Next Article in Journal
Anaerobic Digested Wastewater CO2 Sequestration Using a Biophotocatalytic System with a Magnetized Photocatalyst (Fe-TiO2)
Next Article in Special Issue
Contribution of Reliable Chromatographic Data in QSAR for Modelling Bisphenol Transport across the Human Placenta Barrier
Previous Article in Journal
Recent Advances of Anderson-Type Polyoxometalates as Catalysts Largely for Oxidative Transformations of Organic Molecules
Previous Article in Special Issue
An Alternative Strategy for Screening and Confirmation of 330 Pesticides in Ground- and Surface Water Using Liquid Chromatography Tandem Mass Spectrometry
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Characterizing Powdered Activated Carbon Treatment of Surface Water Samples Using Polarity-Extended Non-Target Screening Analysis

1
Urban Water Systems Engineering, Technical University of Munich, 85748 Garching, Germany
2
Analytisches Forschungsinstitut für Non-Target Screening (AFIN-TS) GmbH, 86167 Augsburg, Germany
*
Author to whom correspondence should be addressed.
Molecules 2022, 27(16), 5214; https://doi.org/10.3390/molecules27165214
Submission received: 21 July 2022 / Revised: 8 August 2022 / Accepted: 11 August 2022 / Published: 16 August 2022
(This article belongs to the Special Issue Chemometrics for Environmental Analysis)

Abstract

:
Advanced wastewater treatment such as powdered activated carbon (PAC) reduces the load of organic micropollutants entering the aquatic environment. Since mobile and persistent compounds accumulate in water cycles, treatment strategies need to be evaluated for the removal of (very) polar compounds. Thereby, non-targeted analysis gives a global picture of the molecular fingerprint (including these very polar molecules) of water samples. Target and non-target screening were conducted using polarity-extended chromatography hyphenated with mass spectrometry. Samples treated with different types and concentrations of PAC were compared to untreated samples. Molecular features were extracted from the analytical data to determine fold changes, perform a principal component analysis and for significance testing. The results suggest that a part of the polar target analytes was adsorbed but also some byproducts might be formed or desorbed from the PAC.

1. Introduction

Chemicals of emerging concern (CECs), i.e., pharmaceuticals, chemicals of daily usage and others, enter the aquatic environment frequently through municipal or industrial wastewater and have already been detected in drinking water at ng L−1 levels [1,2]. In urbanized areas in Germany, the contributions of wastewater effluents to streams can in some cases be >50% in the summer months [3]. The de facto reuse of wastewater could increase the risk of introducing CECs into drinking water sources. Conventional water treatment processes have been reported to ineffectively remove certain CECs, such as hydrophilic compounds [1]. Measures that aim at further reducing the discharge of CECs include upgrading wastewater treatment plants (WWTPs) with advanced treatments such as ozonation or powdered activated carbon (PAC), a strategy currently pursued by Switzerland [4,5].
PAC is commonly used for drinking water purification due to its adsorptive qualities: in batch experiments, Hernández-Leal et al. treated ultrapure water spiked with CECs of personal care products (20–1600 µg L−1) with PAC at a dose of 1.25 g L−1 and found the removal efficiency to be >94% for all compounds after 5 min of contact time [6]. The log KOW values of the investigated CECs ranged from −0.16 to 6.90 [7]. Kovalova et al. showed that 23 mg L−1 of PAC (hydraulic residence time of one day) removed 62% of the load of 56 CECs including pharmaceuticals, metabolites and industrial chemicals from hospital wastewater pre-treated by a membrane bioreactor. In their investigations, all compounds with a log D > 2 (pH 9) were eliminated completely or fell below the limit of quantification [8]. Adsorption capacities of polar CECs are expected to be lower than those of non-polar or mid-polar ones, in some cases up to one order of magnitude [9]. However, depending on the pH, some charged or zwitterionic polar compounds exhibit strong sorption, probably due to electrostatic interactions [8].
The occurrence and fate of CECs throughout treatment processes are typically assessed via targeted analyses. Focusing on a limited number of prioritized or probably already regulated target analytes could cause a partially distorted view of the molecular fingerprint of water samples as unknown or unexpected CECs are missed. At this point, it cannot be determined whether a decrease in concentration of a target analyte indicates its removal or merely its transformation. This leads to transformation products (TPs) being underestimated, even though they tend to be more polar (and thus mobile) and might even have enhanced toxicity potential compared to their parent compounds [10,11,12,13]. The non-target screening (NTS) approach, powered by high-resolution mass spectrometry (HRMS) often in combination with liquid chromatography (LC), allows a more comprehensive assessment of CECs as it refrains from preselecting substances [14,15]. When evaluating NTS data, features characterized by their retention time (RT), mass and signal intensity are extracted from full scan HRMS data and further processed following two general objectives:
  • Identification with the methodological consequence of prioritizing features based on, i.e., the amount and quality of information available and their environmental relevance [16,17,18].
  • Bulk characterization by a non-discriminatory feature extraction workflow followed by statistical analysis [19].
The latter is a suitable method to globally evaluate water treatment processes, often based on features’ signal intensities, that gives indications of the removal of CECs and formation of TPs [20,21,22].
In this study, the removal efficiency of PAC for CECs from water samples is investigated by NTS. Therefore, three different types of PAC at three different concentrations are assessed in batch experiments using the non-target screening approach. The polarity range of the analytical method is extended by coupling reversed-phase (RP) LC to hydrophilic interaction liquid chromatography (HILIC) in series and screening a mass window for small molecules using HRMS equipped with electrospray ionization (ESI). It should be noted that this study does not provide adsorption kinetics but rather presents a non-target screening method that enables a comprehensive comparison of untreated and treated samples, explicitly considering (very) polar compounds.

2. Results and Discussion

In the following, the results of the target evaluation are presented considering the internal standards and the polar standard compounds. Moreover, features extracted before and after PAC treatment are globally assessed with the objective of getting indications on removal and/or formation of CECs. Therefore, fold changes (fc) and multiple hypothesis tests were interpreted. Variabilities throughout the measurement sequence were evaluated based on the feature lists of all samples including pooled QC samples.

2.1. Targeted Evaluation

A targeted search for the internal standards (n = 10, added prior to sample analysis) and the polar standard compounds (n = 10, added prior to PAC treatment) was performed on the ESI(+) data. The RT and mass precision as well as the mass accuracy of the RPLC-HILIC-HRMS system was evaluated on the internal standards. In the positive ionization mode, the precision was better than 1.5 % for RT (n = 35 injections) and better than 1.5 ppm for mass. The mass accuracy was <2.1 ppm for all compounds. One measurement (absorbent H118 at 7 mg L−1, third replicate) appeared to be an outlier in the three parameters mass, RT and signal intensity, and was therefore discarded. In negative ionization mode, precision values for RT and mass were <0.8 % and 1.5 ppm, respectively. The masses of the internal standards were detected more accurately than 2.0 ppm (n = 36 injections). Based on these findings, the instrumental setup performed reproducibly and accurately.
Furthermore, the consistency interval for the fold changes of the signal intensities was validated on the internal standards. It was defined as −1.00 ≤ log2(fc) ≤ 1.00, which corresponds to 0.50 ≤ fc ≤ 2.00 [22]. Since the internal standards were spiked into the samples after treatment and prior to analysis, the log2(fc) was expected to fall into the consistency interval and be close to 0. In all of the sample comparisons in the positive ionization mode, all internal standards fell into the consistency interval with at least ±9 standard deviations (see Table 1). The same was observed in the negative ionization mode with at least ±8 standard deviations (see Table S1). The polar standard compounds were spiked into the samples prior to PAC treatment. In the positive ionization mode, their mean log2(fc) decreased with increasing PAC load and the variability of fold changes increased (Table 1). It can therefore be assumed that some of the standard compounds absorbed onto the activated carbon. Figure 1 represents the log2(fc)-RT plots of the PAC type H121 at different concentrations (2, 7 and 30 mg L−1). The internal standards (orange circles) all fall into the consistency interval (dashed lines). The (very) polar standard compounds (blue diamonds) eluted earlier than 15.6 min have log D values ≤ 0.30 and are thus expected to be primarily retained by the HILIC column. At 2 mg L−1 of PAC H121, they did not show any increase or decrease. At 7 mg L−1, a decrease in signal intensity (log2(fc) < −1) was observed for famotidine and 2,4-diamino-6-(hydroxymethyl)pteridine, complemented by 2-aminopyridine and 3-pyridinemethanol at 30 mg L−1. The respective plots for PAC H118 and H120 are given in Figure S1. Only miglitol and acamprosate were detected in all measurements conducted in the negative ionization mode and their fcs did not suggest any removal. Other polar standard compounds were partially filtered out during data processing.

2.2. Non-Targeted Evaluation

When processing NTS data for process evaluation (Section 3), rather than filtering for features that could readily be matched with real compounds, it is important to curate the data to be statistically reliable. Thus, the data need to be cleaned from noise, artefacts and redundant peaks.
Regions of increased data density were observed in the total ion chromatogram (Figure S2), i.e., approximately RT 25.6 min, where non-target peaks accumulate. In these regions the probability for the algorithm to erroneously extract artefacts might be increased. As a result, peaks were filtered by shape and number of data points before further processing. However, parameter ranges had to be set relatively wide to avoid false negative features. As a result, the bands of high feature density persist even after completing the full data processing regime (Figure 2). This matter is arguably better resolved by adapting the chromatographic separation rather than on a data processing level.
Chromatographic peaks which were identified as isotopologues or adducts of features were removed in order to reduce redundancy in the feature lists. An alignment method was chosen based on the iterative RANSAC algorithm [23,24]. It allows to correct for some RT deviations which were observed for some HILIC-influenced analytes in the past [18]. In an attempt to eliminate misaligned features, the list was searched for duplicates which were then merged to consensus rows.
There are features which are only detected in one or the other sample, either because they were only present in one, their concentrations fell below the limit of detection or the feature was not extracted from the data. Missing values pose an issue in comparative non-target screening analysis, as they hamper the calculation of fold changes. Missing value imputation to replace zeros should be chosen with care as different univariate or multivariate analysis require different methods [25]. Instead, a gap-filling step was implemented into the data processing workflow that conducts a targeted search to retrospectively pick missed peaks or baseline from the raw data.
The results were nine final feature lists for each ionization mode (three PAC types, three concentrations). For each feature list, the mean log2(fc)-value and the respective standard deviation were calculated and are listed in Table 2 for the positive ionization mode and Table S2 for the negative mode. In contrast to what was suggested by the target compounds, no general decrease in signal intensity was observed as result of PAC treatment. On the contrary, the portion of non-target features with a log2(fc) > 1 increased at higher concentrations to a maximum of 13.4% at 30 mg L−1 of PAC H121 (for PAC H118 and H120 at the same concentration please refer to Figures S3 and S4). This result is visualized in Figure 2, showing that the portion of predominantly HILIC-influenced features with RTs < 17 min, exhibits an increase in signal intensity rather than a decrease. This could be attributed to contaminations desorbing from the PAC or, even though less likely, formation of transformation products. For post-treatments with granulated activated carbon and PAC dosed onto a sand filter, formation of transformation products has been observed before, possibly due to biological degradation [20,21]. Nevertheless, variability in the gap-filling results due to a noisy baseline could impair a clear distinction of consistent and increasing features. Our partner laboratory investigated the adsorption characteristics of (among others) the same PAC types under comparable experimental conditions, except they did not include a HILIC separation into their analytical platform [26]. For samples treated with 7 mg L−1 PAC compared to untreated samples, they found median fold changes of 0.54, 0.49 and 0.69 for H118, H120 and H121, respectively. Since these values are in or close to the defined lower limit of the consistency interval, no or little adsorption can be derived from these findings as well.

2.3. Further Statistical Analysis

A PCA was computed using the feature lists prior to alignment in order to determine how individual samples group together based on the normalized peak heights. As an unsupervised, multivariate method, it further served the purpose to examine the pooled QC samples which were measured at the beginning and the end of sequence, as suggested by Sangster et al. [27]. Figure 3 displays the score plot for the data acquired in the positive ionization mode. The equivalent for the negative mode is given in Figure S5 of the Supplementary Material. The first dimension (principal component 1, PC 1) explains 22.4% of the variation and the second dimension (PC 2) 16.5 %. Technical replicates are overall grouped closely together in both ionization modes. Samples subjected to different treatments are mostly separated on the first principal component whereas the H118 isotherm is here also separated on the second dimension. The samples of the H120 isotherm (green) are grouped together suggesting little treatment effects. It was expected that the QC samples cluster tightly together and thus the variability among the test samples reflects the differences caused by treatment effects. However, the samples (green crosses) are positioned far apart on the PC 1 axis. In summation, these findings indicate that the run order might be possibly responsible for some variability between measurements. That a gradual change throughout the analytical sequence rendered the results less reproducible cannot be fully excluded at this point. On the other hand, the feature lists were investigated by PCA prior to alignment and the recursive gap filling. Consequently, some of the observed effects might be softened later on in the data processing regime.
If the features’ signal intensities were significantly different between the untreated and the treated sample, a Welch’s test was performed on signal intensities. Since the probability of wrongfully rejecting a true hypothesis increases for multiple comparisons (here approximately 3000 features), p-values need to be corrected [28,29]. In this study, the Benjamini–Hochberg method for controlling the false discovery rate was applied [30]. Finally, an exemplary volcano plot for PAC H121 at a concentration of 30 mg L−1 was constructed and presented in Figure 4. The features’ negative log10-transformed and adjusted p-values were plotted against the corresponding log2-transformed fold changes. Features with adjusted p-values < 0.05 (horizontal dashed line) and log2(fc)-values < −1 and log2(fc) > 1 (vertical dashed lines) were considered to be of significant decrease and significant increase, respectively. Out of a total of 439 features that fell outside the consistency interval, 340 were statistically significant, of which four were annotated with standard compounds. This additional level of security substantiates the assumption that either compounds are newly formed or desorbed from the H121 PAC material. However, for PAC H118 only 13 significant features with decreasing or increasing feature intensities were detected (Table 2), even though the PCA suggested otherwise. Features with signal intensities close to the limit of detection as well as background signals deconvoluted by the gap-filling algorithm, tend to introduce variability into the data. Since the standard error factors into the denominator of the t-statistic, the t-value decreases and the degrees of freedom increase which leads to higher p-values. For this reason, fold changes need to be considered as a criterion besides significance testing.

3. Material and Methods

3.1. Chemicals

Ultrapure water and acetonitrile were obtained at LC–MS grade from Supelco (Darmstadt, Germany) and Honeywell (Morristown, NJ, USA). Ammonium acetate was purchased from Sigma-Aldrich (Seelze, Germany). Information on (internal) standard compounds is given in Tables S3 and S4 of the electronic Supplementary Material. The polar standard compounds for spike-in (log D at pH 7 of 0.30 to −4.10) were obtained from Neochema (Bodenheim, Germany), handled in four stock solutions at 10 µg mL−1 in methanol and stored at −18 °C. The compounds that served as internal standards were purchased from Sigma-Aldrich (Seelze, Germany) and Dr. Ehrenstorfer (Augsburg, Germany). They were prepared in individual stock solutions at 1000 µM, except for sotalol (586 µM), vidarabine (337 µM) and monuron (970 µM). They were dissolved in acetonitrile (etilefrine, sotalol, chlortoluron and metobromuron), acetonitrile/water (50/50, v/v; 6-amino-1,3-dimethyl-5-(formylamino)uracil, vidarabine and chloridazon) or methanol (chlorbromuron, metconazol and monuron) and stored at 4 °C.

3.2. Samples

For bench-scale batch sorption experiments, a surface water sample was taken from a German reservoir and filtered with a glass fiber filter (type GF 9; Schleicher & Schuell GmbH, Whatman, NH, USA) The dissolved organic carbon (DOC) concentration of the sample was 6.1 mg L−1, the pH 8.02 and the conductivity 526 µS cm−1 at 25 °C. The sample was spiked with 12 polar standard compounds at a final concentration of 50 µg L−1 per compound prior to treatment. The sample was separated into aliquots and treated with three different types of PAC (Table 3) at different concentrations. For each of the three PAC types, four 1 L-batches were prepared in beakers, adding no carbon (blank), 2 mg, 7 mg and 30 mg to the surface water sample. The batches were stirred at 250 rpm for 4 h at room temperature. Aliquots of all the batches were passed through a syringe filter (GF and 0.45 µm cellulose acetate, Minisart NML; Sartorius, Göttingen Germany) and transferred to baked-out vials (450 °C, 2 h). Treated samples were spiked with internal standards to a final concentration of 5 µM per compound prior to LC–MS analysis. Aliquots of all the samples were combined at equal volumes and also spiked with internal standards. Two triplicates of the pooled quality control (QC) samples were measured at the beginning and the end of the sequence.

3.3. LC–MS Analysis

The LC setup consisted of a HILIC and a RPLC system coupled in series via a T-piece with a mixing frit (Upchurch, IDEX Europe GmbH, Erlangen, Germany) [31].
Each LC system (1260 Infinity series; Agilent Technologies, Waldbronn, Germany) consisted of a binary pump, an online degasser and a mixing chamber. The RP separation was carried out on a Poroshell 120 EC-C18 column (50.0 × 3.0 mm, 2.7 µm; Agilent Technologies). The mobile phase consisted of 10 mM ammonium acetate in aqueous solution and acetonitrile at volumetric ratios of 90/10 and 10/90. For the HILIC subsystem, a ZIC-HILIC column was employed (150.0 × 2.1 mm, 5 µm, 200 Å; Merck Sequant, Umeå, Sweden) and the mobile phase consisted of acetonitrile and water. Information on the gradients can be found in previous publications [32,33]. The injection volume was 10 µL.
The chromatographic system was connected to an Orbitrap Exploris 120 mass spectrometer (Thermo Fisher Scientific GmbH; Dreieich, Germany) equipped with an electrospray ionization (ESI) source. The source was operated at spray voltages of 3.5 and −2.5 kV in the positive and negative modes, respectively. Sheath gas, auxiliary gas and sweep gas were set to 50, 8 and 0 (arbitrary units). The capillary temperature and the vaporizer temperature were set to 320 and 400 °C, respectively. In order to obtain NTS data, a mass range of 70–1000 Da was scanned at a resolution of 60,000 (full width at half maximum at m/z 200). MS2 spectra were acquired in the data-dependent acquisition mode at a resolution of 30,000 by employing a collision energy ramp of 15–45 eV. The four most abundant precursor ions were selected to trigger after one scan cycle and afterwards excluded for 7 s.

3.4. Data Analysis

3.4.1. Extracting Target Compounds

The targeted analysis was performed in MZmine 2 [23]. The internal standards as well as the polar standard compounds were extracted at a mass tolerance of 3 ppm and a RT tolerance of 5 min. For further confirmation, detected MS2 spectra were compared to experimental MS2 spectra stored in the mass spectral databases MassBank [34,35] and MassBank of North America (MoNA) [36]. For the retention times (RT) and masses of the internal standards, the precision values were calculated, expressing the closeness of observed values to each other. Additionally, the mass accuracy was determined, expressing the closeness of detected masses to theoretical masses.

3.4.2. Extracting aligned non-target feature lists

The non-targeted data analysis was performed in MZmine 2 [23] as well. Figure 5 schematically shows the workflow for extracting features from NTS data, filtering and bringing the feature lists of the treated and untreated sample together for comparative analysis. The workflow is shortly outlined in the following:
First, masses were detected on MS1 and MS2 level whereas the RPLC as well as the HILIC retention intervals (5 min–33 min) were considered. Then, chromatograms were built and peaks deconvoluted using the ADAP algorithm (“Automated Data Analysis Pipeline”) [37]. Chromatograms were smoothed by applying a Savitzky-Golay filter. Peaks were filtered out if the number of data points, tailing factor and/or asymmetry factor fell outside a predefined range. Thereafter, isotopic patterns of singly charged molecular ions were detected and isotopic peaks as well as sodium, potassium and ammonium adducts were removed. The feature lists were aligned across the three technical replicates using the RANSAC (“random sample consensus”) aligner [23,24] and subsequently duplicates were removed. Finally, the features that were detected in less than three replicates were eliminated.
The cleaned-up feature lists of the treated and the untreated sample were subsequently aligned for comparative analysis. For features which were only present or detected in one of the two samples, the respective gap was filled by a recursive targeted search in the raw data. Feature rows whose gaps could not be filled were removed from the list. Finally, the features’ signal intensities (peak heights) were normalized using the internal standards. All the standard compounds contributed to the normalization factor, but were weighted based on the m/z and RT distance to the feature [38]. The final feature list was exported to comma-separated values (CSV) format. The processing steps along with the parameter settings and explanatory comments are listed in Table S5 of the Supplementary Material. Some parameters were optimized in a previous study on two synthetic water samples containing environmentally relevant and polar standard compounds at different concentration levels [18]. Other parameters were adjusted to fit the requirements of the present study, given for example by a different type of high-resolution mass spectrometer or research objective. Mass and RT tolerances were set based on the targeted analysis of the internal standards as well as the spiked HILIC standards. The parameter settings were validated and if necessary optimized on the three blank samples with the objective of obtaining full recall of the standard compounds and reducing the total feature number as well as the processing time.

3.4.3. Further Statistical Analysis

The fold changes (fc) of the target compounds and features were calculated as the ratio of the mean signal intensities across the technical replicates in the treated and the untreated sample. For each comparison (e.g., H118 (7/0) mg), the mean fc and the standard deviation of all the features were calculated. The threshold values for signal increase and decrease in a feature after activated carbon treatment were defined at log2(fc) >1 and log2(fc) <−1, respectively [22]. Further statistical evaluations were performed in R (version 4.0.2) [39] and RStudio (version 1.3.959) [40]. A principal component analysis (PCA) was carried out for the three blank samples, the nine treated samples and the six pooled QC samples for each ionization mode. The data comprised normalized peak heights and “0” was put for missing values. The data were scaled and centered before being submitted to the prcomp function. Whether the feature intensities are significantly different between the treated and the untreated sample was evaluated using Welch’s t-test [41] and the Benjamini–Hochberg adjustment [30] at a significance level of 0.05.

4. Conclusions

Non-target screening was carried out in order to evaluate the treatment of surface water samples with powdered activated carbon.
A target screening in the ESI(+) mode of the polar standard compounds, spiked into the samples prior to PAC treatment, indicated the reduction of famotidine, 2,4-diamino-6-(hydroxymethyl)pteridine, 2-aminopyridine and 3-pyridinemethanol. No reduction was observed for 4-(2-hydroxyethyl)morpholine when treated with PAC H120 in contrast to PAC types H118 and H121. 2,2,6,6-tetramethyl-4-piperidone was only reduced by PAC H118. According to criteria defined within this study, PAC treatment did not affect 1,3-dimethyl-2-imidazolidinone, ectoine, miglitol and N,N′-ethylenebisacetamide. According to the presented results half of the highly polar standard compounds (log D < 0.30 at pH 7) were (partially) removed by at least one of the investigated PAC types. However, more and quantitative data are necessary to substantiate these findings.
Non-target screening data were evaluated using fold changes, a principal component analysis and multiple hypothesis testing. Derived from the results, features exhibit repeatable peak heights across technical replicates; however, the run order appeared to introduce some variability throughout the sequence of measurements. The mean fold changes of non-target features fall into the consistency interval, overall suggesting no elimination or formation. However, the variability increases with PAC concentration indicating removal or formation of individual compounds. The fraction of features with increasing signal intensities (fold change > 2.00) was elevated at the highest tested PAC concentration of 30 mg L−1 with up to 13% and the highest number of significant features (based on differences in their signal intensities) observed for PAC type H121. Either compounds were newly formed or desorbed from the PAC material. It should be noted that the results could be affected by the feature extraction and data clean-up process, especially the gap-filling step.
For future reference, the priority features (showing significant increase or decrease in feature intensities) could be identified and their (eco)toxicological relevance assessed. The applied chromatographic technology widens the view of molecules observed by mass spectrometric non-target screening. Thus, adsorption processes (and polarities of adsorbed molecules) can be monitored in a better way and classified more sustainably.

Supplementary Materials

The following Supplementary Materials can be downloaded at: https://www.mdpi.com/article/10.3390/molecules27165214/s1, Table S1: Means and standard deviations of log2(fc) values for the internal standards and the polar standard compounds measured in negative ionization mode; Table S2: Means and standard deviations of log2(fc) values for the non-target features in negative ionization mode; Table S3: Table of internal standards spiked into samples after treatment and prior to LC-MS analysis; Table S4: Polar standard compounds spiked into samples prior to PAC treatment; Table S5: Parameter settings of each processing step of the non-target screening workflow for comparative analysis of an untreated and a treated sample; Figure S1: The base-2 logarithm of the fold changes of the polar standard compounds (blue diamonds) and the internal standards (orange circles) are plotted versus their retention times; Figure S2: The total ion chromatograms (blue line, primary y-axis) and the non-target peaks (black crosses, secondary y-axis) are exemplarily displayed for the sample treated with the H118 PAC at 30 mg L−1, third replicate; Figure S3: Non-target features (green crosses) and polar standard compounds (blue diamonds); Figure S4: Non-target features (green crosses) and polar standard compounds (blue diamonds) of the sample treated with 30 mg L-1 of PAC H120 compared to the untreated blank sample, both measured in positive ionization mode; Figure S5: Scores plot of the PCA based on the normalized peak heights of the features extracted from each individual measurement in negative ionization mode.

Author Contributions

Conceptualization, S.M.; formal analysis, S.M.; investigation, S.B.; data curation, S.M.; writing-original draft preparation, S.M.; writing-review and editing, T.L., S.B.; supervision, T.L., S.B.; project administration, T.L.; funding acquisition, T.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research was partially funded by the “Westfälische Wasser- und Umweltanalytik GmbH”.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

We especially thank André Liesener, Anna Maria Reineke, Karin Hilscher, Jacqueline Porada and Anna-Lisa Kowalski as well as other colleagues at the Westfälische Wasser- und Umweltanalytik GmbH in Gelsenkirchen. Without their contribution to the laboratory work as well as the discussion, this study would not have been possible. The authors acknowledge Jörg E. Drewes and the Chair of Urban Water Systems Engineering from the Technical University of Munich for creating the research infrastructure.

Conflicts of Interest

The authors declare no conflict of interest.

Sample Availability

Samples of the compounds applied in this study are available from the authors.

References

  1. Luo, Y.; Guo, W.; Ngo, H.H.; Nghiem, L.D.; Hai, F.I.; Zhang, J.; Liang, S.; Wang, X.C. A review on the occurrence of micropollutants in the aquatic environment and their fate and removal during wastewater treatment. Sci. Total. Env. 2014, 473–474, 619–641. [Google Scholar] [CrossRef] [PubMed]
  2. Mompelat, S.; Bot, B.L.; Thomas, O. Occurrence and fate of pharmaceutical products and by-products, from resource to drinking water. Env. Int. 2009, 35, 803–814. [Google Scholar] [CrossRef] [PubMed]
  3. Karakurt, S.; Schmid, L.; Hu, U.; Drewes, J.E. Dynamics of Wastewater Effluent Contributions in Streams and Impacts on Drinking Water Supply via Riverbank Filtration in Germany-A National Reconnaissance. Env. Sci. Technol. 2019, 53, 6154–6161. [Google Scholar] [CrossRef] [PubMed]
  4. Eggen, R.I.L.; Hollender, J.; Joss, A.; Schärer, M.; Stamm, C. Reducing the discharge of micropollutants in the aquatic environment: The benefits of upgrading wastewater treatment plants. Env. Sci. Technol. 2014, 48, 7683–7689. [Google Scholar] [CrossRef]
  5. 814.201 Gewässerschutzverordnung (GSchV) of 10/28/1998, status of 01/01/2021. Der Schweizerische Bundesrat. Available online: https://www.fedlex.admin.ch/eli/cc/1998/2863_2863_2863/de. (accessed on 10 August 2022).
  6. Hernández-Leal, L.; Temmink, H.; Zeeman, G.; Buisman, C.J.N. Removal of micropollutants from aerobically treated grey water via ozone and activated carbon. Water Res. 2011, 45, 2887–2896. [Google Scholar] [CrossRef]
  7. Leal, L.H.; Vieno, N.; Temmink, H.; Zeeman, G.; Buisman, C.J.N. Occurrence of xenobiotics in gray water and removal in three biological treatment systems. Env. Sci. Technol. 2010, 44, 6835–6842. [Google Scholar] [CrossRef]
  8. Kovalova, L.; Siegrist, H.; Gunten, U.V.; Eugster, J.; Hagenbuch, M.; Wittmer, A.; Moser, R.; McArdell, C.S. Elimination of micropollutants during post-treatment of hospital wastewater with powdered activated carbon, ozone, and UV. Env. Sci. Technol. 2013, 47, 7899–7908. [Google Scholar] [CrossRef]
  9. Kovalova, L.; Knappe, D.R.U.; Lehnberg, K.; Kazner, C.; Hollender, J. Removal of highly polar micropollutants from wastewater by powdered activated carbon. Environ. Sci. Pollut. Res. 2013, 20, 3607–3615. [Google Scholar] [CrossRef]
  10. Boxall, A.; Sinclair, C.; Fenner, K.; Koplin, D.; Maund, S. When synthetic chemicals degrade in the environment. Env. Sci. Technol. 2004, 38, 368–375. [Google Scholar] [CrossRef]
  11. Schmitt-Jansen, M.; Bartels, P.; Adler, N.; Altenburger, R. Phytotoxicity assessment of diclofenac and its phototransformation products. Anal. Bioanal. Chem. 2007, 387, 1389–1396. [Google Scholar] [CrossRef]
  12. Petrie, B.; Barden, R.; Kasprzyk-Hordern, B. A review on emerging contaminants in wastewaters and the environment: Current knowledge, understudied areas and recommendations for future monitoring. Water Res. 2015, 72, 3–27. [Google Scholar] [CrossRef]
  13. Gómez, M.J.; Sirtori, C.; Mezcua, M.; Fernández-Alba, A.R.; Agüera, A. Photodegradation study of three dipyrone metabolites in various water systems: Identification and toxicity of their photodegradation products. Water Res. 2008, 42, 2698–2706. [Google Scholar] [CrossRef]
  14. Krauss, M.; Singer, H.; Hollender, J. LC-high resolution MS in environmental analysis: From target screening to the identification of unknowns. Anal. Bioanal. Chem. 2010, 397, 943–951. [Google Scholar] [CrossRef]
  15. Schulz, W.; Lucke, T. Non-Target Screening in Water Analysis—Guideline for the Application of LC-ESI-HRMS for Screening; German Water Chemistry Society: Mülheim an der Ruhr, Germany, 2019; Available online: https://www.wasserchemische-gesellschaft.de/images/HAIII/NTS-Guidline_EN_s.pdf (accessed on 10 August 2022).
  16. Hollender, J.; Schymanski, E.L.; Singer, H.P.; Ferguson, P.L. Nontarget Screening with High Resolution Mass Spectrometry in the Environment: Ready to Go? Env. Sci. Technol. 2017, 51, 11505–11512. [Google Scholar] [CrossRef]
  17. Schymanski, E.L.; Singer, H.P.; Longrée, P.; Loos, M.; Ruff, M.; Stravs, M.A.; Vidal, C.R.; Hollender, J. Strategies to characterize polar organic contamination in wastewater: Exploring the capability of high resolution mass spectrometry. Env. Sci. Technol. 2014, 48, 1811–1818. [Google Scholar] [CrossRef]
  18. Minkus, S.; Bieber, S.; Letzel, T. (Very) polar organic compounds in the Danube river basin: Non-target screening workflow and prioritization strategy for extracting highly confident features. Anal. Methods. 2021, 13, 2044–2054. [Google Scholar] [CrossRef]
  19. Schollée, J.E.; Schymanski, E.L.; Hollender, J. Statistical Approaches for LC-HRMS Data to Characterize, Prioritize, and Identify Transformation Products from Water Treatment Processes. ACS Symp. Ser. 2016, 4, 45–65. [Google Scholar]
  20. Brunner, A.M.; Bertelkamp, C.; Dingemansm, M.M.L.; Kolkman, A.; Wols, B.; Harmsen, D.; Siegers, W.; Martijn, B.J.; Oorthuizen, W.A.; ter Laak, T.L. Integration of target analyses, non-target screening and effect-based monitoring to assess OMP related water quality changes in drinking water treatment. Sci. Total. Env. 2020, 705, 135779. [Google Scholar] [CrossRef]
  21. Schollée, J.E.; Hollender, J.; McArdell, C.S. Characterization of advanced wastewater treatment with ozone and activated carbon using LC-HRMS based non-target screening with automated trend assignment. Water Res. 2021, 200, 117209. [Google Scholar] [CrossRef]
  22. Bader, T.; Schulz, W.; Kümmerer, K.; Winzenbacher, R. LC-HRMS Data Processing Strategy for Reliable Sample Comparison Exemplified by the Assessment of Water Treatment Processes. Anal. Chem. 2017, 89, 13219–13226. [Google Scholar] [CrossRef]
  23. Pluskal, T.; Castillo, S.; Villar-Briones, A.; Orešič, M. MZmine 2: Modular framework for processing, visualizing, and analyzing mass spectrometry-based molecular profile data. BMC Bioinform. 2010, 11, 1–11. [Google Scholar] [CrossRef]
  24. Fischler, M.A.; Bolles, R.C. Random Sample Consensus: A Paradigm for Model Fitting with Applications to Image Analysis and Automated Cartography. Commun. ACM 1981, 24, 381–395. [Google Scholar] [CrossRef]
  25. Guida, R.D.; Engel, J.; Allwood, J.W.; Weber, R.J.M.; Jones, M.R.; Sommer, U.; Viant, M.R.; Dunn, W.B. Non-targeted UHPLC-MS metabolomic data processing methods: A comparative investigation of normalisation, missing value imputation, transformation and scaling. Metabolomics 2016, 12, 1–14. [Google Scholar] [CrossRef]
  26. Reineke, A.; Liesener, A.; Betz, L. Anwendung von Nontarget-Screening und Wirkungsbezogener Analytik in der Charakterisierung von Aktivkohle zur Aufbereitung von Trinkwasser. Vom Wasser. 2021, 119, 78–81. [Google Scholar] [CrossRef]
  27. Sangster, T.; Major, H.; Plumb, R.; Wilson, A.J.; Wilson, I.D. A pragmatic and readily implemented quality control strategy for HPLC-MS and GC-MS-based metabonomic analysis. Analyst 2006, 131, 1075–1078. [Google Scholar] [CrossRef]
  28. Chen, S.Y.; Feng, Z.; Yi, X. A general introduction to adjustment for multiple comparisons. J. Thorac. Dis. 2017, 9, 1725–1729. [Google Scholar] [CrossRef]
  29. Bender, R.; Lange, S. Adjusting for multiple testing—when and how? Ralf. J Clin. Epidemiol. 2001, 54, 343–349. [Google Scholar] [CrossRef]
  30. Benjamini, Y.; Hochberg, Y. Controlling the False Discovery Rate: A Practical and Powerful Approach to Multiple Testing. J. R. Stat. Soc. Ser. B 1995, 57, 289–300. [Google Scholar] [CrossRef]
  31. Greco, G.; Grosse, S.; Letzel, T. Serial coupling of reversed-phase and zwitterionic hydrophilic interaction LC/MS for the analysis of polar and nonpolar phenols in wine. J. Sep. Sci. 2013, 36, 1379–1388. [Google Scholar] [CrossRef]
  32. Bieber, S.; Greco, G.; Grosse, S.; Letzel, T. RPLC-HILIC and SFC with Mass Spectrometry: Polarity-Extended Organic Molecule Screening in Environmental (Water) Samples. Anal. Chem. 2017, 89, 7907–7914. [Google Scholar] [CrossRef] [PubMed]
  33. Minkus, S.; Grosse, S.; Bieber, S.; Veloutsou, S.; Letzel, T. Optimized hidden target screening for very polar molecules in surface waters including a compound database inquiry. Anal. Bioanal. Chem. 2020, 412, 4953–4966. [Google Scholar] [CrossRef] [PubMed]
  34. Horai, H.; Arita, M.; Kanaya, S.; Nihei, Y.; Ikeda, T.; Suwa, K.; Ojima, Y.; Tanaka, K.; Tanaka, S.; Aoshima, K.; et al. MassBank: A public repository for sharing mass spectral data for life sciences. J. Mass Spectrom. 2010, 45, 703–714. [Google Scholar] [CrossRef] [PubMed]
  35. European MassBank Mass Spectral Database of the NORMAN Network. Available online: https://massbank.eu/MassBank/ (accessed on 2 February 2022).
  36. MassBank of North America. Available online: https://mona.fiehnlab.ucdavis.edu/ (accessed on 2 February 2022).
  37. Myers, O.D.; Sumner, S.J.; Li, S.; Barnes, S.; Du, X. One Step Forward for Reducing False Positive and False Negative Compound Identifications from Mass Spectrometry Metabolomics Data: New Algorithms for Constructing Extracted Ion Chromatograms and Detecting Chromatographic Peaks. Anal. Chem. 2017, 89, 8696–8703. [Google Scholar] [CrossRef] [PubMed]
  38. Katajamaa, M.; Orešič, M. Processing methods for differential analysis of LC/MS profile data. BMC Bioinform. 2005, 6, 1–12. [Google Scholar] [CrossRef]
  39. Team, R.C. R: A Language and Environment for Statistical Computing; R Foundation for Statistical Computing: Vienna, Austria, 2013; p. 201. [Google Scholar]
  40. Team, R.S. RStudio: Integrated Development for R. RStudio, PBC. Boston MA 2020, 770, 165–171. [Google Scholar]
  41. Welch, B.L. The Generalization of “Student’s” Problem when Several Different Population Variances are Involved. Biometrika 1947, 34, 28–35. [Google Scholar] [CrossRef]
Figure 1. The base-2 logarithm of the fold changes of the HILIC standards (added prior to PAC treatment, blue diamonds) and the internal standards (added prior to analysis, orange circles) are plotted versus their RTs. The dashed lines mark the consistency interval where no compound removal is assumed.
Figure 1. The base-2 logarithm of the fold changes of the HILIC standards (added prior to PAC treatment, blue diamonds) and the internal standards (added prior to analysis, orange circles) are plotted versus their RTs. The dashed lines mark the consistency interval where no compound removal is assumed.
Molecules 27 05214 g001
Figure 2. Non-target features (green crosses) and polar standard compounds (blue diamonds) are plotted by their base-2 logarithmic fold changes and RT. The dashed lines mark the consistency interval. Log2(fc) values <−1 and >1 are defined as a decrease and increase in signal intensity, respectively. Here, the sample treated with 30 mg L−1 of PAC H121 was compared to the untreated blank sample, both measured in the positive ionization mode.
Figure 2. Non-target features (green crosses) and polar standard compounds (blue diamonds) are plotted by their base-2 logarithmic fold changes and RT. The dashed lines mark the consistency interval. Log2(fc) values <−1 and >1 are defined as a decrease and increase in signal intensity, respectively. Here, the sample treated with 30 mg L−1 of PAC H121 was compared to the untreated blank sample, both measured in the positive ionization mode.
Molecules 27 05214 g002
Figure 3. Scores plot of the PCA based on the normalized peak heights of the feature extracted from each individual measurement.
Figure 3. Scores plot of the PCA based on the normalized peak heights of the feature extracted from each individual measurement.
Molecules 27 05214 g003
Figure 4. The volcano plot shows features extracted during the comparative analysis of the sample treated with PAC type H121 at 30 mg L−1 and the respective blank sample. The horizontal dashed line marks the cut-off p-value of −log10(0.05) and the vertical dashed lines the consistency interval. Features which were annotated with polar standard compounds are marked with blue diamonds.
Figure 4. The volcano plot shows features extracted during the comparative analysis of the sample treated with PAC type H121 at 30 mg L−1 and the respective blank sample. The horizontal dashed line marks the cut-off p-value of −log10(0.05) and the vertical dashed lines the consistency interval. Features which were annotated with polar standard compounds are marked with blue diamonds.
Molecules 27 05214 g004
Figure 5. NTS data processing workflow for comparing a treated and an untreated surface water sample. The processing steps for extracting and filtering features from the technical replicates of a single sample are depicted in black. The orange boxes are processing steps that enable the comparative analysis of the treated and the untreated sample.
Figure 5. NTS data processing workflow for comparing a treated and an untreated surface water sample. The processing steps for extracting and filtering features from the technical replicates of a single sample are depicted in black. The orange boxes are processing steps that enable the comparative analysis of the treated and the untreated sample.
Molecules 27 05214 g005
Table 1. Means and standard deviations of log2(fc) values for the internal standards and the polar standard compounds measured in the positive ionization mode. H118, H120 and H121 are the laboratory names of the different PAC types which were tested for surface water treatment at three different concentrations.
Table 1. Means and standard deviations of log2(fc) values for the internal standards and the polar standard compounds measured in the positive ionization mode. H118, H120 and H121 are the laboratory names of the different PAC types which were tested for surface water treatment at three different concentrations.
Concentration [mg L−1]H118H120H121
log2(fc)log2(fc)log2(fc)
Internal standardsn = 10n = 10n = 10
2 0.09 ± 0.05−0.20 ± 0.070.03 ± 0.04
7 −0.04 ± 0.09−0.25 ± 0.140.08 ± 0.04
30 −0.03 ± 0.10−0.21 ± 0.16−0.01 ± 0.10
Polar standard compoundsn = 10n = 10n = 10
2 −0.40 ± 0.49−0.21 ± 0.49−0.17 ± 0.22
7 −1.22 ± 1.46−0.70 ± 1.30−0.65 ± 0.95
30 −2.31 ± 2.21−2.07 ± 2.45−1.59 ± 1.88
Table 2. Means and standard deviations of log2(fc) values for the non-target features in the positive ionization mode. H118, H120 and H121 are the laboratory names of the different PAC types (Table 3) which were tested for surface water treatment at three different concentrations.
Table 2. Means and standard deviations of log2(fc) values for the non-target features in the positive ionization mode. H118, H120 and H121 are the laboratory names of the different PAC types (Table 3) which were tested for surface water treatment at three different concentrations.
Concentration [mg L−1]Number of FeaturesMean log2(fc)Increasing/Decreasing Features [%]Significant Features
H118
2 2981−0.16 ± 0.380.6/2.738
7 3366−0.26 ± 0.430.4/5.10
30 30000.07 ± 0.574.5/2.713
H120
2 29410.11 ± 0.392.1/0.840
7 28560.11 ± 0.391.7/1.229
30 30580.17 ± 0.423.0/0.795
H121
2 2842−0.04 ± 0.371.1/1.828
7 28860.17 ± 0.413.5/0.72
30 30990.36 ± 0.6213.4/0.8336
Table 3. The types of powdered activated carbon (PAC) used for the batch sorption experiments. Values were partially published before [26].
Table 3. The types of powdered activated carbon (PAC) used for the batch sorption experiments. Values were partially published before [26].
Laboratory NameH118H120H121
Manufacturer Supplier ASupplier ASupplier B
Water content8.1 %1.5 %2.0 %
Ash content6.7 %13.6 %10.2 %
Contact pH10.89.910.1
Iodine number1088 mg g−11019 mg g−1944 mg g−1
Particle size distribution (wet sieving)
<150 µm99.199.199.7
<50 µm72.088.670.2
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Minkus, S.; Bieber, S.; Letzel, T. Characterizing Powdered Activated Carbon Treatment of Surface Water Samples Using Polarity-Extended Non-Target Screening Analysis. Molecules 2022, 27, 5214. https://doi.org/10.3390/molecules27165214

AMA Style

Minkus S, Bieber S, Letzel T. Characterizing Powdered Activated Carbon Treatment of Surface Water Samples Using Polarity-Extended Non-Target Screening Analysis. Molecules. 2022; 27(16):5214. https://doi.org/10.3390/molecules27165214

Chicago/Turabian Style

Minkus, Susanne, Stefan Bieber, and Thomas Letzel. 2022. "Characterizing Powdered Activated Carbon Treatment of Surface Water Samples Using Polarity-Extended Non-Target Screening Analysis" Molecules 27, no. 16: 5214. https://doi.org/10.3390/molecules27165214

Article Metrics

Back to TopTop