Next Article in Journal
Innovative Use of Ultra-Low-Frequency Dynamic Electronic Impulses for Sustainable Performance of Drippers Applying Produced Water
Previous Article in Journal
Adaptive CNN Ensemble for Apple Detection: Enabling Sustainable Monitoring Orchard
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Development of a Portable LED-Based Photometer for Quality Assessment of Red Palm Oil in SMEs

by
Kamonpan Wongyai
1,2,
Suttirak Kaewpawong
1,2,
Karaket Wattanasit
1,2,
Dhammanoon Srinoum
1,2,
Mudtorlep Nisoa
1,2,3,*,
Parawee Rattanakit
1,2,3,
Arlee Tamman
4 and
Dheerawan Boonyawan
5
1
Functional Materials and Nanotechnology Center of Excellence, Walailak University, Nakhon Si Thammarat 80160, Thailand
2
Hub of Knowledge in Microwave Heating and Applications, Walailak University, Nakhon Si Thammarat 80160, Thailand
3
School of Science, Walailak University, Nakhon Si Thammarat 80160, Thailand
4
Center of Excellence in Engineering and Nuclear Technology, Thailand Institute of Nuclear Technology, Nakhon Nayok 26120, Thailand
5
Department of Physics and Materials Science, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand
*
Author to whom correspondence should be addressed.
AgriEngineering 2025, 7(11), 370; https://doi.org/10.3390/agriengineering7110370
Submission received: 10 September 2025 / Revised: 22 October 2025 / Accepted: 27 October 2025 / Published: 3 November 2025
(This article belongs to the Section Sustainable Bioresource and Bioprocess Engineering)

Abstract

This study presents the development of a portable DOBI meter prototype designed for the rapid, low-cost evaluation of crude red palm oil (RPO) quality. The device employs two narrow-spectrum LEDs (UV at 269 nm and visible at 446 nm) as light sources, paired with a broadband photodiode (PD) detector to measure light absorption in a quartz cuvette containing 95% hexane-diluted oil samples. Dedicated LED driver circuits, a PD receiver module, and microcontroller-based data acquisition and display systems were integrated into a compact enclosure. Calibration procedures involved the measurement of LED emission spectra and PD responses, followed by standard curve generation using known RPO concentrations. The results from the DOBI meter were validated against a commercial spectrophotometer (Merck Prove 600), demonstrating high accuracy with less than 5% deviation. Further analysis of RPO extracted from microwave-treated mesocarps showed consistent DOBI values and carotenoid concentrations across both instruments. The developed device offers a reliable, accessible alternative for assessing palm oil quality, particularly in field or small-scale industrial settings.

1. Introduction

Red palm oil, as illustrated in Figure 1, is derived from the mesocarp of the oil palm fruit (Elaeis guineensis) [1,2,3,4]. It is a rich source of valuable phytochemicals, including carotenoids, vitamin E, phytosterols, coenzyme Q10, polyphenols, and phenolic acids. Among these, carotenoids are particularly important due to their antioxidant properties and their role as precursors for vitamin A biosynthesis. These compounds support visual health, enhance visual acuity, and help reduce the risk of cardiovascular diseases. Vitamin E, comprising both tocopherols and tocotrienols, also plays a vital role through its anti-inflammatory properties and contributions to neuroprotection, cardio protection, hepatoprotection, and anticancer activity.
Red palm oil is produced at various scales—industrial, community, and laboratory levels [4,5,6,7,8]. In industrial settings, it is refined from crude palm oil using specialized processes aimed at maximizing the retention of carotenoids and vitamin E. In contrast, community-level production, especially in African countries, typically involves boiling the palm fruit followed by mechanical pounding and oil extraction [8].
Since 2007, laboratory research in Malaysia, Indonesia, and Thailand has explored the use of microwave technology for red palm oil production. This technique employs rapid heating to inactivate lipase enzymes, rupture oil-containing cells, and facilitate the release of triglycerides from the mesocarp fibers. The resulting oil is characterized by low free fatty acid (FFA) content and high concentrations of carotenoids and vitamin E [9,10,11,12,13,14].
The chemical composition of red palm oil is typically analyzed using UV-visible spectrophotometry [15,16,17,18,19,20,21,22,23,24,25], which measures the absorption of light by specific compounds. This method helps determine the deterioration of bleachability index (DOBI) and carotenoid content. Each chemical constituent exhibits characteristic absorbance patterns, and the absorbance level is proportional to its concentration. Traditional UV-visible spectrophotometers, which use gas discharge tubes as light sources and sensors capable of detecting a wide range of wavelengths, are effective but often expensive, complex, and require significant maintenance.
In recent years, light-emitting diode (LED) and photodiode (PD) technologies have emerged as promising alternatives to conventional light sources in spectrophotometry [26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48]. These diode-based systems offer numerous advantages, including compactness, durability, lower energy consumption, extended service life, and the ability to emit or detect specific wavelengths. As a result, LED/PD-based systems have been increasingly adopted in portable instruments for health, environmental, and food safety applications, especially for field use due to their affordability and ease of use.
The Plasmas and Electromagnetic Wave Center of Excellence (PEwave) at Walailak University has been developing microwave heating technologies for diverse applications, such as plasma generation, drying, and extraction of agricultural products, since 2006 [49,50,51,52,53,54,55,56,57]. In 2020, PEwave initiated the development of microwave-based technologies for small-scale red palm oil production in Thailand [14]. The project aims to construct a prototype facility capable of processing 1–2 tons of fresh fruit bunches and yielding 150–300 kg of red palm oil per day, with a budget under THB 10 million. The effort also focuses on quality monitoring and optimization of production processes to preserve carotenoids and ensure desirable DOBI values.
To support this initiative, a portable DOBI and carotenoid meter has been developed using UV and visible LEDs in combination with photodiodes. This compact, user-friendly, and cost-effective device serves as an alternative to conventional UV-visible spectrophotometers. It demonstrates significant potential for community-level producers, particularly those with limited capital, by enabling rapid on-site assessment of red palm oil quality.

2. UV and Visible Light Absorption Spectra of Red Palm Oil and It Quality Measurement

Figure 2 presents the absorption spectrum of red palm oil, measured using a spectrophotometer (Spectroquant Prove 600, Merck KGaA, Darmstadt, Germany) across the ultraviolet (UV) to visible light range. The absorption values correspond to the concentrations of various chemical constituents. According to the Beer–Lambert Law, higher absorbance indicates higher substance concentration. The absorption peak between 210–230 nm is attributed to oleic and linoleic acids, which are unsaturated fatty acids containing one and two double bonds, respectively. These fatty acids account for approximately 50% of crude palm oil content [1,2]. Meanwhile, the absorption range between 260–280 nm is associated with ketones and aldehydes, which are secondary oxidation by-products [58,59].
Carotenoids, key indicators of red palm oil quality, exhibit strong absorption in the range of 400–500 nm [59]. These spectral characteristics form the basis for determining the deterioration of bleachability index (DOBI) and carotenoid concentrations, as shown in Equations (1) and (2) [18,24]:
D O B I = A 446 A 269
C a r o t e n o i d p p m = 383 A 446 V 100   W
In these equations, A 446 and A 269 refer to the absorbance at 446 nm (due to carotenoids) and 269 nm (due to aldehydes), respectively. The value 383 is the specific extinction coefficient for carotenoids in hexane. V is the volume of hexane in liters (L), and W is the mass of the red palm oil sample in grams (g).
The DOBI value serves as an indicator of oil quality, reflecting the ripeness and freshness of the processed palm fruit. A high DOBI value implies that the fruits are optimally ripe, rich in carotenoids, and have undergone minimal oxidative degradation, resulting in low aldehyde content. Thus, simultaneous analysis of absorbance at 269 nm and 446 nm provides a rapid and effective means of assessing red palm oil quality.

3. Instrumentation

Figure 3 illustrates the working principle of the DOBI meter. The device uses light-emitting diodes (LEDs) to emit light and photodiodes (PDs) to detect the intensity of the transmitted light. It is designed to measure the absorbance properties of red palm oil solutions, which are directly related to their concentration c . When a constant electric current I L E D is supplied to an LED, it emits light at a specific wavelength λ and initial intensity I 0 . As the light passes through a sample with concentration c , part of it is absorbed, reducing the intensity to I . This reduction in intensity is detected by the photodiode, which generates an electric current I P D proportional to the transmitted light I . A signal conditioning circuit then converts this current into a voltage signal V , which is subsequently digitized by an analog-to-digital converter (ADC).
The DOBI meter consists of several key components, as shown in Figure 4. It includes a power supply unit that accepts a 9–12 V DC input and provides regulated voltages of +12 V, +5 V, and −5 V to power the LED circuit, PD circuit, and microcontroller, respectively. The system employs two LEDs—designated LED269 and LED446—emitting light at wavelengths of 269 nm and 446 nm, respectively. When forward biased with a steady current from the LED circuit, each LED generates a consistent light output that increases with the supplied current.
The emitted light passes through a palm oil sample contained within a quartz cuvette. Absorption occurs primarily due to aldehydes (at 269 nm) and carotenoids (at 446 nm), leading to a reduction in light intensity. Two photodiodes—PD269 and PD446—are positioned to measure the transmitted light at the corresponding wavelengths. The signals from the PDs are amplified by the PD circuit to generate suitable voltage levels. These analog signals are then digitized by a 16-bit ADC and processed by a microcontroller. Finally, the calculated DOBI values and carotenoid concentrations are displayed on a digital screen.

3.1. UV and Visible LEDs

Light-emitting diodes (LEDs) are electronic devices composed of semiconductor materials with a P–N junction structure. When an electric current flows through the junction, electron–hole recombination occurs, resulting in the emission of light [60]. LEDs can emit light across a broad spectral range—from deep ultraviolet (UV) to far infrared—spanning wavelengths from approximately 255 to 4600 nm [31,41]. Each diode is characterized by a single emission wavelength with a typical spectral bandwidth or full width at half maximum (FWHM) of 10–20 nm. LEDs that operate in the UV region (λ < 350 nm) tend to be more expensive due to their fabrication using wide bandgap semiconductor materials such as aluminum gallium nitride (AlGaN). LEDs are widely used as light sources in spectrophotometric analysis of chemical components in both liquids and gases, utilizing principles of molecular absorption and fluorescence spectroscopy [31]. Compared to conventional incandescent or discharge lamps, LEDs offer numerous advantages, including compact size, low power consumption, long operational lifespan, cost-effectiveness, and emission at specific wavelengths. Furthermore, because LEDs emit narrowband light, they eliminate the need for a monochromator in many photometric systems. Currently, LEDs are increasingly employed as light sources in high-performance liquid chromatography (HPLC) [28], capillary liquid chromatography, portable field-analysis instruments [30], and compact photometers used for chemical analysis [34].
In this study, we used a UV LED (model SWDC-T306-DNN-U1930) manufactured by Harvatek Corporation (Hsinchu City, Taiwan) [61]. This device emits UV light with a peak wavelength of 269 nm, an optical power density of 0.8 mW/cm2, and a forward voltage of 5–6 V. It has an operational lifespan exceeding 7000 h and was procured via the online platform AliExpress. Additionally, a visible LED (model MTE4600N), produced by Marktech Optoelectronics [62], was employed. This LED emits at a peak wavelength of 450 nm, delivers an optical output power of 43.2 mW at 50 mA, and has a forward voltage of 2.8 V. It was obtained from Mouser Electronics, an online electronics retailer. These two LEDs are shown in Figure 5a and Figure 5b, respectively.

3.2. Photodiode

Photodiodes (PDs) are used to measure the intensity of light passing through a solution, thereby enabling the determination of light absorption by the sample. As the sample absorbs part of the incident light, the remaining transmitted light intensity I is detected by the PD. Photodiodes are widely used as light sensors due to their fast response time, low voltage requirements, compact size, long operational lifespan, and cost-effectiveness [26,31,43]. Structurally, photodiodes are composed of N-type and P-type semiconductor layers, similar to light-emitting diodes, but include a lens that focuses incident light onto the PN junction. When light interacts with the junction, it generates electron–hole pairs, thereby increasing the conductivity of the photodiode. The resulting photocurrent is directly proportional to the intensity of the incident light and varies with the wavelength. Silicon-based photodiodes typically respond to a broad wavelength range, spanning from 190 to 1100 nm.
In this study, we used the PC10-2-TO5 photodiode model (Silicon Sensor International AG, Berlin, Germany), which features an active area of 3.57 mm in diameter and operates across a wavelength range of 200–1100 nm [63]. It supports temperatures from −40 to 100 °C, has a responsivity of 0.17–0.42 A/W depending on the wavelength, and provides up to 10 mA of current output. The photodiode is packaged in a TO-5 metal can, as shown in Figure 6a, which offers mechanical protection and stability for consistent optical performance. In this work, we specifically used the fused silica window type (vendor part number P/N 03-146), which exhibits higher responsivity in the UV region. Its characteristics are illustrated in Figure 6b, where two types of window materials—UV-transmitting glass and fused silica—are compared. The fused silica window is particularly suitable for detecting light at the selected wavelengths of 269 nm and 446 nm in this study. These photodiodes, manufactured by TE Connectivity and procured from Mouser Electronics, were used in the DOBI meter to quantify red palm oil quality.

3.3. Power Supply, LED Driver, and PD Receiver Circuits

3.3.1. Power Supply

The power supply circuit of the DOBI meter, shown in Figure 7, operates on a DC voltage input of 9–12 V. The on/off control is managed by switch S1, which activates regulated outputs of −5 V, +5 V, and +9 V to power the LED driver, photodiode (PD) receiver circuits, and the microcontroller. This ensures consistent functionality and stable operation of the device.
To maintain constant light intensity from the LEDs, a stable current source is essential. The LM317 voltage regulator is employed for this purpose. This widely used three-terminal integrated circuit—comprising input (IN), output (OUT), and adjustment (ADJ) terminals—maintains a voltage drop of 3 V between IN and OUT and 1.25 V between OUT and ADJ during operation [64,65]. When configured as an adjustable constant current source, a variable resistor is connected between OUT and ADJ to set the desired current. The LM317 is a proven solution in photometric applications for driving LEDs.
In this work, LM317 regulators (designated as IC10 and IC13) were used to drive LED446 and LED269, respectively, as shown in Figure 8a and Figure 8b [66]. The power supply provides 12 V, which is sufficient for both LEDs. LED269, with a forward voltage of 6 V, requires a minimum input of 10.25 V (3 + 1.25 + 6 V), while LED446, with a forward voltage of 2.8 V, requires 7.05 V [67]. The current through LED269 is regulated by a combination of resistor R19 and variable resistor R20; likewise, current through LED446 is controlled by resistor R4 and variable resistor R17.

3.3.2. PD Receiver and Signal Conditioning Circuits

Light absorption at 269 nm and 446 nm was measured using photodiodes PD269 and PD446. These were integrated with signal conditioning circuits comprising a transimpedance amplifier (TIA) and a voltage follower (buffer amplifier), as illustrated in Figure 9a,b. The photodiodes were operated in photovoltaic mode to maintain zero bias across their terminals, enhancing measurement accuracy [68].
When light of intensity I 0 passes through a sample, the PDs generate a photocurrent I P D proportional to I [31]. This current is converted into a voltage signal V by the TIA. For 269 nm and 446 nm light, the respective output voltages are V 446 = I P D ( 446 ) R 3 and V 269 = I P D ( 269 ) R 5 [69,70] where R 3 and R 5 are variable resistors used to adjust the gain of the TIA. The voltage signals are then buffered using LM358N operational amplifiers in voltage follower configuration, which maintain signal integrity due to their high input impedance and low output impedance, thereby ensuring unity gain [71].
The conditioned analog signals are digitized by an ADS1115 16-bit ADC module (Adafruit). During measurements, the Programmable Gain Amplifier (PGA) of the ADS1115 was configured with a ± 2.048   V full-scale range, resulting in an effective Least Significant Bit (LSB) size of 62.5 µV (2.048 V/32,768). These digital values are processed by an Arduino MEGA 2560 microcontroller, which calculates the DOBI and carotenoid concentration. The results are displayed in real time on a 3.5-inch TFT screen ( 320 × 480   r e s o l u t i o n ) via an I2C interface.

3.3.3. Printed Circuit Board (PCB), Parts and Fabrication of DOBI Meter

The DOBI meter was designed with careful consideration for both electronic functionality and mechanical integration. The printed circuit board (PCB) (Bangkok, Thailand) layout and component arrangement, as illustrated in Figure 10a,b, were designed using Eagle PCB software (version 8.2). Two LEDs (LED269 and LED446) and their corresponding photodiodes (PD269 and PD446) were mounted orthogonally on either side of a custom-designed cuvette holder. This orthogonal alignment and fixed positioning inside the cuvette help reduce signal crosstalk and minimize stray light interference. Key electronic components, including transimpedance amplifiers (TIAs) and buffer circuits, were positioned near the cuvette region to ensure minimal signal degradation. To enhance user interface and real-time control, a 3.5-inch LCD display and four push buttons were installed adjacent to the cuvette. Additionally, a small fan was integrated beneath the LCD to prevent temperature-induced fluctuations in LED intensity, contributing to stable and reliable device performance. The overall circuitry and component assembly were optimized for compact integration within the system housing. The entire assembled system was enclosed in custom-designed housing, modeled using AutoCAD 2023 and fabricated via 3D printing using black polylactic acid (PLA). The enclosure, with dimensions 11.5   c m × 17.0   c m × 4.5   c m and wall thickness of 4.5 mm, was designed to block external light sources. A 3 mm-thick aluminum lid was fitted to the top of the unit to further enhance light shielding. The cuvette holder itself was fabricated as a 5.6 mm-thick square tube, raised 30 mm above the base, and designed with four optical ports to accommodate the LED and photodiode pairs. A black PLA cuvette cap, 4 mm thick, was employed to eliminate any remaining external light interference during measurement.
To ensure thermal stability and maintain consistent optical output, the enclosure was equipped with a temperature control system comprising a cooling fan and embedded temperature sensor. This system maintains a stable internal environment, ensuring consistent LED output and overall performance reliability of the DOBI meter during extended use.
To address potential safety concerns related to UV leakage from the 269 nm source, measurements were conducted using a UV-Vis spectrophotometer. The results confirmed that the intensity of emitted UV light outside the enclosure was comparable to ambient UV levels, thereby indicating that the system is safe for operation under normal laboratory conditions.
The total cost of components and materials used to fabricate the DOBI meter is approximately 10,000 THB. In contrast, the price of commercial UV-Vis spectrophotometers available in Thailand typically ranges from 40,000 to 300,000 THB, depending on the instrument’s quality, specifications, and country of manufacture. The affordability of the DOBI meter makes it a cost-effective and accessible alternative for small and medium-sized enterprises (SMEs) in Thailand, enabling them to adopt in-house quality control tools for red palm oil production.

3.4. Software and Measurement Procedure

3.4.1. Evaluation of Absorbance

The absorbance of a solution is governed by the Beer–Lambert law [31,43]:
A = ε c L
where A is absorbance, ε is the molar absorption coefficient, c is the concentration of the absorbing species, and L is the optical path length traversed by light of wavelength λ through the solution. Let I represent the light intensity at wavelength λ after it has passed through a sample of concentration c , and I 0 be the initial light intensity. The absorbance of the solvent is given by:
A = l o g I 0 I
When the light interacts with a photodiode (PD), the transmitted intensity I produces an electrical current I P D , which is then converted into a voltage signal by a transimpedance amplifier (TIA). The voltage signal V is directly proportional to the light intensity and can be described as:
V = k I
where k is a constant determined by the characteristics of the TIA circuit. Using this relationship, the absorbance can be expressed as:
A = l o g V 0 V
where V 0 is the voltage measured for the blank (solvent), and V is the voltage measured for the sample. This equation enables precise determination of absorbance at specified wavelengths and forms the basis for calculating DOBI values and carotenoid concentrations in red palm oil.

3.4.2. Measurement Procedure and Software

The measurement process implemented in the DOBI meter is illustrated in Figure 11. The operational algorithm was developed using the Arduino Integrated Development Environment (IDE version 1.8.19) and uploaded to the Arduino MEGA 2560 microcontroller, Arduino store, https://store.arduino.cc/products/arduino-mega-2560-rev3?srsltid=AfBOopH4CJxkEayCxWEbIu-OsJfAoxJlcwZ7ZnTkqLIHmO3Vc0slfRb (asccessed on 22 October 2025). The procedure is as follows:
1.
Upon startup, the display presents the menu: “Blank Set Enter Exit”. The user selects the “Blank” option using the control button.
2.
The screen then prompts “Insert cuvette of the blank”. The user inserts a cuvette containing 95% hexane (the solvent) and presses “Enter”.
3.
The system measures and stores the stores the blank voltages V 0 of the blank at 269 nm and 446 nm.
4.
Next, the user is prompted to “Insert cuvette of the palm oil”. A cuvette containing the red palm oil solution is inserted and confirmed by pressing “Enter”.
5.
The absorbance of the red palm oil is measured at both wavelengths. Using Equations (1) and (2), the device calculates the DOBI value and carotenoid concentration.
6.
The results are displayed in real time on the TFT screen.
This automated measurement procedure ensures consistent, accurate, and repeatable assessment of red palm oil quality, using the integrated hardware–software platform of the DOBI meter.

3.5. Calibration

3.5.1. Measurements of LED Emission Spectral and PD Responses

Figure 12 illustrates the experimental setup used to measure the emission spectra of the LEDs and the response characteristics of the photodiodes (PDs). The setup includes the DOBI meter, an Ocean Optics USB4000 fiber optic spectrometer, and a Fluke 8808A benchtop digital multimeter. During LED spectral measurements, both PDs were removed from the DOBI meter, and optical fibers were inserted into their positions. These fibers were aligned within the cuvette holder to collect light emitted by each LED and transmit it to the spectrometer, which was connected to a computer via USB.
Spectral data were acquired using SpectraSuite software v2 and visualized using KaleidaGraph. Each LED was driven by three different constant electrical currents to produce varying emission intensities, and the corresponding currents were recorded using the digital multimeter. To evaluate the PDs’ responses to LED emissions, a procedure similar to that used for spectral measurements was followed, as summarized below:
1.
Measure the emission spectra of LED269 at I L E D = 6 ,   10   a n d   14.8   m A .
2.
Repeat for LED446 at I L E D = 3 ,   6   a n d   8   m A .
3.
Plot I L E D versus normalized peak intensity I 0 for each LED and determine the slope (emission intensity response).
4.
Reinstall PD269 and PD446 to measure their response to the light from LED269 and LED446, respectively.
5.
Measure the I L E D and resulting output voltage V for each LED–PD pair.
6.
Plot voltage V versus I 0 to determine the photodiode response constants.
7.
Determine the system gain constants from the slope of these plots, representing the relationship between photocurrent and voltage output for each LED–PD pair.

3.5.2. DOBI Meter Calibration

The DOBI meter was calibrated by comparing its measurements with those obtained from a commercial UV-visible spectrophotometer (Merck Prove 600). A series of red palm oil solutions with known concentrations was prepared by weighing between 2.5 and 1000 mg of red palm oil and diluting each sample to a final volume of 25 mL using 95% hexane. Absorbance at 269 nm and 446 nm was measured for all samples using both the DOBI meter and the spectrophotometer to calculate the DOBI values and carotenoid concentrations, as illustrated in Figure 13.
Following calibration with standard solutions, red palm oil extracted from fresh palm fruits was analyzed using both instruments. The extraction was performed using a microwave prototype developed by the Plasma and Electromagnetic Wave Research Laboratory (PEwave) at Walailak University [14]. Fresh fruit bunches were obtained from a local plantation in Tha Sala District, Nakhon Si Thammarat, and separated into spikelets containing fruitlets. These were loaded onto a rotating tray and heated at a constant power of 1000 W for durations of 0, 10, 20, 30, and 40 min.
After heating, the mesocarp was manually separated and pressed using a screw press to obtain red palm oil. The samples were then analyzed using both the DOBI meter and the Prove 600 spectrophotometer. Measurements followed the ISO 17932:2011 standard set by the Malaysian Palm Oil Board (MPOB) [72]. Specifically, 0.1 g of each oil sample was filtered and diluted with isooctane or 95% hexane to 25 mL. Absorbance values at 269 nm and 446 nm were used to calculate DOBI and carotenoid contents, as shown in Figure 14.

4. Results and Discussion

4.1. LED Emission Spectral

The UV LED exhibits peak emission at approximately 277 nm with a full width at half maximum (FWHM) of about 12 nm, while the visible LED peaks around 449 nm with a FWHM of approximately 10 nm, as shown in Figure 15a and Figure 15b, respectively. The UV LED requires a higher driving current than the visible LED to operate. Additionally, increasing the input current enhances the peak emission intensity for both LEDs. During calibration and measurement, the driving currents for LED269 and LED446 were fixed at 10 mA and 6 mA, respectively.

4.2. PD Response

As the driving current increases, both the UV and visible LEDs emit more intense light. Figure 16a demonstrates that the normalized peak intensity I 0 of each LED increases linearly with the driving current I . The slope of the UV LED is higher than that of the visible LED due to its wider energy bandgap, which requires more energy for excitation and light emission [60].
When this light interacts with the photodiode (PD), it generates a voltage difference V . Figure 16b shows the linear relationship between V and I 0 . The constants 0.696 and 1.1096 represent the responses of PD269 and PD446 to emissions from LED269 and LED446, respectively. These values align with the responsivity characteristics of the PC10-2-TO5 photodiode shown in Figure 6b, which has nearly twice the responsivity at 446 nm compared to 269 nm.

4.3. DOBI Meter Calibration and Measurements

According to the Beer–Lambert Law, the absorbance A of a solution is directly proportional to its concentration c and is wavelength-dependent: A = ε ( λ ) c L . Figure 17a and Figure 18a present standard curves demonstrating a linear correlation between red palm oil concentration and absorbance measured at 269 nm and 446 nm using both the DOBI meter and a commercial spectrophotometer (Prove 600). The DOBI meter exhibited higher sensitivity to concentration changes than the Prove 600, indicating superior responsiveness.
By comparing the slopes of the standard curves from both devices, a correction factor was derived to align the DOBI meter’s readings with those from the Prove 600, as shown in Figure 17b and Figure 18b. This calibration confirms the DOBI meter’s capability to accurately measure DOBI values and carotenoid concentrations in red palm oil, making it suitable for both research and quality control. The derived correction factors enable the DOBI meter to measure samples of red palm oil produced via various processing methods, ensuring reliable and consistent results across different production sources.
Following calibration, the DOBI meter was used to measure DOBI values and carotenoid concentrations in red palm oil extracted from mesocarp heated using a 1000 W microwave at various durations (0 to 40 min). Initially, fresh fruit samples contained approximately 820 ppm of carotenoids. After heating for 10 and 20 min, concentrations decreased to 800 ppm and 600 ppm, respectively, likely due to thermal degradation. However, with prolonged heating (30–40 min), the carotenoid content increased again to 700 and 800 ppm. This rise is attributed to the enhanced rupture of oil cells, which released more oil and carotenoids.
As shown in Figure 19a, the carotenoid concentrations measured by both the Prove 600 and DOBI meter were within the industrially accepted range of 500–900 ppm, while the DOBI values shown in Figure 19b were consistently above 2 throughout the heating period. These values meet the quality standards typically required for crude palm oil as reported in Reference [14], confirming that the tested samples remained within acceptable limits for carotenoid retention and oxidation stability during microwave heating.
Despite its advantages in affordability and ease of use, the DOBI meter has some limitations. Firstly, its performance under variable ambient conditions, such as fluctuating temperature, humidity, or lighting, has not yet been validated. The current design assumes stable laboratory environments, which may differ significantly from field or industrial settings. Further testing is needed to confirm robustness across diverse usage conditions and extended periods. Secondly, the DOBI meter uses two specific-wavelength LEDs (269 nm and 446 nm), optimized for measuring absorbance associated with carotenoids and oxidation products (DOBI value). This means it is limited to applications targeting these specific compounds. If users wish to detect other chemical constituents or absorption peaks, they will need to replace the LEDs with those of appropriate wavelengths and recalibrate the system accordingly. This restricts the flexibility of the DOBI meter compared to full-spectrum spectrophotometers, which can scan a wide range of wavelengths without hardware modifications. Thirdly, the system’s accuracy is sensitive to the purity of solvents used in the measurements. In this study, 95% hexane was used consistently due to its availability and practicality in field settings. However, any variations in solvent purity can influence absorbance readings and thus affect the calculated DOBI values. This highlights the need for quality control in solvent preparation and storage, especially in field deployments where solvent quality may vary.
These limitations should be considered when evaluating the suitability of the DOBI meter for different environments or extended applications.

5. Conclusions

This research successfully developed a portable DOBI meter for evaluating the quality of crude red palm oil based on the determination of DOBI values and carotenoid concentrations. The device utilizes narrow-band UV and visible LEDs at 269 nm and 446 nm, respectively, along with broadband photodiodes to detect light absorption in diluted oil samples. Supporting electronics, including LED driver circuits, a photodiode receiver module, and a microcontroller-based interface, were integrated into a compact system suitable for field applications.
The system was thoroughly calibrated by measuring LED emission characteristics, photodiode responses, and standard curves using known concentrations of red palm oil. The calibration results revealed a linear relationship between absorbance and concentration, with the DOBI meter exhibiting higher sensitivity than a commercial spectrophotometer (Merck Prove 600). After calibration, the DOBI meter was employed to measure red palm oil samples extracted from microwave-treated mesocarps. The results showed good agreement between the DOBI meter and the spectrophotometer, with measurement discrepancies below 5%.
The findings demonstrate that the developed DOBI meter is a reliable, cost-effective, and user-friendly alternative for assessing red palm oil quality, particularly in community-based or small-scale production settings where access to laboratory equipment is limited. Future work may focus on enhancing the device’s spectral selectivity, automating calibration, and expanding its application to other edible oils.

Author Contributions

Conceptualization and methodology, M.N.; fabrication of prototype, S.K. and D.S.; investigation and validation, K.W. (Kamonpan Wongyai) and K.W. (Karaket Wattanasit), writing—original draft preparation, K.W. (Kamonpan Wongyai) and M.N.; writing—review and editing, P.R., A.T. and D.B.; visualization, M.N.; supervision, M.N.; funding acquisition, M.N. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by Walailak University, Research Assistant Grant Number WU 02.

Data Availability Statement

Data is contained within the article.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Tan, C.H.; Lee, C.J.; Tan, S.N.; Poon, D.T.S.; Chong, C.Y.E.; Pui, L.P. Red Palm Oil: A Review on Processing, Health Benefits and Its Application in Food. J. Oleo Sci. 2021, 70, 1201–1210. [Google Scholar] [CrossRef]
  2. Nagendran, B.; Unnithan, U.R.; Choo, Y.M.; Sundram, K. Characteristics of red palm oil, a carotene- and vitamin E-rich refined oil for food uses. Food Nutr. Bull. 2000, 21, 189–194. [Google Scholar] [CrossRef]
  3. Loganathan, R.; Subramaniam, K.M.; Radhakrishnan, A.K.; Choo, Y.M.; Teng, K.T. Health-promoting effects of red palm oil: Evidence from animal and human studies. Nutr. Rev. 2017, 75, 98–113. [Google Scholar] [CrossRef]
  4. Purnama, K.O.; Setyaningsih, D.; Hambali, E.; Taniwiryono, D. Processing, Characteristics, and Potential Application of Red Palm Oil—A Review. Int. J. Oil Palm 2020, 3, 40–55. [Google Scholar] [CrossRef]
  5. Nainggolan, M.; Sinaga, A.G.S. Characteristics of fatty acid composition and minor constituents of red palm olein and palm kernel oil combination. J. Adv. Pharm. Technol. Res. 2021, 12, 22–26. [Google Scholar] [CrossRef] [PubMed]
  6. Gao, Z.; Zhu, Y.; Jin, J.; Jin, Q.; Wang, X. Chemical-Physical Properties of Red Palm Oils and Their Application in the Manufacture of Aerated Emulsions with Improved Whipping Capabilities. Foods 2023, 12, 3933. [Google Scholar] [CrossRef] [PubMed]
  7. Nur Sulihatimarsyila, A.W.; Lau, H.L.N.; Nabilah, K.M.; Nur Azreena, I. Production of refined red palm-pressed fibre oil from physical refining pilot plant. Case Stud. Chem. Environ. Eng. 2020, 2, 100035. [Google Scholar] [CrossRef]
  8. Uckert, G.; Hoffmann, H.; Graef, F.; Grundmann, P.; Sieber, S. Increase without Spatial Extension: Productivity in Small-Scale Palm Oil Production in Africa—The Case of Kigoma, Tanzania. Reg. Environ. Change 2015, 15, 1229–1241. [Google Scholar] [CrossRef]
  9. Chow, M.C.; Ma, A.N. Processing of Fresh Palm Fruits Using Microwaves. J. Microw. Power Electromagn. Energy 2016, 40, 165–173. [Google Scholar] [CrossRef]
  10. Sarah, M.; Taib, M.R. Microwave Sterilization of Oil Palm Fruits: Effect of Power, Temperature and D-value on Oil Quality. J. Med. Bioeng. 2013, 2, 153–156. [Google Scholar] [CrossRef]
  11. Sarah, M.; Widyastuti, S.; Ningsih, D. Red Palm Oil Production by Microwave Irradiation. In Proceedings of the TALENTA–CEST 2017 (TALENTA Conference on Engineering, Science and Technology), IOP Conference Series: Materials Science and Engineering. Medan, Indonesia, 7–8 September 2017; Volume 309, p. 012091. [Google Scholar] [CrossRef]
  12. Zamanhuri, N.A.; Rahman, N.A.; Bakar NF, A. Effect of Microwave Power and Extraction Time on Crude Palm Oil Quality Using Microwave-Assisted Extraction Process. Int. J. Renew. Energy Dev. 2021, 10, 495–505. [Google Scholar] [CrossRef]
  13. Sarah, M.; Ramadhan, M.R.; Zahra, A.; Madinah, I.; Maulina, S.; Misran, E. Sterilization of oil palm fruit utilizing continuous microwave sterilizer. Case Stud. Therm. Eng. 2023, 52, 103698. [Google Scholar] [CrossRef]
  14. Wongyai, K.; Kaewpawong, S.; Srinoum, D.; Kongsawat, W.; Polprasarn, K.; Rathore, V.; Nisoa, M. Scale-Up and Development of a Community Industrial Prototype for Red Palm Oil Production Using Advanced Microwave Technology. AgriEngineering 2025, 7, 113. [Google Scholar] [CrossRef]
  15. Tan, Y.-A.; Kuntom, A.; Lee, C.K.; Low, K.S. Comparative evaluation of palm oil color measurement using a prototype palm oil colorimeter. J. Am. Oil Chem. Soc. 2004, 81, 733–736. [Google Scholar] [CrossRef]
  16. Neo, Y.P.; Ariffin, A.; Tan, C.P.; Tan, Y.A. Determination of oil palm fruit phenolic compounds and their antioxidant activities using spectrophotometric methods. Int. J. Food Sci. Technol. 2008, 43, 1832–1837. [Google Scholar] [CrossRef]
  17. Saad, B.; Ling, C.W.; Jab, M.S.; Lim, B.P.; Mohamad Ali, A.S.; Wai, W.T.; Saleh, M.I. Determination of free fatty acids in palm oil samples using non-aqueous flow injection titrimetric method. Food Chem. 2007, 102, 1407–1414. [Google Scholar] [CrossRef]
  18. Azeman, N.H.; Yusof, N.A.; Abdullah, J.; Yunus, R.; Hamidon, M.N.; Hajian, R. Study on the spectrophotometric detection of free fatty acids in palm oil utilizing enzymatic reactions. Molecules 2015, 20, 12328–12340. [Google Scholar] [CrossRef]
  19. Nokkaew, R.; Punsuvon, V.; Inagaki, T.; Tsuchikawa, S. Determination of Carotenoids and Dobi Content in Crude Palm Oil by Spectroscopy Techniques: Comparison of Raman and Ft-Nir Spectroscopy. Int. J. Geomate 2019, 16, 92–98. [Google Scholar] [CrossRef]
  20. Edo, G.I.; Makinde, M.G.; Nwosu, L.C.; Ozgor, E.; Akhayere, E. Physicochemical and Pharmacological Properties of Palm Oil: An Approach for Quality, Safety, and Nutrition Evaluation of Palm Oil. Food Anal. Methods 2022, 15, 2290–2305. [Google Scholar] [CrossRef]
  21. Puteh, A.Q.; Mohd Shahrome, A.A.; Haji Razali, M.H.; Sulaiman, A. Comparative Study of Carotenoids Content in Ripe and Unripe Oil Palm Fresh Fruit Bunches. Int. J. Integr. Eng. 2022, 14, 240–246. [Google Scholar] [CrossRef]
  22. Syamsuddin, J. Determination of Bleachability Index (DOBI) Olein Fraction Crude Palm Oil Using Spectrophotometer UV-VIS. Int. J. Res. Publ. Rev. 2023, 4, 3411–3412. [Google Scholar] [CrossRef]
  23. Muangpratoom, P.; Suriyasakulpong, C.; Maneerot, S.; Vittayakorn, W.; Pattanadech, N. Experimental Study of the Electrical and Physiochemical Properties of Different Types of Crude Palm Oils as Dielectric Insulating Fluids in Transformers. Sustainability 2023, 15, 14269. [Google Scholar] [CrossRef]
  24. Susik, J.; Ptasznik, S. The first stage of refining of post-fermentation corn oil with a high content of free fatty acids and phytosterols—Comparison of neutralisation by an ion-exchange resin without solvent and base neutralisation. Food Res. Int. 2023, 164, 112302. [Google Scholar] [CrossRef] [PubMed]
  25. Putri, G.S.; Saudia, T.N.; Paramita, V.; Ariyanto, H.D. Effect of different composition and emulsifier types on the stability of red palm oil emulsified powder. Malays. J. Chem. Eng. Technol. 2023, 6, 160–173. [Google Scholar] [CrossRef]
  26. Toole, M.O.; Diamond, D. Absorbance Based Light Emitting Diode Optical Sensors and Sensing Devices. Sensors 2008, 8, 2453–2479. [Google Scholar] [CrossRef]
  27. da Silva, M.B.; Crispino, C.C.; Reis, B.F. Automatic photometric titration procedure based on multicommutation and flow-batch approaches employing a photometer based on twin LEDs. J. Braz. Chem. Soc. 2010, 21, 1854–1860. [Google Scholar] [CrossRef]
  28. Bomastyk, B.; Petrovic, I.; Hauser, P.C. Absorbance detector for high-performance liquid chromatography based on light-emitting diodes for the deep-ultraviolet range. J. Chromatogr. A 2011, 1218, 3750–3756. [Google Scholar] [CrossRef]
  29. Zhang, M.; Zhang, Z.; Yuan, D.; Feng, S.; Liu, B. An automatic gas-phase molecular absorption spectrometric system using a UV-LED photodiode based detector for determination of nitrite and total nitrate. Talanta 2011, 84, 443–450. [Google Scholar] [CrossRef]
  30. de Lima, K.M.G. A portable photometer based on LED for the determination of aromatic hydrocarbons in water. Microchem. J. 2012, 103, 62–67. [Google Scholar] [CrossRef]
  31. Bui, D.A.; Hauser, P.C. Analytical devices based on light-emitting diodes—A review of the state-of-the-art. Anal. Chim. Acta 2015, 853, 46–58. [Google Scholar] [CrossRef]
  32. Macka, M.; Piasecki, T.; Dasgupta, P.K. Light-emitting diodes for analytical chemistry. Annu. Rev. Anal. Chem. 2014, 7, 183–207. [Google Scholar] [CrossRef] [PubMed]
  33. Bui, D.A.; Hauser, P.C. Absorbance detector for capillary electrophoresis based on light-emitting diodes and photodiodes for the deep-ultraviolet range. J. Chromatogr. A 2015, 1421, 203–208. [Google Scholar] [CrossRef] [PubMed]
  34. Kvittingen, E.V.; Kvittingen, L.; Sjursnes, B.J.; Verley, R. Simple and Inexpensive UV-Photometer Using LEDs as Both Light Source and Detector. J. Chem. Educ. 2016, 93, 1814–1817. [Google Scholar] [CrossRef]
  35. Noori, A.; Mahbub, P.; Dvorak, M.; Lucieer, A.; Macka, M. Radiometric analysis of UV to near infrared LEDs for optical sensing and radiometric measurements in photochemical systems. Sens. Actuators B Chem. 2018, 262, 171–179. [Google Scholar] [CrossRef]
  36. Murray, E.; Roche, P.; Harrington, K.; McCaul, M.; Moore, B.; Morrin, A.; Diamond, D.; Paull, B. Low cost 235 nm ultra-violet light-emitting diode-based absorbance detector for application in a portable ion chromatography system for nitrite and nitrate monitoring. J. Chromatogr. A 2019, 1603, 8–14. [Google Scholar] [CrossRef]
  37. Grazioli, C.; Faura, G.; Dossi, N.; Toniolo, R.; Abate, M.; Terzi, F.; Bontempelli, G. 3D printed portable instruments based on affordable electronics, smartphones and open-source microcontrollers suitable for monitoring food quality. Microchem. J. 2020, 159, 105584. [Google Scholar] [CrossRef]
  38. Al-Ameri, T.; Prairie, M.W.; Frisbie, S.H.; Rao, K.K.; Saksri, A.H.; Parbat, S.; Mitchell, E.J. An accurate, precise, and affordable light emitting diode spectrophotometer for drinking water and other testing with limited resources. PLoS ONE 2020, 15, e0226761. [Google Scholar] [CrossRef]
  39. Gonzalez-Morales, D.; Valencia, A.; Díaz-Nunez, A.; Fuentes-Estrada, M.; Lopez-Santos, O.; Garcia-Beltran, O. Development of a Low-Cost UV-Vis Spectrophotometer and Its Application for the Detection of Mercuric Ions Assisted by Chemosensors. Sensors 2020, 20, 906. [Google Scholar] [CrossRef]
  40. Li, Y.; Dvorak, M.; Nesterenko, P.N.; Nuchtavorn, N.; Macka, M. High power deep UV-LEDs for analytical optical instrumentation. Sens. Actuators B Chem. 2018, 255, 1238–1243. [Google Scholar] [CrossRef]
  41. Mukunda, D.C.; Joshi, V.K.; Mahato, K.K. Light emitting diodes (LEDs) in fluorescence-based analytical applications: A review. Appl. Spectrosc. Rev. 2020, 57, 1–38. [Google Scholar] [CrossRef]
  42. Lam, S.C.; Coates, L.J.; Gupta, V.; Wirth, H.J.; Gooley, A.A.; Haddad, P.R.; Paull, B. Ultraviolet absorbance detector based on a high output power 235 nm surface mounted device-type light-emitting diode. J. Chromatogr. A 2020, 1631, 461540. [Google Scholar] [CrossRef] [PubMed]
  43. Poh, J.-J.; Wu, W.-L.; Goh, N.W.-J.; Tan, S.M.-X.; Gan, S.K.-E. Spectrophotometer on-the-go: The development of a 2-in-1 UV–Vis portable Arduino-based spectrophotometer. Sens. Actuators A Phys. 2021, 325, 112698. [Google Scholar] [CrossRef]
  44. Al-Sabbagh, B.; Abdulrazzaq, N.N. Measuring Dyes Concentration Using a Low-Cost Visible-Light Spectrophotometer. Iraqi J. Chem. Pet. Eng. 2022, 23, 27–33. [Google Scholar] [CrossRef]
  45. Ye, Z.T.; Kuo, H.-C.; Tseng, S.F.; Chung, S.-R.; Tsou, S.-X. Using Blue Mini-LEDs as a Light Source Designed a Miniaturized Optomechanical Device for the Detection of Direct Bilirubin. Nanoscale Res. Lett. 2022, 17, 111. [Google Scholar] [CrossRef]
  46. Bastan, N.; Ahmadi, M.; Madrakian, T.; Afkhami, A.; Khalili, S.; Majidi, M.; Moradi, M. A paired emitter-detector diode-based photometer for the determination of sodium hypochlorite adulteration in milk. Sci. Rep. 2023, 13, 6217. [Google Scholar] [CrossRef]
  47. Dangkulwanich, M.; Thitaparun, P.; Yimkosol, W.; Chalermsongsak, T. Characterization of factors influencing absorbance measurements in LED-based colorimeters. Eur. J. Phys. 2023, 44, 025801. [Google Scholar] [CrossRef]
  48. Mikhail, I.E.; Hemida, M.; Lebanov, L.; Astrakhantseva, S.; Gupta, V.; Hortin, P.; Parry, J.S.; Macka, M.; Paull, B. Multi-wavelength deep-ultraviolet absorbance detector based upon program-controlled pulsing light-emitting diodes. J. Chromatogr. A 2023, 1709, 464382. [Google Scholar] [CrossRef]
  49. Nisoa, M.; Wattanasit, K.; Tamman, A.; Sirisathitkul, Y.; Sirisathitkul, C. Microwave Drying for Production of Rehydrated Foods: A Case Study of Stink Bean (Parkia speciosa) Seed. Appl. Sci. 2021, 11, 2918. [Google Scholar] [CrossRef]
  50. Pianroj, Y.; Kerdthongmee, M.N.; Kerdthongmee, J.G. Development of a microwave system for highly-efficient drying of fish. Walailak J. Sci. Technol. 2006, 3, 237–250. [Google Scholar]
  51. Kerdtongmee, P.; Srinoum, D.; Nisoa, M. Development of compact high voltage switched mode power supply for microwave plasma sources supply for low pressure plasma. J. Instrum. 2011, 6, P08014. [Google Scholar] [CrossRef]
  52. Suwanchote, C.; Weerakul, J.; Sirisathitkul, C.; Nisoa, M. Color and hardness of durian chips irradiated by controlled low power microwave. Food Sci. Biotechnol. 2012, 21, 1767–1770. [Google Scholar] [CrossRef]
  53. Jaroenkit, P.; Matan, N.; Nisoa, M. Microwave drying of cooked brown rice and the effect on the nutrient composition and trace elements. Int. Food Res. J. 2013, 20, 351–355. [Google Scholar]
  54. Chaijan, M.; Panpipat, W.; Nisoa, M. Chemical deterioration and discoloration of semi-dried tilapia processed by sun drying and microwave drying. Dry. Technol. 2016, 35, 642–649. [Google Scholar] [CrossRef]
  55. Choopan, W.; Panpipat, W.; Nisoa, M.; Cheong, L.Z.; Chaijan, M. Physico-chemical aspects of Thai fermented fish viscera, Tai-Pla, curry powder processed by hot air drying and hybrid microwave-infrared drying. PLoS ONE 2021, 16, e0253834. [Google Scholar] [CrossRef] [PubMed]
  56. Nisoa, M.; Plodkaew, A.; Sirisathitkul, C.; Wattanasit, K.; Somjit, B.; Pacdeepin, P.; Sirisathitkul, Y. Simulation and experimentation on parameters influencing microwave-assisted extraction of bioactive compounds from Kaempferia parviflora rhizomes. Alex. Eng. J. 2023, 65, 357–366. [Google Scholar] [CrossRef]
  57. Sommano, S.; Kerdtongmee, P.; Chompoo, M.; Nisoa, M. Fabrication and characteristics of phase control microwave power for jasmine volatile oil extraction. J. Essent. Oil Res. 2015, 27, 227–233. [Google Scholar] [CrossRef]
  58. Tan, C.H.; Ariffin, A.A.; Ghazali, H.M.; Tan, C.P.; Kuntom, A.; Choo, A.C. Changes in oxidation indices and minor components of low free fatty acid and freshly extracted crude palm oils under two different storage conditions. J. Food Sci. Technol. 2017, 54, 1757–1764. [Google Scholar] [CrossRef]
  59. Loganathan, R.; Tarmizi, A.H.A.; Vethakkan, S.R.; Teng, K.T. A review on lipid oxidation in edible oils. Malays. J. Anal. Sci. 2022, 26, 1378–1393. [Google Scholar]
  60. Schubert, E.F. Light-Emitting Diodes, 3rd ed.; Cambridge University Press: Cambridge, UK, 2018. [Google Scholar]
  61. Harvatek Corporation. SWDC-T306-DNN-U1930: UV-C LED Sterilization Module for Water Tank; Technical Datasheet, Version 1.0; Harvatek Corporation: Hsinchu, Taiwan, 2020–2023. [Google Scholar]
  62. Marktech Optoelectronics. Visible Emitter Product No: MTE4600N—450 nm LED Datasheet; Marktech Optoelectronics: Latham, NY, USA, 2019. [Google Scholar]
  63. Pacific Silicon Sensor. PC10-2-TO5 Silicon Photodiode Data Sheet; Pacific Silicon Sensor Inc.: Westlake Village, CA, USA, 2010. [Google Scholar]
  64. Singh, B.P.; Singh, R. Electronic Devices and Integrated Circuits; Pearson Education: New Delhi, India, 2009. [Google Scholar]
  65. Winder, S. Power Supplies for LED Driving; Newnes (Elsevier): Burlington, MA, USA, 2008. [Google Scholar]
  66. Texas Instruments. LM317: 3-Terminal Adjustable Regulator (Rev. AB); Texas Instruments: Singapore, 2021. [Google Scholar]
  67. Chen, Y.; Su, Y.; Chang, C.; Liang, S. A practical portable photometer using LEDs as inspection light source. Sens. Actuators A Phys. 2017, 263, 579–588. [Google Scholar] [CrossRef]
  68. Wei, Y.; Lehmann, T.; Silvestri, L.; Wang, H.; Ladouceur, F. Photodiode Working in Zero-Mode: Detecting Light Power Change with DC Rejection and AC Amplification. Opt. Express 2021, 29, 18915–18931. [Google Scholar] [CrossRef]
  69. Cretu, V.F.; Kehl, F.; Metz, B.C.; Willis, P.A. Open-Source Lab Hardware: Low Noise Adjustable Two-Stage Gain Transimpedance Amplifier with DC Offset for Low-Light Detection. HardwareX 2021, 10, e00233. [Google Scholar] [CrossRef] [PubMed]
  70. Bielecki, Z.; Achtenberg, K.; Kopytko, M.; Mikołajczyk, J.; Wojtas, J.; Rogalski, A. Review of photodetectors characterization methods. Bull. Pol. Acad. Sci. Tech. Sci. 2022, 70, e140534. [Google Scholar] [CrossRef]
  71. Fang, L.H.; Rahim, R.B.A.; Isa, M.; Syed Hasan, S.I.; Ismail, B.B. The Design of Operational Amplifier for Low Voltage and Low Current Sound Energy Harvesting System. IOP Conf. Ser. Mater. Sci. Eng. 2018, 318, 012035. [Google Scholar] [CrossRef]
  72. ISO 17932:2011; Palm Oil—Determination of the Deterioration of Bleachability Index (DOBI) and Carotene Content. International Organization for Standardization: Geneva, Switzerland, 2011.
Figure 1. Red palm oil obtained from mesocarp of palm oil fruits.
Figure 1. Red palm oil obtained from mesocarp of palm oil fruits.
Agriengineering 07 00370 g001
Figure 2. Absorption spectra of red palm oil.
Figure 2. Absorption spectra of red palm oil.
Agriengineering 07 00370 g002
Figure 3. Principles of LED and PD based DOBI meter.
Figure 3. Principles of LED and PD based DOBI meter.
Agriengineering 07 00370 g003
Figure 4. Schematic diagram of DOBI meter.
Figure 4. Schematic diagram of DOBI meter.
Agriengineering 07 00370 g004
Figure 5. (a) UV LED and (b) visible LED.
Figure 5. (a) UV LED and (b) visible LED.
Agriengineering 07 00370 g005
Figure 6. (a) Photodiode and (b) UV and its responsivity for different wavelength.
Figure 6. (a) Photodiode and (b) UV and its responsivity for different wavelength.
Agriengineering 07 00370 g006
Figure 7. Schematic diagram of the power supply circuit.
Figure 7. Schematic diagram of the power supply circuit.
Agriengineering 07 00370 g007
Figure 8. Schematic diagram of the LED driver circuits (a) for LED 446 nm and (b) for LED 269 nm.
Figure 8. Schematic diagram of the LED driver circuits (a) for LED 446 nm and (b) for LED 269 nm.
Agriengineering 07 00370 g008
Figure 9. Schematic diagram of the PD receiver circuits (a) for λ = 446   n m and (b) for λ = 269   n m .
Figure 9. Schematic diagram of the PD receiver circuits (a) for λ = 446   n m and (b) for λ = 269   n m .
Agriengineering 07 00370 g009
Figure 10. (a) Printed circuit board (PCB), (b) PCB layout, (c) assembled PCB, LCD display and control button in the DOBI meter’s housing and (d) completely assembled DOBI meter.
Figure 10. (a) Printed circuit board (PCB), (b) PCB layout, (c) assembled PCB, LCD display and control button in the DOBI meter’s housing and (d) completely assembled DOBI meter.
Agriengineering 07 00370 g010
Figure 11. Flowchart of the software DOBI meter.
Figure 11. Flowchart of the software DOBI meter.
Agriengineering 07 00370 g011
Figure 12. Experimental setup for measurements of LED spectral and PD response.
Figure 12. Experimental setup for measurements of LED spectral and PD response.
Agriengineering 07 00370 g012
Figure 13. Calibration of the DOBI meter using standard red palm oil solutions and comparison with a commercial spectrophotometer (Prove 600) at 269 and 446 nm.
Figure 13. Calibration of the DOBI meter using standard red palm oil solutions and comparison with a commercial spectrophotometer (Prove 600) at 269 and 446 nm.
Agriengineering 07 00370 g013
Figure 14. Analysis of microwave-extracted red palm oil using the DOBI meter and Prove 600 spectrophotometer at 269 and 446 nm after heating fruitlets at 1000 W for various durations.
Figure 14. Analysis of microwave-extracted red palm oil using the DOBI meter and Prove 600 spectrophotometer at 269 and 446 nm after heating fruitlets at 1000 W for various durations.
Agriengineering 07 00370 g014
Figure 15. (a,b) show the relative emission spectra of UV and Visible LEDs.
Figure 15. (a,b) show the relative emission spectra of UV and Visible LEDs.
Agriengineering 07 00370 g015
Figure 16. (a) Linear dependence of peak intensities I0 on driving current I for LED269 and LED446, and (b) linear dependence of voltage difference Vabs on I0 for PD269 and PD446.
Figure 16. (a) Linear dependence of peak intensities I0 on driving current I for LED269 and LED446, and (b) linear dependence of voltage difference Vabs on I0 for PD269 and PD446.
Agriengineering 07 00370 g016
Figure 17. UV (a) The linear dependence of absorbance on concentration (standard curve) obtained by DOBI meter and spectrophotometer Prove 600 for λ = 269 nm and (b) the modified standard curves obtained by the ratio between the slopes in (a).
Figure 17. UV (a) The linear dependence of absorbance on concentration (standard curve) obtained by DOBI meter and spectrophotometer Prove 600 for λ = 269 nm and (b) the modified standard curves obtained by the ratio between the slopes in (a).
Agriengineering 07 00370 g017
Figure 18. UV (a) The linear dependence of absorbance on concentration (standard curve) obtained by DOBI meter and spectrophotometer Prove 600 for λ = 446 nm and (b) the modified standard curves obtained by the ratio between the slopes in (a).
Figure 18. UV (a) The linear dependence of absorbance on concentration (standard curve) obtained by DOBI meter and spectrophotometer Prove 600 for λ = 446 nm and (b) the modified standard curves obtained by the ratio between the slopes in (a).
Agriengineering 07 00370 g018
Figure 19. (a) The concentration of carotenoids and (b) DOBI value measured in red palm oil heated using a 1000 W microwave for varying durations using Prove 600 and DOBI meter.
Figure 19. (a) The concentration of carotenoids and (b) DOBI value measured in red palm oil heated using a 1000 W microwave for varying durations using Prove 600 and DOBI meter.
Agriengineering 07 00370 g019
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Wongyai, K.; Kaewpawong, S.; Wattanasit, K.; Srinoum, D.; Nisoa, M.; Rattanakit, P.; Tamman, A.; Boonyawan, D. Development of a Portable LED-Based Photometer for Quality Assessment of Red Palm Oil in SMEs. AgriEngineering 2025, 7, 370. https://doi.org/10.3390/agriengineering7110370

AMA Style

Wongyai K, Kaewpawong S, Wattanasit K, Srinoum D, Nisoa M, Rattanakit P, Tamman A, Boonyawan D. Development of a Portable LED-Based Photometer for Quality Assessment of Red Palm Oil in SMEs. AgriEngineering. 2025; 7(11):370. https://doi.org/10.3390/agriengineering7110370

Chicago/Turabian Style

Wongyai, Kamonpan, Suttirak Kaewpawong, Karaket Wattanasit, Dhammanoon Srinoum, Mudtorlep Nisoa, Parawee Rattanakit, Arlee Tamman, and Dheerawan Boonyawan. 2025. "Development of a Portable LED-Based Photometer for Quality Assessment of Red Palm Oil in SMEs" AgriEngineering 7, no. 11: 370. https://doi.org/10.3390/agriengineering7110370

APA Style

Wongyai, K., Kaewpawong, S., Wattanasit, K., Srinoum, D., Nisoa, M., Rattanakit, P., Tamman, A., & Boonyawan, D. (2025). Development of a Portable LED-Based Photometer for Quality Assessment of Red Palm Oil in SMEs. AgriEngineering, 7(11), 370. https://doi.org/10.3390/agriengineering7110370

Article Metrics

Back to TopTop