Next Article in Journal
Engineering Design and Evaluation of the Process Evaluation Method of Auto Repair Professional Training in Virtual Reality Environment
Previous Article in Journal
Special Issue on Requirements Engineering, Practice and Research
Previous Article in Special Issue
Field Experience for Determination of Formaldehyde in Stack Emissions
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Interlaboratory Comparison of 226Ra and 228Ra Activity Concentrations in Groundwater and Surface Water

Worthington Miller Environmental, LLC, 1027 W. Horsetooth Rd., Suite 200, Fort Collins, CO 80526, USA
Appl. Sci. 2022, 12(23), 12198; https://doi.org/10.3390/app122312198
Submission received: 10 November 2022 / Revised: 21 November 2022 / Accepted: 25 November 2022 / Published: 29 November 2022
(This article belongs to the Special Issue Monitoring and Analysis of Environmental Pollution)

Abstract

:
226Ra and 228Ra are typically monitored for groundwater and surface water compliance at legacy U mining and milling sites. Groundwater monitoring results for combined Ra (226Ra + 228Ra) reported by the existing contract laboratory at a former U mine and mill Site (Converse County, WY, USA) have been highly variable and with increasing trends at the Site compliance and background wells since the method was changed in 2005. Sample reanalysis has indicated poor reproducibility and significant analytical error in 228Ra measurements. An interlaboratory comparison was conducted to evaluate the potential causes of the high variability and analytical error. Two different methods were used for 226Ra (M903.0 and M903.1) and 228Ra (M904.0 and Ra-05). 226Ra results were less variable compared to 228Ra, and 228Ra data from the existing laboratory were qualified as estimated with high bias due to detection of 228Ra in the field blank. Compliance with the 226Ra + 228Ra groundwater standard was either met or not met, depending on which laboratory conducted the analyses. Specific laboratory techniques, rather than the analytical method, are contributing to elevated 228Ra values being reported. It was recommended that samples whose 226Ra + 228Ra results exceed the Site standard in the future be reanalyzed by the existing laboratory with a sample split also being sent to an outside laboratory for confirmatory analysis.

1. Introduction

An interlaboratory comparison (ILC) is an exercise carried out by a group of laboratories to compare their performance or to assess a measurement standard, and is typically used to (1) assess random variation in measurements across a population of laboratories, (2) determine the systematic differences in results among a fixed set of laboratories, or (3) determine the value of a physical property of an artifact or a population of artifacts [1]. An ILC was conducted to evaluate the variability of 226Ra and 228Ra activity concentrations (hereafter referred to as concentrations) in groundwater and an adjacent surface water impoundment (Impoundment) at a former U mine and mill (Site) located in Converse County, WY, USA. Assessment of 226Ra + 228Ra trends at the Site compliance well and selected groundwater monitoring wells has shown that 226Ra + 228Ra concentrations are highly variable, and there is some uncertainty regarding the quality of isotopic Ra data reported for the Site.
The most common isotopes of Ra are 226Ra (produced by radioactive decay of 238U) and 228Ra (produced during radioactive decay of 232Th) [2,3]. At the Site, 228Ra is the dominate Ra isotope in groundwater, whereas 226Ra is the dominant Ra isotope in the adjacent Impoundment. At the Site compliance well (Well 176), the historic concentrations of 226Ra + 228Ra have periodically exceeded the compliance standard (USEPA Maximum Contaminant Level, MCL) of 5 pCi/L; since approximately 2005, however, the concentrations have shown an increasing trend and with more frequent exceedances of the MCL. In addition to a downgradient well (Well 179), an unexpected increasing trend and high variability in 226Ra + 228Ra concentrations have also been measured in the unimpacted Site background well (Well 182) since 2005 (Figure 1). The implication of 226Ra + 228Ra concentrations exceeding the MCL at the compliance location (Well 176) is that the Site boundary may potentially need to be expanded to demonstrate that downgradient groundwater 226Ra + 228Ra concentrations will remain protective of human health and the environment.
A preliminary internal evaluation of 226Ra + 228Ra results for Site groundwater indicated that increasing trends and high variability could be the result of analytical error by the contract laboratory based on several key observations: (1) increases in both 226Ra + 228Ra and 228Ra/226Ra ratios occurred subsequent to a change in the 228Ra analytical method from Method 904.0 (M904.0) to Method Ra-05 (Ra-05) in 2005 (Figure 1 and Figure 2), (2) ratios of 228Ra/226Ra have been increasing in the non-impacted Site background Well 182 concurrently with several other monitoring wells and the Impoundment (Figure 2), and (3) re-analysis of samples with unusually elevated 228Ra has demonstrated at least one occurrence of poor reproducibility and significant analytical error in 228Ra measurements. For example, the reported 228Ra value for Site Well 174 (3rd Quarter in 2016) was 8.6 pCi/L, while the field duplicate value was reported as <1.3 pCi/L. Re-analysis results for these two samples were reported as 2.8 pCi/L 228Ra for the primary sample and 2.7 pCi/L 228Ra for the duplicate sample. In addition to the poor reproducibility of the field duplicate, additional 228Ra results appeared to be inaccurate. For instance, 228Ra for the unimpacted background well (Well 182) was reported as 27 pCi/L which is not consistent with much lower historical results for the background location. Reanalysis of the sample from Well 182 produced a revised result of 2.0 pCi/L. Out of nine total samples which were reanalyzed, the initial results from five samples exceeded the MCL for 226Ra + 228Ra based on 228Ra alone, but were reported as below the MCL after reanalysis. The contract laboratory was unable to provide an explanation as to why the reanalysis produced significantly different results from those initially reported.
Therefore, an ILC of 226Ra and 228Ra in selected groundwater wells and the adjacent Impoundment was conducted to evaluate the potential role of random variability and/or systematic error in producing the high variability in 226Ra + 228Ra concentrations and 228Ra/226Ra ratios observed during recent years. Four separate laboratories were evaluated which are designated as Lab A (Site contract laboratory), Lab B, Lab C, and Lab D. The specific objectives were to (1) evaluate potential differences between 226Ra results when using a screening method (M903.0) and compare those results to a 226Ra-specific method (M903.1) and (2) evaluate potential differences between 228Ra between laboratories using M904.0 and compare those results to 228Ra when using Method Ra-05. Additional questions include the following: (1) Are there detectable levels of 232Th in groundwater which could be migrating and influencing ratios of 228Ra/226Ra? (2) Are there differences between total (unfiltered) and dissolved (filtered, 0.45 µm) 226Ra or 228Ra, and if so, how do those fractions vary across laboratories? And (3) does M903.0 for 226Ra (which measures all alpha-emitting isotopes) overestimate 226Ra concentrations compared to M903.1 which is specific to 226Ra? If the presence of additional alpha emitters were to produce higher apparent 226Ra concentrations, it may be recommended to use M903.1, rather than M903.0, for future compliance monitoring at the Site.

2. Materials and Methods

The quality of 226Ra and 228Ra data reported by the existing contract laboratory (Lab A) was investigated using an ILC which included analysis of primary samples, a duplicate sample, and field blanks. Groundwater samples were collected from the Site compliance well (Well 176), the Site background well (Well 182), and from an intermediate well located outside of the Site boundary (Well 179). Surface water was collected from the adjacent Impoundment where the U ore was originally mined. Samples were collected for analysis of 226Ra and 228Ra, in addition to 232Th to evaluate the potential role of 232Th as a source of 228Ra. Water samples were split and processed accordingly to evaluate their dissolved (filtered, 0.45 µm pore-size filter) and total (unfiltered) forms. A duplicate was collected from Well 176, and field blanks (filtered and unfiltered) were prepared using deionized H2O supplied by each laboratory. All sample collection, preservation, and storage protocols followed standard methods for water sample collection [4]. At each location, the filtered and non-filtered samples were split into four groups and submitted to four laboratories designated as Lab A (existing contract laboratory), Lab B, Lab C, and Lab D. Each laboratory and their corresponding analytical methods are summarized in Table 1.
The methods used for 232Th varied among laboratories but are all based on alpha spectroscopy; the methods used for 226Ra and 228Ra have been approved by the U.S. Environmental Protection Agency (USEPA) [5] and also varied among the laboratories (Table 2). For 226Ra, three of the laboratories (Labs A, B, C) used Method 903.0 (M903.0). Although the results are reported as 226Ra, M903.0 measures the total alpha emitting isotopes of Ra (223Ra, 224Ra, 226Ra) and is therefore considered a screening method. Method 903.1 (M903.1), which is specific to 226Ra, was used by Lab D and can be compared to results from M903.0 to determine if other alpha emitting isotopes contribute to the reported 226Ra concentrations for the Site. For 228Ra, two of the laboratories used Method 904.0 (M904.0) (Labs B, C), while Lab A and Lab D used Method Ra-05. Each Ra method in Table 2 has been approved for drinking water compliance analysis by the USEPA and therefore should not produce significantly different results [6].

3. Results

The analytical results for total and dissolved 226Ra and 228Ra at all locations reported by the four laboratories are provided in Table 3. All dissolved 232Th results were below the detection limit (0.0131 to 1.11 pCi/L) and only two total 232Th measurements were detectable at low concentrations (0.0074 and 0.017 pCi/L); therefore, the 232Th results are not tabulated. This section provides (1) a summary of the quality assurance/quality control (QA/QC) results (Section 3.1), (2) a general overview of the dissolved and total 226Ra and 228Ra results from the various laboratories, (3) a more detailed comparison of dissolved and total 226Ra and 228Ra, and (4) an evaluation of the combined 226Ra + 228Ra results between laboratories relative to the MCL.

3.1. Quality Assurance/Quality Control (QA/QC) Summary

Laboratory data quality was evaluated based on results from method blanks (MB), laboratory control samples (LCS), LCS duplicates (LCSD), matrix spikes (MS), and MS duplicates (MSD) using USEPA guidelines [7]. In the MBs, all 226Ra and 228Ra results were below their minimum detectable concentration (MDC). Data accuracy evaluation showed the percent recovery values for the LCS, LCSD, MS, and MSD were within recommended limits. Data precision evaluation showed that the relative percent differences (RPDs) for analytical duplicate samples and field duplicate samples were also within laboratory control limits for all analyses. Total 228Ra was detected above the MDC in the field blank by Lab A and Lab B, and dissolved 226Ra was detected above the MDC in the field blank by Lab C. Dissolved 228Ra was also detected above the MDC in the field blank by Lab D. Sample results containing less than 10× the amount found in the field blanks were qualified as being estimated with high bias (Table 3) and were assigned a J+ qualifier.
Radiological data such as 226Ra and 228Ra are typically reported as the measured value ± the degree of precision. When a value is below the MDC, the actual value is still reported but the datum is flagged as non-detect and the MDC is provided. The sum of the 226Ra and 228Ra is subsequently calculated to determine the combined Ra (226Ra + 228Ra) that is typically used for compliance in the USA. If either of the 226Ra or 228Ra values are below detection, the value of the MDC for the non-detect value is summed with the detectable value to calculate the 226Ra + 228Ra concentration. When both 226Ra and228Ra values are below detection, the respective MDC values are summed to obtain the MDC for the combined 226Ra and228Ra, which is then reported as below detection. The sum of individual precision values for 226Ra and 228Ra is used as the overall precision for 226Ra + 228Ra.

3.2. Summary of 226Ra Results

The dissolved and total 226Ra results for all samples are shown in Figure 3. The error bars represent the precision in the analyses. Any reported value which was below the MDC is indicated with the MDC shown in black, and any J+ qualified value (estimated with high bias) is shown in red. None of the dissolved 226Ra results were significantly different based on their range in precision. The results from Lab C were J+ qualified and were the highest reported at each location, except for the Impoundment. The values reported by Lab D using M903.1, which is specific to 226Ra, tended to be lower for all samples, with the exception of Well 176 where the Lab A value was slightly lower (Figure 3).
The trend in lower values using M903.1 was not observed for total 226Ra (Figure 3), except in Well 176 where total 226Ra was below detection. However, the value of <0.36 pCi/L for total 226Ra in Well 176 should be equal to or greater than the dissolved value because the sample was not filtered; therefore, the value of <0.36 pCi/L is suspect. No 226Ra was detected in the unfiltered field blanks, and therefore none of total 226Ra values were qualified as estimated with high bias (Table 3, Figure 3).

3.3. Summary of 228Ra Results

Overall, the 228Ra results (Figure 4) tended to be more variable across locations and between laboratories compared to 226Ra (Figure 3). A comparison of these results shows that 226Ra tends to be higher in the Impoundment, whereas 228Ra is higher in the groundwater, consistent with historical Site observations. However, no regular trends are apparent based on the method or laboratory used for dissolved 228Ra. For example, although Lab A and Lab D use Ra-05, the Lab D results were notably higher than Lab A in samples where 228Ra concentrations were overall higher (≥2 pCi/L). The Lab D results were qualified as estimated with high bias due to detection of 228Ra in the field blank. Although a higher concentration of 228Ra was reported by Lab A for the field blank, the detection limit (<1.4 pCi/L) was higher compared to other laboratories and methods (Figure 4, Table 2). In samples containing overall lower 228Ra concentrations (≤2 pCi/L), the results from Lab A were notably higher compared to Lab D. The results from Lab B and Lab C using M904.0 tended to be intermediate between those from Lab A and Lab D, where Ra-05 was used (Figure 4).
Similar trends were observed for total 228Ra compared to dissolved 228Ra (Figure 4). However, for total 228Ra, the results from Lab A, rather than Lab D, were qualified as estimated with high bias due to detection of 228Ra in the field blank (Table 3, Figure 4). Lab A reported 2 pCi/L of 228Ra in the field blank, and similar to 226Ra values, results from Lab A were notably higher in samples containing overall low 228Ra (Impoundment, Well 182).

3.4. Detailed Comparison of Total vs. Dissolved Ra

A comparison of total and dissolved 226Ra from each laboratory is shown in Figure 5. The error bars represent the precision in the analyses, with non-detect values shown in black and all J+ qualified values shown in red. Theoretically, total concentrations should be equal to or greater than dissolved concentrations, with the extent of Ra partitioning to suspended particles depending on factors such as total suspended solids concentration and pH [2]. Total and dissolved 226Ra were generally the same in all samples based on the range in precision of the analyses (Figure 5). However, the difference between the actual reported values in total vs. dissolved at a given location were not consistent between laboratories. For example, the reported total 226Ra in the Impoundment was higher than dissolved for the Lab D data, lower than dissolved for Lab A and Lab B, and almost equal for Lab C. In the background Well 182, the reported total 226Ra was greater than dissolved for Lab D and Lab B, less than dissolved for Lab C, and approximately equal for Lab A.
For the 228Ra data shown in Figure 6, total 228Ra was more consistently greater than or equal to the dissolved fraction when compared to 226Ra in Figure 5. It is also apparent that both total and dissolved 228Ra values from Lab A are elevated in the samples containing low Ra (Impoundment, Well 182, and field blank) compared to the other lab results, including those from Lab D where the same method (Ra-05) was used. In addition, the values reported for total 228Ra in the field blank were higher compared to non-detectable dissolved 228Ra from Lab A and Lab B. Differences between total and dissolved 228Ra in the field blank suggests that processing of the sample for total analysis could have produced false positive results.

3.5. Combined 226Ra + 228Ra Results

The 226Ra + 228Ra values for the dissolved and total fractions are shown in comparison to the 5 pCi/L compliance value in Figure 7. The dissolved 226Ra + 228Ra values reported by Lab C and Lab D exceeded the MCL at Well 176. Total 226Ra + 228Ra reported by Lab C and Lab D also exceeded the MCL at Well 176, in addition to total 226Ra + 228Ra reported by Lab A for the Impoundment and total 226Ra + 228Ra reported by Lab B and Lab D at Well 179. Overall, the trends are not entirely consistent between laboratories. For example, there was little difference between dissolved and total 226Ra + 228Ra for the Impoundment and Well 176 in the data reported by Lab B, Lab C, and Lab D, while the total 226Ra + 228Ra values reported by Lab A were higher than dissolved for the same two locations. Conversely, values were higher for total 226Ra + 228Ra compared to dissolved 226Ra + 228Ra at Well 179 for the data reported by Lab B add Lab D, while total vs. dissolved 226Ra + 228Ra concentrations reported by Lab A were similar. The dissolved 226Ra + 228Ra in the field blank was below detection as reported by Lab A and Lab B, while total 226Ra + 228Ra was detectable; the opposite trend was observed for the data reported by Lab C and Lab D (Figure 7).

4. Summary and Discussion

This ILC of 226Ra and 228Ra was conducted to understand the potential reasons for high variability and increasing trends in 226Ra + 228Ra concentrations that have been observed in Site monitoring wells, including the compliance Well 176 and background Well 182. The results for 226Ra were fairly consistent between laboratories, and with the exception of dissolved 226Ra reported by Lab C, all 226Ra concentrations were below detection in the field blanks. The concentrations of 228Ra were more variable between laboratories and with a higher frequency of detection in the field blanks.
For both 226Ra and 228Ra, there were no consistent trends with respect to the total vs. dissolved concentrations across laboratories, which were essentially equal in most samples when the range in measured precision was considered. The largest differences between reported values for total vs. dissolved Ra were observed in the Lab A data for 228Ra, where the total 228Ra data were qualified as estimated with high bias. For both dissolved and total 226Ra + 228Ra, the results from Lab C and Lab D for the compliance Well 176 exceeded the MCL, whereas the results from Lab A and Lab B were below the MCL. Therefore, compliance with the MCL was either achieved or not achieved, depending on which laboratory conducted the analyses.
Results for 226Ra obtained using M903.1 (specific to 226Ra) were usually equal to or slightly lower than those generated using M903.0; however, the differences are insignificant and may be due to analytical variability rather than method specificity. Results for 228Ra reported by Lab D when using Method Ra-05 were more consistent with those from Lab B and Lab C when using M904.0. The 228Ra results reported by Lab A when using Method Ra-05 tended to be higher in samples where the remaining laboratories reported low 228Ra concentrations (<2 pCi/L), and tended to be lower in samples where the remaining laboratories reported higher 228Ra concentrations (>2 pCi/L). Lab A also reported the highest measured 228Ra concentration of all field blanks (2 pCi/L), and consequently those data for total 228Ra were qualified as estimated with high bias.
A prior study relating to radionuclide variability in water samples [6] showed that radionuclide activity in groundwater can vary up to four-fold due to seasonal effects, but also noted that the variability could reflect issues related to method and laboratory variability. The greatest source of error for 226Ra is counting uncertainty, but the overall accuracy and precision for 226Ra results was acceptable among various laboratories. For 228Ra, however, the methods are generally not as robust as those used for 226Ra. No single laboratory was able to provide consistently accurate 228Ra results, and due to the much greater relative standard deviation (RSD) values, it was concluded that the accuracy of 228Ra results is dependent on laboratory technique and skill and “may also reflect less rugged analytical methods” [5].

5. Conclusions and Recommendations

The results obtained from this ILC of Site groundwater and surface water samples are generally consistent with the findings from the prior 2011 study [6] with respect to observed variability in 226Ra compared to 228Ra. The current ILC indicate that 228Ra data reported by the existing contract laboratory (Lab A) are biased high relative to those reported by Lab B, Lab C, Lab D, and this may be due to differences in laboratory skill and technique among the various laboratory technicians [6]. However, the resulting upward trends in 228Ra/226Ra ratios (observed in both impacted wells and the background location) after the 228Ra method was changed from 904.0 to Ra-05 are more difficult to explain.
Although the trends were not entirely consistent, the results from this study show a tendency for total Ra to be higher than dissolved Ra, especially for 228Ra. For samples where the majority of the laboratories reported low 228Ra concentrations (<2 pCi/L), it appears that the method and techniques used by Lab A for total 228Ra is contributing to higher apparent 228Ra concentrations. Therefore, it was recommended that 226Ra and 228Ra analyses (in additional to uranium and other radionuclides) be conducted on a filtered sample for Site compliance evaluations. Analysis for dissolved concentrations would also then be consistent with remaining metals analyses which are also analyzed as dissolved.
For 226Ra, it was recommended that M903.0 be replaced with M903.1 which is specific to 226Ra. Method 903.0 includes all alpha emitters and therefore can result in over-reporting of 226Ra concentrations. The primary additional alpha emitter which could contribute to overestimation of 226Ra is 224Ra, which originates from 232Th. Although 232Th concentrations were very low to non-detect in the Site samples, the presence of 232Th in the solid phase of aquifer materials could release 224Ra into solution. Even though this study did not indicate that M903.0 results in consistent over-reporting of 226Ra concentrations, use of M903.1 would better ensure that the appropriate analysis for 226Ra is being conducted. Alternatively, it could be requested that the laboratory employ a 14-day waiting time which will effectively reduce the contribution of 224Ra to the reported 226Ra [6].
For 228Ra, this study indicates that specific laboratory techniques, rather than the method (Ra-05) used by Lab A are potentially contributing to excessively elevated values being reported for 228Ra at the Site. It should be noted that even though standard methods [5] are being used, many laboratories have developed their own customized Standard Operating Procedures (SOPs) which are often proprietary and difficult to obtain. However, it is recommended that any location whose reported 226Ra + 228Ra concentrations exceed groundwater or surface water standards in the future be immediately re-sampled and re-analyzed by the existing laboratory, with a sample split also being collected and sent to an outside laboratory for additional confirmatory analysis.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data are contained within the article.

Acknowledgments

Thanks to Jill Richards (Worthington Miller Environmental, LLC) for her efforts with laboratory coordination and data validation.

Conflicts of Interest

The author declares no conflict of interest.

References

  1. Guthrie, W.F. Interlaboratory Comparisons. In Encyclopedia of Statistics in Quality and Reliability; Ruggeria, F., Kenett, R.S., Faltin, F.W., Eds.; National Institute of Standards and Technology: Gaithersburg, MD, USA, 2008. [Google Scholar] [CrossRef]
  2. Langmuir, D.L. Aqueous Environmental Chemistry, 1st ed.; Prentice-Hall, Inc.: Upper Saddle River, NJ, USA, 1997; 600p. [Google Scholar]
  3. International Atomic Energy Agency (IAEA). The Environmental Behaviour of Radium: Revised Edition; International Atomic Energy Agency: Vienna, Austria, 2014; 267p. [Google Scholar]
  4. Baird, R.; Bridgewater, L. Standard Methods for the Examination of Water and Wastewater, 23rd ed.; American Public Health Association: Washington, DC, USA, 2017; 1545p. [Google Scholar]
  5. U.S. Environmental Protection Agency (USEPA). Compendium of EPA-Approved Analytical Methods for Measuring Radionuclides in Drinking Water; U.S. Department of Energy, Office of Environmental Policy and Assistance, Air, Water, and Radiation Division: Washington, DC, USA, 1998; 62p.
  6. Eaton, A.Y.; Cha, Y.; Morley, K.M. Evaluation of variability in radionuclide measurements in drinking water. J. Am. Water Resour. Assoc. 2011, 103, 119–130. [Google Scholar] [CrossRef]
  7. U.S. Environmental Protection Agency (USEPA). National Functional Guidelines for Inorganic Superfund Methods Data Review; USEPA Office of Superfund Remediation and Technology Innovation: Washington, DC, USA, 2020; 158p.
Figure 1. 226Ra + 228Ra concentrations in Site monitoring wells and the Impoundment.
Figure 1. 226Ra + 228Ra concentrations in Site monitoring wells and the Impoundment.
Applsci 12 12198 g001
Figure 2. 228Ra/226Ra ratios in selected Site monitoring wells and the Impoundment.
Figure 2. 228Ra/226Ra ratios in selected Site monitoring wells and the Impoundment.
Applsci 12 12198 g002
Figure 3. Comparison of (a) Dissolved 226Ra; (b) Total 226Ra. Values in red are qualified as estimated with high bias (J+) when sample result contained <10× the amount measured in the field blank. Values in black signify the MDC when the reported value is less than the MDC.
Figure 3. Comparison of (a) Dissolved 226Ra; (b) Total 226Ra. Values in red are qualified as estimated with high bias (J+) when sample result contained <10× the amount measured in the field blank. Values in black signify the MDC when the reported value is less than the MDC.
Applsci 12 12198 g003
Figure 4. Comparison of (a) Dissolved 228Ra; (b) Total 228Ra. Values in red are qualified as estimated with high bias (J+) when sample result contained <10× the amount measured in the field blank. Values in black signify the MDC when the reported value is less than the MDC.
Figure 4. Comparison of (a) Dissolved 228Ra; (b) Total 228Ra. Values in red are qualified as estimated with high bias (J+) when sample result contained <10× the amount measured in the field blank. Values in black signify the MDC when the reported value is less than the MDC.
Applsci 12 12198 g004
Figure 5. Comparison of total and dissolved 226Ra concentrations: (a) Lab A; (b) Lab B; (c) Lab C; (d) Lab D.
Figure 5. Comparison of total and dissolved 226Ra concentrations: (a) Lab A; (b) Lab B; (c) Lab C; (d) Lab D.
Applsci 12 12198 g005
Figure 6. Comparison of total and dissolved 228Ra concentrations: (a) Lab A; (b) Lab B; (c) Lab C; (d) Lab D.
Figure 6. Comparison of total and dissolved 228Ra concentrations: (a) Lab A; (b) Lab B; (c) Lab C; (d) Lab D.
Applsci 12 12198 g006
Figure 7. Interlaboratory comparison of (a) Dissolved 226Ra + 228Ra; (b) Total 226Ra + 228Ra. Values in black signify the MDC when the reported values were less than the MDC.
Figure 7. Interlaboratory comparison of (a) Dissolved 226Ra + 228Ra; (b) Total 226Ra + 228Ra. Values in black signify the MDC when the reported values were less than the MDC.
Applsci 12 12198 g007
Table 1. List of laboratories and methods used in the ILC.
Table 1. List of laboratories and methods used in the ILC.
Lab232Th226Ra228Ra
AM908.0M903.0Ra-05
BESM4506M903.0M904.0
CASTM D3972M903.0M904.0
DDOE EML HASL-300
Th-01-RC Modified
M903.1Ra-05
Table 2. Comparison of USEPA-approved methods for Ra [5].
Table 2. Comparison of USEPA-approved methods for Ra [5].
AnalyteMethod/MDL 1Method PrincipleMethod
Notes
226Ra903.0/0.5Radiochemical precipitation; alpha counting by scintillation or gas-flow proportional counting.Does not always provide an accurate measurement of 226Ra with other alpha emitters present and often used for screening. Method states: “When the total radium alpha activity of a drinking water sample is greater than 5 pCi/L, then the 226Ra analysis (Method 903.1) is required.”
226Ra903.1/0.5Radiochemical precipitation; radon emanation; alpha counting by scintillation.No radioactive interferences. 226Ra in solution is determined by co-precipitation with BaSO4 and sample analyzed using de-emanation.
228Ra904.0/1.0Radiochemical precipitation; counting by gas-flow proportional beta counterMeasures 228Ra alone or in conjunction with 226Ra. Ra in solution from 228Ra determination is saved, and the Ra is reprecipitated for 226Ra analysis. Ra is precipitated as Ra-BaSO4, which is dissolved and purified from EDTA solution. After ingrowth of 228Ac, the 228Ac is precipitated with Y(OH)3, reprecipitated with Y, and counted by beta counter.
228RaRa-05\NR 2Radiochemical precipitation; beta counting in a low-background proportional counter.Ra is precipitated as Ra-BaSO4 which is dissolved in a pentasodium diethylenetriamine pentacetate solution. After ingrowth of 228Ac, 228Ac is extracted with Di 2-ethylhexyl phosphoric acid and back-extracted with HNO3.
1 MDL = minimum detectable level (pCi/L). 2 NR = not reported.
Table 3. Comparison of ILC results (pCi/L) for 226Ra and 228Ra 1.
Table 3. Comparison of ILC results (pCi/L) for 226Ra and 228Ra 1.
LocationLab226Ra
(Dissolved)
226Ra
(Total)
228Ra
(Dissolved)
228Ra
(Total)
ImpoundmentA1.81.51.73.7 J+
B2.21.5<0.691 J+
C1.52 J+1.56<0.61<0.69
D1.41.66<0.7<0.7
Well 176
(Compliance)
A0.70.8<1.94.1 J+
B0.970.743.23.2 J+
C1.37 J+1.24.64.8
D0.965<0.3565.2 J+5.9
Well 176
(Duplicate)
A0.70.834.1 J+
B1.30.833.84.7 J+
C1.08 J+0.963.95.1
D0.8590.7775.5 J+7
Well 179A0.50.51.91.5 J+
B<0.660.581.74.8 J+
C0.77 J+0.732.072.68
D0.3340.6393.7 J+5.4
Well 182
(Background)
A0.30.32.93.3 J+
B<1<0.57<0.75<0.55
C0.4 J+0.32<0.62<0.64
D<0.3280.4870.8 J+1.8
Field BlankA<0.2<0.2<1.42
B<1.1<0.55<0.631.4
C0.28<0.25<0.51<0.63
D<0.249<0.2340.8<0.7
1 Values with a J+ qualifier indicates the value is qualified as an estimate with high bias when detected in the field blank and the sample result is <10× the amount measured in the field blank.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Levy, D.B. Interlaboratory Comparison of 226Ra and 228Ra Activity Concentrations in Groundwater and Surface Water. Appl. Sci. 2022, 12, 12198. https://doi.org/10.3390/app122312198

AMA Style

Levy DB. Interlaboratory Comparison of 226Ra and 228Ra Activity Concentrations in Groundwater and Surface Water. Applied Sciences. 2022; 12(23):12198. https://doi.org/10.3390/app122312198

Chicago/Turabian Style

Levy, David B. 2022. "Interlaboratory Comparison of 226Ra and 228Ra Activity Concentrations in Groundwater and Surface Water" Applied Sciences 12, no. 23: 12198. https://doi.org/10.3390/app122312198

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop