Next Article in Journal
Overdurability and Technical Wear of Materials Used in the Construction of Old Buildings
Next Article in Special Issue
Structural and Mechanical Properties of a-BCN Films Prepared by an Arc-Sputtering Hybrid Process
Previous Article in Journal
Effect of Flue Gas Desulfurization Gypsum on the Properties of Calcium Sulfoaluminate Cement Blended with Ground Granulated Blast Furnace Slag
Previous Article in Special Issue
Properties and Classification of Diamond-Like Carbon Films
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Surface Reformation of Medical Devices with DLC Coating

1
Department of Electrical and Electronic Engineering, Tokyo Denki University, Tokyo 120-8551, Japan
2
Nanotec Co., Ltd., Chiba 277-0872, Japan
3
Department of Biomedical Engineering, King Saud University, Riyadh 11362, Saudi Arabia
4
Namiki-Mi Co., Ltd., Tokyo 132-0035, Japan
*
Author to whom correspondence should be addressed.
Materials 2021, 14(2), 376; https://doi.org/10.3390/ma14020376
Submission received: 16 December 2020 / Revised: 30 December 2020 / Accepted: 7 January 2021 / Published: 14 January 2021
(This article belongs to the Special Issue DLC (Diamond-Like Carbon) Film Formation and Application)

Abstract

:
We evaluated the adhesion, friction characteristics, durability against bodily acids, sterilization, cleaning, and anti-reflection performance of diamond-like carbon (DLC) coatings formed as a surface treatment of intracorporeal medical devices. The major coefficients of friction during intubation in a living body in all environments were lower with DLC coatings than with black chrome plating. DLC demonstrated an adhesion of approximately 24 N, which is eight times stronger than that of black chrome plating. DLC-coated samples also showed significant stability without being damaged during acid immersion and high-pressure steam sterilization, as suggested by the results of durability tests. In addition, the coatings remained unpeeled in a usage environment, and there was no change in the anti-reflection performance of the DLC coatings. In summary, DLC coatings are useful for improving intracorporeal device surfaces and extending the lives of medical devices.

Graphical Abstract

1. Introduction

In recent years, owing to the continuous development of medical technology, the average length of life has increased significantly in developed countries [1]. As a result, the number of deaths caused by illnesses related to aging is continuously increasing, and treatments for early detection of illnesses and less invasive medical procedures are attracting attention [2]. The early detection of cancer is of particular importance because it significantly reduces the mortality rate [3]. However, medical devices have material-related problems. Speculum and cervical dilators that are widely used for the examination and treatment of cervical and vaginal cancers are made from stainless steel (SUS) and brass [4]. When performing examinations and surgery, such as a speculum diagnosis, the reflection of surgical light or a laser scalpel on metal surfaces is a cause of interference in treatment and examination [5]. Therefore, current intracorporeal medical devices, such as speculums, need to be coated with a matte black coating, such as black chrome plating, that suppresses light reflection [4]. However, metal materials, such as black chrome plating, have a high coefficient of friction on the contact surface with the living body, causing discomfort and pain stemming from the friction experienced by the patient [6]. In addition, depending on the part of the biologic organ into which a device is inserted, the metal material may elute because of acid in the body, thereby causing problems like the deterioration of the black chrome plating and skin inflammation in patients with metal allergies [7]. It is therefore highly important to treat medical devices after use to prevent infectious diseases, and, once used, medical devices must be sterilized in actual clinical settings [8]. The deterioration of coating materials during the medical processes is also an issue [9]. For these reasons, there is a demand for the development of surface treatment technology for medical devices as an alternative to black chrome plating.
Diamond-like carbon (DLC) is a non-crystalline or amorphous carbon-based material (lacking a crystalline structure) consisting of sp2-bonded carbon atoms of a graphite structure and sp3-bonded carbon atoms of a diamond structure [10]. Some coatings contain hydrogen at 0–40 atm%, and their properties depend on the ratio between sp2 bonding and sp3 bonding with the hydrogen content [11]. Therefore, DLC coatings are produced under conditions suitable for intended use and the properties on the material surfaces. By reducing the hydrogen content, a coating that is nearly as hard as diamond will be formed, whereas in the presence of hydrogen, a coating with gas barrier properties and chemical resistance will be formed [12]. Owing to its high hardness, gas-barrier properties, low-friction properties, and other features depending on the coating formation methods, DLC coating continues to be developed in industrial applications, such as surface coating of cutting tools and optical parts [13]. In addition, because DLC has properties like chemical stability and biocompatibility, it is attracting attention as a surface modification for medical devices with DLC coated on biomaterials [14,15,16,17,18].
This study is intended to utilize DLC surface coating technology to impart functions such as sliding properties, chemical resistance, sterilization resistance, and surgical light reflection prevention in various intracorporeal medical devices. Therefore, in this experiment, film adhesion was evaluated using a scratch test as a mechanical property of DLC coatings. To evaluate the sliding property within a living body under conditions simulating clinical settings, the coefficient of friction was measured in a simulated body environment using lubricating liquids. To evaluate durability, an immersion test simulating the acid in the body and a high-pressure steam sterilization test were conducted repeatedly. Furthermore, regarding the reflection characteristics for surgical light, a sponge-rubbing test simulating the usage environment was performed.

2. Materials and Methods

2.1. Sample Preparation and Surface Treatment

In this experiment, we used SUS304 samples (Namiki-Mi Co., Ltd., Tokyo, Japan), which have a low cost and high workability, and are widely used as actual medical devices [19]. In the scratch test, durability test, and evaluation of light reflection characteristics, SUS304 (20 mm × 20 mm × 1.5 mm) samples were used as analysis samples. In the simulated body friction coefficient test, SUS304 (40 mm × 100 mm × 1.5 mm) samples were used to match the size with the simulated living body tester.
Black chrome plating was applied using electroplating [20]. Film formation was performed with SUS used as a cathode and the plating material (Cr) used as an anode. A schematic diagram of the electroplating method is shown in Figure 1. The sample on which black chrome was formed on SUS304 is denoted as Black Cr/SUS. In the Black Cr/SUS samples, film formation with a thickness of 1.8 µm was confirmed by cross-sectional scanning electron microscopy (SEM) (JEOL, Tokyo, Japan).
DLC was formed via the ionized evaporation method using C6H6 as a source gas [4,21]. DLC was deposited under the conditions of base pressure 2 × 10−3 (Pa), gas pressure 0.2 (Pa), filament current 30 (A), substrate voltage 2 (kV), temperature 200–250 (°C), and deposition time 220 (min). The samples for film formation were subjected to ultrasonic cleaning in an acetone bath as a pretreatment [10]. The sample in which DLC was formed on SUS304 is denoted as DLC/SUS. In the DLC/SUS sample, film formation with a thickness of 1.5 µm was confirmed by cross-sectional SEM. Friction strongly depends on the surface condition (roughness and electrical potential) and the usage environment (dry or wet) of the material [4,22,23,24,25]. DLC coatings are water repellent and, as such, cannot be expected to effectively reduce the coefficient of friction when using lubricating liquids [26]. Therefore, to measure the coefficient of friction in a simulated body environment, the DLC/SUS samples were subjected to oxygen plasma surface treatment using the high-frequency plasma method to improve the hydrophilicity of the outermost surface [27]. The oxygen plasma surface treatment reformed the DLC/SUS surface with plasmaized oxygen. The processing conditions were pressure 10 (Pa), power supply 200 (W), and treatment time 2 (min). The DLC/SUS sample subjected to oxygen plasma treatment was denoted as O-DLC/SUS. A schematic of the plasma reactor and the surface treatment conditions have also been shown in previous research [26].

2.2. Evaluation of Film Adhesion by Scratch Testing

To ensure reliability, thin-film coatings must not peel off in a usage environment when used as a surface treatment for medical devices. Films with weak adhesion will be peeled off, and these films will not be able to exhibit their functions in the peeled area. In this study, the adhesion was evaluated for the black chrome plating and the DLC-coated SUS sample using a scratch tester (Revetest Scratch Tester) (CSM Instruments, Switzerland). The conditions were set with reference to the ISO 20502 standard in the experiment [28]. In the scratch test, Black Cr/SUS and DLC/SUS samples were measured three times each, and the measured values were averaged to determine the peeling load. A schematic of the scratch test is shown in Figure 2, and the conditions for the scratch test are shown in Table 1.

2.3. Measurement of the Coefficient of Friction in a Simulated Body Environment

To clarify the effect of friction between the Black Cr/SUS and DLC/SUS samples and biomaterials, a simulated body environment was prototyped using urethane to measure the friction coefficient [26]. To measure the coefficient of friction in a simulated body environment, the DLC/SUS samples were subjected to oxygen plasma treatment using the high-frequency plasma method and then surface-treated. To calculate the coefficient of static friction, both sides of the sample were deposited in DLC and were treated with oxygen plasma. In this test, a comparative study of two samples, Black Cr/SUS and O-DLC/SUS, was conducted. In addition to the measurement under dry conditions, measurements were also performed under wet conditions containing blood, physiologic saline, and hyaluronic acid solution to simulate an actual clinical setting. Blood (Tokyo Shibaura Zoki Ltd., Cow blood, Tokyo, Japan), physiologic saline (SIGMA, sodium chloride solution 0.9%), and hyaluronic acid solution (FUJIFILM, hyaluronic acid sodium 0.1%) were used as lubricating liquids. These lubricating liquids were selected because they have high biocompatibility and are commonly used in clinical tests [29].
The specific test procedures for measuring the coefficient of friction in a simulated body environment were the same as those used in previous research [26].

2.4. Immersion and Sterilization Tests with Hydrochloric Acid

Immersion and sterilization tests with hydrochloric acid were conducted to study the acid resistance and durability of black chrome coatings and DLC coatings in a simulated clinical setting.
The Black Cr/SUS and DLC/SUS samples were used in these tests. A schematic of a high-pressure steam sterilizer and the sterilization conditions have been shown in previous research [26]. A schematic of the immersion test with hydrochloric acid and the immersion conditions have also been shown in previous research [26].
The chemical composition of DLC/SUS was analyzed by using X-ray photoelectron spectroscopy (XPS) (JEOL, Tokyo, Japan). The surface conditions of the samples were observed with a Schottky field-emission scanning electron microscope (FE-SEM: JEOL, JSM-7100F) (JEOL, Tokyo, Japan). In addition, surface elements were measured with an energy-dispersive X-ray spectrometer (EDS; JEOL, JED-2300) (JEOL, Tokyo, Japan). The results of these tests using Black Cr/SUS and DLC/SUS samples were compared to investigate the utility of the DLC coatings.

2.5. Light Reflection Evaluation before and after the Cleaning Test

A black coating is applied to medical devices used in tests and treatments to prevent the reflection of surgical light [6]. However, depending on the cleaning and usage environment, the coating material may peel off and subsequently become unable to prevent the reflection of surgical light. Therefore, to study the surgical light reflection characteristics before and after the use of medical devices, a cleaning test with a sponge (ICHIGUCHI Ltd., outer diameter: 50 mm) (ICHICUCHI ltd., Yokohama, Japan) was conducted on each sample before and after use. The surgical light reflection characteristics were evaluated via specular gloss measurement using a gloss meter (HORIBA, IG-331) (HORIBA, Kyoto, Japan). The SUS, Black Cr/SUS, and DLC/SUS samples were used in the test. Using the glossiness of SUS as a reference, the glossiness of Black Cr/SUS and DLC/SUS was measured before and after use. The specific test procedures are shown below. A schematic of the test is shown in Figure 3. The test conditions are listed in Table 2.
(1)
Each substrate was fixed to a drill press (MECANIX Co., Ltd. Shop-Ace M18A) (MECANIX Co., Ltd., Japan).
(2)
The sponge attached to the drill press was slid at a constant load.
(3)
The reflection characteristics of each sample after sliding were measured using a gloss meter.
After sliding, the surface condition of the sample on each substrate was visually observed. The results of these tests using the Black Cr/SUS and DLC/SUS samples were compared to investigate the utility of the DLC coatings.

3. Results and Discussion

3.1. Film Adhesion Evaluation by Scratch Testing

As previously mentioned, to ensure reliability, thin-film coatings must not peel off in a usage environment when used as a surface treatment for medical devices. A scratch test was conducted on the Black Cr/SUS and DLC/SUS samples to measure the film adhesion. Table 3 shows the delamination loads measured in the scratch test. Each sample was measured three times to calculate the average. The results in Table 3 show that the delamination loads of Black Cr/SUS and DLC/SUS are 3.04 and 24.01 (N), respectively. The DLC coatings demonstrated a load withstanding capacity that was approximately eight times larger than that of black chrome plating. Delamination loads of DLC coating was roughly the same value as other research paper [30]. In addition, delamination loads of DLC coating was higher than other materials [31]. These results suggest that when DLC coatings are formed on medical devices, their mechanical adhesion rate and usage life are higher than those of the substrate coated with black chrome plating.

3.2. Simulated Body Coefficient of Friction Measurement in Each Lubricating Liquid Environment

To clarify the effect of friction between the Black Cr/SUS and DLC/SUS samples and biomaterials, a simulated body environment was prototyped using urethane to measure the friction coefficient. Simulated body friction coefficient testing was conducted using blood, physiologic saline, and a hyaluronic acid solution, as well as under dry conditions. The measurement results are shown in Figure 4. The friction coefficient in this study was measured by sandwiching the substrate with urethane rubber. Therefore, the friction coefficient in this study is higher than the normal friction coefficient. The coefficient of friction under dry conditions was 1.391 for Black Cr/SUS and 0.763 for O-DLC/SUS. In addition to its lower friction in dry conditions, O-DLC/SUS showed lower frictional properties than Black Cr/SUS under any lubricant conditions. In addition, the coefficient of friction of O-DLC/SUS under physiological saline and hyaluronic acid conditions was lower than that under dry conditions. These results suggest that O-DLC coatings have a lower friction within the living body than black chrome plating, and it is highly possible that physiological saline and hyaluronic acid can suppress the invasiveness of certain medical procedures caused by friction. There is no specific value for the optimum friction coefficient in the body. For a medical device to be inserted into the body, the lower the friction coefficient, the better.

3.3. Immersion and High-Pressure Steam Sterilization Tests with Hydrochloric Acid

To study the acid resistance and durability of Black Cr/SUS and DLC/SUS in a simulated clinical setting, immersion and sterilization tests with hydrochloric acid were performed to evaluate the durability before and after the immersion and sterilization tests. Chemical composition was analyzed by using X-ray photoelectron spectroscopy (XPS). The surface state and film composition were measured using field-emission scanning electron microscopy (FE-SEM) and energy-dispersive X-ray spectroscopy (EDS). The C1s spectra of DLC/SUS before and after the immersion and sterilization tests are shown in Figure 5. These spectrum were deconvoluted to find the ratio of sp3/(sp2 + sp3). Binding energies corresponding to both the diamond and graphite were found at 284.3 eV (sp2) and 285.0 eV (sp3) before and after the immersion and sterilization tests. The sp3/(sp3 + sp2) ratio governing the quality of the DLC films was calculated from the C1s peak (a greater quantity of sp3 bonds makes the films more similar to diamond-like film, while more sp2 bonds make them more graphite-like). The sp3/(sp2 + sp3) before and after the immersion and sterilization tests showed 0.23 to 0.26, and there was almost no change before and after testing. The results of XPS confirmed that the surface of DLC/SUS was stable before and after immersion and sterilization.
SEM images before and after the tests are shown in Figure 6. The figure demonstrates that the surface of Black Cr/SUS after the immersion and sterilization tests was changed and damaged by acid and sterilization. On the other hand, no surface deterioration was confirmed with the DLC/SUS sample before and after the immersion and sterilization tests. This is presumably ascribable to the chemical stability, which is a characteristic of DLC coatings.
To analyze significant morphological changes in the Black Cr/SUS surface before and after the immersion and sterilization tests, a quantitative analysis of elements was performed using EDS. In addition to Fe, Cr, and Ni, which are the main elements of SUS, as well as C, which is the element of DLC, studies were mainly conducted on O to analyze the possibility of oxidation. The analysis results are shown in Table 4. For the Black Cr/SUS sample, the proportion of Cr decreased from 63.85 to 50.25, whereas the proportion of O increased after the immersion and sterilization tests. This is because the passive film was dissolved by the hydrochloric acid used in the immersion test. Moreover, the moisture from the sterilization process invaded the boundary between the material and the film, causing the Black Cr/SUS sample to corrode [32,33]. Next, we considered the DLC/SUS sample. Neither an exposure of metal material nor a change was confirmed with the DLC/SUS sample before and after the immersion and sterilization tests. In other words, the surface of DLC/SUS was stable before and after the immersion and sterilization tests, and its durability as a coating material was confirmed. The results of SEM and EDS suggest that DLC coatings are more durable than conventional coatings as a surface treatment technology for medical devices.

3.4. Light Reflection Characteristic Evaluation before and after the Cleaning Test

To study the surgical light reflection characteristics before and after the use of medical devices, a cleaning test with a cleaning sponge was conducted on each substrate before and after use. The surgical light reflection characteristics were evaluated with a gloss meter. The measurement results of the glossiness are shown in Table 5, and the surface conditions before and after sliding are shown in Figure 7. The glossiness of SUS was 75.4, which was used as the reference. The glossiness of the Black Cr/SUS sample was 6.8 before sliding and 63.8 after sliding. It is plausible that the function as a light reflection inhibitor deteriorated because we confirmed that the black chrome plating was peeled off and that the SUS was exposed by cleaning. The coating material with the lowest glossiness indicated approximately 0 to 30 [34]. On the other hand, the glossiness of the DLC/SUS sample before and after sliding was 15.8 and 15.4, respectively. No change was observed in the reflection characteristics before and after sliding, thereby suggesting that the DLC coatings are useful as a light reflection inhibitor in a usage environment.

4. Summary and Conclusions

In this study, we evaluated the adhesion, friction characteristics, durability against bodily acids, sterilization and cleaning, and light reflection resistance of diamond-like carbon (DLC) formed as a surface treatment for intracorporeal devices. The coefficient of friction was lower with DLC coatings than with black chrome plating in all environments. Furthermore, DLC showed an adhesion of approximately 24 N, which is eight times stronger than that of black chrome plating. The DLC samples were also significantly stable without being damaged during immersion in acid and high-pressure steam sterilization, which were conducted as a durability test. In addition, the coatings remained unpeeled in a usage environment, and there was no change in the anti-reflection performance of the DLC coatings. In summary, the use of DLC coatings is a useful technique for improving the surfaces of intracorporeal devices and extending the life of medical devices.

Author Contributions

Conceptualization, M.K., K.N. and K.H.; methodology, M.K. and K.H.; software, M.H. and H.N.; validation, M.K., M.H. and K.H.; formal analysis, A.A. and K.H.; investigation, M.K.; resources, M.H., H.N. and K.N.; data curation, A.A. and K.H.; writing—original draft preparation, M.K.; writing—review and editing, M.K., A.A. and K.H.; visualization, M.K.; supervision, K.H.; project administration, M.K., M.H., K.N. and K.H. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by JAPAN Keirin Autorace, grant number 2018-113.

Data Availability Statement

The date presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Curtis, C.C.; Lugauer, S.; Mark, N.C. Demographics and aggregate household saving in Japan, China, and India. J. Macroecon. 2017, 51, 175–191. [Google Scholar] [CrossRef]
  2. Merali, N.; Singh, S. Abdominal access techniques (including laparoscopic access). Surgeons (Oxford) 2018, 36, 220–227. [Google Scholar] [CrossRef]
  3. Li, C.; Liu, Y.; Xue, D.; Chan, W.H. Effects of nurse-led interventions on early detection of cancer: A systematic re-view and meta-analysis. Int. J. Nur. Stud. 2020, 110, 103684. [Google Scholar] [CrossRef] [PubMed]
  4. Perez, R. Medical Speculum and Procedure. U.S. Patent 4597382A, 1 July 1986. [Google Scholar]
  5. Pan, C.; Chen, Y.; Yang, T.; Lin, P.-H.; Huang, J.C.; Lin, P.-H. Study of reflection-typed LED surgical shadowless lamp with thin film Ag-based metallic glass. Optic 2016, 127, 2193–2196. [Google Scholar] [CrossRef]
  6. Usui, T.; Iwata, T.; Miyake, S.; Otsuka, T.; Koizumi, S.; Shirakawa, N.; Kawata, T. Mechanical and frictional proper-ties of aesthetic orthodontic wires obtained by hard chrome carbide plating. J. Dent. Sci. 2018, 13, 151–159. [Google Scholar] [CrossRef]
  7. Karthik, C.; Ramkumar, V.S.; Pugazhendhi, A.; Gopalakrishnan, K.; Arulselvi, P.I. Biosorption and biotrans-formation of Cr (VI) by novel Cellulosimicrobium funkei strain AR6. J. Taiwan Inst. Chem. Eng. 2017, 70, 282–290. [Google Scholar] [CrossRef]
  8. Moradi, L.; Jobania, B.M.; Jafarnezhad-Ansariha, F.; Ghorbani, F.; Esmaeil-Pour, R.; Zolbina, M.M.; Kajbafzadeh, A.-M. Evaluation of different sterilization methods for decellularized kidney tissue. Tissue Cell 2020, 66, 101396. [Google Scholar] [CrossRef]
  9. Mousavi, S.M.; Shamohammadi, M.; Moradi, M.; Hormozi, E.; Rakhshan, V. Effects of cold chemical (glutaraldehyde) versus autoclaving sterilization on the rate of coating loss of aesthetic archwires: A double-blind randomized clinical trial Comparaison des effets de la stérilisation à froid (glutaraldéhyde) et à l’autoclave sur le taux de perte de revêtement des arcs esthétiques: Essai clinique randomisé, contrôlé, en double-aveugle. Int. Orthod. 2020, 18, 380–388. [Google Scholar]
  10. Fontaine, J.; Donnet, C.; Erdemir, A. Fundamentals and Application of DLC Films, Tribology of Diamond-Like Carbon Films; Springer: Berlin, Germany, 2008; pp. 139–154. [Google Scholar]
  11. Mabuchi, Y.; Higuchi, T.; Weihnacht, V. Effect of sp2 /sp3 bonding ratio and nitrogen content on friction proper-ties of hydrogen-free DLC coatings. Tribol. Int. 2013, 62, 130–140. [Google Scholar] [CrossRef]
  12. Tashiro, H.; Nakaya, M.; Hotta, A. Enhancement of the gas barrier property of polymers by DLC coating with orga-nosilane interlayer. Diam. Relat. Mater. 2013, 35, 7–13. [Google Scholar] [CrossRef]
  13. Kumar, C.S.; Majumder, H.; Khan, A.; Patel, S.K. Applicability of DLC and WC/C low friction coatings on Al2O3/TiCN mixed ceramic cutting tools for dry machining of hardened 52100 steel. Ceram. Int. 2020, 46, 11889–11897. [Google Scholar] [CrossRef]
  14. Hauert, R. A review of modified DLC coatings for biological applications. Diam. Relat. Mater. 2003, 12, 583–589. [Google Scholar] [CrossRef]
  15. Dearnaley, G.; Arps, J.H. Biomedical applications of diamond-like carbon (DLC) coatings: A review. Surf. Coatings Technol. 2005, 200, 2518–2524. [Google Scholar] [CrossRef]
  16. Kobayashi, S.; Ohgoe, Y.; Ozeki, K.; Sato, K.; Sumiya, T.; Hirakuri, K.K.; Aoki, H. Diamond-like carbon coatings on or-thodontic archwires. Diam. Relat. Mater. 2005, 14, 1094–1097. [Google Scholar] [CrossRef]
  17. Banghard, M.; Hartmann, R.; Silmy, K.; Nisch, W.; Von Metzen, R.; Stett, A.; Bucher, V. High voltage insulation properties of DLC-Parylene multilayer films for microsurgery instruments. Microelectron. Eng. 2016, 153, 126–131. [Google Scholar] [CrossRef]
  18. Mano, T. Improvement of adhesion of diamond-like carbon films and its medical application. J. Surf. Finish. Soc. Jpn. 2007, 58, 18–22. [Google Scholar] [CrossRef]
  19. Ho, C. New device for Pentax-Airway Scope in pediatric intubation. Acta Anaesthesiol. Taiwanica 2010, 48, 178–179. [Google Scholar] [CrossRef]
  20. Lim, D.; Ku, B.; Seo, D.; Lim, C.; Oh, E.; Shim, S.E.; Baeck, S.-H. Pulse-reverse electroplating of chromium from Sargent baths: Influence of anodic time on physical and electrochemical properties of electroplated Cr. Int. J. Refract. Met. Hard Mater. 2020, 89, 105213. [Google Scholar] [CrossRef]
  21. Mayama, T.; Hiratsuka, M.; Nakamori, H.; Homma, A.; Hirakuri, K.; Ohgoe, Y. Effect of oxygen plasma treatment for different types of DLC films. In Proceedings of the 79th JSAP Autumn Meeting, Nagoya, Japan, 18–21 September 2016. [Google Scholar]
  22. Rahaman, M.L.; Zhang, L.; Liu, M.; Liu, W. Surface roughness effect on the friction and wear of bulk metallic glasses. Wear 2015, 332–333, 1231–1237. [Google Scholar] [CrossRef]
  23. Menezes, P.L.; Kishore; Kailas, S.V. Influence of surface texture and roughness parameters on friction and transfer layer formation during sliding of aluminium pin on steel plate. Wear 2009, 267, 1534–1549. [Google Scholar] [CrossRef]
  24. Tian, S.F.; Jiang, L.T.; Guo, Q.; Wu, G.H. Effect of surface roughness on tribological properties of TiB2/Al composites. Mater. Design 2014, 53, 129–136. [Google Scholar] [CrossRef]
  25. Lee, J.; He, M.; Yeo, C.-D.; Kumar, G.; Hu, Z.; Quitevis, E.L.; Thalangamaarachchige, V.D. Friction and wear of Pd-rich amorphous alloy (Pd43Cu27Ni10P20) under dry and ionic liquid (IL) lubricated conditions. Wear 2018, 190–199. [Google Scholar] [CrossRef]
  26. Sakurai, K.; Hiratsuka, M.; Nakamori, H.; Namiki, K.; Hirakuri, K. Evaluation of sliding properties and durability of DLC coating for medical devices. Diam. Relat. Mater. 2019, 96, 97–103. [Google Scholar] [CrossRef]
  27. Onodera, S.; Fujii, S.; Moriguchi, H.; Tsujioka, M.; Hirakuri, K. Antibacterial property of F doped DLC film with plasma treatment. Diam. Relat. Mater. 2020, 107, 107835. [Google Scholar] [CrossRef]
  28. International Organization for Standardization 20502:2005. Fine Ceramics (Advanced Ceramics, Advanced Technical Ceramics)—Determination of Adhesion of Ceramic Coatings by Scratch Testing; ISO: Geneva, Switzerland, 2005. [Google Scholar]
  29. Wang, J.; Ma, L.; Li, W.; Zhou, Z. Influence of different lubricating fluids on friction trauma of small intestine during enteroscopy. Tribol. Int. 2018, 126, 29–38. [Google Scholar] [CrossRef]
  30. Xu, P.; Wang, Y.; Cao, X.; Nie, X.; Yue, W.; Zhang, G. The tribological properties of DLC/SiC and DLC/Si3N4 under different relative humidity: The transition from abrasive wear to tribo-chemical reaction. Ceram. Int. 2021, 47, 3901–3910. [Google Scholar] [CrossRef]
  31. Sander, T.; Tremmel, S.; Wartzack, S. A modified scratch test for the mechanical characterization of scratch re-sistance and adhesion of thin hard coating on soft substrates. Surf. Coatings Technol. 2011, 206, 1873–1878. [Google Scholar] [CrossRef]
  32. Kato, M.; Sakaki, T. Corrosion behavior of metallic chromium in aqueous hydrochloric acid solution. J. Surf. Finish. Soc. Jpn. 1991, 55, 1108–1114. [Google Scholar]
  33. Kato, M.; Sakaki, T. Effect of alloying elements on corrosion behavior of Cr in HCl solutions. J. Surf. Finish. Soc. Jpn. 1992, 56, 74–80. [Google Scholar]
  34. Uribe-Padilla, J.; Graells, M.; Salgado-Valle, J.; Serrano, J.L. A viscosity-mediated model for relating gross and film thickness of coatings. Prog. Org. Coating 2019, 136, 105195. [Google Scholar] [CrossRef]
Figure 1. Schematic diagram of the black chrome (Black Cr) plating.
Figure 1. Schematic diagram of the black chrome (Black Cr) plating.
Materials 14 00376 g001
Figure 2. Schematic of the scratch test to measure adhesion.
Figure 2. Schematic of the scratch test to measure adhesion.
Materials 14 00376 g002
Figure 3. Schematic of the cleaning test.
Figure 3. Schematic of the cleaning test.
Materials 14 00376 g003
Figure 4. Coefficient of friction of each sample in a simulated body environment.
Figure 4. Coefficient of friction of each sample in a simulated body environment.
Materials 14 00376 g004
Figure 5. X-ray photoelectron spectroscopy (XPS) spectra for diamond-like coating (DLC)/SUS (a) before and (b) after the immersion and sterilization tests.
Figure 5. X-ray photoelectron spectroscopy (XPS) spectra for diamond-like coating (DLC)/SUS (a) before and (b) after the immersion and sterilization tests.
Materials 14 00376 g005
Figure 6. Scanning electron microscopy (SEM) images of (a) Black Cr/SUS; (b) DLC/SUS before and after immersion and sterilization tests.
Figure 6. Scanning electron microscopy (SEM) images of (a) Black Cr/SUS; (b) DLC/SUS before and after immersion and sterilization tests.
Materials 14 00376 g006
Figure 7. Surface image of (a) Black Cr/SUS and (b) DLC/SUS after cleaning test.
Figure 7. Surface image of (a) Black Cr/SUS and (b) DLC/SUS after cleaning test.
Materials 14 00376 g007
Table 1. Scratch test conditions.
Table 1. Scratch test conditions.
SamplesBlack Cr/SUS, DLC/SUS
Scratch speed (mm/min)10
Applied load increase rate (N/min)100
Initial load (N)0.9
End load (N)60
Temperature (°C)24
Humidity (%)48
Table 2. Conditions of the cleaning test.
Table 2. Conditions of the cleaning test.
SubstrateBlack Cr/SUS, DLC/SUS
Diameter of polishing sponge (mm)50
Load (N)10
Rotation speed (rpm)1290
Time (s)20
Table 3. Delamination loads of each sample measured in the scratch test.
Table 3. Delamination loads of each sample measured in the scratch test.
SampleDelamination Load (N)
N1N2N3AverageStandard Deviation
Black Cr/SUS2.853.632.643.040.43
DLC/SUS21.8123.9826.2524.011.81
Table 4. Surface element analysis of samples using energy-dispersive spectroscopy (EDS).
Table 4. Surface element analysis of samples using energy-dispersive spectroscopy (EDS).
wt.%Black Cr/SUSDLC/SUS
BeforeAfterBeforeAfter
C2.5 ± 0.212.3 ± 0.5698.1 ± 0.3198.7 ± 0.28
O32.2 ± 0.4747.2 ± 0.431.9 ± 0.121.3 ± 0.34
Fe1.2 ± 0.460.3 ± 0.20NDND
Cr63.9 ± 0.3850.3 ± 0.64NDND
Ni0.2 ± 0.27NDNDND
Mean ± SD, Not Detected (ND): < 0.02%.
Table 5. Glossiness of samples before and after the cleaning test.
Table 5. Glossiness of samples before and after the cleaning test.
SampleGlossiness (%)
beforeafter
SUS75.4 (standard)
Black Cr/SUS6.863.8
DLC/SUS15.815.4
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Kaneko, M.; Hiratsuka, M.; Alanazi, A.; Nakamori, H.; Namiki, K.; Hirakuri, K. Surface Reformation of Medical Devices with DLC Coating. Materials 2021, 14, 376. https://doi.org/10.3390/ma14020376

AMA Style

Kaneko M, Hiratsuka M, Alanazi A, Nakamori H, Namiki K, Hirakuri K. Surface Reformation of Medical Devices with DLC Coating. Materials. 2021; 14(2):376. https://doi.org/10.3390/ma14020376

Chicago/Turabian Style

Kaneko, Mao, Masanori Hiratsuka, Ali Alanazi, Hideki Nakamori, Kazushige Namiki, and Kenji Hirakuri. 2021. "Surface Reformation of Medical Devices with DLC Coating" Materials 14, no. 2: 376. https://doi.org/10.3390/ma14020376

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop