Article

Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study

by
Stefano Marinari
1,*,
Pasqualina Volpe
2,
Marzia Simoni
3,
Matteo Aventaggiato
4,
Fernando De Benedetto
5,
Stefano Nardini
6,
Claudio M. Sanguinetti
7 and
Paolo Palange
8
1
Respiratory Disease Unit, G. Mazzini Hospital, 64100 Teramo, Italy
2
Presidio PneumoTisiologico Territoriale ASL 02, 66100 Chieti, Italy
3
CNR Institute of Clinical Physiology, 56124 Pisa, Italy
4
Independent Researcher, 73100 Lecce, Italy
5
FISAR Foundation (Fondazione Salute Ambiente e Respiro), 66100 Chieti, Italy
6
Italian Multidisciplinary Respiratory Society, 20127 Milan, Italy
7
Editorial Activities, Italian Multidisciplinary Respiratory Society, 20127 Milan, Italy
8
Department of Public Health and Infectious Diseases, Division of Respiratory Diseases, Umberto I Hospital, Sapienza University, 00161 Rome, Italy
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(13), 5031; https://doi.org/10.3390/s22135031
Submission received: 28 April 2022 / Revised: 20 June 2022 / Accepted: 28 June 2022 / Published: 3 July 2022
(This article belongs to the Section Wearables)

Abstract

Early diagnosis and continuous monitoring of respiratory failure (RF) in the course of the most prevalent chronic cardio-vascular (CVD) and respiratory diseases (CRD) are a clinical, unresolved problem because wearable, non-invasive, and user-friendly medical devices, which could grant reliable measures of the oxygen saturation (SpO2) and heart rate (HR) in real-life during daily activities are still lacking. In this study, we investigated the agreement between a new medical wrist-worn device (BrOxy M) and a reference, medical pulseoximeter (Nellcor PM 1000N). Twelve healthy volunteers (aged 20–51 years, 84% males, 33% with black skin, obtaining, during the controlled hypoxia test, the simultaneous registration of 219 data pairs, homogeneously deployed in the levels of Sat.O2 97%, 92%, 87%, 82% [ISO 80601-2-61:2017 standard (paragraph EE.3)]) were included. The paired T test 0 and the Bland-Altman plot were performed to assess bias and accuracy. SpO2 and HR readings by the two devices resulted significantly correlated (r = 0.91 and 0.96, p < 0.001, respectively). Analyses excluded the presence of proportional bias. For SpO2, the mean bias was −0.18% and the accuracy (ARMS) was 2.7%. For HR the mean bias was 0.25 bpm and the ARMS3.7 bpm. The sensitivity to detect SpO2 ≤ 94% was 94.4%. The agreement between BrOxy M and the reference pulse oximeter was “substantial” (for SpO2 cut-off 94% and 90%, k = 0.79 and k = 0.80, respectively). We conclude that BrOxy M demonstrated accuracy, reliability and consistency in measuring SpO2 and HR, being fully comparable with a reference medical pulseoxymeter, with no adverse effects. As a wearable device, Broxy M can measure continually SpO2 and HR in everyday life, helping in detecting and following up CVD and CRD subjects.
Keywords: wrist-watch pulsoximeter; chronic cardiac diseases; chronic respiratory diseases; heart rate; pulse oxygen saturation; reflective pulseoximeter; SpO2 wrist-watch pulsoximeter; chronic cardiac diseases; chronic respiratory diseases; heart rate; pulse oxygen saturation; reflective pulseoximeter; SpO2

Share and Cite

MDPI and ACS Style

Marinari, S.; Volpe, P.; Simoni, M.; Aventaggiato, M.; De Benedetto, F.; Nardini, S.; Sanguinetti, C.M.; Palange, P. Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study. Sensors 2022, 22, 5031. https://doi.org/10.3390/s22135031

AMA Style

Marinari S, Volpe P, Simoni M, Aventaggiato M, De Benedetto F, Nardini S, Sanguinetti CM, Palange P. Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study. Sensors. 2022; 22(13):5031. https://doi.org/10.3390/s22135031

Chicago/Turabian Style

Marinari, Stefano, Pasqualina Volpe, Marzia Simoni, Matteo Aventaggiato, Fernando De Benedetto, Stefano Nardini, Claudio M. Sanguinetti, and Paolo Palange. 2022. "Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study" Sensors 22, no. 13: 5031. https://doi.org/10.3390/s22135031

APA Style

Marinari, S., Volpe, P., Simoni, M., Aventaggiato, M., De Benedetto, F., Nardini, S., Sanguinetti, C. M., & Palange, P. (2022). Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study. Sensors, 22(13), 5031. https://doi.org/10.3390/s22135031

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop