1. Introduction
Leishmaniasis is a vector-borne disease that is prevalent worldwide, particularly in countries with tropical and subtropical climates. Annually, it is estimated that 30,000 new cases of visceral leishmaniasis and over 1 million new cases of cutaneous leishmaniasis occur. The clinical presentation of leishmaniasis depends on the species of
Leishmania that infects and the immune status of the host. The disease can manifest in three main forms: visceral, cutaneous, and mucosal [
1]. In the Americas, leishmaniasis is present in 18 countries, with more than 53,000 new cases recorded annually. The countries with the highest prevalence of the disease are Brazil, Colombia, and Peru [
2].
Of the more than 20 species that cause the disease, about 13 are endemic to the Americas. Of these, only
Leishmania (Leishmania) infantum is a viscerotropic species found in 13 countries in the region, including Argentina, Bolivia, Colombia, Costa Rica, El Salvador, Guatemala, Honduras, Mexico, Nicaragua, Paraguay, Uruguay, Venezuela, and Brazil, the latter accounting for more than 95% of the total number of cases. In contrast, Brazil, Colombia, and Peru account for over 70% of the cases of the American tegumentary leishmaniasis. The most frequently reported species are
Leishmania (Viannia) braziliensis,
Leishmania (Viannia) guyanensis,
Leishmania (Viannia) panamensis,
Leishmania (Viannia) peruviana,
Leishmania (Viannia) lainsoni, and
Leishmania (Leishmania) mexicana [
3,
4,
5,
6].
A definitive diagnosis of the disease is reached through the application of parasitological methods, including direct microscopic examination (DME) by smear or impression,
in vitro culture, microculture, and histopathology. The sensitivity and specificity of these methods have been reported to vary between 60% and 100% [
7]. Conversely, molecular methods based on the amplification of nucleic acids have facilitated the evaluation of a number of genetic markers, including
18S rDNA,
kDNA,
Hsp70,
ITS1 and
ITS2, and
AAP3, among others [
8,
9], all of which are directed towards the detection of the genus
Leishmania. However, only a limited number of studies have evaluated their suitability for a significant number of clinical samples [
10,
11,
12,
13].
Among the targets evaluated, the
heat shock protein 20 gene (
Hsp20), identified in
Leishmania major,
Leishmania infantum, and
Leishmania braziliensis, showed high conservation among
Leishmania species but divergence from other organisms. This enabled the development of a conventional PCR assay targeting
Hsp20, demonstrating superior analytical sensitivity and high diagnostic specificity across all
Leishmania species [
14,
15,
16]. Our objective was to develop and validate a real-time PCR assay targeting the
Hsp20 gene, given its advantages such as sensitivity, specificity, and throughput.
In the present study, real-time PCR targeting the Hsp20 region was evaluated for the detection of Leishmania species using SYBR Green and TaqMan probes with DNA from six different Leishmania strains. The real-time PCR method was validated on tissue samples from patients with confirmed leishmaniasis, obtained from Giemsa-stained slides and lesion scrapings collected with lancets.
2. Materials and Methods
2.1. Reference Strains
The present retrospective observational study was conducted at the Laboratorio de Metaxénicas y Zoonosis Parasitaria of the Instituto Nacional de Salud (INS), Perú. The study focused on
Leishmania strains belonging to the subgenus
Viannia and
Leishmania species (
Table 1), which were cryopreserved in liquid nitrogen; the strains were thawed and cultured in a biphasic medium consisting of blood agar and Schneider’s liquid medium (Ref. 21720024, Gibco, Paisley, UK) with 20% heat-inactivated fetal bovine serum (HI-FBS) (Ref 10082-147, Gibco, Grand Island, NY, USA) and an antibiotic-antimycotic solution (Ref. 15240-062, Gibco, Carlsbad, CA, USA) (supplemented Schneider’s medium). Daily microscopic observation was employed during the reactivation process.
The medium containing
Leishmania parasites was subjected to centrifugation at 8000 rpm for 5 min. The supernatant was then carefully removed, and the sediment was retained for genomic DNA extraction. The DNA samples from the strains were confirmed by
Hsp70 PCR-RFLP, using the primers
Hsp70sen 5′-GACGGTGCCTGCCTGCCTACTACTTCAA-3′ and
Hsp70ant 5′-CCGCCCATGCTCTCTGGTACATC-3′ [
17,
18]. The methodology previously described by Fraga et al. (2012) was employed, utilizing the restriction enzymes HaeIII and RsaI (Cat. ER0151 and ER1122, Thermo Scientific, Vilnius, Lithuania).
2.2. Sample Collection
The tissue samples included in this study were obtained between the years 2011 and 2019, as part of the process for the confirmatory diagnosis of the disease in patients who were referred to the National Institute of Health, or in the context of the quality control process for the direct microscopic examination conducted by the National Reference Laboratory for Metaxenic and Parasitic Zoonoses and implemented by regional reference laboratories.
The sample size was determined using the free software EPIDAT 4.2 with an expected sensitivity and specificity greater than 95% [
16]; we used a precision of 5%, a healthy-to-sick ratio of 1, and a 95% confidence interval (CI), employing the following formulas:
where
n represents the sample size,
Z denotes the confidence level,
S corresponds to the expected sensitivity,
E corresponds to the expected specificity, and
L represents the acceptable margin of error.
Group I included Giemsa-stained slides sent by local laboratories in endemic areas to the Laboratorio de Referencia Nacional de Metaxénicas y Zoonosis Parasitaria for quality control in direct microscopic examination. Only slides that showed positive results by visualizing at least one amastigote form in one hundred fields using a 100x oil-immersion objective was included. Additionally, Giemsa-stained slides that showed negative results, a low presence of tissue, poor conservation quality, or evidence of contamination were excluded from the group.
Group II included tissue samples obtained from patients by scraping with sterile lancets, which were used to prepare Giemsa-stained slides for direct microscopic examination or to inoculate a biphasic medium for in vitro culture. These samples showed positive results, either by visualizing at least one amastigote form in one hundred fields using the 100x oil-immersion objective, by visualizing promastigote forms in liquid medium with an inverted microscope using a 40x objective, or by both methods. The tissue obtained from the lancets was preserved in sterile polypropylene conical tubes immersed in 96% alcohol.
Group III included tissue samples from lancets of patients who showed negative results by assessing all fields of the Giemsa-stained slides in direct microscopic examination with a 100x oil-immersion objective, or who did not exhibit promastigote forms in liquid medium when evaluated using an inverted microscope with a 40x objective. Additionally, samples from patients who showed a positive result when assessed in the Montenegro skin test were excluded.
To ensure a rigorous analysis process and minimize bias, samples failing to meet the inclusion criteria for any group were excluded. Furthermore, the remaining samples underwent initial assessment for human RNase P detection using real-time PCR before Hsp20 assessment. Samples yielding neither human RNase P nor Hsp20 amplification were excluded from the final analysis.
2.3. DNA Extraction and Purification
Genomic DNA was extracted from the samples using the Purelink Genomic DNA Mini Kit (Ref. K1820-02, Invitrogen, Life Technologies, Carlsbad, CA, USA). Reference strains were processed in accordance with the kit instructions and were eluted in a final volume of 100 µL.
Prior to genomic DNA extraction, the slides were immersed in methanol to eliminate any residual dye. The tissue samples from each slide were removed individually with the aid of a No. 15 scalpel blade. The biological material was transferred to conical polypropylene tubes and subjected to centrifugation at 8000 rpm for 10 min. Subsequently, the supernatant was then removed and 180 µL of the PureLinkTM Genomic Digestion Buffer was added.
To extract DNA from tissue obtained by lancing, the tubes containing the samples were subjected to centrifugation at 8000 rpm for 10 min. Subsequently, the supernatant was removed and 180 µL of the digestion buffer was added.
The tubes containing the samples, obtained from either Giemsa-stained slide or lancets, were homogenized and incubated at 55 °C for two hours. Subsequently, the samples were processed in a manner like that used for the extraction of genomic DNA from Leishmania strains. The DNA was quantified using an Eon5 spectrophotometer (BioTek Instruments, Winooski, VT, USA), and then aliquoted to a volume of 50 µL for storage at −20 °C until required.
2.4. Primer Design and Performance Evaluation of Real-Time PCR
The real-time PCR targeting the region of the gene encoding Hsp20 was developed using the primers initially described by Montalvo A., et al. (2020): Hsp20S F: 5′-GCCRGARGTGARRAAGGAGGAGGAC-3′ and Hsp20S R: 5′-GYAGCTGGYKYTCGTCCTGC-3′. Likewise, a specific FAM/TAMRA-linked probe sequence was designed, using the program Geneious Prime 2022.1 based on the sequences of Leishmania (Leishmania) aethiopica JX630112, Leishmania (Leishmania) tropica JX630114, Leishmania (Leishmania) major JX630117, Leishmania (Leishmania) donovani JX630120, Leishmania (Leishmania) infantum JX630122, Leishmania (Leishmania) chagasi JX630125, Leishmania (Leishmania) mexicana JX630126, Leishmania (Leishmania) amazonensis JX630128, Leishmania (Leishmania) garnhami JX630129, Leishmania (Viannia) braziliensis JX630135, Leishmania (Viannia) peruviana JX630139, Leishmania (Viannia) guyanensis JX63011242, Leishmania (Viannia) panamensis JX63011243, and Leishmania (Viannia) lainsoni JX630148 from the GenBank database. Real-time PCR targeting Hsp20 was evaluated on MicPCR (Biomolecular Systems, Upper Coomera, Australia) and Rotor-Gene Q (Qiagen, Hilden, Germany) thermal cyclers in a reaction volume of 10 µL, as well as during the standardization process.
The performance of real-time PCR for the detection of a specific region of Hsp20 of Leishmania was evaluated using an intercalating fluorophore with an affinity for DNA (SYBR Green), as well as Taqman-labeled probes.
To detect Leishmania with intercalating fluorophores, a 1x solution of Luminaris Color HiGreen qPCR Master Mix 2x solution (Cat. K0392, Thermofisher, Vilnius, Lithuania) was employed, along with primer Hsp20S F and Hsp20S R at 0. 3 µM each and 3 µL of genomic DNA. The total volume was completed with molecular biology-grade water. The amplification conditions included an initial denaturation step at 95 °C for five minutes for polymerase activation. Subsequently, two stages were conducted. The initial stage comprised five cycles without data acquisition: denaturation at 95 °C for 15 s, followed by a hybridization/extension at 62 °C for 30 s. The second stage comprised 45 cycles with data acquisition, which consisted of denaturation at 95 °C for 15 s and hybridization/extension at 62 °C for 30 s.
The detection of Leishmania using Taqman-labeled probes was achieved using 1x Platinum™ Quantitative PCR SuperMix-UDG 2x (Invitrogen, Life Technologies, Carlsbad, CA, USA), primers Hsp20S F and Hsp20S R at 0.5 µM, and probe Hsp20SS at 0.3 µM. The amplification conditions were similar to those used with the intercalating fluorophore, with the exception that the hybridization/extension step was developed at 57 °C. The final concentration of the magnesium chloride in the reaction tubes was 3.5 mM, the concentration of the primer and probe was 0.3 µM, and the volume of DNA was 3 µL. The remaining components were molecular biology-grade water.
2.5. Real-Time PCR Efficient, Analytical Sensitivity, Inclusivity, and Exclusivity
To determine the efficiency of the real-time PCR, as well as the analytical sensitivity (limit of detection, LoD) of the method, genomic parasite DNA was prepared at concentrations equivalent to 1 × 10
4 up to 1 × 10
−1, parasite/mL by serial dilution from different
Leishmania species [
7,
19].
The inclusivity was determined by the detection of
Leishmania species described in
Table 1, while exclusivity was evaluated using genomic DNA samples from related pathogens, including
Crithidia spp.,
Trypanosoma cruzi,
Plasmodium vivax, and
Plasmodium falciparum. All parameters were assessed in triplicate.
2.6. Validation Process
The validation of the real-time PCR method for the detection of
Leishmania included the parameters of diagnostic sensitivity and specificity, predictive values, and precision of the method. All parameters were determined in accordance with the current Peruvian standards [
20].
To guarantee the adequate DNA extraction process, the samples were previously processed for human RNase P region detection through real-time PCR. The TaqMan® RNase P Detection Reagents Kit (Cat. 4316831, ThermoFisher Scientific, Vilnius, Lithuania) was used. The amplification was developed with denaturation at 95 °C for 2 min, followed by 40 cycles of denaturation at 95 °C for 3 s and hybridization/extension at 60 °C for 30 s. The data acquisition was performed using the green channel. A cutoff cycle threshold (Ct) value of ≤35 was established to minimize possible nonspecific amplifications and decrease the occurrence of false positives.
Samples were processed using real-time PCR, and those that amplified or exhibited a Ct value ≤ 35 were considered positive. In this context, samples that amplified solely for the Hsp20 region or for both Hsp20 and RNase P were classified as positive. Samples that amplified human RNase P but did not achieve a Ct value of 35 or lower for the Hsp20 target were deemed negative. Conversely, samples that did not amplify any targets for either Hsp20 or human RNase P were excluded from subsequent analysis.
The diagnostic sensitivity and specificity, along with the predictive values and likelihood ratios, were determined through the evaluation of groups I, II, and III. The repeatability of the method was assessed by processing a positive and a negative control sample on five separate occasions over a five-day period, under the same conditions concerning the analyst, instrument, and environment.
Similarly, intermediate precision was determined from the results by two analysts who employed the same methodology utilized for repeatability.
2.7. Analysis
Efficiency of real-time PCR was determined using the following formula [
19]:
In this context, the slope reflects the steepness of the standard curve, as determined by plotting the Ct values against the logarithm of the starting quantity of target DNA [
19].
The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), likelihood ratio (LR), and area under the curve (AUC) were determined using the parasitological tests as the reference standards. Statistical analysis was conducted using Fisher’s exact test in SPSS v.26.0 software, with a confidence interval of 95%.
2.8. Ethics
The present study was developed from samples that were obtained between 2011 and 2019 within the framework of the specialized diagnosis of the disease in patients and the Quality Control Program of the Direct Microscopic Examination of the National Institute of Health. Furthermore, the study was approved by the Ethics in Research Committee of Instituto Nacional de Salud—Peru, by Memorandum No. 028-2022-CIEI-INS.
3. Results
Real-time PCR targeting the Hsp20 region was optimized for Leishmania detection. A slope ranging from −3.40 to −3.89 was observed when SYBR Green was used, with lower values for Leishmania (Leishmania) mexicana and higher values for Leishmania (Leishmania) amazonensis. Additionally, R2 values were determined between 0.94 and 0.99, and an efficiency of the real-time PCR greater than 90%.
In contrast, the TaqMan probe exhibited a slope ranging from −3.27 to −4.096, with lower values for
Leishmania (Viannia) guyanensis and higher values for
Leishmania (Viannia) lainsoni, respectively, with R
2 values of 0.99, and an efficiency of 100%, as shown in
Figure 1 and
Table 2.
Similarly, the LoD was found to be 1 parasite/µL in both systems. However, Leishmania (Viannia) peruviana was not amplified using the TaqMan probe, indicating a limitation in the detection range of this specific probe compared to using SYBR Green.
Regarding the inclusivity of the method, all species of
Leishmania described in
Table 1 were detected using SYBR Green, whereas with the TaqMan probe,
Leishmania (Viannia) peruviana did not show amplification. With respect to exclusivity, both detection systems demonstrated specificity for
Leishmania, while neither detection system exhibited amplification when sourced from DNA from
Crithidia spp.,
Plasmodium vivax,
Plasmodium falciparum, and
Trypanosoma cruzi.
In the validation process, a total of 323 samples were included in this study. Specifically, 128 samples from Giemsa-stained slides were categorized as Group I, while 129 samples obtained from lancets were classified as Group II. Additionally, 66 tissue samples from lancets identified as negative for Leishmania were categorized as Group III.
During the analysis process, 32 samples were excluded based on specific criteria to ensure reliability and accuracy. These excluded samples exhibited a Ct value for the RNase P that exceeded a Ct value of 35. Of these excluded samples, 15 were from Group I, and the remaining 17 were from Group II.
The results are presented in
Table 3. The real-time polymerase chain reaction (PCR) targeting the
Hsp20 region with SYBR Green was performed on a total of 225 samples that tested positive in parasitological assays. Of these, 198 were accurately identified, while 27 yielded false-negative results; of these, 17 were from Giemsa-stained slides and 10 from lancet samples. The overall sensitivity of the method was 88.00% (95% CI: 83.53–92.47), with sensitivities of 84.96% (95% CI: 77.9–91.9) and 91.20% (95% CI: 84.3–95.7) for Giemsa-stained slides and lancets, respectively. Importantly, no false positives were observed, leading to a specificity of 100%.
Conversely, given the limitations of
Leishmania (Viannia) peruviana and
Leishmania (Viannia) lainsoni detected by the TaqMan probe, this analysis was conducted on 36 samples that were randomly selected from Group II. Seventeen samples were identified as positive, while nineteen were classified as false negatives, resulting in a sensitivity of 47.22% (95% CI: 29.53–64.92). Notably, the TaqMan probe exhibited no false positives, as shown in
Table 4.
Finally, the AUC for real-time PCR targeting the
Hsp20 gene region using SYBR Green considering positive samples from both Giemsa-stained slides and lancets was 0.94 (95% CI: 0.91–0.97). Independently, samples from Giemsa-stained slides showed an AUC of 0.92 (95% CI: 0.88–0.97), while samples obtained from lancets demonstrated an AUC of 0.96 (95% CI: 0.93–0.98). In the case of the TaqMan probe, the method assessed using samples from lancets reached an AUC of 0.74 (95% CI: 0.62–0.85), as shown in
Figure 2.
4. Discussion
To date, direct microscopic examination remains the most widely used method for diagnostic leishmaniasis, which involves the microscopic visualization of parasites from tissue samples that are fixed in glass slides and stained with Giemsa, allowing for time-efficient results. However, despite this advantage, this method presents limitations associated with sensitivity, which is closely related to the quality of the sample. Poor sample collection can lead to false-negative results.
Moreover, the performance of direct microscopic examination could be influenced by factors such as the stage of the disease, the clinical presentation, parasite loads, and other variables that significantly affect diagnostic outcomes. It is inherently limited by the subjective nature of the examination process. The specificity of the test depends mainly on the experience and skill of the microscopist [
21]. Those with limited experience may confuse the parasitic form of
Leishmania with other pathogens, such as
Histoplasma or
Sporothrix schenckii, leading to misdiagnosis and inappropriate management of the condition. On the other hand, patients with chronic cutaneous or mucosal lesions represent a greater diagnostic challenge because of their reduced parasitic load, which may result in false negatives, delayed treatment, and adverse patient outcomes [
22,
23].
Integrating molecular diagnostic techniques, such as PCR, can significantly improve sensitivity and specificity, even in samples with reduced parasitic load, thereby addressing one of the key limitations of microscopy. This approach can optimize patient management strategies, particularly in endemic regions. In this context, real-time PCR is a tool with increased sensitivity and specificity compared to conventional diagnostic methods, capable of detecting even low concentrations of parasitic DNA, down to a single copy, as demonstrated in this study with the
Hsp20 gene. This capability is especially important for the early and accurate diagnosis of patients with chronic skin lesions or those with mucosal lesions. As a result, the likelihood of false-negative results is significantly reduced. In this context, molecular methods can facilitate the diagnosis of the disease, such as real-time PCR targeting the
18S rDNA,
kDNA,
ITS1, and
Hsp70 regions, among others, genetic markers for the detection of
Leishmania [
10,
12,
24,
25].
The
Hsp20 gene, a conserved marker in
Leishmania, was initially identified in
Leishmania amazonensis in 2009 [
14]. Previous reports have demonstrated that the target exhibits a high level of sensitivity and specificity in the detection of
Leishmania when employing conventional PCR [
15,
16].
In this study, we utilized real-time PCR targeting
Hsp20 for the detection of
Leishmania. We evaluated the efficiencies of the SYBR Green and TaqMan detection systems in terms of their ability to amplify the target DNA during each cycle [
19], thereby ensuring accurate and consistent detection of
Leishmania in the analyzed samples. Our findings revealed that the molecular method exhibits amplification efficiencies of 91% and 100%, respectively, while the LoD was of one parasite per microliter for both systems.
The estimated diploid genome size of
Leishmania is 83.15 femtograms (fg). Previous reports have demonstrated that the real-time PCR that targets the SSU-rRNA region allows for the detection of up to 0.036 picograms (pg) of
Leishmania DNA [
24], equivalent to 36 fg. When compared with the total DNA concentration of
Leishmania, this represents an ability to identify DNA quantities equivalent to less than one parasite. Meanwhile, the
18S rRNA target allows for the detection of approximately 8.32 × 10
−3 fg, predominantly in
Leishmania (Viannia) guyanensis and
Leishmania (Viannia) panamensis, which is equivalent to the total genomic DNA of a single parasite diluted to a ratio of 1:1000. In the case of
Leishmania (Viannia) braziliensis and
Leishmania (Leishmania) amazonensis, the assay exhibited an LoD of up to one parasite [
10].
Similar to the 18S rRNA, the Hsp70 gene region utilized in real-time PCR enabled the detection of up to one parasite in Leishmania (Viannia) guyanensis, Leishmania (Viannia) braziliensis, and Leishmania (Leishmania) amazonensis. Leishmania (Viannia) panamensis could be detected at a concentration equivalent to 1000 parasites [
10]. In another study, both 18S rRNA and kDNA demonstrated the ability to detect a thousandths part of a parasite, while Hsp70 and ITS were able to detect a concentration equivalent to a hundredth of Leishmania [
9].
In relation to inclusivity, the assessed Hsp20 region in real-time PCR using SYBR Green exhibited the ability to detect different Leishmania species, which is concordant with targets such as Hsp70, kDNA, ITS, and 18S rRNA. Conversely, the use of a TaqMan probe did not show the ability to detect Leishmania (Viannia) peruviana; additionally, Leishmania (Viannia) lainsoni yielded a slope of −4.0. This characteristic was possibly related to specific variations in the complementary DNA sequences of the TaqMan probe.
The
Hsp20 region in both detection systems (SYBR Green and TaqMan probe) were shown to be specific to
Leishmania, as demonstrated in studies developed by Montalvo A (2020), a characteristic similar to that of
Hsp70 and
ITS. Meanwhile, the
18S rRNA and
kDNA, which demonstrated an ability to detect quantities lower than one parasite, exhibited cross-reactivity, particularly with
Trypanosoma cruzi [
9,
10].
During the validation process, we found that the detection of
Leishmania spp. by real-time PCR targeting the
Hsp20 region exhibited 100% specificity (95% CI: 99.2–100), with no significant differences compared to the gold standard, consistent with previous reports [
26]. Conversely, the sensitivity of the molecular method showed a significant difference when the SYBR Green system and the TaqMan probe were compared.
Sensitivity values for
Hsp70 and
18S rRNA were reported at 83.6% and 90.9%, respectively, accompanied by a specificity of 61% for Hsp70 and 18% for
18S rRNA in real-time PCR using the TaqMan probe [
10]. In the case of
kDNA, sensitivity values ranged from 73% to 88% using SYBR Green and from 54% to 79% with the TaqMan probe, both demonstrating specificity values of 100% [
27,
28]. Our real-time PCR targeting the
Hsp20 gene region using SYBR Green reached a sensitivity value of 88% (95% CI 83.53–92.47) with a specificity of 100% (95% CI 94.2–100).
Using SYBR Green, a total of 27 samples were identified as false negatives; 17 DNA samples from Giemsa-stained slides exhibited discordant results compared to the reference parasitological results. Additionally, ten samples from tissue obtained from lancets were also false negatives. This mismatch did not correlate with the parasite load and is probably associated with the preservation of the DNA samples or a probable genetic variability of Leishmania, which does not allow adequate hybridization between the oligonucleotide sequence of the labeled TaqMan probe and the specific sequence of Leishmania.
In contrast, the use of the TaqMan probe reported sensitivity and specificity values of 47% (95% CI 29.52–64.92) and 100% (95% CI 99.24–100.00), respectively. The limited clinical sensitivity could be attributed to variations in the sequences of certain Leishmania species, mainly Leishmania (Viannia) peruviana and Leishmania (Viannia) lainsoni. Using this detection system, from 36 samples randomly selected from Group II, 19 samples identified as false negatives. This mismatch was probably associated with the genetic variability of the parasite that limits the capacity of the TaqMan probe to detect all species of Leishmania. These difficulties underscore the need for a new design of the TaqMan probe to improve the performance of the method.
These sensitivity and specificity values of the real-time PCR-Hsp20 method for the detection of Leishmania found using SYBR Green were comparable to those previously reported by Montalvo A. et al. (2020), despite differences in the number of samples used in each study. These results demonstrated the robustness of the real-time PCR-Hsp20 method. The analysis showed that the sensitivity observed in the samples obtained from lancets was slightly higher than in the DNA samples from Giemsa-stained slides; however, this difference was not statistically significant (p-value > 0.05), suggesting that both sample types allow for the effective detection of Leishmania. This slight variation could be related to the conservation of tissue samples and DNA in Giemsa-stained slides compared to samples from lancets that were preserved in alcohol, which allows for better conservation of tissue samples and DNA.
5. Conclusions
In conclusion, the real-time PCR method targeting the Hsp20 gene using SYBR Green for the detection of Leishmania spp. has demonstrated high sensitivity and specificity, achieving values of 88% and 100%, respectively, in DNA samples extracted from DME slides and lancets. This method can detect low concentrations of Leishmania down to a single cell, which underscores its potential for early and accurate diagnosis, which is particularly important for patients presenting with chronic cutaneous lesions or mucosal involvement, where the parasite load may be minimal and traditional diagnostic methods may fall short.
Furthermore, the application of this diagnostic method could facilitate access to timely treatment, significantly improving patient outcomes, as well as initiating appropriate strategies for controlling the disease in conditions where progression may lead to severe complications.
However, it is essential to continue research focused on identifying specific sequences associated with the TaqMan probe that enhance the method and enable the development of a duplex real-time PCR assay, allowing for the simultaneous detection of both the Hsp20 region and an endogenous control. This dual-target approach will not only improve diagnostic precision but also assess the quality of the samples being tested, contributing to the accuracy of the method.
Additionally, further investigations should aim to validate the real-time PCR-Hsp20 method in larger and more diverse populations to ensure its applicability across various demographic groups and clinical presentations. Expanding the scope of the study will help ascertain the method’s robustness and adaptability to different geographical settings.