Next Article in Journal
The Formation of β-Strand Nine (β9) in the Folding and Insertion of BamA from an Unfolded Form into Lipid Bilayers
Previous Article in Journal
Incorporation of Functionalized Halloysite Nanotubes (HNTs) into Thin-Film Nanocomposite (TFN) Nanofiltration Membranes for Water Softening
Previous Article in Special Issue
Sr2Fe1.575Mo0.5O6-δ Promotes the Conversion of Methane to Ethylene and Ethane
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Design and Validation of an Experimental Setup for Evaluation of Gas Permeation in Ceramic Membranes

by
Sabrina G. M. Carvalho
1,2,*,
Eliana N. S. Muccillo
1 and
Reginaldo Muccillo
1
1
Center of Science and Technology of Materials, Energy and Nuclear Research Institute, Cidade Universitária, Av. Prof. Lineu Prestes, 2242, São Paulo 05508-000, SP, Brazil
2
Institute of Physics, University of São Paulo, São Paulo 05508-090, SP, Brazil
*
Author to whom correspondence should be addressed.
Membranes 2023, 13(2), 246; https://doi.org/10.3390/membranes13020246
Submission received: 17 November 2022 / Revised: 6 February 2023 / Accepted: 14 February 2023 / Published: 18 February 2023
(This article belongs to the Special Issue Development and Application of Ceramic Membranes)

Abstract

:
An experimental setup for the evaluation of permeation of gaseous species with the possibility of simultaneously collecting electrochemical impedance spectroscopy data in disk-shaped ceramic membranes was designed and assembled. It consists of an alumina sample holder with thermocouple tips and platinum electrodes located close to both sides of the sample. Water-cooled inlet and outlet gas connections allowed for the insertion of the sample chamber into a programmable split tubular furnace. Gas permeation through a ceramic membrane can be monitored with mass flow controllers, a mass spectrometer, and an electrochemical impedance analyzer. For testing and data validation, ceramic composite membranes were prepared with the infiltration of molten eutectic compositions of alkali salts (lithium, sodium, and potassium carbonates) into porous gadolinia-doped ceria. Values of the alkali salt melting points and the permeation rates of carbon dioxide, in agreement with reported data, were successfully collected.

1. Introduction

1.1. Membranes

Ceramic membranes are important components of devices for the separation/production of gaseous species, such as oxygen, hydrogen, and carbon dioxide. There is a continuous demand in novel and efficient devices for oxygen production for the application in several industrial sectors, for hydrogen to be used in clean energy production, and for the capture and separation of carbon dioxide to achieve a pollution-free environment. The electroceramics that play an important role for the success of producing those gases are oxygen ion-, proton-, and carbon dioxide-ion conductors.

1.1.1. Membranes for Oxygen

Oxygen is an important commodity due to its frequent use in several industrial sectors, including in food, agriculture, medical, pharmaceutical, and metallurgical industries. Many efforts have been carried out on applied research looking for promising techniques for its production. Ceramic dual-phase composite membranes with suitable performance and stability may be used for the separation of oxygen from air, with a good selectivity for producing a high purity (>99.99%). A thorough review of all developed ceramic membranes, from the basic properties to the industrial application, may be found in reference [1].

1.1.2. Membranes for Hydrogen

Hydrogen has attracted attention, mainly for its use in renewable and clean energy sources. One of the sources of its production is a mixed-conducting ceramic membrane. Water is introduced to the ceramic surface and, after dissociation, oxygen is removed, resulting in a flow of hydrogen-enriched carrier gas [2]. Low-coast proton and electron conducting dual-phase composite membranes, besides presenting high selectivity to hydrogen, show high stability [3].

1.1.3. Membranes for Carbon Dioxide

One of the main sources of greenhouse gas emissions, particularly carbon dioxide, is the combustion of fossil fuels. One way to minimize this effect, while the non-pollutant generation of electricity and conversion of energy are not cost-effectively applicable, is carbon capture by the separation of carbon dioxide emitted from power plants fired with coal or natural gas, from industrial plants, and from non-electrical cars and trucks [4,5].
Ceramic membranes have been widely proposed for carbon dioxide separation processes [6,7,8,9,10,11,12]. They are two-phase composite membranes: a matrix consisting of porous oxygen ion conductor (e.g., gadolinia-doped ceria, samaria-doped ceria, yttria-stabilized zirconia) impregnated with a eutectic composition of alkali salts (Li-Na, Li-K, Na-K, and Li-Na-K carbonates), with 501, 498, 710, and 397 °C melting points, respectively [13]. Those membranes operate at the melting point of the eutectic composition for long times with reasonable carbon dioxide flow rates, ranging from 0.1 to 1.3 mL cm−2 min−1, depending on the physical characteristics of the membrane, the measured temperature, and the atmosphere of the experiments [14,15,16,17,18,19].

1.2. Carbon Dioxide Permeation

Permeation is an important parameter for evaluating the performance of a ceramic membrane. One of the requirements for designing ceramic membranes for carbon dioxide separation is a relatively high permeation of CO32− ions. The permeation mechanism is based on the surface reaction between the CO2 from the feed side and oxide anions, O2−, from the solid oxide (SO) ceramic phase to form CO32−. The carbonate ion is transported through the molten carbonate (MC) phase to the other side of the membrane, according to the reaction below:
CO2 (gas) + O2− (SO) Membranes 13 00246 i001 CO32− (MC)
On the permeated side of the membrane, with low CO2 partial pressure, the reverse reaction occurs, releasing CO2 back into the gas phase. This reaction also releases the oxide anion in the reverse path through the ceramic phase, to the feed side of the membrane with a high CO2 partial pressure, to restart the reaction [20,21]. For the carbon dioxide to be transported through the membrane, the flow of carbonate ions needs to be balanced by a counter flow of O2− ions. If one of the ionic phase resistances is higher than the other, the membrane efficiency will be limited by the most resistive phase.
The permeation flux (J) is defined as the volume of gas flowing through the membrane per unit area and unit time (mL cm−2 min−1). Permeability (mol m−1 s−1 Pa−1) is the permeation flux normalized by the partial pressure difference of the permeated gas and membrane thickness, while permeance (mol m−2 s−1 Pa−1) is normalized only for the partial pressure driving force [14,20]. Both parameters can be used to compare membrane performances at a given temperature, as they are an intrinsic characteristic of the membrane.
The CO2 flux depends strongly on the CO32− ionic conductivity through the molten carbonate mixture and the O2− ionic conductivity through the ceramic phase. A commonly used molten carbonate mixture (400 °C melting point) is the ternary eutectic carbonate (Li-Na-K)2CO3 with a 43.5:31.5:25 molar ratio concentration [14,18,20,22,23,24,25,26].
The most common experimental setup for measuring CO2 permeation flux consists of a membrane fixed to an alumina tube, through which the feed gas is inserted and delivered to the sample [10,12,15,16,18,20,26]. This tube-sample system is placed inside another alumina or quartz tube with a thermocouple fixed to monitor the temperature and the sweep gas carrying the permeated CO2 to the gas analyzer, usually a mass spectrometer or a gas chromatograph.
Here, we show details of an experimental setup designed for the qualitative and quantitative measurement of the content of gaseous species that flow through single ceramic membranes, together with the possibility of simultaneously collecting data on the electrochemical impedance spectroscopy of the membrane. The analysis of the electrical properties during permeation provides information on the ionic conductivity and transport of charged species of the membrane under study. The performance of the experimental setup was ascertained by monitoring carbon dioxide permeation through gadolinium-doped ceria/lithium-sodium carbonates (GDC-LNC) dual phase composite membranes.

2. Materials and Methods

2.1. Membrane Preparation

For testing the performance of the experimental setup, two ceramic compositions were prepared: (a) alumina (α-Al2O3, Alcoa, Pittsburgh, PA, USA, 0.6 μm average particle size, 6.5 m2 g−1 specific surface area), cold pressed uniaxially at 10 MPa and isostatically at 100 MPa, followed by sintering at 1600 °C/3 h, and this dense alumina sample was used to evaluate and validate the sealing process; and (b) CeO2: 20 mol.% Gd2O3-Li2CO3/Na2CO3 (gadolinia-doped ceria, GDC-eutectic lithium sodium carbonate, LNC) membrane by the infiltration of molten LNC into a porous GDC ceramic pellet. The porous 20GDC (Ceramic Powder Technology AS, Tiller, Norway) matrix was prepared by tape casting using 10 wt.% of rice starch (Remy Ind., Leuven, Belgium) as a pore former and sintered at 1450 °C/5 h with low heating rates to remove all organic additives. A eutectic mixture of sodium and lithium carbonates (99.99%, Alfa Aesar, Tewksbury, MA, USA) was prepared by mixing the powders 52 mol.% Li2CO3 with 48 mol.% Na2CO3 and infiltrating it to the porous membrane at 600 °C/1 h. That membrane was attached to the alumina tube of the experimental setup to evaluate the CO2 permeation at different temperatures. The leakage test, using the alumina sample, was performed at 550, 610, and 665 °C, and the permeation measurement was performed at 390, 600, and 705 °C, both experiments were carried out by injecting 50% Ar-50% CO2 as feed gas and N2 as the sweep gas.

2.2. Experimental Setup

Figure 1 shows a drawing of the components of the experimental setup used for the analysis of CO2 pressure gradient-assisted permeation through a ceramic membrane. The setup consists basically of a three-atmosphere sample chamber (one gas at each side of the ceramic membrane, another gas for sweeping out any leakage at the membrane sealing cement), mass flow controllers, and a mass spectrometer. A ceramic paste containing aluminum oxide (58 wt.%), titanium dioxide (7 wt.%), and calcium oxide (35 wt.%) was mixed with water and used to attach the surface of a disk-shape membrane (see Figure 1) to the end of two alumina tubular pieces positioned in the center of a programmable split tubular furnace. After curing the ceramic sealant at 120 °C for 3 h, a selected mixture of carbon dioxide and argon (feed gas) is injected to the surface of the membrane through a 1/2” alumina tube. A similar 1/2” alumina tube is used for introducing nitrogen (sweep gas) to the other surface of the membrane. The gas permeated to the other side of the membrane is quantitatively and qualitatively analyzed with a mass spectrometer (Thermostar®, Pfeiffer Vacuum, Germany). Controlled fluxes of gases are injected with mass flow controllers (MKS Instruments, Inc., Andover, MA, USA).
Figure 2 shows details of input and output gases and water (for cooling the stainless-steel flanges) and a picture of the sample chamber positioned inside a split tubular furnace.
Two thermocouples are positioned close to each parallel face of the membrane to accurately monitor the temperature during permeation experiments. The feed gas is introduced into the sample chamber at the right side (see Figure 2, top) and is delivered close to the membrane surface through a thin alumina tube. Excess feed gas is exhausted through the outlet, labeled retentate. The sweep gas inlet is located on the opposite side, which also delivers gas close to the sample surface and carries the permeated gas to the mass spectrometer through the permeate outlet. The figure also shows the inlet and outlet of a third gas/gas mixture, identified as carrier gas, which can be used to support the possible removal of gas leakage.
The sample chamber is provided with Pt terminal leads for connecting both sides of the ceramic membrane to a 4192A Hewlett-Packard impedance analyzer (Yokogawa-Hewlett-Packard, Tokyo, Japan) connected to a series 360 Hewlett-Packard controller to collect electrical impedance spectroscopy data Z(ω) in the 10–107 Hz frequency range. Z(ω) = [−Z″(ω) × Z′(ω)], where Z′ and Z″ are the real and the imaginary components of the electrical impedance; ω = 2 π f, f standing for the frequency of the input signal; Z(ω) is deconvoluted for the evaluation of the total electric resistivity with a special software [27]. Simultaneous electrochemical impedance spectroscopy and permeation data can be collected, making it possible to analyze, in situ, the membrane ionic conductivity, one of the parameters responsible for the efficiency of the membrane. Depending on the temperature of the measurement, the electrical resistivity of bulk and interfaces (grain boundary, second phases, and pores) of the ceramic membrane could be evaluated. The total electrical resistivity values, i.e., the sum of the electrical resistivity of bulk and interfaces, were used to obtain the Arrhenius plot, which is useful for determining the thermal activation energy of the solid electrolyte (membrane matrix) and the impregnated second phase (carbonates).

3. Results and Discussion

Before performing the permeation measurements, a leakage test was carried out using an alumina pellet instead of a ceramic membrane. Figure 3 shows a picture of a disk-shape dense alumina ceramic piece attached with cement to the alumina tube of the sample holder in the sample chamber.
The dense alumina pellet was prepared to evaluate possible gas leakage through the ceramic sealant. For this test, an equal flux of carbon dioxide and argon was used as a feed gas, and nitrogen as a sweep gas, both at a 100 mL min−1 flow rate (fgas). The permeated gas (Cf. Figure 1) was analyzed with the mass spectrometer. The Ar, CO2, and N2 fluxes were evaluated at 550, 610, and 665 °C. These temperature values were selected for being close to the ones used during the permeation experiments (above 500 °C), so it was possible to analyze not only the amount of gas leaked, but also the proportion of the chemical species with increasing temperature. Figure 4 shows these results. Increasing the temperature leads to an increase in the mobility of CO2 (3.04 to 3.16 to 3.30%, respectively) and Ar (3.68 to 3.82 to 4.0%), increasing their content, as expected. The sweeping gas, N2, decreases from 93.3 to 93.0 to 92.7%, accordingly (Table 1).
Considering that the percentage of N2, detected by the mass spectrometer, represents 100 mL min−1 (same amount injected at the sweep side), Table 1 shows the measured average fluxes for CO2 and Ar collected at different temperatures.
An enhancement of Ar and CO2 contents is evaluated for increasing temperature. It is worth noting that, although the same amounts of CO2 and Ar were used as feed gas, the leakage of these two gases was not similar; these values should then be taken into account when analyzing the data collected after the permeation experiments. From Table 1 data, the evaluation of the ratio of CO2/Ar leakage to the permeated side yields close to 0.83 and does not change considerably with temperature.
After the gas leakage experiment, the three-atmosphere sample chamber was used to measure the CO2 permeation through a dual-phase ceramic-carbonate membrane (GDC-LNC). The temperature of the furnace was programmed to three temperature stages, the first one below the melting point of the carbonate mixture to evaluate sealing behavior of the ceramic glue, and the other two temperatures above 500 °C to measure the effective permeation; the feed gas was a mixture of 50% CO2 and 50% Ar, and N2 was the sweep gas, both at a 100 mL min−1 flow rate (fgas).
Figure 5 shows data of the CO2, Ar, and N2 concentrations measured with the mass spectrometer at the permeated side of the GDC-LNC membrane.
The measurement started at a temperature below the melting point of Li-Na carbonates, showing a significant increase in the contents of CO2 and Ar when the temperature of the membrane exceeds 500 °C. This effect is observable, assuming there is a chemical interaction of the ceramic sealant with the molten carbonates. Small portions of the carbonate end up being absorbed by the sealant, leading to self-healing of the gas leakage regions in the bulk of the sealant; this is favorable for the permeation process, but can also decrease the efficiency of the membrane, due to a decrease in the percentage of the carbonate phase in the bulk of the membrane.
There is a significant increment in the detection of CO2 and Ar for increasing the temperature to 700 °C. That increase shows that gas leakage also increased, since the ceramic membrane is not permeable to argon. There was probably an accumulation of molten carbonate at the membrane-ceramic adhesion paste interface, promoting open porosity.
Table 2 shows data of the contents of CO2 and Ar measured at three different temperatures. An evaluation of those values, which were detected with the mass spectrometer, did not allow for discriminating the CO2 permeated content from the leaked content; therefore, it was necessary to use the detected argon to determine the effective permeation.
Since the permeation occurs only at temperatures higher than the melting point of the Li-Na carbonate, the value obtained at 390 °C was used to determine the ratio (R) between CO2 and Ar leaked contents. This ratio is important to evaluate the fraction of CO2 detected with the mass spectrometer that was obtained through permeation, and the fraction resulting from the leakage. Since the ratio does not change considerably in this temperature range (as shown in the leakage test), the value detected at 390 °C was used to calculate the permeation above 500 °C.
The fraction (P) of CO2, Ar, and N2 detected at the permeated side and the total flow of the carrier gas (fgas) were used in the following equations to evaluate the permeation flux (J) of CO2 in mL cm−2 min−1 through the membrane [19]:
J C O 2 t o t a l = P C O 2 · f g a s P N 2 ·   S
J C O 2 l e a k = R · P A r · f g a s P N 2 · S
J C O 2   p e r m e a t e d = J C O 2 t o t a l J C O 2 l e a k
R stands for the ratio between the CO2 and Ar concentrations detected below 500 °C at the permeated side, and S stands for the surface area of the membrane in contact with the feed gas. After using the above equations for the initial data obtained during the permeation experiments (Figure 5), the permeated CO2 could then be evaluated at 660 °C and 705 °C, Figure 6.
There is a negligible decrease in the permeation value during the isothermal stages, probably related to the interaction between the sealant and the molten carbonate, assumed above. Considering the active area of the membrane, the permeation flux of the membrane was estimated as 0.49 mL min−1 cm−2 and normalizing the partial pressure of CO2 on each side, the permeance of the membrane at 705 °C was estimated as 8.58 × 10−8 mol m−2 s−1 Pa−1.
Table 3 shows permeance data of some membranes with oxygen ion conduction under different experimental conditions, e.g., feed gas and operation temperature. The data obtained with our experimental setup are in the same range of results found in the literature, within 10−8–10−7 mol m−2 s−1 Pa−1.
Impedance spectroscopy measurements were performed on porous and infiltrated GDC membranes at temperatures below and above the melting point of the carbonate mixture to evaluate the electrical resistivity of the membrane over a wide temperature range. Figure 7 shows the impedance diagrams.
The electrical data, represented by impedance spectroscopy plots, show the resistive and capacitive behavior of porous CeO2: 20 mol.% Gd2O3 ceramic matrix (Figure 7a) and of the ceramic composite (porous matrix infiltrated with 52 mol.% Li2CO3/48 mol.% Na2CO3) (Figure 7b) at 450 °C and 580 °C, below and above the melting point of the eutectic mixture of lithium and sodium carbonates, respectively. The total electrical resistivity is higher in porous ceramics, as expected [32]. For the composite membrane, the contribution of the CO32− ions is predominant at 575 °C, Figure 7b.
Figure 8 shows Arrhenius plots of the total electric conductivity of porous gadolinia-doped ceria (GDC) before and after infiltration with the eutectic composition of lithium and sodium carbonates. The conductivity data of the porous membrane exhibits a linear behavior for the entire frequency range, as expected, with activation energy of 82.0 kJ mol−1, highlighting the conduction of oxide ions [33]. The sudden increase of the electrical conductivity of infiltrated membranes at 500 °C represents the contribution of the carbonate ion of the molten carbonate phase to the total conductivity of the composite ceramic pellet [13,27].
There is a significant increase in the total ionic conductivity of the infiltrated membrane (red dots) when the temperature approaches 500 °C; this was expected and occurs due to the fast increase of the ionic conductivity of the carbonate phase as a result of its melting. The thermal activation energy of the electrical conductivity for temperatures higher than the melting point of the carbonates was evaluated as 24 kJ mol−1. This value is typical for molten carbonates; the reported value for the activation energy of the conductivity of a eutectic mixture of lithium-sodium carbonates is ~19 kJ mol−1 [34]. The phase change measured near 500 °C sets the beginning of CO2 ionic transport through the membrane.
The total conductivity of the infiltrated membrane above 500 °C shows little variation since the conductivity of molten carbonate is nearly constant with increasing temperature. However, the total electrical conductivity of the ceramic phase (blue circles) keeps its linear increase during the entire temperature range. Although the increase in the electrical conductivity generated by the ceramic phase of the membrane is negligible, it is very important for the simultaneous ionic current of CO32− ions (through the molten carbonate phase) and O2− ions (through the ceramic phase).
The knowledge of the electrical behavior of membranes contributes significantly to improving the design of the experimental setup, allowing for the collection of electrical data simultaneously with gas permeation.

4. Conclusions

An experimental setup was projected and assembled using an alumina sample chamber connected to mass flow controllers, a mass spectrometer, and an impedance analyzer. The sample chamber, capable of monitoring the temperature on both sides of the membrane and in situ electrochemical impedance spectroscopy measurements, was inserted in a split tubular furnace for high temperature testing of gas leakage and carbon dioxide ion permeation through a CeO2: 20 mol.% Gd2O3 porous ceramic infiltrated with a (Li,Na)2CO3 eutectic composition. The evaluated CO2 permeability data were in good agreement with data collected in the scientific literature. Impedance spectroscopy data were collected to evaluate the electrical conductivity of mobile ions at temperatures below and above the melting point of the alkali carbonates. The setup could also be used to evaluate the permeation of different gaseous species through other ceramic membranes with simultaneous monitoring electrochemical impedance spectroscopy behavior.

Author Contributions

Conceptualization, R.M.; methodology, S.G.M.C.; data collection, S.G.M.C.; writing—original draft preparation, S.G.M.C.; writing—review and editing, E.N.S.M. and R.M.; supervision, R.M. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by CNEN, CINE-SHELL (ANP) (FAPESP Proc. 2017/11937-4), CDMF-CEPID (FAPESP Proc. 2013/07296-2), FAPESP 2020/05250-9, and CNPq (Procs. 402966/2021-0, 302357/2018-1, 305889/2018-4).

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Kiebach, R.; Pirou, S.; Aguilera, L.M.; Haugen, A.B.; Kaiser, A.; Hendriksen, P.V.; Balaguer, M.; Garciıa-Fayos, J.; Serra, J.M.; Schulze-Küppers, F.; et al. A review on dual-phase oxygen transport membranes: From fundamentals to commercial deployment. J. Mater. Chem. A 2022, 10, 2152–2195. [Google Scholar] [CrossRef]
  2. Balachandran, U.; Lee, T.H.; Dorris, S.E. Hydrogen production by water dissociation using mixed conducting dense ceramic membranes. Int. J. Hydrogen Energy 2007, 32, 451–456. [Google Scholar] [CrossRef]
  3. Cheng, H. Dual-Phase Mixed Protonic-Electronic Conducting Hydrogen Separation Membranes: A Review. Membranes 2022, 12, 647. [Google Scholar] [CrossRef]
  4. Rochelle, G.T. Amine scrubbing for CO2 capture. Science 2009, 325, 1652–1654. [Google Scholar] [CrossRef]
  5. Ozkan, M.; Custelcean, R. The status and prospects of materials for carbon capture technologies. MRS Bull. 2022, 47, 390–394. [Google Scholar] [CrossRef]
  6. Phair, J.W.; Badwal, S.P.S. Materials for separation membranes in hydrogen and oxygen production and future power generation. Sci. Technol. Adv. Mater. 2006, 7, 792–805. [Google Scholar] [CrossRef]
  7. Rondão, A.I.B.; Patrício, S.G.; Figueiredo, F.M.L.; Marques, F.M.B. Impact of ceramic matrix functionality on composite electrolytes performance. Electrochim. Acta 2013, 109, 701–709. [Google Scholar] [CrossRef]
  8. Kenarsari, S.D.; Yang, D.; Jiang, G.; Zhang, S.; Wang, J.; Russell, A.G.; Wei, Q.; Fan, M. Review of recent advances in carbon dioxide separation and capture. RSC Adv. 2013, 3, 22739–22773. [Google Scholar] [CrossRef]
  9. Pera-Titus, M. Porous inorganic membranes for CO2 capture: Present and prospects. Chem. Rev. 2014, 114, 1413–1492. [Google Scholar] [CrossRef]
  10. Patrício, S.G.; Papaioannou, E.I.; Ray, B.M.; Metcalfe, I.S.; Marques, F.M.B. Composite CO2 separation membranes: Insights on kinetics and stability. J. Membr. Sci. 2017, 541, 253–261. [Google Scholar] [CrossRef] [Green Version]
  11. Zhang, A.P.; Tong, J.; Huang, K.; Zhu, X.; Yang, W. The current status of high temperature electrochemistry-based CO2 transport membranes and reactors for direct CO2 capture and conversion. Progr. Energy Combust. Sci. 2021, 82, 100888. [Google Scholar] [CrossRef]
  12. Starykevich, M.; Jamale, A.; Yasakau, K.A.; Marques, F.M.B. Novel molten phase route for composite CO2 separation membranes. J. Membr. Sci. 2022, 659, 120806. [Google Scholar] [CrossRef]
  13. Frangini, S.; Masi, A. Molten carbonates for advanced and sustainable energy applications: Part I. Revisiting molten carbonate properties from a sustainable viewpoint. Int. J. Hydrogen Energy 2016, 41, 18739–18746. [Google Scholar] [CrossRef]
  14. Li, Y.D.; Rui, Z.B.; Xia, C.; Anderson, M.; Lin, Y.S. Performance of ionic- conducting ceramic/carbonate composite material as solid oxide fuel cell electrolyte and CO2 permeation membrane. Catal. Today 2009, 148, 303–309. [Google Scholar] [CrossRef]
  15. Anderson, M.; Lin, Y.S. Carbonate-ceramic dual-phase membrane for carbon dioxide separation. J. Membr. Sci. 2010, 357, 122–129. [Google Scholar] [CrossRef]
  16. Patricio, S.G.; Papaioannou, E.I.; Zhang, G.; Metcalfe, I.S.; Marques, F.M.B. High performance composite CO2 separation membranes. J. Membr. Sci. 2014, 471, 211–218. [Google Scholar] [CrossRef] [Green Version]
  17. Papaioannou, E.I.; Qi, H.; Metcalfe, I.S. ‘Uphill’ permeation of carbon dioxide across a composite molten salt-ceramic membrane. J. Membr. Sci. 2015, 485, 87–93. [Google Scholar] [CrossRef] [Green Version]
  18. Kazakli, M.; Mutch, G.A.; Triantafyllou, G.; Gil, A.G.; Li, T.; Wang, B.; Bailey, J.J.; Brett, D.J.L.; Shearing, P.R.; Li, K.; et al. Controlling molten carbonate distribution in dual-phase molten salt-ceramic membranes to increase carbon dioxide permeation rates. J. Membr. Sci. 2021, 617, 118640. [Google Scholar] [CrossRef]
  19. Carvalho, S.G.M.; Muccillo, E.N.S.; Fonseca, F.C.; Müller, M.; Schulze-Küppers, F.; Baumann, S.; Meulenberg, W.A.; Guillon, O.; Muccillo, R. Tape-casting and freeze drying gadolinia-doped ceria composite membranes for carbon dioxide permeation. J. Membr. Sci. 2022, 648, 120355. [Google Scholar] [CrossRef]
  20. Wade, J.L.; Lee, C.; West, A.C.; Lackner, K.S. Composite electrolyte membranes for high temperature CO2 separation. J. Membr. Sci. 2011, 369, 20–29. [Google Scholar] [CrossRef]
  21. Mutch, G.A.; Qu, L.; Triantafyllou, G.; Xing, W.; Fontaine, M.L.; Metcalfe, I.S. Supported molten-salt membranes for carbon dioxide permeation. J. Mater. Chem. 2019, 7, 12951–12973. [Google Scholar] [CrossRef]
  22. Xia, C.; Li, Y.; Tian, Y.; Liu, Q.; Zhao, Y.; Jia, L.; Li, Y. A high performance composite ionic conducting electrolyte for intermediate temperature fuel cell and evidence for ternary ionic conduction. J. Power Source 2009, 188, 156–162. [Google Scholar] [CrossRef]
  23. Norton, T.T.; Lu, B.; Lin, Y.S. Carbon dioxide permeation properties and stability of samarium-doped-ceria carbonate dual-phase membranes. J. Membr. Sci. 2014, 467, 244–252. [Google Scholar] [CrossRef]
  24. Lu, B.; Lin, Y.S. Asymmetric thin samarium doped cerium oxide-carbonate dual-phase membrane for carbon dioxide separation. Ind. Eng. Chem. Res. 2014, 53, 13459–13466. [Google Scholar] [CrossRef]
  25. Nikolaeva, E.V.; Bovet, A.L.; Zakiryanova, I.D. Electrical conductivity of molten carbonate and carbonate-chloride systems coexisting with aluminum oxide powder. Z. Naturforsch. 2018, 73, 79–83. [Google Scholar] [CrossRef]
  26. Grilo, J.P.F.; Macedo, D.A.; Nascimento, R.M.; Marques, F.M.B. Electronic conductivity in Gd-doped ceria with salt additions. Electrochim. Acta 2019, 318, 977–988. [Google Scholar] [CrossRef]
  27. Kleitz, M.; Kennedy, J.H. Resolution of multicomponent impedance diagrams. In Fast Ion Transport in Solids, Electrodes and Electrolytes; Mundy, J.N., Shenoy, G.K., Vashishta, P., Eds.; Elsevier: Amsterdam, The Netherlands, 1979; pp. 185–188. [Google Scholar]
  28. Chen, T.; Wang, Z.; Hu, J.; Wai, M.H.; Kawi, S.; Lin, Y.S. High CO2 permeability of ceramic-carbonate dual-phase hollow fiber membrane at medium-high temperature. J. Membr. Sci. 2020, 597, 117770. [Google Scholar] [CrossRef]
  29. Rui, Z.; Anderson, M.; Li, Y.; Lin, Y.S. Ionic conducting ceramic and carbonate dual phase membranes for carbon dioxide separation. J. Membr. Sci. 2012, 417–418, 174–182. [Google Scholar] [CrossRef]
  30. Chen, T.; Wu, H.C.; Li, Y.; Lin, Y.S. Poisoning effect of H2S on CO2 permeation of samarium-doped-ceria/carbonate dual-phase membrane. Ind. Eng. Chem. Res. 2017, 56, 14662–14669. [Google Scholar] [CrossRef]
  31. Dong, X.; Wu, H.C.; Lin, Y.S. CO2 permeation through asymmetric thin tubular ceramic-carbonate dual-phase membranes. J. Membr. Sci. 2018, 564, 73–81. [Google Scholar] [CrossRef]
  32. Muccillo, R. Impedance spectroscopy analysis of zirconia:8 mol% yttria solid electrolytes with graphite pore former. J. Mater. Res. 2022, 24, 1780–1784. [Google Scholar] [CrossRef]
  33. Horovistiz, A.L.; Rocha, R.A.; Muccillo, E.N.S. Electrical conductivity and microstructure relationship in ternary systems based on cerium oxide. Ceram. Int. 2013, 39, 5887–5892. [Google Scholar] [CrossRef]
  34. Kojima, T.; Miyazaki, Y.; Nomura, K.; Tanimoto, K. Electrical conductivity of molten Li2CO3–X2CO3 (X: Na, K, Rb, and Cs) and Na2CO3–Z2CO3 (Z: K, Rb, and Cs). J. Electrochem. Soc. 2007, 154, F222. [Google Scholar] [CrossRef]
Figure 1. Overall structure of the components of the whole experimental setup for analysis of CO2 permeation through a ceramic membrane.
Figure 1. Overall structure of the components of the whole experimental setup for analysis of CO2 permeation through a ceramic membrane.
Membranes 13 00246 g001
Figure 2. Top: schematics of the sample chamber for measuring gas permeation in a ceramic membrane; bottom: picture of the sample chamber positioned inside the split furnace.
Figure 2. Top: schematics of the sample chamber for measuring gas permeation in a ceramic membrane; bottom: picture of the sample chamber positioned inside the split furnace.
Membranes 13 00246 g002
Figure 3. Picture of the dense alumina piece (white) cemented on top of the alumina tube of the sample chamber.
Figure 3. Picture of the dense alumina piece (white) cemented on top of the alumina tube of the sample chamber.
Membranes 13 00246 g003
Figure 4. CO2, Ar, and N2 concentrations as a function of time at the permeated side of the setup shown in Figure 1. The colored areas represent the dwell time of the membrane temperature.
Figure 4. CO2, Ar, and N2 concentrations as a function of time at the permeated side of the setup shown in Figure 1. The colored areas represent the dwell time of the membrane temperature.
Membranes 13 00246 g004
Figure 5. Percentage amounts of CO2, Ar, and N2 at the permeated side of the GDC-LNC ceramic membrane during a permeation experiment. The colored areas represent the period of membrane temperature stability and the vertical dotted line points to the time the membrane reaches 500 °C.
Figure 5. Percentage amounts of CO2, Ar, and N2 at the permeated side of the GDC-LNC ceramic membrane during a permeation experiment. The colored areas represent the period of membrane temperature stability and the vertical dotted line points to the time the membrane reaches 500 °C.
Membranes 13 00246 g005
Figure 6. CO2 permeation flux of dense dual-phase GDC-LNC membrane with tape-casted porous support. The colored areas represent the period of membrane temperature stability.
Figure 6. CO2 permeation flux of dense dual-phase GDC-LNC membrane with tape-casted porous support. The colored areas represent the period of membrane temperature stability.
Membranes 13 00246 g006
Figure 7. Impedance spectroscopy diagrams of CeO2: 20 mol.% Gd2O3 (a) porous and (b) infiltrated with 52 mol.% Li2CO3/48 mol.% Na2CO3, measured at temperatures below (450 °C) and above (575 °C) the melting point of 52 mol.% Li2CO3/48 mol.% Na2CO3. Numbers stand for log f (f:Hz).
Figure 7. Impedance spectroscopy diagrams of CeO2: 20 mol.% Gd2O3 (a) porous and (b) infiltrated with 52 mol.% Li2CO3/48 mol.% Na2CO3, measured at temperatures below (450 °C) and above (575 °C) the melting point of 52 mol.% Li2CO3/48 mol.% Na2CO3. Numbers stand for log f (f:Hz).
Membranes 13 00246 g007
Figure 8. Arrhenius plots of the electrical conductivity of CeO2: 20 mol.% Gd2O3 porous (blue) and infiltrated (red) with 52 mol.% Li2CO3/48 mol.% Na2CO3.
Figure 8. Arrhenius plots of the electrical conductivity of CeO2: 20 mol.% Gd2O3 porous (blue) and infiltrated (red) with 52 mol.% Li2CO3/48 mol.% Na2CO3.
Membranes 13 00246 g008
Table 1. Gas fluxes measured with a mass spectrometer during the leakage test.
Table 1. Gas fluxes measured with a mass spectrometer during the leakage test.
Temperature (°C)Gas Flux (mL min−1)
CO2Ar
5503.23.9
6103.44.1
6653.64.3
Table 2. Gas fluxes measured with a mass spectrometer during the permeation test.
Table 2. Gas fluxes measured with a mass spectrometer during the permeation test.
Temperature (°C)Gas Flux (mL min−1)
CO2Ar
3905.16.0
6004.65.4
7057.58.3
Table 3. Permeance data obtained after testing several membranes (GDC: gadolinia-doped ceria; SDC: samaria-doped ceria; LSCF: lanthanum strontium cobalt ferrite; BYS: bismuth yttrium samarium oxide). Adapted from Ref. [19].
Table 3. Permeance data obtained after testing several membranes (GDC: gadolinia-doped ceria; SDC: samaria-doped ceria; LSCF: lanthanum strontium cobalt ferrite; BYS: bismuth yttrium samarium oxide). Adapted from Ref. [19].
Ceramic MatrixFeed GasTemperature (°C)Permeance (mol m−2 s−1 Pa−1)Ref.
GDC20%CO2/15%H2O/62.4%Ar/2.6% H2800 (heating)5.35 × 10−7[19]
800 (cooling)3.44 × 10−7
SDC5%H2/47.5%CO2/47.5%N27008.56 × 10−7[28]
LSCF50%CO2/50%Ar9005.36 × 10−8[15]
BYS50%CO2/50%Ar6501.10 × 10−8[29]
SDC10%H2/45%CO2/45%N27501.36 × 10−7[30]
SDC50%CO/35%CO2/10%H2/5%N29001.68 × 10−7[23]
SDC-BYS49.5%CO/36%CO2/4.5%N2/10%H27001.05 × 10−7[31]
GDC50%N2/50%CO26504.40 × 10−8[16]
GDC50% Ar-50% CO27058.58 × 10−8This work
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Carvalho, S.G.M.; Muccillo, E.N.S.; Muccillo, R. Design and Validation of an Experimental Setup for Evaluation of Gas Permeation in Ceramic Membranes. Membranes 2023, 13, 246. https://doi.org/10.3390/membranes13020246

AMA Style

Carvalho SGM, Muccillo ENS, Muccillo R. Design and Validation of an Experimental Setup for Evaluation of Gas Permeation in Ceramic Membranes. Membranes. 2023; 13(2):246. https://doi.org/10.3390/membranes13020246

Chicago/Turabian Style

Carvalho, Sabrina G. M., Eliana N. S. Muccillo, and Reginaldo Muccillo. 2023. "Design and Validation of an Experimental Setup for Evaluation of Gas Permeation in Ceramic Membranes" Membranes 13, no. 2: 246. https://doi.org/10.3390/membranes13020246

APA Style

Carvalho, S. G. M., Muccillo, E. N. S., & Muccillo, R. (2023). Design and Validation of an Experimental Setup for Evaluation of Gas Permeation in Ceramic Membranes. Membranes, 13(2), 246. https://doi.org/10.3390/membranes13020246

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop