Article

Analytical Validation of the Cxbladder® Triage Plus Assay for Risk Stratification of Hematuria Patients for Urothelial Carcinoma

1
Pacific Edge Diagnostics NZ, Ltd., 87 St. David Street, Dunedin 9016, New Zealand
2
David Fletcher Consulting Ltd., 67 Stornoway Street, Karitane 9471, New Zealand
3
Pacific Edge Diagnostics USA, Ltd., 1214 Research Boulevard, Hummelstown, PA 17036, USA
*
Author to whom correspondence should be addressed.
Diagnostics 2025, 15(14), 1739; https://doi.org/10.3390/diagnostics15141739
Submission received: 1 May 2025 / Revised: 2 July 2025 / Accepted: 4 July 2025 / Published: 8 July 2025
(This article belongs to the Special Issue Opportunities in Laboratory Medicine in the Era of Genetic Testing)

Abstract

Background/Objectives: Cxbladder® Triage Plus is a multimodal urinary biomarker assay that combines reverse transcription-quantitative analysis of five mRNA targets and droplet-digital polymerase chain reaction (ddPCR) analysis of six DNA single-nucleotide variants (SNVs) from two genes (fibroblast growth factor receptor 3 (FGFR3) and telomerase reverse transcriptase (TERT)) to provide risk stratification for urothelial carcinoma (UC) in patients with hematuria. This study evaluated the analytical validity of Triage Plus. Methods: The development dataset used urine samples from patients with microhematuria or gross hematuria that were previously stabilized with Cxbladder solution. Triage Plus was evaluated for predicted performance, analytical criteria (linearity, sensitivity, specificity, accuracy, and precision), extraction efficiency, and inter-laboratory reproducibility. Results: The development dataset included 987 hematuria samples. Compared with cystoscopy (standard of care), Triage Plus had a predicted sensitivity of 93.6%, specificity of 90.8%, positive predictive value (PPV) of 46.5%, negative predictive value of 99.4%, and test-negative rate of 84.1% (score threshold 0.15); the PPV increased to 74.6% for the 0.54 score threshold. For the individual FGFR3 and TERT SNVs, the limit of detection (analytical sensitivity) was a mutant-to-wild type DNA ratio of 1:440–1:1250 copies/mL. Intra- and inter-assay variance was low, while extraction efficiency was high. All other pre-specified analytical criteria (linearity, specificity, and accuracy) were met. Triage Plus showed good reproducibility (87.9% concordance between laboratories). Conclusions: Cxbladder Triage Plus accurately and reproducibly detected FGFR3 and TERT SNVs and, in combination with mRNA expression, provides a non-invasive, highly sensitive, and reproducible tool that aids in risk stratification of patients with hematuria.

1. Introduction

The American Urological Association (AUA) microhematuria guideline recommends risk stratification of patients with microhematuria for the presence or absence of urothelial carcinoma (UC) [1]. Based on the 2025 AUA risk criteria, clinicians should engage low-risk patients in repeat urinalysis within 6 months, while intermediate- and high-risk patients should undergo full work-up (i.e., cystoscopy and renal ultrasound in intermediate-risk patients, or cystoscopy and axial upper tract imaging in high-risk patients). According to the AUA 2025 guidelines, validated urinary biomarker tests or cytology may be used to guide shared decision making for deferred work-up in intermediate-risk patients who are reluctant to undergo full work-up, with repeat urinalysis conducted within 1 year [1].
Cystoscopy, while considered the gold standard for UC diagnosis, may produce false-negative and false-positive results and has variable performance [2]. Historically, cytology and fluorescence in situ hybridization have also been used to evaluate patients with suspected UC; however, both of these methods have poor sensitivity [3,4]. Urine cytology may be used to provide clinical resolution in samples with equivocal cystoscopy findings, but it may miss some high-grade or muscle-invasive tumors due to low sensitivity [5]. Therefore, there is an unmet need for non-invasive tests with high sensitivity and specificity to assess the risk of UC in patients with hematuria.
Cxbladder® Triage and Detect assays are validated reverse transcription quantitative-polymerase chain reaction (PCR)-based urinary biomarker tests that are used for risk stratification of patients with hematuria [6,7]. Both assays quantify mRNA expression of five biomarkers, including four genes that are known to be associated with UC (cyclin-dependent kinase 1 (CDK1), midkine (MDK), insulin-like growth factor binding protein 5 (IGFBP5), and homeobox A13 (HOXA13)) [8] and one known marker of inflammation (C-X-C motif chemokine receptor 2 (CXCR2)) [6,7]. Cxbladder Triage also incorporates clinical risk factors to increase the assay sensitivity (age, sex, smoking history, and presence of frequent gross hematuria) [9].
The multimodal Cxbladder Triage Plus assay was subsequently developed with ongoing research. It combines the five mRNA biomarkers with six DNA single-nucleotide variants (SNVs) from two other genes that are known to be associated with UC: fibroblast growth factor receptor 3 (FGFR3) and telomerase reverse transcriptase (TERT) [10]. Hotspot mutations in the FGFR3 gene (R248C, S249F/C, G372C, and Y375C (missense gain-of-function mutations)) and upstream of the TERT promoter (C228T and C250T (gain-of-function mutations)) are often identified in urothelial carcinoma samples [11,12,13]. In a previous clinical validation study, an enhanced Triage assay showed improved diagnostic performance over the original Detect and Triage assays [14]. Of note, an earlier version of Triage Plus was named “Detect+” in the Lotan et al. 2023 publication [14]; this should not be confused with “Triage+”, which contained clinical factors. Further refinement of the enhanced Triage assay resulted in development of the Cxbladder Triage Plus assay (hereafter referred to as Triage Plus), which provides higher specificity and sensitivity than earlier versions of the assays while solely relying on genomic biomarkers (i.e., without the need for clinical factors).
For biomarker tests to be integrated into routine use in clinical practice, they must show strong analytical validity, clinical validity, and clinical utility [15]. A previous analytical validation study showed accurate and reproducible quantification of mRNA expression for the five biomarker genes with both Cxbladder Triage and Detect assays [16]; as Triage Plus also uses these biomarkers, the analytical performance of mRNA detection was not changed and so will only be referred to here (not reiterated). This analytical validation study focused on the linearity, analytical sensitivity, specificity, accuracy, and precision, as well as the extraction efficiency and inter-laboratory reproducibility, for the analysis of the six DNA SNVs in Triage Plus, and we reference the earlier publication [16] for the mRNA analysis.

2. Materials and Methods

2.1. Samples and Study Design

A development dataset was used to develop the Triage Plus algorithm and to perform the analytical validation. The dataset included −80 °C stored Cxbladder-stabilized urine samples from patients with gross hematuria or microhematuria who had participated in a previous clinical validation study [14] and from patients with hematuria who participated in the STRATA: Safe Testing of Risk for Asymptomatic Microhematuria study [17]. Samples were blinded prior to their use in the development dataset.
The previous clinical validation study included two cohorts of patients aged ≥18 years with gross hematuria (United States (US)) or aged >21 years with gross hematuria or microhematuria (Singapore) who were scheduled to undergo evaluation for possible UC [14]. The study protocols were approved in Singapore by the SingHealth Centralized Institutional Review Board (IRB; 2016/2471 (approved 19 July 2016, 13 June 2017, and 23 February 2019)) and NHG Domain Specific Review Board (2018/00234-SRF0002 (approved 3 August 2019)), and in the US by local relevant IRBs (Chesapeake IRB, Pro00009623 (approved 13 December 2012, 29 January 2013, 6 February 2013, 3 June 2014, 24 February 2016, 9 May 2016, and 20 May 2016); Florida Hospital IRB, 394399 (approved 19 March 2013); UT Southwestern Medical Center IRB, STU-112012-018 (approved 4 April 2013); and PennState Hershey IRB, 41719EP (approved 5 June 2013)). These studies were conducted in accordance with the Good Clinical Practice requirements and the Declaration of Helsinki (1975, revised 2013), and all patients provided informed consent before any study procedures were undertaken [14].
The STRATA study was a multicenter US study of patients aged > 18 years who were referred for evaluation of microhematuria [17]. Patients with a prior history of urologic malignancy or pelvic radiotherapy were excluded. The study protocol was approved by the PennState IRB (STUDY00010988 (approved 9 September 2019)), the University of Southern California IRB (HS-19-00766 (approved 14 June 2020)), the UT Southwestern Medical Center IRB (STU-2019-1020 (approved 11 September 2019)), the University of British Columbia Clinical Research Ethics Board (H19-01797 (approved 7 November 2019)), the University of Minnesota IRB (STUDY00008103 (approved 27 April 2020)), the Vanderbilt IRB (200304 (approved 16 June 2020)), the Western Research Health Sciences Research Ethics Board (114112 (approved 3 February 2020)), and the WCG IRB (20202112 (approved 20 July 2020, 2 August 2021, 21 November 2021, 19 October 2022, and 13 March 2023)). This study was also conducted in accordance with the Good Clinical Practice requirements and the Declaration of Helsinki (1975, revised 2013), and all eligible patients signed IRB-approved consent prior to study entry [17].

2.2. Algorithm Development

In the original Detect+ algorithm, quantification of mRNA expression was encapsulated by two variables in the algorithm: X1, which combined the four biomarkers known to be associated with UC (IGFBP5, HOXA13, MDK, and CDK1) into a single predictor; and X2, the inflammation target (CXCR2). The DNA component of the original Detect+ algorithm was summarized as an “FGFR3 and TERT DNA-positive call” or a DNA-positive result.
Based on a thorough review of the statistical and machine-learning literature, which was conducted to select potentially useful algorithms, it was concluded that the best predictions were obtained using a single algorithm (i.e., Bayesian additive regression tree (BART)) [18]. However, the BART algorithm could not improve the performance of the original Detect+ algorithm.
After re-evaluation of this approach, while keeping the FGFR3 and TERT DNA-positive call, the mRNA quantification component of the Detect+ algorithm was improved to create the new algorithm, hereafter referred to as the Triage Plus algorithm.
In the Triage Plus algorithm, all five biomarkers appeared separately in the second-order polynomial equation (X1 = IGFBP5, X2 = HOXA13, X3 = MDK, X4 = CDK1, and X5 = CXCR2), as follows:
l o g i t p = a 0 + a 1 X 1 + a 2 X 2 + a 3 X 3 + a 4 X 4 + a 5 X 5 + a 11 X 1 2 + a 22 X 2 2 + a 33 X 3 2 + a 44 X 4 2 + a 55 X 5 2 + a 12 X 1 X 2 + a 13 X 1 X 3 + a 14 X 1 X 4 + a 15 X 1 X 5 + a 23 X 2 X 3 + a 24 X 2 X 4 + a 25 X 2 X 5 + a 34 X 3 X 4 + a 35 X 3 X 5 + a 45 X 4 X 5
The coefficients (a0, a1, …, a45) were obtained by fitting a logistic regression model, with the confirmed diagnosis as the response variable and the linear predictor as given by the equation above. The Triage Plus score, which was given by the value of logit (p) determined by the above equation using the inverse of the logit function, can be interpreted as an estimate of the probability of cancer. The calculated composite Cxbladder Triage Plus score ranged from 0.00 to 1.00. Two test thresholds were set, with one threshold optimized for maximum sensitivity and negative predictive value (NPV) and the second threshold chosen to maximize positive predictive value (PPV), to create three zones to risk-stratify patients on the probability of having UC: low, intermediate, or high probability.
The development dataset was used to determine the thresholds for the predicted performance of Triage Plus and obtained thresholds of 0.15 and 0.54. Using these thresholds and compared with cystoscopy, the predicated performance parameters were predicted sensitivity, specificity, PPV, NPV, and test-negative rate (TNR). All performance outcomes were based on “leave one out” cross validation (i.e., the score for each sample was found by training the relevant algorithm on all other samples).

2.3. Analytical Validation

Assessment of the analytical performance of Triage Plus for detection of RNA from IGFBP5, HOXA13, MDK, CDK1, and CXCR2 genes was conducted as previously described for the Cxbladder Triage and Detect assays [16]. The methods for the analytical validation of Triage Plus for detection of the six DNA SNVs from FGFR3 (R248C, S249F/C, G372C, Y375C) and TERT (C228T, C250T) are described in the following sections.

2.3.1. Linearity

The linearity of Triage Plus was assessed for each of the six DNA SNVs of FGFR3 and TERT, with an upper limit being 2.21 λ and 1.76 λ, respectively (where λ is the mean number of target DNA molecules per droplet in droplet-digital PCR (ddPCR) using Poisson distribution). A 10-point dilution curve (33,000, 6600, 1320, 264, 132, 66, 33, 16.5, 8.25, and 4.12 DNA copies/well, with at least four replicates at each concentration) was performed to confirm the linearity of each analytic target. Statistical testing was performed to determine the concentration at which the assay became non-linear for each analytic target. Data were fitted using a linear regression model, and linearity was assessed using the regression coefficient (R2) and mean squared error (MSE). The null hypothesis was that Triage Plus had valid linear regression through the whole diluted range; this null hypothesis was rejected if R2 was <0.9. The MSE was used to compare linearity between different analytic targets.

