Assessment of Gesture Accuracy for a Multi-Electrode EMG-Sensor-Array-Based Prosthesis Control System

by
Vinod Sharma
1,*,
Erik Lloyd
1,
Mike Faltys
1,
Max Ortiz-Catalan
2,3 and
Connor Glass
1
1
Phantom Neuro Inc., Austin, TX 78704, USA
2
Prometei Pain Rehabilitation Center, 21000 Vinnytsia, Ukraine
3
Center for Complex Endoprosthetics, Osseointegration, and Bionics, 04136 Kyiv, Ukraine
*
Author to whom correspondence should be addressed.
Prosthesis 2025, 7(4), 99; https://doi.org/10.3390/prosthesis7040099
Submission received: 17 April 2025 / Revised: 26 July 2025 / Accepted: 8 August 2025 / Published: 13 August 2025

Abstract

Background: Upper limb loss significantly impacts quality of life, and whereas myoelectric prostheses restore some function, conventional surface electromyography (sEMG) systems face challenges like poor signal quality, high cognitive burden, and suboptimal control. Phantom X, a novel implantable electrode-array-based system leveraging machine learning (ML), aims to overcome these limitations. This feasibility study assessed Phantom X’s performance using non-invasive surface EMG electrodes to approximate implantable system behavior. Methods: This single-arm, non-randomized study included 11 participants (9 able-bodied, 2 with transradial amputation) fitted with a 32-electrode cutaneous array around the forearm. EMG signals were processed through an ML algorithm to control a desk-mounted prosthesis. Performance was evaluated via gesture accuracy (GA), modified gesture accuracy (MGA), and classifier gesture accuracy (CGA) across 11 hand gestures in three arm postures. User satisfaction was also assessed among the two participants with transradial amputation. Results: Phantom X achieved an average GA of 89.0% ± 6.8%, MGA of 96.8% ± 2.0%, and CGA of 93.6% ± 4.1%. Gesture accuracy was the highest in the Arm Parallel posture and the lowest in the Arm Perpendicular posture. Thumbs Up had the highest accuracy (100%), while Index Point and Index Tap gestures showed lower performance (70% and 79% GA, respectively). The mean latency between EMG onset and gesture detection was 250.5 ± 145.9 ms, with 91% of gestures executed within 500 ms. The amputee participants reported high satisfaction. Conclusions: This study demonstrates Phantom X’s potential to enhance prosthesis control through multi-electrode EMG sensing and ML-based gesture decoding. The non-invasive evaluation suggests high accuracy and responsiveness, warranting further studies with the implantable system to assess long-term usability and real-world performance. Phantom X may offer a superior alternative to conventional sEMG-based control, potentially reducing cognitive burden and improving functional outcomes for upper limb amputees.

1. Introduction

Arms, hands, and fingers are vital for numerous activities of daily living, and thus the loss of one or both upper limbs is a devastating disability and significantly impacts an individual’s quality of life. Globally, an estimated 180 million people live with limb loss due to various causes (e.g., vascular conditions, trauma, and congenital factors) [1]. Of these, approximately 58 million people are affected by traumatic limb amputation, with around 39% experiencing unilateral or bilateral upper limb loss [2]. In a recent publication, Rivera et al. [3] estimated that approximately 2.3 million people in the U.S. are living with limb loss, with around 200,000 affected by upper limb amputation. Furthermore, this number is projected to double by 2050.
Commercially available prostheses can help to restore functionality in individuals with limb deficiencies and enable them to perform activities of daily living that they otherwise would not be able to accomplish. Advancements in prosthetic technologies, especially myoelectric control, have enabled modern prostheses to make significant improvements in the pursuit of mimicking the versatility of natural limbs. Electromyography (EMG) sensing from the residual limb stump using cutaneous surface electrodes is the most ubiquitous approach to controlling robotic prostheses. In this approach, dating back to the 1970s, cutaneous electrodes are placed within the inner surface of a prosthetic socket and sense EMG signals from the surface of the skin associated with muscle contractions in the residual limb [4,5]. In the most simplistic embodiment, which is still the most commonly used, two oppositely placed bipolar electrodes are used for prosthesis control; one electrode pair overlies the flexor muscles, and the second electrode pair overlies the extensor muscles. However, this first-generation embodiment only captures a small subset of EMG signals that are involved in hand and wrist movements and, hence, can only be used for very limited control of a prosthetic device (even though the prosthetic device itself is capable of far more complex movements). To execute a greater set of movements with this electrode-pair embodiment, the user must initiate specific muscle contractions (e.g., two rapid muscle contractions) to change the state of the prosthesis (e.g., change the mode, change the selected gesture, etc.). While this approach does technically allow the user to perform a greater number of gestures, it is unnatural, time consuming, and cognitively burdensome. This contributes heavily to a large abandonment rate of myoelectric prostheses, ranging from ~30–40%, observed to this day [6,7].
Thus, more advanced prosthesis control systems using a greater number of surface electrodes combined with machine learning algorithms, commonly referred to as myoelectric pattern recognition (MPR), have been proposed and shown to achieve high classification accuracy in the literature [8,9,10,11,12,13,14]. However, with a few exceptions, the gesture set tested in these studies has been relatively small and often included limited involvement of individual fingers (e.g., one recent study only tested six gestures, and none involved individual fingers [8]). While there is consistent evidence that classification performance improves with increasing electrode count [9,10], most of these systems use experimental electrode arrays suitable for acute lab settings. They remain at an early academic research stage, and only a few have progressed to commercial viability [15,16,17]. These commercially available sensor arrays and accompanying MPR algorithms for decoding muscle signals can, in principle, enable more versatile and complex prosthesis movements. This is expected to translate into greater prosthesis control, and, indeed, various controlled studies have demonstrated that MPR provides superior control compared to conventional dual-electrode direct control in real-world settings [18,19].
However, both MPR and conventional prosthesis control systems suffer from the same challenges arising from sEMG. For instance, the fact that cutaneous electrodes sit farther from the source of the EMG signal (i.e., underlying muscle) reduces the signal-to-noise ratio (SNR), sometimes to unusable levels when abundant soft tissues or scar tissues are present. The sEMG interface is also highly susceptible to electromagnetic interference, reducing the SNR even further. In addition, the non-polarizable, high-impedance electrode–skin interface also makes it susceptible to motion artifacts that often result in unwanted activation of the prosthesis. Low electrode density, typically eight or fewer electrode pairs, and placement challenges make it difficult to isolate discrete EMG signals from specific muscle groups. Instead, crosstalk results in recorded sEMG being a summation of signals from multiple muscles. Moreover, relative movement of surface electrodes on the skin, especially with socket rotation, can reduce the reliability of MPR controllers. Finally, skin perspiration alters the electrical impedance of the source–sense interface, adding variability to the recordings [19]. These challenges make current sEMG-controlled prostheses bulky due to the integration of sensors and wiring into the socket, unreliable due to inconsistent signal detection, and inconvenient, as they often create an unnatural user experience and require retraining. Therefore, current myoelectric prosthesis performance fails to meet users’ needs, leading to a high abandonment rate. Salminger et al. [7] observed overall abandonment rates of upper limb prostheses of ~44% in a population of mainly myoelectric prosthesis users (~92%). This, combined with high rates of compensation injuries, depression, and loss of autonomy observed in populations with limb loss, demonstrates a clear need for better solutions that can provide more intuitive, accurate, and reliable prosthesis control to broadly improve patient outcomes and reduce prosthesis abandonment [20,21,22].
The Phantom X prosthesis control system (Phantom Neuro Inc., Austin, TX, USA) aims to address several of the aforementioned limitations of the current sEMG-based prosthesis control systems, with a goal of enabling a clinically scalable solution. At the heart of the Phantom X system (Figure 1) is a multi-electrode implantable sensor array that is circumferentially placed around a user’s forearm and senses EMG signals at a high spatial resolution. Considerable evidence exists that implantable electrodes result in more accurate prosthesis control. For example, the concept of invasive recording electrodes that directly record from the surface of the muscle (epimysial) or from within the muscle itself (intramuscular) is well known and has yielded the most advanced motor control of upper limb prostheses in humans to date, albeit in an experimental, non-scalable setting [23,24,25]. However, the invasive nature of the Phantom X system warrants that the development of associated algorithms be conducted in phases to minimize risk to patients. In the first phase, largely constituted by this Assessment of Gesture Accuracy for a Multi-electrode EMG-Sensor-Array-Based Prosthesis Control System (ASCENT) study, we evaluated the performance of the Phantom X system using a non-invasive EMG sensor array. In a subsequent phase, longitudinal clinical studies will be undertaken with the implantable version containing the aforementioned electrode array(s). The objective of this study was to evaluate the performance of the Phantom X system, a novel platform with strong potential for rapid clinical translation. The Phantom X system employs a computationally efficient machine-learning-based control algorithm driven by input from a high-density array of 32 unipolar electrodes, designed to replicate the form factor of its implantable array, across a broader set of gestures intended to meaningfully enhance daily task performance for users. These results establish a critical foundation for future human studies with the implanted system and represent a meaningful step toward scalable, high-performance myoelectric control.

2. Materials and Methods

The ASCENT study was a single-arm, single-center, non-randomized, non-blinded study requiring one in-office visit by the participating subjects. This study was conducted in accordance with the principles of Good Clinical Practice (GCP) and the Declaration of Helsinki. Institutional Review Board (IRB) approval was obtained prior to study initiation, and all the participants provided written informed consent before any study-related procedures were performed. This study was registered at clinicaltrials.gov before first subject enrollment (NCTNCT06446037).
We enrolled able-bodied and amputee participants 21 years of age or older who signed the informed consent and were able to follow the study directions. The amputee subjects were required to have amputation at the transradial level either in one or both arms. Individuals meeting one or more of the following criteria were excluded from this study: (1) previously diagnosed muscle pathology, (2) cognitive impairment limiting their ability to follow study directions, (3) impaired muscle function and/or ability to perform normal phantom hand movements (applicable to amputees only), (4) allergies to skin adhesive materials necessary for cutaneous electrode placement, (5) excessive hair growth on arms and inability to shave off the hair for electrode placement, (6) arms or residual limbs with insufficient diameter to accommodate the wearing of two cutaneous sensor arrays, (7) injuries, bruises, or open wounds on the arm that need to be instrumented for the study, and (8) pregnant women.
After obtaining informed consent, the subjects came in for a single in-office visit that lasted for up to 3 h. The subjects were instrumented with cutaneous electrode arrays with 32 electrodes (Figure 2) circumferentially wrapped around the subject’s forearm. Each electrode of the array was gold-plated and 7.5 mm in diameter. The adjacent electrodes were spaced 7 mm apart and staggered by 12 mm. The array was adhered to the skin with a 1.1 mm thick foam pad (3M PE Foam, 3M, Minneapolis, MN, USA) and terminated into an EMG hardware board. The openings over the gold-plated contacts were cut into the foam, forming wells for electrode paste (Weaver and Company, Aurora, CO, USA). In addition, reference and ground electrodes (H124SG Kendall—Cardinal Health, Dublin, OH, USA) were attached near the array and then to the reference and ground terminal of the EMG board.
The EMG signals sensed by the electrodes were input into custom electronics hardware that amplified and filtered the signals. The amplifiers had a gain of 20×. The hardware used a bandpass filter of 5 Hz–4 kHz. The pre-processed EMG signals were wirelessly transferred to an edge computing component of the Phantom X system known as the Fusion Port. The Fusion Port was connected to a Windows desktop computer via a universal serial bus (USB) running machine learning (ML) algorithm that decoded the EMG signals into prosthesis control signals (Figure 3). The computer running the ML algorithm was also used for data storage and as a user interface to provide gesture instructions. The control signals delivered by the Fusion Port controlled a desk-mounted prosthesis (Psyonic, San Diego, CA, USA) via a wired connection.
A block diagram of the ML algorithm is shown in Figure 4. In brief, once the EMG signals were sensed, they were amplified and transmitted to the Fusion Port. The data were then passed from the Fusion Port to the PC, where time synchronization was performed to correct for any lost data packets during transmission. The signals were then band-pass filtered between 20 Hz and 500 Hz to remove motion artifacts, followed by spatial filtering to reduce crosstalk between channels. A proprietary algorithm was subsequently used to extract both time-domain and frequency-domain features from the EMG data. During the training phase, these features were mapped to specific gestures, enabling the development of a trained model that could recognize intended gestures based on incoming EMG features. In the final stage, post-processing was applied to reduce transient misclassifications and enhance overall system reliability.
After the subjects were fully instrumented, they started a training session that lasted ~30 min. During this session, the subjects were first instructed to perform eleven distinct gestures in three distinct postures (Figure 5) with an objective to train the machine learning (ML) algorithm. The three postures included (i) arm straight down, parallel to the body, referred to as ‘Arm Parallel’; (ii) arm perpendicular to the body, referred to as ‘Arm Perpendicular’; and (iii) arm bent at the elbow and hand pointing to the sky, referred to as ‘Arm Right-Angled’. The instructor-led training session used an interface to prompt the user to perform the gestures at the correct time. The gesture was initiated in the Arm Parallel posture and maintained as the subject transitioned into other two postures. During the training session, the desktop prosthesis was disconnected from the desktop computer.
Once the algorithm had been trained on all eleven gestures and in three postures, the desk-mounted prosthesis was connected to the control system, and the subjects were required to perform the various gestures in all three postures. The motion and gestures performed by the desk-mounted prosthesis were monitored. The reproduction of the gesture by the desk-mounted prosthesis was graded by assigning letter grades A, B, or C. The letters signified the varying fidelity of gesture execution as follows:
A: Gesture reproduced accurately;
B: Gesture reproduced with minor difficulty (e.g., digit mislocated initially but finally located in the right location);
C: Gesture reproduced with major difficulty (e.g., multiple attempts before correct gesture is attained).
In situations where clear grade assignment was ambiguous, agreement of two investigators was used to arrive at the final grade. The gesture accuracy was calculated in two different ways by assigning different weights to the gesture execution grades: A = 1.0, B = 0.75, and C = 0.5. The following formulae were used for the accuracy computation:
Gesture   Accuracy   ( GA )   =   G e s t u r e s   a c c u r a t e l y   p e r f o r m e d   w i t h   G r a d e   A T o t a l   n u m b e r   o f   G e s t u r e s   p e r f o r m e d   i n   a l l   p o s t u r e s
Modified Gesture Accuracy (MGA) =
((Gestures with Grade A) + (0.75 * Gestures with Grade B)+(0.5 * Gestures with Grade C))/(Total number of Gestures performed in all postures)
The gesture accuracy was also computed by observing the output of the ML classifying algorithm and is referred to as Classifier Gesture Accuracy (CGA). The ML classifier produced decisions on 128 sample data windows from 32 channels with data collected at a sampling rate of 1250 Hz per channel. Proprietary features were computed on each window of data. The classifier produced 25 decisions per second. The participant performed each gesture in each posture for 5 s. This produced approximately 125 decisions per gesture. To calculate the classifier gesture accuracy (CGA) for a given posture, the number of correct classifier inferences was divided by the total number of decisions produced for a specific gesture in each posture:
Classifier   Gesture   Accuracy   ( CGA )   =   N u m b e r   o f   c o r r e c t   d e c i s i o n s T o t a l   n u m b e r   o f   d e c i s i o n s
The performance of the algorithm was also assessed by constructing confusion matrices for all the postures combined and for each individual posture. These matrices were constructed using classifier gesture accuracies (CGAs) for each gesture. The gesture accuracy and misclassifications for each gesture were averaged across all the participants to produce the confusion matrices.
We also calculated the latency in gesture detection by the Phantom X algorithm. For this calculation, EMG signals from all the channels were superimposed, and latency was measured from the onset of EMG activity to the point of gesture detection by the algorithm. This latency calculation does not include the delay introduced by the prosthetic hand in physically executing the gesture.
At the end of the study session, the amputee subjects were required to complete a user satisfaction survey rating their satisfaction with the performance of Phantom X system relative to their current prosthesis, if they had one, on a scale from 1 to 5 (5 being most satisfied). They were also asked whether they would use the prosthesis for more, less, or the same amount of time if the prosthesis was equipped with the Phantom X control system. They also had an option to leave any additional descriptive comments.
Wherever applicable means were compared using one-way ANOVA. A significance level of α = 0.05 was set, and analyses were conducted in Microsoft Excel.

3. Results

3.1. Gesture Accuracy for All Postures and Individual Postures

This study enrolled 11 subjects over a duration of approximately six months, of which two subjects (18%) had amputation at the transradial level, and the remaining subjects (n = 9; 82%) were able-bodied. Seven subjects (64%) were male, and four subjects (36%) were female. Of the two amputee subjects, one was male and another was female. The GA, MGA, and CGA for each subject are shown in Table 1. The average (± SD) GA, MGA, and CGA for the entire cohort were 89.0 ± 6.8%, 96.8 ± 2.0%, and 93.6% ± 4.1%, respectively, with GA being the lowest, MGA being the highest, and CGA falling in between (p = 0.002). Both GA and MGA were highly correlated with CGA (r = 0.99 for both GA vs. CGA and MGA vs. CGA).
The GA, MGA, and CGA were also calculated for each of the three postures separately and are shown in Table 2. The Arm Perpendicular posture had the lowest accuracy of all three postures, although the difference was not statistically significant from other postures for the three accuracy metrics, i.e., GA, MGA, and CGA (p > 0.15).
One amputee had a fitted DC prosthesis, while the second amputee did not use a prosthesis. The amputee with an existing prosthesis reported greater satisfaction with the Phantom X system compared to their current device and indicated they would use it more frequently if available. Both amputees provided positive feedback in the open comment section (see Appendix A).

3.2. Accuracy for Individual Gestures

Figure 6 shows the gesture accuracy for each of the 11 gestures for all three postures combined. The GA ranged from 70% to 100%, MGA from 90% to 100%, and CGA from 85% to 99%. The highest accuracy (GA = 100%, MGA = 100%, and CGA = 97%) was for the Thumbs Up gesture, and the lowest accuracy was for the Index Point gesture (GA = 70%, MGA = 90%, and CGA = 85%). The Index Tap gesture accuracy was also at the lower end of the range (GA = 79%, MGA = 92%, and CGA = 86%).
Figure 7 shows GA, MGA, and CGA for each of the 11 gestures for three postures separately. Even though the Arm Parallel posture tended to have slightly higher accuracy on average (e.g., average GA for Arm Parallel was 93% vs. 86% and 88% for Arm Perpendicular and Arm Right-Angled, respectively), this difference was not statistically significant (p > 0.3). Consistent with the results for all the postures pooled together (Figure 6), Index Tap and Index Point had the lowest accuracy of all the gestures, and the accuracy for these two gestures was particularly compromised in the Arm Perpendicular posture (e.g., GA for Index Point was 55%).

3.3. Confusion Matrices

Figure 8 presents the confusion matrices for all the postures combined, as well as for each posture individually. These matrices illustrate the extent to which one gesture was misclassified as another. For instance, when considering all the postures combined, the Open Hand gesture was misclassified as the Rest gesture 2.7% of the time.
Among all the misclassifications, those involving a gesture being mistaken for the Rest gesture were the most common. This trend was particularly pronounced for the Index Point and Index Tap gestures, with the Index Point being misclassified as Rest 8.5% of the time, and Index Tap being misclassified as Rest 9.1% of the time for all the postures combined (Figure 8a). Additionally, gestures that shared visual or functional similarities were more prone to being confused with each other. For example, Index Point was frequently misclassified as Two Count, with a misclassification rate of 6.2% observed in the Arm Right-Angled posture (Figure 8d).

3.4. Latency Between EMG Onset and Initiation of Gesture Execution

We investigated the ML algorithm-detected time delay between the onset of EMG and initiation of a gesture. Figure 9 shows a representative example of EMG signals, along with the corresponding latency in gesture detection. The delay ranged from a minimum of 68.0 ms to a maximum of 900.8 ms, with a mean ± SD of 250.5 ± 145.9 ms in a total of n = 120 latency measurements in 11 subjects. The latency for one gesture (Open Hand) in one subject could not be computed because the EMG was too noisy, hence its onset could not be ascertained accurately.
Figure 10 shows the distribution of latency times. Note that the vast majority of latency times (91%) were less than 500 ms.
The latency for various gestures was also investigated separately and is shown in Figure 11. The average latency was longest for the Key Hold gesture (364.4 ± 252.6 ms), followed by Index point gesture (291.0 ± 163.8 ms). The Key Hold gesture also showed the largest variation among subjects with a large standard deviation.

4. Discussion

The results of our ASCENT study demonstrate the potential effectiveness of the Phantom X system in improving myoelectric prosthesis control through its multi-electrode EMG sensor array and ML algorithm. The system exhibited high overall accuracy in executing a broad range of gestures, with an average GA of 89.0%, MGA of 96.8%, and CGA of 93.6%. These findings indicate that the Phantom X control system can provide highly accurate gesture recognition, even in a non-invasive testing environment. The strong correlation between GA, MGA, and CGA suggests that the system’s machine learning algorithm reliably translates EMG signals into control commands for the prosthesis.
The Gesture Accuracy (GA) only included perfectly executed gestures. When partially executed gestures were included, the Modified Gesture Accuracy (MGA) increased from a GA of 89.0% to an MGA of 96.8%. These results, achieved with minimal training, likely represent a conservative estimate of the system’s potential performance. It is noteworthy to mention that the signal-to-noise ratio is expected to be significantly higher with the implantable electrodes because the amplitude of sensed EMG signals scales with the reciprocal of the square of the distance between the source (i.e., muscle) and electrodes [26,27]. Furthermore, an additional improvement in performance is anticipated as users gain experience with the Phantom X system. Thus, the performance observed in the ASCENT study is likely a lower limit of the performance the Phantom X system with implantable electrodes might demonstrate in a real-world at-home setting. However, the ASCENT study only included two amputee participants, with most subjects being able-bodied, which limits the generalizability of the findings to the broader amputee population.
A notable observation from this study was the variation in gesture accuracy across different arm postures. The Arm Perpendicular posture exhibited the lowest accuracy, although the difference was not statistically significant. This may be attributed to biomechanical factors, such as muscle recruitment variations and changes in skin–electrode contact due to arm positioning. Future iterations of the Phantom X system could incorporate adaptive algorithms that account for posture-related variations to further enhance performance. It is worth noting that the use of implanted electrodes, as would be the case for Phantom X system, has been shown to mitigate problems related to limb positioning without training for all possible positions [28].
Gesture-specific analysis revealed that most gestures had high recognition rates, with Thumbs Up achieving the highest accuracy (GA = 100%). However, certain gestures, such as Index Point and Index Tap, exhibited lower accuracy (GA of 70% and 79% GA, respectively). These gestures were particularly challenging in the arm perpendicular posture, where Index Point accuracy dropped to 55%. The reasons for lower accuracy for these gestures and deterioration in the Arm Perpendicular posture may be several. First, these gestures involve isolated index finger movement and tend to produce more localized and lower-amplitude EMG signals, which may be harder to distinguish from baseline or other gestures. Second, the fact that surface electrodes are more susceptible to crosstalk may make the small differences associated with these gestures more difficult to discern [29,30]. Finally, the arm posture has been shown to significantly affect EMG pattern recognition performance due to changes in muscle length, orientation, and electrode–skin impedance [31,32]. Specifically, the Arm Perpendicular posture may introduce mechanical shifts in the muscle belly and residual limb geometry, leading to reduced signal fidelity or altered spatial activation patterns.
The confusion matrices provided deeper insight into gesture misclassification patterns, revealing that the majority of the misclassifications involved gestures being misclassified as Rest. Gestures that engaged minimal muscle mass, such as Index Point and Index Tap, which involve activation of only one or two fingers, tend to produce lower-amplitude EMG signals. As discussed above, this likely led to missed detections of finger activity, resulting in these gestures being more frequently classified as Rest. Misclassification was also more common between gestures that were visually and functionally similar. For instance, confusion between Index Point and Two Count occurred relatively often, with each being misidentified as the other at a higher rate compared to dissimilar gesture pairs.
The lower accuracy observed for certain gestures, such as Index Point and Index Tap, along with the frequent misclassification of gestures that are functionally and visually similar suggest a need for further refinement in signal decoding to improve robustness across postures. One potential strategy is to incorporate additional training data along with clearer and more standardized user instructions to enhance model performance. Alternatively, integrating advanced signal processing techniques may help better distinguish between overlapping muscle activations and improve gesture classification accuracy. Finally, EMG signals from smaller and deeper muscles are expected to be significantly stronger when recorded with implanted electrodes compared to surface electrodes, which could improve the classification accuracy of gestures that rely on these lower-mass muscle groups.
Latency analysis showed a mean response time of 250.5 ms, with the vast majority of gestures initiated within 500 ms. While this latency is within an acceptable range for real-time prosthesis control, minimizing delays could further enhance the user experience. The longest average latency was observed for the Key Hold gesture (364.4 ms), followed by Index Point (291.0 ms). These findings may be indicative of the complexity of muscle activation patterns required for certain gestures or similarities in muscle activation patterns relative to other gestures, necessitating further optimization of signal processing and classification algorithms for certain specific gestures. Moreover, the participants were not instructed to perform the gestures at a particular speed. Therefore, the rate at which a participant contracted their muscles to achieve a gesture could have greatly affected the calculated latency between EMG onset and the detection of the correct gesture and likely contributed to the observed spread in latency times. Finally, additional factors likely contributed to the variability in latency times, including the signal-to-noise ratio (SNR) of the recorded EMGs, which may have been affected by the quality of surface electrode contact with the skin, and the pattern and discriminability of the EMG signals, which influence the processing time required by the algorithm to classify gestures accurately. It is worth noting that our algorithm was designed using fully executed gestures, which involved longer EMG segments. An alternative approach would be to use partially executed gestures and shorter EMG segments, which could reduce latency. However, this would likely come at the cost of reduced algorithm performance and lower classification accuracy. From a practical, real-world task performance perspective, delays in the range of 300–400 ms appear to be acceptable and may not significantly impact functional performance [33,34,35]. It is possible that improvement in the signal-to-noise ratio (SNR) with Phantom X’s implantable sensor array may reduce algorithm processing time for gesture detection and narrow the distribution of latency, thereby minimizing the occurrence of prolonged delays.
Comparing the performance of the Phantom X system in the ASCENT study to gesture control outcomes reported in the literature for transradial amputees with direct control (DC) and more advanced pattern recognition systems is essential and valuable. However, data specifically addressing gesture accuracy in two-electrode DC systems, which are currently the standard of care in myoelectric control, are limited. Studies that report such metrics highlight significant limitations of DC systems, including their constrained ability to support multiple gestures without mode switching and their cognitively demanding operation. For instance, Kuiken et al. [36] evaluated gesture accuracy in three transradial amputees, reporting results ranging from 84% to 91%. Other studies have assessed functional outcomes using functional assessment instruments, such as the Assessment of Capacity of Myoelectric Control (ACMC) [37,38] and the Southampton Hand Assessment Procedure (SHAP) [39] but did not include gesture accuracy metrics for direct comparison with the ASCENT study [40]. Nonetheless, DC systems face challenges due to their non-intuitive, time-consuming, and cognitively taxing nature. Despite this, Phantom X achieves comparable accuracy to DC systems while offering a more extensive gesture set, greater ease of use, and reduced cognitive burden for the user.
When comparing Phantom X to newer pattern-recognition systems, which are not yet widely adopted as standard care, data remain sparse. Simon et al. [19] reported an average gesture accuracy of 81.5% (range: 75% to 95%) for eight patients performing up to six gestures in an at-home setting using a control system called Complete Control System (Coapt LLC, Chicago, IL, USA). However, it is important to note that the ASCENT study was conducted in an in-office environment, making direct comparisons difficult. In another in-office study using a research-grade system (Cyberhand [41] combined with a pattern-recognition algorithm), Cipriani et al. [42] reported 79% accuracy for seven hand gestures in seven transradial amputees. Additional research with non-commercial systems and various classifier techniques has reported classification accuracy ranging from 60% to 96% [43,44,45]. These findings underscore the need for standardized benchmarks while highlighting Phantom X’s potential to outperform traditional and MPR systems using sEMG in terms of accuracy and reliability.
As our study predominantly included able-bodied participants, generalizing the findings to a broader amputee population is limited. Muscle activation in transradial upper limb amputees differs from that of able-bodied individuals due to anatomical changes, altered load distribution, and compensatory strategies. Amputees often rely on residual forearm and proximal muscles, which may exhibit atypical activation patterns or co-contractions [43,44,45]. Furthermore, EMG signal quality in amputees can be more variable due to changes in muscle mass, scarring, and altered limb geometry [32]. However, there are also similarities; for instance, amputees retain voluntary control over their residual muscles and, with appropriate training, can generate activation patterns that closely approximate those of able-bodied individuals [46].
Finally, the ASCENT study evaluated the performance of the Phantom X system in a controlled, in-office setting. Thus, its performance in real-world, ambulatory environments remains unknown. Future studies will evaluate the performance of the Phantom X system with an implantable array in a larger cohort of the amputee patient population using longitudinal study designs, where participants will use the system in an at-home setting over extended durations (e.g., several weeks to months).

5. Limitations

Despite the promising accuracy results, it is important to acknowledge the limitations of our study. First, this study utilized non-invasive surface EMG electrodes rather than the fully implanted Phantom X system. While the surface electrodes provided a useful approximation of expected performance, they are subject to signal degradation due to skin impedance, movement artifacts, and crosstalk from adjacent muscles. Transition to the fully implantable system is expected to mitigate these issues, likely improving signal fidelity and, consequently, control accuracy.
Second, this study was conducted in a relatively controlled in-office environment, which does not fully replicate real-world conditions, where users may encounter varying levels of physical activity, environmental noise, and prolonged usage challenges. Future studies incorporating long-term, at-home testing will be important in assessing the real-world usability and durability of the Phantom X system.
Third, the sample size was relatively small (n = 11), with two amputee participants. While the high accuracy observed in both the able-bodied and amputee participants is promising, as discussed above, larger studies with a greater representation of amputees are needed to validate the system’s effectiveness in its target user population.
Finally, while gesture accuracy for the two amputees of different genders was among the highest observed across all the subjects, variations in residual limb anatomy and muscle recruitment patterns among a broader amputee population may introduce unique challenges that warrant further investigation.

6. Conclusions

In conclusion, this study provides evidence that the Phantom X system has the potential to significantly enhance myoelectric prosthesis control. Its high gesture accuracy and reasonable latency times suggest that it could offer a more intuitive and reliable user experience compared to existing sEMG-based prostheses control systems.
Future work will include additional testing of the Phantom X system using the cutaneous version in its final form factor, targeting a broader amputee population and incorporating endpoints that better reflect real-world utility, such as performance in activities of daily living rather than gesture accuracy alone. A new study, titled EMPOWER (NCT07071584), is planned to accomplish this. Following EMPOWER, clinical trials with the fully implantable system, larger participant cohorts, and real-world testing will be essential to further validate its clinical utility and address any remaining technical challenges.

7. Trial Registration

This study was registered in the ClinicalTrials.gov database (NCT06446037; date of registration: 28 May 2024).

Author Contributions

All the authors contributed to the main manuscript text. V.S., E.L., M.F. and C.G. contributed to the study design and protocol development. V.S. and E.L. conducted the study and gathered and analyzed study data. All authors have read and agreed to the published version of the manuscript.

Funding

This study was supported by Phantom Neuro Inc.

Institutional Review Board Statement

This study was approved by the Biomedical Research Alliance of New York (BRANY) institutional review board (No. 24-02-222-1716), 2 February 2024, and all the subjects volunteered to participate and signed informed consent forms.

Informed Consent Statement

Informed consent was obtained from all the subjects involved in this study.

Data Availability Statement

All the data generated or analyzed during this study are included in this published article.

Acknowledgments

We would like to thank Erik Larson, Masoud Farshchian, and Chris Reiger of Phantom Neuro Inc. for building the necessary hardware and contributing to the ML algorithm evaluated in this study.

Conflicts of Interest

V.S., E.L., M.F. and C.G. are employees of Phantom Neuro Inc. and receive salaries and have equity in the company. M.O.-C. received consulting fee from Phantom Neuro Inc.

Abbreviations

The following abbreviations are used in this manuscript:
ACMCAssessment of Capacity of Myoelectric Control
DCDirect Control
EMGElectromyography
GAGesture Accuracy
MGAModified Gesture Accuracy
CGAClassifier Gesture Accuracy
MLMachine Learning
SDStandard Deviation
sEMGSurface Electromyography
SHAPSouthampton Hand Assessment Procedure
SNRSignal-to-Noise ratio
USBUniversal Serial Bus

Appendix A

Comment provided by the first amputee subject :
I would wear the prosthesis with the Phantom X about the same amount, but I would ‘use’ the functions of the hand probably 3× as much each day.
Comment provided by the second amputee subject:
I am unlikely to use a prosthesis without a Phantom X.

References

  1. World Health Organization. Musculoskeletal Health. 2022. Available online: https://www.who.int/news-room/fact-sheets/detail/musculoskeletal-conditions (accessed on 8 April 2025).
  2. McDonald, C.L.; Westcott-McCoy, S.; Weaver, M.R.; Haagsma, J.; Kartin, D. Global prevalence of traumatic non-fatal limb amputation. Prosthet. Orthot. Int. 2021, 45, 105–114. [Google Scholar] [CrossRef] [PubMed]
  3. Rivera, J.A.; Churovich, K.; Anderson, A.B.; Potter, B.K. Estimating Recent US Limb Loss Prevalence and Updating Future Projections. Arch. Rehabil. Res. Clin. Transl. 2024, 6, 100376. [Google Scholar] [CrossRef] [PubMed]
  4. Ison, M.; Artemiadis, P. The role of muscle synergies in myoelectric control: Trends and challenges for simultaneous multifunction control. J. Neural Eng. 2014, 11, 051001. [Google Scholar] [CrossRef]
  5. Farina, D.; Enoka, R.M. Evolution of surface electromyography: From muscle electrophysiology towards neural recording and interfacing. J. Electromyogr. Kinesiol. 2023, 71, 102796. [Google Scholar] [CrossRef]
  6. Resnik, L.J.; Borgia, M.L.; Clark, M.A. Prevalence and predictors of unmet need for upper limb prostheses: An observational cohort study. J. Prosthet. Orthot. 2023, 36, 80–88. [Google Scholar] [CrossRef]
  7. Salminger, S.; Stino, H.; Pichler, L.H.; Gstoettner, C.; Sturma, A.; Mayer, J.A.; Szivak, M.; Aszmann, O.C. Current rates of prosthetic usage in upper-limb amputees—Have innovations had an impact on device acceptance? Disabil. Rehabil. 2022, 44, 3708–3713. [Google Scholar] [CrossRef]
  8. Gasparini, E.; Cerqueira, R.; Preziuso, C.; Angelini, L.; Guachi, R.; Controzzi, M.; Cipriani, C.; Mastinu, E. LimbMATE: A Versatile Platform for Closed-Loop Research in Prosthetics. IEEE Trans. Neural Syst. Rehabil. Eng. 2025, 33, 2112–2122. [Google Scholar] [CrossRef]
  9. Lara, J.E.; Cheng, L.K.; Rohrle, O.; Paskaranandavadivel, N. Muscle-Specific High-Density Electromyography Arrays for Hand Gesture Classification. IEEE Trans. Biomed. Eng. 2022, 69, 1758–1766. [Google Scholar] [CrossRef]
  10. Montazerin, M.; Rahimian, E.; Naderkhani, F.; Atashzar, S.F.; Yanushkevich, S.; Mohammadi, A. Transformer-based hand gesture recognition from instantaneous to fused neural decomposition of high-density EMG signals. Sci. Rep. 2023, 13, 11000. [Google Scholar] [CrossRef]
  11. Jaramillo-Yánez, A.; Benalcázar, M.E.; Mena-Maldonado, E. Real-Time Hand Gesture Recognition Using Surface Electromyography and Machine Learning: A Systematic Literature Review. Sensors 2020, 20, 2467. [Google Scholar] [CrossRef]
  12. Kadavath, M.R.K.; Nasor, M.; Imran, A. Enhanced Hand Gesture Recognition with Surface Electromyogram and Machine Learning. Sensors 2024, 24, 5231. [Google Scholar] [CrossRef]
  13. Pradhan, A.; He, J.; Jiang, N. Multi-day dataset of forearm and wrist electromyogram for hand gesture recognition and biometrics. Sci. Data 2022, 9, 733. [Google Scholar] [CrossRef] [PubMed]
  14. Simpetru, R.C.; Marz, M.; Del Vecchio, A. Proportional and Simultaneous Real-Time Control of the Full Human Hand From High-Density Electromyography. IEEE Trans. Neural Syst. Rehabil. Eng. 2023, 31, 3118–3131. [Google Scholar] [CrossRef] [PubMed]
  15. Ottobock. MyoPlus. Available online: https://shop.ottobock.us/Prosthetics/Upper-Limb-Prosthetics/Myo-Plus/c/2901 (accessed on 25 July 2025).
  16. Chen, Z.; Min, H.; Wang, D.; Xia, Z.; Sun, F.; Fang, B. A Review of Myoelectric Control for Prosthetic Hand Manipulation. Biomimetics 2023, 8, 328. [Google Scholar] [CrossRef] [PubMed]
  17. IBT Glide. Available online: https://www.i-biomed.com/glide (accessed on 25 July 2025).
  18. Hargrove, L.J.; Miller, L.A.; Turner, K.; Kuiken, T.A. Myoelectric Pattern Recognition Outperforms Direct Control for Transhumeral Amputees with Targeted Muscle Reinnervation: A Randomized Clinical Trial. Sci. Rep. 2017, 7, 13840. [Google Scholar] [CrossRef]
  19. Simon, A.M.; Turner, K.L.; Miller, L.A.; Potter, B.K.; Beachler, M.D.; Dumanian, G.A.; Hargrove, L.J.; Kuiken, T.A. User Performance With a Transradial Multi-Articulating Hand Prosthesis During Pattern Recognition and Direct Control Home Use. IEEE Trans. Neural Syst. Rehabil. Eng. 2023, 31, 271–281. [Google Scholar] [CrossRef]
  20. Ostlie, K.; Franklin, R.J.; Skjeldal, O.H.; Skrondal, A.; Magnus, P. Musculoskeletal pain and overuse syndromes in adult acquired major upper-limb amputees. Arch. Phys. Med. Rehabil. 2011, 92, 1967–1973.e1. [Google Scholar] [CrossRef]
  21. Rosca, A.C.; Baciu, C.C.; Burtaverde, V.; Mateizer, A. Psychological Consequences in Patients With Amputation of a Limb. An Interpretative-Phenomenological Analysis. Front. Psychol. 2021, 12, 537493. [Google Scholar] [CrossRef]
  22. Sahu, A.; Sagar, R.; Sarkar, S.; Sagar, S. Psychological effects of amputation: A review of studies from India. Ind. Psychiatry J. 2016, 25, 4–10. [Google Scholar] [CrossRef]
  23. Ortiz-Catalan, M.; Zbinden, J.; Millenaar, J.; D’Accolti, D.; Controzzi, M.; Clemente, F.; Cappello, L.; Earley, E.J.; Mastinu, E.; Kolankowska, J.; et al. A highly integrated bionic hand with neural control and feedback for use in daily life. Sci. Robot. 2023, 8, eadf7360. [Google Scholar] [CrossRef]
  24. Vu, P.P.; Vaskov, A.K.; Irwin, Z.T.; Henning, P.T.; Lueders, D.R.; Laidlaw, A.T.; Davis, A.J.; Nu, C.S.; Gates, D.H.; Gillespie, R.B.; et al. A regenerative peripheral nerve interface allows real-time control of an artificial hand in upper limb amputees. Sci. Transl. Med. 2020, 12, eaay2857. [Google Scholar] [CrossRef]
  25. Zbinden, J.; Sassu, P.; Mastinu, E.; Earley, E.J.; Munoz-Novoa, M.; Brånemark, R.; Ortiz-Catalan, M. Improved control of a prosthetic limb by surgically creating electro-neuromuscular constructs with implanted electrodes. Sci. Transl. Med. 2023, 15, eabq3665. [Google Scholar] [CrossRef] [PubMed]
  26. Plonsey, R. The active fiber in a volume conductor. IEEE Trans. Biomed. Eng. 1974, 21, 371–381. [Google Scholar] [CrossRef] [PubMed]
  27. Rodriguez-Falces, J. A novel approach to teach the generation of bioelectrical potentials from a descriptive and quantitative perspective. Adv. Physiol. Educ. 2013, 37, 327–336. [Google Scholar] [CrossRef] [PubMed]
  28. Ortiz-Catalan, M.; Hakansson, B.; Branemark, R. An osseointegrated human-machine gateway for long-term sensory feedback and motor control of artificial limbs. Sci. Transl. Med. 2014, 6, 257re6. [Google Scholar] [CrossRef]
  29. Kong, Y.K.; Hallbeck, M.S.; Jung, M.C. Crosstalk effect on surface electromyogram of the forearm flexors during a static grip task. J. Electromyogr. Kinesiol. 2010, 20, 1223–1229. [Google Scholar] [CrossRef]
  30. Winter, D.A.; Fuglevand, A.J.; Archer, S.E. Crosstalk in surface electromyography: Theoretical and practical estimates. J. Electromyogr. Kinesiol. 1994, 4, 15–26. [Google Scholar] [CrossRef]
  31. Campbell, E.; Phinyomark, A.; Scheme, E. Current Trends and Confounding Factors in Myoelectric Control: Limb Position and Contraction Intensity. Sensors 2020, 20, 1613. [Google Scholar] [CrossRef]
  32. Young, A.J.; Smith, L.H.; Rouse, E.J.; Hargrove, L.J. Classification of simultaneous movements using surface EMG pattern recognition. IEEE Trans. Biomed. Eng. 2013, 60, 1250–1258. [Google Scholar] [CrossRef]
  33. Englehart, K.; Hudgins, B. A robust, real-time control scheme for multifunction myoelectric control. IEEE Trans. Biomed. Eng. 2003, 50, 848–854. [Google Scholar] [CrossRef]
  34. Hefftner, G.; Zucchini, W.; Jaros, G.G. The electromyogram (EMG) as a control signal for functional neuromuscular stimulation—Part I: Autoregressive modeling as a means of EMG signature discrimination. IEEE Trans. Biomed. Eng. 1988, 35, 230–237. [Google Scholar] [CrossRef] [PubMed]
  35. Blackett, C.; Fernandes, A.; Teigen, E.; Thoresen, T. Effects of signal latency on human performance in Teleoperations. In Proceedings of the International Conference on Human Interaction and Emerging Technologies, Strasbourg, France, 28–30 April 2021; Springer: Berlin/Heidelberg, Germany, 2021. [Google Scholar]
  36. Kuiken, T.A.; Miller, L.A.; Turner, K.; Hargrove, L.J. A Comparison of Pattern Recognition Control and Direct Control of a Multiple Degree-of-Freedom Transradial Prosthesis. IEEE J. Transl. Eng. Health Med. 2016, 4, 2100508. [Google Scholar] [CrossRef] [PubMed]
  37. Lindner, H.Y.; Linacre, J.M.; Norling Hermansson, L.M. Assessment of capacity for myoelectric control: Evaluation of construct and rating scale. J. Rehabil. Med. 2009, 41, 467–474. [Google Scholar] [CrossRef] [PubMed]
  38. Lindner, H.Y.; Langius-Eklof, A.; Hermansson, L.M. Test-retest reliability and rater agreements of assessment of capacity for myoelectric control version 2.0. J. Rehabil. Res. Dev. 2014, 51, 635–644. [Google Scholar] [CrossRef] [PubMed]
  39. Light, C.M.; Chappell, P.H.; Kyberd, P.J. Establishing a standardized clinical assessment tool of pathologic and prosthetic hand function: Normative data, reliability, and validity. Arch. Phys. Med. Rehabil. 2002, 83, 776–783. [Google Scholar] [CrossRef]
  40. Luchetti, M.; Cutti, A.G.; Verni, G.; Sacchetti, R.; Rossi, N. Impact of Michelangelo prosthetic hand: Findings from a crossover longitudinal study. J. Rehabil. Res. Dev. 2015, 52, 605–618. [Google Scholar] [CrossRef]
  41. Carrozza, M.C.; Cappiello, G.; Micera, S.; Edin, B.B.; Beccai, L.; Cipriani, C. Design of a cybernetic hand for perception and action. Biol. Cybern. 2006, 95, 629–644. [Google Scholar] [CrossRef]
  42. Cipriani, C.; Antfolk, C.; Controzzi, M.; Lundborg, G.; Rosén, B.; Carrozza, M.C.; Sebelius, F. Online myoelectric control of a dexterous hand prosthesis by transradial amputees. IEEE Trans. Neural Syst. Rehabil. Eng. 2011, 19, 260–270. [Google Scholar] [CrossRef]
  43. Chateaux, M.; Rossel, O.; Vérité, F.; Nicol, C.; Touillet, A.; Paysant, J.; Jarrassé, N.; De Graaf, J.B. New insights into muscle activity associated with phantom hand movements in transhumeral amputees. Front. Hum. Neurosci. 2024, 18, 1443833. [Google Scholar] [CrossRef]
  44. Tenore, F.V.; Ramos, A.; Fahmy, A.; Acharya, S.; Etienne-Cummings, R.; Thakor, N.V. Decoding of individuated finger movements using surface electromyography. IEEE Trans. Biomed. Eng. 2009, 56, 1427–1434. [Google Scholar] [CrossRef]
  45. Yadav, D.; Veer, K. Recent trends and challenges of surface electromyography in prosthetic applications. Biomed. Eng. Lett. 2023, 13, 353–373. [Google Scholar] [CrossRef]
  46. Powell, M.A.; Thakor, N.V. A Training Strategy for Learning Pattern Recognition Control for Myoelectric Prostheses. J. Prosthet. Orthot. 2013, 25, 30–41. [Google Scholar] [CrossRef]
Figure 1. Complete Phantom X System. The major components of the Phantom X system include Implantable Sensor Electrode Arrays, Fusion Port and Power Coils, Phantom Mobile App, and associated Machine Learning Algorithms that are embedded in the Fusion Port and continuously run to decode EMG signals and generate prosthesis control signals.
Figure 1. Complete Phantom X System. The major components of the Phantom X system include Implantable Sensor Electrode Arrays, Fusion Port and Power Coils, Phantom Mobile App, and associated Machine Learning Algorithms that are embedded in the Fusion Port and continuously run to decode EMG signals and generate prosthesis control signals.
Prosthesis 07 00099 g001
Figure 2. Cutaneous sensor array for sensing EMG signals. A 32-channel array made of polyamide was custom fabricated and wrapped around the circumference of a subject’s arm. See text for additional details.
Figure 2. Cutaneous sensor array for sensing EMG signals. A 32-channel array made of polyamide was custom fabricated and wrapped around the circumference of a subject’s arm. See text for additional details.
Prosthesis 07 00099 g002
Figure 3. The schematic of the experimental set-up and the process for converting EMG signals to control signals required for controlling a desk-mounted prosthesis. EMG signals were sensed by a multi-electrode array strapped on the subject’s forearm. These EMG signals are amplified and filtered by EMG hardware (EMG HW) and transmitted to another device, known as the Fusion Port. The Fusion Port ran the Phantom X machine learning (ML) control algorithm. The Fusion Port was wirelessly connected to a computer that recorded data. The output of the Phanom X algorithm was a set of control signals that were transmitted over a wired link to a desk-mounted prosthesis to articulate its various digits and generate a gesture.
Figure 3. The schematic of the experimental set-up and the process for converting EMG signals to control signals required for controlling a desk-mounted prosthesis. EMG signals were sensed by a multi-electrode array strapped on the subject’s forearm. These EMG signals are amplified and filtered by EMG hardware (EMG HW) and transmitted to another device, known as the Fusion Port. The Fusion Port ran the Phantom X machine learning (ML) control algorithm. The Fusion Port was wirelessly connected to a computer that recorded data. The output of the Phanom X algorithm was a set of control signals that were transmitted over a wired link to a desk-mounted prosthesis to articulate its various digits and generate a gesture.
Prosthesis 07 00099 g003
Figure 4. Machine Learning (ML) Algorithm block diagram. The EMG data from 32 channels were amplified, time-synchronized to correct for lost packets, and band-pass filtered to remove motion artifacts. Spatial filtering was then applied to reduce inter-channel crosstalk. The resulting signals were then used to extract both time-domain and frequency-domain features to decode intended gestures. In the final stage, a logic-based post-processing step was applied to reduce intermittent misclassifications and enhance overall system reliability.
Figure 4. Machine Learning (ML) Algorithm block diagram. The EMG data from 32 channels were amplified, time-synchronized to correct for lost packets, and band-pass filtered to remove motion artifacts. Spatial filtering was then applied to reduce inter-channel crosstalk. The resulting signals were then used to extract both time-domain and frequency-domain features to decode intended gestures. In the final stage, a logic-based post-processing step was applied to reduce intermittent misclassifications and enhance overall system reliability.
Prosthesis 07 00099 g004
Figure 5. Various gestures and postures for which system performance was assessed. Panel (a) shows eleven gestures, and Panel (b) shows three distinct postures (leftmost: Arm Parallel, middle: Arm Perpendicular, and rightmost Arm Right-Angled) for subject training and assessment.
Figure 5. Various gestures and postures for which system performance was assessed. Panel (a) shows eleven gestures, and Panel (b) shows three distinct postures (leftmost: Arm Parallel, middle: Arm Perpendicular, and rightmost Arm Right-Angled) for subject training and assessment.
Prosthesis 07 00099 g005
Figure 6. Accuracy for various gestures. GA, MGA, and CGA for each of the 11 gestures are shown separately. The accuracy ranged from 70 to 100%. The Index Point and Index Tap gestures exhibited accuracies that were towards the lower end of the range.
Figure 6. Accuracy for various gestures. GA, MGA, and CGA for each of the 11 gestures are shown separately. The accuracy ranged from 70 to 100%. The Index Point and Index Tap gestures exhibited accuracies that were towards the lower end of the range.
Prosthesis 07 00099 g006
Figure 7. Accuracy for various gestures for three postures. Panels (ac) show the GA, MGA, and CGA for Arm Parallel, Arm Perpendicular, and Arm Right-Angled postures, respectively.
Figure 7. Accuracy for various gestures for three postures. Panels (ac) show the GA, MGA, and CGA for Arm Parallel, Arm Perpendicular, and Arm Right-Angled postures, respectively.
Prosthesis 07 00099 g007
Figure 8. Confusion matrices. Panels (ad) show confusion matrix for all the postures combined, Arm Parallel, Arm Perpendicular, and Arm Right-Angled postures, respectively.
Figure 8. Confusion matrices. Panels (ad) show confusion matrix for all the postures combined, Arm Parallel, Arm Perpendicular, and Arm Right-Angled postures, respectively.
Prosthesis 07 00099 g008
Figure 9. Representative EMG recorded by Phantom X system showing gesture detection latency. The inset displays superimposed EMG signals from all 32 channels during the Open Hand gesture for a subject. The first 8 s of EMG data are expanded to highlight the latency measurement from EMG onset to gesture detection, which, in this example, was approximately 254 ms. Upon gesture detection, a command was sent to the prosthetic hand to initiate gesture execution.
Figure 9. Representative EMG recorded by Phantom X system showing gesture detection latency. The inset displays superimposed EMG signals from all 32 channels during the Open Hand gesture for a subject. The first 8 s of EMG data are expanded to highlight the latency measurement from EMG onset to gesture detection, which, in this example, was approximately 254 ms. Upon gesture detection, a command was sent to the prosthetic hand to initiate gesture execution.
Prosthesis 07 00099 g009
Figure 10. Distribution of latency time. Latency times were measured from onset of EMG and initiation of gesture execution and ranged from 68 ms to 968 ms. The vast majority of gestures (~91%) had a latency time of less than 500 ms.
Figure 10. Distribution of latency time. Latency times were measured from onset of EMG and initiation of gesture execution and ranged from 68 ms to 968 ms. The vast majority of gestures (~91%) had a latency time of less than 500 ms.
Prosthesis 07 00099 g010
Figure 11. Latency time for various gestures separately. The mean latency delays for various gestures are shown along with standard deviation as vertical bars.
Figure 11. Latency time for various gestures separately. The mean latency delays for various gestures are shown along with standard deviation as vertical bars.
Prosthesis 07 00099 g011
Table 1. Summary of gesture accuracy metrics in all study subjects.
Table 1. Summary of gesture accuracy metrics in all study subjects.
Subject #Subject TypeSubject GenderGesture
Accuracy (GA)
Modified Gesture
Accuracy (MGA)
Classifier Gesture
Accuracy (CGA)
01Able-bodiedMale100%100%98.4%
02Able-bodiedFemale87.9%96.2%89.2%
03Able-bodiedMale81.8%95.4%92.2%
04Able-bodiedMale97.0%99.2%96.2%
05Able-bodiedFemale81.8%93.2%84.4%
06Able-bodiedMale93.9%98.5%97.2%
07AmputeeMale90.9%97.7%95.1%
08Able-bodiedMale93.9%96.9%94.3%
09Able-bodiedFemale78.8%94.7%95.6%
10Able-bodiedFemale87.9%97.0%96.1%
11AmputeeFemale84.8%96.2%90.9%
Average ±
SD
89.0% ±
6.8%
96.8% ±
2.0%
93.6% ±
4.1%
Table 2. Average (±SD) accuracy for each posture separately.
Table 2. Average (±SD) accuracy for each posture separately.
Arm ParallelArm PerpendicularArm Right-Angled
GA92.6 ± 6.8%86.0 ± 11.0%88.4 ± 8.2%
MGA97.9 ± 1.9%95.9 ± 3.3%96.7 ± 2.1%
CGA94.0 ± 4.8%92.4 ± 4.7%94.4 ± 4.5%
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Sharma, V.; Lloyd, E.; Faltys, M.; Ortiz-Catalan, M.; Glass, C. Assessment of Gesture Accuracy for a Multi-Electrode EMG-Sensor-Array-Based Prosthesis Control System. Prosthesis 2025, 7, 99. https://doi.org/10.3390/prosthesis7040099

AMA Style

Sharma V, Lloyd E, Faltys M, Ortiz-Catalan M, Glass C. Assessment of Gesture Accuracy for a Multi-Electrode EMG-Sensor-Array-Based Prosthesis Control System. Prosthesis. 2025; 7(4):99. https://doi.org/10.3390/prosthesis7040099

Chicago/Turabian Style

Sharma, Vinod, Erik Lloyd, Mike Faltys, Max Ortiz-Catalan, and Connor Glass. 2025. "Assessment of Gesture Accuracy for a Multi-Electrode EMG-Sensor-Array-Based Prosthesis Control System" Prosthesis 7, no. 4: 99. https://doi.org/10.3390/prosthesis7040099

APA Style

Sharma, V., Lloyd, E., Faltys, M., Ortiz-Catalan, M., & Glass, C. (2025). Assessment of Gesture Accuracy for a Multi-Electrode EMG-Sensor-Array-Based Prosthesis Control System. Prosthesis, 7(4), 99. https://doi.org/10.3390/prosthesis7040099

Article Metrics

Back to TopTop