A Novel Markerless Lung Tumor-Tracking Method Using Treatment MV Beam Imaging

1
Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75390, USA
2
Department of Computer Science, University of Texas at Dallas, Richardson, TX 75080, USA
3
Department of Radiation Oncology, Henry Ford Hospital, Detroit, MI 48202, USA
*
Author to whom correspondence should be addressed.
Received: 30 October 2018 / Revised: 27 November 2018 / Accepted: 3 December 2018 / Published: 6 December 2018
(This article belongs to the Section Applied Biosciences and Bioengineering)
Full-Text   |   PDF [2022 KB, uploaded 6 December 2018]   |  

Abstract

A novel method was developed to track lung tumor motion in real time during radiation therapy with the purpose to allow target radiation dose escalation while simultaneously reducing the dose to sensitive structures, thereby increasing local control without increasing toxicity. This method analyzes beam’s eye view radiation therapy treatment megavoltage (MV) images with simulated digitally reconstructed radiographs (DRRs) as references. Instead of comparing global DRRs with projection images, this method incorporates a technique that divides the global composite DRR and the corresponding MV projection into sub-images called tiles. Registration is performed independently on tile pairs in order to reduce the effects of global discrepancies due to scattering or imaging modality differences. This algorithm was evaluated by phantom studies while simulated tumors were controlled to move with various patterns in a complex humanoid torso. Approximately 15,000 phantom MV images were acquired at nine gantry angles, with different tumors moving within ranges between 10 and 20 mm. Tumors were successfully identified on every projection with a total maximum/average error of 1.84/0.98 mm. This algorithm was also applied to over 5,000 frames of MV projections acquired during radiation therapy of five lung cancer patients. This tumor-tracking methodology is capable of accurately locating lung tumors during treatment without implanting any internal fiducial markers nor delivering extra imaging radiation doses. View Full-Text
Keywords: tumor tracking; MV imaging; markerless tumor tracking; MV imaging; markerless
Figures

Figure 1

This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited (CC BY 4.0).
SciFeed

Share & Cite This Article

MDPI and ACS Style

Rozario, T.; Chiu, T.D.; Chen, M.; Jia, X.; Lu, W.; Bereg, S.; Mao, W. A Novel Markerless Lung Tumor-Tracking Method Using Treatment MV Beam Imaging. Appl. Sci. 2018, 8, 2525.

Show more citation formats Show less citations formats

Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Related Articles

Article Metrics

Article Access Statistics

1

Comments

[Return to top]
Appl. Sci. EISSN 2076-3417 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert
Back to Top