.U.; Ali, W.; Zamzami, M.A.; Bashiri, A.H.; Zakri, W. Serum CRP biomarker detection by using carbon nanotube field-effect transistor (CNT-FET) immunosensor. Bioelectrochemistry 2023, 153, 108493. [Google Scholar] [CrossRef]
  36. Santos, L.A. An overview on bipolar junction transistor as a sensor for X-ray beams used in medical diagnosis. Sensors 2022, 22, 1923. [Google Scholar] [CrossRef]
  37. Basov, M. Pressure sensor with novel electrical circuit utilizing bipolar junction transistor. In Proceedings of the 2021 IEEE Sensors, Sydney, Australia, 31 October–3 November 2021; pp. 1–4. [Google Scholar]
  38. Zafar, S.; D’Emic, C.; Ning, T. Silicon Device-based Electronic Biosensors for Sustainability and Healthcare. In Proceedings of the 2024 31st International Workshop on Active-Matrix Flatpanel Displays and Devices (AM-FPD), Kyoto, Japan, 2–5 July 2024; pp. 155–157. [Google Scholar]
  39. Formisano, N.; Bhalla, N.; Heeran, M.; Martinez, J.R.; Sarkar, A.; Laabei, M.; Jolly, P.; Bowen, C.R.; Taylor, J.T.; Flitsch, S. Inexpensive and fast pathogenic bacteria screening using field-effect transistors. Biosens. Bioelectron. 2016, 85, 103–109. [Google Scholar] [CrossRef]
  40. Bakhshi, T.; Yousaf, F. Securing Bio-FET Interfaces in IoBNT Systems using Deep Learning Techniques. In Proceedings of the 2023 17th International Conference on Open Source Systems and Technologies (ICOSST), Lahore, Pakistan, 20–21 December 2023; pp. 1–6. [Google Scholar]
  41. Pindoo, I.A.; Sinha, S.K. Increased sensitivity of biosensors using evolutionary algorithm for bio-medical applications. Radioelectron. Commun. Syst. 2020, 63, 308–318. [Google Scholar] [CrossRef]
  42. Cheng, J.C. Acinetobacter Baumannii Detection Using A Bio-FET Device: Towards Healthcare Infection Management. ECS Trans. 2023, 111, 17. [Google Scholar] [CrossRef]
  43. Nguyen, T.T.-H.; Nguyen, C.M.; Huynh, M.A.; Vu, H.H.; Nguyen, T.-K.; Nguyen, N.-T. Field effect transistor based wearable biosensors for healthcare monitoring. J. Nanobiotechnol. 2023, 21, 411. [Google Scholar] [CrossRef]
  44. Chen, C.H.; Perl, J.; Teitelbaum, I. Prescribing high-quality peritoneal dialysis: The role of preserving residual kidney function. Perit. Dial. Int. 2020, 40, 274–281. [Google Scholar] [CrossRef]
  45. Hao, R.; Liu, L.; Yuan, J.; Wu, L.; Lei, S. Recent advances in field effect transistor biosensors: Designing strategies and applications for sensitive assay. Biosensors 2023, 13, 426. [Google Scholar] [CrossRef]
  46. Panahi, A.; Sadighbayan, D.; Forouhi, S.; Ghafar-Zadeh, E. Recent advances of field-effect transistor technology for infectious diseases. Biosensors 2021, 11, 103. [Google Scholar] [CrossRef] [PubMed]
  47. Alnaji, N.; Wasfi, A.; Awwad, F. The design of a point of care FET biosensor to detect and screen COVID-19. Sci. Rep. 2023, 13, 4485. [Google Scholar] [CrossRef] [PubMed]
  48. Park, C.; Choi, W.; Kim, D.; Jin, B.; Lee, J.-S. Highly sensitive detection of influenza A (H1N1) virus with silicon nanonet BioFETs. IEEE Sens. J. 2019, 19, 10985–10990. [Google Scholar] [CrossRef]
  49. Sung, D.; Koo, J. A review of BioFET’s basic principles and materials for biomedical applications. Biomed. Eng. Lett. 2021, 11, 85–96. [Google Scholar] [CrossRef]
  50. Zhao, W.; Hu, J.; Liu, J.; Li, X.; Sun, S.; Luan, X.; Zhao, Y.; Wei, S.; Li, M.; Zhang, Q. Si nanowire Bio-FET for electrical and label-free detection of cancer cell-derived exosomes. Microsyst. Nanoeng. 2022, 8, 57. [Google Scholar] [CrossRef]
  51. Al-Younis, Z.K.; Almajidi, Y.Q.; Mansouri, S.; Ahmad, I.; Turdialiyev, U.; Alsaab, H.O.; Ramadan, M.F.; Joshi, S.; Alawadi, A.H.; Alsaalamy, A. Label-Free Field Effect Transistors (FETs) for Fabrication of Point-of-Care (POC) Biomedical Detection Probes. Crit. Rev. Anal. Chem. 2024, 1–22. [Google Scholar] [CrossRef]
  52. Zhang, Z.; Hu, J.J.; Lin, S.; Wu, J.; Xia, F.; Lou, X. Field effect transistor biosensors for healthcare monitoring. Interdiscip. Med. 2024, 2, e20240032. [Google Scholar] [CrossRef]
  53. Sadighbayan, D.; Hasanzadeh, M.; Ghafar-Zadeh, E. Biosensing based on field-effect transistors (FET): Recent progress and challenges. TrAC Trends Anal. Chem. 2020, 133, 116067. [Google Scholar] [CrossRef]
  54. Tsai, S.-Y.; Huang, C.-C.; Chen, P.-H.; Tripathi, A.; Wang, Y.-R.; Wang, Y.-L.; Chen, J.-C. Rapid drug-screening platform using field-effect transistor-based biosensors: A study of extracellular drug effects on transmembrane potentials. Anal. Chem. 2021, 94, 2679–2685. [Google Scholar] [CrossRef]
  55. Bonini, A.; Vivaldi, F.M.; Herrera, E.; Melai, B.; Kirchhain, A.; Sajama, N.V.P.; Mattonai, M.; Caprioli, R.; Lomonaco, T.; Di Francesco, F. A graphenic biosensor for real-time monitoring of urea during dialysis. IEEE Sens. J. 2020, 20, 4571–4578. [Google Scholar] [CrossRef]
  56. Purwidyantri, A.; Lai, H.-C.; Tsai, S.-H.; Luo, J.-D.; Chiou, C.-C.; Tian, Y.-C.; Cheng, C.-H.; Lin, Y.-T.; Lai, C.-S. Sensing performance of fibronectin-functionalized Au-EGFET on the detection of S. epidermidis biofilm and 16S rRNA of infection-related bacteria in peritoneal dialysis. Sens. Actuators B Chem. 2015, 217, 92–99. [Google Scholar] [CrossRef]
  57. Manimekala, T.; Sivasubramanian, R.; Dharmalingam, G. Nanomaterial-based biosensors using field-effect transistors: A review. J. Electron. Mater. 2022, 51, 1950–1973. [Google Scholar] [CrossRef] [PubMed]
  58. Masola, V.; Bonomini, M.; Borrelli, S.; Di Liberato, L.; Vecchi, L.; Onisto, M.; Gambaro, G.; Palumbo, R.; Arduini, A. Fibrosis of peritoneal membrane as target of new therapies in peritoneal dialysis. Int. J. Mol. Sci. 2022, 23, 4831. [Google Scholar] [CrossRef] [PubMed]
  59. Albarghouthi, F.M.; Semeniak, D.; Khanani, I.; Doherty, J.L.; Smith, B.N.; Salfity, M.; MacFarlane, Q.; Karappur, A.; Noyce, S.G.; Williams, N.X. Addressing Signal Drift and Screening for Detection of Biomarkers with Carbon Nanotube Transistors. ACS Nano 2024, 18, 5698–5711. [Google Scholar] [CrossRef]
  60. Shafi, N.; Parmaar, J.S.; Porwal, A.; Bhat, A.M.; Sahu, C.; Periasamy, C. Gate All around junctionless dielectric modulated biofet based hybrid biosensor: Design, simulation and performance investigation. Silicon 2021, 13, 2041–2052. [Google Scholar] [CrossRef]
  61. Garg, A.; Ratnesh, R.K.; Chauhan, R.K.; Mittal, N.; Shankar, H. Current advancement and progress in BioFET: A review. In Proceedings of the 2022 International Conference on Signal and Information Processing (IConSIP), Pune, India, 26–27 August 2022; pp. 1–7. [Google Scholar]
  62. Vu, C.-A.; Chen, W.-Y. Field-effect transistor biosensors for biomedical applications: Recent advances and future prospects. Sensors 2019, 19, 4214. [Google Scholar] [CrossRef]
  63. Hamed, A.; Asad, M.; Wei, M.-D.; Vorobiev, A.; Stake, J.; Negra, R. Integrated 10-GHz graphene FET amplifier. IEEE J. Microw. 2021, 1, 821–826. [Google Scholar] [CrossRef]
  64. Elli, G.; Ciocca, M.; Lugli, P.; Petti, L. Field-effect-transistor based biosensors: A review of their use in environmental monitoring applications. In Proceedings of the 2021 IEEE International Workshop on Metrology for Agriculture and Forestry (MetroAgriFor), Trento-Bolzano, Italy, 3–5 November 2021; pp. 102–107. [Google Scholar]
  65. Mitta, S.B.; Choi, M.S.; Nipane, A.; Ali, F.; Kim, C.; Teherani, J.T.; Hone, J.; Yoo, W.J. Electrical characterization of 2D materials-based field-effect transistors. 2D Mater. 2020, 8, 012002. [Google Scholar] [CrossRef]
  66. Veeralingam, S.; Badhulika, S. Surface functionalized β-Bi2O3 nanofibers based flexible, field-effect transistor-biosensor (BioFET) for rapid, label-free detection of serotonin in biological fluids. Sens. Actuators B Chem. 2020, 321, 128540. [Google Scholar] [CrossRef]
  67. Wadhera, T.; Kakkar, D.; Wadhwa, G.; Raj, B. Recent advances and progress in development of the field effect transistor biosensor: A review. J. Electron. Mater. 2019, 48, 7635–7646. [Google Scholar] [CrossRef]
  68. Farahmandpour, M.; Ansari, H.R.; Kordrostami, Z. Flexible Enzyme-Free Gate Engineered Bio-FET Glucose Sensor based on Nickel-Tungstate Microcrystals. IEEE Sens. J. 2024, 24, 9308–9316. [Google Scholar] [CrossRef]
  69. Devi, K.S.; Keshwani, G.; Thakur, H.R.; Dutta, J.C. Fabrication and Physical Characterization of Different Layers of CNT-BioFET for Creatinine Detection. In Proceedings of the Pattern Recognition and Machine Intelligence: 8th International Conference, PReMI 2019, Tezpur, India, 17–20 December 2019; pp. 535–542. [Google Scholar]
  70. Ghasemi, F.; Salimi, A. Advances in 2d based field effect transistors as biosensing platforms: From principle to biomedical applications. Microchem. J. 2023, 187, 108432. [Google Scholar] [CrossRef]
  71. Baldacchini, C.; Montanarella, A.F.; Francioso, L.; Signore, M.A.; Cannistraro, S.; Bizzarri, A.R. A reliable biofet immunosensor for detection of p53 tumour suppressor in physiological-like environment. Sensors 2020, 20, 6364. [Google Scholar] [CrossRef] [PubMed]
  72. Kajisa, T.; Sakata, T. Molecularly imprinted artificial biointerface for an enzyme-free glucose transistor. ACS Appl. Mater. Interfaces 2018, 10, 34983–34990. [Google Scholar] [CrossRef]
  73. Piccinini, E.; Alberti, S.; Longo, G.S.; Berninger, T.; Breu, J.; Dostalek, J.; Azzaroni, O.; Knoll, W. Pushing the boundaries of interfacial sensitivity in graphene FET sensors: Polyelectrolyte multilayers strongly increase the Debye screening length. J. Phys. Chem. C 2018, 122, 10181–10188. [Google Scholar] [CrossRef]
  74. Thriveni, G.; Ghosh, K. Advancement and challenges of biosensing using field effect transistors. Biosensors 2022, 12, 647. [Google Scholar] [CrossRef]
  75. Sakata, T. Signal transduction interfaces for field-effect transistor-based biosensors. Commun. Chem. 2024, 7, 35. [Google Scholar] [CrossRef]
  76. Cheng, Z.; Pang, C.-S.; Wang, P.; Le, S.T.; Wu, Y.; Shahrjerdi, D.; Radu, I.; Lemme, M.C.; Peng, L.-M.; Duan, X. How to report and benchmark emerging field-effect transistors. Nat. Electron. 2022, 5, 416–423. [Google Scholar] [CrossRef]
  77. Zhu, Y.; Wei, Q.; Jin, Q.; Li, G.; Zhang, Q.; Xiao, H.; Li, T.; Wei, F.; Luo, Y. Polyethylene Glycol Functionalized Silicon Nanowire Field-Effect Transistor Biosensor for Glucose Detection. Nanomaterials 2023, 13, 604. [Google Scholar] [CrossRef]
  78. Firoozbakhtian, A.; Rezayan, A.H.; Hajghassem, H.; Rahimi, F.; Ghazani, M.F.; Kalantar, M.; Mohamadsharifi, A. Buried-gate MWCNT FET-based nanobiosensing device for real-time detection of CRP. ACS Omega 2022, 7, 7341–7349. [Google Scholar] [CrossRef]
  79. Papamatthaiou, S.; Estrela, P.; Moschou, D. Printable graphene BioFETs for DNA quantification in Lab-on-PCB microsystems. Sci. Rep. 2021, 11, 9815. [Google Scholar] [CrossRef] [PubMed]
  80. Kwong Hong Tsang, D.; Lieberthal, T.J.; Watts, C.; Dunlop, I.E.; Ramadan, S.; del Rio Hernandez, A.E.; Klein, N. Chemically functionalised graphene FET biosensor for the label-free sensing of exosomes. Sci. Rep. 2019, 9, 13946. [Google Scholar] [CrossRef] [PubMed]
  81. Yojo, L.S.; Rangel, R.C.; Duarte, P.H.; Sasaki, K.R.A.; Martino, J.A. An enzymatic glucose biosensor using the BESOI MOSFET. Solid State Electron. 2024, 211, 108830. [Google Scholar] [CrossRef]
  82. Guo, S.; Wu, K.; Li, C.; Wang, H.; Sun, Z.; Xi, D.; Zhang, S.; Ding, W.; Zaghloul, M.E.; Wang, C. Integrated contact lens sensor system based on multifunctional ultrathin MoS2 transistors. Matter 2021, 4, 969–985. [Google Scholar] [CrossRef]
  83. Lee, H.W.; Kang, D.-H.; Cho, J.H.; Lee, S.; Jun, D.-H.; Park, J.-H. Highly sensitive and reusable membraneless field-effect transistor (FET)-type tungsten diselenide (WSe2) biosensors. ACS Appl. Mater. Interfaces 2018, 10, 17639–17645. [Google Scholar] [CrossRef]
  84. Wang, Z.; Yu, H.; Zhao, Z. Silk fibroin hydrogel encapsulated graphene filed-effect transistors as enzyme-based biosensors. Microchem. J. 2021, 169, 106585. [Google Scholar] [CrossRef]
  85. Shibata, K.; Nakamura, A. An extended gate field-effect transistor (EG-FET) type non-enzymatic glucose sensor with inkjet-printed copper oxide nanoparticles. SN Appl. Sci. 2022, 4, 253. [Google Scholar] [CrossRef]
  86. Saikia, O.; Dutta, J.C.; Thakur, H.R. Carbon nanotube based biologically modified field effect transistors (CNT-BioFETs): A research review. Mater. Today Proc. 2023; in press. [Google Scholar] [CrossRef]
  87. Poghossian, A.; Schöning, M.J. Recent progress in silicon-based biologically sensitive field-effect devices. Curr. Opin. Electrochem. 2021, 29, 100811. [Google Scholar] [CrossRef]
  88. Panahi, A.; Ghafar-Zadeh, E. Emerging Field-Effect Transistor Biosensors for Life Science Applications. Bioengineering 2023, 10, 793. [Google Scholar] [CrossRef]
  89. Hajian, R.; Balderston, S.; Tran, T.; DeBoer, T.; Etienne, J.; Sandhu, M.; Wauford, N.A.; Chung, J.-Y.; Nokes, J.; Athaiya, M. Detection of unamplified target genes via CRISPR–Cas9 immobilized on a graphene field-effect transistor. Nat. Biomed. Eng. 2019, 3, 427–437. [Google Scholar] [CrossRef]
  90. Liang, Y.; Xiao, M.; Xie, J.; Li, J.; Zhang, Y.; Liu, H.; Zhang, Y.; He, J.; Zhang, G.; Wei, N. Amplification-Free Detection of SARS-CoV-2 Down to Single Virus Level by Portable Carbon Nanotube Biosensors. Small 2023, 19, 2208198. [Google Scholar] [CrossRef] [PubMed]
  91. Salzer, W.L. Peritoneal dialysis-related peritonitis: Challenges and solutions. Int. J. Nephrol. Renov. Dis. 2018, 11, 173–186. [Google Scholar] [CrossRef] [PubMed]
  92. Ramil-Gómez, O.; Rodríguez-Carmona, A.; Fernández-Rodríguez, J.A.; Pérez-Fontán, M.; Ferreiro-Hermida, T.; López-Pardo, M.; Pérez-López, T.; López-Armada, M.J. Mitochondrial dysfunction plays a relevant role in Pathophysiology of peritoneal membrane damage induced by peritoneal dialysis. Antioxidants 2021, 10, 447. [Google Scholar] [CrossRef] [PubMed]
  93. Lavecchia di Tocco, F.; Botti, V.; Cannistraro, S.; Bizzarri, A.R. Detection of miR-155 using peptide nucleic acid at physiological-like conditions by surface plasmon resonance and bio-field effect transistor. Biosensors 2024, 14, 79. [Google Scholar] [CrossRef]
  94. Paulose, A.K.; Hou, Y.-J.; Huang, Y.-S.; Chakkalaparambil Dileep, N.; Chiu, C.-L.; Pal, A.; Kalaimani, V.M.; Lin, Z.-H.; Chang, C.-R.; Chen, C.-P. Rapid Escherichia coli cloned DNA detection in serum using an electrical double layer-gated field-effect transistor-based DNA sensor. Anal. Chem. 2023, 95, 6871–6878. [Google Scholar] [CrossRef]
  95. Obermueller, M.; Traby, L.; Weiss-Tessbach, M.; Kriz, R.; Spettel, K.; Schneider, L.; Hohl, L.; Burgmann, H.; Kussmann, M. Staphylococcus aureus small colony variants: A potentially underestimated microbiological challenge in peritoneal dialysis. Int. J. Antimicrob. Agents 2024, 63, 107135. [Google Scholar] [CrossRef]
  96. Cheung, K.M.; Abendroth, J.M.; Nakatsuka, N.; Zhu, B.; Yang, Y.; Andrews, A.M.; Weiss, P.S. Detecting DNA and RNA and differentiating single-nucleotide variations via field-effect transistors. Nano Lett. 2020, 20, 5982–5990. [Google Scholar] [CrossRef]
  97. Piro, B. Electronic devices for biomarker monitoring. In The Detection of Biomarkers; Elsevier: Amsterdam, The Netherlands, 2022; pp. 183–207. [Google Scholar]
  98. Bhattacharyya, I.M.; Shalev, G. Electrostatically governed debye screening length at the solution-solid interface for biosensing applications. ACS Sens. 2019, 5, 154–161. [Google Scholar] [CrossRef]
  99. Tian, M.; Qiao, M.; Shen, C.; Meng, F.; Frank, L.A.; Krasitskaya, V.V.; Wang, T.; Zhang, X.; Song, R.; Li, Y. Highly-sensitive graphene field effect transistor biosensor using PNA and DNA probes for RNA detection. Appl. Surf. Sci. 2020, 527, 146839. [Google Scholar] [CrossRef]
  100. Yoo, H.; Jo, H.; Oh, S.S. Detection and beyond: Challenges and advances in aptamer-based biosensors. Mater. Adv. 2020, 1, 2663–2687. [Google Scholar] [CrossRef]
  101. An, J.; Park, H.; Kim, J.; Park, H.; Kim, T.-H.; Park, C.; Kim, J.; Lee, M.-H.; Lee, T. Extended-Gate Field-Effect Transistor Consisted of a CD9 Aptamer and MXene for Exosome Detection in Human Serum. ACS Sens. 2023, 8, 3174–3186. [Google Scholar] [CrossRef] [PubMed]
  102. Himori, S.; Nishitani, S.; Sakata, T. Aptamer-based nanofilter interface for small-biomarker detection with potentiometric biosensor. Electrochim. Acta 2021, 368, 137631. [Google Scholar] [CrossRef]
  103. Shin, S.; Kim, S.; Choi, W.; Do, J.; Son, J.; Kim, K.; Jang, S.; Lee, J.-S. Sensing Characteristics of SARS-CoV-2 Spike Protein Using Aptamer-Functionalized Si-Based Electrolyte-Gated Field-Effect Transistor (EGT). Biosensors 2024, 14, 124. [Google Scholar] [CrossRef]
  104. Tong, Y.; Fang, J.-Y.; Song, A.-H.; Deng, H.; Li, P.; Huang, Z.-H.; Ji, O.-Y.; Ge, X.-L.; Zhu, T.-Y.; Liu, Y.-L. Peritoneal dialysis effluent-derived exosomal miR-432-5p: An assessment tool for peritoneal dialysis efficacy. Ann. Transl. Med. 2022, 10, 242. [Google Scholar] [CrossRef]
  105. Keskin Gözmen, Ş.; Serdaroğlu, E. High C-reactive protein and number of previous episodes at diagnosis increase the risk of catheter removal in peritoneal dialysis-related peritonitis in children. Ther. Apher. Dial. 2023, 27, 328–334. [Google Scholar] [CrossRef]
  106. Lou, B.; Liu, Y.; Shi, M.; Chen, J.; Li, K.; Tan, Y.; Chen, L.; Wu, Y.; Wang, T.; Liu, X. Aptamer-based biosensors for virus protein detection. TrAC Trends Anal. Chem. 2022, 157, 116738. [Google Scholar] [CrossRef]
  107. Röthlisberger, P.; Hollenstein, M. Aptamer chemistry. Adv. Drug Deliv. Rev. 2018, 134, 3–21. [Google Scholar] [CrossRef]
  108. Dai, Y.; Cao, Y.; Köhler, J.; Lu, A.; Xu, S.; Wang, H. Unbiased RNA-Seq-driven identification and validation of reference genes for quantitative RT-PCR analyses of pooled cancer exosomes. BMC Genom. 2021, 22, 27. [Google Scholar] [CrossRef]
  109. Han, C.-H.; Jang, J. Integrated microfluidic platform with electrohydrodynamic focusing and a carbon-nanotube-based field-effect transistor immunosensor for continuous, selective, and label-free quantification of bacteria. Lab A Chip 2021, 21, 184–195. [Google Scholar] [CrossRef]
  110. Lee, K.; Shao, H.; Weissleder, R.; Lee, H. Acoustic purification of extracellular microvesicles. ACS Nano 2015, 9, 2321–2327. [Google Scholar] [CrossRef]
  111. Alharbi, M.A. Is Low Serum Albumin a Predictor Sign of the Incidence of Peritoneal Dialysis-Associated Peritonitis? A Quasi-systematic Review. Saudi J. Kidney Dis. Transplant. 2020, 31, 320–334. [Google Scholar] [CrossRef] [PubMed]
  112. Yahya, I.; Hassan, M.A.; Maidin, N.N.M.; Mohamed, M.A. Swcnt network-fet device for human serum albumin detection. Sensors 2022, 22, 8212. [Google Scholar] [CrossRef] [PubMed]
  113. Gao, Z.; Ducos, P.; Ye, H.; Zauberman, J.; Sriram, A.; Yang, X.; Wang, Z.; Mitchell, M.W.; Lekkas, D.; Brisson, D. Graphene transistor arrays functionalized with genetically engineered antibody fragments for Lyme disease diagnosis. 2D Mater. 2020, 7, 024001. [Google Scholar] [CrossRef]
  114. Zhang, J.; Sun, H.; Pei, W.; Jiang, H.; Chen, J. Nanobody-based immunosensing methods for safeguarding public health. J. Biomed. Res. 2021, 35, 318. [Google Scholar] [CrossRef]
  115. Filipiak, M.S.; Rother, M.; Andoy, N.M.; Knudsen, A.C.; Grimm, S.; Bachran, C.; Swee, L.K.; Zaumseil, J.; Tarasov, A. Highly sensitive, selective and label-free protein detection in physiological solutions using carbon nanotube transistors with nanobody receptors. Sens. Actuators B Chem. 2018, 255, 1507–1516. [Google Scholar] [CrossRef]
  116. Farahmandpour, M.; Kordrostami, Z.; Rajabzadeh, M.; Khalifeh, R. Flexible Bio-Electronic Hybrid Metal–Oxide Channel FET as a Glucose Sensor. IEEE Trans. NanoBiosci. 2023, 22, 855–862. [Google Scholar] [CrossRef]
  117. Wang, B.; Luo, Y.; Gao, L.; Liu, B.; Duan, G. High-performance field-effect transistor glucose biosensors based on bimetallic Ni/Cu metal-organic frameworks. Biosens. Bioelectron. 2021, 171, 112736. [Google Scholar] [CrossRef]
  118. Piccinini, E.; Fenoy, G.E.; Knoll, W.; Marmisollé, W.A.; Azzaroni, O. Polyelectrolyte-Enzyme Assemblies Integrated into Graphene Field-Effect Transistors for Biosensing Applications. In Graphene Field-Effect Transistors: Advanced Bioelectronic Devices for Sensing Applications; Wiley-VCH: Weinheim, Germany, 2023; pp. 285–299. [Google Scholar] [CrossRef]
  119. Liu, W.; Wang, X.; Dong, B.; Liu, Y.; Wei, D. Enzymatic cascade reactors on carbon nanotube transistor detecting trace prostate cancer biomarker. Biosens. Bioelectron. 2024, 263, 116603. [Google Scholar] [CrossRef]
  120. Wang, H.; Xie, J.; Xiao, M.; Ke, Y.; Li, J.; Nie, Z.; Chen, Q.; Zhang, Z. Spherical Nucleic Acid Probes on Floating-Gate Field-Effect Transistor Biosensors for Attomolar-Level Analyte Detection. ACS Nano 2024, 18, 34391–34402. [Google Scholar] [CrossRef]
  121. Halima, H.B.; Bellagambi, F.G.; Alcacer, A.; Pfeiffer, N.; Heuberger, A.; Hangouët, M.; Zine, N.; Bausells, J.; Elaissari, A.; Errachid, A. A silicon nitride ISFET based immunosensor for tumor necrosis factor-alpha detection in saliva. A promising tool for heart failure monitoring. Anal. Chim. Acta 2021, 1161, 338468. [Google Scholar] [CrossRef]
  122. Sinha, K.; Chakraborty, B.; Chaudhury, S.S.; Chaudhuri, C.R.; Chattopadhyay, S.; Mukhopadhyay, C.D. Selective, ultra-sensitive, and rapid detection of serotonin by optimized ZnO nanorod FET biosensor. IEEE Trans. NanoBiosci. 2021, 21, 65–74. [Google Scholar] [CrossRef] [PubMed]
  123. Dou, C.; Wu, Z.; Chen, W.; Yan, H.; Li, D.; You, X.-Q.; Chen, Y.-S.; Zhou, C.; Chen, S.; Zhuang, P. Au-functionalized wrinkle graphene biosensor for ultrasensitive detection of Interleukin-6. Carbon 2024, 216, 118556. [Google Scholar] [CrossRef]
  124. Hu, J.; Li, Y.; Zhang, X.; Wang, Y.; Zhang, J.; Yan, J.; Li, J.; Zhang, Z.; Yin, H.; Wei, Q. Ultrasensitive Silicon Nanowire Biosensor with Modulated Threshold Voltages and Ultra-Small Diameter for Early Kidney Failure Biomarker Cystatin C. Biosensors 2023, 13, 645. [Google Scholar] [CrossRef] [PubMed]
  125. Wei, S.; Dou, Y.; Li, T. Ultra-sensitive and label-free detection of Escherichia coli O157: H7 using graphene-based field effect transistor modified with heat-denatured casein. Microchem. J. 2023, 193, 109049. [Google Scholar] [CrossRef]
  126. Capua, L.; Sprunger, Y.; Elettro, H.; Grammoustianou, A.; Midahuen, R.; Ernst, T.; Barraud, S.; Gill, R.; Ionescu, A. Double-gate Si nanowire FET sensor arrays for label-free C-reactive protein detection enabled by antibodies fragments and pseudo-super-nernstian back-gate operation. In Proceedings of the 2021 IEEE International Electron Devices Meeting (IEDM), San Francisco, CA, USA, 11–16 December 2021; pp. 16.12.11–16.12.14. [Google Scholar] [CrossRef]
  127. He, J.; Cao, X.; Liu, H.; Liang, Y.; Chen, H.; Xiao, M.; Zhang, Z. Power and Sensitivity Management of Carbon Nanotube Transistor Glucose Biosensors. ACS Appl. Mater. Interfaces 2023, 16, 1351–1360. [Google Scholar] [CrossRef]
  128. Hao, Z.; Pan, Y.; Huang, C.; Wang, Z.; Lin, Q.; Zhao, X.; Liu, S. Modulating the linker immobilization density on aptameric graphene field effect transistors using an electric field. ACS Sens. 2020, 5, 2503–2513. [Google Scholar] [CrossRef]
  129. Wang, B.; Zhao, C.; Wang, Z.; Yang, K.-A.; Cheng, X.; Liu, W.; Yu, W.; Lin, S.; Zhao, Y.; Cheung, K.M. Wearable aptamer-field-effect transistor sensing system for noninvasive cortisol monitoring. Sci. Adv. 2022, 8, eabk0967. [Google Scholar] [CrossRef]
  130. Huang, C.; Hao, Z.; Qi, T.; Pan, Y.; Zhao, X. An integrated flexible and reusable graphene field effect transistor nanosensor for monitoring glucose. J. Mater. 2020, 6, 308–314. [Google Scholar] [CrossRef]
  131. Naorem, M.; Paily, R.P. Non-enzymatic glucose detection based on go/ag nanocomposite in sio 2 trench embedded field effect transistor. IEEE Sens. J. 2022, 22, 15614–15621. [Google Scholar] [CrossRef]
  132. Mukherjee, P.; Dutta, P.; Sinha, K.; Sen, S.; Shirke, T.; Ganguly, R.; Barui, A.; RoyChaudhuri, C. A direct MIP-interfaced FET sensor for sensitive, selective, and real time biomolecule detection in unamplified samples: Toward POC bioelectronic transduction. Appl. Phys. Lett. 2023, 123, 193701. [Google Scholar] [CrossRef]
  133. Yang, J.C.; Lim, S.J.; Cho, C.H.; Hazarika, D.; Park, J.P.; Park, J. Determination of tumor necrosis factor-α in serum using extended-gate field-effect transistor-based chemosensors with molecularly imprinted polymer-coated gold dendrites. Sens. Actuators B Chem. 2023, 390, 133982. [Google Scholar] [CrossRef]
  134. Layasree, S.; Deepak, A. Simulation and Comparison of Current Voltage Characteristics of Si and Ge based Bio Field Effect Transistor by Varying Oxide Thickness to Get Better Sensitivity. Rev. Geintec Gest. Inov. E Tecnol. 2021, 11, 1066–1083. [Google Scholar] [CrossRef]
  135. Khushaim, W.; Vijjapu, M.T.; Yuvaraja, S.; Mani, V.; Salama, K.N. Graphitic carbon nitride and IGZO Bio-FET for rapid diagnosis of myocardial infarction. Biosensors 2022, 12, 836. [Google Scholar] [CrossRef] [PubMed]
  136. Lee, S.-E.; Choi, Y.; Oh, Y.; Lee, D.; Kim, J.; Hong, S. Black Phosphorus-Based Reusable Biosensor Platforms for the Ultrasensitive Detection of Cortisol in Saliva. ACS Appl. Mater. Interfaces 2024, 16, 11305–11314. [Google Scholar] [CrossRef] [PubMed]
  137. Dalila, N.; Arshad, M.M.; Gopinath, S.C.; Nuzaihan, M.; Fathil, M. Molybdenum disulfide—Gold nanoparticle nanocomposite in field-effect transistor back-gate for enhanced C-reactive protein detection. Microchim. Acta 2020, 187, 588. [Google Scholar] [CrossRef]
  138. de Freitas Martins, E.; Pinotti, L.F.; de Carvalho Castro Silva, C.; Rocha, A.R. Addressing the theoretical and experimental aspects of low-dimensional-materials-based FET immunosensors: A review. Chemosensors 2021, 9, 162. [Google Scholar] [CrossRef]
  139. Choi, W.; Jin, B.; Shin, S.; Do, J.; Son, J.; Kim, K.; Lee, J.-S. Highly Sensitive Detection of Urea Using Si Electrolyte-Gated Transistor with Low Power Consumption. Biosensors 2023, 13, 565. [Google Scholar] [CrossRef]
  140. Ogurcovs, A.; Kadiwala, K.; Sledevskis, E.; Krasovska, M.; Plaksenkova, I.; Butanovs, E. Effect of DNA Aptamer Concentration on the Conductivity of a Water-Gated Al: ZnO Thin-Film Transistor-Based Biosensor. Sensors 2022, 22, 3408. [Google Scholar] [CrossRef]
  141. Gwyther, R.E.; Côté, S.; Lee, C.-S.; Miao, H.; Ramakrishnan, K.; Palma, M.; Dafydd Jones, D. Optimising CNT-FET biosensor design through modelling of biomolecular electrostatic gating and its application to β-lactamase detection. Nat. Commun. 2024, 15, 7482. [Google Scholar] [CrossRef]
  142. Meng, L.; Xin, N.; Hu, C.; Sabea, H.A.; Zhang, M.; Jiang, H.; Ji, Y.; Jia, C.; Yan, Z.; Zhang, Q. Dual-gated single-molecule field-effect transistors beyond Moore’s law. Nat. Commun. 2022, 13, 1410. [Google Scholar] [CrossRef]
  143. Dai, C.; Liu, Y.; Wei, D. Two-dimensional field-effect transistor sensors: The road toward commercialization. Chem. Rev. 2022, 122, 10319–10392. [Google Scholar] [CrossRef] [PubMed]
  144. Porwal, A.; Sahu, C.; Periasamy, C. Comparative Analog Analysis of Si, Ge and Si 0.7 Ge 0.3 Channel Based DG-JLFET. In Proceedings of the 2021 IEEE International Symposium on Smart Electronic Systems (iSES), Jaipur, India, 18–22 December 2021; pp. 59–63. [Google Scholar]
  145. Park, H.-Y.; Dugasani, S.R.; Kang, D.-H.; Yoo, G.; Kim, J.; Gnapareddy, B.; Jeon, J.; Kim, M.; Song, Y.J.; Lee, S. M-DNA/transition metal dichalcogenide hybrid structure-based bio-FET sensor with ultra-high sensitivity. Sci. Rep. 2016, 6, 35733. [Google Scholar] [CrossRef] [PubMed]
  146. Simionescu, O.-G.; Avram, A.; Adiaconiţă, B.; Preda, P.; Pârvulescu, C.; Năstase, F.; Chiriac, E.; Avram, M. Field-effect transistors based on single-layer graphene and graphene-derived materials. Micromachines 2023, 14, 1096. [Google Scholar] [CrossRef] [PubMed]
  147. Masurkar, N.; Varma, S.; Mohana Reddy Arava, L. Supported and suspended 2D material-based FET biosensors. Electrochem 2020, 1, 260–277. [Google Scholar] [CrossRef]
  148. Liu, Q.; Zhao, C.; Chen, M.; Liu, Y.; Zhao, Z.; Wu, F.; Li, Z.; Weiss, P.S.; Andrews, A.M.; Zhou, C. Flexible multiplexed In2O3 nanoribbon aptamer-field-effect transistors for biosensing. iScience 2020, 23, 101469. [Google Scholar] [CrossRef]
  149. Amen, M.T.; Pham, T.T.T.; Cheah, E.; Tran, D.P.; Thierry, B. Metal-oxide FET biosensor for point-of-care testing: Overview and perspective. Molecules 2022, 27, 7952. [Google Scholar] [CrossRef]
  150. Kumaresan, Y.; Ma, S.; Ozioko, O.; Dahiya, R. Soft capacitive pressure sensor with enhanced sensitivity assisted by ZnO NW interlayers and airgap. IEEE Sens. J. 2022, 22, 3974–3982. [Google Scholar] [CrossRef]
  151. Yamada, S.; Sato, T.; Toshiyoshi, H. A pressure sensitive ionic gel FET for tactile sensing. Appl. Phys. Lett. 2017, 110, 253501. [Google Scholar] [CrossRef]
  152. Ren, H.; Liang, K.; Li, D.; Zhao, M.; Li, F.; Wang, H.; Miao, X.; Zhou, T.; Wen, L.; Lu, Q. Interface Engineering of Metal-Oxide Field-Effect Transistors for Low-Drift pH Sensing. Adv. Mater. Interfaces 2021, 8, 2100314. [Google Scholar] [CrossRef]
  153. Ren, H.; Zhang, S.; Li, D.; Tang, Y.; Chen, Y.; Wang, Y.; Liu, G.; Li, F.; Liu, L.; Huang, Q. Wearable and Multiplexed Biosensors based on Oxide Field-Effect Transistors. Small Methods 2024, 8, 2400781. [Google Scholar] [CrossRef]
  154. Côté, S.; Bouilly, D.; Mousseau, N. The molecular origin of the electrostatic gating of single-molecule field-effect biosensors investigated by molecular dynamics simulations. Phys. Chem. Chem. Phys. 2022, 24, 4174–4186. [Google Scholar] [CrossRef] [PubMed]
  155. Chehelamirani, M.; da Silva, M.C.; Salahub, D.R. Electronic properties of carbon nanotubes complexed with a DNA nucleotide. Phys. Chem. Chem. Phys. 2017, 19, 7333–7342. [Google Scholar] [CrossRef] [PubMed]
  156. Tintelott, M.; Pachauri, V.; Ingebrandt, S.; Vu, X.T. Process variability in top-down fabrication of silicon nanowire-based biosensor arrays. Sensors 2021, 21, 5153. [Google Scholar] [CrossRef]
  157. Shaleen; Singh, S.; Kumar, P. Ultrasensitive label-free electrical detection of charged biomolecules using a metal–semiconductor–metal Schottky silicon nanowire biristor. J. Comput. Electron. 2022, 21, 86–93. [Google Scholar] [CrossRef]
  158. Alias, N.N.; Yaacob, K.A.; Cheong, K.Y. Fabrication of SiNWs-FET Nanostructure Via Atomic Force Microscopy Lithography. Solid State Phenom. 2020, 301, 103–110. [Google Scholar] [CrossRef]
  159. Yusoh, S.N.; Yaacob, K.A. Study on the Physical Properties of a SiNW Biosensor to the Sensitivity of DNA Detection. Materials 2021, 14, 5716. [Google Scholar] [CrossRef]
  160. Hossain, M.M.; Shabbir, B.; Wu, Y.; Yu, W.; Krishnamurthi, V.; Uddin, H.; Mahmood, N.; Walia, S.; Bao, Q.; Alan, T. Ultrasensitive WSe2 field-effect transistor-based biosensor for label-free detection of cancer in point-of-care applications. 2D Mater. 2021, 8, 045005. [Google Scholar] [CrossRef]
  161. Yang, M.; Perera, G.; Dhanabalan, S.S.; Mahasivam, S.; Dong, D.; Cheong, Y.Z.; Xu, C.; Elango, P.F.M.; Borkhatariya, S.; Sriram, S. Rapid Conductometric Sensing of Chronic Kidney Disease Biomarkers: Specific and Precise Detection of Creatinine and Cystatin C in Artificial Saliva. Adv. Sens. Res. 2024, 3, 2400042. [Google Scholar] [CrossRef]
  162. Moutaouakil, A.E.; Poovathy, S.; Belmoubarik, M.; Peng, W.K. Graphene-based biosensor for Viral Detection. arXiv 2020, arXiv:2006.11881. [Google Scholar]
  163. Szunerits, S.; Rodrigues, T.; Bagale, R.; Happy, H.; Boukherroub, R.; Knoll, W. Graphene-Based Field-Eff. Transistors Biosensing: Where Is Field Head? Anal. Bioanal. Chem. 2024, 416, 2137–2150. [Google Scholar] [CrossRef]
  164. Bahri, M.; Shi, B.; Djebbi, K.; Elaguech, M.; Zhou, D.; Ali, M.B.; Tlili, C.; Wang, D. Toward clean and crackless polymer-assisted transfer of CVD-grown graphene and its recent advances in GFET-based biosensors. Mater. Today Chem. 2021, 22, 100578. [Google Scholar] [CrossRef]
  165. Salehirozveh, M.; Dehghani, P.; Zimmermann, M.; Roy, V.A.; Heidari, H. Graphene field effect transistor biosensors based on aptamer for amyloid-β detection. IEEE Sens. J. 2020, 20, 12488–12494. [Google Scholar] [CrossRef]
  166. Kong, X.; Bahri, M.; Djebbi, K.; Shi, B.; Zhou, D.; Tlili, C.; Wang, D. Graphene-based liquid gated field-effect transistor for label-free detection of DNA hybridization. In Proceedings of the 2021 18th International Multi-Conference on Systems, Signals & Devices (SSD), Monastir, Tunisia, 22–25 March 2021; pp. 387–391. [Google Scholar]
  167. Krishnan, S.K.; Nataraj, N.; Meyyappan, M.; Pal, U. Graphene-based field-effect transistors in biosensing and neural interfacing applications: Recent advances and prospects. Anal. Chem. 2023, 95, 2590–2622. [Google Scholar] [CrossRef] [PubMed]
  168. Vieira, N.C.; Ribeiro, B.C.; Blasques, R.V.; Janegitz, B.C.; dos Santos, F.A.; Zucolotto, V. Solution-Gated Reduced Graphene Oxide FETs: Device Fabrication and Biosensors Applications. In Graphene Field-Effect Transistors: Advanced Bioelectronic Devices for Sensing Applications; Wiley-VCH: Weinheim, Germany, 2023; pp. 43–70. [Google Scholar]
  169. Zhang, Y.; Ding, Y.; Li, C.; Xu, H.; Liu, C.; Wang, J.; Ma, Y.; Ren, J.; Zhao, Y.; Yue, W. An optic-fiber graphene field effect transistor biosensor for the detection of single-stranded DNA. Anal. Methods 2021, 13, 1839–1846. [Google Scholar] [CrossRef] [PubMed]
  170. Kwon, S.S.; Shin, J.H.; Choi, J.; Nam, S.; Park, W.I. Defect-mediated molecular interaction and charge transfer in graphene mesh–glucose Sensors. ACS Appl. Mater. Interfaces 2017, 9, 14216–14221. [Google Scholar] [CrossRef]
  171. Zhang, Y.; Feng, D.; Xu, Y.; Yin, Z.; Dou, W.; Habiba, U.E.; Pan, C.; Zhang, Z.; Mou, H.; Deng, H. DNA-based functionalization of two-dimensional MoS2 FET biosensor for ultrasensitive detection of PSA. Appl. Surf. Sci. 2021, 548, 149169. [Google Scholar] [CrossRef]
  172. Park, H.; Baek, S.; Sen, A.; Jung, B.; Shim, J.; Park, Y.C.; Lee, L.P.; Kim, Y.J.; Kim, S. Ultrasensitive and selective field-effect transistor-based biosensor created by rings of MoS2 nanopores. ACS Nano 2021, 16, 1826–1835. [Google Scholar] [CrossRef]
  173. Thriveni, G.; Ghosh, K. Covalent functionalization in graphene nanoribbon: Theoretical modeling and sensitivity analysis. J. Appl. Phys. 2021, 129, 114301. [Google Scholar] [CrossRef]
  174. Sen, A.; Shim, J.; Bala, A.; Park, H.; Kim, S. Boosting Sensitivity and Reliability in Field-Effect Transistor-Based Biosensors with Nanoporous MoS2 Encapsulated by Non-Planar Al2O3. Adv. Funct. Mater. 2023, 33, 2301919. [Google Scholar] [CrossRef]
  175. Liu, Y.; Zhang, Q.; Li, J.; Wu, C.; Cao, L.; Luo, Y.; Zhang, Z.; Wei, S.; Wei, Q.; Yao, J. Investigation of Vertically Stacked Horizontal Gate-All-Around SI Nanosheet Ion Sensitive Field Effect Transistor For Detection of C-Reactive Protein. In Proceedings of the 2023 China Semiconductor Technology International Conference (CSTIC), Shanghai, China, 26–27 June 2023; pp. 1–3. [Google Scholar]
  176. Sun, Y.; Peng, Z.; Li, H.; Wang, Z.; Mu, Y.; Zhang, G.; Chen, S.; Liu, S.; Wang, G.; Liu, C. Suspended CNT-Based FET sensor for ultrasensitive and label-free detection of DNA hybridization. Biosens. Bioelectron. 2019, 137, 255–262. [Google Scholar] [CrossRef]
  177. Xia, Y.; Sun, Y.; Li, H.; Chen, S.; Zhu, T.; Wang, G.; Man, B.; Pan, J.; Yang, C. Plasma treated graphene FET sensor for the DNA hybridization detection. Talanta 2021, 223, 121766. [Google Scholar] [CrossRef] [PubMed]
  178. Huang, C.; Hao, Z.; Wang, Z.; Zhao, X.; Wang, H.; Li, F.; Liu, S.; Pan, Y. A fully integrated graphene-polymer field-effect transistor biosensing device for on-site detection of glucose in human urine. Mater. Today Chem. 2022, 23, 100635. [Google Scholar] [CrossRef]
  179. Cai, B.; Xia, Z.; Wang, J.; Wu, S.; Jin, X. Reduced graphene oxide-based field effect transistor biosensors for high-sensitivity mirna21 detection. ACS Appl. Nano Mater. 2022, 5, 12035–12044. [Google Scholar] [CrossRef]
  180. Sebastian, A.; Pendurthi, R.; Choudhury, T.H.; Redwing, J.M.; Das, S. Benchmarking monolayer MoS2 and WS2 field-effect transistors. Nat. Commun. 2021, 12, 693. [Google Scholar] [CrossRef]
  181. Boschetto, G.; Carapezzi, S.; Todri-Sanial, A. Supported Pt Nanoclusters on Single-Layer MoS2 for the Detection of Cortisol: From Atomistic Scale to Device Modeling. ACS Appl. Electron. Mater. 2023, 5, 2977–2987. [Google Scholar] [CrossRef]
  182. Syu, Y.-C.; Hsu, W.-E.; Lin, C.-T. Field-effect transistor biosensing: Devices and clinical applications. ECS J. Solid State Sci. Technol. 2018, 7, Q3196. [Google Scholar] [CrossRef]
  183. Béraud, A.; Sauvage, M.; Bazán, C.M.; Tie, M.; Bencherif, A.; Bouilly, D. Graphene field-effect transistors as bioanalytical sensors: Design, operation and performance. Analyst 2021, 146, 403–428. [Google Scholar] [CrossRef]
  184. Ju, W.; Lee, S. Al back-gated graphene field-effect transistors for capacitive sensing applications based on quantum capacitance effect. AIP Adv. 2022, 12, 095210. [Google Scholar] [CrossRef]
  185. Sarcina, L.; Macchia, E.; Tricase, A.; Scandurra, C.; Imbriano, A.; Torricelli, F.; Cioffi, N.; Torsi, L.; Bollella, P. Enzyme based field effect transistor: State-of-the-art and future perspectives. Electrochem. Sci. Adv. 2023, 3, e2100216. [Google Scholar] [CrossRef]
  186. Sheibani, S.; Capua, L.; Kamaei, S.; Akbari, S.S.A.; Zhang, J.; Guerin, H.; Ionescu, A.M. Extended gate field-effect-transistor for sensing cortisol stress hormone. Commun. Mater. 2021, 2, 10. [Google Scholar] [CrossRef]
  187. Hao, Z.; Pan, Y.; Shao, W.; Lin, Q.; Zhao, X. Graphene-based fully integrated portable nanosensing system for on-line detection of cytokine biomarkers in saliva. Biosens. Bioelectron. 2019, 134, 16–23. [Google Scholar] [CrossRef] [PubMed]
  188. Vasilijević, S.; Boukraa, R.; Battaglini, N.; Piro, B. Graphene-based materials and their applications in electrolyte-gated transistors for sensing. Synth. Met. 2023, 295, 117355. [Google Scholar] [CrossRef]
  189. Islam, A.E.; Martineau, R.; Crasto, C.M.; Kim, H.; Rao, R.S.; Maruyama, B.; Kim, S.S.; Drummy, L.F. Graphene-based electrolyte-gated field-effect transistors for potentiometrically sensing neuropeptide Y in physiologically relevant environments. ACS Appl. Nano Mater. 2020, 3, 5088–5097. [Google Scholar] [CrossRef]
  190. Sant, W.; Pourciel-Gouzy, M.; Launay, J.; Do Conto, T.; Colin, R.; Martinez, A.; Temple-Boyer, P. Development of a creatinine-sensitive sensor for medical analysis. Sens. Actuators B Chem. 2004, 103, 260–264. [Google Scholar] [CrossRef]
  191. Aufricht, C.; Beelen, R.; Eberl, M.; Fischbach, M.; Fraser, D.; Jörres, A.; Kratochwill, K.; LópezCabrera, M.; Rutherford, P.; Schmitt, C.-P. Biomarker research to improve clinical outcomes of peritoneal dialysis: Consensus of the European Training and Research in Peritoneal Dialysis (EuTRiPD) network. Kidney Int. 2017, 92, 824–835. [Google Scholar] [CrossRef]
  192. Faria, B.; Gaya da Costa, M.; Lima, C.; Willems, L.; Brandwijk, R.; Berger, S.P.; Daha, M.R.; Pestana, M.; Seelen, M.A.; Poppelaars, F. Soluble CD59 in peritoneal dialysis: A potential biomarker for peritoneal membrane function. J. Nephrol. 2021, 34, 801–810. [Google Scholar] [CrossRef]
  193. Rago, C.; Lombardi, T.; Di Fulvio, G.; Di Liberato, L.; Arduini, A.; Divino-Filho, J.C.; Bonomini, M. A new peritoneal dialysis solution containing L-carnitine and xylitol for patients on continuous ambulatory peritoneal dialysis: First clinical experience. Toxins 2021, 13, 174. [Google Scholar] [CrossRef]
  194. Roumeliotis, S.; Dounousi, E.; Salmas, M.; Eleftheriadis, T.; Liakopoulos, V. Unfavorable effects of peritoneal dialysis solutions on the peritoneal membrane: The role of oxidative stress. Biomolecules 2020, 10, 768. [Google Scholar] [CrossRef]
  195. Zhao, H.; Zhang, H.-L.; Jia, L. High glucose dialysate-induced peritoneal fibrosis: Pathophysiology, underlying mechanisms and potential therapeutic strategies. Biomed. Pharmacother. 2023, 165, 115246. [Google Scholar] [CrossRef]
  196. Sitter, T.; Sauter, M. Impact of glucose in peritoneal dialysis: Saint or sinner? Perit. Dial. Int. 2005, 25, 415–425. [Google Scholar] [CrossRef]
  197. Ng, J.K.; Ling, J.; Luk, A.O.; Lau, E.S.; Ma, R.C.; Li, P.K.; Szeto, C.C.; Chan, J.C.; Chow, E. Evaluation of a fourth-generation subcutaneous real-time continuous glucose monitor (CGM) in individuals with diabetes on peritoneal dialysis. Diabetes Care 2023, 46, 1191–1195. [Google Scholar] [CrossRef] [PubMed]
  198. Hamdan, Z.; Abdel-Hafez, Y.; Enaya, A.; Sarsour, A.; Kharraz, L.; Nazzal, Z. Dapagliflozin in peritoneal dialysis patients: A pilot study evaluating peritoneal membrane function. BMC Nephrol. 2024, 25, 37. [Google Scholar] [CrossRef] [PubMed]
  199. Zeggai, O.; Belarbi, M.; Ouledabbes, A.; Mouloudj, H. Modeling of a micro-biological sensor field effect for the enzymatic detection of glucose. Int. J. Mod. Phys. B 2019, 33, 1950289. [Google Scholar] [CrossRef]
  200. Ba, M.; Diallo, A.K.; Launay, J.; Temple-Boyer, P. Numerical Modeling of Glucose Biosensor With pH-Based Electrochemical Field-Effect Transistor Device. IEEE Trans. Electron Devices 2020, 67, 1787–1792. [Google Scholar] [CrossRef]
  201. Ma, M.; Zhou, Y.; Li, J.; Ge, Z.; He, H.; Tao, T.; Cai, Z.; Wang, X.; Chang, G.; He, Y. Non-invasive detection of glucose via a solution-gated graphene transistor. Analyst 2020, 145, 887–896. [Google Scholar] [CrossRef]
  202. Ravariu, C.; Parvulescu, C.C.; Manea, E.; Tucureanu, V. Optimized Technologies for Cointegration of MOS Transistor and Glucose Oxidase Enzyme on a Si-Wafer. Biosensors 2021, 11, 497. [Google Scholar] [CrossRef]
  203. Moakhar, R.S.; Mirzaei, M.; Flynn, S.E.; Jalali, M.; Sanati, A.; Mahshid, S. Photoelectrochemical sensing of titanium oxide nanostructures for the detection of glucose: Fabrication methods and signal enhancement strategies. Microchem. J. 2024, 201, 110528. [Google Scholar] [CrossRef]
  204. Fumagall, G.; Panichi, V. Biocompatibility of the dialysis system. In Critical Care Nephrology; Elsevier: Amsterdam, The Netherlands, 2019; pp. 918–922.e912. [Google Scholar]
  205. Kajisa, T.; Sakata, T. Glucose-responsive hydrogel electrode for biocompatible glucose transistor. Sci. Technol. Adv. Mater. 2017, 18, 26–33. [Google Scholar] [CrossRef]
  206. Tonomura, S.; Uchiyama, K.; Nakayama, T.; Mitsuno, R.; Kojima, D.; Hama, E.Y.; Nagasaka, T.; Nishimura, E.S.; Kusahana, E.; Takahashi, R. Clinical significance of serum urea-to-creatinine ratio in patients undergoing peritoneal dialysis. Ther. Apher. Dial. 2023, 27, 1103–1112. [Google Scholar] [CrossRef]
  207. Cánovas, R.; Cuartero, M.; Crespo, G.A. Modern creatinine (Bio) sensing: Challenges of point-of-care platforms. Biosens. Bioelectron. 2019, 130, 110–124. [Google Scholar] [CrossRef]
  208. Brookes, E.M.; Power, D.A. Elevated serum urea-to-creatinine ratio is associated with adverse inpatient clinical outcomes in non-end stage chronic kidney disease. Sci. Rep. 2022, 12, 20827. [Google Scholar] [CrossRef] [PubMed]
  209. van der Slikke, E.C.; Star, B.S.; de Jager, V.D.; Leferink, M.B.; Klein, L.M.; Quinten, V.M.; Olgers, T.J.; Ter Maaten, J.C.; Bouma, H.R. A high urea-to-creatinine ratio predicts long-term mortality independent of acute kidney injury among patients hospitalized with an infection. Sci. Rep. 2020, 10, 15649. [Google Scholar] [CrossRef] [PubMed]
  210. Davies, S.J.; Finkelstein, F.O. Accuracy of the estimation of V and the implications this has when applying K t/V urea for measuring dialysis dose in peritoneal dialysis. Perit. Dial. Int. 2020, 40, 261–269. [Google Scholar] [CrossRef]
  211. Cullis, B.; Al-Hwiesh, A.; Kilonzo, K.; McCulloch, M.; Niang, A.; Nourse, P.; Parapiboon, W.; Ponce, D.; Finkelstein, F.O. ISPD guidelines for peritoneal dialysis in acute kidney injury: 2020 update (adults). Perit. Dial. Int. 2021, 41, 15–31. [Google Scholar] [CrossRef]
  212. Doğan, B.; Elik, A.; Altunay, N. Determination of paracetamol in synthetic urea and pharmaceutical samples by shaker-assisted deep eutectic solvent microextraction and spectrophotometry. Microchem. J. 2020, 154, 104645. [Google Scholar] [CrossRef]
  213. Castellarnau, A.; Werner, M.; Günthner, R.; Jakob, M. Real-time Kt/V determination by ultraviolet absorbance in spent dialysate: Technique validation. Kidney Int. 2010, 78, 920–925. [Google Scholar] [CrossRef]
  214. Silva, G.; Mulato, M. Urea detection using commercial field effect transistors. ECS J. Solid State Sci. Technol. 2018, 7, Q3014. [Google Scholar] [CrossRef]
  215. Lin, Y.-H.; Chu, C.-P.; Lin, C.-F.; Liao, H.-H.; Tsai, H.-H.; Juang, Y.-Z. Extended-gate field-effect transistor packed in micro channel for glucose, urea and protein biomarker detection. Biomed. Microdevices 2015, 17, 111. [Google Scholar] [CrossRef]
  216. Park, M.; Kim, J.; Kim, K.; Pyun, J.-C.; Sung, G.Y. Parylene-coated polytetrafluoroethylene-membrane-based portable urea sensor for real-time monitoring of urea in peritoneal dialysate. Sensors 2019, 19, 4560. [Google Scholar] [CrossRef]
  217. Singh, S.; Sharma, M.; Singh, G. Recent advancements in urea biosensors for biomedical applications. IET Nanobiotechnol. 2021, 15, 358–379. [Google Scholar] [CrossRef]
  218. Pijanowska, D.G.; Torbicz, W. pH-ISFET based urea biosensor. Sens. Actuators B Chem. 1997, 44, 370–376. [Google Scholar] [CrossRef]
  219. Cho, S.-K.; Cho, W.-J. Highly sensitive and transparent urea-EnFET based point-of-care diagnostic test sensor with a triple-gate a-IGZO TFT. Sensors 2021, 21, 4748. [Google Scholar] [CrossRef] [PubMed]
  220. Kashani, K.; Rosner, M.H.; Ostermann, M. Creatinine: From physiology to clinical application. Eur. J. Intern. Med. 2020, 72, 9–14. [Google Scholar] [CrossRef] [PubMed]
  221. Dissanayake, R.K.; Ranaweera, K.K.P.T.; Dias, P.; Priyadarshani, A. The effect of bilirubin on laboratory investigations on serum creatinine: A comparison study between jaffe reaction and creatinase enzymatic method with creatinine in phosphate buffered saline solution and serum. Clin. Exp. Health Sci. 2022, 12, 11–17. [Google Scholar] [CrossRef]
  222. Boobphahom, S.; Ruecha, N.; Rodthongkum, N.; Chailapakul, O.; Remcho, V.T. A copper oxide-ionic liquid/reduced graphene oxide composite sensor enabled by digital dispensing: Non-enzymatic paper-based microfluidic determination of creatinine in human blood serum. Anal. Chim. Acta 2019, 1083, 110–118. [Google Scholar] [CrossRef]
  223. Narimani, R.; Esmaeili, M.; Rasta, S.H.; Khosroshahi, H.T.; Mobed, A. Trend in creatinine determining methods: Conventional methods to molecular-based methods. Anal. Sci. Adv. 2021, 2, 308–325. [Google Scholar] [CrossRef]
  224. Saddique, Z.; Faheem, M.; Habib, A.; UlHasan, I.; Mujahid, A.; Afzal, A. Electrochemical Creatinine (Bio) Sensors for Point-of-Care Diagnosis of Renal Malfunction and Chronic Kidney Disorders. Diagnostics 2023, 13, 1737. [Google Scholar] [CrossRef]
  225. Gonzalez-Gallardo, C.L.; Arjona, N.; Álvarez-Contreras, L.; Guerra-Balcázar, M. Electrochemical creatinine detection for advanced point-of-care sensing devices: A review. RSC Adv. 2022, 12, 30785–30802. [Google Scholar] [CrossRef]
  226. Temple-Boyer, P.; Benyahia, A.; Sant, W.; Pourciel-Gouzy, M.; Launay, J.; Martinez, A. Modelling of urea-EnFETs for haemodialysis applications. Sens. Actuators B Chem. 2008, 131, 525–532. [Google Scholar] [CrossRef]
  227. Cassedy, A.; Parle-McDermott, A.; O’Kennedy, R. Virus detection: A review of the current and emerging molecular and immunological methods. Front. Mol. Biosci. 2021, 8, 637559. [Google Scholar] [CrossRef]
  228. Hazarika, K.; Thakur, H.R.; Dutta, J.C. Fabrication and characterization of different polymer doped CNT nanocomposites for creatinine detection. Mater. Today Proc. 2023, in press. [CrossRef]
  229. Wen, T.; Zhu, W.; Xue, C.; Wu, J.; Han, Q.; Wang, X.; Zhou, X.; Jiang, H. Novel electrochemical sensing platform based on magnetic field-induced self-assembly of Fe3O4@ Polyaniline nanoparticles for clinical detection of creatinine. Biosens. Bioelectron. 2014, 56, 180–185. [Google Scholar] [CrossRef] [PubMed]
  230. Ozansoy Kasap, B.; Marchenko, S.V.; Soldatkin, O.O.; Dzyadevych, S.V.; Akata Kurc, B. Biosensors based on nano-gold/zeolite-modified ion selective field-effect transistors for creatinine detection. Nanoscale Res. Lett. 2017, 12, 162. [Google Scholar] [CrossRef] [PubMed]
  231. Marchenko, S.V.; Soldatkin, O.O.; Kasap, B.O.; Kurc, B.A.; Soldatkin, A.P.; Dzyadevych, S.V. Creatinine deiminase adsorption onto silicalite-modified pH-FET for creation of new creatinine-sensitive biosensor. Nanoscale Res. Lett. 2016, 11, 173. [Google Scholar] [CrossRef]
  232. Kumar, R.R.; Shaikh, M.O.; Chuang, C.-H. A review of recent advances in non-enzymatic electrochemical creatinine biosensing. Anal. Chim. Acta 2021, 1183, 338748. [Google Scholar] [CrossRef]
  233. Tsai, H.-H.; Lin, C.-F.; Juang, Y.-Z.; Wang, I.-L.; Lin, Y.-C.; Wang, R.-L.; Lin, H.-Y. Multiple type biosensors fabricated using the CMOS BioMEMS platform. Sens. Actuators B Chem. 2010, 144, 407–412. [Google Scholar] [CrossRef]
  234. Nashruddin, S.N.A.B.M.; Salleh, F.H.M.; Raub, A.A.M. Early detection of kidney problem through voltammetry, potentiometry, amperometry, and impedance electrochemical techniques: A comprehensive review. Measurement 2024, 230, 114475. [Google Scholar] [CrossRef]
  235. Morisi, N.; Virzì, G.M.; Ferrarini, M.; Alfano, G.; Zanella, M.; Ronco, C.; Donati, G. Exploring the Role of Cell-Free Nucleic Acids and Peritoneal Dialysis: A Narrative Review. Genes 2024, 15, 553. [Google Scholar] [CrossRef]
  236. Lousa, I.; Reis, F.; Beirão, I.; Alves, R.; Belo, L.; Santos-Silva, A. New potential biomarkers for chronic kidney disease management—A review of the literature. Int. J. Mol. Sci. 2020, 22, 43. [Google Scholar] [CrossRef]
  237. Corciulo, S.; Nicoletti, M.C.; Mastrofrancesco, L.; Milano, S.; Mastrodonato, M.; Carmosino, M.; Gerbino, A.; Corciulo, R.; Russo, R.; Svelto, M. AQP1-containing exosomes in peritoneal dialysis effluent as biomarker of dialysis efficiency. Cells 2019, 8, 330. [Google Scholar] [CrossRef]
  238. Araújo, J.; Jorge, S.; e Costa, F.T.; Ramos, A.; Lodeiro, C.; Santos, H.; Capelo, J. A cost-effective method to get insight into the peritoneal dialysate effluent proteome. J. Proteom. 2016, 145, 207–213. [Google Scholar] [CrossRef] [PubMed]
  239. Barreto, D.L.; Krediet, R.T. Current status and practical use of effluent biomarkers in peritoneal dialysis patients. Am. J. Kidney Dis. 2013, 62, 823–833. [Google Scholar] [CrossRef] [PubMed]
  240. van Gelder, M.K.; Jong, J.A.; Folkertsma, L.; Guo, Y.; Blüchel, C.; Verhaar, M.C.; Odijk, M.; Van Nostrum, C.F.; Hennink, W.E.; Gerritsen, K.G. Urea removal strategies for dialysate regeneration in a wearable artificial kidney. Biomaterials 2020, 234, 119735. [Google Scholar] [CrossRef] [PubMed]
  241. Maharjan, S.R.S.; Davenport, A. Comparison of sodium removal in peritoneal dialysis patients treated by continuous ambulatory and automated peritoneal dialysis. J. Nephrol. 2019, 32, 1011–1019. [Google Scholar] [CrossRef]
  242. Du, Y.; Zong, M.; Guan, Q.; Huang, Z.; Zhou, L.; Cai, J.; da Roza, G.; Wang, H.; Qi, H.; Lu, Y. Comparison of mesenchymal stromal cells from peritoneal dialysis effluent with those from umbilical cords: Characteristics and therapeutic effects on chronic peritoneal dialysis in uremic rats. Stem Cell Res. Ther. 2021, 12, 398. [Google Scholar] [CrossRef]
  243. Krediet, R.T. Acquired decline in ultrafiltration in peritoneal dialysis: The role of glucose. J. Am. Soc. Nephrol. 2021, 32, 2408–2415. [Google Scholar] [CrossRef]
  244. Brito, C.; Esteves, M.; Peixoto, H.; Abelha, A.; Machado, J. A data mining approach to classify serum creatinine values in patients undergoing continuous ambulatory peritoneal dialysis. Wirel. Netw. 2022, 28, 1269–1277. [Google Scholar] [CrossRef]
  245. Goodlad, C.; George, S.; Sandoval, S.; Mepham, S.; Parekh, G.; Eberl, M.; Topley, N.; Davenport, A. Measurement of innate immune response biomarkers in peritoneal dialysis effluent using a rapid diagnostic point-of-care device as a diagnostic indicator of peritonitis. Kidney Int. 2020, 97, 1253–1259. [Google Scholar] [CrossRef]
  246. Idei, M.; Abe, M.; Tanaka, M.; Nakata, J.; Isshiki, M.; Hino, O.; Miida, T. Effluent N-terminal expressed in renal cell carcinoma/mesothelin predicts increased peritoneal permeability in patients undergoing peritoneal dialysis. Ther. Apher. Dial. 2022, 26, 1014–1022. [Google Scholar] [CrossRef]
  247. Martins, A.C.; Pimenta, G.F.; Calça, R.; Matias, P. CA-125 Is a Good Biomarker of Overhydration in Peritoneal Dialysis Patients: SA-PO672. J. Am. Soc. Nephrol. 2023, 34, 915. [Google Scholar] [CrossRef]
  248. Carreras-Planella, L.; Soler-Majoral, J.; Rubio-Esteve, C.; Morón-Font, M.; Franquesa, M.; Bonal, J.; Troya-Saborido, M.I.; Borràs, F.E. Proteomic profiling of peritoneal dialysis effluent-derived extracellular vesicles: A longitudinal study. J. Nephrol. 2019, 32, 1021–1031. [Google Scholar] [CrossRef] [PubMed]
  249. Yan, H.; Ma, D.; Yang, S.; Ni, Z.; Fang, W. Effluent lipopolysaccharide is a prompt marker of peritoneal dialysis-related gram-negative peritonitis. Perit. Dial. Int. 2020, 40, 455–461. [Google Scholar] [CrossRef] [PubMed]
  250. Capua, L.; Sprunger, Y.; Elettro, H.; Risch, F.; Grammoustianou, A.; Midahuen, R.; Ernst, T.; Barraud, S.; Gill, R.; Ionescu, A. Label-free C-reactive protein Si nanowire FET sensor arrays with super-Nernstian back-gate operation. IEEE Trans. Electron. Devices 2022, 69, 2159–2165. [Google Scholar] [CrossRef]
  251. Than, W.H.; Ng, J.K.-C.; Fung, W.W.-S.; Chan, G.C.-K.; Lai, K.-B.; Luk, C.C.-W.; Cheng, P.M.-S.; Chow, K.-M.; Szeto, C.-C. Prognostic significance of peritoneal dialysis effluent mitochondrial DNA level. Clin. Chim. Acta 2021, 519, 1–9. [Google Scholar] [CrossRef]
  252. Zhang, S.; Li, H.; Cai, X.; Zhao, C.; Cao, J. The association between serum cystatin C and residual renal function in peritoneal dialysis patients. Ther. Apher. Dial. 2022, 26, 1241–1246. [Google Scholar] [CrossRef]
  253. Fernando, S.; Polkinghorne, K.R. Cystatin C: Not just a marker of kidney function. Braz. J. Nephrol. 2020, 42, 6–7. [Google Scholar] [CrossRef]
  254. Davies, S.J. Peritoneal solute transport and inflammation. Am. J. Kidney Dis. 2014, 64, 978–986. [Google Scholar] [CrossRef]
  255. Chen, H.; Choo, T.K.; Huang, J.; Wang, Y.; Liu, Y.; Platt, M.; Palaniappan, A.; Liedberg, B.; Tok, A.I.Y. Label-free electronic detection of interleukin-6 using horizontally aligned carbon nanotubes. Mater. Des. 2016, 90, 852–857. [Google Scholar] [CrossRef]
  256. Mukherjee, P.; Sen, S.; Das, A.; RoyChaudhuri, C. Real Time Estimation of Corneal Damage Using IL-6 in Tears based on Flexi-Graphene FET Sensor Array. IEEE Sens. Lett. 2024, 8, 4503104. [Google Scholar] [CrossRef]
  257. Li, D.; Ren, Y.; Chen, R.; Wu, H.; Zhuang, S.; Zhang, M. Label-free MXene-assisted field effect transistor for the determination of IL-6 in patients with kidney transplantation infection. Microchim. Acta 2023, 190, 284. [Google Scholar] [CrossRef]
  258. Yung, S.; Lui, S.L.; Ng, C.K.; Yim, A.; Ma, M.K.; Lo, K.Y.; Chow, C.C.; Chu, K.H.; Chak, W.L.; Lam, M.F. Impact of a low-glucose peritoneal dialysis regimen on fibrosis and inflammation biomarkers. Perit. Dial. Int. 2015, 35, 147–158. [Google Scholar] [CrossRef] [PubMed]
  259. Redahan, L.; Davenport, A. Peritoneal dialysate effluent and serum CA125 concentrations in stable peritoneal dialysis patients. J. Nephrol. 2016, 29, 427–434. [Google Scholar] [CrossRef] [PubMed]
  260. Danford, C.J.; Lin, S.C.; Smith, M.P.; Wolf, J.L. Encapsulating peritoneal sclerosis. World J. Gastroenterol. 2018, 24, 3101. [Google Scholar] [CrossRef]
  261. Majd, S.M.; Salimi, A. Ultrasensitive flexible FET-type aptasensor for CA 125 cancer marker detection based on carboxylated multiwalled carbon nanotubes immobilized onto reduced graphene oxide film. Anal. Chim. Acta 2018, 1000, 273–282. [Google Scholar] [CrossRef]
  262. Ji, H.; Wang, Z.; Wang, S.; Wang, C.; Zhang, K.; Zhang, Y.; Han, L. Highly stable InSe-FET biosensor for ultra-sensitive detection of breast cancer biomarker CA125. Biosensors 2023, 13, 193. [Google Scholar] [CrossRef]
  263. Panorchan, K.; Davenport, A. Diagnostic and prognostic role of peritoneal CA 125 in peritoneal dialysis patients presenting with acute peritonitis. BMC Nephrol. 2014, 15, 149. [Google Scholar] [CrossRef]
  264. Wijayaratne, D.; Muthuppalaniappan, V.M.; Davenport, A. Serum CA125 a potential marker of volume status for peritoneal dialysis patients? Int. J. Artif. Organs 2021, 44, 1029–1033. [Google Scholar] [CrossRef]
  265. Oliveira, W.V.d.; Turani, S.D.; Marinho, M.A.S.; Pinto, S.W.L.; Otoni, A.; Figueiredo, R.C.; Rios, D.R.A. CA-125 and CCL2 may indicate inflammation in peritoneal dialysis patients. Braz. J. Nephrol. 2021, 43, 502–509. [Google Scholar] [CrossRef]
  266. Wang, Z.; Hao, Z.; Zhao, X.; Lin, Q. A Mechanically Flexible Aptamer-Based Graphene Nanosensor for Biomarker Monitoring. In Proceedings of the 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems & Eurosensors XXXIII (TRANSDUCERS & EUROSENSORS XXXIII), Berlin, Germany, 23–27 June 2019; pp. 2428–2431. [Google Scholar]
  267. Samota, S.; Rani, R.; Chakraverty, S.; Kaushik, A. Biosensors for simplistic detection of pathogenic bacteria: A review with special focus on field-effect transistors. Mater. Sci. Semicond. Process. 2022, 141, 106404. [Google Scholar] [CrossRef]
  268. Kim, K.H.; Park, S.J.; Park, C.S.; Seo, S.E.; Lee, J.; Kim, J.; Lee, S.H.; Lee, S.; Kim, J.-S.; Ryu, C.-M. High-performance portable graphene field-effect transistor device for detecting Gram-positive and-negative bacteria. Biosens. Bioelectron. 2020, 167, 112514. [Google Scholar] [CrossRef]
  269. Purwidyantri, A.; Kamajaya, L.; Chen, C.-H.; Luo, J.-D.; Chiou, C.-C.; Tian, Y.-C.; Lin, C.-Y.; Yang, C.-M.; Lai, C.-S. A colloidal nanopatterning and downscaling of a highly periodic au nanoporous EGFET biosensor. J. Electrochem. Soc. 2018, 165, H3170. [Google Scholar] [CrossRef]
  270. Wang, H.; Chen, R.; He, Y.; Zhu, X.; Yu, Z.; Feng, Z.; Pan, D.; Yang, L.; Tang, X.; Xiong, B. Controllable self-cleaning FET self-assembled RNA-cleaving DNAzyme based DNA nanotree for culture-free Staphylococcus aureus detection. J. Nanobiotechnol. 2024, 22, 414. [Google Scholar] [CrossRef] [PubMed]
  271. Mishra, A.K.; Jarwal, D.K.; Mukherjee, B.; Kumar, A.; Ratan, S.; Jit, S. CuO nanowire-based extended-gate field-effect-transistor (FET) for pH sensing and enzyme-free/receptor-free glucose sensing applications. IEEE Sens. J. 2020, 20, 5039–5047. [Google Scholar] [CrossRef]
  272. Khan, M.; Nagal, V.; Masrat, S.; Tuba, T.; Alam, S.; Bhat, K.S.; Wahid, I.; Ahmad, R. Vertically oriented zinc oxide nanorod-based electrolyte-gated field-effect transistor for high-performance glucose sensing. Anal. Chem. 2022, 94, 8867–8873. [Google Scholar] [CrossRef] [PubMed]
  273. Xu, T.; Ren, H.; Fang, Y.; Liang, K.; Zhang, H.; Li, D.; Chen, Y.; Zhu, B.; Wang, H. Glucose sensing by field-effect transistors based on interfacial hydrogelation of self-assembled peptide. Appl. Mater. Today 2023, 30, 101713. [Google Scholar] [CrossRef]
  274. Middya, S.; Bhattacharjee, M.; Bandyopadhyay, D. Reusable nano-BG-FET for point-of-care estimation of ammonia and urea in human urine. Nanotechnology 2019, 30, 145502. [Google Scholar] [CrossRef]
  275. Rayanasukha, Y.; Pratontep, S.; Porntheeraphat, S.; Bunjongpru, W.; Nukeaw, J. Non-enzymatic urea sensor using molecularly imprinted polymers surface modified based-on ion-sensitive field effect transistor (ISFET). Surf. Coat. Technol. 2016, 306, 147–150. [Google Scholar] [CrossRef]
  276. Kuo, P.-Y.; Chen, Y.-Y.; Lai, W.-H.; Chang, C.-H. An extended-gate field-effect transistor applied to resistive divider integrated with the readout circuit using 180nm CMOS process for uric acid detection. IEEE Sens. J. 2021, 21, 20229–20238. [Google Scholar] [CrossRef]
  277. Nasiruddin, M.; Waizumi, H.; Takaoka, T.; Wang, Z.; Sainoo, Y.; Al Mamun, M.S.; Ando, A.; Fukuyama, M.; Hibara, A.; Komeda, T. A microfluidic approach for the detection of uric acid through electrical measurement using an atomically thin MoS 2 field-effect transistor. Analyst 2023, 148, 4091–4098. [Google Scholar] [CrossRef]
  278. Vozgirdaite, D.; Ben Halima, H.; Bellagambi, F.G.; Alcacer, A.; Palacio, F.; Jaffrezic-Renault, N.; Zine, N.; Bausells, J.; Elaissari, A.; Errachid, A. Development of an ImmunoFET for analysis of tumour necrosis factor-α in artificial saliva: Application for heart failure monitoring. Chemosensors 2021, 9, 26. [Google Scholar] [CrossRef]
  279. Wang, Z.; Hao, Z.; Yu, S.; Huang, C.; Pan, Y.; Zhao, X. A wearable and deformable graphene-based affinity nanosensor for monitoring of cytokines in biofluids. Nanomaterials 2020, 10, 1503. [Google Scholar] [CrossRef] [PubMed]
  280. Chen, S.; Sun, Y.; Xia, Y.; Lv, K.; Man, B.; Yang, C. Donor effect dominated molybdenum disulfide/graphene nanostructure-based field-effect transistor for ultrasensitive DNA detection. Biosens. Bioelectron. 2020, 156, 112128. [Google Scholar] [CrossRef] [PubMed]
  281. Villiger, M.; Stoop, R.; Vetsch, T.; Hohenauer, E.; Pini, M.; Clarys, P.; Pereira, F.; Clijsen, R. Evaluation and review of body fluids saliva, sweat and tear compared to biochemical hydration assessment markers within blood and urine. Eur. J. Clin. Nutr. 2018, 72, 69–76. [Google Scholar] [CrossRef] [PubMed]
  282. Dossin, T.; Goffin, E. When the color of peritoneal dialysis effluent can be used as a diagnostic tool. Semin. Dial. 2019, 32, 72–79. [Google Scholar] [CrossRef]
  283. Gasparyan, L.; Gasparyan, F.; Simonyan, V. Internal Electrical Noises of BioFET Sensors Based on Various Architectures. Open J. Biophys. 2021, 11, 177–204. [Google Scholar] [CrossRef]
  284. Bhattacharyya, I.M.; Ron, I.; Chauhan, A.; Pikhay, E.; Greental, D.; Mizrahi, N.; Roizin, Y.; Shalev, G. A new approach towards the Debye length challenge for specific and label-free biological sensing based on field-effect transistors. Nanoscale 2022, 14, 2837–2847. [Google Scholar] [CrossRef]
  285. Ravariu, C.; Srinivasulu, A.; Mihaiescu, D.E.; Musala, S. Generalized analytical model for enzymatic BioFET transistors. Biosensors 2022, 12, 474. [Google Scholar] [CrossRef]
  286. Sakata, T.; Nishitani, S.; Kajisa, T. Molecularly imprinted polymer-based bioelectrical interfaces with intrinsic molecular charges. RSC Adv. 2020, 10, 16999–17013. [Google Scholar] [CrossRef]
  287. Kim, K.; Lee, J.; Moon, B.M.; Seo, Y.B.; Park, C.H.; Park, M.; Sung, G.Y. Fabrication of a urea biosensor for real-time dynamic fluid measurement. Sensors 2018, 18, 2607. [Google Scholar] [CrossRef]
  288. Molinnus, D.; Beging, S.; Lowis, C.; Schöning, M.J. Towards a multi-enzyme capacitive field-effect biosensor by comparative study of drop-coating and nano-spotting technique. Sensors 2020, 20, 4924. [Google Scholar] [CrossRef]
  289. Lin, Y.-H.; Wang, S.-H.; Wu, M.-H.; Pan, T.-M.; Lai, C.-S.; Luo, J.-D.; Chiou, C.-C. Integrating solid-state sensor and microfluidic devices for glucose, urea and creatinine detection based on enzyme-carrying alginate microbeads. Biosens. Bioelectron. 2013, 43, 328–335. [Google Scholar] [CrossRef] [PubMed]
  290. Poghossian, A.; Schöning, M.J. Capacitive field-effect EIS chemical sensors and biosensors: A status report. Sensors 2020, 20, 5639. [Google Scholar] [CrossRef] [PubMed]
  291. Daurai, B.; Ramchiary, S.S.; Gogoi, M. Enzymatic Biosensors for Healthcare Applications. In Enzyme-based Biosensors: Recent Advances and Applications in Healthcare; Springer: Berlin/Heidelberg, Germany, 2024; pp. 1–29. [Google Scholar]
  292. Li, Y.; Luo, L.; Nie, M.; Davenport, A.; Li, Y.; Li, B.; Choy, K.-L. A graphene nanoplatelet-polydopamine molecularly imprinted biosensor for Ultratrace creatinine detection. Biosens. Bioelectron. 2022, 216, 114638. [Google Scholar] [CrossRef]
  293. Jang, H.-J.; Lee, T.; Song, J.; Russell, L.; Li, H.; Dailey, J.; Searson, P.C.; Katz, H.E. Electronic cortisol detection using an antibody-embedded polymer coupled to a field-effect transistor. ACS Appl. Mater. Interfaces 2018, 10, 16233–16237. [Google Scholar] [CrossRef]
  294. Zhang, A.; Lieber, C.M. Nano-bioelectronics. Chem. Rev. 2016, 116, 215–257. [Google Scholar] [CrossRef]
  295. Nishitani, S.; Sakata, T. Polymeric nanofilter biointerface for potentiometric small-biomolecule recognition. ACS Appl. Mater. Interfaces 2019, 11, 5561–5569. [Google Scholar] [CrossRef]
  296. Bao, Z.; Sun, J.; Zhao, X.; Li, Z.; Cui, S.; Meng, Q.; Zhang, Y.; Wang, T.; Jiang, Y. Top-down nanofabrication of silicon nanoribbon field effect transistor (Si-NR FET) for carcinoembryonic antigen detection. Int. J. Nanomed. 2017, 12, 4623–4631. [Google Scholar] [CrossRef]
  297. Yakkala, B.; Azariah, C.R.; Narendran, S.; Sivagami, A. Enhancing the sensitivity of a BIOFET by varying the oxide thickness and nanowire length for water sensing application. IOP Conf. Ser. Earth Environ. Sci. 2021, 945, 012011. [Google Scholar] [CrossRef]
  298. Shoorideh, K.; Chui, C.O. On the origin of enhanced sensitivity in nanoscale FET-based biosensors. Proc. Natl. Acad. Sci. USA 2014, 111, 5111–5116. [Google Scholar] [CrossRef]
  299. Ren, R.; Zhang, Y.; Nadappuram, B.P.; Akpinar, B.; Klenerman, D.; Ivanov, A.P.; Edel, J.B.; Korchev, Y. Nanopore extended field-effect transistor for selective single-molecule biosensing. Nat. Commun. 2017, 8, 586. [Google Scholar] [CrossRef]
  300. Murugasenapathi, N.K.; Ghosh, R.; Ramanathan, S.; Ghosh, S.; Chinnappan, A.; Mohamed, S.A.J.; Esther Jebakumari, K.A.; Gopinath, S.C.; Ramakrishna, S.; Palanisamy, T. Transistor-Based Biomolecule Sensors: Recent Technological Advancements and Future Prospects. Crit. Rev. Anal. Chem. 2023, 53, 1044–1065. [Google Scholar] [CrossRef] [PubMed]
  301. Kim, S.; Lee, R.; Kwon, D.; Kim, T.-H.; Park, T.J.; Choi, S.-J.; Mo, H.-S.; Kim, D.H.; Park, B.-G. Multiplexed silicon nanowire tunnel FET-based biosensors with optimized multi-sensing currents. IEEE Sens. J. 2021, 21, 8839–8846. [Google Scholar] [CrossRef]
  302. Li, D.; Chen, H.; Fan, K.; Labunov, V.; Lazarouk, S.; Yue, X.; Liu, C.; Yang, X.; Dong, L.; Wang, G. A supersensitive silicon nanowire array biosensor for quantitating tumor marker ctDNA. Biosens. Bioelectron. 2021, 181, 113147. [Google Scholar] [CrossRef]
  303. Li, C.; Liu, F.; Han, R.; Zhuang, Y. A vertically stacked Nanosheet gate-all-around FET for biosensing application. IEEE Access 2021, 9, 63602–63610. [Google Scholar] [CrossRef]
  304. Eswaran, M.; Chokkiah, B.; Pandit, S.; Rahimi, S.; Dhanusuraman, R.; Aleem, M.; Mijakovic, I. A Road Map toward Field-Effect Transistor Biosensor Technology for Early Stage Cancer Detection. Small Methods 2022, 6, 2200809. [Google Scholar] [CrossRef]
  305. Ouyang, J.; Li, Y.; Yang, F.; Wu, X.; Qiu, Z.; Shu, J. Electrochemical and Field Effect Transistor Dual-Mode Biosensor Chip for Label-Free Detection of Cytokine Storm Biomarker with High Sensitivity within a Wide Range. Adv. Funct. Mater. 2024, 34, 2405212. [Google Scholar] [CrossRef]
  306. Xing, Y.; Hu, J.; Wu, Z.; Huang, P.; Li, B.; Liu, L. Ultrasensitive detection of miRNA based on single-walled carbon nanotube@ MoS2 field-effect transistor regulated by atomic layer deposition. Carbon 2024, 223, 119022. [Google Scholar] [CrossRef]
  307. Hideshima, S.; Hayashi, H.; Takeuchi, R.; Wustoni, S.; Kuroiwa, S.; Nakanishi, T.; Momma, T.; Osaka, T. Improvement in long-term stability of field effect transistor biosensor in aqueous environments using a combination of silane and reduced graphene oxide coating. Microelectron. Eng. 2022, 264, 111859. [Google Scholar] [CrossRef]
  308. Zhou, W.; Dai, X.; Fu, T.-M.; Xie, C.; Liu, J.; Lieber, C.M. Long term stability of nanowire nanoelectronics in physiological environments. Nano Lett. 2014, 14, 1614–1619. [Google Scholar] [CrossRef]
  309. Albarghouthi, F.M.; Williams, N.X.; Doherty, J.L.; Lu, S.; Franklin, A.D. Passivation strategies for enhancing solution-gated carbon nanotube field-effect transistor biosensing performance and stability in ionic solutions. ACS Appl. Nano Mater. 2022, 5, 15865–15874. [Google Scholar] [CrossRef]
  310. Wang, H.-Y.; Tian, Y.-F.; Chien, C.-C.; Kan, W.-C.; Liao, P.-C.; Wu, H.-Y.; Su, S.-B.; Lin, C.-Y. Differential proteomic characterization between normal peritoneal fluid and diabetic peritoneal dialysate. Nephrol. Dial. Transplant. 2010, 25, 1955–1963. [Google Scholar] [CrossRef]
  311. Olivares-Gandy, H.J.; Domínguez-Isidro, S.; López-Domínguez, E.; Hernández-Velázquez, Y.; Tapia-McClung, H.; de-la-Calleja, J. A telemonitoring system for nutritional intake in patients with chronic kidney disease receiving peritoneal dialysis therapy. Comput. Biol. Med. 2019, 109, 1–13. [Google Scholar] [CrossRef] [PubMed]
  312. Chae, Y.J.; Kim, H.S. Effects of a mobile application on improving self-management of adult patients receiving peritoneal dialysis: A randomized controlled trial. Jpn. J. Nurs. Sci. 2024, 21, e12555. [Google Scholar] [CrossRef] [PubMed]
  313. Sayyad, P.W.; Park, S.-J.; Ha, T.-J. Recent advances in biosensors based on metal-oxide semiconductors system-integrated into bioelectronics. Biosens. Bioelectron. 2024, 259, 116407. [Google Scholar] [CrossRef] [PubMed]
  314. Yang, Y.; Zeng, B.; Li, Y.; Liang, H.; Yang, Y.; Yuan, Q. Construction of MoS2 field effect transistor sensor array for the detection of bladder cancer biomarkers. Sci. China Chem. 2020, 63, 997–1003. [Google Scholar] [CrossRef]
  315. Tawade, P.; Mastrangeli, M. Integrated Electrochemical and Optical Biosensing in Organs-on-Chip. ChemBioChem 2024, 25, e202300560. [Google Scholar] [CrossRef]
  316. Rothberg, J.M.; Hinz, W.; Rearick, T.M.; Schultz, J.; Mileski, W.; Davey, M.; Leamon, J.H.; Johnson, K.; Milgrew, M.J.; Edwards, M. An integrated semiconductor device enabling non-optical genome sequencing. Nature 2011, 475, 348–352. [Google Scholar] [CrossRef]
  317. Zhang, R.; Hao, T.; Hu, S.; Wang, K.; Ren, S.; Tian, Z.; Jia, Y. Electrolyte-Gated Graphene Field Effect Transistor-Based Ca2+ Detection Aided by Machine Learning. Sensors 2022, 23, 353. [Google Scholar] [CrossRef]
  318. Xue, M.; Mackin, C.; Weng, W.-H.; Zhu, J.; Luo, Y.; Luo, S.-X.L.; Lu, A.-Y.; Hempel, M.; McVay, E.; Kong, J. Integrated biosensor platform based on graphene transistor arrays for real-time high-accuracy ion sensing. Nat. Commun. 2022, 13, 5064. [Google Scholar] [CrossRef]
  319. Cullis, B.; Abdelraheem, M.; Abrahams, G.; Balbi, A.; Cruz, D.N.; Frishberg, Y.; Koch, V.; McCulloch, M.; Numanoglu, A.; Nourse, P. Peritoneal dialysis for acute kidney injury. Perit. Dial. Int. 2014, 34, 494–517. [Google Scholar] [CrossRef]
  320. Panahi, A.; Ghafar-Zadeh, E. A hybrid microfluidic electronic sensing platform for life science applications. Micromachines 2022, 13, 425. [Google Scholar] [CrossRef] [PubMed]
  321. Dai, C.; Xiong, H.; He, R.; Zhu, C.; Li, P.; Guo, M.; Gou, J.; Mei, M.; Kong, D.; Li, Q. Electro-Optical Multiclassification Platform for Minimizing Occasional Inaccuracy in Point-of-Care Biomarker Detection. Adv. Mater. 2024, 36, 2312540. [Google Scholar] [CrossRef] [PubMed]
  322. Xie, Z.; Sun, P.; Cao, S.; Yang, Y.; Wang, X.; Xiao, G.; Yan, G.; Bi, J.; Ji, J.; Yue, Z. Multi-channel detection of dopamine and glucose utilizing graphene field effect transistor electrochemical sensor and efficient data fusion algorithm. J. Electroanal. Chem. 2023, 950, 117901. [Google Scholar] [CrossRef]
  323. Sarfraz, N.; Khan, I. Plasmonic gold nanoparticles (AuNPs): Properties, synthesis and their advanced energy, environmental and biomedical applications. Chem. Asian J. 2021, 16, 720–742. [Google Scholar] [CrossRef]
  324. Lee-Thorp, J.P.; Weinstein, M.I.; Zhu, Y. Elliptic operators with honeycomb symmetry: Dirac points, edge states and applications to photonic graphene. Arch. Ration. Mech. Anal. 2019, 232, 1–63. [Google Scholar] [CrossRef]
  325. Zheng, Y.; Omar, R.; Zhang, R.; Tang, N.; Khatib, M.; Xu, Q.; Milyutin, Y.; Saliba, W.; Broza, Y.Y.; Wu, W. A wearable microneedle-based extended gate transistor for real-time detection of sodium in interstitial fluids. Adv. Mater. 2022, 34, 2108607. [Google Scholar] [CrossRef]
  326. Garcia-Cordero, E.; Bellando, F.; Zhang, J.; Wildhaber, F.; Longo, J.; Guérin, H.; Ionescu, A.M. Three-dimensional integrated ultra-low-volume passive microfluidics with ion-sensitive field-effect transistors for multiparameter wearable sweat analyzers. ACS Nano 2018, 12, 12646–12656. [Google Scholar] [CrossRef]
  327. Sakata, T.; Hagio, M.; Saito, A.; Mori, Y.; Nakao, M.; Nishi, K. Biocompatible and flexible paper-based metal electrode for potentiometric wearable wireless biosensing. Sci. Technol. Adv. Mater. 2020, 21, 379–387. [Google Scholar] [CrossRef]
  328. Ku, M.; Kim, J.; Won, J.-E.; Kang, W.; Park, Y.-G.; Park, J.; Lee, J.-H.; Cheon, J.; Lee, H.H.; Park, J.-U. Smart, soft contact lens for wireless immunosensing of cortisol. Sci. Adv. 2020, 6, eabb2891. [Google Scholar] [CrossRef]
  329. McQuarrie, E.P.; Traynor, J.P.; Taylor, A.H.; Freel, E.M.; Fox, J.G.; Jardine, A.G.; Mark, P.B. Association between urinary sodium, creatinine, albumin, and long-term survival in chronic kidney disease. Hypertension 2014, 64, 111–117. [Google Scholar] [CrossRef]
  330. Zhang, J.; Rupakula, M.; Bellando, F.; Garcia Cordero, E.; Longo, J.; Wildhaber, F.; Herment, G.; Guerin, H.; Ionescu, A.M. Sweat biomarker sensor incorporating picowatt, three-dimensional extended metal gate ion sensitive field effect transistors. ACS Sens. 2019, 4, 2039–2047. [Google Scholar] [CrossRef] [PubMed]
  331. Oh, H.G.; Jeon, D.C.; Gianti, M.S.; Cho, H.S.; Jo, D.A.; Indriatmoko, M.N.; Jang, B.K.; Lim, J.M.; Cho, S.; Song, K.S. Two–dimensional disposable graphene sensor to detect na+ ions. Nanomaterials 2021, 11, 787. [Google Scholar] [CrossRef] [PubMed]
  332. Park, S.-C.; Jeong, H.J.; Heo, M.; Shin, J.H.; Ahn, J.-H. Carbon nanotube-based ion-sensitive field-effect transistors with an on-chip reference electrode toward wearable sodium sensing. ACS Appl. Electron. Mater. 2021, 3, 2580–2588. [Google Scholar] [CrossRef]
  333. Ito, K.; Satake, H.; Mori, Y.; Tseng, A.C.; Sakata, T. Biocompatible and Na+-sensitive thin-film transistor for biological fluid sensing. Sci. Technol. Adv. Mater. 2019, 20, 917–926. [Google Scholar] [CrossRef]
  334. Hu, C.; Wang, L.; Liu, S.; Sheng, X.; Yin, L. Recent Development of Implantable Chemical Sensors Utilizing Flexible and Biodegradable Materials for Biomedical Applications. ACS Nano 2024, 18, 3969–3995. [Google Scholar] [CrossRef]
  335. Li, G.; Wei, Q.; Zhang, Q.; Wei, F. Biocompatible Hydrogel Modified Silicon Nanowire Field Transistor for Highly Sensitive pH Detection. J. Phys. Conf. Ser. 2022, 2248, 012005. [Google Scholar] [CrossRef]
  336. Jiang, Z.; Ye, D.; Xiang, L.; He, Z.; Dai, X.; Yang, J.; Xiong, Q.; Ma, Y.; Zhi, D.; Zou, Y. A drug-mediated organic electrochemical transistor for robustly reusable biosensors. Nat. Mater. 2024, 23, 1547–1555. [Google Scholar] [CrossRef]
Figure 1. Schematic overview of BioFETs components and their applications in PD monitoring.
Figure 1. Schematic overview of BioFETs components and their applications in PD monitoring.
Biosensors 15 00193 g001
Figure 2. Schematic of BioFETs structure and workflow.
Figure 2. Schematic of BioFETs structure and workflow.
Biosensors 15 00193 g002
Figure 3. Evaluation of the performance of BioFETs and the characteristics of BioFETs through a semiconductor characterization system.
Figure 3. Evaluation of the performance of BioFETs and the characteristics of BioFETs through a semiconductor characterization system.
Biosensors 15 00193 g003
Figure 4. Common bioreceptors used in BioFETs for PD-related biomarkers monitoring.
Figure 4. Common bioreceptors used in BioFETs for PD-related biomarkers monitoring.
Biosensors 15 00193 g004
Figure 5. (A) Glucose oxidase and PEG layers were co-modified on the SiNW surface for the glucose detection through catalytic principle. (B) Transfer curves of SiNW-BioFETs for detecting glucose in serum samples [77]. (C) Schematic of the MWCNT-FET modified with aptamer for capturing target CRP. (D) The response demonstrated by BioFETs indicated its high affinity for the target CRP despite interference from three other proteins commonly present during PD inflammation [78]. (E) Schematic of a PNA-based EGFET fabricated on PCB for DNA hybridization. Blue lines are the PNA probes, while the red ones are the target DNA. (F) Mean values of Vdirac shifts of BioFETs used for DNA hybridization (bars represent standard deviations of 5 BioFETs) [79]. (G) SEM images of bare graphene and graphene layer conjugated with probes and exosomes. (H) Schematic of a microfluidic GFET modified with CD63 antibody for detecting the surface protein of exosome [80] (images used under Creative Commons Licenses).
Figure 5. (A) Glucose oxidase and PEG layers were co-modified on the SiNW surface for the glucose detection through catalytic principle. (B) Transfer curves of SiNW-BioFETs for detecting glucose in serum samples [77]. (C) Schematic of the MWCNT-FET modified with aptamer for capturing target CRP. (D) The response demonstrated by BioFETs indicated its high affinity for the target CRP despite interference from three other proteins commonly present during PD inflammation [78]. (E) Schematic of a PNA-based EGFET fabricated on PCB for DNA hybridization. Blue lines are the PNA probes, while the red ones are the target DNA. (F) Mean values of Vdirac shifts of BioFETs used for DNA hybridization (bars represent standard deviations of 5 BioFETs) [79]. (G) SEM images of bare graphene and graphene layer conjugated with probes and exosomes. (H) Schematic of a microfluidic GFET modified with CD63 antibody for detecting the surface protein of exosome [80] (images used under Creative Commons Licenses).
Biosensors 15 00193 g005
Figure 6. Comparison of various probes in solutions with different ionic strengths.
Figure 6. Comparison of various probes in solutions with different ionic strengths.
Biosensors 15 00193 g006
Figure 7. Common transducing materials used in BioFETs.
Figure 7. Common transducing materials used in BioFETs.
Biosensors 15 00193 g007
Figure 8. (A) XRD pattern of CuO NPs. The insert is a photo of CuO NPs ink. (B) The effect of Nafion coating concentration on responding to various glucose concentration at the 0.5 V potential [85]. (C) AFM analysis of Ag surface functionalized with urease. (D) The curve of ID/ID0 value when FET is exposed to 1 × PBS with and without target urea at a fixed Vg of 0.3 V [139]. (E) Schematic of the principle of MOS-BioFET-based detection for DNA molecules. (F) SEM image of FET AI:ZnO channel. (G) XPS analysis of AI:ZnO film indicated the formation of AI-O-Zn bonds [140] (images used under Creative Commons licenses).
Figure 8. (A) XRD pattern of CuO NPs. The insert is a photo of CuO NPs ink. (B) The effect of Nafion coating concentration on responding to various glucose concentration at the 0.5 V potential [85]. (C) AFM analysis of Ag surface functionalized with urease. (D) The curve of ID/ID0 value when FET is exposed to 1 × PBS with and without target urea at a fixed Vg of 0.3 V [139]. (E) Schematic of the principle of MOS-BioFET-based detection for DNA molecules. (F) SEM image of FET AI:ZnO channel. (G) XPS analysis of AI:ZnO film indicated the formation of AI-O-Zn bonds [140] (images used under Creative Commons licenses).
Biosensors 15 00193 g008
Figure 9. (A) The immobilization of the non-specific receptor protein (BLIP2) on the SWCNT channel of FET leads to binding events exceeding the Debye screening length. Green parts represent lysine residues. (B) Poor design part indicates that the binding site of BLIP2 is obstructed by the CNT or that the binding reaction occurs outside the Debye length. The proposed modeling-led design optimizes analyte binding within the Debye length and introduces an electrostatic profile to the SWCNT. (C) I-VDS before (black) and after (red) the attachment of two viable rotamers for SWCNT docking. The shaded region is the standard error of the average values [141]. (D) SEM image of SiNW channel. Inset I and II are the SEM images of cross section and top view, respectively. (E) Ids-Vgs curves of BioFET measured with various cell concentrations and PBS. (F) SEM image of exosomes captured by the CD63 antibody-modified FET channel and control channel [50]. (G) Schematic of single-molecule FET structure with a single dinuclear ruthenium-diarylethene (Ru-DAE) complex connects the nanogapped graphene point contacts for optimized functions. (H) Left: STEM analysis of the cross section of the Al/Al2O3/HfO2 multilayer structure under 200 kV. Right: XPS analysis of the dielectric layer with 5 nm Al2O3 and 5 nm HfO2 [142] (images used under Creative Commons licenses).
Figure 9. (A) The immobilization of the non-specific receptor protein (BLIP2) on the SWCNT channel of FET leads to binding events exceeding the Debye screening length. Green parts represent lysine residues. (B) Poor design part indicates that the binding site of BLIP2 is obstructed by the CNT or that the binding reaction occurs outside the Debye length. The proposed modeling-led design optimizes analyte binding within the Debye length and introduces an electrostatic profile to the SWCNT. (C) I-VDS before (black) and after (red) the attachment of two viable rotamers for SWCNT docking. The shaded region is the standard error of the average values [141]. (D) SEM image of SiNW channel. Inset I and II are the SEM images of cross section and top view, respectively. (E) Ids-Vgs curves of BioFET measured with various cell concentrations and PBS. (F) SEM image of exosomes captured by the CD63 antibody-modified FET channel and control channel [50]. (G) Schematic of single-molecule FET structure with a single dinuclear ruthenium-diarylethene (Ru-DAE) complex connects the nanogapped graphene point contacts for optimized functions. (H) Left: STEM analysis of the cross section of the Al/Al2O3/HfO2 multilayer structure under 200 kV. Right: XPS analysis of the dielectric layer with 5 nm Al2O3 and 5 nm HfO2 [142] (images used under Creative Commons licenses).
Biosensors 15 00193 g009
Figure 10. Various configurations of BioFETs. (a) back-gate; (b) dual-gate; (c) electrolyte-gate; and (d) extended-gate (edited from Nguyen et al. [43], under Creative Commons license).
Figure 10. Various configurations of BioFETs. (a) back-gate; (b) dual-gate; (c) electrolyte-gate; and (d) extended-gate (edited from Nguyen et al. [43], under Creative Commons license).
Biosensors 15 00193 g010
Figure 11. Schematic of some strategies for enhancing sensitivity. (A) A general model of EnFET [285]. (B) Silicalite was used as adsorbent at the top for the creatinine deiminase immobilization [69]. (C) Schematic diagram of the PCN-Au NPs IGZO TFT aptamer sensor, in which the aptamer is immobilized on Au NPs. The inset shows that under a constant bias voltage, the output characteristics of the PCN-Au NPs-modified IGZO are greatly optimized [135]. (D) Schematic of MIP-BioFET structure and illustration of MIP binding principle with paromomycin template as an example [286] (edited from images under Creative Commons licenses).
Figure 11. Schematic of some strategies for enhancing sensitivity. (A) A general model of EnFET [285]. (B) Silicalite was used as adsorbent at the top for the creatinine deiminase immobilization [69]. (C) Schematic diagram of the PCN-Au NPs IGZO TFT aptamer sensor, in which the aptamer is immobilized on Au NPs. The inset shows that under a constant bias voltage, the output characteristics of the PCN-Au NPs-modified IGZO are greatly optimized [135]. (D) Schematic of MIP-BioFET structure and illustration of MIP binding principle with paromomycin template as an example [286] (edited from images under Creative Commons licenses).
Biosensors 15 00193 g011
Figure 12. (A) The photograph of a flexible sensor array stretched to fit human skin. (B) Schematic of an In2O3 nanoribbon FET for the simultaneous detection of serotonin and dopamine. (C) Sensor array simultaneously senses temperature, pH, serotonin, and dopamine changes in artificial cerebrospinal fluid. The two devices were exposed to 1 pM, 1 nM serotonin, 1 pM, and 1 nM dopamine. Both devices responded to the pH change, but only the FET functionalized with the respective target responded to the corresponding target (working under a Vds of 0.2 V and a Vgs of 0.25 V). (D) Mobility and threshold voltage change curves of the FET sensor in the relaxed state after different bending cycles [148]. (E) Schematic of FET-based calibration-free detection of Ca2+. The electrical properties of the aptamer-functionalized GFET were measured using a source measurement unit (SMU). (F) The measured characteristic curves were analyzed using four machine learning-based algorithms to directly predict the analyte concentration [317]. (G) Ids-Vgs curve of sensing chip measured in electrolyte solution containing 100 μM K+, 30 mM Na+, and 1 mM Ca2+ ions. The solid line is the average value, and the shaded area represents the standard error. (H) Single sensing unit with specific ISM. The right curve shows the relationship between the electrostatic potential and the distance from the FET graphene surface. ISM: ion sensitive membrane. (I) Schematic diagram of a FET chip composed of N × N sensing units. Different ISMs are printed into different areas for multiplexed detection of different ions [318] (images used under Creative Commons licenses).
Figure 12. (A) The photograph of a flexible sensor array stretched to fit human skin. (B) Schematic of an In2O3 nanoribbon FET for the simultaneous detection of serotonin and dopamine. (C) Sensor array simultaneously senses temperature, pH, serotonin, and dopamine changes in artificial cerebrospinal fluid. The two devices were exposed to 1 pM, 1 nM serotonin, 1 pM, and 1 nM dopamine. Both devices responded to the pH change, but only the FET functionalized with the respective target responded to the corresponding target (working under a Vds of 0.2 V and a Vgs of 0.25 V). (D) Mobility and threshold voltage change curves of the FET sensor in the relaxed state after different bending cycles [148]. (E) Schematic of FET-based calibration-free detection of Ca2+. The electrical properties of the aptamer-functionalized GFET were measured using a source measurement unit (SMU). (F) The measured characteristic curves were analyzed using four machine learning-based algorithms to directly predict the analyte concentration [317]. (G) Ids-Vgs curve of sensing chip measured in electrolyte solution containing 100 μM K+, 30 mM Na+, and 1 mM Ca2+ ions. The solid line is the average value, and the shaded area represents the standard error. (H) Single sensing unit with specific ISM. The right curve shows the relationship between the electrostatic potential and the distance from the FET graphene surface. ISM: ion sensitive membrane. (I) Schematic diagram of a FET chip composed of N × N sensing units. Different ISMs are printed into different areas for multiplexed detection of different ions [318] (images used under Creative Commons licenses).
Biosensors 15 00193 g012
Table 1. Comparison of various probe types in FETs for target detection.
Table 1. Comparison of various probe types in FETs for target detection.
Binding PrincipleBiosensing ElementsTargetLOD
Enzyme-basedGODglucose 200 nM [84]
GOD/Zno/Cuoglucose30 nM [116]
GOD/MOFsglucose0.51 μM [117]
Ureaseurea1 μM [118]
SOXsarcosine105 zM [119]
papainCys C0.05 ag/uL [30]
DNA-basedssDNAE. coli DNA1 fM [94]
SNAvirus RNA0.13 copies/uL [120]
DNAmiRNA100 aM [99]
PNAmiRNA0.1 aM [99]
Immuno-basedantibodyTNF-α1 pg/mL [121]
antibodyserotonin0.1 fM [122]
antibody/AuNPsIL-62.2 fM[123]
antibodyCys C0.25 ag/mL [124]
antibody/caseinE. coli1 CFU/mL [125]
antibodyCRP0.06 μg/mL [35]
antibody fragmentCRP0.73 μg/mL [126]
Aptamer-basedaptamerglucose0.5 fM [127]
aptamerIL-6618 fM [128]
aptamer-SHcortisol1 pM [129]
Other binding events-basedPBAglucose0.15 μM [130]
GO/Agglucose1 μM [131]
MIPserotonin0.05 fM [132]
MIPTNF-α0.55 pg/mL [133]
GOD, glucose oxidase; LOD: limit of detection; MOF, metal-organic frameworks; SOX, sarcosine enzyme; Cys C, cystatin C; ssDNA, single-stranded DNA; SNA, spherical nucleic acids; PNA, peptide nucleic acid; CRP, C-reactive protein; PBA, phenylboronic acid; GO, graphene oxide; and MIP, molecularly imprinted polymer.
Table 2. Examples of transducing materials and modifications in enhancing FET performance.
Table 2. Examples of transducing materials and modifications in enhancing FET performance.
TypeTransducing Material (Modification)TargetRemarkSensitivity
BulkSi
(Ge)
/Si0.7Ge0.3Electric performance: 2 × higher sensitivity vs. pure Si [144]
Bulk graphene
(AuNPs)
/Higher defect density of bulk grapheneON/OFF current ratio: 4 × higher [146]
MOSIn2O3 (Al2O3)/Surface passivationlow operation voltage (0.05 V) [152]
In2O3 (Al2O3/SU-8)/Passivation and protection layerlow operation voltage (0.005 V) [153]
1DSi NWsExosome45 nm width poly-Si nanowires2159 particles/mL [50]
Si NWsVirus DNA10 Si nanowires1.93 fM [159]
Si NWsCRPAntibody fragment as probe0.73 μg/mL [126]
SWCNTCRPSuitable for minute quantities of analytes0.06 μg/mL [35]
Si nanosheetCRPVertically stacked channels100 pg/mL [175]
SCNT
(liquid silver)
DNASuspended material mitigates the impact of the substrate10 aM [176]
MWCNT
(PEI/ZrO2)
/ZrO2: decrease current leakage; PEI: increase CNT conductivity [69]/
2DGraphene
(plasma)
DNAPlasma treatment removes residues and enhances the hydrophilicity10 aM (an order of magnitude higher vs. unmodified) [177]
Graphene
(Monolayer)
DNACVD-grown graphene15 fM [166]
Graphene
(defect-engineered)
GlucoseMore binding sites introduced with lower energy barriers Glucose sensitivity: 0.16 mV/mM higher vs. unmodified [170]
Graphene
(polymerized)
GlucoseReversible reaction between glucose and polymer1.9 μM [178]
rGOmiRNAEasier to modify1 fM [179]
WS2
(monolayer)
/MOCVD minimizes the contamination during film transfer33 cm2 V−1 s−1 of carrier mobility (1.5 times higher than the best reported) [180]
MoS2
(nanoporous)
CortisolMore binding sites and edges 1 ag/mL [172]
Single-layer MoS2
(Pt)
CortisolStronger charge transfer effect [181]/
MXeneCD9/ExosomeTunable properties and the numerous functional groups on MXeneCD9: 10.64 pM
Exosome: 6.41 × 102 exosomes/mL [101]
Black phosphorusCortisolAdjustable band gap1 aM [136]
Si NWs, silicon nanowires; CRP, C-reactive protein; SWCNT, single-walled carbon nanotube; SCNT, suspended carbon nanotube; MWCNT, multi-walled carbon nanotube; PEI, polyethylene imine; CVD, chemical vapor deposition; rGO, reduced graphene oxide; and MOCVD, metal–organic chemical vapor deposition.
Table 3. Examples of transducing materials and modifications in enhancing the performance of BioFETs.
Table 3. Examples of transducing materials and modifications in enhancing the performance of BioFETs.
BiofluidsBiomarkerDisease/ApplicationRemarksRef.
PDEGlucoseUltrafiltration failure/PFLong-term effects[243]
PDE/plasma/serumUreaUrea removal/renal function<20 mM during PD[240]
PDE/serumCreatinineCreatinine clearance/renal function/[244]
PDEIL-6Acute inflammation/PD peritonitis/solute transport rate/EPSSharply increase during inflammation[245]
PDECA125Overhydration/EPSPeritoneal mesothelial cell count[246,247]
PDE/serumHAPFCharacteristic of PF and wound healing in the peritoneum[26]
PDEWater channel Aquaporin 1Ultrafiltration failureIncrease[237]
PDEMMP-2/VEGFInflammation/PF/solute transport rate/[19]
PDE/urineNa+Cardiovascular eventsLow sodium clearance rate[241]
PDE/tears/salivaTNF-αInflammation/membrane damage/[242]
PDEExtracellular vesiclesIndicate peritoneal membrane functionProteome difference[248]
PDElipopolysaccharideGNB infection/[249]
PDE/urineDecoy receptor 2PF/[17]
PDE/serumCRPInflammationSharply increase[250]
PDEDNA and miRNA (bacterial and mitochondrial)Local and systemic inflammation/solute transport rate/[251]
PDE/serumCysCAcute kidney injury/residual renal function/cardiovascular diseases/[252,253]
PDE, peritoneal dialysis effluent; PF, peritoneal fibrosis; PD, peritoneal dialysis; EPS, encapsulating peritoneal sclerosis; HA, hyaluronan; MMP, metalloproteinase; GNB, Gram-negative bacterial; CRP, C-reactive protein; and CysC, cystatin C.
Table 4. Examples of BioFETs used for PD-related monitoring.
Table 4. Examples of BioFETs used for PD-related monitoring.
ApplicationTargetProbeTransducing MaterialUsed SamplesLODRef.
Key IndicatorsGlucosePBACopolymerized hydrogelBuffer solution (PBS)5 μM[205]
GlucoseCuOCuO nanowiresSlightly basic medium (pH = 7.4)1 mM[271]
GlucoseEnzymePEG/SiNWBuffer solution (PBS)10 nM[77]
GlucoseEnzymeVertical ZnO nanorodsBuffer solution (PBS)0.05 mM[272]
GlucosePeptide hydrogelIn2O3Glucose samples10 nM[273]
GlucosePBAMIPBuffer solution (sodium phosphate)3 μM[72]
CreatinineEnzymeMWCNTBuffer solution (pH = 7.4)/[228]
CreatinineEnzymeSilicaliteBuffer solution (KH2PO4-NaOH)5 μM[231]
Urea/ammoniaCdS/TiO2CdS/TiO2Urine samples0.85 ppm[274]
UreaUreaseSnO2/IGZOBuffer solution (PBS)/[219]
UreaMIPSi3N4Urea samples1.0 × 10−4 M[275]
UreaUreaseSiBuffer solution (PBS)/[139]
Uric acidUricaseRuO2Buffer solution (PBS)0.082 mg/dL[276]
Uric acidMoS2MoS2IPA solvent60 nM[277]
Potential Infection BiomarkersIL-6AntibodySWCNTBuffer solution (PBS)1.37 pg/mL[255]
IL-6AptamerHfO2/grapheneSaliva12 pM[187]
IL-63’-thiolated aptamersMXene/10 fg/mL[257]
CA125AptamerMWCNTBuffer solution (PBS)0.5 nU/mL[261]
CA125AntibodyInSe/0.01 U/mL[262]
AlbuminAntibodySWCNTHuman serum0.47 fg/mL[112]
TNF-αAntibodySi3N4Artificial saliva and Buffer solution (PBS)/[278]
TNF-αAptamerGrapheneArtificial tears1 nM[266]
TNF-α and IFN-γAptamerGrapheneArtificial tears2.75 pM/2.89 pM[279]
CRP FabSiNWBuffer solution (PBS)0.6 μg/mL[250]
CRPAntibodyCNTBuffer solution (PBS)0.06 μg/mL[35]
Cystatin C Papain/AuNPsLIGUrine samples0.05 ag/μL[30]
DNADNA probe MoS2/grapheneBuffer solution (PBS)10 aM[280]
Germs Infection S. aureus and S. epidermidisFibronectinSAMsBuffer solution (PBS)9 × 105 CFU/mL[56]
S. aureusPolystyrene nanospheresAu nanoporous structureBuffer solution (PBS)1 pM[269]
S. aureusAntibodySWCNTBuffer solution (PBS)150 CFU/mL[109]
GPB and GNB AntibioticsGrapheneBacterial sample1–9 CFU/mL[268]
S. aureusRCDSWCNTBuffer solution (PBS)1 CFU/mL[270]
PBA, phenylboronic acid; PEG, polyethylene glycol; MIP, molecularly imprinted polymer; MWCNT, multi-walled carbon nanotube; SWCNT, single-wall carbon nanotube; CRP, C-reactive protein; LIG, laser-induced graphene; SAMs, self-assembled monolayers; GPB, Gram-positive bacterial; GNB, Gram-negative bacterial; and RCD, RNA-cleaving DNAzyme.
Table 5. Comparison and solutions to overcome technical barriers.
Table 5. Comparison and solutions to overcome technical barriers.
PrincipleSolutionsAdvantagesDisadvantages
Loading matrixMicrobeadsreplaceablePartial depletion of enzyme
AuNPsImproved stability/catalytic efficacyHigh cost for mass-manufacturing
Nafion/silicaliteImproved stabilitySubstrate diffusion affected/enzyme activity affected due to pH
Artificial membranePolymer membrane (e.g., MIPs, PSMA)Higher sensitivity/specificity/reduced ion screening effectBiocompatibility concerns/complex synthesis
Reduction in non-specific signalsDilutionEasy to manipulateAffinity of biomolecules/stability/sensitivity might be affected
Polymer nanofilterReduced non-specific signalsConcerns of membrane thickness/size/orientation control/poor reproducibility
Larger reaction areaPorous structureMore binding sites/improved sensitivity
Nanowires/nanoribbon. etc.Fit in multichannel design/higher sensitivity due to larger surface
Improved stabilityExtended-gate designImproved stability and reusabilityPotential signal loss/complex integration
Deposition technology (e.g., ALD)Optimized thickness/morphology control/interface qualityHigher cost and complexity/slow deposition speed
Passivation/protection layer (e.g., rGO, HfO2, and AI2O3)Improved stability and biofunctionalization/high dielectric constant/mature technologyNon-uniform deposition/dielectric and interface trade-offs/thickness control
AuNPs, gold nanoparticles; MIPs, molecularly imprinted polymers; PSMA, poly (styrene-co-methacrylic acid); ALD, atomic layer deposition; and rGO, reduced graphene oxide.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Wang, Q.; Zhao, Z.-A.; Yao, K.-Y.; Cheng, Y.-L.; Wong, D.S.-H.; Wong, D.W.-C.; Cheung, J.C.-W. The Versatility of Biological Field-Effect Transistor-Based Biosensors (BioFETs) in Point-of-Care Diagnostics: Applications and Future Directions for Peritoneal Dialysis Monitoring. Biosensors 2025, 15, 193. https://doi.org/10.3390/bios15030193

AMA Style

Wang Q, Zhao Z-A, Yao K-Y, Cheng Y-L, Wong DS-H, Wong DW-C, Cheung JC-W. The Versatility of Biological Field-Effect Transistor-Based Biosensors (BioFETs) in Point-of-Care Diagnostics: Applications and Future Directions for Peritoneal Dialysis Monitoring. Biosensors. 2025; 15(3):193. https://doi.org/10.3390/bios15030193

Chicago/Turabian Style

Wang, Quan, Zi-An Zhao, Ke-Yu Yao, Yuk-Lun Cheng, Dexter Siu-Hong Wong, Duo Wai-Chi Wong, and James Chung-Wai Cheung. 2025. "The Versatility of Biological Field-Effect Transistor-Based Biosensors (BioFETs) in Point-of-Care Diagnostics: Applications and Future Directions for Peritoneal Dialysis Monitoring" Biosensors 15, no. 3: 193. https://doi.org/10.3390/bios15030193

APA Style

Wang, Q., Zhao, Z.-A., Yao, K.-Y., Cheng, Y.-L., Wong, D. S.-H., Wong, D. W.-C., & Cheung, J. C.-W. (2025). The Versatility of Biological Field-Effect Transistor-Based Biosensors (BioFETs) in Point-of-Care Diagnostics: Applications and Future Directions for Peritoneal Dialysis Monitoring. Biosensors, 15(3), 193. https://doi.org/10.3390/bios15030193

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop