1. Introduction
The ubiquitous occurrence of microplastics (MP) in surface waters around the globe, even in remote areas, is widely studied [
1,
2,
3]. Concern about their potential adverse effects on human and environmental health reinforces the need for reliable concentration data [
4,
5,
6]. Different spot sampling techniques (e.g., nets, pumps, containers) have been used to identify MP. While many snapshot investigations prove that MP occur in water bodies, there is little long-term data and basically no information on spatial and temporal variation [
7]. Overall, sampling and quantifying different polymers in water bodies is challenging due to low concentrations, the microscopic size, and the need for a reliable distinction from “natural” suspended particulate matter (SPM).
Several MP-focused meta-studies highlight the large variety of sampling, sample preparation, and detection methods (e.g., Lu et al. [
8] or Hidalgo-Ruz et al. [
9]). These different methods significantly influence the obtained results, leading to studies not being comparable with each other (see, e.g., Primpke et al. [
10] for the vast result differences between pyrolysis gas chromatography/mass spectrometry and hyperspectral Fourier-transform infrared spectroscopy (FTIR) imaging spectroscopy).
The commonly used sampling methods for MP from water bodies cover a limited water volume and usually a rather short time span (minutes to hours). Frequent spot sampling would be required to determine a time-averaged MP concentration, which is needed to calculate MP fluxes in rivers. Alternatively, time-integrated sampling methods such as passive samplers cover larger time spans (from days to weeks). Within this study, the term “passive sampler” refers to a sampling device through which water flows and SPM is separated from the water phase without using manual or automated labor in the device itself.
Passive samplers for particulate matter are devices that generally work with the principle of sedimentation due to a decreased flow velocity within the sampler induced by an expansion of the inner diameter. They are well-known for the sampling of SPM [
11] and may also be suitable to sample MP. Nevertheless, current literature reveals that passive samplers are insufficiently validated to provide quantitative results on both SPM and MP in surface waters.
The most commonly used passive sampler is the TIMS (time-integrated mass-flux sediment) sampler, which is directly applied in flowing water and has been developed by Phillips et al. [
12]. The sampler has been used to sample SPM from rivers and creeks. However, only a few studies have assessed the retention rates under controlled laboratory conditions [
12,
13,
14,
15]. A comparison of the data reveals that retention rates decrease with increasing flow velocity and decreasing particle size. Hence, they indicate a systematic response of the devices’ retention performance to altering parameters. Other parameters that are suspected to have an impact on retention rates have not been considered yet, such as the density or shape of the used material or the influence of organic material or flocs.
Furthermore, there have been attempts to validate the TIMS sampler in the field [
12,
14,
15,
16,
17,
18,
19,
20,
21,
22]. For instance, Perks et al. [
19] compared TIMS samples to SPM concentrations estimated with a previously calibrated turbidity sensor. Although there is a statistically significant correlation between both methods, they determine a level of underestimation of 66–96% for the concentration of suspended solids by the TIMS sampler. However, given the typically high variability of the flow velocity in rivers, retention rates can be expected to change during the sampling period, and the total sampled volume cannot be determined reliably. Without having information on the flow, it is impossible to determine the concentration or flux of SPM.
Besides the TIMS sampler, there are other passive samplers which are directly placed in flowing water, such as the BiSam sampler [
23,
24,
25] or are typically operated in monitoring stations, such as the sedimentation tank [
23,
24].
In general, passive samplers like the sedimentation box can also be operated with a pump to sample SPM from water bodies (compare ISO 5667-23, 2011 [
26]). This allows for a constant flow velocity inside the sampling device and the monitoring of the sampled water volume. Sedimentation boxes have been designed by the German Environmental Specimen Bank and are applied in the routine monitoring of SPM from rivers [
11].
Harhash et al. [
27] investigate the retention performance of sedimentation boxes operated with a pump. They performed laboratory experiments with different mineral, organic, and MP particles of varying sizes. Masson et al. [
28] compared sedimentation box samples to continuous flow centrifuge samples. Both studies found significantly coarser particles in the sedimentation box, underscoring the higher retention rate of larger particles compared to smaller particles.
The suitability of passive samplers to collect SPM is documented in many studies and we hypothesize that these experiences are transferrable to MP. SPM does not include or exclude specific materials; it rather describes the physical state of particulate matter. Hence, MP and their different properties add to the range of physical properties of SPM. SPM includes a wide range of materials and their respective properties (e.g., see Koelmans et al. [
29] for different categories of natural particles), and we state that the differences between MP and other materials in SPM are in line with differences between materials described as SPM.
The density of organic matter such as wood, leaves, or algal debris ranges from 0.9 to 1.3 g/cm
3 [
30]. This range is similar to the density range of polymers in general and to the densities of our focus polymers: polypropylene (PP, 0.91–0.94 g/cm
3, [
31]), polyethylene (PE, 0.91–0.97 g/cm
3, [
32]), PS (1.05–1.06 g/cm
3, [
33]), and tire and road wear particles (TRWP, around 1.8 g/cm
3, [
34]). In contrast to particulate organic matter, inorganic particles are mostly minerals and have densities between 2.5 g/cm
3 and 3.0 g/cm
3. However, biofouling may alter the density of particles. Several studies have demonstrated that biofilm growth leads to both the sinking of formerly buoyant MP particles and the resuspension of formerly sinking MP particles [
35,
36,
37,
38]. Furthermore, the adhesion of minerals on MP particles or the formation of aggregates can increase their density and facilitate sedimentation. It has been found that this process may be enhanced through biofouling [
39] or the presence of surfactants [
40].
In the UK, Walling et al. [
41] found over 95% of the SPM in most of the investigated rivers to have a particle size below 63 µm. This is in good accordance with other studies, e.g., Rügner et al. [
42], finding median particle sizes between 5 and 34 µm for SPM in rivers in South-West Germany, or Le et al. [
43], who determined median SPM sizes between 10 and 20 µm in the lower Mekong, mostly as flocs. By definition, MP particles are in range with these particle sizes, and investigations on MP particle size in water bodies show similar size ranges. Crew et al. [
44] found that 95% of all sampled microplastic particles had a size below 400 μm; in the MiWa project, even 96% were below 20 μm [
45].
The flow characteristics (e.g., velocity, turbulence) of the surrounding water define both the size and density characteristics of SPM and thus, the transport mechanisms of MP particles [
46]. MP is present in various shapes. Fibers, foils, and particles most likely behave differently in similar flow conditions. This adds to the difficulty of assessing polymer-specific transport patterns. All that considered, it seems easier to directly measure MP in SPM than to predict if certain MP will end up in SPM.
The aim of our study is to investigate the potential of sedimentation boxes as time-integrative sampling devices for MP in surface waters. To approximate actual MP concentrations in surface waters and to reduce the uncertainty in quantitative MP results, it is pivotal to understand the retention of MP in the sedimentation boxes. We operate them with a pump to reduce the limitations of currently used passive samplers.
We conducted laboratory experiments under controlled conditions to gain additional insights into MP retention rates. Then we carried out several sampling campaigns at rivers. The first step was to apply sedimentation boxes directly in the river, which provided qualitative MP results. We then operated the boxes with a pump in regular river monitoring stations. Taken together with the laboratory experiments, we aim to quantify MP concentrations in the sampled rivers.
2. Materials and Methods
Between 2019 and 2021, we applied sedimentation boxes in various experiments to sample MP from different rivers in Central and Eastern Europe. In this study, we present the results from three different monitoring campaigns as examples of the application of sedimentation boxes. The first sampling campaign was part of the Joint Danube Survey 4 (JDS4) in 2019. The survey was carried out to assess water quality parameters along the River Danube and its main tributaries. Sedimentation boxes were used to assess MP abundance and its regional variation. Further information can be found in Kittner et al. [
47]. While they used the MP concentration data to compare the pollution of the rivers, we evaluated the data to derive information on the sampling performance of the sedimentation boxes, taking the properties of the sampled rivers into account.
In 2020, we carried out further monitoring campaigns at the River Elbe and Teltowkanal. The aim of these campaigns was to assess the sampling performance of the boxes and develop a robust sampling strategy in terms of handling and workload. In addition, we determined the retention rates of MP particles under controlled conditions in laboratory experiments.
2.1. Sedimentation and Operating Principle of Passive Samplers
To elucidate the sensitivity of different parameters affecting the retention of SPM in sedimentation boxes, we give a brief overview of the physical basics of sedimentation and the operating principle of passive samplers.
The behavior of a suspended particle in a stationary fluid can be described by Stokes’ law [
48]. Due to gravity, different forces act on the particle and its terminal sinking or rising velocity is determined by the relation between gravitational forces (F
G) and the sum of buoyant (F
B) and frictional (F
F) forces (see
Figure 1a). Among other parameters, these forces are not only dependent on the size (volume V
p and diameter d
p) but also on the density ρ of the particle (only particles with a density larger than water are supposed to sink). While frictional forces are proportional to the particle diameter, gravitational forces are proportional to the particle diameter in third power. Hence, the terminal sinking velocity of a particle is proportional to the particle diameter in second power (see
Figure 1b), and larger particles settle down or rise faster than smaller particles. Moreover, in turbulent fluids, the smaller the particle, the stronger its movement is influenced by the surrounding flow.
In passive samplers with a horizontal design, particles are transported horizontally with the same velocity as the surrounding water. Simultaneously, they move vertically with their terminal sinking (or rising) velocity v
s. Hence, depending on their specific sinking velocity, particles are retained if they sediment before reaching the outlet of the sampler, i.e., if their retention time is long enough (see
Figure 1c). In vertical passive samplers, particles are retained if their terminal sinking velocity v
s is larger than the vertical flow velocity v
d (see
Figure 1b).
In a sedimentation box, the flow direction of the water is continuously changed by vertical baffles.
Figure 2 shows a simulation of the flow inside the box. In the first chamber, the jet hits the first baffle, and the flow becomes fully turbulent. Due to the expansion of the box in comparison to the inlet, the flow velocity decreases and the flow becomes more laminar in the subsequent chambers. In the second and third chambers, the flow alters between a horizontal and vertical direction. Areas with a particularly low flow velocity are close to the walls and in the corners of the box. If a particle reaches these areas, it can be trapped in the sampler without being fully sedimented. This is illustrated in
Figure 3, where the particles accumulate in the corners of the device.
In consequence, the retention of particles in the sedimentation box depends on the size and density of the particles as well as the flow velocity inside the box. The sinking velocity, and thus, the expected retention of particles increases with increasing diameter and density. Furthermore, the retention time of a particle increases with decreasing flow velocity inside the box. If it is lower than in the sampled river, formerly suspended particulate matter can be retained by sedimentation (provided their density is higher than the density of water).
We used Stokes’ law to estimate a suitable flow within the sedimentation box. With an expansion of the inner diameter by the factor 34, our chosen flow rate of 5.2 L/min produces a flow velocity of 0.004 m/s inside the box and should lead to a complete retention of PS particles > 454 µm (considering a density of 1.04 g/cm3). A flow rate of 2.6 L/min produces a flow velocity of 0.002 m/s inside the box and should lead to the complete retention of PS particles > 321 µm. Smaller particles should be retained partially with the respective flow rates.
The aforementioned calculations strongly simplify the fate of SPM in the sedimentation box and neglect factors such as different particle shapes, adhesion to walls, position within the flow, or particle concentration. Furthermore, when using passive samplers directly in the river, the flow velocity and flow pattern inside the device might not be constant due to a varying flow in the river.
2.2. Laboratory Experiment Methods
We conducted laboratory experiments to determine the retention rates of sedimentation boxes for PS and PE particles. For the experiments with PS, we produced artificial suspensions of reverse osmosis water and PS particles with a median particle size of 250 µm and a density of 1.04 g/cm3. To produce a stock suspension, we heated 2 L of reverse osmosis water, added 100 g Tween80®, left it to boil for approx. 5 minutes, and stirred it for 24 h at ambient temperature. Tween80© is a surfactant which is based on Polysorbate 80 and facilitates the suspension of the hydrophobic PS particles. Subsequently, we added 90 g PS particles. For each experiment, we added the 2 L stock suspension to 2 m3 reverse osmosis water, resulting in a concentration of 45 mg/L PS. NaCl was added until it reached a concentration of 0.1 g/L to increase electrical conductivity. During the experiments, the suspension was automatically stirred (agitator: Kiesel, Heilbronn, Germany, 1400–1680 rpm) in two stainless-steel barrels, which were connected to each other at the bottom.
The experiments with PE were carried out similarly; however, we produced a stock suspension with surface water. The water was taken at the surface of a local lake in Berlin-Marienfelde, Germany. We added 45 g of UHMW-PE (Ultra High Molecular Weight Polyethylene) with a median particle size of 147 µm and a density of 0.94 g/cm3 to 2 L surface water. At first, the PE particles remained on the water surface but were suspended after one day of constant stirring at ambient temperature. We added the stock suspension to 1 m3 surface water, resulting in a concentration of 45 mg/L PE. The suspension was automatically stirred in a stainless-steel barrel.
We used three parallel sedimentation boxes for the PS experiments. Due to the reproducible results of the parallel boxes, we conducted the PE experiment with a single sedimentation box to reduce the workload. For the PS experiments, a submersible stainless-steel pump (Optima, Ebara pumps) was connected to a stainless-steel distribution piece, which divides the flow into three approximately equal parts. For the PE experiment, the submersible pump was connected directly. One-inch needle valves made of brass were used to regulate the flow at the inlet of the sedimentation boxes. Behind the sedimentation boxes, electromagnetic flowmeters (Picomag, Endress & Hauser, Reinach, Switzerland) measured flow and total volume. Sieves with a mesh size of 50 µm (Bückmann, Mönchengladbach, Germany) were used at the outlet of the system. These were placed approx. 50 cm above ground to guarantee completely filled sedimentation boxes.
A total of 600 L of the MP suspension was pumped through each of the sedimentation boxes at a flow of 5.2 L/min for run 1 and 2.6 L/min for run 2 of the PS experiments and with a flow rate of 5.2 L/min for the PE experiment. Before and after each run, the sedimentation boxes were flushed with 100 L MP-free reverse osmosis water for PS and surface water for PE. After the completion of each experiment, the retained material on the outlet sieves was transferred into glass bottles. The content of each sedimentation box was directed over 5 µm sieves using a vacuum pump. The remaining suspension in the barrels was directed over 50 µm sieves. All samples were frozen at −20 °C, freeze-dried at −60 °C and 1 mbar (Delta 1–24 LSCplus, Christ, Osterode am Harz, Germany), weighed, and the PSD was determined (Partica LA 960, HORIBA, Kyoto, Japan).
2.3. Field Experiment Methods
For the monitoring campaigns, the boxes were either placed directly in the stream, using the natural flow of the river (method “in situ”) or operated with a pump (method “pump”). We used slightly different models at the different sampling sites. All samples obtained within the Joint Danube Survey 4 (JDS4) in 2019 were taken using the in situ method. Due to practical reasons (e.g., sampling sites without access to electricity), the JDS4 team placed the boxes directly into the stream. The used sedimentation boxes had a volume of 67 L, six 1″ inlets, and four 1″ outlets. They were placed approx. 0.5 m below the water surface and operated for 14 days. The flow velocity inside the box decreases with a factor of 37. Further information can be found in Kittner et al. [
47].
To evaluate whether the JDS4 sampling method is transferable to the Teltowkanal, we investigated the flow velocity in the river. Therefore, we used an acoustic Doppler velocimeter and measured the flow velocity at five spots in front of an installed sedimentation box for a total of 65 min. The flow velocity showed significant differences with a minimum minute average of −0.22 m/s, a maximum minute average of 0.09 m/s, and a mean minute average flow velocity of −0.01 m/s during the sampling period.
Based on these results, the corresponding uncertainty about the flow through the device and the associated loss of information when using the sedimentation boxes directly in the stream, we used a pump-based setup in river monitoring stations for the sampling campaigns at the River Elbe and Teltowkanal in Germany. The sampling locations were the monitoring station Cumlosen, lying on the right side of the River Elbe at km 470, and Kleinmachnow, lying on the left side of Teltowkanal at km 6 (kilometers at Teltowkanal are counted opposite to the flow direction, from west to east). Three parallel boxes were operated at each sampling location to assess the reproducibility.
For the pump method, we used a setup, which is similar to the one used for the laboratory experiments. We used boxes with a single 1″ inlet, to facilitate the connection with the hose. At Teltowkanal, we used boxes with a volume of approx. 60 L and an expansion factor of 78; at the River Elbe, we used smaller boxes with a volume of approx. 20 L and an expansion factor of 34. We connected the boxes to the existing circular pipe system of the monitoring station. Using a ball valve, we set a constant flow velocity of 5.2 L/min (or 0.002 m/s in the large and 0.004 m/s in the small box) for the sampling duration of 48 h and controlled it with an electromagnetic flowmeter (Picomag, Endress & Hauser, Reinach, Switzerland). We chose this flow velocity and sampling duration based on our initial calculation and the subsequent results of preliminary experiments, where we could show that enough sedimentary material for all further analyses could be retained within the sedimentation box. Furthermore, lower flow velocities (e.g., 2.6 L/min) sometimes caused clogging in the ball valve, leading to an irregular flow.
After the completion of the sampling period, we separated the content of the boxes by fractionated filtration, using a sequence of stainless-steel sieves (mesh size 500 and 100 µm for the in situ method; 500, 100, 50, and 10 µm for the pump method). For the filtration over 10 µm, the application of a vacuum was needed, and due to the high solids content and filter cake formation, we used aliquots representing 10% of the total sample volume for this fraction. We analyzed the filtrate of this step to obtain the <10 µm fraction. Samples taken with the pump method were frozen at −20 °C and subsequently freeze-dried at −60 °C and 1 mbar (Delta 1–24 LSCplus freeze-dryer, Christ, Osterode am Harz, Germany). Samples from the in situ method were air-dried in an oven at 40 °C. We weighed all subsamples and homogenized them using a mortar and pestle after drying.
The total organic carbon (TOC) of the samples was determined individually for most fractions, following DIN EN 15936 and using the “varioTOC” (Elementar, Langenselbold, Germany). Between 2.6 and 43.8 mg of the subsamples were weighed for triplicate determination onto silver foil and treated with 10% hydrochloric acid to remove inorganic carbon. After air-drying at 40 °C, samples were combusted at 950 °C under oxygen, and the combustion products were transferred to a catalyst (copper oxide) to achieve complete oxidation to CO2. After a drying step and removal of other volatile compounds by adsorption, CO2 is quantified by an infrared detector (NDIR, nondispersive infrared sensor, at 38 °C). The NDIR was calibrated using a soil reference material (0.651% TOC, calibration range 0.0058–0.916 mg TOC) and potassium hydrogen phthalate (47.05% TOC, calibration range 0.48–3.79 mg TOC). The standard deviation of the method is typically less than 10%.
To quantify MP content, we carried out the polymer analysis with a thermogravimetric detection method, namely the TED-GC/MS (thermal extraction desorption-gas chromatography/mass spectrometry). Although the method is suitable for analyzing a larger number of polymer types, we could only unambiguously detect PE, PP, PS, and styrene-butadiene rubber (SBR), a marker for TRWP. The sample analysis with TED-GC/MS provides information on polymer types and masses, yet information on the number, size, or shape is lost. However, we can roughly approximate the size according to the mass results of the fractionated filtration. Due to the analysis process, the minimum mass reflecting the limit of detection is different for each polymer. Detailed information on polymer analysis with TED-GC/MS is provided by Kittner et al. [
47].
To characterize the river during the sampling period, information on the annual discharge at or close to the sampling site was derived from different sources (e.g., Cumlosen: River Basin Community Elbe [
49] or Lanžhot: Czech Hydrometeorological Institute [
50]). The actual discharge during the sampling period might have been very different from the annual average due to seasonal or event-induced variation; however, it was not assessed during the sampling.
4. Discussion
Sedimentation boxes have a simple and robust design, are easy to handle and install, and require no maintenance. Hence, sampling SPM and MP from surface waters with the device is cost-effective and may enable researchers to sample at a greater number of spots in comparison to other devices.
In this study, we aimed to evaluate the sampling performance of sedimentation boxes. We were able to identify different aspects which have an impact on the sampling performance and drive the uncertainty when determining MP concentrations in rivers. First, the retention rates of sedimentation boxes have a random or intrinsic error, which defines the reproducibility of the results. We quantified this error by comparing the results of three boxes operated in parallel. Furthermore, we expected the particle size and the flow velocity in the device to be decisive impact factors on retention rates. By conducting laboratory experiments under controlled conditions, we were able to quantify these factors and their variability. There are additional factors, such as the interaction with natural SPM, which likely increase the uncertainty of MP retention rates when sedimentation boxes are used to sample rivers. However, their influence could not be quantified with our setup. Moreover, the applied analysis method has a method-specific intrinsic error, which adds to the uncertainty of the derived MP concentrations. This error was not further investigated.
The coefficients of variation of the retained total mass are around 4%, and for the mass of different size fractions, they are well below 20%, regarding the samples provided by the three parallel sedimentation boxes. This did not change when switching from laboratory to field experiments. The MP concentrations obtained from rivers are subject to larger variation. Within a size fraction and between the parallel sedimentation boxes, the overall mean coefficient of variation is 36%. The uncertainty is larger for large-size fractions with a lower total mass and lower MP concentrations. Here, single MP particles may be decisive for the total concentration, which increases the variability between parallel boxes. However, the variation between MP concentrations cannot fully be attributed to the intrinsic error of sampling with the sedimentation box, as it cannot be distinguished from variability introduced within the process of sample preparation and MP detection with TED-GC/MS. To the best of our knowledge, there are no studies available which quantify the intrinsic error of MP analysis with TED-GC/MS.
Only a few studies report the intrinsic error of commonly applied MP sampling methods. When operating parallel manta nets, different variability between the replicate samples has been found, e.g., a mean coefficient of variation of 35% [
53] or 89% [
54] between MP concentrations. The replicate samples in Campanale et al. [
55] differed with a coefficient of variation of 35% for the total number of MP particles and 52% after fractionation. However, due to different MP detection methods, the comparison of the results of the sedimentation boxes is limited.
Hence, when maintaining constant conditions (i.e., flow rate, particle composition, and concentration), the intrinsic error of sampling with sedimentation boxes is similar to or lower than the one derived for manta nets.
Our laboratory experiments have shown that the retention rates of particles with the same characteristics (e.g., density, shape) increase with increasing particle size. This is in line with the principle of Stokes’ law, which states that the sinking velocity of particles increases with their size. The retention rates in our experiments closely follow a square root function of the particle diameter, as expected from Stokes’ law. However, complete retention of particles < 454 µm (for a flow of 5.2 L/s) and <321 µm (for a flow of 2.6 L/s), as calculated using the equations in
Figure 1, was not observed. For both flow rates, particles of that and larger sizes were only retained partially. This may be due to the particle shape, which was not spherical or areas with non-laminar flow in the box (see
Figure 2). Size-specific retention rates confirm the findings of many studies, which observe a shift in particle size towards larger particles for the material retained with passive samplers in comparison to the input or surrounding material [
12,
13,
15,
16,
17,
18,
19,
21,
23].
Our laboratory experiments have shown that retention rates increase with a decreasing flow rate, as expected. By halving the flow velocity, retention rates increase by an average factor of 1.3, i.e., almost
. Lower flow rates lead to a lower flow velocity and longer retention times inside the sedimentation box. Therefore, smaller particles with lower sinking velocities can be retained to a higher degree in comparison to higher flow rates. This effect has also been observed in laboratory experiments operating the TIMS sampler with different flow velocities [
12,
13]. When using the pump method, the flow velocity inside the box needs to be smaller than in the sampled river to allow the retention of formerly suspended particles. Although flow velocities in the Teltowkanal were very low (0.01 m/s on average), the flow velocity inside the box was lower (0.002 m/s); hence, sampling due to enhanced sedimentation was still feasible. A constant flow rate is crucial for quantitative results. Our velocity measurements have shown that an operation of sedimentation boxes in situ, i.e., with the natural flow of a river is prone to error due to highly variable flow rates inside the box. This effect is especially pronounced in rivers with unsteady flow regimes, such as the Teltowkanal, but certainly extends to other rivers.
In advance of the experiments, we expected the retention rates of PS particles to be higher, considering the retention rates determined by Harhash et al. [
27], who conducted similar laboratory experiments, yet with a higher flow rate. For PS particles with a diameter of 228 µm and a flow rate of 8.5 L/min, they determined a retention rate of approx. 67%, while our experiments showed retention rates of 71% (2.6 L/min) and 51% (5.2 L/min) for PS particles with a diameter of 229 µm. A possible explanation could be different particle shapes or different MP concentrations in the suspension.
Remarkably, PE particles have been retained very similarly to PS particles in the laboratory experiments, although their density is lower than water (0.94 g/cm
3). Due to their low density, it cannot be expected that they sediment in the boxes, as this is against Stokes’ law. With the sedimentation box being closed, PE particles are probably retained in the upper section of the sedimentation box. Positively buoyant particles can be trapped between the baffles of the sedimentation box or before the outlet, which is approx. 4 cm below the water level. As the density difference between the polymer and water is the same as for PS, the same mechanism applies in the opposite direction, leading to similar size-specific retention rates. Additionally, the low flow velocity close to the walls and especially in the corners of the box may favor the adhesion of the particles (see
Figure 3).
Our laboratory experiments have shown that sampling with sedimentation boxes is sensitive to sampling conditions and alters the sample composition compared to the input material. However, the retention of particles using an artificial MP suspension is highly reproducible, and the retention rates are in an approx. square root function in relation to the particle diameter. Hence, in the given setup, particle size distribution and concentration can be recalculated from the obtained sample.
In the field experiments, polymers with lower densities than water (PE, PP) were retained by the sedimentation boxes. Furthermore, MP particles with a size < 10 µm were retained at the River Elbe and the Teltowkanal. In addition to possible sampling through positive buoyancy, these observations emphasize the relevance of additional factors, which are not considered when regarding an isolated MP particle. Possible explanations are the advection to the device’s walls, biofouling, or aggregation with larger and high-density minerals. In addition, Mancini et al. [
56] found that the fast-settling (natural) particles scavenged MP particles and thereby increased their sedimentation velocity. This may favor the retention of small and low-density MP particles.
Although retention rates could only be averaged roughly for an entire size fraction, together with the controlled sampling conditions they allow the estimation of PS concentrations at River Elbe and Teltowkanal. Laermanns et al. [
57] found PS concentrations in the water phase of River Elbe at 0.02–0.56 µg/L, which is in the same order of magnitude as our results. In the project “MiWa”, the Teltowkanal was sampled with a continuous flow centrifuge and a PS concentration of 2.9 µg/L was determined [
58], which is in good accordance with our results too. However, due to different sampling and analysis methods, a direct comparison to other studies is limited. The MP concentration detected here and reported in the literature indicates that their fraction in SPM is typically below 1% (see also
Table 1).
The determination of quantitative MP data is impossible using the in situ method of the sedimentation boxes due to the missing information on the precise flow velocity and volume that was flowing through the box. In rivers with very low or alternating flow velocities, such as the Teltowkanal, the operation with a pump ensures a constant flow through the box, which could otherwise come to a standstill. It is still unclear if polymers that have not been detected are not retained by the sedimentation box or if these polymers are not present in the sampled water bodies. Moreover, the retained polymer mass needs to exceed the limit of detection to be quantifiable. Although our monitoring campaigns, which operate sedimentation boxes in situ are not suitable to provide quantitative results, qualitative information on MP abundance in the rivers can be derived. Assuming similar physical properties and thus retention rates, the relatively constant MP concentrations in all size fractions indicate a similar particle size distribution for MP and natural SPM.
The PE concentrations determined with TED-GC/MS must be treated with caution. Several studies have shown that during combustion, natural organic material such as leaves or certain fats may form thermal decomposition products which are similar to markers used for PE quantification with TED-GC/MS or pyrolysis-GC/MS [
47,
59,
60]. Lykkemark et al. [
61] did not find similar matrix effects leading to an overestimation of PS and PP with pyrolysis-GC/MS. A comparison between the displayed PE concentrations and TOC values in
Table 1 leads to a Pearson correlation coefficient of 0.78, indicating a good linear correlation. The correlation between PP, PS, and SBR concentrations with TOC is considerably weaker, with Pearson correlation coefficients ranging between 0.15 and 0.45. A good correlation between PE and TOC may be caused by a misinterpretation of organic compounds as PE. Thus, an overestimation of PE cannot be ruled out when applying these detection methods.
When sedimentation boxes are applied for the sampling of natural media such as rivers, we expect additional factors (other than flow velocity and particle size and density) to increase the uncertainty of retention rates. In the laboratory experiments, we determined the retention rates for artificial suspensions with pristine PS and PE particles of a limited size range (77–517 µm). Previously mentioned factors, such as interaction with natural SPM, weathering, biofouling, or different shapes, likely have an impact on the retention of MP in sedimentation boxes, but we did not include these factors. Moreover, the retention rates of particles < 77 µm remain unknown. These additional factors could not be assessed with the setup of our experiments. Hence, we cannot quantify the resulting increase in uncertainty for retention rates. Further investigations are needed to assess the impact of these additional factors on the retention of SPM and MP.
5. Conclusions
To date, there are no harmonized sampling methods for the monitoring of MP in water bodies because MP sampling is still challenging due to the low concentration, the high fluctuation, and the microscopic size. It has been shown that neither sampling techniques nor definitions distinguish between “natural” SPM and MP and that the properties of MP are in range with those of “natural” SPM. Hence, MP are already an integrated part of reported SPM.
Passive integrative sampling devices have been used to sample SPM in many different water bodies around the world. Yet, only a few studies have investigated their retention performance or identified factors that have an impact on retention. We investigated one type of passive sampler, sedimentation boxes, and operated them with a pump to achieve constant flow velocities, record the total sampled volume, and calculate retention rates.
Our laboratory experiments show that highly reproducible retention rates can be determined for the sedimentation boxes under controlled conditions. The retention rates increase with increasing particle size and a decreasing flow rate. Hence, both factors need to be considered in sampling strategies and when recalculating MP concentrations in the sampled water body.
We resume data from several experiments with sedimentation boxes in rivers across Europe. Despite the diverse river characteristics, we found MP in all samples and we conclude that sedimentation boxes (and most likely other passive samplers) are suitable to sample surface waters for MP qualitatively. Based on the determined retention rates, we could calculate PS concentrations in the sampled rivers, where we have operated the sedimentation boxes with a pump. The concentrations are in good accordance with the findings of previous studies.
Enclosed sedimentation boxes are suitable for retaining low-density polymers (<1 g/cm3) and small MP (<50 µm). Although particles may be underrepresented by sedimentation boxes due to the particle-size-specific retention rate, their actual concentration can be recalculated through their respective retention rates. Therefore, a standardized series of experiments with sedimentation boxes is needed to determine polymer- and size-specific retention rates under the effect of varying parameters such as flow velocity or SPM concentration.
In consequence, sampling MP with sedimentation boxes requires a low workload and provides highly reproducible results. In contrast to commonly used MP sampling methods such as manta nets, samples are time-integrated, and MP particles are not excluded due to their density or size. Although we were able to quantify some parameters that have an impact on the retention rates, the influence of others remains unknown (e.g., weathering, interaction with SPM, advection to the devices’ walls). Hence, we cannot fully determine the sensitivity of the method toward varying sampling conditions. Nevertheless, sedimentation boxes could be a valuable tool to obtain long-term information on MP in water bodies.