Next Article in Journal
PUF-PSS: A Physically Secure Privacy-Preserving Scheme Using PUF for IoMT-Enabled TMIS
Next Article in Special Issue
A Compact 10–14.5 GHz Quadrature Hybrid with Digitally Reconfigurable I/Q Phase in SiGe BiCMOS Process
Previous Article in Journal
Compact Ultra-Wideband Wilkinson Power Dividers Using Linearly Tapered Transmission Lines
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Real-Time Non-Uniformity Correction without TEC for Microbolometer Array

1
School of Electrical and Computer Engineering, UNIST, Ulsan 44919, Korea
2
School of Information, Communications and Electronics Engineering, The Catholic University of Korea, Gyeonggi-do 14662, Korea
*
Author to whom correspondence should be addressed.
Electronics 2022, 11(19), 3083; https://doi.org/10.3390/electronics11193083
Submission received: 8 September 2022 / Revised: 22 September 2022 / Accepted: 24 September 2022 / Published: 27 September 2022
(This article belongs to the Special Issue CMOS Chips for Sensing and Communication)

Abstract

:
This paper describes a new readout integrated circuit and non-uniformity correction (NUC) method that ensures that the bolometer array has low non-uniformity over a wide operating temperature range without a thermoelectric cooler (TEC). The proposed NUC minimizes the circuit and memory required for signal processing, making it suitable for compact and power-efficient portable infrared cameras. It corrects the aging phenomenon through start-up calibration and corrects non-uniformities without a TEC through calibration during operation mode. It minimizes the calibration process during operation mode and uses a pixel-level analog-to-digital converter to enable real-time NUC. A 0.18 μm standard CMOS process is applied to the proposed NUC. The frame rate for calibration during the operation mode is approximately 14.3 Hz. The proposed NUC demonstrates excellent uniformity with a non-uniformity of less than 0.12% over a wide operating temperature range (−20 to 50 °C).

1. Introduction

Infrared (IR) cameras, which detect objects with infrared radiation instead of visible light, are widely used in various fields [1]. In particular, the demand for compact and power-efficient portable infrared cameras has increased for use in ultralight portable devices, such as drones or head-mounted displays (HMDs). A cooled IR camera lowers the sensor temperature to cryogenic temperatures, which is necessary to reduce thermally induced noise. However, cryogenic cooling systems consume high power and occupy a large volume. Therefore, considerable research has been conducted on microbolometers, which are uncooled infrared sensors that do not require cryogenic cooling systems [2,3,4,5]. The microbolometer is a representative thermal infrared sensor that uses a heat-sensitive material whose resistance changes with temperature. Amorphous silicon (a-Si), thin-film vanadium oxide (VOx), and titanium oxide (TiOx) are widely used as heat-sensitive materials.
Since two-dimensional microbolometer arrays have large variations in characteristics between pixels, non-uniformity correction is essential for excellent infrared images [6]. However, the uniformity of a microbolometer array is highly sensitive to changes in operating temperature. Figure 1 shows the variation in resistance with temperature for two bolometers. As the change in bolometer resistance according to the operating temperature is highly nonlinear, the offset and gain differences between the two bolometers vary according to the operating temperature. To solve this problem, a thermoelectric cooler (TEC) can be used to maintain the bolometer array at a constant operating temperature. However, because TECs consume high power and occupy a large volume, they cannot be used for applications in ultralight portable devices [7,8].
Because of the well-established silicon manufacturing methods and the simplicity of its structure, the a-Si-based microbolometer array exhibits uniform characteristics [9]. Therefore, good uniformity can be maintained without a TEC by simply adjusting the offset correction coefficient according to the operating temperature [10]. VOx and TiOx microbolometers show superior 1/f noise compared with a-Si microbolometers. However, they show a relatively higher non-uniformity because metal oxides have various phases, and their characteristics change with time due to high bias current. Therefore, good uniformity cannot be obtained only by adjusting the offset correction coefficient for VOx or TiOx microbolometers. The adjustment of the gain correction coefficient is additionally required according to the operating temperature [11]. Calibration for each sensor can be performed without a TEC using linear interpolation to adjust the gain and offset correction coefficients according to the current operating temperature [1]. However, there is an additional cost to find a table of correction coefficients for the operating temperature point. Moreover, time-varying characteristics cannot be calibrated.
In this study, we propose a novel readout integrated circuit (ROIC) with a simple structure capable of efficient non-uniformity correction (NUC) over a wide operating temperature range without using a TEC. Offset and gain can be corrected in real time and even the ‘aging’ phenomenon, in which characteristics change over time, can be corrected. In addition, the use of a pixel-level analog-to-digital converter (ADC) enables fast and accurate correction according to the surrounding environment.

2. Basic Concept

2.1. Monolithic IR Camera Chip

Figure 2 shows an outline of the monolithic IR camera chip to which the proposed ROIC can be applied. To reduce the size, weight, cost, and power consumption of the camera, considerable research on wafer-level cameras has been conducted, and microlens and thin-film packaging technologies are needed to realize it [12,13,14,15]. The goal of this study is to implement an ROIC suitable for the application shown in Figure 2, enabling fast and efficient NUC without a TEC. The conventional shutter, which is essential in a general IR camera, can be replaced with a shutter using a liquid crystal (LC). The LC shutter can be used for the NUC of the microbolometer array along with the IR organic light-emitting diode (OLED). IR OLED has a fixed power in the mid-wave infrared (MWIR) band, and it causes a temperature change of about 1 K in the bolometer.

2.2. Calibration Sequence of the Proposed NUC

Figure 3 shows the calibration sequence of the proposed NUC method, which is divided into calibration in the production phase (a), and calibration when power is supplied during use (b). The IR OLED in Figure 2 is the light source for the start-up calibration in Figure 3b, and there is a variation in the amount of light from the IR OLED detected by each sensor. Therefore, a correction coefficient (βn) to calibrate this variation is necessary and is obtained through the process in Figure 3a. First, as shown in Figure 3a, the operating temperature is adjusted to the standard temperature T0 using a TEC without any light source, and the bias voltage VB0n of each sensor is determined by applying a bias current IB to it. Thereafter, while maintaining the operating temperature T0 and VB0n of each sensor, two currents sensed by each sensor are sequentially obtained using the standard light source ΔT0 and IR OLED, respectively. The standard light is blackbody radiation, and ΔT0 is approximately 1 K. The correction coefficient βn (=ΔT0nT0) of each sensor is obtained by comparing the two currents.
The start-up calibration shown in Figure 3b begins when the camera is turned on for use after the production is completed. As this application does not use a TEC, the operating temperature of the sensor is the same as the external temperature T, and the external light is blocked by the LC shutter during the start-up calibration. First, the operating temperature T is sensed, and the bias voltage VBn of each sensor and correction coefficient γ are determined. Coefficient γ is a function of T and is necessary for real-time NUC. The process of offset correction by maintaining the bias voltage VBn of each sensor if the operating temperature does not change significantly will be explained later. Next, using βn and the current of each sensor obtained using the IR OLED, a coefficient αn for gain correction is calculated. When the operating temperature changes significantly during operation, only VBn and γ must be updated, enabling real-time NUC. Since the correction coefficients are updated each time the camera is turned on, the aging phenomenon of the sensor can also be corrected. As the IR OLEDs are in use only for a short time, they can maintain constant characteristics regardless of camera usage time.

3. Proposed ROIC for Real-Time NUC without a TEC

3.1. Circuit Operation for Calibration in the Production

Figure 4 and Figure 5 show the schematic of the unit cell circuit and block diagram of the overall arrangement for the proposed ROIC, respectively. As the pixel size (17 μm × 17 μm) of the sensor is very small, it is difficult to implement the unit cell circuit shown in Figure 4 within one pixel. Accordingly, 16 (4 × 4) adjacent bolometers used as IR sensors share one unit cell circuit using 16 switches (ϕSR1ϕSR16) in a time-division manner. The bolometer resistance RB at temperature T is given by the following equation:
R B = R o e b k T ,
where Ro and b are constants that are determined by the physical properties of the bolometer materials, and k is the Boltzmann constant [16,17]. Ro and b in Equation (1) show a large difference between bolometers.
During the first phase of Figure 3a for determining VB0n, signals ϕs and ϕB in Figure 4 are set to the logic ‘1’ state, and amplifier A1 is used as a comparator. In this case, the connections depicted by dotted lines among those depicted by arrows in Figure 4 are valid. The bias current IB is supplied to the bolometer of each unit cell using the global current source IB while the operating temperature is T0. VB0n is determined using the following equation:
V B 0 n = I B R o e b k T 0 .
The VB0n of each bolometer is input into a column multiplexer and stored in the VBn data memory outside the unit cell by a single-slope (SS) analog-to-digital (A/D) conversion process. For the SS A/D conversion process, a rising ramp signal is applied to the positive terminal of A1 using a column-level bias circuit. A1 detects when the ramp signal equals VB0n and enables the digital value of VB0n from a 12-bit counter to be stored in 12-bit memory.
During the second phase of Figure 3a for determining βn, signal ϕs is set to the logic ‘0’ state, and amplifiers A1 and A2 are used for the integrator and comparator, respectively. In this case, connections depicted by red and blue lines among those depicted by arrows are valid. Figure 6 shows the timing diagram of the operation mode after the calibration sequence. The same timing is also applied to the second phase of Figure 3a. Since the 16 bolometers share a unit cell circuit, the timing diagram in Figure 6 is for 1/16-th of a frame. When the frame rate of the operation mode is set to 30 Hz, the maximum time that one bolometer can use the unit cell circuit is 2.08 ms. As certain operations cannot be performed on all unit cells simultaneously, all operations are performed with row-level timing for efficiency.
To maintain the VB0n of each bolometer, the value stored in the VBn data memory is periodically transmitted to the capacitance CB of each unit cell circuit using the ramp signal generator of the bias circuit. The transmission of VB0n proceeds row by row, as shown in Figure 6, using signal ϕB and the row number, shown in square brackets ([]) in Figure 6. As the circuit used in the process of determining and transmitting VB0n is the same, an accurate VB0n can be transmitted without additional errors. First, a standard light source ΔT0 is applied. The integrator input current IS0 generated for the bolometer resistance RB01 is given by Equation (3). When ΔT0 (≈1 K) is sufficiently small compared to T0 (≈300 K), Equation (3) can be simply expressed through two approximations as follows:
I S 0 = V B 0 n R B 01 I B = I B R o e b k T 0 R o e b k ( T 0 + Δ T 0 ) I B = I B e b k T 0 e b k ( T 0 + Δ T 0 ) 1 ,
I S 0 I B e b k T 0 e b k T 0 ( 1 Δ T 0 T 0 ) 1 = I B e b k T 0 Δ T 0 T 0 1           I B b k T 0 2 Δ T 0 .
IS0 is integrated into the capacitance CINT through an integrator composed of A1 and several MOSs. A buffered direct injection (BDI) input circuit using a feedback amplifier in a common gate configuration has a very low input resistance and stable bias voltage. A current mirror after BDI is used such that the range of the integration voltage can be maximized regardless of the VB0n value. Since CINT is reset by using A2 and ϕRST, the effect of the A2 offset can be eliminated. After integration, an SS A/D conversion is performed using the row-level circuit for the ADC and A2, and the digitally converted IS0 is transferred to the outside of the unit cell.
Next, the light source is replaced by an IR OLED ΔT0n, and the process in Figure 6 is repeated to obtain the digital value of the corresponding input current IS0n. An approximate expression for IS0n can be obtained from Equations (3) and (4); accordingly, the correction coefficient βn is determined by the following equation:
I S 0 n I B b k T 0 2 Δ T 0 n ,
β n = I S 0 n I S 0 = Δ T 0 n Δ T 0 .
As seen from Equation (6), βn is the normalized value of ΔT0n detected by each sensor, which is stored in a non-volatile memory, such as flash memory.

3.2. Start-Up Calibration

Except for the operating temperature being the external temperature T, the process for determining VBn in the start-up calibration of Figure 3b is the same as that for determining VB0n, as described in Section 3.1. Coefficient γ is required for real-time NUC, and the equations of VBn and γ for temperature T are as follows:
V B n = I B R o e b k T , γ = T 0 2 T 2 .
To determine the coefficient γ, the temperature sensor shown in Figure 5 is used.
Next, ΔT0n is applied, and the process shown in Figure 6 is performed as described in Section 3.1. An approximate expression for the input current IS1n can be derived in the same manner as Equation (5), using the gain correction coefficient αn, which is determined by the following equation:
I S 1 n I B b k T 2 Δ T 0 n ,
α n = 1 I S 1 n β n γ = k I B b T 0 2 Δ T 0 .
Coefficient αn is defined as the reciprocal of the final expression in Equation (4) and includes the gain deviation of each bolometer with respect to the standard light source ΔT0. In Equation (9), IB and b include the deviations for each bolometer, whereas the other parameters are independent of the deviations. As the coefficient αn is maintained only when power is supplied, it is stored in volatile memory, such as SRAM.

3.3. Normal Operation Mode

The LC shutter is kept open and the IR OLED is not used in the normal operation mode. Therefore, the amount of light ΔTn corresponding to the temperature of the target object is transmitted to each bolometer. Using the bias voltage VBn of Equation (7) while the operating temperature does not change at value T, the process shown in Figure 6 is performed as described in Section 3.1. An approximate expression for the input current IS can be derived in the same manner as Equation (5). Consequently, the final result IS_f, corrected using αn and γ, is as follows:
I S I B b k T 2 Δ T n ,
I S _ f = I S × α n × 1 γ = Δ T n Δ T 0 .
As shown in Equation (10), the Ro and offset deviations are eliminated in the expression for IS, which is obtained while maintaining the VBn determined according to the method described above. As shown in Equation (11), the effects of the operating temperature T, b, IB, and gain deviation are eliminated to obtain a result proportional to the temperature of the target object.

3.4. VBn Adjustment and Real-Time NUC

When the operating temperature changes significantly in the operation mode, a large offset current is added to Equation (11). Thus, the dynamic range of the input current is greatly reduced. Therefore, the bias voltage VBn should be adjusted if the operating temperature changes by more than 2 K relative to the previous temperature. If the operating temperature changes to T’, VBn and γ are adjusted as in the start-up calibration described in Section 3.2, and the following equation is obtained from Equation (7):
V B n = I B R o e b k T , γ = T 0 2 T 2 .
The adjustment of αn is not required, unlike in the start-up calibration. Therefore, it is possible to reduce the calibration time during the operation and enable real-time NUC.
Subsequently, the operation mode proceeds in the same manner as described in Section 3.3. The equations of the bolometer resistance RB′ for ΔTn at the operating temperature T′ and IS obtained from RB′ are as follows:
R B = R o e b k ( T + Δ T n ) ,
I S = V B n R B I B I B b k T 2 Δ T n .
Next, the equation for IS_f is corrected using the value of αn in Equation (9), determined during the start-up correction, as follows:
I S _ f = I S × α n × 1 γ = Δ T n Δ T 0 .
As shown in Equation (15), all non-uniformities can be corrected by adjusting only VBn and γ as the operating temperature changes.

4. Simulation Results

The proposed ROIC is designed using a 0.18 μm standard CMOS process. Table 1 summarizes the design parameters and characteristics of the target bolometer and the ROIC.
The resistance RB of the target VOx bolometer is modeled using Equation (1). Four bolometers are set up with a variation in Ro and b in Equation (1) of 10% each, and their characteristics are shown in the lower part of Figure 7a. If IB is independent of the operating temperature, the change in VBn according to the operating temperature is equal to the change in RB, as shown in the upper part of Figure 7a. In this case, it is difficult to design a bias circuit and integrator because the range of VBn is very wide. As shown in Figure 7b, the range of VBn can be significantly reduced by dividing the range of operating temperatures into four sections and using different IB values for each. IB values for each section do not need to be constant because the deviation of IB is corrected, as demonstrated in Section 3.
Figure 8 shows the process of determining VBn during start-up calibration. After VBn is determined by IB, a rising ramp signal, VRamp, is supplied from the bias circuit. When the VRamp value becomes equal to VBn, the output voltage VA1 of comparator A1 changes, and the digital value of VBn is stored in the 12-bit memory of each unit circuit. Considering the speed of comparator A1, the step time of VRamp is set to 1 μs, and the time taken to determine the VBn of the entire array is less than 70 ms. Therefore, the frame rate for determining VBn is approximately 14.3 Hz, and real-time NUC is possible, as mentioned in Section 3.4. The DC offset effect of comparator A1 observed in Figure 8 compensates for itself when fed to the corresponding bolometer after VBn transmission.
Figure 9 shows the simulated waveforms for verifying the unit cell circuit in the operation mode, which were previously described in Figure 4 and Figure 6. The period for VBn transmission is set using signal ϕB, and the IS integration time is determined using signals ϕRST and ϕH. When ϕH changes to logic ‘0′, the integration voltage V(CINT) remains constant. After the integration of IS, a 12-bit SS A/D conversion is performed using a row-level circuit for the ADC and comparator A2. The output voltage VA2 of A2 controls the enabled terminal of the 12-bit memory.
Figure 10 shows the mask layout of the proposed unit cell circuit shown in Figure 4. As the pixel size is (17 μm)2, the size of the unit cell circuit is {68 (= 17 × 4) μm}2. The integrator and comparator in Figure 10 are marked based on the operation mode, and the capacitors occupy most of the layout.
To verify the proposed NUC method, corner simulations are performed using the Cadence Virtuoso. For a conventional calibration method that uses a single correction coefficient table to perform a two-point correction, if the operating temperature changes in the absence of a TEC, the correction table will not fit, and the non-uniformity will increase. Figure 11 shows the non-uniformity of the final digital signal obtained by applying the conventional calibration method to a bolometer with non-uniformity, as shown in Figure 7. The correction coefficient is set according to the operating temperature of 20 °C, and Figure 11 shows that the greater the difference between the operating temperature and 20 °C, the greater the non-uniformity. The offset correction in Figure 11 uses the VBn adjustment method, as in the proposed method. In this case, the charge feed-through and leakage current caused by the ϕB switch generated during the VBn transmission-and-hold process may affect the correction accuracy. Therefore, three types of ϕB switches are designed, which are denoted as A, B, and C in Figure 11, indicating an NMOS switch, CMOS switch, and NMOS switch with a dummy switch, respectively.
Figure 12a shows that non-uniformity is greatly decreased by the proposed NUC. A table for βn is obtained by the calibration in production, as shown in Figure 3a, at an operating temperature of 25 °C, and αn is obtained by the start-up calibration shown in Figure 3b, at an operating temperature of 20 °C. Table 2 shows the βn and αn for the four bolometers shown in Figure 7. Subsequently, the non-uniformity according to the operating temperature is observed by applying the same αn, regardless of the operating temperature. When using the C switch, the transmission and hold of VBn are most accurate, and non-uniformity also decreases relatively. The noise equivalent temperature difference (NETD), which can be used as a measure of non-uniformity, is the smallest possible temperature difference that a thermal imager can resolve. The smaller the NETD value, the higher the image resolution. The most frequently used limiting criteria for NETD are 50 and 100 mK, and the corresponding non-uniformities for these two values are shown in Figure 12a. Figure 12b compares the non-uniformity between the conventional NUC and the proposed NUC and shows the superiority of the proposed NUC.
Table 3 shows the performance comparison between the proposed NUC and other methods.

5. Conclusions

We propose a novel NUC capable of accurate and efficient calibration without a TEC that is suitable for compact and power-efficient portable infrared cameras. It corrects not only the basic offset and gain non-uniformity but also the aging phenomenon. The input current is integrated after the bias current is skimmed during offset correction. Thus, high-frequency noise is reduced by using sufficient integration time in a limited pixel area. The basic feasibility of the proposed NUC is verified mathematically. The ROIC and driving method are proposed to efficiently implement the proposed NUC and improve the calibration speed and NETD. The frame rate for the in-use calibration is approximately 14.3 Hz; therefore, real-time NUC is possible. The proposed NUC shows excellent non-uniformity characteristics corresponding to an NETD of less than 50 mK over a wide operating temperature range (−20 to 50 °C).

Author Contributions

D.W. supervised the research. J.D.Y. and D.W. proposed the idea and designed the circuit. J.D.Y. performed the simulation and mask layout. J.D.Y. and D.W. wrote the initial manuscript. D.W. revised and finalized the paper. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Korean government (MSIT), grant number 2020R1F1A1052571.

Data Availability Statement

Not applicable.

Acknowledgments

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (No. 2020R1F1A1052571). This work was supported by the Catholic University of Korea, Research Fund, 2021. The EDA tool was supported by the IC Design Education Center (IDEC), Korea.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Li, C.; Han, C.J.; Skidmore, G.D. Overview of DRS uncooled VOx infrared detector development. Opt. Eng. 2011, 50, 061017. [Google Scholar] [CrossRef]
  2. Akin, T. Low-Cost LWIR-Band CMOS Infrared (CIR) Microbolometers for High Volume Applications. In Proceedings of the IEEE 33rd International Conference on MEMS, Vancouver, BC, Canada, 18–22 January 2020; pp. 147–152. [Google Scholar]
  3. Abbasi, S.; Shafique, A.; Ceylan, O.; Gurbuz, Y. A Digital Readout IC for Microbolometer Imagers Offering Low Power and Improved Self-Heating Compensation. IEEE Sens. J. 2020, 20, 909–917. [Google Scholar] [CrossRef]
  4. Yen, P.; Su, G.J. Uncooled Long-Wavelength Infrared Sensors Using Cytochrome C Protein on Suspension Electrodes with CMOS Readout Circuits. IEEE Sens. J. 2019, 19, 10221–10227. [Google Scholar] [CrossRef]
  5. Jiang, Z.; Men, L.; Wan, C.; Xiao, P.; Jiang, C.; Tu, X.; Jia, X.; Kang, L.; Li, L.; Chen, J.; et al. Low-Noise Readout Integrated Circuit for Terahertz Array Detector. IEEE Trans. Terahertz Sci. Technol. 2018, 8, 350–356. [Google Scholar] [CrossRef]
  6. Li, J.; Qin, H.; Yan, X.; Zeng, Q.; Yang, T. Temporal-Spatial Nonlinear Filtering for Infrared Focal Plane Array Stripe Nonuniformity Correction. Symmetry 2019, 11, 673. [Google Scholar] [CrossRef]
  7. Howard, P.E.; Clarke, J.E.; Bradley, M.G.; Ionescu, A.C.; Li, C. Progress in uncooled focal plane sensor technology at Boeing. In Proceedings of the International Symposium on Optical Science and Technology, San Diego, CA, USA, 29 July–3 August 2001; pp. 168–174. [Google Scholar]
  8. Parrish, W.J.; Woolaway, J.T. Improvements in uncooled systems using bias equalization. In Proceedings of the AeroSense ‘99, Orlando, FL, USA, 26 July 1999; pp. 748–755. [Google Scholar]
  9. Alicandro, C.J.; DeMarco, R.W. 1024 × 768 XGA uncooled camera core achieves new levels of performance in a small package. In Proceedings of the SPIE Defense, Security, and Sensing, Orlando, FL, USA, 20 May 2011; p. 80121H. [Google Scholar]
  10. Dupont, B.; Dupret, A.; Becker, S.; Hamelin, A.; Imperinetti, P.; Rabaud, W. A [10 °C; 70 °C] 640 × 480 17 μm pixel pitch TEC-less IR Bolometer Imager with below 50mK and below 4V power supply. In Proceedings of the IEEE International Solid-State Circuits Conference, San Francisco, CA, USA, 17–21 February 2013; pp. 394–395. [Google Scholar]
  11. Kwon, M.H.; Yang, K.; Park, Y.S.; Kim, Y.H.; Chung, H. Investigations of reactively sputtered TiO2-δ films for microbolometer application. In Proceedings of the SPIE Security + Defense, Cardiff, Wales, UK, 2 October 2008; p. 711306. [Google Scholar]
  12. Zoberbier, R.; Hennemeyer, M.; Toennies, D.; Kraft, A.; Eisner, M.; Völkel, R. Technology trends in the manufacturing and packaging of Wafer Level Cameras. In Proceedings of the IEEE 3rd Electronics System Integration Technology Conference ESTC, Berlin, Germany, 13–16 September 2010; pp. 1–5. [Google Scholar]
  13. Symmons, A.; Pini, R. A practical approach to LWIR wafer-level optics for thermal imaging systems. In Proceedings of the SPIE Defense, Security, and Sensing, Baltimore, MD, USA, 18 June 2013; p. 870425. [Google Scholar]
  14. Kim, K.; Jang, K.W.; Ryu, J.K.; Jeong, K.H. Biologically inspired ultrathin arrayed camera for high-contrast and high-resolution imaging. Light Sci. Appl. 2020, 9, 28. [Google Scholar] [CrossRef] [PubMed]
  15. Jeon, G.J.; Kim, W.Y.; Lee, H.C. Thin-film vacuum packaging based on porous anodic alumina (PAA) for infrared (IR) detection. In Proceedings of the IEEE SENSORS, Taipei, Taiwan, 28–31 October 2012; pp. 1–4. [Google Scholar]
  16. Rogalski, A. Infrared Detectors; Gordon and Breach Science Publishers: Amsterdam, The Netherlands, 2000. [Google Scholar]
  17. Abdel-Rahman, M.; Zia, M.; Alduraibi, M. Temperature-Dependent Resistive Properties of Vanadium Pentoxide/Vanadium Multi-Layer Thin Films for Microbolometer & Antenna-Coupled Microbolometer Applications. Sensors 2019, 19, 1320. [Google Scholar]
  18. Lv, J.; Que, L.; Wei, L.; Zhou, Y.; Liao, B.; Jiang, Y. Uncooled Microbolometer Infrared Focal Plane Array Without Substrate Temperature Stabilization. IEEE Sens. J. 2014, 14, 1533–1544. [Google Scholar] [CrossRef]
Figure 1. Non-uniform resistance changes in two bolometers according to the operating temperature.
Figure 1. Non-uniform resistance changes in two bolometers according to the operating temperature.
Electronics 11 03083 g001
Figure 2. Monolithic IR camera chip.
Figure 2. Monolithic IR camera chip.
Electronics 11 03083 g002
Figure 3. Calibration sequence of the proposed NUC method: (a) calibration in the production phase; (b) start-up calibration.
Figure 3. Calibration sequence of the proposed NUC method: (a) calibration in the production phase; (b) start-up calibration.
Electronics 11 03083 g003
Figure 4. Schematic of the unit cell circuit for the proposed ROIC.
Figure 4. Schematic of the unit cell circuit for the proposed ROIC.
Electronics 11 03083 g004
Figure 5. Block diagram of the overall arrangement for the proposed ROIC.
Figure 5. Block diagram of the overall arrangement for the proposed ROIC.
Electronics 11 03083 g005
Figure 6. Timing diagram of the operation mode.
Figure 6. Timing diagram of the operation mode.
Electronics 11 03083 g006
Figure 7. Variations and changes in RB and VBn according to operating temperature: (a) When IB is independent of operating temperature; (b) when IB varies according to temperature section.
Figure 7. Variations and changes in RB and VBn according to operating temperature: (a) When IB is independent of operating temperature; (b) when IB varies according to temperature section.
Electronics 11 03083 g007
Figure 8. Simulated waveforms for determining VBn.
Figure 8. Simulated waveforms for determining VBn.
Electronics 11 03083 g008
Figure 9. Simulated waveforms for the operation mode of the unit cell circuit.
Figure 9. Simulated waveforms for the operation mode of the unit cell circuit.
Electronics 11 03083 g009
Figure 10. Mask layout of the unit cell circuit.
Figure 10. Mask layout of the unit cell circuit.
Electronics 11 03083 g010
Figure 11. Non-uniformity of the conventional NUC.
Figure 11. Non-uniformity of the conventional NUC.
Electronics 11 03083 g011
Figure 12. Non-uniformity according to operating temperature: (a) for the proposed NUC; (b) for the two methods using the ‘C’ switch.
Figure 12. Non-uniformity according to operating temperature: (a) for the proposed NUC; (b) for the two methods using the ‘C’ switch.
Electronics 11 03083 g012
Table 1. Design parameters and characteristics of the target bolometer and the ROIC.
Table 1. Design parameters and characteristics of the target bolometer and the ROIC.
ParameterValue
Bolometer materialVOx
Resistance200 kΩ (@ 300 K)
Thermal conductance0.5 × 10−8 W/K
Emissivity80%
Temperature coefficient of resistance2.0%/K
Optics F-number1
Spectral range3–8 μm
Pixel size17 μm × 17 μm
Array size640 × 512
Frame rate30 Hz
Power consumption per pixel <0.61 µW
Table 2. βn and αn for the four bolometers shown in Figure 7.
Table 2. βn and αn for the four bolometers shown in Figure 7.
Ro of RBβnαn
3151.0513.197 × 107
3250.94893.547 × 107
3300.94913.366 × 107
3451.0503.193 × 107
Table 3. Performance comparison.
Table 3. Performance comparison.
[1][10][18]This Work
Pixel pitch17 μm17 μm25 μm17 μm
BolometerVOxa-SiVOxVOx
Power supplyN/A4 V9 V1.8 V
NETDN/A<47 mK<63 mK<50 mK
Operating temperature [°C]N/A10 to 70 −20 to 60−20 to 50
Aging correctionNoNoNoYes
Applicability to monolithic IR camera chipNoYesYesYes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Yeo, J.D.; Woo, D. Real-Time Non-Uniformity Correction without TEC for Microbolometer Array. Electronics 2022, 11, 3083. https://doi.org/10.3390/electronics11193083

AMA Style

Yeo JD, Woo D. Real-Time Non-Uniformity Correction without TEC for Microbolometer Array. Electronics. 2022; 11(19):3083. https://doi.org/10.3390/electronics11193083

Chicago/Turabian Style

Yeo, Jun Dong, and DooHyung Woo. 2022. "Real-Time Non-Uniformity Correction without TEC for Microbolometer Array" Electronics 11, no. 19: 3083. https://doi.org/10.3390/electronics11193083

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop