3.1. Seasonal Characteristics of Fog at CYOD
Figure 4 shows the monthly and time of day statistical distribution of fog at CYOD. As indicated in
Figure 4a, the most frequent fog type at CYOD was radiation fog and this was mostly observed in the summer season. The precipitation, advection and cloud-base-lowering fogs normally occurred in the fall and winter seasons (
Figure 4a).
Figure 4b shows the frequency of occurrence of fog events for a given month and the time of the day, indicating that most of the fog events occurred at night and early morning near sunrise, and usually dissipated after sunrise, as illustrated by the white and pink lines that show the local sunrise and sunset times, respectively. As shown in
Figure 4b,c, most of the fog occurred during the night, but the fog could extend to the late morning during cloudy conditions that blocked the incoming solar radiation. During the study period, there were 59 fog events reported, representing a total of 115.3 h. Based on this study, the frequencies of fog events that occurred were 70%, 15%, 7%, 5% and 3% for radiation, precipitation, advection, cloud-base-lowering and unknown fog types, respectively.
Figure 5 shows the distributions of fog events onset (
Figure 5a), dissipation time (
Figure 3b), and duration (
Figure 3c) for different fog types. Radiation fog could last from several minutes to more than 8 h (
Figure 5c). Based on this study, the mean fog duration for all fog types was 1.9 h, and the mean duration for radiation fog alone was 2 h, for precipitation fog 1.5 h, advection fog 2.7 h, cloud-base-lowering fog 2 h, while the unknown fog type only lasted for 0.6 h.
Figure 6 shows the time series of surface pressure (
Figure 6a), temperature (
Figure 6b) and relative humidity (
Figure 6c) measured at 2 m height. The fog types are also plotted using five different colors. The surface pressure pattern depicted in
Figure 6a shows that 25% of the fog events were associated with a high-pressure system (surface pressure over 1015 hPa [
37], and 75% were associated with a low-pressure system. The time series of surface temperature (
Figure 6b) suggested that fog events occurred at the coldest temperature of the day. A separate analysis confirmed that fog normally occurred near the minimum temperature of the day, except for some precipitation and cloud-base-lowering events. The corresponding relative humidity data shown in
Figure 6c shows that the RH was usually greater than 95% during fog events, and only in very cold temperature cases (T ≈ −30 °C) was the relative humidity with respect to water near 70%, which would translate to 138% humidity with respect to ice, so these cases may be associated with ice fog.
Figure 7 shows the frequency distribution of the observed visibility (
Figure 7a), surface level temperature (
Figure 7b) and relative humidity (
Figure 7c), and the wind speed and direction (
Figure 7d). The data in
Figure 7a demonstrate that 94% of the low visibility (<0.4 km) cases were caused by radiation fog and the remainder, 6%, were caused by precipitation, advection, and cloud-base-lowering fogs. The distribution of surface wind speed and direction measured at about 2 m height (
Figure 7d) shows that the wind speeds were greater than 2.5
during the cloud-base-lowering and advection fog events, and the corresponding wind directions were mainly coming from the southeast and east. Most of the radiation fogs occurred under calm wind conditions (<2
), with the predominant wind coming from northwest, west, southwest, and south directions.
3.2. Microphysical Characteristics of Fog at CYOD
To study the microphysical characteristics of the fog events observed at CYOD, the one-minute averaged observation data collected using the FM-120, PWD22 and Rotronic probes between January and September 2017 were used. The fog events were identified based on the METAR report when the visibility was less than 1 km and no precipitation was detected by the PWD22. According to the METAR reports, there were 18 fog events from January to September in 2017. The observed bulk microphysical parameters, such as the , droplet concentration and spectra, were used to investigate the microphysical characteristics of the different types of fog under various atmospheric conditions.
Figure 8 shows the one-minute averaged measured visibility using the PWD22 plotted against the one-minute averaged
(
Figure 8a), MVD (
Figure 8b), and concentration N
d (
Figure 8c), derived using the particle spectra measured using the FM-120 probe. The histogram of one-minute averaged concentration for different fog types (
Figure 8d), the one-minute averaged fog particle spectra for different fog types (
Figure 8e), the mean particle spectra of the period from January to September 2017 for a number of fog types (
Figure 8f) with varying temperature (
Figure 8g), and
(
Figure 8h) are also shown in the figure. There was no cloud-base-lowering fog event and only one unknown fog event was identified during this study period. The associated fog type is also shown in the figure (
Figure 8a–f). As indicated in
Figure 8a, radiation fog was dominant at CYOD, as discussed earlier, and was normally characterized by higher
(>0.1 gm
−3) and dense fog with visibility less than 400 m. The LWC during the advection and precipitation fog events was less than 0.06 gm
−3 while the maximum one-minute averaged LWC during radiation fogs could exceed 1.2 gm
−3. As indicated in
Figure 8b, the observed MVD in advection and precipitation fog events were between 17 μm and 35 μm, while the MVD observed in the radiation fog events varied more widely between 7 μm and 45 μm.
Figure 8c,d show that
in radiation fog varied from near 5 cm
−3 to more than 230 cm
−3, the lower end being the most frequent, but, for precipitation and advection fog events, the values of
were less than 40 cm
−3. It was found that the visibility decreased when
and
increased (as seen in
Figure 8a,c). The one-minute averaged particle spectra for different fog types (
Figure 8e) indicated that the particle size and N
d of radiation fog were much greater compared to the other fog types. Each of the observed one-minute averaged drop spectra showed bimodal size distributions, with one mode around 4 μm and the other near the higher end of the spectrum, that varied from 17 μm to 25 μm depending on the meteorological condition. The maximum total concentrations were around 100 cm
−3 and 20 cm
−3 for the two modes, respectively. The observed mean particle spectra during the fog events (
Figure 8f–h) showed clearly bimodal size distributions, as indicated in
Figure 8e.
Figure 8f illustrates that the second mode of the mean spectra of radiation fog shifted to large particle sizes compared to the other fog types.
Figure 8g shows that there was no significant difference in droplet particle spectra between warm (T > 0 °C) and freezing (−10 < T ≤ 0 °C) conditions (possibly super-cooled), but, at colder temperatures (T ≤ −10 °C), where mixed phase conditions were expected, smaller particle concentrations were observed and the second mode of the bimodal distribution peaked at smaller sizes around 17 μm; however, the accuracy of the FM-120 probe when sizing non-spherical ice particles may be questionable since the probe assumes perfectly spherical water drops [
38]. As shown in
Figure 8h, the first mode of the distribution of large
(>0.1 gm
−3) was almost overlapped by those of small LWC (0.01 < LWC ≤ 0.1 gm
−3), but the second mode of large LWC shifted to larger size with increasing
, indicating that large LWC was mainly comprised of the large particle size (D > 7 μm).
3.3. Parametrization of Extinction/Visibility
To investigate the relationships between visibility/extinction, the relevant bulk microphysics parameters, such as LWC, drop spectra and concertation, were used. These parameters were measured using the DMT fog monitor (FM-120) that outputs data at a one-second temporal resolution. The Vaisala PWD22 weather probe measured the visibility, precipitation intensity and type. Additionally, the Rotronic MP102 T/RH probes measured the temperature and humidity data every minute. For this study, a one-minute averaged integrated dataset collected between January and September 2017 was used. The visibility reduction due to precipitation events was excluded using the present weather probe reports. The Vaisala PWD22 sensor estimated Vis using Equation (1) assuming a visual contrast of 0.05 [
11,
16].
A lower visual threshold of 2% has been used (e.g., [
15]). In this study, the visual threshold of 5% was adopted since this is consistent with the PWD22 data and WMO recommendations. The more sophisticated current NWP models predict the bulk microphysical quantities, such as extinction and LWC, based on an assumed single-modal fog particle size distribution [
28,
39,
40]. When such information is available, most models mainly rely on parameterization of extinction or visibility, mainly using surface-based observation data. Extinction or visibility in fog has been parameterized in terms of RH and dewpoint depression [
11,
21,
41], LWC and N
d [
2,
14,
15,
22]. Based on definitions of the parameters, [
14] determined the visibility as
where
is the density of water and
is the number concentration. The exponent of LWC in Equation (7) matches that proposed by [
21] and the relationship shows that visibility is more sensitive to changes in LWC than
. The authors of [
2,
21] and [
15] parameterized visibility as
where c and d are constants; their values c (d) are given as 1.002 (0.87706) and 0.6473 (0.49034) in [
21] and [
2], respectively. On the other hand, in [
15], the coefficients for all combined fog events were 0.797 and 0.535, respectively, for c and d. It is worth mentioning that when the measured visibility using the present weather sensors are used instead of the calculated visibility based on FM fog detector measurements, the relationship given in Equation (8) gives very different coefficients (see ref. [
15] and references therein). Without performing regression, using Equations (3)–(5), and assuming
extinction can be directly related to LWC and
in a form as (see ref. [
14]):
where the
and the density of water
are given in gm
−3 and the area weighed diameter
is given in m. The extinction is given in m
−1. In order to investigate the dependence of
on
,
,
, and
, scatterplots for each case are given in
Figure 9 and the associated best-fit regression lines and the correlation coefficients are also given. Based on these results, although the correlation coefficients were close, the error variance (EV), or the square of root mean square error (RMSE), were higher when only
was used (
Figure 9b), by a factor of three compared to when LWC was used (
Figure 9a), and by more than a factor of 10 when both Nd and LWC were used (
Figure 9c).
Figure 9d shows
derived in this study; as indicated in the plot, there is excellent agreement with the FM-120 observation with R close to 1 and a mean square error (MES) of near zero.
Figure 10 compares the relationships based on (I-2020) [
14], (G-2009) [
2], and (Z-2014) [
15] and this investigation. All the relationships showed good agreement; however, based on these results, I-2020 and
(Equation (10)) appeared to be closer, although I-2020 estimated slightly higher extinction compared to the other parameterizations. The better agreement of the extinction parameterization that included
, other than
, was because
is inversely proportional to the cross-sectional area (Equation (7)), which is directly related to extinction (Equation (5)).
3.4. Comparisons of FM-120 and PWD22
The Vaisala PWD22 probe measures extinction and then converts it to visibility using Equation (1) as stated earlier and hence it can be directly compared with FM-120 (
) data.
Figure 11 shows extinction
plotted against the extinction measured with the PWD22 (
. As shown in the plot, the FM-120 estimated higher extinction as compared to the PWD22 for (
0.002 m
-1) or visibility less than approximately 1.5 km which is very close to moderate to heavy fog (FG). The mist (BR) and drizzle (DZ) case reported by the PWD22 are also shown and according these results the PWD22 detects more of the mist cases as compared to the FM-120. As mentioned earlier mist was identified mainly based on visibility and hence it cannot be directly related to size based on the PWD22 measurement alone.
Figure 12 shows the ratio
plotted against
(
Figure 12a),
(
Figure 12b),
(
Figure 12c) and
(
Figure 12d) for both fog and mist cases. As indicated in the plots, there was some size dependence for the fog case showing higher ratios related to larger particle sizes, but for the mist case the FM-120 probe measured lower extinction for all sizes, except near 6 µm, where it showed a stronger peak, suggesting that the FM-120, under some misty conditions, detects more small particles (5 µm <
< 7 µm) (
Figure 12a). However, there was a clear
,
, and
dependence indicating that the detection efficiency of
increased with increasing
,
and
(
Figure 12b–d). In addition, there was a clear distinction between the fog and mist cases, particularly for
> 0.002 gm
−3,
> 1 cm
−3, and
> 200 gm
−4. As demonstrated in
Figure 12a, all the ratios > 1 indicated in
Figure 12b,c, and d during the mist events were associated with small particles in the size range mentioned above, and these were also characterized by higher LWC and
thus, the discrepancy between these probes during the mist events cannot be explained based on the particle size alone. For the same mass, small and numerous particles are known to have higher effective surface area and, hence, higher extinction. According to these results
,
and
appear to be more important than
alone for both fog and mist cases. Based on these results, there was no clear evidence of particle losses for (D < 10 μm) and (D~50 μm), as suggested in [
31]. In fact, as mentioned previously, the collection efficiency of the FM-120 probe, particularly in fog, increased with increasing particle size compared to the PWD22 probe. According to these results, the ratio showed a clearer relationship with
, as would be expected from Equation (10); the best-fit lines for both mist and cases are shown in
Figure 12d. This information could be used to make corrections to the visibility parameterization, rather than developing two separate parameterizations, one with visibility estimated using the present weather sensors, and the other estimated based on particle spectra measured, using probes such as the FM-120 probes.
However, further investigations are necessary to understand the causes of these discrepancies between these two probes. One way to test the accuracy of these probes would be to compare them against human vision, but there are a number of difficulties associated with doing this, including that there are some differences between nighttime and daytime human vision, as discussed earlier. One example is given in
Figure 13 for 22 July 2017 that shows visibility estimated using both FM-120 and PWD22, and human observation based on METAR (
Figure 13a) during nighttime, as indicated by the observed luminance measured using the Vaisala FS11P [
42] (
Figure 13c). The relative humidity was close to saturation during the fog event and the temperature was radiatively cooled overnight from 20 °C to close to 12 °C. As indicated in
Figure 13a, visibility based on human observation is much greater compared to visibility calculated based on the two probes. As mentioned earlier, this is because the human eye sees further under an artificial light at night and, hence, needs to be corrected by using Equation (3). After corrections were applied to the values of measured visibilities, they were much closer to human observations, particularly that based on the PWD22 (
Figure 13b).
The FM-120 probe estimated higher extinction when both
and
were relatively high (
Figure 13c). Similar plots for a wintertime case on 20 January 2017 are shown in
Figure 14. In this case, the fog event also occurred during the night (
Figure 14c). The PWD22 also agreed better with human observation (
Figure 14b), but, because of cold temperatures (−12 °C < T < −5 °C), it was a freezing fog event, potentially due to radiative cooling during the night. The RH sensors gave conflicting results; however, based on the one reported by the human observer (METAR), the RH values exceeded 95%. Similar to the 22 July case, the PWD22 data agreed with the human observation data, and the FM-120 results appeared to fluctuate more (a and b).
As mentioned earlier, by combining the LWC and
measured using the FM-120 and visibility obtained from the PWD22, a regression equation similar to Equation (8) can be derived.
Figure 15 shows the scatterplots of visibility against LWC for different temperatures (
Figure 15a), for warm fogs with binned data in red circles (
Figure 15b), the same as
Figure 15b but based on calculated visibility (
Figure 15c), and, for warm fogs, the observed visibility is plotted against the inverse of the product of LWC and
(
Figure 15d). The parameterization equation with correlation coefficient R, standard deviation σ and relative error ε are shown in each panel. The best-fit lines for each panel are also shown in the figure. The binned data in red circles represent the mean value of LWC; visibility in each bin is matched with the fit line in
Figure 15b. The regression relationships of observed visibility versus LWC for all temperatures and for warm fog events (T ≥ 0 °C) are given in Equations (11) and (12), respectively. For warm fogs, the regression relationship of calculated visibility versus LWC is given in Equation (13), and the relationship between observed visibility and both LWC and N
d is given in Equation (14).
The parameter b (see Equation (4)), derived using the observed visibility and LWC in this study (Equations (11) and (12)), was close to 0.1 that would represent dense fog, but the value for a = 0.3 derived in this study deviates from the two-thirds assumption based on the gamma size distribution [
21,
22] or the theoretical derivation of [
14]. In comparison to the results reported in [
15], based on the observed visibility, the parameter b found in this study lies between their values for light and dense fog cases, but the parameter a derived in this study was higher than their value of 0.215 for dense fogs. The parameters (0.031, 0.893) shown in Equation (13), based on calculated visibility, were comparable to the values (0.027, 0.88) suggested by [
41] and (0.0219, 0.960) by [
24], and (0.017, 0.871) by [
15] using calculated visibility. The parameters of parameterization given in Equation (14) (0.270 and 0.195) were closer to those for dense fogs (0.212 and 0.147) proposed by [
15].
The relative error of observed visibility parameterization as a function of
for all temperature and fog events (Equation (11)) was 36% and for warm fog events (Equation (12)) was 34%, with correlation coefficients of 0.51 and 0.53 and standard deviations of 0.47 km and 0.46 km, respectively (see
Figure 15a,b). The relative error of the parameterization, as a function of both
and
(Equation (14)), as shown in
Figure 15d, was 32%, with a correlation coefficient of 0.55 and standard deviation of 0.45 km, representing small improvements over the use of just
(
Figure 15b). In
Figure 15c, the best-fit line, as shown in Equation (13), for the calculated visibility that included the fog particle spectra, gave a better correlation 0.98 and relatively lower relative error 20%, as would be expected since both
and extinction were derived using the FM-120 probe data. However, the calculated visibilities based on the FM-120 data were smaller than the measured visibilities using the PWD22 present weather sensor when the
> 0.1 gm
−3, and larger when LWC < 0.1 gm
−3, but the reasons are not well understood. According to this result, there was no indication that the FM-120 suffered from losing some particles as suggested by [
31]. The authors of [
15] found that the observed visibility was similar to the calculated visibility for dense fogs and smaller than the calculated visibility using the DMT FM-100 probe for light fogs. They attributed this discrepancy to the possible effect of aerosol concentrations which were not measured by the DMT FM-100 probe but were included in the extinction seen by the PWD22 probe. This is probably true for heavily polluted locations, but the effect seen in this study was the opposite when dense fogs occurred, and no clear explanation can be offered at this stage. Using a lower visual threshold of 2%, as shown in Equation (5), for visibility calculation (shown in
Figure 15c as blue dots) as performed by [
15], can enhance the visibility by a certain amount, but not enough to explain the phenomena.
No effect was found when the first modes (2–15 μm), which may have contained some aerosols, were removed from the observed spectra, with no significant effect observed on the calculated visibility. The FM-120 probe was installed at 2 m height, which was lower than where the PWD22 was installed (z~3 m). If there was a decrease in LWC within this layer, it is possible that the PWD22 may measure larger visibility, but this cannot be validated based on the available measurements. Another possible explanation is that when the
was significant during heavy fog, the forward scattered radiation was obscured by a high concentration of drop particles that may have led to lower extinction data being recorded by the PWD22 probe. Furthermore, the PWD22 uses infrared light instead of visible light and different suspended particulates and water droplets in the air have different scattering properties; hence, the use of a fixed wavelength of light and a fixed scattering angle may make a particular forward-scatter sensor less suitable in certain weather conditions [
43].