Disposable Non-Enzymatic Glucose Sensors Using Screen-Printed Nickel/Carbon Composites on Indium Tin Oxide Electrodes

1
Department of Nanobiomedical Sciences and BK21 PLUS NBM Global Research Center for Regenerative Medicine, Dankook University, Anseo-Dong, Cheonan, Chungnam 330-714, Korea
2
Department of Chemistry, College of Natural Science, Dankook University, Anseo-Dong, Cheonan, Chungnam 330-714, Korea
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Academic Editors: Nicole Jaffrezic-Renault and Mike McShane
Sensors 2015, 15(12), 31083-31091; https://doi.org/10.3390/s151229846
Received: 1 October 2015 / Revised: 24 November 2015 / Accepted: 8 December 2015 / Published: 10 December 2015
(This article belongs to the Section Biosensors)
Disposable screen-printed nickel/carbon composites on indium tin oxide (ITO) electrodes (DSPNCE) were developed for the detection of glucose without enzymes. The DSPNCE were prepared by screen-printing the ITO substrate with a 50 wt% nickel/carbon composite, followed by curing at 400 °C for 30 min. The redox couple of Ni(OH)2/NiOOH was deposited on the surface of the electrodes via cyclic voltammetry (CV), scanning from 0–1.5 V for 30 cycles in 0.1 M NaOH solution. The DSPNCE were characterized by field-emission scanning electron microscopy (FE-SEM), X-ray photoelectron spectroscopy (XPS), and electrochemical methods. The resulting electrical currents, measured by CV and chronoamperometry at 0.65 V vs. Ag/AgCl, showed a good linear response with glucose concentrations from 1.0–10 mM. Also, the prepared electrodes showed no interference from common physiologic interferents such as uric acid (UA) or ascorbic acid (AA). Therefore, this approach allowed the development of a simple, disposable glucose biosensor. View Full-Text
Keywords: nickel/carbon composite; biosensor; non-enzymatic glucose sensor; disposable glucose sensor nickel/carbon composite; biosensor; non-enzymatic glucose sensor; disposable glucose sensor
Show Figures

Figure 1

MDPI and ACS Style

Jeon, W.-Y.; Choi, Y.-B.; Kim, H.-H. Disposable Non-Enzymatic Glucose Sensors Using Screen-Printed Nickel/Carbon Composites on Indium Tin Oxide Electrodes. Sensors 2015, 15, 31083-31091.

Show more citation formats Show less citations formats

Article Access Map by Country/Region

1
Back to TopTop