2.3.2. Analytical Sensitivity

Analytical sensitivity (or limit of detection (LOD)) was defined as the lowest analyte concentration that could be consistently detected with 95% probability. The LOD for FGFR3 and TERT DNA SNVs was determined by logistic regression and the three-concentrations approach. For logistic regression, a model was fitted to explain the relationship between the dependent (positive/negative) and independent (analyte concentration) variables, and the LOD was predicted based on this fitted logistic model. For the three-concentrations approach, a fraction of positives was calculated from highest to lowest concentration, and the concentration that matched 95% of positives ( c 0 ) was the boundary of LOD. Samples with concentrations > c 0 were grouped into three concentration levels (i.e., c 1 , c 2 , and c 3 , where c 1 > c 2 > c 3 > c 0 ). If n i was the number of positive samples at concentration c i , it was assumed that n i ~ P o i s s o n ( λ c i ) (where i = 1, 2, 3); the expected value E n i = λ c i was the mean number of positive samples at concentration c i . If p i was the probability of no positive samples at concentration c i , then p i = P n i = 0 = e λ c i . If c * was the LOD, by definition, 1 p i = 0.95 and p i = e λ c * = 0.05 ; hence, c * = log 0.05 / λ ^ . If X i was the observed number of positive samples at concentration c i , the maximum likelihood estimator for λ ( λ ^ ) could be computed using the Triage Plus algorithm. The three concentrations ( c 1 , c 2 , and c 3 ) could be selected in multiple ways and the corresponding number of samples at each concentration changed accordingly, with both having an impact on the LOD estimate. The most conservative value was used as the estimated LOD.

2.3.3. Analytical Specificity

Analytical specificity was defined as the ability of Triage Plus to detect FGFR3 and TERT mutant DNA SNVs in the presence of potentially interfering substances, which may have been carried over from the extraction reagent or present in the patient’s urine sample. The effect on analytical specificity of the assay for both sample- and process-derived interfering substances was assessed.
The sample-derived substances assessed were red blood cells (RBCs; 8 × 105, 4 × 106, 2 × 107, and 1 × 108 cells/mL), bacteria (1 × 106 cells/mL Escherichia coli), yeast (1 × 104 colony-forming units (CFU)/mL), urea (60 mg/mL), glucose (0.5 mg/mL), and protein (1.25, 2.50, 5.00, and 10.00 mg/mL serum albumin). The selected amount of each substance was mixed with eight high- and low-concentration FGFR3 and TERT extraction controls. The contaminated controls were then extracted and compared with high- and low-concentration controls that did not contain interfering substances. Samples within the expected level of gene variance and with a Triage Plus score that was within the 95% confidence interval (CI) for the control were considered acceptable.
The process-derived substances assessed were ethanol (4%), MagMAX wash buffer (2%), Cxbladder stabilizing reagent (1%), MagMAX magnetic beads (5%), and acetone (10%). The substance percentage was calculated per 64 μL of elution volume and each substance was mixed with high and low concentrations of FGFR3 and TERT extraction controls as 12 replicates and compared with controls (without potentially interfering substances) to determine if the samples were affected by the process-derived contaminants.

2.3.4. Analytical Accuracy

As there was no analytical standard for FGFR3 and TERT DNA, multiplex ddPCR was considered the most accurate method to define absolute quantitation in Triage Plus. To validate the analytical accuracy of Triage Plus, ddPCR of the SNVs of FGFR3 (R248C, S249F/C, G372C, and Y375C) and TERT (C228T, C250T) as single analytes were compared with combined analyte samples (of mutant plus wild type (WT) DNA). A combination of TERT C228T and C250T mutant DNA was also assessed.
Mutant DNA alone and mutant DNA combined with WT DNA were manufactured in parallel to contain equivalent concentrations of the target DNA. High-extraction controls (HECs) had a DNA concentration of ~1 × 106 copies/μL and low-extraction controls (LECs) had a DNA concentration of ~1 × 104 copies/μL. Each mutant was combined with WT DNA at a high (1:10) and low (1:200) mutant-to-WT ratio for the HECs and LECs, respectively. Mutant DNA samples were compared with mutant + WT DNA samples for each SNV of FGFR3 and TERT in the multiplex ddPCR assays.
Accuracy was determined as the percentage of the expected mutant DNA concentration compared with the mutant + WT DNA sample. Quantification of mutant DNA within the combined sample was then reviewed against its corresponding 95% CI.

2.3.5. Analytical Precision

Analytical precision was defined as reproducibility within a single run or between separate runs for replicate samples. Variance in the ddPCR assay was assessed using HECs and LECs for each FGFR3 and TERT SNV. A fitted linear random effects model was used to estimate intra-assay and inter-assay variance for the mutant fraction, presented as standard deviation (SD) and 95% CI; the coefficient of variation (CV%) was also calculated. Inter-assay, intra-assay, and total variance for the mutant fraction were assessed by four operators over >60 days by reviewing 46 ddPCR plate controls split over 22 plates (per FGFR3 SNV) or 48 ddPCR plate controls over 23 plates (per TERT SNV).
The lot-to-lot reagent variation was also assessed by testing three independent manufacture lots of MagMAX magnetic beads (BD2206339, BD2206338, and BD2302342) and MagMAX wash buffer (WB2405057, WB2405056, and WB2312054) for both FGFR3 and TERT SNVs using HECs and LECs. Each manufacture lot was run in parallel on a single plate with at least eight replicates of each control for each lot of reagents.

2.3.6. Extraction Efficiency

To confirm that Triage Plus was not biased for extraction of either FGFR3 or TERT mutant DNA, synthetic urine samples were prepared for each SNV at high (1:10) and low (1:200) mutant-to-WT ratios. The extraction efficiency was tested by processing eight replicates of each synthetic urine sample from extraction to ddPCR. These results were then compared with the expected mutant fraction and copy number. The absolute quantity of the extracted sample was used to define the extraction efficiency of DNA.

2.3.7. Inter-Laboratory Evaluation

The inter-laboratory comparison between the NZ laboratory (PEDNZ) and the US laboratory (PEDUSA) was assessed for Cxbladder Triage Plus. A random sample set from the PEDNZ validation was used to confirm the reproducibility of Triage Plus at PEDUSA. Acceptable variability was defined as achieving ≥80% concordance for all clinical results.

2.4. Statistical Analysis

Statistical analyses were conducted using R, version 4.3.1.

3. Results

3.1. Algorithm Development Samples

Of the original 1073 samples, 63 had DNA results that were neither positive nor negative and 23 had data missing for at least one of the predictors (IGFBP5, HOXA13, MDK, CDK1, and CXCR2) and could not be used to fit the Triage Plus algorithm. Therefore, 987 samples were used for the development of Triage Plus (Table S1).

3.2. Predicted Performance

Using this development dataset, the performance of the Triage Plus algorithm (score threshold 0.15) gave a sensitivity of 93.6%, a specificity of 90.8%, a PPV of 46.5%, an NPV of 99.4%, and a TNR of 84.1%. Using the upper score threshold (0.54), Triage Plus had higher specificity (98.2%) and PPV (74.6%). Confusion matrices for the predicted performance Triage Plus algorithm versus tumor status confirmed by pathology are shown in Table S2.
When the lower and upper score thresholds for Triage Plus were used to define the probability of UC, the actual incidence of UC (confirmed by pathology) was 0.6% in low-probability samples (score < 0.15; 84.1% of samples), 27.7% in intermediate-probability samples (score ≥ 0.15 to <0.54; 9.5% of samples), and 74.6% in high-probability samples (score ≥ 0.54; 6.4% of samples).

3.3. Analytical Validation

The analytical validation of Triage Plus for detection of mRNA from IGFBP5, HOXA13, MDK, CDK1, and CXCR2 genes was the same as previously described for Cxbladder Detect and Triage [16]. The analytical validation of Triage Plus for detection of mutant DNA SNVs from FGFR3 and TERT is described in the following sections.

3.3.1. Linearity

Triage Plus demonstrated linearity across all analyzed FGFR3 and TERT SNVs, with an R2 value of >0.99 for all targets (Figure 1). The MSE values for individual SNVs ranged from 0.006 to 0.014. The tested range was 1 × 105 to 1 × 101 DNA copies/well, which corresponded to a maximum linear concentration of 5λ. The linearity of the assay was considered acceptable as it met the criteria for linearity in diagnostic testing.

3.3.2. Analytical Sensitivity

The predicted LOD of Triage Plus for DNA detection using the logistic regression approach was a mutant-to-WT DNA ratio of 1:840, 1:1200, 1:1250, and 1:970 for FGFR3 R248C, S249F/C, G372C, and Y375C, respectively, and a mutant-to-WT DNA ratio of 1:440 and 1:740 for TERT C228T and C250T, respectively (Table 1). Using the three concentrations approach, the predicted LOD was a mutant-to-WT DNA ratio of 1:632, 1:1220, 1:946, and 1:439 for FGFR3 R248C, S249F/C, G372C, and Y375C, respectively, and a mutant-to-WT DNA ratio of 1:319 and 1:418 for TERT C228T and C250T, respectively. There was no significant difference between the logistic regression and three concentrations approach for LOD prediction; however, the logistic regression approach was preferred for all SNVs except FGFR3 G372C, for which the logistic regression approach did not work well as there were too few negative samples.

3.3.3. Analytical Specificity

For sample-derived substances, RBCs had no significant effect on the extraction of HECs or LECs of FGFR3 and TERT at RBC concentrations below 4 × 106 cells/mL (Table 2). At RBC concentrations of 2 × 107 cells/mL, the FGFR3 and TERT mutant count per well was increased for HECs and decreased for LECs. At these RBC concentrations, all controls returned a positive Triage Plus result. At RBC concentrations of 1 × 108 cells/mL, there was significant impact on extraction efficiency; however, there was no loss of sensitivity as all HECs returned a positive Triage Plus result. LECs had some loss of sensitivity at RBC concentrations of 1 × 108 cells/mL, with the potential to return a false-negative test result.
The presence of clinically high levels of bacteria (E. coli 1 × 106 cells/mL), glucose (0.5 mg/mL), or yeast (1 × 104 CFU/mL) had no significant effect on the extraction of HECs and LECs of FGFR3 and TERT or assay performance. However, the presence of urea (60 mg/mL) had an impact on the extraction efficiency of HEC and increased the FGFR3 mutant count per well, but did not impact Triage Plus results. The presence of protein had an effect on the extraction efficiency and DNA detection, with the results for 1.25 mg/mL serum albumin showing a significant (but not meaningful) effect on extraction efficiency. For HECs, a positive result was returned at any protein (serum albumin) concentration; however, LECs were not guaranteed to return a positive result at serum albumin concentrations of ≥2.5 mg/mL. The required minimum WT count was not reached for samples containing ≥2.5 mg/mL serum albumin. For process-derived substances, all reagents showed a slight increase in assay performance, but were within the expected level of variation.
The substances that had the greatest impact on the assay (MagMAX wash buffer and Cxbladder stabilizing reagent) are the first and second reagents used in the extraction and purification steps, and were considered to have a low risk of being present at high concentrations at the elution step (Table S3). The substances that are closest to the elution step (acetone and MagMAX magnetic beads) did not significantly impact the assay at the highest concentrations tested.

3.3.4. Analytical Accuracy

The absolute quantification of FGFR3 mutant DNA versus mutant + WT DNA were within the 95% CI for the expected intra-plate variance for HECs of R248C and S249F/C and LECs of S249F/C and G372C (Table 3). However, higher intra-plate variance was observed for the LEC of R248C, HEC of G372C, and the HEC and LEC of Y373.
TERT mutant DNA versus mutant + WT DNA quantification showed acceptable intra-plate variance for HECs of C228T and C228T + C250T, but higher acceptable intra-plate variance for all LECs and the HEC of C250T.

3.3.5. Analytical Precision

Inter-assay variability showed mutant fraction CV% of ≤4.33% for HECs, with mutant fraction CV%s of 1.35–4.33% for FGFR3 and 1.01–2.11% for TERT; however, mutant fraction variance was slightly higher for LECs (mutant fraction CV%s of 4.62–6.97% and 3.42–6.86%, respectively; Table 4). Based on a maximum mutant fraction CV% of 6.97% for FGFR3 C228T and 6.86% for TERT C228T + C250T, the inter-assay variability was deemed acceptable.
For intra-assay variability, HECs had mutant fraction CV%s of ≤3.54% for FGFR3 and ≤2.95% for TERT (Table 4). LECs had higher mutant fraction CV%s than HECs (11.56–15.90% for FGFR3 and 11.79–20.27% for TERT). Based on a maximum mutant fraction CV% of 15.90% for FGFR3 G372C and 20.27% for TERT C228T + C250T, the intra-assay variability was deemed acceptable.
Total assay mutant fraction CV% for FGFR3 and TERT was ≤5.59% for HECs, and 12.45–17.24% and 12.27–21.40%, respectively, for LECs (Table 4). The maximum total mutant fraction variance was considered excellent for high-extraction controls and acceptable for low-extraction controls.
Lot-to-lot reagent variance was low for FGFR3 HECs (mutant fraction CV%s 1.14–2.53%), and within an acceptable range (CV%s 13.10–17.64%) for LECs (Supplementary Table S4). Similarly, lot-to-lot variance was low for TERT HECs (mutant fractions CV%s 0.84–1.74%), but was higher for LECs (CV%s 11.46–32.15%).

3.3.6. Extraction Efficiency

The extraction efficiency of Triage Plus for FGFR3 and TERT SNVs was lower in HECs versus LECs (Table 5). The FGFR3 samples had a mean extraction efficiency of 72.9% for high mutant-to-WT ratio controls and 88.0% for low mutant-to-WT ratio controls. The TERT samples had a mean extraction efficiency of 83.5% for high mutant-to-WT ratio controls and 95.5% for low mutant-to-WT ratio controls. No difference was observed between extracted samples and input controls, confirming that the extraction method did not introduce sampling bias.

3.3.7. Inter-Laboratory Evaluation

The inter-laboratory comparison between PEDNZ and PEDUSA was based on data from 33 samples. There was 87.9% concordance in clinical results for Triage Plus between the two laboratories using assay-derived CIs as the concordance metric.

4. Discussion

This analytical validation study showed that Cxbladder Triage Plus can accurately and reproducibly quantify the presence of six DNA SNVs of FGFR3 and TERT, as well as mRNA expression of five biomarker genes, to provide risk stratification for UC in patients with hematuria. All pre-specified analytical criteria were met, including analytical linearity, sensitivity, specificity, accuracy, and precision. The most notable impact on specificity was observed with protein contamination, which showed an effect on the accuracy of the quantitative ddPCR assay, but did not meaningfully change the ability to detect the presence of DNA mutations at concentrations of <2.5 mg/mL.
The original Cxbladder Triage assay was developed to provide high sensitivity and NPV when risk stratifying patients with hematuria for possible UC, while Cxbladder Detect was designed for high specificity [6,7]. The predicted performance of Triage Plus showed sensitivity and NPV that was as good as, or better than, Cxbladder Triage, as well as significantly higher specificity and PPV. Triage Plus also had improved performance over Cxbladder Detect. Furthermore, clinical validation of Triage Plus found that this assay provides improved specificity versus Cxbladder Triage and improved sensitivity versus Cxbladder Detect in a Veterans Affairs population with hematuria [19].
In the current study, Triage Plus showed acceptable accuracy for measurement of high and low controls containing FGFR3 and TERT mutant DNA. Analytical precision analysis showed acceptable inter-assay, intra-assay, and total assay variance, although variance was greater in low controls than in high controls. A study by Liu and colleagues similarly showed increased variance as the mutant frequency decreased [20]. Lot-to-lot reagent variance was low for HECs and acceptable for LECs. The analytical validity of Triage Plus was confirmed at a second laboratory (PEDUSA), with concordance between the two laboratories of 87.9%, consistent with that reported for Cxbladder Triage and Detect [16].
Of note, the presence of protein was found to have an impact on analytical specificity of Triage Plus and ddPCR-based DNA detection. While a DNA-positive call was returned at all protein concentrations for HECs, this could not be confidently asserted for LECs containing protein concentrations of ≥2.5 mg/mL. Although this is considered to be a very high urine protein concentration, patients with severe proteinuria can have protein concentrations as high as 30 mg/mL. The inhibitory effect of protein was observed for all DNA SNVs and resulted in test failure for WT only samples. For all samples that were impacted by protein contamination (i.e., patients with proteinuria), positive samples would return either a positive or “No Result” finding for Triage Plus. In clinical practice, the protein-inhibited result would not affect patient outcomes, as a “No Result” finding would require further evaluation for UC as part of standard of care.
The limitations of this study include its analytical validation design, which means that it does not provide clinical validation or describe the clinical utility of this assay. However, data from the recent DRIVE study demonstrated that Triage Plus has clinical validity for the risk stratification of patients presenting with hematuria [19].

5. Conclusions

This analytical validation study demonstrated that the urinary biomarker Cxbladder Triage Plus assay can accurately and reproducibly detect six DNA SNVs of FGFR3 and TERT, as well as mRNA expression of five biomarker genes that are associated with UC, from urine samples of patients with hematuria. This assay will provide clinicians with a non-invasive method of risk stratification in patients presenting for evaluation of hematuria, thereby allowing for more accurate assessment of UC risk in these patients.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics15141739/s1. Table S1: Demographics of patients with hematuria who were included in the algorithm development dataset (N = 987); Table S2: Confusion matrix of Cxbladder Triage Plus algorithm at the upper and lower threshold compared with tumor status confirmed by pathology in the algorithm development dataset (N = 987); Table S3: Analytical specificity of Cxbladder Triage Plus with control samples when mixed with process-derived interfering substances; Table S4: Lot-to-lot Cxbladder reagent variance for the six DNA single-nucleotide variants of FGFR3 and TERT in Cxbladder Triage Plus.

Author Contributions

Conceptualization, J.C.H., C.W.E., M.C. and J.A.H.; formal analysis, J.C.H., D.F., C.W.E., X.Z. and J.M.N.; methodology, J.C.H., D.F., C.W.E., M.C., J.A.H. and X.Z.; validation, J.C.H., C.W.E., M.C., J.A.H. and J.M.N.; writing—original draft, J.C.H.; writing—review and editing, J.C.H., D.F., C.W.E., M.C., J.A.H., X.Z. and J.M.N. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Pacific Edge Diagnostics, Ltd.

Institutional Review Board Statement

The previous clinical validation study and the STRATA study were both conducted in accordance with the Good Clinical Practice requirements and the Declaration of Helsinki (1975, revised 2013), and all patients provided informed consent prior to study entry. The clinical validation study protocols were approved in Singapore by the SingHealth Centralized Institutional Review Board (IRG; 2016/2471 (approved 19 July 2016, 13 June 2017, and 23 February 2019)) and NHB Doman Specific Review Board (2018/00234-SRF002 (approved 3 August 2019)) and in the US by relevant local IRBs (Chesapeake IRB, Pro00009623 (approved 13 December 2012, 29 January 2013, 6 February 2013, 3 June 2014, 24 February 2016, 9 May 2016, and 20 May 2016); Florida Hospital IRB, 394399 (approved 19 March 2013); UT Southwestern Medical Center IRB, STU-112012-018 (approved 4 April 2013); and PennState Hershey IRB, 41719EP (approved 5 June 2013)). The STRATA study protocol was approved by the PennState IRB (STUDY00010988 (approved 9 September 2019)), the University of Southern California IRB (HS-19-00766 (approved 14 June 2020)), the UT Southwestern Medical Center IRB (STU-2019-1020 (approved 11 September 2019)), the University of British Columbia Clinical Research Ethics Board (H19-01797 (approved 7 November 2019)), the University of Minnesota IRB (STUDY00008103 (approved 27 April 2020)), the Vanderbilt IRB (200304 (approved 16 June 2020)), the Western Research Health Sciences Research Ethics Board (114112 (approved 3 February 2020)), and the WCG IRB (20202112 (approved 20 July 2020, 2 August 2021, 21 November 2021, 19 October 2022, and 13 March 2023)).

Informed Consent Statement

All sample collection was performed with written informed consent obtained from the participants of the study.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors upon reasonable request.

Acknowledgments

We would like to thank Sarah Greig (Springer Health+), who assisted in the preparation of the outline and subsequent drafts of the manuscript and with post-submission revisions. This medical writing assistance was funded by Pacific Edge Diagnostics, Ltd.

Conflicts of Interest

Justin C. Harvey, Charles W. Ellen, Megan Colonval, Jody A. Hazlett, Xin Zhou, and Jordan Newell are employees of Pacific Edge Diagnostics, Ltd. or its subsidiaries. David Fletcher was a contractor for Pacific Edge Diagnostics, Ltd. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
AUAAmerican Urological Association
BARTBayesian additive regression tree
CDK1Cyclin-dependent kinase 1
CXCR2C-X-C motif chemokine receptor 2
CV%Coefficient of variation
ddPCRDroplet-digital PCR
FGFR3Fibroblast growth factor receptor 3
HECHigh-extraction control
HOXA13Homeobox A13
IGFBP5Insulin-like growth factor binding protein 5
LECLow-extraction control
LODLimit of detection
MDKMidkine
MSEMean squared error
NPVNegative predictive value
PPVPositive predictive value
R2Regression coefficient
RBCRed blood cell
SNVSingle-nucleotide variant
TERTTelomerase reverse transcriptase
TNRTest-negative rate
UCUrothelial carcinoma
WTWild type

References

  1. Barocas, D.A.; Lotan, Y.; Matulewicz, R.S.; Raman, J.D.; Westerman, M.E.; Kirkby, E.; Pak, L.J.; Souter, L.H. Updates to Microhematuria: AUA/SUFU guideline (2025). J. Urol. 2025, 213, 547–557. [Google Scholar] [PubMed]
  2. Mowatt, G.; Zhu, S.; Kilonzo, M.; Boachie, C.; Fraser, C.; Griffiths, T.R.; N’Dow, J.; Nabi, G.; Cook, J.; Vale, L. Systematic review of the clinical effectiveness and cost-effectiveness of photodynamic diagnosis and urine biomarkers (FISH, ImmunoCyt, NMP22) and cytology for the detection and follow-up of bladder cancer. Health Technol. Assess. 2010, 14, iii–iv. [Google Scholar]
  3. Freifeld, Y.; Lotan, Y. Effect of blue-light cystoscopy on contemporary performance of urine cytology. BJU Int. 2019, 124, 251–257. [Google Scholar] [PubMed]
  4. Soria, F.; Droller, M.J.; Lotan, Y.; Gontero, P.; D’Andrea, D.; Gust, K.M.; Roupret, M.; Babjuk, M.; Palou, J.; Shariat, S.F. An up-to-date catalog of available urinary biomarkers for the surveillance of non-muscle invasive bladder cancer. World J. Urol. 2018, 36, 1981–1995. [Google Scholar] [PubMed]
  5. Yafi, F.A.; Brimo, F.; Steinberg, J.; Aprikian, A.G.; Tanguay, S.; Kassouf, W. Prospective analysis of sensitivity and specificity of urinary cytology and other urinary biomarkers for bladder cancer. Urol. Oncol. 2015, 33, 66.e25–66.e31. [Google Scholar] [PubMed]
  6. Kavalieris, L.; O’Sullivan, P.J.; Suttie, J.M.; Pownall, B.K.; Gilling, P.J.; Chemasle, C.; Darling, D.G. A segregation index combining phenotypic (clinical characteristics) and genotypic (gene expression) biomarkers from a urine sample to triage out patients presenting with hematuria who have a low probability of urothelial carcinoma. BMC Urol. 2015, 15, 23. [Google Scholar]
  7. O’Sullivan, P.; Sharples, K.; Dalphin, M.; Davidson, P.; Gilling, P.; Cambridge, L.; Harvey, J.; Toro, T.; Giles, N.; Luxmanan, C.; et al. A multigene urine test for the detection and stratification of bladder cancer in patients presenting with hematuria. J. Urol. 2012, 188, 741–747. [Google Scholar] [PubMed]
  8. Holyoake, A.; O’Sullivan, P.; Pollock, R.; Best, T.; Watanabe, J.; Kajita, Y.; Matsui, Y.; Ito, M.; Nishiyama, H.; Kerr, N.; et al. Development of a multiplex RNA urine test for the detection and stratification of transitional cell carcinoma of the bladder. Clin. Cancer Res. 2008, 14, 742–749. [Google Scholar] [PubMed]
  9. Davidson, P.J.; McGeoch, G.; Shand, B. Assessment of a clinical pathway for investigation of haematuria that reduces the need for cystoscopy. N. Z. Med. J. 2020, 133, 71–82. [Google Scholar] [PubMed]
  10. Hayashi, Y.; Fujita, K.; Matsuzaki, K.; Matsushita, M.; Kawamura, N.; Koh, Y.; Nakano, K.; Wang, C.; Ishizuya, Y.; Yamamoto, Y.; et al. Diagnostic potential of TERT promoter and FGFR3 mutations in urinary cell-free DNA in upper tract urothelial carcinoma. Cancer Sci. 2019, 110, 1771–1779. [Google Scholar] [PubMed]
  11. Pietzak, E.J.; Bagrodia, A.; Cha, E.K.; Drill, E.N.; Iyer, G.; Isharwal, S.; Ostrovnaya, I.; Baez, P.; Li, Q.; Berger, M.F.; et al. Next-generation Sequencing of Nonmuscle Invasive Bladder Cancer Reveals Potential Biomarkers and Rational Therapeutic Targets. Eur. Urol. 2017, 72, 952–959. [Google Scholar] [PubMed]
  12. Hayashi, Y.; Fujita, K.; Matsuzaki, K.; Eich, M.L.; Tomiyama, E.; Matsushita, M.; Koh, Y.; Nakano, K.; Wang, C.; Ishizuya, Y.; et al. Clinical significance of hotspot mutation analysis of urinary cell-free DNA in urothelial bladder cancer. Front. Oncol. 2020, 10, 755. [Google Scholar]
  13. Knowles, M.A. FGFR3—A central player in bladder cancer pathogenesis? Bladder Cancer 2020, 6, 403–423. [Google Scholar]
  14. Lotan, Y.; Raman, J.D.; Konety, B.; Daneshmand, S.; Schroeck, F.; Shariat, S.F.; Black, P.; de Lange, M.; Asroff, S.; Goldfischer, E.; et al. Urinary analysis of FGFR3 and TERT gene mutations enhances performance of Cxbladder tests and improves patient risk stratification. J. Urol. 2023, 209, 762–772. [Google Scholar] [PubMed]
  15. Teutsch, S.M.; Bradley, L.A.; Palomaki, G.E.; Haddow, J.E.; Piper, M.; Calonge, N.; Dotson, W.D.; Douglas, M.P.; Berg, A.O.; EGAPP Working Group. The Evaluation of Genomic Applications in Practice and Prevention (EGAPP) Initiative: Methods of the EGAPP Working Group. Genet. Med. 2009, 11, 3–14. [Google Scholar] [PubMed]
  16. Harvey, J.C.; Cambridge, L.M.; Ellen, C.W.; Colonval, M.; Hazlett, J.A.; Newell, J.; Zhou, X.; Guilford, P.J. Analytical validation of Cxbladder® Detect, Triage, and Monitor: Assays for detection and management of urothelial carcinoma. Diagnostics 2024, 14, 2061. [Google Scholar] [CrossRef] [PubMed]
  17. Lotan, Y.; Daneshmand, S.; Shore, N.; Black, P.; Scarpato, K.R.; Patel, A.; Lough, T.; Shoskes, D.A.; Raman, J.D. A multicenter prospective randomized controlled trial comparing Cxbladder Triage to cystoscopy in patients with microhematuria: The Safe Testing of Risk for Asymptomatic Microhematuria trial. J. Urol. 2024, 212, 41–51. [Google Scholar] [PubMed]
  18. Hill, J.; Linero, A.; Murray, J. Bayesian additive regression trees: A review and look forward. Ann. Rev. Stat. Appl. 2020, 7, 251–278. [Google Scholar]
  19. Savage, S.J.; Ercole, C.E.; Hemstreet, C.E.; Leone, A.; Masterson, T.; McWilliams, G.; Risk, M.; Schroeck, F.; Stratton, K.; Lough, T.; et al. The prognostic performance of Cxbladder Triage Plus for the identification and priority evaluation of veterans at risk for urothelial carcinoma: The DRIVE study [podium presentation #8]. In Proceedings of the Urological Society for American Veterans, Las Vegas, NV, USA, 28 April 2025. [Google Scholar]
  20. Liu, Y.; Han, C.; Li, J.; Xu, S.; Xiao, Z.; Guo, Z.; Rao, S.; Yao, Y. Laboratory-developed droplet digital PCR assay for quantification of the JAK2 (V617F) mutation. Glob. Med. Genet. 2024, 11, 132–141. [Google Scholar] [PubMed]
Figure 1. Measured versus known DNA concentrations for the six single-nucleotide variants of FGFR3 and TERT. conc, concentration; FGFR3, fibroblast growth factor receptor 3; MSE, mean squared error; R2, regression coefficient; TERT, telomerase reverse transcriptase.
Figure 1. Measured versus known DNA concentrations for the six single-nucleotide variants of FGFR3 and TERT. conc, concentration; FGFR3, fibroblast growth factor receptor 3; MSE, mean squared error; R2, regression coefficient; TERT, telomerase reverse transcriptase.
Diagnostics 15 01739 g001
Table 1. Limit of detection (analytical sensitivity) of Cxbladder Triage Plus for the six DNA single nucleotide variants of FGFR3 and TERT.
Table 1. Limit of detection (analytical sensitivity) of Cxbladder Triage Plus for the six DNA single nucleotide variants of FGFR3 and TERT.
SNVEstimated LOD (95% CI)
Logistic RegressionThree Concentrations
FGFR3, mutant-to-WT DNA ratio
 R248C1:840 (1:600, 1:980)1:632 (1:468, 1:978)
 S249F/C1:1200 (1:860, 1:1370)1:1220 (1:806, 1:2779)
 G372C1:1250 a1:946 (1:630, 1:2263)
 Y375C1:970 (NE b, 1:1220)1:439 (1:286, 1:2128)
TERT, mutant-to-WT DNA ratio
 C228T1:440 (1:250, 1:520)1:319 (1:217, 1:1494)
 C250T1:740 (1:560, 1:810)1:418 (1:300, 1:651)
a The logistic regression approach did not work well for FGFR3 G372C as there were too few negative samples. b The lower 95% CI was NE due to large variations in the data. CI, confidence interval; FGFR3, fibroblast growth factor receptor 3; LOD, limit of detection; NE, not estimable; SNV, single-nucleotide variant; TERT, telomerase reverse transcriptase; WT, wild type.
Table 2. Analytical specificity of Cxbladder Triage Plus for FGFR3 and TERT DNA control samples when mixed with sample-derived interfering substances.
Table 2. Analytical specificity of Cxbladder Triage Plus for FGFR3 and TERT DNA control samples when mixed with sample-derived interfering substances.
SubstanceFGFR3 Mutant Count per WellTERT Mutant Count per Well
MeanDifference ap-Value bMeanDifference ap-Value b
RBCs, cells/mL
 HECs767.60692.14
  8 × 105763.13−4.480.894729.00+36.860.120
  4 × 106781.38+13.780.694733.63+41.480.120
  2 × 107847.00+79.400.049811.25+119.110.000
  1 × 108535.63−231.980.000468.50−223.640.000
 LECs33.0031.71
  8 × 10531.43−1.570.73030.50−1.210.808
  4 × 10624.33−8.670.07426.00−5.710.043
  2 × 10723.86−9.140.05320.50−11.210.003
  1 × 1086.13−26.880.0003.17−28.550.000
Bacteria (E. coli), cells/mL
 HECs662.20691.38
  1 × 106716.63+54.430.063694.43+3.050.834
 LECs14.2512.71
  1 × 10614.250.001.00010.63−2.090.205
Yeast, CFU/mL
 HECs662.20691.38
  1 × 104696.63+34.430.064667.00−24.380.157
 LECs14.2512.71
  1 × 10414.14−0.110.9359.38−3.340.020
Urea, mg/mL
 HECs662.20691.38
  60736.75+74.550.010701.43+10.050.470
 LECs14.2512.71
  6014.63+0.380.81313.50+0.790.638
Glucose, mg/mL
 HECs662.20691.38
  0.5756.57+94.370.000706.71+15.340.365
 LECs14.2512.71
  0.517.38+3.130.16914.71+2.000.217
Protein (serum albumin), mg/mL
 HECs683.50754.88
  1.25594.25−89.250.000630.13−124.750.000
  2.508.88−674.630.0007.63−747.250.000
  5.006.29−677.210.0003.88−751.000.000
  10.004.29−679.210.0004.63−750.250.000
 LECs15.0010.83
  1.2511.75−3.250.25613.13+2.290.174
  2.500.00−15.000.0020.00−10.830.000
  5.000.00−15.000.0020.00−10.830.000
  10.000.00−15.000.0020.00−10.830.000
a Calculated by subtracting the mean score for the control sample from that of each contaminated sample. b Two-sided t test. CFU, colony-forming unit; FGFR3, fibroblast growth factor receptor 3; HECs, high-extraction controls; LECs, low-extraction controls; RBC, red blood cell; TERT, telomerase reverse transcriptase.
Table 3. Analytical accuracy of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
Table 3. Analytical accuracy of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
SNVDNA Concentration, Copies/μLDifferenceInaccuracy
(95% CI), %
Mutant DNAMutant + WT DNA
FGFR3
 R248C
  HEC a237.69239.081.380.6 (−11.8, 13.0)
  LEC b11.1614.383.2228.9 (14.5, 43.3)
 S249F/C
  HEC a193.48201.648.174.2 (−6.7, 15.2)
  LEC b10.1810.500.323.1 (−11.0, 17.3)
 G372C
  HEC a197.82225.1527.3413.8 (3.6, 24.0)
  LEC b11.2510.09−1.16−10.3 (−22.4, 1.8)
 Y375C
  HEC a232.29247.0614.776.4 (1.3, 11.5)
  LEC b8.7612.463.7042.2 (18.2, 66.3)
TERT
 C228T
  HEC a229.43237.828.393.7 (−2.5, 9.9)
  LEC b10.4113.433.0229.0 (14.5, 43.4)
 C250T
  HEC a187.32216.3529.0315.5 (9.5, 21.5)
  LEC b9.0811.812.7430.2 (19.9, 40.4)
 C228T + C250T
  HEC a226.62240.6214.06.2 (−12.8, 25.1)
  LEC b9.4115.165.7561.1 (28.7, 93.5)
a DNA concentration of ~1 × 106 copies/μL and mutant-to-WT ratio of 1:10 for combined sample. b DNA concentration of ~1 × 104 copies/μL and mutant-to-WT ratio of 1:200 for combined sample. CI, confidence interval; FGFR3, fibroblast growth factor receptor 3; HEC, high-extraction control; LEC, low-extraction control; SNV, single-nucleotide variant; TERT, telomerase reverse transcriptase; WT, wild type.
Table 4. Analytical precision of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
Table 4. Analytical precision of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
SNVMutant Fraction Variance
Inter-Assay VarianceIntra-Assay VarianceTotal Assay Variance
Mean (SD)CV%Mean (SD)CV%Mean (SD)CV%
FGFR3
 R248C
  HEC a0.1047 (0.0014)1.350.1100 (0.0025)2.390.1047 (0.0029)2.74
  LEC b0.0063 (0.0003)4.640.0063 (0.0008)13.220.0063 (0.0009)14.01
 S249F/C
  HEC a0.0808 (0.0035)4.330.0808 (0.0029)3.540.0808 (0.0045)5.59
  LEC b0.0043 (0.0002)4.620.0043 (0.0005)11.560.0043 (0.0005)12.45
 G372C
  HEC a0.0909 (0.0014)1.500.0909 (0.0029)3.150.0909 (0.0032)3.49
  LEC b0.0053 (0.0004)6.680.0053 (0.0008)15.900.0053 (0.0009)17.24
 Y375C
  HEC a0.0965 (0.0028)2.890.0964 (0.0028)2.900.0964 (0.0040)4.10
  LEC b0.0058 (0.0004)6.970.0057 (0.0009)15.020.0057 (0.0009)16.56
TERT
 C228T
  HEC a0.1033 (0.0014)1.380.1033 (0.0025)2.450.1033 (0.0029)2.82
  LEC b0.0063 (0.0004)6.180.0063 (0.0008)12.420.0063 (0.0009)13.87
 C250T
  HEC a0.0913 (0.0019)2.110.0913 (0.0023)2.540.0913 (0.0030)3.30
  LEC b0.0055 (0.0002)3.420.0055 (0.0007)11.790.0055 (0.0007)12.27
 C228T + C250T
  HEC a0.0933 (0.0009)1.010.0933 (0.0027)2.950.0933 (0.0029)3.11
  LEC b0.0054 (0.0004)6.860.0054 (0.0011)20.270.0054 (0.0012)21.40
a DNA concentration of ~1 × 106 copies/μL and mutant-to-WT ratio of 1:10. b DNA concentration of ~1 × 104 copies/μL and mutant-to-WT ratio of 1:200. CV%, coefficient of variation; FGFR3, fibroblast growth factor receptor 3; HEC, high-extraction control; LEC, low-extraction control; SD, standard deviation; SNV, single-nucleotide variant; TERT, telomerase reverse transcriptase; WT, wild type.
Table 5. Extraction efficiency of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
Table 5. Extraction efficiency of Cxbladder Triage Plus for the six DNA single-nucleotide variants of FGFR3 and TERT.
SNVExtraction Efficiency (95% CI), %
FGFR3
 R248C
  HEC a72.7 (69.3, 76.0)
  LEC b82.4 (51.4, 100.0)
 S249F/C
  HEC a62.9 (53.3, 72.5)
  LEC b84.2 (59.6, 100.0)
 G372C
  HEC a77.6 (69.5, 85.7)
  LEC b100.0 (100.0, 100.0)
 Y375C
  HEC a78.6 (70.1, 87.2)
  LEC b85.6 (60.7, 100.0)
TERT
 C228T
  HEC a86.4 (76.1, 96.6)
  LEC b97.5 (71.1, 100.0)
 C250T
  HEC a83.4 (76.7, 90.1)
  LEC b95.2 (69.8, 100.0)
 C228T + C250T
  HEC a80.7 (70.5, 90.9)
  LEC b93.8 (52.6, 100.0)
a DNA concentration of ~1 × 106 copies/μL and mutant-to-WT ratio of 1:10. b DNA concentration of ~1 × 104 copies/μL and mutant-to-WT ratio of 1:200. CI, confidence interval; FGFR3, fibroblast growth factor receptor 3; SNV, single-nucleotide variant; TERT, telomerase reverse transcriptase; WT, wild type.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Harvey, J.C.; Fletcher, D.; Ellen, C.W.; Colonval, M.; Hazlett, J.A.; Zhou, X.; Newell, J.M. Analytical Validation of the Cxbladder® Triage Plus Assay for Risk Stratification of Hematuria Patients for Urothelial Carcinoma. Diagnostics 2025, 15, 1739. https://doi.org/10.3390/diagnostics15141739

AMA Style

Harvey JC, Fletcher D, Ellen CW, Colonval M, Hazlett JA, Zhou X, Newell JM. Analytical Validation of the Cxbladder® Triage Plus Assay for Risk Stratification of Hematuria Patients for Urothelial Carcinoma. Diagnostics. 2025; 15(14):1739. https://doi.org/10.3390/diagnostics15141739

Chicago/Turabian Style

Harvey, Justin C., David Fletcher, Charles W. Ellen, Megan Colonval, Jody A. Hazlett, Xin Zhou, and Jordan M. Newell. 2025. "Analytical Validation of the Cxbladder® Triage Plus Assay for Risk Stratification of Hematuria Patients for Urothelial Carcinoma" Diagnostics 15, no. 14: 1739. https://doi.org/10.3390/diagnostics15141739

APA Style

Harvey, J. C., Fletcher, D., Ellen, C. W., Colonval, M., Hazlett, J. A., Zhou, X., & Newell, J. M. (2025). Analytical Validation of the Cxbladder® Triage Plus Assay for Risk Stratification of Hematuria Patients for Urothelial Carcinoma. Diagnostics, 15(14), 1739. https://doi.org/10.3390/diagnostics15141739

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop