Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method
Abstract
1. Introduction
2. Theoretical Foundation and Model Establishment
2.1. Theoretical Foundation
2.1.1. Detection Efficiency
2.1.2. Energy Resolution
2.1.3. Peak-to-Total Ratio
2.2. Monte Carlo Model Establishment
3. Analysis of Simulation Calculation Results
3.1. The Effect of Crystal Size on Detection Efficiency
3.2. The Effect of Ray Energy on Detection Efficiency
3.3. The Effect of Cladding Thickness on Detection Efficiency
3.4. The Effect of Distance on Detection Efficiency
3.5. Energy Resolution
3.6. Peak-to-Total Ratio
4. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Cengiz, A. An approximation for response function to γ-rays of NaI(Tl) detectors up to 1.5MeV. Appl. Radiat. Isot. 2008, 66, 1371–1376. [Google Scholar] [CrossRef] [PubMed]
- Hajheidari, M.T.; Safari, M.J.; Afarideh, H.; Rouhi, H. Experimental validation of response function of a NaI(Tl) detector modeled with Monte Carlo codes. J. Instrum. 2016, 11, 6011. [Google Scholar] [CrossRef]
- Ahmed, Z.; Darweesh, M. Estimation of detection efficiency for NaI detector using MCNP. IOSR J. Appl. Phys. 2016, 8, 130–133. [Google Scholar]
- Mouhti, I.; Elanique, A.; Messous, M.Y. Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code. J. Mater. Environ. Sci. 2017, 8, 4560–4565. [Google Scholar] [CrossRef]
- Olaniyi, H.B.; Owojori, A.A.; Olise, F.S. Monte Carlo simulation of a NaI[Tl] detector’s response function. J. Instrum. 2017, 12, 12015. [Google Scholar] [CrossRef]
- Nilgün, D.; Nur, K.Z. Determination of energy resolution for a NaI(Tl) detector modeled with FLUKA code. Nucl. Eng. Technol. 2021, 53, 3759–3763. [Google Scholar]
- Grozdanov, D.N.; Fedorov, N.A.; Kopatch, Y.N.; Ruskov, I.N.; Skoy, V.R.; Tretyakova, T.Y.; Dabylova, S.B. Semi-empirical gamma-ray response function of BGO, NaI(Tl) and LaBr3(Ce) scintillation detectors. Nucl. Inst. Methods Phys. Res. A 2021, 1014, 165741. [Google Scholar] [CrossRef]
- Ozan, T.H.; Ghada, A.; Issa, S.; Zakaly, H.M.H.; Kilic, G.; Ene, A. Calculation of NaI(Tl) detector efficiency using 226Ra, 232Th, and 40K radioisotopes: Three-phase Monte Carlo simulation study. Open Chem. 2022, 20, 541–549. [Google Scholar] [CrossRef]
- Wang, X.J.; Yu, G.B.; Guo, J.Y.; Xu, F.; Gu, X.B.; Pang, Y. Study on the method of measuring air absorbed dose rate with NaI(Tl) detector. J. Radiat. Res. Radiat. Process. 2023, 41, 83–90. [Google Scholar]
- Sadeghi, N.; Gholamzadeh, Z.; Hassanzadeh, M. Monte Carlo simulation of photo-peak efficiency response function and calculation of total efficiency for a NaI(Tl) scintillator detector. Kerntechnik 2024, 89, 774–780. [Google Scholar] [CrossRef]
- Arhouni, F.E.; Hakkar, M.; Ouakkas, S.; Abdo, M.A.S.; Boukhair, A. Experimental Measurements and GEANT4 Simulation of NaI(Tl) Scintillator Detector. Mater. Sci. Forum 2025, 1173, 83–96. [Google Scholar] [CrossRef]
- Chen, S.Q.; Wu, B.; Dong, Z.Y.; Zhou, S.H.; Sun, G.Y.; He, L.; Yao, D.; Yu, S.P.; Gan, Q.; Hao, L.J.; et al. Development of TopMC 1.0 for nuclear technology applications. EPJ Nucl. Sci. Technol. 2025, 11, 6. [Google Scholar] [CrossRef]
- Yuan, H.; San, W.; Zhao, M.W.; Zheng, X.; Zhong, S.J. The influence of reflector on the detection efficiency of NaI(Tl) scintillator detector. Horiz. Sci. Technol. 2021, 10, 90–92. [Google Scholar]
- Li, F.L.; Zhang, X.J.; Wang, R.B. Methodology and Validation for Determining Gaussian Broadening Coefficients in MCNP. Nucl. Electron. Detect. Technol. 2013, 33, 1266–1270. [Google Scholar]
- Metwally, W.A.; Gardner, R.P.; Sood, A. Gaussian broadening of MCNP pulse height spectra. Trans. Am. Nucl. Soc. 2004, 91, 789–790. [Google Scholar]
- Sun, H.; Gu, X.B.; Yu, G.B.; Wang, H.; Wen, D.Y.; Zhang, W.L.; Liu, B. Measurement and Comparative Analysis of Nuclear Spectroscopy Performance for Common Gamma-ray Spectrometers. J. Radiat. Res. Radiat. Process. 2025, 43, 129–136. [Google Scholar]
- Guo, C.Y.; Luo, Y.F.; Lin, Y.G. Monte Carlo calculation of peak-to-total ratio for NaI(Tl) scintillation spectrometer. Nucl. Electron. Detect. Technol. 2002, 5, 453–455. [Google Scholar]









| Detect Crystal Type | Crystal Size/Inch × Inch * | εabe (%) | εp (%) | εin (%) |
|---|---|---|---|---|
| NaI(Tl) | Φ1 × 1 | 0.52 | 0.15 | 37.1 |
| Φ2 × 2 | 2.39 | 1.03 | 47.8 | |
| Φ3 × 3 | 5.19 | 2.71 | 54 |
| Source | Energy/keV | Energy Resolution | |||
|---|---|---|---|---|---|
| Simulated Value | Experimental Value [16] | ||||
| FWHM/keV | R/% | FWHM/keV | R/% | ||
| 241Am | 59.5 | 15.9 | 26.7 | 9.57 | 16.1 |
| 137Cs | 662 | 46 | 6.95 | 47.53 | 7.2 |
| 60Co | 1332.5 | 79 | 5.93 | 66.87 | 5 |
| Nuclide | γ-Ray Energy/keV | Peak-to-Total Ratio εf | ||
|---|---|---|---|---|
| Simulated Value | Experimental Value [17] | Error/% | ||
| 47Sc | 160 | 0.900 | 0.960 | 6.25 |
| 139Ce | 166 | 0.900 | 0.950 | 5.26 |
| 51Cr | 320 | 0.780 | 0.820 | 4.88 |
| 7Be | 478 | 0.643 | 0.668 | 3.74 |
| 137Cs | 662 | 0.533 | 0.536 | 0.56 |
| 95Nb | 766 | 0.490 | 0.500 | 2.00 |
| 54Mn | 835 | 0.467 | 0.474 | 1.48 |
| 65Zn | 1115 | 0.400 | 0.395 | 1.27 |
| 28Al | 1779 | 0.302 | 0.290 | 4.14 |
| 88Y | 1836 | 0.296 | 0.280 | 5.71 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Share and Cite
Du, Q.; Yang, Y.; Zhao, X.; Lv, Q.; Tang, Y.; He, J.; Liu, Y.; Li, G. Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method. Sensors 2026, 26, 3913. https://doi.org/10.3390/s26123913
Du Q, Yang Y, Zhao X, Lv Q, Tang Y, He J, Liu Y, Li G. Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method. Sensors. 2026; 26(12):3913. https://doi.org/10.3390/s26123913
Chicago/Turabian StyleDu, Qingbo, Yapeng Yang, Xiaoyu Zhao, Qi Lv, Yuyao Tang, Jiapeng He, Yier Liu, and Guoqiang Li. 2026. "Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method" Sensors 26, no. 12: 3913. https://doi.org/10.3390/s26123913
APA StyleDu, Q., Yang, Y., Zhao, X., Lv, Q., Tang, Y., He, J., Liu, Y., & Li, G. (2026). Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method. Sensors, 26(12), 3913. https://doi.org/10.3390/s26123913