3.1. Structural Characterization of SP/e-spunCL Membranes
In order to confirm that the applied high electrical field does not cause any change in the chemical structure of SP/e-spunCL membranes, the FTIR spectra of a cast SPEEK/Cloisite membrane and SP/e-spunCL membranes are compared (
Figure 3). The two spectra are identical showing the presence of characteristic peak of 1600 cm
−1–600 cm
−1 [
11]. The details of transmission bands are shown in
Table 2.
Since it is crucially vital to examine the morphological types (intercalated and exfoliated) alongside the effects of the two highlights on the membrane properties, surface structure of the membranes were investigated with three basic strategies which are Scanning electron microscopy (SEM), X-Ray diffraction (XRD) and Field Emission Scanning Electron Microscope (FESEM).
XRD investigation was used in the research to look at the microstructure and portray the clay scattering with SP/e-spunCL layers. Note that the presence of organic materials (SPEEK) lead the featureless peak happen at the typical pinnacle that unadulterated closite 15A was regularly seen at 2
θ = 2.6° [
11]. Thus, the examples of SP/e-spunCL layers were seen at 2
θ = 5°. The XRD patterns of the corresponding clay in
Figure 4 were only observed in a small angle part in the range of 2° to 10° of 2
θ scale because the dispersion state of the clay particles can simply be obtained by analyzing the (001) lattice spacing of the clay.
Referring to the Bragg’s law equation, the expanding of d-spacing causes the widening and shifting of the XRD patterns. The peaks likewise characterized the morphology as intercalated. From the two peaks, it very well may be seen that SP/e-spunCL10 demonstrated lower angle when contrasted with SP/e-spunCL30. This can be clarified that the layer spacing was expanded because of exfoliation or intercalation effects [
12,
13,
14]. The other membranes which are SP/e-spunCL15, SP/e-spunCL20 and SP/e-spunCL25 demonstrated no peak showed up [
9]. The result demonstrated that cloisite were effectively dispersed in polymer matrix because of the electro spinning procedure and the hydrophilicity of cloisite [
15]. The missing peak of the three membrane samples can also be induced that the completely exfoliated structure obtained from the introduction of electro spun fibers of SPEEK/cloisite with 15 wt.%–25 wt.% of cloisite loadings into SPEEK polymer matrix.
Excluding SP/e-spunCL10 and SP/e-spunCL30 membranes, the intensity of the peaks that corresponds to the plane (001) increased as the loading of cloisite was below 0.15 wt.% and above 0.25 wt.%. The high intensity of the corresponding peak for the SP/e-spunCL10 and SP/e-spunCL30 membranes occurred probably because of lack of cloisite dispersion on the surface of the membranes as confirmed by the EDX images. SP/e-spunCL10 and SP/e-spunCL30 membranes showed the diffraction peak at 2
θ = 5.8° (
d001 = 1.52 nm) and 6.10° (
d001 = 1.45 nm), respectively. Since that both particular composite membranes showed higher
d001 than the pure cloisite membrane (2
θ = 7.1° (
d001 = 1.25 nm) [
15,
16], it means that the intercalated nanocomposite is obtained.
Even though XRD is a good conventional method that offers a technique which quantitatively analyzes morphology of the membranes, subsequently, another qualitative investigation to finish the morphological examination was conducted to support the existing results. The methods were checking electron microscopy (SEM) and Field Emission Scanning Electron Microscope (FESEM).
The methods not only provided energy dispersive X-ray (EDX) in determining the presence of important element which was cloisite in SP/e-spunCL membrane but also the exfoliation images of surface membranes. Although the methods only provided the information on a small area which was not representative of the overall microstructure, regardless of all, SEM, FESEM, and XRD are very crucial analyses in evaluating morphology of nanocomposites [
17].
Energy dispersive X-ray (EDX) was applied to observe the elements that presented in membranes together with the dispersion percentage of the desired element which was cloisite. The highest percentage of cloisite dispersion was found in SP/e-spunCL15 (13.89%) followed by SP/e-spunCL25 (13.36%), SP/e-spunCL20 (11.36%), SP/e-spunCL10 (11.29%), and SP/e-spunCL30 (11.23%) respectively. The effect of dispersion can clearly observed quantitatively and qualitatively by SEM and FESEM images (
Table 3 and
Figure 5).
To further the investigation on the exfoliation of SP/e-spunCL membranes, FESEM analysis with few thousand magnifications was carried out. The FESEM and SEM surface and cross sectional images of all SP/e-spunCL membranes were presented in
Figure 6(a–e2).
SP/e-spunCL15, SP/e-spunCL20 and SP/e-spunCL25 showed the fully exfoliation structures occurred on the both surface and cross section which were agreed with XRD result. All samples showed crumpled and fibrous patterns on the surfaces which resulted from the addition of electrospun fibers on the SPEEK matrix. As for the cross section images, all three samples had proven the electrospun fibers did not only exist on the surface but also inside the membranes. However the fibrous structures that formed did consistently be found in the whole section of cross sectional areas or it can be found more at the surface rather than the bottom part. The structure can be related as sewing pattern like structures.
SP/-spunCL10 which more exfoliated compared to SP/e-spunCL 30 as defined by XRD also agreed with result in FESEM images. SP/e-spunCL10 showed the existence of cylindrical bacterial like structures on the surface with huge agglomerated sectional areas which SP/e-spunCL30 showed lowest existence of neither fibrous structure nor cylindrical structure. Only few long threads like structures can be observed on the SP/e-spunCL30 surface. From the observation, the different of morphological structures can be explained that the concentration of dope solution with specific electro spinning setting played important role in governing the patterns and performance of membranes.
Although many researchers observed different dispersion of cloisite was proven to give the different impacts on membrane morphologies [
18,
19,
20,
21,
22,
23,
24], the study on size of cloisite together with volume of water that can be retained by the particle is scarce. The observation of closite size reduction as a result of electrospinning was done by using AFM.
3.2. The Effects of Morphological Structures and Closite Dispersion on Barrier Property of Void-Free SP/e-spunCL
The dramatic improvements in barrier properties were observed as the impacts of addition of electrospun fibers into SPEEK polymer matrix. Clay sheets are regularly impermeable and the existence of clay helps to extend the barrier properties of polymers by making an all the more winding way that retards the dissemination of gas molecules through the polymer matrix (
Figure 6) [
25,
26].
Figure 7 showed the distinction of methanol path between SPEEK/cloisite membrane and SP/e-spunCL membrane.
The improvements of barrier properties rely upon the level of tortuosity made by cloisite layers in the molecules path way in polymeric membranes [
27,
28]. The tortuous path is controlled by the proportion of real separation which diffusive molecules (methanol and water) are transferred to the most limited separation to diffuse (thickness of membranes). Previous study reported that the barrier properties of polymer/clay nanocomposites against the dissemination of gases and vapors [
29,
30,
31,
32]. This was because of the dispersion of clay in the polymeric matrix. Other than this, the exfoliation factor and degree of dispersion cause to the more barrier improvement in the polymer matrix.
Aside from creating more winding path and exfoliated structure to a membrane, another addition essential study on the effect of electrospun fibers on membrane morphology is the reduction of the cloisite size and volume of water that can be retained by cloisite. The retained water can be greatly effect proton conductivity as well as performance of membranes as PEMs in DMFC application.
Table 4 presents the summary of information obtained from 10 µm
2 of AFM images.
As compared to the original size of commercialized cloisite 15A (2–13 µm) [
33], after electro spinning applied to the SPEEK/cloisite solution, the sizes of the particle had been reduced to 0.286–0.390 µm. This had proven the electro spinning is the effective method for inorganic particle size reduction. The number of cloisite as well as volume and water retained of particle agreed with the dispersion trend in
Table 4 which showed the highest for SP/e-spunCL15 (1668 with volume of 2.914 × 10
−3 µm
3, water retained of 4.854 µm
3 per 10 µm
2) and the lowest for SP/e-spunCL 30 (1097 with volume of 4.505 × 10
−4 µm
3, water retained of 0.494 µm
3 per 10 µm
2). The existence of cloisite as fillers had improve the proton conductivity of membrane by the ability of retaining water which is a proton transport medium [
33].
3.4. Long Term Stability of SP/e-spunCL Membranes in Hydrated State
Proton conductivity, mechanical properties, and barrier properties are affected by amount of absorbed water [
37]. Thus, study of long-term stability of membrane in hydrated state is crucial for DMFC application. A good PEM requires a dimensionally stable membrane without high percentage of swelling. Swelling caused by excessive water absorbed by hydrophilic group in sulfonic acid group presence in membrane. The excessive absorption contributes on morphological instability, mechanical fragility, and dimensional instability [
38].
Besides swelling, dimensional change affects performance of PEMs. The high dimensional change in plane is not required for good PEMs due to a weak contact between catalyst and electrolyte membrane [
39].
Figure 8 showed the water uptake and dissolution time for SP/e-spunCL membranes, while
Figure 9a,b demonstrated the swelling ratio of SP/e-spunCL membranes in the directions of plane and thickness.
From
Figure 8, the water uptake of SP/e-spunCL15 was the highest throughout the experimental period among the five tested membranes. It can be suggested that the less-uniform distribution of Cloisite15A
® particles in SP/e-spunCL10, SP/e-spunCL20, SP/e-spunCL25, and SP/e-spunCL30 (see
Figure 5) allowed the sulfonic acid groups in the SPEEK polymer matrix to absorb water. This is because when a large number of O atoms in sulfonic acid groups are left unattached to Cloisite15A
® via hydrogen bonding, a large amount of water will be taken into the membrane via bonding between the O atoms in sulfonic groups and hydrogen atoms in water [
40].
The swelling data reported in
Figure 9b showed similar trend to water uptake. The water uptake and dimensional change in thickness of SP/e-spunCL15 was observed the highest among other membranes. However, SP/e-spunCL15 showed the smallest value for dimensional change in plane which is 25% (
Figure 9a). This study indicates that SP/e-spunCL10, SP/e-spunCL20, SP/e-spunCL25, and SP/e-spunCL30 membrane became more fragile with time and mechanically less stable in water than the SP/e-spunCL15 membrane However, all membranes had shown increment dimensionally as well as plane throughout the test period. This indicated that, the free volume for water adsorption as well as the mobility of polymer chains increased with time, which contributed to the increment of water absorbability of membranes. All membranes presented a much higher swelling ratio in thickness than in plane. Similar observation was reported by Juhana 2011 [
41].
A PEM with water uptake less than 50% is considered stable and low degree of swelling [
42]. Hence, it can be concluded that the swelling effect on SP/e-spunCL15 and SP/e-spunCL25 can be considered as low degree of swelling. Thus, SP/e-spunCL15 and SP/e-spunCL25 membrane is stable enough to be applied in the real DMFC application in hydrated state.
3.5. DMFC Performance
The impact of methanol permeability and proton conductivity were verified by polarization measurement [
43].
Figure 10 presented the cell voltages as a function of current density for the single cell of DMFC prepared from SP/e-spunCL membranes along with the performance of and Nafion
®115 membranes for comparison. It was observed that SP/e-spunCL15 exhibited highest performance as compared to other SP/e-spunCL membranes and Nafion
®115 membranes. Hence, it can be concluded that the performance of the DMFC was improved by exfoliated structure of membranes.
The maximum current densities of SP/e-spunCL15, SP/e-spunCL25, SP/e-spunCL20, SP/e-spunCL10 and SP/e-spunCL30 membranes measured are 1.0422 Acm
−2, 1.0132 Acm
−2, 0.7773 Acm
−2, 0.7216 Acm
−2 and 0.3472 Acm
−2, respectively, and 0.9615 Acm
−2 for Nafion
®115 as a comparison. The lower current densities of the SP/e-spunCL10 and SP/e-spunCL30 membranes were due to the lower proton conductivity of those membranes than that of other SP/e-spunCL nanocomposite membranes. Lower proton conductivity attributes in obvious ohmic losses causes by high membrane resistance and shows mass transfer limitations [
44].
The open circuit voltage (OCV) for SP/e-spunCL15, SP/e-spunCL25, SP/e-spunCL20, SP/e-spunCL10, and SP/e-spunCL30 membranes measured are 0.412 V, 0.223 V, 0.323 V, 0.330 V, and 0.289 V, respectively, and 0.403 V for Nafion
®115. OCV defines methanol permeation and it increases when the methanol crossover is decreased. Methanol crossover avoids oxygen reduction at the anode and consequently causes to a drastic decrease in the OCV [
45]. SP/e-spunCL15 showed the highest methanol barrier property (as presented in
Table 4), it reduced the possibility for the methanol to transport towards the opposite electrode thus resulted in higher OCV.
The OCV result especially for Nafion
® 115 membrane found a low value which can be related not only caused by methanol crossover. As compared to other comparable materials and fabrications in
Table 5, it can be deduced that MEA set up also plays important role in determining a high OCV value. Hence, the inappropriate MEA set up was a reason of the low OCV value of Nafion
® 115 membrane. The incompatibility between the membrane and electrode caused methanol leakage directly from anode to cathode without passing through the electrolyte membrane first [
46].
Table 6 presented the OCV value together with current and power density of newly developed SP/e-spunCL 15 compared to other researches. It showed that the newly developed membrane has a potential as a high performance PEM in the future due to the highest current density presented.
Figure 10 presented the polarization curve of power densities as a function of current density for SP/e-spunCL membranes and Nafion
®115 membranes. The membrane morphology which is exfoliated structure effected power density. The highest power density 0.00118 Wcm
−2 found in SP/e-spunCL15 which was observed morphologically exfoliated among all membranes. SP/e-spunCL25 (0.000710 Wcm
−2) and (0.000772 Wcm
−2) SP/e-spunCL20 smaller value of power density were observed in less exfoliated samples as compared to SP/e-spun17. Sp/e-spunCL10 and SP/e-spunCL30 with half-exfoliated structure, which resulted 0.000744 Wcm
−2 and 0.000315 Wcm
−2 respectively. Nevertheless, all SP/e-spunCL membranes observed higher power density than that of Nafion
® 115 with 0.000550 Wcm
−2. The result has shown the potential of SP/e-spunCL membranes, notably SP/e-spunCL15, as alternatives for Nafion
® 115 due to better performance. However, the value obtained was rather lower than the standard test. The improvement in future should be studied not only on membranes but also optimization of the system such as methanol concentration, cooling system and MEA set up.
Beside the exfoliated structure desired for high power density PEMs, a good distribution also contributed to a high performance PEMs for DMFC as well. A good distribution causes a low ohmic resistance. Hence, in more oriented electrolyte, the proton from the anode side could transport more efficient to the cathode side and this would accelerate the kinetic of the reactions in both electrodes. This explains the essential of oriented nanostructures that contribute ion diffusion in the proper direction for the electrochemical processes occurring at each electrode.