The Alveolar Gas Monitor: An Alternative to Pulse Oximetry for the Noninvasive Assessment of Impaired Gas Exchange in Patients at Risk of Respiratory Deterioration

Health Division of Pulmonary, Critical Care, Sleep Medicine, and Physiology, University of California San Diego, La Jolla, CA 92093, USA
*
Author to whom correspondence should be addressed.
J. Clin. Med. 2025, 14(16), 5880; https://doi.org/10.3390/jcm14165880
Submission received: 24 April 2025 / Revised: 29 May 2025 / Accepted: 13 August 2025 / Published: 20 August 2025
(This article belongs to the Section Respiratory Medicine)

Abstract

Background/Objectives: The COVID-19 pandemic highlighted the limitations of pulse oximetry in detecting occult hypoxemia. The superiority of the alveolar gas monitor (AGM) compared to pulse oximetry (SpO2) in predicting respiratory deterioration among COVID-19-positive individuals has previously been demonstrated. Here, we combine COVID-19 and non-COVID-19 individuals as a combined cohort of participants to determine if the AGM has similar utility across a larger, more generalizable cohort. Methods: Adult patients (n = 75) at risk of respiratory deterioration in the emergency department (ED) underwent prospective assessments of their oxygen deficit (OD) and SpO2, simultaneously measured during quiet breathing on the AGM. The OD and SpO2 were then compared for their ability to predict the dichotomous outcome of the need for supplemental oxygen. The administration of supplemental oxygen was ordered by the clinical care team with no knowledge of the patients’ enrollment in this study. Results: In the logistic regression analysis, both SpO2 and OD significantly predicted the need for supplemental oxygen among COVID-19-negative individuals. However, in the multivariable regression, only OD (p < 0.001) significantly predicted the need for supplemental oxygen, while SpO2 (p = 0.05) did not in the combined cohort of COVID-19-negative and -positive individuals. Receiver operating characteristic (ROC) curve analysis demonstrated the superior discriminative ability of OD (area under ROC curve = 0.937) relative to SpO2 (area under ROC curve = 0.888) to predict the need for supplemental oxygen. Conclusions: The noninvasive AGM, which combines the measurement of exhaled partial pressures of gas with SpO2, outperforms SpO2 alone in predicting the need for supplemental oxygen among individuals in the ED at risk of respiratory deterioration regardless of the etiology for their symptoms (COVID-19-positive or -negative).

1. Introduction

The COVID-19 pandemic exposed unexpected limitations in the current healthcare system’s resource availability, both at the supply and personnel levels. The overwhelming burden that COVID-19 placed on the healthcare system underscored the need for efficient and safe tools to quickly triage and prioritize the care of patients at risk of respiratory failure. Although a very useful tool for assessing hypoxemia, pulse oximetry has many limitations compared to arterial blood gas (ABG) sampling, potentially resulting in inaccurate measurements due to metabolic disturbances, skin pigmentation, and vascular function [1].
We recently reported that the alveolar gas monitor, or AGM (MediPines, Yorba Linda, CA, USA), outperformed SpO2 as a noninvasive method for analyzing important physiological endpoints useful in predicting the risk of respiratory failure in patients with COVID-19 [2]. Such endpoints provide prognostic information that can be used to anticipate those who are likely to deteriorate versus those who are likely to recover spontaneously. The AGM is further advantageous because it does not require ABG sampling [3,4,5], making it highly accessible in under-resourced areas or in the field away from healthcare systems. However, the prior report stemmed from a study with a modest sample size, leaving a need for further data collection.
John B. West and others developed the AGM to obviate the need for ABG sampling in some cases as ABGs are invasive, costly, not without risk [6], and are being performed less frequently [7]. The AGM has been used with excellent accuracy and reliability in healthy outpatients, both young and old people, hypoxic individuals, volunteers exercising at altitude and in hypoxic conditions, patients with parenchymal lung disease in the outpatient setting, and COVID-19 patients with dyspnea in the ED. In these prior studies, the gPaO2 has correlated closely with the PaO2 from an ABG, and the OD has correlated closely with the AaDO2.
The AGM measures the exhaled tensions (partial pressure in mmHg) of oxygen (O2) and carbon dioxide (CO2), along with other parameters such as the respiratory quotient, oxygen saturation, respiratory rate, barometric pressure, inspired pressure of oxygen, and pulse rate. The O2 deficit (OD) is an AGM output that is used as a surrogate for the alveolar–arterial O2 gradient (AaDO2 or Aa gradient) based on the difference between the alveolar pressure of oxygen (PAO2) and the calculated arterial oxygen pressure (gPaO2). The calculation of gPaO2 makes use of measured pulse oximetry (SpO2), as detailed below in Section 2.
Use of the OD as a surrogate for the AaDO2 is relevant as ventilation–perfusion (VQ) disruption in the lung, specifically shunt and low VQ, will lead to widening of the Aa gradient due to lower PaO2 and increasing PAO2. By focusing on this difference, we are potentially more sensitive to VQ disruption than by being reliant on SpO2 alone. Moreover, even if SpO2 is adversely affected by the aforementioned factors, PAO2 is not; therefore, by focusing on a widened gradient, we are potentially more sensitive to disruptions in VQ. Abnormalities of the parenchyma or vasculature could lead to pathological VQ inequity, and this likely occurs prior to abnormalities on imaging. Therefore, in acute respiratory illnesses presenting with dyspnea, early sensitivity to VQ disruption could have diagnostic and therapeutic implications. Finally, the exhaled gas values may circumvent issues around skin pigmentation, a recent focus related to pulse oximetry limitations [1].
Despite the positive results in our AGM COVID-19 manuscript, with waning cases and ongoing debate as to whether COVID-19 respiratory failure is different from other causes of respiratory failure, we felt it was important to broaden our investigation to non-COVID patients. This is particularly germane as methods for efficient patient triage could be used in future disaster situations regardless of the etiology of respiratory failure.
Based on this conceptual framework, we sought to test the hypothesis that the OD was superior to SpO2 in predicting the need for supplemental oxygen in a real-world setting among patients with respiratory symptoms at risk of respiratory deterioration. Such findings would provide validation of our prior COVID-19 AGM report [2] and improve the generalizability of our findings to non-COVID-19 respiratory illnesses.

2. Methods

2.1. Study Design

This was a diagnostic accuracy study performed in a prospective fashion comparing the OD to SpO2 to predict respiratory decline, as defined by requiring supplemental O2 administration within 24 h after measuring the OD. Requiring supplemental oxygen was defined as any amount of oxygen support above room air, and included nasal cannulas, face masks, high-flow devices, noninvasive and invasive ventilation, and extracorporeal membrane oxygen. This study was approved by our Institutional Review Board, and all participants provided signed informed consent. The decision to administer supplemental O2 was made by the clinical care team, who had no knowledge of patients’ enrollment in this study. Further, the research team had no knowledge of the treating physician’s decision to administer supplemental O2. Administration of supplemental oxygen was ascertained by medical record review at least 24 h after measurement of the OD.

2.2. Participants

All patients were at least 18 years old and were approached for enrollment in the Emergency Department (ED) of two academic medical centers between November 2020 and November 2023. Patients were initially identified by screening the ED census for a chief symptom of dyspnea, as this is a symptom that is more likely to have impairments in gas exchange than, for example, abdominal pain. The medical record was then assessed for inclusion and exclusion criteria prior to approaching the patient for enrollment in this study. Inclusion criteria were afebrile adults who were not on chronic oxygen therapy with a recorded SpO2 (on the hospital pulse oximeter) < 97% on room air with a chief symptom of dyspnea. Exclusion criteria included being on any amount of supplemental oxygen, a resting SpO2 < 88% even if on room air, at-risk patients as defined by the Department of Health and Human Services Common Rule (i.e., 45 CFR 46 Subparts A–D), chronic O2 therapy, use of medications that could interfere with the accuracy of SpO2 (e.g., intravenous dyes, vasodilators, and vasopressors), facial trauma precluding the use of a nose clip, and advanced neuromuscular weakness preventing an adequate mouthpiece seal. It is important to note that the chief symptom of dyspnea or shortness of breath was only used to screen the ED patient board for possible candidates for this study, not to draw any conclusions about a patient’s dyspnea and their impairments in gas exchange. Of note, many patients had an increased O2 requirement between the time of medical record screening and the investigator’s arrival in the ED for enrollment, which precluded such patients from participating in this study. This issue, combined with investigator availability, necessitated convenience sampling. Participant safety was ensured by stopping the experimental protocol if a patient’s SpO2 fell below 88% on the hospital pulse oximeter, regardless of the SpO2 on the AGM.

2.3. Test Methods

In this diagnostic accuracy study, the index test (AGM) was compared to a reference test (pulse oximetry). The index variable of interest (OD) was compared to the reference variable of interest (SpO2), as displayed on the AGM.

2.4. Device Details

The expired pressures of CO2 and O2 are measured using rapid gas analyzers built into the housing of the AGM using a pump-driven, active sampling line connected at the midpoint of the mouthpiece, with data sampled at 100 Hz (see Figure 1). Infrared CO2 detection and fuel cell O2 analysis are used in the AGM to measure CO2 and O2 values, respectively.

2.5. Index Test Details

The AGM was used with the participant sitting semi-recumbent in bed to optimize functional residual capacity as much as possible, given their symptomatic state. Participants were instructed to perform quiet breathing as if they were reading a book to minimize breath-to-breath variability. If participants had difficulty maintaining a steady respiratory rate and depth, investigators played a metronome on their cellphones to aid participants in their respiratory pattern. Tidal breathing was performed on room air with a nose clip and a tight mouthpiece seal. The OD was derived by subtracting the calculated arterial partial pressure of oxygen (gPaO2) from the alveolar partial pressure of oxygen (PAO2). The gPaO2 was calculated from SpO2 rather than by direct measurement (PaO2), as is done with an ABG. Previous details about this calculation have been published by our group [8]. In summary, gPaO2 is an algebraic solution of the logarithm of the Hill equation using a Hill n of 2.88 for O2 based on work by Severinghaus [9,10,11]. The gPaO2 calculation also factors in the effects of PACO2 and, by proxy, PaCO2 on the oxygen tension at which point the O2–hemoglobin dissociation curve is 50% saturated (P50) based on research by Kelman [12,13,14]. The PAO2 was calculated by averaging end tidal O2 values from 5 consecutive breaths if and only if the patient and machine had reached a steady state [15] (see Figure 1).
Steady state was determined by the absence of variation in the end-tidal CO2 (etCO2) signal over the preceding 45 s. Stability in the etCO2 signal is contingent on the stability of the respiratory rate, tidal volume, and respiratory quotient. Prior research by our group has demonstrated good stability of etCO2 during steady-state breathing and is therefore usable in this context. A stable respiratory rate and depth (“steady state”) can usually be achieved in between one and four minutes. PACO2 was calculated much like PAO2, by averaging etCO2 values from five consecutive steady-state breaths. All the values from the AGM were saved in a .csv data file, exported onto a secure flash drive, and uploaded onto a research computer for analysis. This method obviated any transcription errors in recording values from the AGM.
A dummy continuous variable, ODFlip, was created by subtracting the OD from 100. Given that a higher OD is indicative of worse gas exchange, and a higher SpO2 is indicative of better gas exchange, ODFlip allowed plotting along the same axis in receiver operating characteristic (ROC) curves. Threshold values for ODFlip were then converted back to threshold values for the OD by subtracting ODFlip from 100.

2.6. Reference Test Details

The reference variable, SpO2, used for statistical analysis was obtained from the AGM rather than the SpO2 of the hospital pulse oximeter. This approach was chosen intentionally so that the only additional variables separating the OD from SpO2 were partial pressures of exhaled gases. For this reason, the superiority of the OD relative to SpO2 in predicting the outcome variable would support the clinical relevance of exhaled gas monitoring. SpO2 was also monitored using hospital pulse oximeters due to standard of care, and the decision to stop early was based on an SpO2 < 88% on the hospital pulse oximeter, irrespective of the SpO2 on the AGM. The hospital pulse oximeters used in this study were manufactured by three different companies (Massimo (Brisbane City, QLD, Australia), Nonin (Plymouth, MN, USA), and Welch Allyn (Auburn, NY, USA)).

2.7. Statistical Analysis

All statistical analyses were performed in R (Version 4.3.3) or SPSS (Version 29.0.2.0). Logistic regression analyses with Bonferroni correction were performed with a statistical significance threshold of p < 0.01. The informedness of the model was based on Youden’s J statistic in the ROC curve analysis.

3. Results

Among the non-COVID-19 cohort (n = 45), three patients desaturated to SpO2 < 88% after enrollment and during data acquisition on the AGM. Since they did not reach a steady state, a usable AGM data file was not generated, and they were censored from the analysis. The remaining non-COVID-19 patients (n = 42) were combined with our previously described COVID-19 patients (n = 30) into a combined cohort (n = 72) for all subsequent analyses. Several variables of interest [gender, age, self-reported race, self-reported ethnicity, and body mass index (BMI)] between those who did and did not have COVID-19 were investigated to assess the generalizability of this study (see Table 1).
Three binomial logistic regressions were performed to ascertain the effects of the continuous index variable (ODFlip) and the continuous reference variable (SpO2) on the likelihood that participants would require supplemental O2 (the dichotomous dependent variable). The first logistic regression was performed on COVID-19-positive participants (n = 30) from our original study [2]. The second was performed on COVID-19-negative participants (n = 42), and the third was performed on the pooled cohort of all participants (n = 72). The linearity of the continuous independent variables with respect to the logit of the dependent variable was assessed via the Box–Tidwell procedure [16]. Bonferroni correction was applied using all five terms in each model, resulting in statistical significance being accepted when p < 0.01. Based on this assessment, both independent variables were found to be linearly related to the logit of the dependent variable. In short, as SpO2 increases and ODFlip increases, the need for supplemental oxygen decreases. Conversely, as OD increases, the need for supplemental oxygen also increases (see Table 2).
In the COVID-19-negative cohort, for each one-point increase in OD, the odds of needing supplemental O2 increased by 1.14 (95% CI, 1.03 to 1.26) (see Table 3). In the overall cohort, for each one-point increase in OD, the odds of needing supplemental O2 increased by 1.17 (95% CI, 1.07 to 1.27) (see Table 4).
A multivariable logistic regression was then run in the combined cohort to evaluate ODFlip and SpO2 in combination, demonstrating strong statistical significance for ODFlip (p < 0.001) and borderline statistical significance for SpO2 (p = 0.05) in predicting the need for supplemental oxygen.
An ROC curve was generated to compare the overall discrimination of SpO2 and ODFlip on the need for supplemental O2 among COVID-19-negative participants (n = 42). The area under the ROC curve (AUROC) for ODFlip was 0.896 (95% CI, 0.712 to 0.966) with a Youden’s Index of 0.664 at an OD of 41 mmHg, resulting in a sensitivity of 0.840 and a 1–specificity of 0.176. The AUROC for SpO2 was 0.839 (95% CI, 0.800 to 0.993) with a Youden’s Index of 0.565 at an SpO2 of 94%, resulting in a sensitivity of 0.800 and a 1–specificity of 0.235.
An ROC curve was generated to compare the overall discrimination of SpO2 and ODFlip on the need for supplemental O2 among all participants (n = 72). The AUROC for ODFlip was 0.937 (95% CI, 0.885 to 0.990) with a Youden’s Index of 0.754 at an OD of 37 mmHg, resulting in a sensitivity of 0.811 and a 1–specificity of 0.057. The AUROC for SpO2 was 0.888 (95% CI, 0.812 to 0.964) with a Youden’s Index of 0.668 at an SpO2 of 94%, resulting in a sensitivity of 0.811 and a 1–specificity of 0.143.
These two ROC curves demonstrate similar discrimination to the ROC curve previously published in COVID-19-positive participants [2] (see Figure 2). In all three ROC curves, the discriminative ability of the OD was statistically superior to SpO2.

4. Discussion

In this study, we have demonstrated that the OD can predict who will require supplemental oxygen based on univariable and multivariable regression and that it is superior to SpO2 for predicting this outcome based on the statistical parameters of receiver operating characteristic curve analysis.
This study is important for several reasons. First, we have extended our previous observations from COVID-19-positive patients by showing a significant AUROC for predicting respiratory deterioration via the AGM method in COVID-19-negative patients. In our cohort of non-COVID-19 patients, the etiologies of dyspnea included anemia, asthma, chronic obstructive pulmonary disease, granulomatous lung disease, heart failure, interstitial lung disease, pleural effusion, pneumonia (bacterial and non-COVID-19 viral), pulmonary embolism, and pulmonary hypertension among those patients whose etiology for their dyspnea could reliably be identified. This noninvasive tool may be a reasonable method to obviate the need for ABGs in some cases. For example, the AGM could be used in various clinical settings to help with care escalation and triage. In the outpatient clinic, it could be used to identify patients with mildly impaired gas exchange from their underlying chronic diseases. These baseline measurements could then be used as reference points for subsequent visits by these same patients. In the emergency department, it could be used to help with decisions about care escalation, such as providing supplemental oxygen, as we did in this study. On the hospital ward, it could be used for frequent, repeated measurements of gas exchange in hospitalized patients to assess for recovery from, or the worsening of, their acute respiratory illness. Moreover, in places without significant healthcare access (e.g., rural locations and under-developed countries), the AGM could be used in lieu of an ABG. Second, combining data from our pilot (COVID-19) dataset and confirmatory (non-COVID-19) dataset, our AUROC is outstanding and exceeds the highest threshold for discrimination according to Hosmer and Lemeshow [17]. We consider this finding an improvement upon the AUROC presented in our first report [2], where we demonstrated a trend toward significance but did not meet the threshold for statistical significance (p = 0.057) and had broad 95% confidence intervals (0.571 to 1.009) in the logistic regression. Moreover, the OD threshold with the maximal Youden’s index from our first manuscript is identical to the OD threshold in the combined dataset, with outstanding sensitivity, specificity, a positive predictive value, and a negative predictive value. Finally, the AGM has great potential to identify impending respiratory failure in a noninvasive manner, irrespective of underlying cause, which is of value in ongoing respiratory infectious diseases, as well as any future pandemics.
In prior studies, the gPaO2 has correlated closely with the PaO2, and the OD has correlated closely with the AaDO2 [2]. Use of the OD as a surrogate for the AaDO2 is relevant as ventilation-perfusion (VQ) disruption in the lung, specifically shunt and low VQ, will lead to widening of the Aa gradient due to a lower PaO2 and increasing PAO2. By focusing on this difference, we are potentially more sensitive to VQ disruption than being reliant on SpO2 alone. Finally, the exhaled values may circumvent issues around skin pigmentation, a recent focus related to pulse oximetry limitations, though more data are needed.
Various techniques have been used to triage patients at risk of respiratory failure. The Lung Injury Prediction Score (LIPS) was developed for acute lung injury (ALI) prediction in 2010 and 2011, prior to the Berlin definition of acute respiratory distress syndrome (ARDS) subsuming ALI in 2012. The AUROC in the single-center derivation study [18] for LIPS was 0.84, while the AUROC in the multi-center validation study [19] was 0.82. However, in the larger validation study, the sensitivity was 69% and the specificity was 78%. Finally, the model could only be applied to patients with a clearly defined risk factor for ALI/ARDS, thereby limiting its generalizability. More recently, the ratio of oxygen saturation (ROX) index has gained attention for its efforts to discriminate who would undergo intubation among patients requiring high-flow nasal oxygen for hypoxemic respiratory failure. Initially derived in 2016 [20], the ROX index had a modest predictive accuracy, with an AUROC of 0.74. The multi-center validation cohort of 2019 [21] demonstrated an AUROC of 0.759. However, the ROX index has limited applications for three reasons. First, it relies only on SpO2, which has the limitations described above. Second, it assumes the fraction of inspired oxygen based on the liters per minute of oxygen flowing through nasal cannula tubing. Third, it necessitates an already very ill patient needing a high degree of respiratory support before it can be applied for predictive modeling.
We and others have been using deep learning methods to identify patients at risk of respiratory failure. In Chest, Shashikumar et al. reported [22] excellent AUROCs (0.895 for derivation and 0.882 for validation) for all patients and even stronger AUROCs for COVID-19 patients (0.918 to 0.943 during prospective validation) for a deep learning model that consistently outperformed expert clinicians and established predictors, including the ROX index. The deep learning model was validated in COVID-19 and non-COVID-19 patients from two major academic medical centers. The implementation of this technique is ongoing but may well be supplemented with the use of the AGM to further improve the predictive models. Moreover, data enrichment using the AGM may improve actionability of the deep learning methods (e.g., clinicians could use the OD to determine optimal interventions prior to obvious respiratory deterioration).
Despite our study’s strengths, we acknowledge a few limitations. First, while noninvasive and rapid, the AGM requires patient engagement and some degree of stability in the respiratory rate and pattern. This issue limits its utility as a continuous, passive screening tool, such as the fingertip pulse oximeter; however, we believe it can still be used in a screening capacity by bedside nurses, respiratory therapists, and physicians at intervals, much like routine vital signs, and advocate for its addition to, not replacement of, pulse oximetry. Second, although our goal was to enroll a cohort with rich diversity, we ultimately included only ten black participants (13.7%) plus three of mixed race, which is more than predicted for San Diego County but could clearly be improved with multicenter studies. We further acknowledge that this is a single-center, two-site study, which may further limit its generalizability. Third, we did not perform an interventional study and thus cannot be confident which findings are clinically actionable. For example, future studies could attempt to identify patients likely to benefit from early empiric antibiotics or noninvasive ventilation prior to obvious clinical deterioration. Fourth, we chose “the need for supplemental oxygen” as a clinically important outcome but recognize that need for intubation or extracorporeal membrane oxygen (ECMO) may well have different predictors. Fifth, given the dynamic nature of patients at risk, we recognize that oxygen requirements within 24 h may differ for AGM readings taken early versus late in this window. However, we would argue that such random misclassifications should bias towards the null hypothesis, potentially making our statistically significant results somewhat stronger. Subsequent studies could assess multiple AGM readings over time, such that the trajectory of deterioration versus improvement could be more carefully defined. Despite these limitations, we view our findings as important, and they could potentially yield salient future studies.

5. Conclusions

The alveolar gas monitor, a noninvasive point-of-care device, was able to predict which patients with acute dyspnea in the ED would require supplemental oxygen with increased accuracy relative to pulse oximetry and regardless of their COVID-19 status. The addition of exhaled gas partial pressure to routine pulse oximetry (with an estimation of PaO2) adds value to the prognostic information provided by pulse oximetry alone and does so with a fidelity that exceeds or matches the accuracy of the best predictive tools for respiratory failure in the published literature.

Author Contributions

W.C.M. had full access to all the data in this study and takes responsibility for the integrity of the data and the accuracy of the data analysis. E.G., T.C.L., R.S., T.M., B.N., A.K.P., A.R.E., J.M.F., G.K.P. and A.M. helped with the experimental design, data analysis, and manuscript preparation. J.B.W., G.K.P. and D.R.C. helped with content expertise regarding the alveolar gas monitor, interpretation of findings, and manuscript preparation. All authors have read and agreed to the published version of the manuscript.

Funding

This grant was funded by NIH/NHLBI, Grant Number 5F32HL167551-02 (WCM).

Institutional Review Board Statement

This study was performed in accordance with the Declaration of Helsinki. This human study was approved by the University of California, San Diego Institutional Review Board—approval: 181700, 31 October 2018.

Informed Consent Statement

All adult participants provided written informed consent to participate in this study.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Conflicts of Interest

J.B.W. helped to develop the technology of the AGM with MediPines, Yorba Linda, CA, USA but was not involved in the experimental design, data analysis, or manuscript preparation (see Contributions above).

Abbreviations

The following abbreviations are used in this manuscript:
AaDO2alveolar to arterial oxygen difference
ABGarterial blood gas
AGMalveolar gas monitor
AUCarea under the curve
AUROCarea under the receiver operating characteristic curve
CO2carbon dioxide
COVID-19Coronavirus Disease 2019
EDemergency department
etCO2end-tidal carbon dioxide
FiO2fraction of inspired oxygen
gPaO2calculated arterial partial pressure of oxygen
ODoxygen deficit
O2oxygen
PaCO2partial pressure of arterial carbon dioxide
PACO2partial pressure of alveolar carbon dioxide
PaO2partial pressure of arterial oxygen
PAO2partial pressure of alveolar oxygen
pHacidity
SDstandard deviation
SpO2oxygen saturation; pulse oximetry
ROCreceiver operating characteristic
VQventilation–perfusion

References

  1. Sjoding, M.W.; Dickson, R.P.; Iwashyna, T.J.; Gay, S.E.; Valley, T.S. Racial Bias in Pulse Oximetry Measurement. N. Engl. J. Med. 2020, 383, 2477–2478. [Google Scholar] [CrossRef] [PubMed]
  2. McGuire, W.C.; Pearce, A.K.; Elliott, A.R.; Fine, J.M.; West, J.B.; Crouch, D.R.; Prisk, G.K.; Malhotra, A. Noninvasive Assessment of Impaired Gas Exchange with the Alveolar Gas Monitor Predicts Clinical Deterioration in COVID-19 Patients. J. Clin. Med. 2023, 12, 6203. [Google Scholar] [CrossRef] [PubMed]
  3. West, J.B.; Crouch, D.R.; Fine, J.M.; Makadia, D.; Wang, D.L.; Prisk, G.K. A New, Noninvasive Method of Measuring Impaired Pulmonary Gas Exchange in Lung Disease: An Outpatient Study. Chest 2018, 154, 363–369. [Google Scholar] [CrossRef] [PubMed]
  4. West, J.B.; Liu, M.A.; Stark, P.C.; Prisk, G.K. Measuring the efficiency of pulmonary gas exchange using expired gas instead of arterial blood: Comparing the “ideal” PO2 of Riley with end-tidal PO2. Am. J. Physiol. Lung Cell Mol. Physiol. 2020, 319, L289–L293. [Google Scholar] [CrossRef] [PubMed]
  5. West, J.B.; Prisk, G.K. A new method for noninvasive measurement of pulmonary gas exchange using expired gas. Respir. Physiol. Neurobiol. 2018, 247, 112–115. [Google Scholar] [CrossRef] [PubMed]
  6. Rowling, S.C.; Fløjstrup, M.; Henriksen, D.P.; Viberg, B.; Hallenberg, C.; Lindholt, J.S.; Alberg-Fløjborg, A.; Nanayakkara, P.W.; Brabrand, M. Arterial blood gas analysis: As safe as we think? A multicentre historical cohort study. ERJ Open Res. 2022, 8, 00535-2021. [Google Scholar] [CrossRef] [PubMed]
  7. DellaVolpe, J.D.; Chakraborti, C.; Cerreta, K.; Romero, C.J.; Firestein, C.E.; Myers, L.; Nielsen, N.D. Effects of implementing a protocol for arterial blood gas use on ordering practices and diagnostic yield. Healthcare 2014, 2, 130–135. [Google Scholar] [CrossRef] [PubMed]
  8. Prisk, G.K.; West, J.B. Deriving the arterial Po(2) and oxygen deficit from expired gas and pulse oximetry. J. Appl. Physiol. 2019, 127, 1067–1074. [Google Scholar] [CrossRef] [PubMed]
  9. Severinghaus, J.W. Pathophysiologic aspects of the regulation of respiration. Bull. Mem. Acad. R. Med. Belg. 1979, 134, 261–271. [Google Scholar] [PubMed]
  10. Severinghaus, J.W.; Crawford, R.D. Regulation of respiration. Acta Anaesth. Scand. Suppl. 1978, 70, 188–191. [Google Scholar] [CrossRef]
  11. Severinghaus, J.W. Blood gas calculator. J. Appl. Physiol. 1966, 21, 1108–1116. [Google Scholar] [CrossRef] [PubMed]
  12. Kelman, G.R. Digital computer subroutine for the conversion of oxygen tension into saturation. J. Appl. Physiol. 1966, 21, 1375–1376. [Google Scholar] [CrossRef] [PubMed]
  13. Kelman, G.R. Digital computer procedure for the conversion of PCO2 into blood CO2 content. Respir. Physiol. 1967, 3, 111–115. [Google Scholar] [CrossRef] [PubMed]
  14. Kelman, G.R.; Nunn, J.F. Nomograms for correction of blood PO2, PCO2, pH, and base excess for time and temperature. J. Appl. Physiol. 1966, 21, 1484–1490. [Google Scholar] [CrossRef] [PubMed]
  15. MediPines. MediPines Gas Exchange Monitor AGM100 User Manual; MediPines: Yorba Linda, CA, USA, 2020. [Google Scholar]
  16. Box, G.E.P.; Tidwell, P.W. Transformation of the Independent Variables. Technometrics 1962, 4, 531–550. [Google Scholar] [CrossRef]
  17. Hosmer, D.W., Jr.; Lemeshow, S.; Sturdivant, R.X. Applied Logistic Regression; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2013. [Google Scholar]
  18. Trillo-Alvarez, C.; Cartin-Ceba, R.; Kor, D.; Kojicic, M.; Kashyap, R.; Thakur, S.; Thakur, L.; Herasevich, V.; Malinchoc, M.; Gajic, O. Acute lung injury prediction score: Derivation and validation in a population-based sample. Eur. Respir. J. 2011, 37, 604–609. [Google Scholar] [CrossRef] [PubMed]
  19. Gajic, O.; Dabbagh, O.; Park, P.K.; Adesanya, A.; Chang, S.Y.; Hou, P.; Anderson, H.; Hoth, J.J.; Mikkelsen, M.E.; Gentile, N.T.; et al. Early identification of patients at risk of acute lung injury: Evaluation of lung injury prediction score in a multicenter cohort study. Am. J. Respir. Crit. Care Med. 2011, 183, 462–470. [Google Scholar] [CrossRef] [PubMed]
  20. Roca, O.; Messika, J.; Caralt, B.; García-De-Acilu, M.; Sztrymf, B.; Ricard, J.-D.; Masclans, J.R. Predicting success of high-flow nasal cannula in pneumonia patients with hypoxemic respiratory failure: The utility of the ROX index. J. Crit. Care. 2016, 35, 200–205. [Google Scholar] [CrossRef] [PubMed]
  21. Roca, O.; Caralt, B.; Messika, J.; Samper, M.; Sztrymf, B.; Hernández, G.; García-De-Acilu, M.; Frat, J.-P.; Masclans, J.R.; Ricard, J.-D. An Index Combining Respiratory Rate and Oxygenation to Predict Outcome of Nasal High-Flow Therapy. Am. J. Respir. Crit. Care Med. 2019, 199, 1368–1376. [Google Scholar] [CrossRef] [PubMed]
  22. Shashikumar, S.P.; Wardi, G.; Paul, P.; Carlile, M.; Brenner, L.N.; Hibbert, K.A.; North, C.M.; Mukerji, S.S.; Robbins, G.K.; Shao, Y.-P.; et al. Development and Prospective Validation of a Deep Learning Algorithm for Predicting Need for Mechanical Ventilation. Chest 2021, 159, 2264–2273. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Example of breath-by-breath analysis as displayed on the AGM. (A) demonstrates a cough in the second breath (arrow) and variable expiratory times (red bars), resulting in a steady state not being achieved (red circle). (B) demonstrates similar breath-by-breath characteristics, resulting in the subject achieving a steady state (second red circle). The asterisk after R (RQ) stands for dimensionless number. It is a feature of the AGM100 that is part of the company (MediPines) programming of the device, not an asterisk to notify readers of additional information.
Figure 1. Example of breath-by-breath analysis as displayed on the AGM. (A) demonstrates a cough in the second breath (arrow) and variable expiratory times (red bars), resulting in a steady state not being achieved (red circle). (B) demonstrates similar breath-by-breath characteristics, resulting in the subject achieving a steady state (second red circle). The asterisk after R (RQ) stands for dimensionless number. It is a feature of the AGM100 that is part of the company (MediPines) programming of the device, not an asterisk to notify readers of additional information.
Jcm 14 05880 g001
Figure 2. Receiver operating characteristic curves demonstrating the superior discriminative ability of an oxygen deficit to predict the need for supplemental oxygen as compared to pulse oximetry among COVID-19-positive participants (A), COVID-19-negative participants, and (B) all participants (C).
Figure 2. Receiver operating characteristic curves demonstrating the superior discriminative ability of an oxygen deficit to predict the need for supplemental oxygen as compared to pulse oximetry among COVID-19-positive participants (A), COVID-19-negative participants, and (B) all participants (C).
Jcm 14 05880 g002
Table 1. Descriptive statistics of the COVID-19-negative and -positive cohorts.
Table 1. Descriptive statistics of the COVID-19-negative and -positive cohorts.
Summary of COVID-19-Negative CohortSummary of COVID-19-Positive Cohort
VariablenMeanStd. Dev.MinPctl. 25Pctl. 75MaxVariablenMeanStd. Dev.MinPctl. 25Pctl. 75Max
Participants42 Participants30
Age (yrs) 601619517284Age (yrs) 611434527285
BMI (kg/m2) 318.218253651BMI (kg/m2) 296.720243249
Resp Rate 186.18132230Resp Rate 23177132799
Heart Rate 8820597297145Heart Rate 8111607389100
Gender Gender
Female1331% Female1240%
Male2969% Male1860%
Race Race
Asian12% Asian310%
Black614% Black413%
Hispanic512% Hispanic723%
Other37% Other0
Native American25% Native American0
White3071% White2377%
Table 2. Univariable logistic regressions and Bonferroni corrections for all three cohorts, comparing OD, ODFlip, and SpO2 to the need for supplemental oxygen.
Table 2. Univariable logistic regressions and Bonferroni corrections for all three cohorts, comparing OD, ODFlip, and SpO2 to the need for supplemental oxygen.
COVID-19-Positive Cohort
Est.Std. Errort-Valuep-ValueIndependent Variable (Predictor)ModelBonferroni Correction
0.250.083.220.00O2DeficitSuppO2~O2Deficit0.05
−0.250.08−3.220.00ODFlipSuppO2~ODFlip0.05
−0.740.26−2.890.01SpO2SuppO2~SpO20.11
COVID-19-Negative Cohort
Est.Std. Errort-Valuep-ValueIndependent Variable (Predictor)ModelBonferroni Correction
0.140.043.450.00O2DeficitSuppO2~O2Deficit0.02
−0.140.04−3.450.00ODFlipSuppO2~ODFlip0.02
−0.580.19−3.030.00SpO2SuppO2~SpO20.07
Pooled Cohorts
Est.Std. Errort-Valuep-ValueIndependent Variable (Predictor)ModelBonferroni Correction
0.170.034.930.00O2DeficitSuppO2~O2Deficit0.00
−0.170.03−4.930.00ODFlipSuppO2~ODFlip0.00
−0.670.15−4.540.00SpO2SuppO2~SpO20.00
Table 3. Logistic regression results for COVID-19-negative participants demonstrating a statistical significance of an oxygen deficit in predicting the need for supplemental oxygen but in the absence of statistical significance for pulse oximetry.
Table 3. Logistic regression results for COVID-19-negative participants demonstrating a statistical significance of an oxygen deficit in predicting the need for supplemental oxygen but in the absence of statistical significance for pulse oximetry.
BS.E.WalddfSig.Exp(B)95% CI
LowerUpper
SpO2−0.3720.2053.30710.0690.6890.4621.029
ODFlip−0.1320.0536.27610.0120.8760.7910.972
Constant42.41519.5074.72810.0302.63 × 1018
Table 4. Logistic regression results for all participants demonstrating a statistical significance of an oxygen deficit in predicting the need for supplemental oxygen but in the absence of statistical significance for pulse oximetry.
Table 4. Logistic regression results for all participants demonstrating a statistical significance of an oxygen deficit in predicting the need for supplemental oxygen but in the absence of statistical significance for pulse oximetry.
BS.E.WalddfSig.Exp(B)95% CI
LowerUpper
SpO2−0.3150.1643.71410.0540.7290.5291.005
ODFlip−0.1530.04511.2991<0.0010.8580.7850.938
Constant38.43615.3016.31010.0124.93 × 1016
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

McGuire, W.C.; Gruenberg, E.; Long, T.C.; Sheth, R.; Marin, T.; Nokes, B.; Pearce, A.K.; Elliott, A.R.; Fine, J.M.; West, J.B.; et al. The Alveolar Gas Monitor: An Alternative to Pulse Oximetry for the Noninvasive Assessment of Impaired Gas Exchange in Patients at Risk of Respiratory Deterioration. J. Clin. Med. 2025, 14, 5880. https://doi.org/10.3390/jcm14165880

AMA Style

McGuire WC, Gruenberg E, Long TC, Sheth R, Marin T, Nokes B, Pearce AK, Elliott AR, Fine JM, West JB, et al. The Alveolar Gas Monitor: An Alternative to Pulse Oximetry for the Noninvasive Assessment of Impaired Gas Exchange in Patients at Risk of Respiratory Deterioration. Journal of Clinical Medicine. 2025; 14(16):5880. https://doi.org/10.3390/jcm14165880

Chicago/Turabian Style

McGuire, W. Cameron, Eli Gruenberg, Tanner C. Long, Richa Sheth, Traci Marin, Brandon Nokes, Alex K. Pearce, Ann R. Elliott, Janelle M. Fine, John B. West, and et al. 2025. "The Alveolar Gas Monitor: An Alternative to Pulse Oximetry for the Noninvasive Assessment of Impaired Gas Exchange in Patients at Risk of Respiratory Deterioration" Journal of Clinical Medicine 14, no. 16: 5880. https://doi.org/10.3390/jcm14165880

APA Style

McGuire, W. C., Gruenberg, E., Long, T. C., Sheth, R., Marin, T., Nokes, B., Pearce, A. K., Elliott, A. R., Fine, J. M., West, J. B., Crouch, D. R., Prisk, G. K., & Malhotra, A. (2025). The Alveolar Gas Monitor: An Alternative to Pulse Oximetry for the Noninvasive Assessment of Impaired Gas Exchange in Patients at Risk of Respiratory Deterioration. Journal of Clinical Medicine, 14(16), 5880. https://doi.org/10.3390/jcm14165880

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop