4.2. Why the Density Axis Resists
Neither an oracle classifier on crops cut from the reference boxes nor four architectural interventions raised the best result available. The physical signature itself is barely there: over the 24 descriptor comparisons of
Section 3.6, the area under the ROC curve departs from chance by at most 0.071 on the merged density column. The five that do separate include the strongest, and that one runs opposite to its intended direction. At this resolution, the operator responds to radiolucency, but not selectively to the radiolucency a clinician would call a void.
The arch split of
Section 3.1 does not settle this and should not be read as though it did. Missed density defects concentrate in the mandible while false density calls concentrate in the maxilla, which is in part what one threshold applied to two base rates produces. At 1280 pixels, the Youden index, which the base rates do not move, separates the two classes less well in the mandible by 0.194, although the same contrast is not resolved at 1024 pixels and the two resolutions do not differ from each other. If that residue is real, the harder jaw on this axis is the one with fewer density defects rather than the one the projection argument nominates. That is the opposite of what superimposition predicts, but it does not refute it, for a reason no subgroup split can get around: the reference standard is itself a reading of the same projection. If superimposition misleads the annotator, it misleads the label with it, and the contrast then measures the model against that error rather than against the anatomy. The superimposition account is neither supported nor excluded here, and a design that could settle it would need a reference standard from a different modality.
The source/label split is the more informative one. The detector is worst on the label defined in
Table 1 by interruptions within the body of the material, but only in one jaw. In the maxilla, it misses 47.6% of those boxes against 16.3% of the deficient condensation label. In the mandible, the two are 44.4% and 45.1%, so the contrast is a maxillary one and not a property of the labels as such. Both jaws are given because reporting only the maxillary pair would make a 31-point gap look like a property of the label pair when it reverses to within a point on the other jaw, on 18 and 102 boxes, respectively. The consequence for the attribution analysis is definitional rather than inferential.
Section 2.8 builds its reference on a radiolucent discontinuity inside radiopaque material, whereas
Table 1 places 410 of the 487 density defects under a label defined chiefly at the filling/dentine interface or by a diffuse reduction of radiopacity, with interior voids admitted only below 1 mm. For most of those, the reference is therefore named for an appearance the label need not carry. What it responds to instead is measurable, and it is not what the phrase first suggests. Running the operator on synthetic lucencies of known diameter and shape, described in
Section 2.8, separates the candidate explanations for that split and, conditional on the split being real, rules two of them out.
Four accounts of that split can be put to the phantom. Position is the first. A lucency straddling the filling/dentine boundary returns a weaker response than the same lucency inside the material, which is the opposite of what a positional account needs. Composition is the second. Taking the 410 as interface lesions alone predicts that the interior-void label should separate the better of the two, and it separates worse. Size is the third, and it divides. At the level of a single lesion, it cannot be tested here at all, because
Table 1 records a lucency by its longest dimension while the operator responds to its shortest, and two lesions of the same recorded size can return responses a factor of five apart. At the level of the population, it is consistent with the observation. An interior lucency below 1 mm cannot be labelled non-homogeneous and falls to deficient condensation, and on the phantom such lesions return the strongest responses measured. The account nevertheless rests on a margin of 2.20 points between those responses and the highest a lesion of the 77 can return, smaller than the 2.45 points the operator loses to lesion length alone. It also rests on a mixing fraction the annotation does not record, and on an assumption about what a reader can see on a panoramic film. Every one of these readings is drawn at a 2.24 mm strip, and each inherits a dependence on canal width that this cohort cannot resolve.
A fourth account needs no mechanism at all. If the 77 interior-void boxes carried exactly the effect the 410 shows, a subgroup of that size would fail to reach significance here about nine times in ten, so a pure difference in power could produce the pattern of significance. The intervals weigh against that account without excluding it. For the strongest descriptor, the 77 interior-void boxes give 0.429 to 0.564, an interval that excludes the 0.573 the 410 reach, but by less than one hundredth, without correction and without a test of the difference between the labels. A power account is therefore disfavoured rather than excluded, and that changes the standing of the other three rather than joining them: each of those is an account of a difference between the labels, so each presupposes that the difference is real. The position, then, is this. If the difference between the two labels is real, a positional account is excluded by a measurement that points the wrong way, and an account that treats the 410 as interface lesions alone is excluded by composition. A size account cannot be tested lesion by lesion, and at the level of the population it is consistent with the observation but rests on a margin of 2.20 points, an unmeasured mixing fraction, and an assumption about what a reader can see. If the difference is not real, power accounts for the whole of it and none of the above is needed, although the interval just given makes that the less likely case. This work does not settle which of those two cases obtains, so the split reported in
Section 3.6 is not explained here. What the phantom does is narrow the field: it says what a cohort would have to record to settle the question, namely the shape of each void and not only its longest dimension, the width of the canal it sits in, and which deficient condensation boxes carry a sub-millimetre interior void.
What the phantom does establish is narrower and worth stating on its own. The reference is not a void detector in any simple sense: its response is governed by the transverse width of a lucency, it saturates once the structuring element reaches about twice that width, and it is very nearly indifferent to the length that
Table 1 actually records. The same 1.04 mm round lucency returns 10.1 at a 9-pixel element, 43.8 at 17, and 47.1 at 31, so even the round-lucency scale is a property of the implementation rather than of root fillings.
The ceiling is not peculiar to this cohort. Against a two-observer cone-beam consensus, the same observers’ consensus reading of the panoramic images scores an F1 of 55.65% for density [
18]. That is concordance between two modalities rather than agreement within this one, but it bounds how far panoramic reading can track the state the reference standard is meant to capture, and it does so before any model is involved. A reference standard whose size criterion and whose operator respond to different axes cannot be expected to agree with that operator. That disagreement is a property of the reference standard, and it adds to, rather than replaces, whatever limit the projection imposes. Two caveats attach to all of it. The phantoms carry no root contour, so they cannot test the explanation
Section 3.6 offers for the inverted direction of the lucency descriptor, which is attributed there to exactly that contour, and every millimetre here rests on the conversion of
Section 2.1, which uses a nominal manufacturer magnification rather than a per-image calibration.
The source/label split reported in
Section 3.6 runs the opposite way to the reading the merged column invites: the separation sits on the deficient condensation label, and none is detected on the label the operator is named for. That strengthens the negative result rather than qualifying it. The reference is not a validated marker of voids, and it is on the boxes a void defines that no separation is detected at all. It does not establish two things. It does not establish a zero, because 77 boxes against 810 give a half-width near 0.068. An effect of up to 6.8 points of area under the curve in either direction remains open for that label, which is less than the 7.3 the deficient condensation label itself reaches, and it does not establish that the two labels differ from each other, which would need a test of the difference and is not run here. What is shown is a separation established on one label and not detected on the other.
Agreement on the density criterion is the weakest of the radiographic criteria in the studies available. Undergraduate readers scored against a gold standard set by four endodontists, who themselves agreed at a Fleiss kappa of 0.81, reached kappa values of 0.41 to 0.48 on the length axis and only 0.15 to 0.17 on the density axis, on periapical radiographs [
8]. An older assessment of extracted teeth on periapical film, pooling length, density, and taper across uncalibrated examiners, returned a kappa of 0.36, which its authors describe as poor agreement [
7]. On panoramic radiographs, formal calibration raised observer agreement for periapical pathology from a range of 0.22 to 0.60 to one of 0.59 to 0.80. In the same study, the objective registrations, among them whether a tooth is root filled at all, were highly reliable in every assessment, including the one before calibration, at intraclass correlations of 0.88 to 0.98 and kappa values of 0.98 to 0.99 [
9]. That contrast is the present study’s own division in human form: what a reader must find is reliable, and what a reader must grade is not. None of the three measured inter-observer agreement for filling density among specialists on panoramic radiographs, which is the quantity at issue here, so the density axis is described here by what these studies do show rather than by an agreement statistic on this cohort. A study dedicated to the reliability of intraoral radiographs for endodontic technical quality found the agreement of its two observers to be good. Recordings of the seal nonetheless differed significantly between clinical radiographs and in vitro images of the same extracted roots taken in the same projection, with overlying tissue masking seal defects [
37], so the imaging conditions contribute to the variation as well as the reader. Calibration is therefore a lever on this modality, even though its effect on filling density has not been measured.
The model’s own errors fall on the same axis. The adequate-against-density exchange carries 244 errors in cross-validation against 79 for adequate-against-length, a ratio of 3.1-to-1, and that ratio reappears at 3.3-to-1 on the locked test. The two observations are consistent with one another, but the second does not establish the first, because a model trained on labels of a given reliability reproduces their structure whatever the cause of that structure is.
Section 4.4 records label noise as an alternative this study cannot exclude.
4.3. What the Attribution Analysis Shows and Does Not Show
A null result is interpretable only if the instrument can detect something. The global maximum of the D-RISE map falls inside the target box in 97.8% and 95.0% of detections against 0.0% and 0.6% for a Sobel gradient map and a seeded random map, at p below 0.0001.
The remaining results are bounded non-detections rather than failures to measure. Against a smooth random map, none of the three region statistics separate at the mask budget used throughout, two of the three being one quantity in two readings, and the excluded effect is bounded at 0.052, 0.008, and 0.026 in the convolutional detector. Raising that budget separates two of the three: relative to a random map, the attribution is shifted apically and away from the coronal restoration. The suite therefore does resolve regional structure at this scale when given enough masks. What it does not resolve at either budget is any preference for the void reference, the bound on which is 0.058 at the higher budget. Evidence below the scale of the occlusion regions would not register, and the criterion sits at or below that scale, so no conclusion is drawn here about a localised substrate at the scale of the criterion.
The void reference is not validated as a marker of voids, and this is the limit that matters most for what the attribution analysis can be asked to show. The descriptor search in
Section 3.6 places its largest departure from chance on the merged density column at 0.0704 and reverses the sign of its strongest descriptor. A model given that reference as an input channel, moreover, does not attend to it more than one that was not. The lesson is transferable and is the one this study would put forward in place of the claim it set out to make. An annotation-free reference derived from the physical definition of a criterion is attractive because it needs no expert opinion, but it must itself be shown to mark the evidence before a null against it carries meaning. The test for that is cheap, since a model can simply be given the reference as an input and its attribution measured.
One limit of the attribution analysis follows from the label definitions rather than from the attribution method. The void reference is defined on a discontinuity inside the material, and
Table 1 separates the two density source labels by location. For the 410 deficient condensation boxes, which
Table 1 defines chiefly at the filling/dentine interface or by a diffuse loss of radiopacity, the reference is therefore measuring something the label need not turn on.
Whether the scale of the structuring element governs these descriptors was checked rather than assumed, by rerunning the response at 17, 31, and 41 pixels. Each statistic keeps its side of chance across an element that more than doubles, so the element scale does not govern the statistics it recomputed, and it leaves the inversion of the lucency descriptor untested.
Two elements of the protocol transfer. Localisation and discrimination belong apart wherever an object is easy to find, and its grade is hard to assign. Occlusion-based attribution on radiographs also needs a high retention rate, since the retention of 0.5 used for natural images [
33,
34] leaves only 16.4% of masks informative on this material. At a high retention, moreover, the classical weighted sum must be read against its near-constant offset, as the contrast of Equation (5) does, which is a correction for the regime rather than a different estimand.
4.4. Comparison with Previous Studies, and Limitations
Table 10 places this study beside the closest published work. Çelik et al. segmented the filling on periapical radiographs and assigned a length category from the segmented tip to a hand-marked apex [
20]. Ayhan et al. trained the detector used here, YOLO11m, on periapical radiographs of extracted teeth, merging both directions of the length axis [
17]. Jin et al. developed YOLOv5 on periapical radiographs from one centre and tested on 500 from a second, folding length and density into one binary decision [
16]. None of them read one annotation at the two levels of the clinical rubric, acceptable or not first and only then deficient along an axis, nor did the two commercial-platform studies score panoramic features as independent binary decisions [
18,
19].
The two platform studies bear on
Section 4.2 and were obtained independently. On 55 panoramic radiographs scored against two observers, the platform reached F1 values of 14.29% for voids and 8.33% for short fillings, with accuracies of 55.81% for obturation adequacy and 62.79% for density adequacy [
19]. The quoted metric differs between the features; the F1 on those last two stands at 62.75% and 75.00%. Those accuracies rest on specificities of 30.77% and 16.67%, so on both of them the platform called almost every filling adequate. Its successor, scored against a two-observer cone beam computed tomography consensus, reported adequate density at an F1 of 42.67% with a 95% interval of 27.69% to 56.31% [
18]. The observers’ consensus reading of the same panoramic radiographs reached 55.65%, and the platform exceeded 94% accuracy on all five features on cone beam computed tomography.
Choi et al. on 1638 panoramic radiographs [
11] and Kazimierczak et al. on 4148 teeth [
38] located root-filled teeth at a similar level. Khurshid et al. on 2235 radiographs from three countries reported a best mAP50 of 82.0% from Faster R-CNN, as a mean over six restoration classes, of which the root canal-treated tooth is one [
12]. Finding a root-filled tooth is therefore reliable across cohorts and countries. Deciding what is wrong with the filling is not.
Three properties of the last row are absent from most of the others. Localisation and discrimination were measured separately, where every other quality-grading study reported one figure carrying both error sources and neither platform study reports grading conditioned on matched teeth [
18,
19]. Measuring them apart is not itself new outside dentistry, where a consortium framework guides the problem-aware selection of validation metrics [
39] and diagnostic tools decompose detection error into localisation error, confusion with similar objects, and confusion with background [
40,
41]. What the last row adds is that decomposition is applied to a clinical grading task read at two nested levels, and the scope of the claim is the dental grading literature rather than detection at large. The test set was locked before training. A review of thirteen studies identifies patient selection and the absence of external validation as the dominant sources of bias in this literature [
15], and this study also lacks the second. Every headline comparison carries an interval, which appeared in only two prior rows, both for a commercial platform the authors had not trained [
18,
38].
Several studies hold advantages this study does not. Jin et al. validated externally at a second centre [
16], and Kazimierczak et al. used a cone beam computed tomography anchor [
18]. On the reference standard, however, the position is level rather than behind. Both panoramic quality studies used the consensus of two independent observers [
18,
19], and the annotation here was read independently by two specialist endodontists and adjudicated by a third. Two other studies used three readers: Khurshid et al., who cross-reviewed 15% of their images and reported a Cohen’s kappa of 0.89 [
12], and Dennis et al. [
42]. What this study offers is scale: 806 panoramic radiographs and 1668 graded roots against 189 root-filled teeth in the panoramic quality study that reports that count [
18].
Against that standing, seven limits bound what the study supports. The phantom work of
Section 4.2 is the first. Its lucencies are uniform ellipses of a single intensity on a uniform strip varied only in width, with no root contour, no trabecular background, and no projection blur, so they characterise the operator on those scenes rather than on radiographs. The response of the operator depends on the width of the strip a lucency sits in, and the cohort records no canal width, so where this cohort falls on that dependence is not established.
The second is the cohort itself. The radiographs come from a single centre, and no external cohort was used. They also come from one panoramic unit, and examinations of poor quality were excluded before analysis, so performance on unselected routine examinations is not established. The tooth number is unrecorded, and the filling material, sealer, and obturation technique are those of the institutional protocol stated in
Section 2.1 rather than a record kept root by root. Each bounds a different claim. Without a tooth number, no canal width can be recovered, even approximately. A root filled outside that protocol would be read against the wrong radiopacity comparator in
Table 1. Moreover, the technique sets how often a void occurs and where, and therefore sets the very ratio of the two labels whose difference
Section 4.2 is about, so that ratio may not transfer to other techniques. The working set of 806 radiographs and 1668 boxes is a modest cohort for a three-class problem.
The third is the label space. It omits overfilling, so the length axis runs in one direction only. Taper was not annotated either, since it is an objective of canal preparation rather than a requirement for the filling itself [
1], although the observer studies discussed in
Section 4.2 do score it beside length and density [
7,
8]. The rubric applied here therefore sits between two poles, the single-level criterion of the Belgian panoramic survey [
43] and the three criteria those observer studies score. Coronal restoration quality is not annotated, although it is associated with periapical health [
2,
4,
44] and, jointly with filling quality, in a further cohort [
3]. Single-label annotation, moreover, cannot express a tooth deficient on two axes at once.
The fourth is label noise this study introduced. Of the 95 separated-instrument boxes dropped before training, 24 were the only annotation their radiograph carried, so those films left the working set, which is what reduces 830 examinations to 806. The other 71 sit on 66 radiographs that carry other annotations and were retained, so those roots stay in the images with no label of their own, and the detector is trained to read a real root filling as background. It could act against detection recall on exactly the teeth a clinician would want flagged. The 71 boxes remain in the source annotation, and
Section 3.9 scores the detector against them. The detector places a box on most of these roots despite that training signal, so the cost falls on measured detection precision rather than on recall, and no reduction in recall is detected on the films that carry them.
The fifth is the projection. Panoramic radiography is not the projection on which endodontic treatment is usually assessed, which is intraoral periapical radiography [
45], so the labels were assigned on a projection a clinician would not choose. It also means that the millimetre thresholds of
Table 1 are operational radiographic criteria, read through magnification, distortion, and superimposition, rather than exact anatomical measurements. The evidence on whether the projection matters is not unanimous. A comparison of 138 endodontically-treated teeth imaged all three ways found a significant difference in obturation quality between each of the two projections and cone beam computed tomography, but none between the two projections themselves. Panoramic and periapical radiographs agreed on lesion identification at a kappa of 0.956 [
46]. A second property of the material bounds the reference standard further. An exploratory clustering of the radiographs on acquisition radiometry, part of the unplanned examination described in
Section 2.10, separates groups that differ in class prevalence at a chi-square
p of 1.89 × 10
−5, so a device signature is associated with the labels in this single-centre material. Nothing in the reported results exploits it, but it is the kind of association a model trained and tested within one centre can learn in place of the pathology.
The sixth is that every value rests on a single draw. Every configuration scored lower on the locked test set than in cross-validation on average precision, when both are read as means over single-fold models, with the axis-resolved baseline falling from 51.84 to 45.23 AP50 and the length class furthest in recall, 46.0% against 56.6%. The fold-fused readings of
Table 8 recover part of that, because fusing five models is a different aggregation and not a different measurement. YOLO11l at 1280 pixels completed three of the five folds, and the framework accepts but does not implement augmented inference for RT-DETR, so the transformer could not enter that comparison. A second seed moved the five-fold mean macro-F1 from 64.74 to 62.10, and the five-fold mean classification accuracy from 67.49% to 63.47%, below the 63.72% of the 1024-pixel baseline, so the ordering across the resolution ladder is not stable to it, and 1280 pixels is not an established optimum. No a priori power or sample size calculation was performed, and the cohort was assembled from examinations already carrying a root filling, so it is enriched relative to an unselected screening stream and the predictive values in
Section 4.5 will differ there even if sensitivity and specificity carry over.
The seventh is the central inference itself. The claim that the density axis is bounded by the evidence in the projection rests on convergent negative results. The measurement that would make it direct is an agreement statistic on this cohort, and although every radiograph was read twice, the two readings were not kept apart, so that statistic cannot be recovered from the annotation as it stands. This is the limitation that bears most directly on the grading results, because label noise lowers any measured agreement between model and reference, and its size on this cohort is unknown. Label noise on the density axis therefore remains an alternative this study cannot exclude. Some formulations train with annotator disagreement rather than around it, among them the regularised estimation of an annotator confusion matrix [
47] and the wider set of remedies surveyed for label noise in medical image analysis [
48]. These were not attempted here and are the most direct response to that alternative.
What was not tested bounds the conclusion as much as what was. Every configuration here is a supervised detector of one generation initialised from COCO-pretrained weights, and four branches were not opened. The first is self-supervised or foundation-model pretraining on dental radiographs. The second is a segmentation-first pipeline that would measure the apical termination against a delineated apex instead of inferring it from a box. The third is a hierarchical or multi-label formulation matching the staged rubric the Introduction describes. The fourth is a cost-sensitive or pairwise-margin loss aimed at the adequate against density exchange. No search over learning rate, loss weighting, class sampling, or augmentation was performed either, and since the mosaic result of
Section 3.2 shows that the augmentation recipe alone moves macro-F1 by more than any module tested, the recipe rather than the architecture may be where the remaining room is. Two estimators of irreducible error that a singly annotated set admits were also left unused: a confident-learning estimate over the stored out-of-fold predicted probabilities and the disagreement between the five per-fold models and the second seed.
Re-reading this cohort with the per-reader labels retained would yield the agreement statistic the present annotation cannot supply, converting the central inference into direct evidence and placing the model against a human upper bound. Repeating the analysis on periapical radiography would test the criterion where it is assessed clinically.
4.5. Clinical Implications
One implication follows from
Section 4.2 rather than from the detection figures. The reference standard records a void by its longest dimension while the operator responds to its shortest, so two lesions that
Table 1 calls the same size can return 8.8 and 46.7. A density flag is therefore not a statement about how large a void is, and a clinician should not read the absence of one as evidence that a void is small. What the flag tracks is transverse width, the dimension
Table 1 does not record. A root canal runs roughly vertically in the image, so a void’s transverse dimension lies along the horizontal axis, and that is the axis along which a moving-slit panoramic system’s magnification varies with the patient’s position relative to the image layer. The dimension the reference standard ignores and the dimension the projection degrades most are therefore the same dimension.
On the locked test set, the fused binary decision, the highest of the ten binary readings scored there, found 93.5% of the root fillings and graded 72.70% of matched boxes correctly, the unit throughout being a treated root and not a tooth. Taking a defective filling as the positive class, sensitivity was 76.25%, specificity 68.75%, the positive predictive value 73.05%, and the negative predictive value 72.26%. In counts, 38 of 160 defective fillings were returned as adequate, and 45 of 144 adequate fillings were flagged. Counting the 9 defective fillings the detector never found as false negatives, end-to-end sensitivity for a defective filling is 72.19%, and the reading the development folds would have selected gives 69.82%. At those figures, the decision could serve as a triage aid within the setting studied. It may be useful for teaching and for flagging teeth that require a periapical view. In an audit, it is useful only where each flagged tooth is confirmed by a reader rather than counted directly, since at this sensitivity and specificity an uncorrected count would overstate a true prevalence of 20% as 40%. It is not a screening instrument and not a prevalence estimator. The margin of 20.07 points over the majority class is in any case a weak reference, since calling every filling defective is not a clinical alternative. What such a margin does not supply is the net benefit of acting on the decision across the range of threshold probabilities a clinician might hold, which is what a decision curve reports [
49] and which this study does not compute. Both criteria it merges are associated with apical periodontitis in independent cross-sectional cohorts, inadequate length and non-homogeneous filling alike [
2,
3,
4,
5].
The two errors do not carry the same cost. A filling wrongly called adequate leaves a tooth that would require a periapical view unexamined, which is the error the use cases in
Section 1 are least able to absorb. A filling wrongly called defective, by contrast, costs a second view and the small exposure it carries. At the operating point reported here, the first error occurs in 23.75% of matched defective fillings, and in 27.81% once the fillings the detector never found are counted, whereas the second occurs in 31.25% of adequate ones. Trading one against the other would need a decision threshold on a calibrated score, which this study does not provide, and not the score threshold of
Section 2.7, which is a detection floor and moves recall as well as grading.
The axis-resolved reading is not deployable. It grades 60.54% of matched boxes correctly against a majority class of 48.83%, a margin of 11.71 points. It returns a positive predictive value of 46.67% on the density axis, so fewer than half of the density flags it raises are correct.
The cost of running the configuration these figures describe is small. Every clinical quantity above comes from the binary decision at 1024 pixels read with the five per-fold models fused, which was timed separately at 9.8 milliseconds per radiograph for one fold model on the card used here, so 49.1 for the five. On a processor alone, with no graphics card, the same single pass takes about 281 milliseconds and the fused reading about 1.4 s, a figure that bounds the order of magnitude and not the value. Both are well inside what a chairside or an audit workflow would absorb, and the second matters more than the first, because a dental clinic is far likelier to hold the processor than the card. The ensemble of
Section 3.5 is the expensive object, at 354 milliseconds for its fold-fused reading against 12.4 for a single pass, and it is not the system characterised here. None of that bears on why the density axis resists, since inference cost and the ceiling on accuracy are different quantities.
The binary decision could suit teaching and reader-confirmed audit once externally validated, but not the indication of retreatment, and the axis along which a defective filling departs should not be reported.
The system cannot report material extruded beyond the apex, which this label space does not carry. The prevalence literature associates extrusion with apical periodontitis more strongly than a short filling, at odds ratios of 2.51 against 1.76 in the largest of these cohorts [
2]. Every comparable study cited here gives extruded material a place, as a separate variable, as a third length category, or folded into an unacceptable length, so its absence here is a difference from the field and not a convention of it. A systematic review of primary root canal treatment outcomes identified four conditions that significantly improved the outcome, among them a root filling extending to within 2 mm of the radiographic apex and a root filling with no voids [
50]. An extruded filling fails the first of those two. Whether it also carries voids is a separate question that extrusion does not settle, and this study does not measure. As
Section 2.1 records, such roots were labelled on the position of the main filling mass, so most of them fall into the adequate class and are counted as adequate wherever an adequate class appears, including the specificity of
Table 9. A system trained on this scheme therefore cannot flag extrusion and will call most extruded roots acceptable.
Radiographic obturation quality is, moreover, one component of endodontic treatment quality and not a measure of it. A filling records the end of a sequence whose earlier steps the radiograph does not show. How well the canal was cleaned and disinfected depends on the irrigation protocol, which governs the removal of debris and smear layer and how far the irrigant penetrates [
51]. Access cavity design in turn influences canal location, instrumentation, irrigation, and the obturation that follows [
52]. A radiographic grade of the filling is therefore not a surrogate for the quality of the treatment as a whole, and the decisions discussed above concern the filling alone.
Every figure above also belongs to the setting in which it was measured: one centre, one panoramic unit, fillings placed by cold lateral condensation with an epoxy resin sealer under one institutional protocol, and examinations that had already passed exclusion for image quality, which removed 212 of the 1129 screened. In routine use, a system of this kind would encounter blurred, distorted, or heavily restored films, and other devices, sealers, and techniques, so its sensitivity and specificity describe this setting and not panoramic radiography in general. External validation on other centres and devices, and on unselected examinations, is required before transfer can be claimed.