Next Article in Journal
Effect of Sunlight-Induced Isomerisation on the Biotransformation of 4′-Hydroxychalcones by Yarrowia lipolytica KCh 71
Previous Article in Journal
Stem Cells and Cell-Free Therapies for Olfactory Epithelium Regeneration: Insights from Experimental Models
Previous Article in Special Issue
Determination of Dronedarone and Debutyldronedarone in Human Plasma by HPLC-UV
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

A Sensitive SPE-LC-MS/MS Method for Determination of Selected Veterinary Drugs and Other Organic Contaminants in Human Urine: Development, Validation, and Application Study

by
Wojciech Rodzaj
1,
Małgorzata Wacławik
1,
Joanna Jurewicz
2 and
Bartosz Wielgomas
1,*
1
Department of Toxicology, Faculty of Pharmacy, Medical University of Gdańsk, 107 Hallera Street, 80-416 Gdańsk, Poland
2
Department of Chemical Safety, Nofer Institute of Occupational Medicine, 8 Teresy Street, 91-348 Łódź, Poland
*
Author to whom correspondence should be addressed.
Int. J. Mol. Sci. 2025, 26(18), 9025; https://doi.org/10.3390/ijms26189025
Submission received: 27 August 2025 / Revised: 11 September 2025 / Accepted: 12 September 2025 / Published: 16 September 2025

Abstract

The complexity of human exposure to surrounding chemicals warrants developing analytical methods that are capable of the simultaneous quantitation of many diverse environmental pollutants and their biomarkers for the needs of human biomonitoring (HBM). Examples include pesticides used in veterinary medicine, such as fipronil (FIP), imidacloprid and pyrethroids, as well as other chemicals, like bisphenols and flame retardants. The goal of this paper was to develop and validate a liquid chromatography-tandem mass spectrometry method for the quantification of selected organic contaminants in human urine. The method was then applied to real samples and used to assess the potential of a new FIP biomarker, fipronil-hydroxy (FIP-OH), for HBM. As a sample preparation protocol, enzymatic deconjugation followed by solid phase extraction were used. The method was successfully developed and validated for 16 organic pollutants and/or their metabolites, with lower limits of quantitation ranging from 0.5 to 2000 pg/mL. FIP-OH could not be included in the method, possibly due to its chemical instability. In an application study among pet owners, the detection rate for FIP was 71% (median: 3.32 pg/mL); several other chemicals were also commonly detected. The results of validation and application experiments confirm that the method can be used in HBM studies to quantify organic pollutants in urine.

1. Introduction

Humans are exposed to many chemicals on a daily basis. The variety of chemicals and sources thereof in the surrounding environment made the single, isolated exposures an exception rather than the rule. As exposure to chemicals may cause negative health effects, monitoring the magnitude is needed in order to determine the associated risk and take action, if necessary [1]. Human biomonitoring (HBM), based on the collection of biological samples from subjects and the determination of compounds of interest, is a crucial tool for exposure assessment [2]. It allows us to track exposure from all sources and routes [3] and its results can be directly translated into an internal dose, which may exert systemic effects in the human body [4]. The chemicals that are highly prioritized for exposure assessment due to their possible toxicity and widespread exposure include pesticides, bisphenols, flame retardants, and per-/polyfluoroalkyl substances (PFASs) [5,6].
Most commonly, pesticides are used to protect crops [7]. In some cases, however, their usage may extend to other areas, such as residential applications and veterinary medicine. That was the case with several flagship insecticides, including fipronil (FIP), a phenylpyrazole [8], imidacloprid (IMI), a neonicotinoid [9], and permethrin, a pyrethroid [10,11]. Other insecticides, such as flumethrin, were designed for animal treatment from the beginning [12].
FIP acts via blocking γ-aminobutyric acid-related neural transmission, leading to the paralysis and death of arthropods such as locusts, cockroaches, ticks, and fleas [13]. After its registration in many countries for crop protection, residential use, and veterinary applications in the 1990s, FIP was banned in the European Union (EU) from agriculture use in 2016 due to its toxicity towards bees [14]. Its approval for biocide use in the EU expired in 2023 [15], making the ectoparasiticidal treatment of household pets the only approved application in that region. In mammals, FIP may cause acute neurotoxic effects, whereas chronic exposure may lead to thyroid, developmental and reproductive toxicity [16].
In humans, no exposure biomarker to FIP has been established yet. In epidemiological studies, FIP itself and its active metabolite, fipronil-sulfone (FIP-sulfone), are mainly detected [17]. These compounds, however, can also be found in the environment [13], which poses a risk of external contamination of the biological samples during collection. Additionally, due to their lipophilicity, both FIP and FIP-sulfone are poorly excreted with urine [18], a matrix often used in HBM [19]. To overcome these problems, fipronil-hydroxy (FIP-hydroxy), hitherto detected only in rats, was proposed as a promising urinary biomarker of FIP exposure in humans [20]. Other FIP derivatives, that are more relevant to environmental research, include fipronil-detrifluoromethylsulfinyl (FIP-dtfms), fipronil-amide (FIP-amide), fipronil-desulfinyl (FIP-desulfinyl), and fipronil-sulfide (FIP-sulfide). FIP and its metabolites/degradation products are collectively known as fiproles (FIPs) [15].
IMI kills arthropods by binding to nicotinic receptors causing the overexcitation of the arthropods’ nervous system [21]. Introduced to the market shortly before FIP, it met a similar fate—worldwide commercial success [22] followed by restrictions of use due to environmental concerns [23]. In the EU, it is still widely used to prevent and treat ectoparasite infestations in household pets [15]. Suspected adverse effects of IMI exposure in humans involve neurotoxicity, hepatotoxicity and reproductive toxicity [24]. Biomarkers of human exposure to IMI include the parent compound and its 5-hydroxy derivative (IMI-OH) [25].
Insecticidal action of pyrethroids is based on the alteration of voltage-sensitive sodium channel kinetics in the arthropod nervous system [26]. Permethrin and flumethrin, both pyrethroid insecticides, are popular ectoparasiticides used in veterinary products for household pets [15]. In humans, pyrethroid poisoning is associated with neurotoxic effects [26]. While many studies track permethrin exposure by measuring urinary 3-phenoxybenzoic acid levels [27], this chemical can be formed in the environment [28], whereas no such reports exist for 4′-hydroxy-3-phenoxybenzoic acid (4OH3PBA), another metabolite of permethrin [29]. To the authors’ knowledge, no HBM study using biomarkers specific to flumethrin has been published yet. Animal studies indicate that 3-(2-chloro-2-(4-chlorophenyl)vinyl)-2,2-dimethylcyclopropanecarboxylic acid (CPhCA) may be used for that purpose, whereas 4-fluoro-3-phenoxybenzoic acid (4F3PBA) is a less specific alternative [12].
In contrast to the insecticides described above, the use of fungicides is heavily focused on crop protection [30]. Important examples include imazalil, boscalid, and tebuconazole [31]. While their mechanisms of action vary, all three are excreted with urine as conjugates of hydroxy metabolites produced during phase I of metabolism: imazalil O-dealkenylation leads to the formation of imazalil-despropenyl (IMZ-OH) [32], whereas aromatic oxidation of boscalid and tebuconazole yields boscalid-5-hydroxy (BOS-OH) [33] and tebuconazole-tert-butylhydroxy (TEB-OH) [34], respectively. In non-target organisms, these fungicides may induce endocrine disruption [35,36] and neurotoxicity [37].
While not a pesticide itself, DEET (N,N-diethyl-meta-toluamide) is often discussed alongside these chemicals. Instead of killing insects, DEET is used to repel them. Since DEET is often applied directly onto human skin, internal exposure can be expected [38]. In general, DEET is considered safe, but may induce neurotoxicity, especially if used in combination with insecticides [39].
Bisphenol S (BPS) is a substitute to bisphenol A, commonly used in the production of polycarbonate materials and epoxy resins. Studies show, however, that BPS exhibits similar endocrine activity to the chemical it is supposed to replace [40,41,42]. In humans, BPS is rapidly excreted with urine, mostly in its conjugated form [43].
Organophosphate flame retardants (OPFRs) are commonly used to reduce flammability of various materials [44]. Aryl OPFRs, such as triphenyl phosphate and 2-ethylhexyl diphenyl phosphate, constitute an important group of these chemicals [45]. These and some other OPFRs are known to be metabolized to diphenyl phosphate (DPhP), which is commonly detected in many populations [46]. Investigations on aryl OPFRs show that these chemicals may exert endocrine- and metabolic-disrupting effects [47].
Perfluorooctanoic acid (PFOA) is a widely known PFAS [48]. Due to its persistence, it can still be expected to be found in humans, despite its use being strongly restricted over the years [6]. PFOA exhibits carcinogenic, endocrine-disrupting, immunotoxic, and hepatotoxic properties [48].
Many chemicals, despite their different structures, may exhibit similar negative health effects [1,49,50]. Synergism between these substances may cause a stronger effect than a single chemical [51], so an effort to monitor as many potentially harmful chemicals as possible is warranted. In order to achieve this goal efficiently, wide-scope analytical methods are needed. Therefore, the main goal of this study was to develop and validate a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the determination of a diverse group of organic contaminants in human urine. Additionally, the developed protocol was used to verify the potential of FIP-hydroxy as a biomarker of human exposure. Finally, the method was used to quantify the compounds of interest in real samples in order to examine its applicability.

2. Results and Discussion

2.1. Hydroxy-Fipronil Investigation

The total ion chromatogram in negative mode of the FIP-hydroxy standard donated by prof. Bruce Hammock from UC Davis is shown below (Figure 1). Several peaks were observed instead of a single one. Since FIP-hydroxy is a hydroxylated derivative of FIP-dtfms, its retention time should be shorter than both FIP-dtfms and FIP. Only one peak, marked with an “X” on Figure 1, matched that criterion.
For the “X” peak, a mass spectrum in negative mode was obtained (Figure 2, left). As a reference, a mass spectrum of [M-H] precursor for a molecular formula corresponding to FIP-hydroxy (C11H5Cl2F3N4O) was generated using Molecular Weight Calculator, version 6.50 (Figure 2, right). Note that an intense signal at m/z 333 is present in the observed spectrum, but not in the predicted one. The intensity ratios of m/z 335, 337, and 339 are different as well. The mismatch might have been caused by coelution with another compound or partial in-source oxidation of the hydroxy group, possibly to a ketone derivative [52,53]. At the same retention time, m/z 381 and 383 in a ratio corresponding to the presence of two chlorine atoms were also observed (Figure 2, left). These m/z match the values expected for a formate adduct of FIP-hydroxy and were used as a proof of FIP-hydroxy formation in rats in the research of Vasylieva et al., 2017 [20]. Since the mobile phase used in the present study contained formate ions (see Section 3.2), such adducts could be formed here as well. However, the discrepancy between the m/z values observed for pseudomolecular ions and potential adducts makes drawing certain conclusions difficult. In the paper proposing FIP-hydroxy for use in HBM, some concerns were raised about the purity of the standard [20]. However, the unexpected signals observed here may also result from the chemical instability of the compound. To rule out the possibility of FIP-hydroxy degradation during storage, a new FIP-hydroxy standard was ordered from one of the largest manufacturers of analytical standards in the industry. However, after repeated attempts, the company failed to synthesize the product.
Finally, the MS/MS conditions for m/z 335 in negative mode (Figure 2, left) were optimized and 42 samples with quantifiable FIP(s) levels were screened for transitions obtained. A transition in positive mode, used by Vasylieva et al., 2017 [20], was included as well. The samples were prepared according to the final protocol (Section 3.4). None of the samples tested were positive for FIP-hydroxy. Consequently, FIP-hydroxy was not included in the method.
Our failure to establish FIP-hydroxy as a urinary biomarker of FIP exposure in humans may stem from several reasons. Firstly, FIP-hydroxy has only been found in rats so far [18,20], and both quantitative and qualitative differences in metabolism between rats and humans are widely reported in the literature [54,55,56,57]. Although a comparative study of human and rat liver microsomes stresses the qualitative similarity between FIP metabolism in both species [58], only one metabolite, FIP-sulfone, was taken into account in that study. Secondly, the insufficient stability of the compound itself might have led to a situation where FIP-hydroxy, despite being produced in the human body and excreted with urine, was still undetectable due to degradation during sample storage and/or preparation. Finally, assay sensitivity may have affected the results. It appears that the trifluoromethylsulfinyl moiety of the FIP molecule plays an important role in ionization efficiency in the electrospray ion source. For instance, at the same molar concentration, FIP generates a signal that is approximately 30 times higher than FIP-dtfms, which is deprived of that functional group. A similar pattern can be expected for FIP-hydroxy, making it undetectable despite being excreted with urine. Since several other urinary metabolites have been reported in laboratory animals [18,59], searching for a different urinary biomarker might yield better results in the future.

2.2. Method Development

The final protocol is described in Section 3.4. Flow-through solid phase extraction (SPE) was chosen as a sample preparation technique because it is an exhaustive extraction method and, in consequence, it provides maximum sensitivity [60]. Some of the steps of sample preparation were adapted from other methods developed in the laboratory. The deconjugation procedure was already used in the research of Klimowska and Wielgomas, 2018 [61], whereas ethyl acetate was demonstrated to be an optimal elution solvent in the research of Klimowska et al., 2023 [62]. FIP-hydroxy was not included in method development due to the reasons described in the previous section.

2.2.1. Filtration Loss Experiment

The goal of a filtration process before chromatographic analysis is to remove particles that are present in the extract which could clog the tubing and/or the column of the LC system. It is typically the final step of sample preparation [63]. However, the wrong choice of filter material and/or filtration conditions may cause substantial analyte loss [63,64]. The process of binding analytes to the filter material may involve van der Waals forces, hydrophobic interactions, hydrogen bonding, electrostatic interactions, or other phenomena. The organic content of the solution is one of the key parameters affecting this process [63]. Since the filtration step was present in all optimization experiments, the investigation on the effect of the solvent composition on analyte loss during filtration was conducted first. Nylon was used as the filter membrane material.
The results of the optimization of solvent composition before filtration are shown below (Figure 3). IMI, FIP and cis-permethrin (cis-PER) were used as model compounds for the experiments. Although cis-PER was not among the analytes quantitated in urine, it was included in this experiment to take into account compounds that are more lipophilic than FIP. As shown in Figure 3, the solvent composition was not an important factor for the recovery of IMI and FIP (average recovery within 96–116%). However, a statistically significant loss was observed for cis-PER at 60% methanol (average recovery of 81%; t-test, p = 0.0168) compared to the unfiltered reference. No statistically significant loss was observed in the case of 80 or 100% methanol. The loss of cis-PER at the lowest methanol content was not unexpected, as the lipophilic analytes are generally more strongly retained by filter membranes if the water content of the extract is high [63]. At the same time, lower methanol content in the final extract would be beneficial to the peak shape of early-eluting analytes [65]. As a compromise, 80% methanol was used to redissolve evaporated extracts before filtration.

2.2.2. Extraction Cartridge Selection

Careful consideration of the stationary phase used during the SPE procedure is necessary for efficient sample preparation [66]. To select an optimal stationary phase for SPE, Bond Elut Plexa 30 mg and Oasis HLB 60 mg extraction cartridges were compared. The results are shown in Figure 4. The recovery for most analytes was close to 100% regardless of the cartridge used. For some polar analytes, however, sharp differences between the sorbents were observed. While better recovery was observed for IMI-OH using Oasis HLB compared to Bond Elut Plexa (99 and 14%, respectively), the former performed much worse in the case of some acidic species, namely BPS, 4OH3PBA, DPhP, and PFOA (Figure 4). For instance, the average recovery of 4OH3PBA extraction was 8 and 102% for Oasis HLB and Bond Elut Plexa, respectively. In the case of FIP-amide, a surprisingly high result was obtained using Oasis HLB (176%). Since Bond Elut Plexa performed better than Oasis HLB in several cases, the former was chosen for the final protocol.
It is not clear why Oasis HLB performed so much worse for several acidic species (Figure 4), especially since its bed mass was two times higher than in Bond Elut Plexa (60 vs. 30 mg, respectively). The discrepancies between tested cartridges might stem from different chemistries of sorbent material. While the Oasis HLB sorbent is described by Waters as a copolymer of divinylbenzene and N-vinylpyrrolidone [66], Agilent describes Bond Elut Plexa as “divinylbenzene-based” [67]. However, further research would be necessary to provide a definitive answer to this question. Finally, an unexpectedly high signal of FIP-amide in the case of Oasis HLB may result from signal enhancement due to the presence of interferences in the blank urine used for the experiment.

2.2.3. Washing Step Optimization

After selecting the SPE cartridge, the sorbent wash conditions for Bond Elut Plexa 30 mg were optimized in order to wash away as many interferences as possible without losing the compounds of interest. The formic acid concentration (1%, v/v) was kept constant, while methanol percentages ranging from 0 to 25% were tested. Figure S1 depicts the results. In most cases, the analyte signal was unaffected by the methanol percentages tested. However, a sharp decrease in recovery was observed for IMZ-OH when the methanol content was increased from 15 to 20% (Figure S1). The change was statistically significant (t-test, p = 0.0004). Additionally, a slow but consistent signal increase for IMI and BPS was observed with increasing methanol content in the wash solution (Figure S1). Since the decreasing IMZ-OH signal turned out to be the limiting factor for sample cleanup, the methanol percentage for the washing step was set at 15%.
The rapid decrease in observed IMZ-OH signal with increasing methanol content was not surprising. Its parent compound, imazalil, is a weak base [68], and similar behavior was expected for IMZ-OH. The presence of formic acid in the wash solution led to IMZ-OH protonation, making it susceptible to elution with an aqueous–organic mixture. The increase in IMI and BPS signal with increasing methanol percentage is probably the result of decreasing signal suppression, since fewer interferents could be found in the final extract of the samples treated with a stronger wash solution.

2.3. Method Validation

Following optimization, the method was validated. The goal of method validation is to ensure that the assay performance is sufficient and the results it provides are reliable. Here, the validation process was conducted based on the ICH M10 guideline on bioanalytical method validation and study sample analysis [69], with the exception of the matrix effect investigation, which was carried out following the procedure described in EMA, 2011 [70]. Since the results for IMZ-OH and PFOA failed to meet the validation criteria, these analytes were omitted.

2.3.1. Selectivity

The selectivity of the analytical method was investigated by injecting blank urine samples and evaluating the analyte signal. No significant amounts of any of the analytes and/or internal standards (ISs) were found. Therefore, the selectivity of the method was assessed as satisfactory.

2.3.2. Internal Standard Selection and Matrix Effect

Mass spectrometric methods are prone to a matrix effect, which is a change in analyte response due to interferences present in the matrix. Since it may reduce the accuracy, precision, and robustness of the method [71], its presence and magnitude were carefully evaluated. Absolute matrix effects (matrix factors) for the compounds of interest are provided in Table 1. Specific gravity (SG) for all urine lots is also shown. In all cases, signal suppression was observed compared to pure solvent. Additionally, an increasing SG value was often associated with a stronger matrix effect. These observations were not unexpected as samples with higher SG contain more matrix components which contribute to matrix effects [71]. According to Ferrer Amate et al., 2010 [72], the uncorrected matrix effect within 80 and 120% corresponds to a soft matrix effect, whereas matrix effects between 50 and 79% and between 121 and 150% can be considered to be medium signal suppression and enhancement, respectively; finally, values below 49 and above 151% indicate a strong matrix effect. Based on this description and the average matrix factors (Table 1), nearly all analytes were subjected to strong signal suppression. Medium signal suppression was observed for DPhP and TEB-OH only.
This observation can be attributed to the fact that urine samples were concentrated one hundred times during sample preparation with relatively little cleanup (wash solvent used during SPE contained only 15% methanol), resulting in high amounts of interferents being injected into the LC-MS/MS system. Even under the assumption that the unconcentrated sample would cause no matrix effect, the gain in sensitivity resulting from sample preconcentration was still higher than the observed signal suppression, although the signal increase varied greatly. For instance, the average signal of TEB-OH in the matrix was 48% lower than in pure solvent (Table 1). However, taking into account the sample preconcentration, its signal was still approximately 50 times higher than the result that could be expected in a sample not subjected to preconcentration. On the other hand, a 98% average decrease in BOS-OH signal due to a matrix effect in the preconcentrated sample lead to an only twofold increase in the analytical response overall, making the signal gain almost insignificant. These two extreme examples show how diverse results can be obtained using high sample preconcentration as a strategy for improving the sensitivity of an LC-MS/MS method.
The IS-corrected matrix factors for optimal analyte–internal standard pairs are provided in Table 2. The coefficients of variation (CVs) of IS-corrected matrix factors, which reflect the relative matrix effect [73], are given in the last column of the Table 2. No clear link between the average value and urine lot SG could be found. Unsurprisingly, analytes with an isotopically labeled analog used as an internal standard exhibited low variability of the IS-corrected matrix factor. For instance, the CVs of IS-corrected matrix factors for IMI, FIP-dtfms and FIP were 6, 3, and 2%, respectively. BPS is a notable exception, with a CV of an IS-corrected matrix factor of 12%, despite the use of a deuterated analog as an internal standard (Table 2). Interestingly, a similar response variability (13%) was observed in the work of Gys et al., 2020 [74], even though the 13C-labeled internal standard was used in that work. For other analytes, CVs ranging from 9 (FIP-sulfone) to 41% (BOS-OH) were observed. In order to mitigate the variability associated with a relative matrix effect, matrix-matched internal calibration was used for quantitative analysis of all analytes [62,75,76].

2.3.3. Linearity and Lower Limit of Quantification

The lower limit of quantification (LLOQ) is “the lowest amount (concentration) of an analyte in a sample that can be quantitatively determined using a method with predefined precision and accuracy” [69]. Linearity is the assumption that the signal–concentration relationship for a given analyte forms a straight line [77]. The results of experiments on assay linearity and sensitivity are shown in Table 3. The LLOQs ranged from 0.5 (FIP-sulfone) to 2000 pg/mL (4OH3PBA). The linear range spanned two or more orders of magnitude for all analytes, except for DPhP (100–5000 pg/mL). The dilution integrity experiment allowed us to extend this range for some analytes even further (see Section 2.3.6.). The calibration curve fit was determined using the coefficient of determination (R2) and its value did not fall below 0.9900 for any analyte (Table 3).
The sensitivity of the method developed here can be compared to those achieved recently in other laboratories. For IMI biomarkers, namely IMI itself and IMI-OH, our results were similar to the limits of quantification (LOQs) observed in Wrobel et al., 2023 (190 and 1000 pg/mL, respectively) [78], but much higher than the limits of detection (LODs) reported in Baker et al., 2019 (30 and 50 pg/mL, respectively) [79]. In our method, a significant loss of IMI-OH occurred during the SPE procedure (Figure 4), strongly affecting the observed LLOQ. Nevertheless, the sensitivity for both analytes seems sufficient to monitor human exposure to IMI after ectoparasiticide application on household pets as observed in the work of Wrobel et al., 2024 [80]. The LLOQ for BPS (Table 3) was close to the values reported by other authors, such as 40 [74] and 250 pg/mL [81]. In contrast, the sensitivity of the hereby presented method for BOS-OH (1000 pg/mL) was poor compared to LODs reported by Rodríguez-Zamora et al., 2024 (50 pg/mL) [75] or Norén et al., 2024 (20 pg/mL) [82]. This discrepancy can be attributed to the very strong signal suppression of BOS-OH, as evident in the results presented in Table 1. For DPhP, the assay performed quite well, although the linear range was narrow (100–5000 pg/mL), and urinary concentrations above the upper limit of quantification (ULOQ) reported here are not out of the ordinary [46]. However, the achieved LLOQ was satisfactory—for comparison, an LOD equal to 20 pg/mL was reported in the work of Norén et al., 2024 [82], whereas in the work of Van den Eede et al., 2013, the LOQ was 300 pg/mL [83]. The 10 pg/mL LLOQ for DEET (Table 3) is considerably lower than the values reported in the recent literature, ranging from 80 [84] to 210 pg/mL [85]. The LLOQ established for TEB-OH (100 pg/mL) lies between the LOD reported by Norén et al., 2024 (16 pg/mL) [82] and the LOQ from the work of Šulc et al., 2022 (270 pg/mL) [86]. For 4F3PBA, other authors usually reported better sensitivity, such as 30 [87] or 25 pg/mL [88]. No data to compare could be found for other pyrethroid metabolites (4OH3PBA and CPhCA). The LLOQs for FIP-desulfinyl, FIP, FIP-sulfide and FIP-sulfone were 5, 1, 1, and 0.5 pg/mL, respectively (Table 3). Quantitation limits reported in the literature for these compounds ranged from 0.1 [76] to 2.5 pg/mL [89]. For FIP-dtfms and FIP-amide, no data from the literature was available for comparison.
Two important things should be mentioned here. Firstly, while the sensitivity of the developed method was usually at a similar level with the ones developed by other authors, the one presented here requires a much higher sample volume (5 mL). In the papers mentioned above, the sample volume ranged from 0.2 to 1 mL, with the exception of the work of Le Grand et al., 2012 [88], where 5 mL of urine was needed. Secondly, the assay sensitivity can be established in many different ways [90], which makes the comparisons difficult. The method used here, based on the accuracy at the LLOQ level, as recommended by the European Medicines Agency (EMA) [69], can be considered conservative compared to other methods, such as the ones based on the analyte signal-to-noise ratio, which is heavily influenced by the data collection process within the software [91].

2.3.4. Accuracy and Precision

Accuracy can be defined as the closeness of the measurement to the nominal value, whereas precision is the degree of agreement between a series of measurements [69]. The results of accuracy and precision experiments on quality control (QC) urine samples are shown in Table 4. For the vast majority of the analytes, the criteria set by the EMA guidelines [69] were fulfilled for QC samples at both low (LQC) and high (HQC) concentration. The only exception was the LQC inter-run precision for IMI-OH and DPhP (20 and 16%, respectively). The use of isotopically labeled analogs of these compounds as internal standards would certainly improve this parameter; it is planned to include them in future studies.

2.3.5. Carry-Over

Carry-over is a change in measured analyte concentration due to its residues from the previous injection still being present in the system. The carry-over experiment revealed no significant transfer of any analyte and/or internal standard residues from one injection to another.

2.3.6. Dilution Integrity

This experiment was performed in order to determine whether sample dilution affected the accuracy and precision of the results at the levels exceeding the highest level of calibration curve. The results are shown in Table S1. The eightfold dilution integrity was confirmed for IMI, BPS, FIP-dtfms, FIP-desulfinyl, FIP, FIP-sulfide, FIP-sulfone, and CPhCA.

2.3.7. Stability

In biomonitoring studies, the samples may undergo many processes before analysis, such as shipping, preparation, or short- and long-term storage [92]. Therefore, analyte stability, understood as a lack of analyte degradation in a given matrix during defined storage conditions [69], is another important validation parameter [93]. The stability studies included the 24 h autosampler stability of processed samples at room temperature and the long-term stability of unprocessed QC samples stored for 30 days and 12 months at −20 °C. The results are presented in Table 5. All analytes were stable in processed samples stored in the autosampler at both QC levels. The long-term stability study, however, revealed significant losses of BOS-OH and 4OH3PBA after 12 months at both LQC (accuracy of 46 and 77%, respectively) and HQC levels (46 and 58%, respectively). Notably, all analytes were stable in a solvent stored for the same period of time.
To the authors’ knowledge, this is the first study to investigate 1-year stability of these compounds in urine. Although it probably does not mimic the real-life scenario perfectly (in animal studies, both BOS-OH [33] and 4OH3PBA [29] are mostly excreted as conjugates), the results presented here raise concerns about their use for human biomonitoring. Spiking the urine samples before storage with stabilizers, such as ascorbic acid or butylated hydroxytoluene [94], may also be considered to prevent the degradation of these analytes. Further research is necessary to determine whether such an approach would be efficient for BOS-OH and 4OH3PBA.

2.4. Method Application

The developed method was applied to 28 urine samples collected during a longitudinal study on pet owners who recently applied an ectoparasiticide on their pet. The medication contained either fipronil, permethrin, or both insecticides. The summary statistics, both uncorrected and adjusted by SG, are presented in Table 6. For IMI-OH, IMI, FIP-dtfms, FIP-amide, 4F3PBA, FIP-desulfinyl, FIP-sulfide, and CPhCA, the detection rates were negligible (≤4%). 4OH3PBA was not commonly detected either (18% of the samples). A moderate detection rate was observed for BOS-OH and FIP-sulfone (50 and 39%, respectively). The geometric mean concentration for BOS-OH was 1238 and 1454 pg/mL (unadjusted and SG-adjusted result, respectively; all concentrations will be given in that order). Finally, BPS, DPhP, DEET, TEB-OH, and FIP were frequently detected (89, 93, 68, 89, and 71% of the samples). The highest median concentration was observed for DPhP (1712 and 1656 pg/mL), whereas the lowest was reported for FIP (3.32 and 3.39 pg/mL).
A near lack of samples that were positive for IMI-OH and IMI was not surprising, as the samples were collected after the EU ban of IMI use in plant protection products [23]. In a German study where samples were collected before the ban was imposed [78], the detection rates for IMI-OH and IMI were also fairly low (21 and 36%, respectively), even though some samples were collected following the application of an IMI-based animal care product on a pet. In contrast, BPS was frequently detected in the present study, with the uncorrected concentration median as high as 663 pg/mL (Table 6). Much lower concentrations were observed in samples collected from Polish children in 2014–2015 (median below the LOQ of 250 pg/mL). This discrepancy can be explained at least partially by the different period of sample collection. In the study on Polish children, the samples were collected before the EU ban of bisphenol A for use in thermal paper, which was imposed in 2020. It was followed by increased use of other bisphenols, such as BPS [95]. Since the samples for the present study were collected in 2021–2022, higher BPS levels could be expected. Indeed, the exposure to BPS has been reported to increase over the past several years in Sweden [82] and Denmark [96]. Limited data exist for urinary levels of BOS-OH in humans; a SG-corrected geometric mean (GM) equal to 330 pg/mL was reported in farmworkers in Costa Rica [75], whereas the 50th percentile in Swedish adolescents was 160 pg/mL (uncorrected concentration) [82]. In our study, a SG-corrected GM of 1454 pg/mL was observed, suggesting much higher exposure. Relatively high urinary levels were also observed for DPhP (uncorrected median: 1712 pg/mL). In comparison, the median concentration in a nationwide US study was 820 pg/mL [97], whereas in Swedish adolescents it was equal to 930 pg/mL [82]. The value observed here, however, was still lower than the median reported in a study on American firefighters (2900 pg/mL) [98]. The DEET detection rate was also higher than in other studies; for instance, a method application study on 75 anonymous adult volunteers reported 12% detection frequency [38]. This difference may stem from the higher LOD than the LLOQ reported here (100 and 10 pg/mL, respectively). TEB-OH prevalence and concentrations observed in our study were slightly lower than in an adult population of the Czech Republic (detection rate ≥ 95%, SG-corrected median of 470 and 440 pg/mL for winter and summer seasons, respectively) [86]. Finally, the relatively high detection frequency and concentrations of FIP and FIP-sulfone observed in our study probably result from the ectoparasiticide application prior to sample collection. This observation suggests that FIP and FIP-sulfone can be used as biomarkers of human exposure to FIP in humans. To compare, FIP-sulfone was detected in less than 4% of Czech adults [86] and in only 10% of a general population in China [89]. A similar detection rate of FIP and FIP-sulfone (48 and 40%, respectively) was observed only in a longitudinal study on urinary excretion variability conducted in Luxembourg/France [99]. Five out of sixteen volunteers who participated in that study reported having a pet treated with ectoparasiticide, and that is probably why the detection rate was comparable to the results presented herein. In line with other biomonitoring studies on fiproles [17], FIP-amide, FIP-desulfinyl, and FIP-sulfide were not detected.
It should be noted that due to the small sample size, the results presented here may not be representative of the general population. Additionally, due to the study intervention mentioned earlier, the levels of fipronil and pyrethroid biomarkers observed in this sample are likely to be elevated compared to others.

3. Materials and Methods

3.1. Chemicals and Materials

Analytical standards used in this work (both analytes and internal standards) are presented in Table S2. β-Glucuronidase from Helix pomatia type HP-2 (low sulfatase activity, catalog number G7017), glacial acetic acid, and formic acid pro analysis were purchased from Sigma-Aldrich (Burlington, MA, USA). Isopropanol, acetonitrile, methanol, water and formic acid (all LC-MS grade) were obtained from Supelco (Bellefonte, PA, USA). Ethyl acetate GC-MS grade and Extran MA 01 laboratory detergent were obtained from that supplier as well. Anhydrous sodium acetate was purchased from POCH (Gliwice, Poland). Technical nitrogen and argon 6.0 were acquired from Oxygen (Gdańsk, Poland) and Air Products (Allentown, PA, USA), respectively.
Bond Elut Plexa (30 mg) and Oasis HLB 3 cc (60 mg) cartridges were purchased from Agilent Technologies (Santa Clara, CA, USA) and Waters (Drinagh, Ireland), respectively. Centrifugal filters (0.2 µm pore size, modified nylon), rimless test tubes, and disposable transfer pipettes were obtained from VWR (Leuven, Belgium). Screw-top borosilicate glass tubes were provided by Corning (Corning, NY, USA), whereas phenolic screw thread caps (15-415) were obtained from Sun-Sri (Rockwood, TN, USA). Sarstedt (Nümbrecht, Germany) was the supplier of all pipette tips. Chromatography vials, along with caps and microinserts, were purchased from La-Pha-Pack (Langerwehe, Germany).

3.2. Instrumental Analysis

The chromatographic separation was performed using a Varian system (Varian, Walnut Creek, CA, USA) consisting of two 212-LC dual piston pumps, a ProStar 420 autosampler, and a high pressure mixer. The heated column compartment from Dionex (Sunnyvale, CA, USA) was used. ACE Excel 3 SuperC18, 75 × 3.0 mm (particle size 3 µm) from Advanced Chromatography Technologies (Aberdeen, Scotland) was used as an analytical column. Mobile phase A consisted of 0.5 mM of ammonium formate buffer at a pH of 3 in water:methanol in a 9:1 ratio (v/v), whereas phase B was the same buffer in pure methanol. The gradient program, along with the list of other chromatographic parameters, is shown in Table S3.
A 320-MS triple quadrupole mass spectrometer with an electrospray interface (Varian) was used for analyte detection. A detailed list of MS parameters is shown in Table S4.
In Table 7, compound-specific parameters for instrumental analysis are shown. MS Workstation software, version 6.9.3 (Varian) was used for system control, data collection, and analysis.

3.3. Fipronil-Hydroxy

As mentioned in the Introduction, FIP-hydroxy is a promising candidate for a urinary biomarker of human FIP exposure. However, it has only been detected in rats so far [20]. After an analytical standard was kindly provided by prof. Bruce Hammock from UC Davis, a series of experiments were conducted in order to investigate the FIP-hydroxy potential for human biomonitoring. Molecular Weight Calculator software, version 6.50 (Matthew Monroe, Richland, WA, USA), was used to generate a theoretical mass spectrum of the FIP-hydroxy parent ion. Since the precursor m/z observed during the infusion of the FIP-hydroxy standard into the mass spectrometer did not match the expected values, additional LC-MS(/MS) analyses of FIP-hydroxy were performed. LC conditions were the same as in Table S3. Scans were run in both positive and negative mode (m/z range 100–920) at 70 V of capillary voltage. Several peaks were observed in negative mode and the MS/MS parameters were determined separately for all peaks selected for further analysis.

3.4. Final Protocol of Sample Preparation

All reusable glassware used during method development and sample preparation was thoroughly cleaned using ultrasonic bath and laboratory detergent, then rinsed with methanol, and baked in a muffle furnace at 350 °C for 4 h; single-use glassware was baked in the furnace as well.
In the final method, 5 mL aliquots of urine samples in glass screw cap tubes were spiked with an IS mixture and incubated overnight at 37 °C with 1250 µL of β-glucuronidase type HP-2 from Helix pomatia dispersed in 1 M of acetate buffer with a pH of 5.0 (glucuronidase and sulfatase activity: 300 and 3 U/mL of buffer, respectively); the process was stopped the next day by the addition of 750 µL of formic acid. The deconjugation procedure described above is an adapted protocol used elsewhere [61]. Following mixing and centrifugation, the supernatants were loaded on Bond Elut Plexa 30 mg cartridges preconditioned sequentially with 1 mL of ethyl acetate, 1 mL of 1% formic acid in methanol (v/v), and 1 mL of 1% formic acid in water (v/v). After loading, the cartridges were washed with 1 mL of 1% formic acid (v/v) in 15% methanol (v/v) and dried for 30 min in an SPE dryer connected to a vacuum pump. Analytes were eluted with 4 × 250 µL of ethyl acetate, similarly to the work of Klimowska et al., 2023 [62], and then carefully evaporated just to dryness at 40 °C under a nitrogen stream. The reconstitution comprised several steps. Firstly, 40 µL of methanol and 10 µL of water were added to the tube. After the addition of each of the solvents, the content was vortexed at 2000 rpm. Then, to prevent the loss of the extract, the tubes were centrifuged for 2 min at 1500 rpm. Finally, the entire extract was transferred to centrifugal filters (nylon, pore size 0.2 µm) and centrifuged for 3 min at 14,000× g. The content of centrifugal tubes was then transferred to amber glass vials containing glass microinserts and injected into an LC-MS/MS system.
The outline of the procedure is shown below (Figure 5); a description of optimization experiments follows.

3.5. Method Development

3.5.1. Filtration Loss Experiment

The following solvent compositions were tested: 60, 80, and 100% methanol (v/v). To cover a wide range of lipophilicity, mixture of three compounds in acetonitrile was prepared: IMI (80,000 pg/mL; logP 0.6), FIP (2000 pg/mL; logP 3.8), and cis-PER (160,000 pg/mL; logP 6.1) [68]. A sample of 100 µL of the mixture was added to glass tubes and evaporated. The reconstitution was performed as follows: first, methanol was added and the tube content was mixed, then water was transferred into the tube followed by short vortexing. For each methanol percentage, three replicates were prepared and filtered. Since the solvent composition of a sample injected into an LC system is known to affect peak shape and height [65], three unfiltered samples with the same methanol content acted as a reference for every tested percentage (100% recovery).

3.5.2. Extraction Cartridge Selection

Pooled urine used for this experiment was split in half; one was left unchanged, whereas the other was spiked before extraction with a mixture of standards so that the final analyte concentrations in urine were as follows:
  • 20 pg/mL: FIP-desulfinyl, FIP, FIP-sulfide, FIP-sulfone;
  • 100 pg/mL: DEET, FIP-amide, PFOA;
  • 500 pg/mL: BPS, DPhP, FIP-dtfms, TEB-OH;
  • 1000 pg/mL: 4F3PBA;
  • 2000 pg/mL: IMI;
  • 5000 pg/mL: IMZ-OH, BOS-OH;
  • 10,000 pg/mL: IMI-OH, 4OH3PBA, CPhCA.
Three aliquots of spiked and unspiked urine were loaded on each type of the cartridges that were preconditioned with 1 mL (Bond Elut Plexa 30 mg) or 2 mL (Oasis HLB 60 mg) of 1% formic acid in methanol (v/v) and 1% formic acid in water (v/v). Additionally, a single reagent and urine blank for both cartridges was prepared, so eight samples in total were run per sorbent. The washing step was performed using 1% formic acid (v/v) in 5% methanol (v/v); again, 1 mL of washing solution was added to Plexa, and 2 mL was added to Oasis cartridges. After the drying step, the analytes were eluted using 1 mL (Plexa) or 2 mL (Oasis) of ethyl acetate; at that moment, the mixture of analytes corresponding to 100% recovery was added to three unspiked urine extracts per sorbent. After the evaporation of ethyl acetate under a nitrogen stream, the dry residue was reconstituted using 80% methanol (v/v) and injected into the LC-MS/MS system.

3.5.3. Washing Step Optimization

This experiment was conducted using Bond Elut Plexa in 30 mg only. Samples of 1% formic acid (v/v) in 0, 5, 10, 15, 20, and 25% methanol (v/v) were tested (n = 3). The same (un)spiked urine as in the previous section was used, and the rest of the SPE procedure remained unchanged.

3.6. Method Validation

Method validation was performed basing on EMA ICH M10 guidelines [69], with the exception of matrix effect investigation, which was performed in accordance with earlier EMA guidelines [70]. A detailed description of performed experiments is provided below.

3.6.1. Selectivity

Selectivity as a capability of an analytical method to differentiate and measure the analyte(s) despite the presence of interferences was assessed by injecting several blank samples from separate sources. The acceptable threshold was less than or equal to 20% of the analyte response at the LLOQ level and no more than 5% of the IS response in the LLOQ sample.

3.6.2. Internal Standard Selection and Matrix Effect

To investigate the matrix effect, seven different lots of urine were prepared in triplicate (SG range 1.006–1.031, determined refractometrically) and fortified post-extraction with analytes. Internal standards were added as well, in amounts corresponding to the following concentrations in urine (pg/mL): IMI-D4, 10,000; BPS-D8, 5000; FIP-dtfms-13C215N2, 25,000; 3PBA-13C4, 50,000; FIP-13C4, 1000. In parallel, three repetitions of analyte and IS mixture in pure solvent were prepared. Following analysis, the matrix factor was calculated as the ratio of analyte peak area in the blank matrix to the analyte peak area in pure solvent and was expressed in %. A result equal to 100% indicates no matrix effect, while results below that indicate signal suppression, and values above 100% reflect signal enhancement. The IS-normalized matrix factor was measured by dividing the matrix factor of a given analyte by the matrix factor of the IS. A CV not greater than 15% for an IS-normalized matrix factor was considered satisfactory. For each analyte, all ISs available were tested; the one that provided the lowest CV of the IS-normalized matrix factor was chosen for routine analysis.

3.6.3. Linearity and Lower Limit of Quantification

The calibration curves were prepared by spiking a blank urine matrix at 10 calibration levels for all analytes. Blank and zero samples (blank sample spiked with IS) were run in parallel. The calibration range was based on pre-validation studies on assay sensitivity and expected concentrations in real samples. The curves were prepared and run in quadruplicate over a four-day period. For LLOQ determination, the acceptable accuracy of each standard was ±20% of the nominal concentration; for other levels, accuracy within ±15% was considered sufficient. A minimum of 75% of the samples at each calibration level had to meet the aforementioned criteria. For each analyte, curve fitting and weighting were assessed using the MS Workstation software, version 6.9.3. The linearity was monitored using R2; a value above 0.9900 was considered acceptable.

3.6.4. Accuracy and Precision

The QC samples were prepared from a single source of a blank matrix at two concentration levels: low (LQC) and high (HQC). Within-run accuracy and precision were determined by analyzing 5 replicates at both concentration levels in a single run; 15 replicates over three days were run and combined to assess between-run accuracy and precision. In both experiments, accuracy within ±15% of the nominal concentration and precision (measured as CV) less than or equal to 15% were considered acceptable.
In non-validation runs, usually consisting of 48 samples, two LQC and HQC samples were run. At least three had to be within ±15% of the nominal values for the run to be accepted.

3.6.5. Carry-Over

Carry-over was assessed by the analysis of blank solvent analysis preceded by the injection of a sample at the highest calibration level. The maximum acceptable carry-over was 20% of the LLOQ for the analytes and 5% of the response for the internal standards.

3.6.6. Dilution Integrity

The same urine lot as was used for the accuracy and precision experiments was spiked with analytes at a level eight times higher than the ULOQ and prepared normally in five replicates. Before instrumental analysis, the samples were diluted eightfold with blank solvent. Accuracy and precision were determined and acceptance criteria were the same as for LQC and HQC samples (see above).

3.6.7. Stability

Several stability studies were conducted. In all cases, the samples were run in triplicates. The accuracy needed to be within 15% of nominal concentration and a precision (CV) less than or equal to 15% for the result to be accepted. Firstly, a 24 h autosampler stability of processed samples at room temperature was performed. The LQC and HQC urine samples prepared according to the final protocol (see Section 3.4.) were injected at t = 0 and t = 24 h. Secondly, a single batch of unprocessed LQC and HQC samples was used to determine the 30-day and 12-month storage stability at −20 °C. Before the 12-month stability study was conducted, the stability of the standard mixture used to prepare calibration curves at the same storage conditions was investigated.

3.7. Method Application

Following validation, the method was used to quantitate the compounds of interest in 28 human urine samples collected in 2021–2022 from Polish pet owners who recently applied an ectoparasiticidal medication on their pet. As pet owners are expected to be dermally exposed to veterinary drugs by direct contact with their pet(s) following ectoparasiticide treatment [80,100], this population seemed particularly fit for the method application study performed here. The detailed methodology and results of that study will be a subject of a separate publication. Apart from volumetric concentrations, the SG-corrected results were also calculated to take the fluctuations in urine dilution into account [101]. Since each of the participants provided several samples, the data cannot be considered as independent; therefore, no correlation analysis was performed.

Supplementary Materials

The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/ijms26189025/s1.

Author Contributions

Conceptualization, B.W., W.R., M.W. and J.J.; formal analysis, W.R.; funding acquisition, B.W. and W.R.; investigation, W.R. and M.W.; methodology, W.R. and B.W.; resources, B.W. and J.J.; supervision, B.W. and J.J.; visualization, W.R.; writing—original draft, W.R.; writing—review and editing, B.W. and J.J. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Science Centre, Poland, grant number 2018/31/N/NZ7/02512.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and approved by the Medical University of Gdańsk Bioethics Committee for Scientific Research (Resolution No. NKBBN/535/2020; approval date: 7 October 2020).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Acknowledgments

A part of this work was included in Wojciech Rodzaj’s doctoral dissertation. Additionally, some of the results were presented at the 12th International Symposium on Biological Monitoring in Occupational and Environmental Health (21–23 June 2023, Porto, Portugal) and the 14th Scientific and Training Conference of the Polish Society of Toxicology (4–6 September 2024, Poznań, Poland).

Conflicts of Interest

The authors declare no conflicts of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Abbreviations

The following abbreviations are used in this manuscript:
HBMHuman biomonitoring
PFASPer-/polyfluoroalkyl substance
FIPFipronil
IMIImidacloprid
EUEuropean Union
FIP-sulfoneFipronil-sulfone
FIP-hydroxyFipronil-hydroxy
FIP-dtfmsFipronil-detrifluoromethylsulfinyl
FIP-amideFipronil-amide
FIP-desulfinylFipronil-desulfinyl
FIP-sulfideFipronil-sulfide
FIPsFiproles
IMI-OHImidacloprid-5-hydroxy
4OH3PBA4′-hydroxy-3-phenoxybenzoic acid
CPhCA3-(2-chloro-2-(4-chlorophenyl)vinyl)-2,2-dimethylcyclopropanecarboxylic acid
4F3PBA4-fluoro-3-phenoxybenzoic acid
IMZ-OHImazalil-despropenyl
BOS-OHBoscalid-5-hydroxy
TEB-OHTebuconazole-tert-butylhydroxy
DEETN,N-Diethyl-meta-toluamide
BPSBisphenol S
OPFRsOrganophosphate flame retardants
DPhPDiphenyl phosphate
PFOAPerfluorooctanoic acid
LC-MS/MSLiquid chromatography-tandem mass spectrometry
SPESolid phase extraction
cis-PERcis-Permethrin
SGSpecific gravity
ISInternal standard
CVCoefficient of variation
LLOQLower limit of quantification
R2Coefficient of determination
LOQLimit of quantification
LODLimit of detection
ULOQUpper limit of quantification
EMAEuropean Medicines Agency
QCQuality control
LQCQuality control sample at low concentration
HQCQuality control sample at high concentration
GMGeometric mean

References

  1. Løkke, H.; Ragas, A.M.J.; Holmstrup, M. Tools and Perspectives for Assessing Chemical Mixtures and Multiple Stressors. Toxicology 2013, 313, 73–82. [Google Scholar] [CrossRef]
  2. Angerer, J.; Ewers, U.; Wilhelm, M. Human Biomonitoring: State of the Art. Int. J. Hyg. Environ. Health 2007, 210, 201–228. [Google Scholar] [CrossRef] [PubMed]
  3. Willey, J.B.; Pollock, T.; Thomson, E.M.; Liang, C.L.; Maquiling, A.; Walker, M.; St-Amand, A. Exposure Load: Using Biomonitoring Data to Quantify Multi-Chemical Exposure Burden in a Population. Int. J. Hyg. Environ. Health 2021, 234, 113704. [Google Scholar] [CrossRef] [PubMed]
  4. Needham, L.L.; Calafat, A.M.; Barr, D.B. Uses and Issues of Biomonitoring. Int. J. Hyg. Environ. Health 2007, 210, 229–238. [Google Scholar] [CrossRef]
  5. Ougier, E.; Ganzleben, C.; Lecoq, P.; Bessems, J.; David, M.; Schoeters, G.; Lange, R.; Meslin, M.; Uhl, M.; Kolossa-Gehring, M.; et al. Chemical Prioritisation Strategy in the European Human Biomonitoring Initiative (HBM4EU)—Development and Results. Int. J. Hyg. Environ. Health 2021, 236, 113778. [Google Scholar] [CrossRef] [PubMed]
  6. Vorkamp, K.; Castaño, A.; Antignac, J.P.; Boada, L.D.; Cequier, E.; Covaci, A.; Esteban López, M.; Haug, L.S.; Kasper-Sonnenberg, M.; Koch, H.M.; et al. Biomarkers, Matrices and Analytical Methods Targeting Human Exposure to Chemicals Selected for a European Human Biomonitoring Initiative. Environ. Int. 2021, 146, 106082. [Google Scholar] [CrossRef]
  7. Bolognesi, C.; Merlo, F.D. Pesticides: Human Health Effects. In Encyclopedia of Environmental Health; Nriagu, J.O., Ed.; Elsevier: Amsterdam, The Netherlands, 2020; pp. 118–132. [Google Scholar] [CrossRef]
  8. Tingle, C.C.D.; Rother, J.A.; Dewhurst, C.F.; Lauer, S.; King, W.J. Health and Environmental Effects of Fipronil; Pesticide Action Network UK: Brighton and Hove, UK, 2000; pp. 1–30. [Google Scholar]
  9. Rust, M.K. Advances in the Control of Ctenocephalides Felis (Cat Flea) on Cats and Dogs. Trends Parasitol. 2005, 21, 232–236. [Google Scholar] [CrossRef]
  10. Matsuo, N. Discovery and Development of Pyrethroid Insecticides. Proc. Japan Acad. Ser. B Phys. Biol. Sci. 2019, 95, 378–400. [Google Scholar] [CrossRef]
  11. Nakagawa, L.E.; do Nascimento, C.M.; Costa, A.R.; Polatto, R.; Papini, S. Persistence of Indoor Permethrin and Estimation of Dermal and Non-Dietary Exposure. J. Expo. Sci. Environ. Epidemiol. 2020, 30, 547–553. [Google Scholar] [CrossRef]
  12. FAO/WHO. Pesticide Residues in Food—1997 Evaluations, Part II—Toxicological; FAO/WHO: Geneva, Switzerland, 1998; Available online: https://www.inchem.org/documents/jmpr/jmpmono/v097pr01.htm (accessed on 5 August 2025).
  13. Gunasekara, A.S.; Truong, T.; Goh, K.S.; Spurlock, F.; Tjeerdema, R.S. Environmental Fate and Toxicology of Fipronil. J. Pestic. Sci. 2007, 32, 189–199. [Google Scholar] [CrossRef]
  14. European Comission. Commission Implementing Regulation (EU) No 2016/2035 of 21 November 2016 Amending Implementing Regulation (EU) No 540/2011 as Regards the Approval Periods of the Active Substances Fipronil and Maneb. Off. J. Eur. Union 2016, L 314, 7–8. [Google Scholar]
  15. EMA. Reflection Paper on the Environmental Risk Assessment of Ectoparasiticidal Veterinary Medicinal Products Used in Cats and Dogs; EMA/CVMP/ERA/31905/2021; EMA: Amsterdam, The Netherlands, 2023. [Google Scholar]
  16. FAO/WHO. Pesticide Residues in Food 2021: Joint FAO/WHO Meeting on Pesticide Residues. Evaluation Part II—Toxicological; FAO/WHO: Geneva, Switzerland, 2022. [Google Scholar]
  17. Chen, D.; Li, J.; Zhao, Y.; Wu, Y. Human Exposure of Fipronil Insecticide and the Associated Health Risk. J. Agric. Food Chem. 2022, 70, 63–71. [Google Scholar] [CrossRef]
  18. Cravedi, J.P.; Delous, G.; Zalko, D.; Viguié, C.; Debrauwer, L. Disposition of Fipronil in Rats. Chemosphere 2013, 93, 2276–2283. [Google Scholar] [CrossRef] [PubMed]
  19. Verner, M.A.; Salame, H.; Housand, C.; Birnbaum, L.S.; Bouchard, M.F.; Chevrier, J.; Aylward, L.L.; Naiman, D.Q.; Lakind, J.S. How Many Urine Samples Are Needed to Accurately Assess Exposure to Non-Persistent Chemicals? The Biomarker Reliability Assessment Tool (BRAT) for Scientists, Research Sponsors, and Risk Managers. Int. J. Environ. Res. Public Health 2020, 17, 9102. [Google Scholar] [CrossRef] [PubMed]
  20. Vasylieva, N.; Barnych, B.; Wan, D.; El-Sheikh, E.S.A.; Nguyen, H.M.; Wulff, H.; McMahen, R.; Strynar, M.; Gee, S.J.; Hammock, B.D. Hydroxy-Fipronil Is a New Urinary Biomarker of Exposure to Fipronil. Environ. Int. 2017, 103, 91–98. [Google Scholar] [CrossRef] [PubMed]
  21. Kasiotis, K.M.; Machera, K. Neonicotinoids and Their Metabolites in Human Biomonitoring: A Review. Hell. Plant Prot. J. 2015, 8, 33–45. [Google Scholar] [CrossRef]
  22. Simon-Delso, N.; Amaral-Rogers, V.; Belzunces, L.P.; Bonmatin, J.M.; Chagnon, M.; Downs, C.; Furlan, L.; Gibbons, D.W.; Giorio, C.; Girolami, V.; et al. Systemic Insecticides (Neonicotinoids and Fipronil): Trends, Uses, Mode of Action and Metabolites. Environ. Sci. Pollut. Res. 2015, 22, 5–34. [Google Scholar] [CrossRef]
  23. European Commission. Commission Implementing Regulation (EU) 2020/1643 of 5 November 2020 Amending Implementing Regulation (EU) No 540/2011 as Regards the Approval Periods of the Active Substances Calcium Phosphide, Denathonium Benzoate, Haloxyfop-P, Imidacloprid, Pencycuron. Off. J. Eur. Union 2020, L 370, 18–20. [Google Scholar]
  24. Han, W.; Tian, Y.; Shen, X. Human Exposure to Neonicotinoid Insecticides and the Evaluation of Their Potential Toxicity: An Overview. Chemosphere 2018, 192, 59–65. [Google Scholar] [CrossRef]
  25. Wrobel, S.A.; Bury, D.; Koslitz, S.; Hayen, H.; Koch, H.M.; Brüning, T.; Käfferlein, H.U. Quantitative Metabolism and Urinary Elimination Kinetics of Seven Neonicotinoids and Neonicotinoid-Like Compounds in Humans. Environ. Sci. Technol. 2023, 57, 19285–19294. [Google Scholar] [CrossRef]
  26. Soderlund, D.M.; Clark, J.M.; Sheets, L.P.; Mullin, L.S.; Piccirillo, V.J.; Sargent, D.; Stevens, J.T.; Weiner, M.L. Mechanisms of Pyrethroid Neurotoxicity: Implications for Cumulative Risk Assessment. Toxicology 2002, 171, 3–59. [Google Scholar] [CrossRef]
  27. Rodzaj, W.; Wileńska, M.; Klimowska, A.; Dziewirska, E.; Jurewicz, J.; Walczak-Jędrzejowska, R.; Słowikowska-Hilczer, J.; Hanke, W.; Wielgomas, B. Concentrations of Urinary Biomarkers and Predictors of Exposure to Pyrethroid Insecticides in Young, Polish, Urban-Dwelling Men. Sci. Total Environ. 2021, 773, 145666. [Google Scholar] [CrossRef] [PubMed]
  28. Sudakin, D.L. Pyrethroid Insecticides: Advances and Challenges in Biomonitoring. Clin. Toxicol. 2006, 44, 31–37. [Google Scholar] [CrossRef] [PubMed]
  29. Kaneko, H. Pyrethroid Chemistry and Metabolism. In Hayes’ Handbook of Pesticide Toxicology; Elsevier: Amsterdam, The Netherlands, 2010; Volume 1, pp. 1635–1663. [Google Scholar] [CrossRef]
  30. Toda, M.; Beer, K.D.; Kuivila, K.M.; Chiller, T.M.; Jackson, B.R. Trends in Agricultural Triazole Fungicide Use in the United States, 1992–2016 and Possible Implications for Antifungal-Resistant Fungi in Human Disease. Environ. Health Perspect. 2021, 129, 055001. [Google Scholar] [CrossRef] [PubMed]
  31. Jeschke, P.; Witschel, M.; Kramer, W.; Schirmer, U. Modern Crop Protection Compounds: Volume 2: Fungicides, 3rd ed.; Wiley-VCH: Weinheim, Germany, 2019. [Google Scholar]
  32. FAO/WHO. Pesticide Residues in Food—2018 Evaluations, Part II—Toxicological; FAO/WHO: Geneva, Switzerland, 2018. [Google Scholar]
  33. FAO/WHO. Pesticide Residues in Food—2006 Evaluations, Part II—Toxicological; FAO/WHO: Geneva, Switzerland, 2008. [Google Scholar]
  34. Mercadante, R.; Polledri, E.; Scurati, S.; Moretto, A.; Fustinoni, S. Identification and Quantification of Metabolites of the Fungicide Tebuconazole in Human Urine. Chem. Res. Toxicol. 2014, 27, 1943–1949. [Google Scholar] [CrossRef]
  35. Jin, C.; Zhang, R.; Fu, Z.; Jin, Y. Maternal Exposure to Imazalil Disrupts the Endocrine System in F 1 Generation Mice. Mol. Cell. Endocrinol. 2019, 486, 105–112. [Google Scholar] [CrossRef]
  36. Taxvig, C.; Hass, U.; Axelstad, M.; Dalgaard, M.; Boberg, J.; Andeasen, H.R.; Vinggaard, A.M. Endocrine-Disrupting Activities In Vivo of the Fungicides Tebuconazole and Epoxiconazole. Toxicol. Sci. 2007, 100, 464–473. [Google Scholar] [CrossRef]
  37. Wang, H.; Meng, Z.; Liu, F.; Zhou, L.; Su, M.; Meng, Y.; Zhang, S.; Liao, X.; Cao, Z.; Lu, H. Characterization of Boscalid-Induced Oxidative Stress and Neurodevelopmental Toxicity in Zebrafish Embryos. Chemosphere 2020, 238, 124753. [Google Scholar] [CrossRef]
  38. Kuklenyik, P.; Baker, S.E.; Bishop, A.M.; Morales-A, P.; Calafat, A.M. On-Line Solid Phase Extraction-High Performance Liquid Chromatography-Isotope Dilution-Tandem Mass Spectrometry Approach to Quantify N,N-Diethyl-m-Toluamide and Oxidative Metabolites in Urine. Anal. Chim. Acta 2013, 787, 267–273. [Google Scholar] [CrossRef]
  39. del-Rahman, A.; Dechkovskaia, A.M.; Goldstein, L.B.; Bullman, S.H.; Khan, W.; El-Masry, E.M.; Abou-Donia, M.B. Neurological Deficits Induced by Malathion, Deet, and Permethrin, Alone or in Combination in Adult Rats. J. Toxicol. Environ. Health Part A 2004, 67, 331–356. [Google Scholar] [CrossRef]
  40. Eladak, S.; Grisin, T.; Moison, D.; Guerquin, M.J.; N’Tumba-Byn, T.; Pozzi-Gaudin, S.; Benachi, A.; Livera, G.; Rouiller-Fabre, V.; Habert, R. A New Chapter in the Bisphenol a Story: Bisphenol S and Bisphenol F Are Not Safe Alternatives to This Compound. Fertil. Steril. 2015, 103, 11–21. [Google Scholar] [CrossRef]
  41. Silva, J.P.A.; Ramos, J.G.; Campos, M.S.; da Silva Lima, D.; de Azevedo Brito, P.V.; Mendes, E.P.; Taboga, S.R.; Biancardi, M.F.; Ghedini, P.C.; Santos, F.C.A. Bisphenol-S Promotes Endocrine-Disrupting Effects Similar to Those Promoted by Bisphenol-A in the Prostate of Adult Gerbils. Reprod. Toxicol. 2019, 85, 83–92. [Google Scholar] [CrossRef]
  42. Frenzilli, G.; Martorell-Ribera, J.; Bernardeschi, M.; Scarcelli, V.; Jönsson, E.; Diano, N.; Moggio, M.; Guidi, P.; Sturve, J.; Asker, N. Bisphenol A and Bisphenol S Induce Endocrine and Chromosomal Alterations in Brown Trout. Front. Endocrinol. 2021, 12, 645519. [Google Scholar] [CrossRef]
  43. Oh, J.; Choi, J.W.; Ahn, Y.A.; Kim, S. Pharmacokinetics of Bisphenol S in Humans after Single Oral Administration. Environ. Int. 2018, 112, 127–133. [Google Scholar] [CrossRef] [PubMed]
  44. Marklund, A.; Andersson, B.; Haglund, P. Screening of Organophosphorus Compounds and Their Distribution in Various Indoor Environments. Chemosphere 2003, 53, 1137–1146. [Google Scholar] [CrossRef] [PubMed]
  45. Reemtsma, T.; García-López, M.; Rodríguez, I.; Quintana, J.B.; Rodil, R. Organophosphorus Flame Retardants and Plasticizers in Water and Air I. Occurrence and Fate. TrAC—Trends Anal. Chem. 2008, 27, 727–737. [Google Scholar] [CrossRef]
  46. Saillenfait, A.M.; Ndaw, S.; Robert, A.; Sabaté, J.P. Recent Biomonitoring Reports on Phosphate Ester Flame Retardants: A Short Review; Springer: Berlin/Heidelberg, Germany, 2018; Volume 92. [Google Scholar] [CrossRef]
  47. Hu, W.; Gao, P.; Wang, L.; Hu, J. Endocrine Disrupting Toxicity of Aryl Organophosphate Esters and Mode of Action. Crit. Rev. Environ. Sci. Technol. 2023, 53, 1–18. [Google Scholar] [CrossRef]
  48. Steenland, K.; Fletcher, T.; Savitz, D.A. Epidemiologic Evidence on the Health Effects of Perfluorooctanoic Acid (PFOA). Environ. Health Perspect. 2010, 118, 1100–1108. [Google Scholar] [CrossRef]
  49. Wang, Y.; Lv, L.; Yu, Y.; Yang, G.; Xu, Z.; Wang, Q.; Cai, L. Single and Joint Toxic Effects of Five Selected Pesticides on the Early Life Stages of Zebrafish (Denio Rerio). Chemosphere 2017, 170, 61–67. [Google Scholar] [CrossRef]
  50. Wei, F.; Wang, D.; Li, H.; You, J. Joint Toxicity of Imidacloprid and Azoxystrobin to Chironomus Dilutus at Organism, Cell, and Gene Levels. Aquat. Toxicol. 2021, 233, 105783. [Google Scholar] [CrossRef]
  51. Martin, O.; Scholze, M.; Ermler, S.; McPhie, J.; Bopp, S.K.; Kienzler, A.; Parissis, N.; Kortenkamp, A. Ten Years of Research on Synergisms and Antagonisms in Chemical Mixtures: A Systematic Review and Quantitative Reappraisal of Mixture Studies. Environ. Int. 2021, 146, 106206. [Google Scholar] [CrossRef]
  52. Kertesz, V.; Van Berkel, G.J. Minimizing Analyte Electrolysis in an Electrospray Emitter. J. Mass Spectrom. 2001, 36, 204–210. [Google Scholar] [CrossRef]
  53. Hassan, I.; Pavlov, J.; Errabelli, R.; Attygalle, A.B. Oxidative Ionization Under Certain Negative-Ion Mass Spectrometric Conditions. J. Am. Soc. Mass Spectrom. 2017, 28, 270–277. [Google Scholar] [CrossRef] [PubMed]
  54. Caldwell, G.W.; Masucci, J.A.; Yan, Z.; Hageman, W. Allometric Scaling of Pharmacokinetic Parameters in Drug Discovery: Can Human CL, Vss and T1/2 Be Predicted from in-Vivo Rat Data? Eur. J. Drug Metab. Pharmacokinet. 2004, 29, 133–143. [Google Scholar] [CrossRef] [PubMed]
  55. Cao, X.; Gibbs, S.T.; Fang, L.; Miller, H.A.; Landowski, C.P.; Shin, H.C.; Lennernas, H.; Zhong, Y.; Amidon, G.L.; Yu, L.X.; et al. Why Is It Challenging to Predict Intestinal Drug Absorption and Oral Bioavailability in Human Using Rat Model. Pharm. Res. 2006, 23, 1675–1686. [Google Scholar] [CrossRef] [PubMed]
  56. Indorf, P.; Patzak, A.; Lichtenberger, F.B. Drug Metabolism in Animal Models and Humans: Translational Aspects and Chances for Individual Therapy. Acta Physiol. 2021, 233, e13734. [Google Scholar] [CrossRef]
  57. Martignoni, M.; Groothuis, G.M.M.; de Kanter, R. Species Differences between Mouse, Rat, Dog, Monkey and Human CYP-Mediated Drug Metabolism, Inhibition and Induction. Expert Opin. Drug Metab. Toxicol. 2006, 2, 875–894. [Google Scholar] [CrossRef]
  58. Tang, J.; Usmani, K.A.; Hodgson, E.; Rose, R.L. In Vitro Metabolism of Fipronil by Human and Rat Cytochrome P450 and Its Interactions with Testosterone and Diazepam. Chem. Biol. Interact. 2004, 147, 319–329. [Google Scholar] [CrossRef]
  59. McMahen, R.L.; Strynar, M.J.; Dagnino, S.; Herr, D.W.; Moser, V.C.; Garantziotis, S.; Andersen, E.M.; Freeborn, D.L.; McMillan, L.; Lindstrom, A.B. Identification of Fipronil Metabolites by Time-of-Flight Mass Spectrometry for Application in a Human Exposure Study. Environ. Int. 2015, 78, 16–23. [Google Scholar] [CrossRef]
  60. Mirnaghi, F.S.; Goryński, K.; Rodriguez-Lafuente, A.; Boyaci, E.; Bojko, B.; Pawliszyn, J. Microextraction versus Exhaustive Extraction Approaches for Simultaneous Analysis of Compounds in Wide Range of Polarity. J. Chromatogr. A 2013, 1316, 37–43. [Google Scholar] [CrossRef]
  61. Klimowska, A.; Wielgomas, B. Off-Line Microextraction by Packed Sorbent Combined with on Solid Support Derivatization and GC-MS: Application for the Analysis of Five Pyrethroid Metabolites in Urine Samples. Talanta 2018, 176, 165–171. [Google Scholar] [CrossRef]
  62. Klimowska, A.; Wynendaele, E.; Wielgomas, B. Quantification and Stability Assessment of Urinary Phenolic and Acidic Biomarkers of Non-Persistent Chemicals Using the SPE-GC/MS/MS Method. Anal. Bioanal. Chem. 2023, 415, 2227–2238. [Google Scholar] [CrossRef]
  63. Michlig, N.; Lehotay, S.J.; Lightfield, A.R. Comparison of Filter Membranes in the Analysis of 183 Veterinary and Other Drugs by Liquid Chromatography-Tandem Mass Spectrometry. J. Sep. Sci. 2024, 47, 2300696. [Google Scholar] [CrossRef]
  64. Hebig, K.H.; Nödler, K.; Licha, T.; Scheytt, T.J. Impact of Materials Used in Lab and Field Experiments on the Recovery of Organic Micropollutants. Sci. Total Environ. 2014, 473–474, 125–131. [Google Scholar] [CrossRef] [PubMed]
  65. VanMiddlesworth, B.J.; Dorsey, J.G. Quantifying Injection Solvent Effects in Reversed-Phase Liquid Chromatography. J. Chromatogr. A 2012, 1236, 77–89. [Google Scholar] [CrossRef] [PubMed]
  66. Waters. Beginner’s Guide to SPE: Solid Phase Extraction; Waters Corporation: Milford, MA, USA, 2014; pp. 1–212. [Google Scholar]
  67. Agilent. Bond Elut Plexa SPE Method Guide; 04034-0712; Agilent Technologies: Melbourne, VIC, Australia, 2012; pp. 1–2. [Google Scholar]
  68. Lewis, K.A.; Tzilivakis, J.; Warner, D.J.; Green, A. An International Database for Pesticide Risk Assessments and Management. Hum. Ecol. Risk Assess. 2016, 22, 1050–1064. [Google Scholar] [CrossRef]
  69. EMA. ICH Guideline M10 on Bioanalytical Method Validation; EMA/CHMP/ICH/172948/2019; EMA: Amsterdam, The Netherlands, 2022. [Google Scholar]
  70. EMA. Guideline on Bioanalytical Method Validation; EMEA/CHMP/EWP/192217/2009 Rev. 1 Corr. 2; EMA: London, UK, 2011. [Google Scholar]
  71. Panuwet, P.; Hunter, R.E.; D’Souza, P.E.; Chen, X.; Radford, S.A.; Cohen, J.R.; Marder, M.E.; Kartavenka, K.; Ryan, P.B.; Barr, D.B. Biological Matrix Effects in Quantitative Tandem Mass Spectrometry-Based Analytical Methods: Advancing Biomonitoring. Crit. Rev. Anal. Chem. 2016, 46, 93–105. [Google Scholar] [CrossRef]
  72. Ferrer Amate, C.; Unterluggauer, H.; Fischer, R.J.; Fernández-Alba, A.R.; Masselter, S. Development and Validation of a LC-MS/MS Method for the Simultaneous Determination of Aflatoxins, Dyes and Pesticides in Spices. Anal. Bioanal. Chem. 2010, 397, 93–107. [Google Scholar] [CrossRef]
  73. Raposo, F.; Barceló, D. Challenges and Strategies of Matrix Effects Using Chromatography-Mass Spectrometry: An Overview from Research versus Regulatory Viewpoints. TrAC—Trends Anal. Chem. 2021, 134, 116068. [Google Scholar] [CrossRef]
  74. Gys, C.; Ait Bamai, Y.; Araki, A.; Bastiaensen, M.; Caballero-Casero, N.; Kishi, R.; Covaci, A. Biomonitoring and Temporal Trends of Bisphenols Exposure in Japanese School Children. Environ. Res. 2020, 191, 110172. [Google Scholar] [CrossRef]
  75. Rodríguez-Zamora, M.G.; Fuhrimann, S.; Winkler, M.S.; Rosa, M.J.; Reich, B.; Lindh, C.; Mora, A.M. Respiratory and Allergic Outcomes among Farmworkers Exposed to Pesticides in Costa Rica. Sci. Total Environ. 2024, 954, 176776. [Google Scholar] [CrossRef]
  76. Gao, B.; Poma, G.; Malarvannan, G.; Dumitrascu, C.; Bastiaensen, M.; Wang, M.; Covaci, A. Development of an Analytical Method Based on Solid-Phase Extraction and LC-MS/MS for the Monitoring of Current-Use Pesticides and Their Metabolites in Human Urine. J. Environ. Sci. 2022, 111, 153–163. [Google Scholar] [CrossRef] [PubMed]
  77. Araujo, P. Key Aspects of Analytical Method Validation and Linearity Evaluation. J. Chromatogr. B Anal. Technol. Biomed. Life Sci. 2009, 877, 2224–2234. [Google Scholar] [CrossRef] [PubMed]
  78. Wrobel, S.A.; Bury, D.; Belov, V.N.; Klenk, J.M.; Hauer, B.; Hayen, H.; Martino-Andrade, A.J.; Koch, H.M.; Brüning, T.; Käfferlein, H.U. Rapid Quantification of Seven Major Neonicotinoids and Neonicotinoid-like Compounds and Their Key Metabolites in Human Urine. Anal. Chim. Acta 2023, 1239, 340680. [Google Scholar] [CrossRef] [PubMed]
  79. Baker, S.E.; Serafim, A.B.; Morales-Agudelo, P.; Vidal, M.; Calafat, A.M.; Ospina, M. Quantification of DEET and Neonicotinoid Pesticide Biomarkers in Human Urine by Online Solid-Phase Extraction High-Performance Liquid Chromatography-Tandem Mass Spectrometry. Anal. Bioanal. Chem. 2019, 411, 669–678. [Google Scholar] [CrossRef]
  80. Wrobel, S.A.; Koslitz, S.; Bury, D.; Hayen, H.; Koch, H.M.; Brüning, T.; Käfferlein, H.U. Human Biomonitoring of Neonicotinoid Exposures: Case Studies after the Use of a Spray-Agent to Ornamental Plants and a Topical Medication to Pets. Front. Public Health 2023, 11, 1321138. [Google Scholar] [CrossRef]
  81. Garí, M.; Bury, D.; Moos, R.K.; Wolniewicz, M.; Jankowska, A.; Brzozowska, A.; Jerzynska, J.; Bose-O’Reilly, S.; Koch, H.M.; Polanska, K. Urinary Concentrations of BPA and Analogous Bisphenols (BPF and BPS) among School Children from Poland: Exposure and Risk Assessment in the REPRO_PL Cohort. Expo. Health 2025, 17, 191–200. [Google Scholar] [CrossRef]
  82. Norén, E.; Lindh, C.; Bengtsson, M.; Rönnholm, A. Urin-Och Serumhalter Av Organiska Miljöföroreningar Hos Unga Vuxna i Skåne År 2000–2022. Resultat Från Den Sjätte Delstudien 2022; Naturvårdsverkets: Lund, Sweden, 2024; Available online: https://www.naturvardsverket.se/verktyg-och-tjanster/data-databaser-och-sokregister/naturvardsverkets-oppna-rapportarkiv---diva/ (accessed on 30 July 2025).
  83. Van den Eede, N.; Neels, H.; Jorens, P.G.; Covaci, A. Analysis of Organophosphate Flame Retardant Diester Metabolites in Human Urine by Liquid Chromatography Electrospray Ionisation Tandem Mass Spectrometry. J. Chromatogr. A 2013, 1303, 48–53. [Google Scholar] [CrossRef]
  84. Fišerová, P.S.; Kohoutek, J.; Degrendele, C.; Dalvie, M.A.; Klánová, J. New Sample Preparation Method to Analyse 15 Specific and Non-Specific Pesticide Metabolites in Human Urine Using LC-MS/MS. J. Chromatogr. B Anal. Technol. Biomed. Life Sci. 2021, 1166, 122542. [Google Scholar] [CrossRef]
  85. Nishihara, N.; Isobe, T.; Takagi, M.; Tajima, T.; Kitahara, Y.; Hayashi, M.; Saito, I.; Watanabe, S.; Iwai-Shimada, M.; Ueyama, J. Determination of Skin-Insect Repellent Icaridin and DEET in Human Urine Using Solid-Phase Extraction and Liquid Chromatography with Tandem Mass Spectrometry and Its Application to a Sample of Japanese Adults. Environ. Health Prev. Med. 2025, 30, 18. [Google Scholar] [CrossRef]
  86. Šulc, L.; Janoš, T.; Figueiredo, D.; Ottenbros, I.; Šenk, P.; Mikeš, O.; Huss, A.; Čupr, P. Pesticide Exposure among Czech Adults and Children from the CELSPAC-SPECIMEn Cohort: Urinary Biomarker Levels and Associated Health Risks. Environ. Res. 2022, 214, 114002. [Google Scholar] [CrossRef]
  87. Davis, M.D.; Wade, E.L.; Restrepo, P.R.; Roman-Esteva, W.; Bravo, R.; Kuklenyik, P.; Calafat, A.M. Semi-Automated Solid Phase Extraction Method for the Mass Spectrometric Quantification of 12 Specific Metabolites of Organophosphorus Pesticides, Synthetic Pyrethroids, and Select Herbicides in Human Urine. J. Chromatogr. B 2013, 929, 18–26. [Google Scholar] [CrossRef]
  88. Le Grand, R.; Dulaurent, S.; Gaulier, J.M.; Saint-Marcoux, F.; Moesch, C.; Lachâtre, G. Simultaneous Determination of Five Synthetic Pyrethroid Metabolites in Urine by Liquid Chromatography-Tandem Mass Spectrometry: Application to 39 Persons without Known Exposure to Pyrethroids. Toxicol. Lett. 2012, 210, 248–253. [Google Scholar] [CrossRef]
  89. Shi, L.; Wan, Y.; Liu, J.; He, Z.; Xu, S.; Xia, W. Insecticide Fipronil and Its Transformation Products in Human Blood and Urine: Assessment of Human Exposure in General Population of China. Sci. Total Environ. 2021, 786, 147342. [Google Scholar] [CrossRef] [PubMed]
  90. Kruve, A.; Rebane, R.; Kipper, K.; Oldekop, M.L.; Evard, H.; Herodes, K.; Ravio, P.; Leito, I. Tutorial Review on Validation of Liquid Chromatography-Mass Spectrometry Methods: Part I. Anal. Chim. Acta 2015, 870, 29–44. [Google Scholar] [CrossRef]
  91. Evard, H.; Kruve, A.; Leito, I. Tutorial on Estimating the Limit of Detection Using LC-MS Analysis, Part I: Theoretical Review. Anal. Chim. Acta 2016, 942, 23–39. [Google Scholar] [CrossRef] [PubMed]
  92. Ye, X.; Bishop, A.M.; Reidy, J.A.; Needham, L.L.; Calafat, A.M. Temporal Stability of the Conjugated Species of Bisphenol A, Parabens, and Other Environmental Phenols in Human Urine. J. Expo. Sci. Environ. Epidemiol. 2007, 17, 567–572. [Google Scholar] [CrossRef] [PubMed]
  93. Kruve, A.; Rebane, R.; Kipper, K.; Oldekop, M.L.; Evard, H.; Herodes, K.; Ravio, P.; Leito, I. Tutorial Review on Validation of Liquid Chromatography-Mass Spectrometry Methods: Part II. Anal. Chim. Acta 2015, 870, 8–28. [Google Scholar] [CrossRef]
  94. Li, W.; Zhang, J.; Tse, F.L.S. Strategies in Quantitative LC-MS/MS Analysis of Unstable Small Molecules in Biological Matrices. Biomed. Chromatogr. 2011, 25, 258–277. [Google Scholar] [CrossRef]
  95. ECHA. BPA Being Replaced by BPS in Thermal Paper, ECHA Survey Finds. Available online: https://echa.europa.eu/pl/-/bpa-being-replaced-by-bps-in-thermal-paper-echa-survey-finds (accessed on 25 August 2025).
  96. Frederiksen, H.; Nielsen, O.; Koch, H.M.; Skakkebaek, N.E.; Juul, A.; Jørgensen, N.; Andersson, A.M. Changes in Urinary Excretion of Phthalates, Phthalate Substitutes, Bisphenols and Other Polychlorinated and Phenolic Substances in Young Danish Men; 2009–2017. Int. J. Hyg. Environ. Health 2020, 223, 93–105. [Google Scholar] [CrossRef]
  97. Ospina, M.; Jayatilaka, N.K.; Wong, L.Y.; Restrepo, P.; Calafat, A.M. Exposure to Organophosphate Flame Retardant Chemicals in the U.S. General Population: Data from the 2013–2014 National Health and Nutrition Examination Survey. Environ. Int. 2018, 110, 32–41. [Google Scholar] [CrossRef]
  98. Jayatilaka, N.K.; Restrepo, P.; Williams, L.T.; Ospina, M.; Valentin-Blasini, L.; Calafat, A.M. Quantification of Three Chlorinated Dialkyl Phosphates, Diphenyl Phosphate, 2,3,4,5-Tetrabromobenzoic Acid, and Four Other Organophosphates in Human Urine by Solid Phase Extraction-High Performance Liquid Chromatography-Tandem Mass Spectrometry. Anal. Bioanal. Chem. 2017, 409, 1323–1332. [Google Scholar] [CrossRef]
  99. Faÿs, F.; Palazzi, P.; Hardy, E.M.; Schaeffer, C.; Phillipat, C.; Zeimet, E.; Vaillant, M.; Beausoleil, C.; Rousselle, C.; Slama, R.; et al. Is There an Optimal Sampling Time and Number of Samples for Assessing Exposure to Fast Elimination Endocrine Disruptors with Urinary Biomarkers? Sci. Total Environ. 2020, 747, 141185. [Google Scholar] [CrossRef]
  100. Dyk, M.B.; Liu, Y.; Chen, Z.; Vega, H.; Krieger, R.I. Fate and Distribution of Fipronil on Companion Animals and in Their Indoor Residences Following Spot-on Flea Treatments. J. Environ. Sci. Health—Part B Pestic. Food Contam. Agric. Wastes 2012, 47, 913–924. [Google Scholar] [CrossRef]
  101. Cone, E.J.; Caplan, Y.H.; Moser, F.; Robert, T.; Shelby, M.K.; Black, D.L. Normalization of Urinary Drug Concentrations with Specific Gravity and Creatinine. J. Anal. Toxicol. 2009, 33, 1–7. [Google Scholar] [CrossRef]
Figure 1. Total ion chromatogram of the FIP-hydroxy standard in negative ionization. The peak that was subjected to further investigation is marked by an “X”. Retention times for FIP-dtfms and FIP are indicated by gray arrows.
Figure 1. Total ion chromatogram of the FIP-hydroxy standard in negative ionization. The peak that was subjected to further investigation is marked by an “X”. Retention times for FIP-dtfms and FIP are indicated by gray arrows.
Ijms 26 09025 g001
Figure 2. Spectrum of peak “X” (left) and theoretical spectrum expected for FIP-hydroxy (right).
Figure 2. Spectrum of peak “X” (left) and theoretical spectrum expected for FIP-hydroxy (right).
Ijms 26 09025 g002
Figure 3. Analyte loss due to filtration. The solid lines represent the mean values. A dotted line at 100% recovery was added for reference. An asterisk denotes a statistically significant difference (p < 0.05).
Figure 3. Analyte loss due to filtration. The solid lines represent the mean values. A dotted line at 100% recovery was added for reference. An asterisk denotes a statistically significant difference (p < 0.05).
Ijms 26 09025 g003
Figure 4. Recovery comparison for compounds of interest obtained using Bond Elut Plexa 30 mg and Oasis HLB 60 mg. The data points represent the mean, whereas the whiskers show the range. A dotted line at 100% efficiency was added for reference. 1 Due to the presence of unlabeled BPS in the matrix, a deuterated analog was used to assess the recovery.
Figure 4. Recovery comparison for compounds of interest obtained using Bond Elut Plexa 30 mg and Oasis HLB 60 mg. The data points represent the mean, whereas the whiskers show the range. A dotted line at 100% efficiency was added for reference. 1 Due to the presence of unlabeled BPS in the matrix, a deuterated analog was used to assess the recovery.
Ijms 26 09025 g004
Figure 5. Workflow of the final sample preparation procedure for urine samples.
Figure 5. Workflow of the final sample preparation procedure for urine samples.
Ijms 26 09025 g005
Table 1. Matrix factors (%) for analytes included in the method for urinalysis.
Table 1. Matrix factors (%) for analytes included in the method for urinalysis.
AnalyteNominal Concentration (pg/mL)Specific Gravity of Urine LotsMean
1.0061.0101.0111.0171.0261.0281.031
Matrix Factors (%)
IMI-OH40,0005052515010444944
IMI80004332404021375138
BPS20001712131115101613
BOS-OH20,00053221132
4OH3PBA40,000166873447
DPhP20008972616425344255
DEET4003820242322211824
FIP-dtfms200038191815910717
FIP-amide4001911141056510
TEB-OH20008158564842453452
4F3PBA400014101112691010
FIP-desulfinyl806740343630322137
FIP807148524834392646
FIP-sulfide80533737365271230
FIP-sulfone806131413528282035
CPhCA40,0005531393328292735
Table 2. IS-corrected matrix factors (%) and their respective CVs for analytes quantitated in urine. The analyte concentrations are provided in Table 1.
Table 2. IS-corrected matrix factors (%) and their respective CVs for analytes quantitated in urine. The analyte concentrations are provided in Table 1.
AnalyteInternal StandardSpecific Gravity of Urine LotsMeanCV
1.0061.0101.0111.0171.0261.0281.031
IS-Corrected Matrix Factors (%)
IMI-OHIMI-D4110146114117381068510230
IMIIMI-D493919095788890896
BPSBPS-D8857474607961747212
BOS-OHBPS-D82416131168141341
4OH3PBAFIP-dtfms-13C215N2483644474244604615
DPhPFIP-13C4118140106125698015011225
DEETFIP-13C4513842466048645018
FIP-dtfmsFIP-dtfms-13C215N21141101051101111061161103
FIP-amideFIP-dtfms-13C215N2576280726569786911
TEB-OHFIP-13C410811296931131041231079
4F3PBA3PBA-13C6527965717466897115
FIP-desulfinylFIP-13C4897858708273767512
FIPFIP-13C495939095939195932
FIP-sulfideFIP-13C4717264711562415735
FIP-sulfoneFIP-13C480607068766672709
CPhCAFIP-13C4736068657668967215
Table 3. Internal standard selection, linearity data and sensitivity obtained for analytes during method validation.
Table 3. Internal standard selection, linearity data and sensitivity obtained for analytes during method validation.
AnalyteISLLOQ (pg/mL)Linear Range 1 (pg/mL)Regression
Equation
Curve FitCurve WeightingCoefficient of Determination (R2)
IMI-OHIMI-D410001000–200,0000.0853x + 0.0065Linear1/x0.9972
IMIIMI-D4100100–20,000 (160,000)1.5555x + 0.0017Linear1/x0.9990
BPSBPS-D85050–5000 (40,000)0.2526x + 0.0056Linear1/x0.9979
BOS-OHBPS-D810001000–200,0000.0138x + 0.0020Linear1/x0.9969
4OH3PBAFIP-dtfms-13C215N220002000–200,0000.0185x − 0.0018Linear1/x0.9907
DPhPFIP-13C4100100–50000.0668x + 0.0238Linear1/x0.9935
DEETFIP-13C41010–20000.8049x + 0.0604Linear1/x0.9991
FIP-dtfmsFIP-dtfms-13C215N2200200–40,000 (320,000)0.8586x − 0.0021Linear1/x0.9995
FIP-amideFIP-dtfms-13C215N25050–50000.2650x − 0.0004Linear1/x0.9948
TEB-OHFIP-13C4100100–20,0000.0499x + 0.0096Linear1/x0.9973
4F3PBA3PBA-13C6100100–10,0003.8030x + 0.0050Linear1/x20.9921
FIP-desulfinylFIP-13C455–1000 (8000)0.9155x + 0.0027Linear1/x0.9988
FIPFIP-13C411–200 (1600)1.2291x + 0.0013Linear1/x0.9996
FIP-sulfideFIP-13C411–200 (1600)0.8579x + 0.0007Linear1/x0.9992
FIP-sulfoneFIP-13C40.50.5–100 (800)1.3612x + 0.0053Linear1/x0.9995
CPhCAFIP-13C4500500–100,000 (800,000)0.0012x + 0.0015Linear1/x0.9993
1 The values in brackets are quantifiable after eightfold dilution (see Section 2.3.6.).
Table 4. Accuracy and precision investigation for analytes in urine at LQC and HQC level.
Table 4. Accuracy and precision investigation for analytes in urine at LQC and HQC level.
AnalyteLQCHQC
Nominal Concentration (pg/mL)Accuracy (%)Precision (CV, %)Nominal Concentration (pg/mL)Accuracy (%)Precision (CV, %)
Intra-Run (n = 5)Inter-Run (n = 15)Intra-Run (n = 5)Inter-Run (n = 15)Intra-Run (n = 5)Inter-Run (n = 15)Intra-Run (n = 5)Inter-Run (n = 15)
IMI-OH300010286132050,000969588
IMI300991006850009710149
BPS75978914141250110102710
BOS-OH300010010091350,00010196712
4OH3PBA300010197121450,000888657
DPhP15091101816250010086714
DEET3010511315145009710434
FIP-dtfms600100989610,00010210233
FIP-amide15095887132500969576
TEB-OH300104100810500010310346
4F3PBA150959311132500979946
FIP-desulfinyl1599976825010010344
FIP3979371250999747
FIP-sulfide3102961014509599711
FIP-sulfone1.51038810132597100714
CPhCA15009399101225,0009910134
Table 5. Results of the stability study for the compounds of interest. The 24 h results refer to the stability of processed samples at room temperature, whereas the 30-day and 12-month values describe the stability of unprocessed samples stored at −20 °C.
Table 5. Results of the stability study for the compounds of interest. The 24 h results refer to the stability of processed samples at room temperature, whereas the 30-day and 12-month values describe the stability of unprocessed samples stored at −20 °C.
AnalyteStudy PeriodLQCHQC
Nominal Concentration (pg/mL)Accuracy
(n = 4, %)
Precision
(n = 4, CV, %)
Nominal Concentration (pg/mL)Accuracy
(n = 4, %)
Precision
(n = 4, CV, %)
24 h 8813 968
IMI-OH30-day3000861550,000956
12-month 9212 904
24 h 997 1039
IMI30-day300991450001005
12-month 10814 1033
24 h 9010 1037
BPS30-day75115412501049
12-month 11314 1086
24 h 1016 853
BOS-OH30-day300080850,000891
12-month 4611 4611
24 h 887 894
4OH3PBA30-day3000107450,000797
12-month 7710 582
24 h 1058 864
DPhP30-day150109132500879
12-month 9714 11410
24 h 1135 1102
DEET30-day3011565001086
12-month 976 985
24 h 9411 1052
FIP-dtfms30-day60097110,0001053
12-month 981 992
24 h 896 916
FIP-amide30-day150105112500936
12-month 1028 993
24 h 9611 1002
TEB-OH30-day300976500010110
12-month 10210 992
24 h 9115 1006
4F3PBA30-day15099142500916
12-month 11513 924
24 h 9411 1064
FIP-desulfinyl30-day159332501022
12-month 9814 903
24 h 905 901
FIP30-day389150904
12-month 960 891
24 h 9113 920
FIP-sulfide30-day3901050924
12-month 9512 855
24 h 8912 894
FIP-sulfone30-day1.586925895
12-month 853 8610
24 h 972 1051
CPhCA30-day150095325,000994
12-month 9612 1082
Table 6. Results of the applicability study (n = 28).
Table 6. Results of the applicability study (n = 28).
Analyte% ≥LLOQ 1Unadjusted (pg/mL)Specific Gravity-Adjusted (pg/mL)
GMP25P50P75P95MaxGMP25P50P75P95Max
IMI-OH0------------
IMI4-----101-----122
BPS894122776637912888394748425753784024372978
BOS-OH501238-538 2273211,95014,1901454-561372613,37413,384
4OH3PBA18----981314,008----881113,350
DPhP93154195217123531>ULOQ 3>ULOQ181093016563692>ULOQ>ULOQ
DEET6820.2-15.478.521221423.7-17.7101190216
FIP-dtfms0------------
FIP-amide0------------
TEB-OH862811342597414565690533010825458030124465
4F3PBA4-----416-----290
FIP-desulfinyl0------------
FIP713.94-3.329.981291504.63-3.3924.9233306
FIP-sulfide0------------
FIP-sulfone39---0.85816.717.7---1.0414.516.9
CPhCA0------------
1 LLOQ, GM, P25-95 and Max stand for lower limit of quantification, geometric mean, 25–95th percentile, and maximum, respectively. 2 Since the detection rate of BOS-OH was exactly 50%, this value is an arithmetic mean of zero and the first non-zero result. 3 ULOQ, upper limit of quantification.
Table 7. Compound-specific parameters of analytes and internal standards used for LC-MS/MS analysis.
Table 7. Compound-specific parameters of analytes and internal standards used for LC-MS/MS analysis.
CompoundStatusRetention Time (min)Precursor IonPrecursor m/zCapillary Voltage (V)Product Ions m/z 1Collision Energy (V) 2
IMI-OHanalyte4.90[M+H]+
[M−H]
272.0
270.0
70
−70
225.0, 228.0
46.1
14, 10
10
IMZ-OHanalyte5.18[M+H]+257.09069.0, 125.0, 136.019, 33, 42
IMIanalyte5.33[M+H]+256.070209.0, 175.2, 212.212, 14, 11
BPSanalyte6.11[M−H]249.1−110155.4, 91.9, 107.921, 33, 27
BOS-OHanalyte8.31[M−H]
[M+H]+
357.0
359.0
−100
100
243.7
140.0, 323.0
18
21, 20
4OH3PBAanalyte8.51[M−H]229.3−70108.9, 109.8, 185.621, 20.5, 12
DPhPanalyte8.52[M−H]249.0−9292.7, 154.626, 19.5
DEETanalyte9.09[M+H]+192.088119.0, 91.014, 26
FIP-dtfmsanalyte9.14[M−H]319.0−70282.9, 262.88, 20
FIP-amideanalyte9.26[M−H]453.0−70347.8, 271.9, 303.815, 41, 25
TEB-OHanalyte10.04[M+H]+324.09070.0, 125.0, 179.120, 37, 19
4F3PBAanalyte10.39[M−H]230.9−7592.7, 186.625, 10
FIP-desulfinylanalyte10.77[M−H]387.0−50350.9, 281.9, 330.812, 30, 28
FIPanalyte10.96[M−H]435.0−70329.7, 249.6, 277.615, 26, 27
FIP-sulfideanalyte11.12[M−H]419.0−70261.8, 313.9, 382.926, 18, 11
FIP-sulfoneanalyte11.34[M−H]451.0−70281.9, 243.8, 414.925, 44, 17
PFOAanalyte11.79[M−H]
[M−CO2−H]
413.0
369.0
−30
−72
369.0, 169.0
219.0
7, 18.5
10.5
CPhCAanalyte12.11[M−H]283.0−90246.6, 35.27, 9
IMI-D4internal standard5.30[M+H]+260.070213.0, 214.0, 216.012, 8, 10
BPS-D8internal standard6.05[M−H]257.0−110112.0, 96.0, 160.027, 33, 21
FIP-dtfms-13C215N2internal standard9.13[M−H]323.0−70287.0, 184.98, 28
3PBA-13C4internal standard10.40[M−H]219.0−9099.0, 175.019, 11
FIP-13C4internal standard10.97[M−H]439.0−70334.0, 250.9, 321.915, 26, 24
1 First ion is the quantifier, the others are the qualifiers (in increasing m/z order). 2 For product ions, respectively.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Rodzaj, W.; Wacławik, M.; Jurewicz, J.; Wielgomas, B. A Sensitive SPE-LC-MS/MS Method for Determination of Selected Veterinary Drugs and Other Organic Contaminants in Human Urine: Development, Validation, and Application Study. Int. J. Mol. Sci. 2025, 26, 9025. https://doi.org/10.3390/ijms26189025

AMA Style

Rodzaj W, Wacławik M, Jurewicz J, Wielgomas B. A Sensitive SPE-LC-MS/MS Method for Determination of Selected Veterinary Drugs and Other Organic Contaminants in Human Urine: Development, Validation, and Application Study. International Journal of Molecular Sciences. 2025; 26(18):9025. https://doi.org/10.3390/ijms26189025

Chicago/Turabian Style

Rodzaj, Wojciech, Małgorzata Wacławik, Joanna Jurewicz, and Bartosz Wielgomas. 2025. "A Sensitive SPE-LC-MS/MS Method for Determination of Selected Veterinary Drugs and Other Organic Contaminants in Human Urine: Development, Validation, and Application Study" International Journal of Molecular Sciences 26, no. 18: 9025. https://doi.org/10.3390/ijms26189025

APA Style

Rodzaj, W., Wacławik, M., Jurewicz, J., & Wielgomas, B. (2025). A Sensitive SPE-LC-MS/MS Method for Determination of Selected Veterinary Drugs and Other Organic Contaminants in Human Urine: Development, Validation, and Application Study. International Journal of Molecular Sciences, 26(18), 9025. https://doi.org/10.3390/ijms26189025

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop