1. Introduction
CT pulmonary angiography (CTPA) is the gold standard for diagnosing pulmonary embolism (PE). However, suboptimal image quality remains a major challenge, as it may lead to diagnostic inaccuracies and potentially serious consequences in patients with suspected PE [
1,
2]. Among the factors contributing to suboptimal CTPA quality, poor vascular opacification is one of the most common causes [
3,
4].
Poor opacification may result from inappropriate bolus timing, transient interruption of contrast inflow due to deep inspiration and breath-holding, hemodynamic alterations, or large patient body habitus [
5,
6]. Adequate control of contrast administration, proper respiratory instruction, and the use of low kVp settings can help maintain sufficient vascular enhancement in CTPA [
6,
7]. Nonetheless, a practical solution is needed for cases where CTPA images have already been acquired with inadequate pulmonary artery opacification, as repeating the examination increases the patient’s exposure to contrast media and radiation.
Dual-energy CT (DECT) has shown promise in improving the diagnostic quality of suboptimally enhanced CTPA studies. Low-keV VMI reconstructions, particularly at 40 keV, significantly increase pulmonary artery attenuation, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) compared to standard images [
8,
9,
10]. These enhancements can substantially improve the diagnostic accuracy of PE detection in cases with inadequate contrast opacification [
8]. However, not all CTPA studies are performed using DECT technology.
Recently, deep learning has emerged as a powerful tool for improving image quality and, consequently, diagnostic accuracy in CT imaging [
11,
12,
13,
14]. Among these approaches, iodine contrast–augmentation algorithms have been developed to enable CT imaging with reduced doses of iodine contrast [
15]. Previous studies have applied this technique to CT scans performed under predefined conditions of reduced radiation and iodine load, demonstrating significant improvements in image quality compared with scans without contrast augmentation [
16,
17,
18]. We hypothesized that this iodine contrast–augmentation algorithm could also improve the image quality of CTPA scans that are unintentionally suboptimally enhanced, a clinically relevant challenge in daily practice.
Therefore, this study aimed to evaluate the effect of a deep learning–based iodine contrast–augmentation (DLCA) algorithm on image quality and diagnostic performance for PE detection in suboptimally enhanced CTPA.
2. Materials and Methods
This study was approved by the Institutional Review Board (Gyeongsang National University Changwon Hospital, IRB No. 2025-07-018). The requirement for informed consent was waived due to the retrospective nature of the study and the use of anonymized data. A schematic flow diagram of the overall study methodology is presented in
Figure 1.
2.1. Study Population
Among 2367 CTPA examinations performed between May 2020 and March 2025, a total of 103 cases with suboptimal pulmonary artery enhancement and available follow-up CTPA within 3 days were retrospectively included (
Figure 2). Suboptimal enhancement was defined as a CT attenuation value of less than 250 Hounsfield units (HU) in the pulmonary trunk, based on the image quality assessment reported during CT interpretation. This 250-HU threshold was chosen because it has been considered optimal for confident evaluation of acute PE in previous studies [
19,
20]. The study cohort comprised 76 male and 27 female patients, with a mean age of 57.6 years (range, 19–92 years).
2.2. CT Image Acquisition and Postprocessing
CTPA examinations were performed using one of three CT scanners: IQon (Philips Healthcare, Best, the Netherlands), Aquilion ONE (Canon Medical Systems, Otawara, Japan), or SOMATOM Force (Siemens Healthcare, Forchheim, Germany). CT scanning parameters were as follows: 120 kVp, 140–250 mA, 2 mm slice thickness, 2 mm reconstruction interval, and a smooth reconstruction filter. Images were obtained during a single breath-hold, covering the lung apex to the costophrenic angles in the craniocaudal direction. For contrast enhancement, 70–90 mL of iodinated contrast medium (iohexol; Omnipaque 300, GE Healthcare, Shanghai, China) was injected through an antecubital vein at 3.5–4 mL/s, followed by a 30 mL contrast–saline mixture and a 20 mL saline flush, both at the same rate. Bolus tracking in the pulmonary trunk was used for individualized timing with a threshold of 150 Hounsfield units (HU).
All suboptimal CTPA images were post-processed using a DLCA algorithm (ClariACE, ClariPi Inc., Seoul, Korea) to enhance pulmonary vascular opacification. The DLCA is a commercially available contrast-boosting model developed for low–contrast-dose CT, employing a two-stage U-net architecture. Further technical details of the model have been described in previous studies [
16,
17].
2.3. Image Quality Evaluation
A radiologist with three years of experience measured attenuation values in HU and standard deviations in the pulmonary trunk, back muscle, and room air on the original CTPA images using circular regions of interest (ROIs) placed on axial images. The ROIs were then copied and pasted at the same anatomical locations on the DLCA-processed images. Each measurement was performed three times, and the mean value was recorded as the representative measurement.
The CNR of the pulmonary artery was calculated according to the following equation: CNR = (SIPA − SImuscle)/BN, where SIPA represents the mean signal intensity of the pulmonary trunk, SImuscle is the mean signal intensity of the back muscle, and BN denotes the background noise, defined as the standard deviation of attenuation in room air. The SNR was calculated as: SNR = SIPA/BN.
2.4. PE Detection
Two thoracic radiologists with 29 and 30 years of experience, respectively, and one pulmonologist jointly reviewed all suboptimal CTPA images, along with follow-up CTPA studies, 40-keV VMIs when available, and relevant clinical data to establish the reference standard for PE. Readers assessed PE by identifying the presence of an endoluminal filling defect partially or completely occluding pulmonary arteries [
21].
For the comparison of PE detection performance, the pulmonary arteries were divided into 20 anatomical segments (1 trunk, 2 main, 2 interlobar, 5 lobar, 5 segmental, and 5 subsegmental) according to vascular anatomy [
22]. PEs from the trunk through the lobar arteries were classified as central, whereas PEs in the segmental and subsegmental arteries were classified as peripheral. PE detection in each segment was independently assessed by two radiologists with 10 and 5 years of experience, first using suboptimal CTPA alone and then, after a 4-week interval, using both the original and DLCA-processed images.
2.5. Statistical Analysis
Continuous variables were expressed as means ± standard deviations, and categorical variables as frequencies or percentages. Original and DLCA-processed CT images were compared using paired t-tests. Diagnostic accuracy for PE detection was assessed with receiver operating characteristic (ROC) analysis, and pulmonary artery attenuation thresholds were evaluated by comparing area under the curve (AUC) values in 10-HU increments. Inter-reader agreement was assessed using kappa statistics, interpreted as follows: ≤0.20, poor; 0.21–0.40, fair; 0.41–0.60, moderate; 0.61–0.80, substantial; and ≥0.81, almost perfect. Statistical analyses were conducted using SPSS (version 27) and MedCalc (version 23.1.7), with p < 0.05 considered statistically significant.
3. Results
Among 103 cases with suboptimal enhancement, the causes included incorrect bolus tracking or delayed scan initiation (n = 37), transient interruption of contrast flow (n = 29), contrast extravasation (n = 11), extremely large body habitus (n = 10), low injection flow rates (n = 9), markedly reduced cardiac function (n = 3), technical errors (n = 4).
DLCA processing improved pulmonary artery opacification by 57.7% (159.0 ± 25.8 HU → 250.8 ± 46.3 HU;
p < 0.001) and reduced noise by 56.7% (12.7 ± 2.3 HU → 5.5 ± 1.0 HU;
p < 0.001). Consequently, SNR and CNR increased markedly (SNR: 13.2 → 47.5; CNR: 8.7 → 37.2; both
p < 0.001) (
Table 1).
According to the standard reference, PEs were identified in 195 segments (83 central, 112 peripheral) across 35 patients, while the remaining 1865 segments from 68 patients were negative.
For PE detection, the use of DLCA-processed images, in conjunction with the original suboptimal CTPA, significantly improved accuracy for overall, central, and peripheral PE detection (all
p < 0.05) by both readers (
Table 2) (
Figure 3,
Figure 4,
Figure 5 and
Figure 6). Inter-reader agreement for PE detection was moderate for both suboptimal CTPA (κ = 0.65) and DLCA-processed images (κ = 0.74).
A pulmonary artery attenuation threshold of 130 HU on the original CTPA was identified, above which DLCA processing enhanced diagnostic performance for PE detection in both readers (
p < 0.001) (
Table 3).
4. Discussion
Deep learning-based techniques in thoracic imaging have achieved state-of-the-art performance in image reconstruction and post-processing, enabling advanced imaging capabilities that were previously unattainable [
23,
24,
25,
26]. The DLCA algorithm generates contrast-augmented images without requiring specialized hardware such as DECT and was originally developed to reduce the amount of contrast needed in CT imaging. Accordingly, previous studies have applied this algorithm to CT images acquired under predefined low-contrast protocols, including pediatric examinations and adult scans performed with low-kVp settings [
15,
16,
27]. In particular, a CTPA study using DLCA [
18] under low-radiation and low-contrast-dose protocols demonstrated that images processed with this algorithm achieved superior vascular opacification, CNR, and SNR compared with CTPA reconstructed using various methods without DLCA processing.
In our study, we specifically evaluated suboptimally enhanced CTPA encountered in routine clinical practice. Consistent with previous reports, DLCA processing improved vascular opacification and reduced image noise, thereby significantly enhancing both CNR and SNR. Moreover, DLCA processing increased diagnostic accuracy for PE detection compared with suboptimal CTPA alone, for overall PEs as well as for both central and peripheral PEs. By contrast, a prior study reported that DLCA was superior for detecting PEs in small vessels, such as segmental and subsegmental arteries [
18]. This discrepancy may be explained by their use of low-kVp protocols, in which pulmonary artery attenuation exceeded 250 HU in all groups; consequently, no intergroup differences were observed in detecting more proximal PEs. Our findings therefore extend the applicability of DLCA by demonstrating its value in real-world suboptimal CTPA, a clinically relevant challenge not addressed in previous studies.
In addition, we identified a pulmonary artery attenuation threshold of 130 HU—comparable to that of a venous-phase image—above which DLCA-processed images enhanced diagnostic performance for PE detection, increasing attenuation by approximately 1.6-fold and improving pulmonary artery CNR by about 4.3-fold. Previously, virtual arterial-phase images generated using low-energy VMI reconstruction from DECT have shown potential as a feasible alternative to true CTPA or aortic angiography [
28,
29,
30]. Considering the potential of DLCA demonstrated by our findings, further studies in larger cohorts are warranted to explore whether routine venous-phase chest CT images could serve as an alternative to CTPA when using DLCA, eliminating the need for rapid contrast injection.
Our study has several limitations. First, it was retrospectively designed and included a relatively small number of cases from a single institution, which may limit the generalizability of the findings. Nevertheless, the data reflect real-world practice over an extended period. Importantly, because DLCA can be applied independently of the type of CT scanner used, the algorithm has the potential to be implemented in other institutions and across different clinical settings. Second, the reference standard for PE detection was not based on optimally enhanced CTPA performed on the same day, but rather on follow-up CTPA performed the following 3 days, 40-keV VMI images when available, and clinical evaluation. Third, because of the limited number of positive PE cases, detection performance was evaluated on a per-segment basis.
5. Conclusions
Suboptimal opacification is a common cause of limited or nondiagnostic CTPA, which may lead to misdiagnosis, repeat examinations, unnecessary radiation or contrast exposure, and delays in patient care. In this study, the application of the DLCA algorithm significantly improved image quality and diagnostic performance for PE detection in suboptimally enhanced CTPA. These findings underscore the clinical value of DLCA as a practical and widely applicable solution for optimizing CTPA scans without additional contrast media or radiation exposure. Future studies with larger cohorts are warranted to validate these results and further define the role of DLCA in routine clinical practice.
Author Contributions
Conceptualization, K.N.J.; methodology, K.N.J. and K.B.; software, K.B.; validation, K.N.J., K.B. and T.H.K.; formal analysis, K.B.; investigation, K.N.J., K.B. and T.H.K.; data curation, K.B.; writing—original draft preparation, K.B. and K.N.J.; writing—review and editing, K.N.J., K.B. and T.H.K.; supervision, K.N.J.; project administration, K.B. All authors have read and agreed to the published version of the manuscript.
Funding
This work was supported by the biomedical research institute fund (GNUCHBRIF-2025-001) from Gyeongsang National University Changwon Hospital.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review Board of Gyeongsang National University Changwon Hospital (protocol code, 2025-07-018; date of approval, 25 July 2025).
Informed Consent Statement
Patient consent was waived by the IRB due to the retrospective design of the study and the use of anonymized data. Written informed consent has been obtained from the patients (in
Figure 4,
Figure 5 and
Figure 6) to publish this paper.
Data Availability Statement
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
Acknowledgments
The authors thank J.I. Moon, S.G. Lim, and N.H. Kim for their support in data analysis and image interpretation.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
CTPA | Computed tomography pulmonary angiography |
PE | Pulmonary embolism |
DECT | Dual-energy computed tomography |
SNR | Signal-to-noise ratio |
CNR | Contrast-to-noise ratio |
DLCA | Deep learning-based iodine contrast augmentation algorithm |
ROC | Receiver operating characteristic |
ROI | Region of interest |
HU | Hounsfield units |
References
- Ahuja, J.; Palacio, D.; Jo, N.; Strange, C.D.; Shroff, G.S.; Truong, M.T.; Wu, C.C. Pitfalls in the imaging of pulmonary embolism. Semin. Ultrasound CT MRI 2022, 43, 221–229. [Google Scholar] [CrossRef]
- Wittenberg, R.; Peters, J.F.; Sonnemans, J.J.; Bipat, S.; Prokop, M.; Schaefer-Prokop, C.M. Impact of image quality on the performance of computer-aided detection of pulmonary embolism. AJR Am. J. Roentgenol. 2011, 196, 95–101. [Google Scholar] [CrossRef]
- Bates, D.D.; Tkacz, J.N.; LeBedis, C.A.; Holalkere, N. Suboptimal CT pulmonary angiography in the emergency department: A retrospective analysis of outcomes in a large academic medical center. Emerg. Radiol. 2016, 23, 603–607. [Google Scholar] [CrossRef] [PubMed]
- Parra Caballero, P.; Ruiz Berraco, F.; Salvador Rodriguez, A.; Sanchez Juez, A.; Ruiz-Gimenez Arrieta, N.; Bustos Carpio, J.; Rodriguez Revillas, A.; Lopez Villar, I.; de Benavides Bernaldo de Quiros, M.D.C.; Delgado Heredia, M.J. Suboptimal CT pulmonary angiography for PE diagnosis, is it worth further study? A retrospective 7.5-year evaluation in a large academic medical center. Thromb. Res. 2025, 250, 109324. [Google Scholar] [CrossRef] [PubMed]
- Sudarski, S.; Haubenreisser, H.; Henzler, T.; Reischauer, C.; Kolokythas, O.; Matoori, S.; Herzog, B.A.; Schonberg, S.O.; Gutzeit, A. Incidence of transient interruption of contrast (TIC)—A retrospective single-centre analysis in CT pulmonary angiography exams acquired during inspiratory breath-hold with the breathing command: “Please inspire gently!”. PLoS ONE 2019, 14, e0210473. [Google Scholar] [CrossRef]
- Chaturvedi, A.; Oppenheimer, D.; Rajiah, P.; Kaproth-Joslin, K.A.; Chaturvedi, A. Contrast opacification on thoracic CT angiography: Challenges and solutions. Insights Imaging 2017, 8, 127–140. [Google Scholar] [CrossRef]
- Viteri-Ramírez, G.; García-Lallana, A.; Simón-Yarza, I.; Broncano, J.; Ferreira, M.; Pueyo, J.C.; Villanueva, A.; Bastarrika, G. Low radiation and low-contrast dose pulmonary CT angiography: Comparison of 80 kVp/60 ml and 100 kVp/80 ml protocols. Clin. Radiol. 2012, 67, 833–839. [Google Scholar] [CrossRef]
- Bae, K.; Jeon, K.N.; Cho, S.B.; Park, S.E.; Moon, J.I.; Baek, H.J.; Choi, B.H. Improved Opacification of a Suboptimally Enhanced Pulmonary Artery in Chest CT: Experience Using a Dual-Layer Detector Spectral CT. AJR Am. J. Roentgenol. 2018, 210, 734–741. [Google Scholar] [CrossRef]
- Bayraktaroglu, S.; Cinkooglu, A.; Ceylan, N.; Savas, R. Salvage of Suboptimal Enhancement of Pulmonary Artery in Pulmonary CT Angiography Studies: Rapid kVp Switch Dual Energy CT Experience. Iran J. Radiol. 2020, 17, e97230. [Google Scholar] [CrossRef]
- Leithner, D.; Wichmann, J.L.; Vogl, T.J.; Trommer, J.; Martin, S.S.; Scholtz, J.E.; Bodelle, B.; De Cecco, C.N.; Duguay, T.; Nance, J.W., Jr.; et al. Virtual Monoenergetic Imaging and Iodine Perfusion Maps Improve Diagnostic Accuracy of Dual-Energy Computed Tomography Pulmonary Angiography with Suboptimal Contrast Attenuation. Investig. Radiol. 2017, 52, 659–665. [Google Scholar] [CrossRef]
- Lenfant, M.; Chevallier, O.; Comby, P.O.; Secco, G.; Haioun, K.; Ricolfi, F.; Lemogne, B.; Loffroy, R. Deep Learning Versus Iterative Reconstruction for CT Pulmonary Angiography in the Emergency Setting: Improved Image Quality and Reduced Radiation Dose. Diagnostics 2020, 10, 558. [Google Scholar] [CrossRef]
- Gong, H.; Peng, L.; Du, X.; An, J.; Peng, R.; Guo, R.; Ma, X.; Xiong, S.; Ma, Q.; Zhang, G.; et al. Artificial Intelligence Iterative Reconstruction in Computed Tomography Angiography: An Evaluation on Pulmonary Arteries and Aorta with Routine Dose Settings. J. Comput. Assist. Tomogr. 2024, 48, 244–250. [Google Scholar] [CrossRef]
- Cheikh, A.B.; Gorincour, G.; Nivet, H.; May, J.; Seux, M.; Calame, P.; Thomson, V.; Delabrousse, E.; Crombe, A. How artificial intelligence improves radiological interpretation in suspected pulmonary embolism. Eur. Radiol. 2022, 32, 5831–5842. [Google Scholar] [CrossRef] [PubMed]
- Langius-Wiffen, E.; de Jong, P.A.; Hoesein, F.A.M.; Dekker, L.; van den Hoven, A.F.; Nijholt, I.M.; Boomsma, M.F.; Veldhuis, W.B. Retrospective batch analysis to evaluate the diagnostic accuracy of a clinically deployed AI algorithm for the detection of acute pulmonary embolism on CTPA. Insights Imaging 2023, 14, 102. [Google Scholar] [CrossRef] [PubMed]
- Shin, D.J.; Choi, Y.H.; Lee, S.B.; Cho, Y.J.; Lee, S.; Cheon, J.E. Correction to: Low-iodine-dose computed tomography coupled with an artificial intelligence-based contrast-boosting technique in children: A retrospective study on comparison with conventional-iodine-dose computed tomography. Pediatr. Radiol. 2024, 54, 1412. [Google Scholar] [CrossRef] [PubMed]
- Lee, T.; Yoon, J.H.; Park, J.Y.; Lee, J.; Choi, J.W.; Ahn, C.; Lee, J.M. Deep learning-based iodine contrast-augmenting algorithm for low-contrast-dose liver CT to assess hypovascular hepatic metastasis. Abdom. Radiol. 2023, 48, 3430–3440. [Google Scholar] [CrossRef]
- Kang, H.J.; Lee, J.M.; Ahn, C.; Bae, J.S.; Han, S.; Kim, S.W.; Yoon, J.H.; Han, J.K. Low dose of contrast agent and low radiation liver computed tomography with deep-learning-based contrast boosting model in participants at high-risk for hepatocellular carcinoma: Prospective, randomized, double-blind study. Eur. Radiol. 2023, 33, 3660–3670. [Google Scholar] [CrossRef]
- Park, M.; Hwang, M.; Lee, J.W.; Kim, K.I.; Ahn, C.; Suh, Y.J.; Jeong, Y.J. Application of a Deep Learning-Based Contrast-Boosting Algorithm to Low-Dose Computed Tomography Pulmonary Angiography with Reduced Iodine Load. J. Comput. Assist. Tomogr. 2025, 49, 288–296. [Google Scholar] [CrossRef]
- Jones, S.E.; Wittram, C. The indeterminate CT pulmonary angiogram: Imaging characteristics and patient clinical outcome. Radiology 2005, 237, 329–337. [Google Scholar] [CrossRef]
- Bae, K.T.; Tao, C.; Gurel, S.; Hong, C.; Zhu, F.; Gebke, T.A.; Milite, M.; Hildebolt, C.F. Effect of patient weight and scanning duration on contrast enhancement during pulmonary multidetector CT angiography. Radiology 2007, 242, 582–589. [Google Scholar] [CrossRef]
- Wittram, C.; Maher, M.M.; Halpern, E.F.; Shepard, J.A. Attenuation of acute and chronic pulmonary emboli. Radiology 2005, 235, 1050–1054. [Google Scholar] [CrossRef] [PubMed]
- Kandathil, A.; Chamarthy, M. Pulmonary vascular anatomy & anatomical variants. Cardiovasc. Diagn. Ther. 2018, 8, 201–207. [Google Scholar] [CrossRef] [PubMed]
- Bae, K.; Lee, J.; Jung, Y.; de Arcos, J.; Jeon, K.N. Deep learning reconstruction for zero echo time lung magnetic resonance imaging: Impact on image quality and lesion detection. Clin. Radiol. 2024, 79, e1296–e1303. [Google Scholar] [CrossRef] [PubMed]
- Zhou, Z.; Gao, Y.; Zhang, W.; Bo, K.; Zhang, N.; Wang, H.; Wang, R.; Du, Z.; Firmin, D.; Yang, G.; et al. Artificial intelligence-based full aortic CT angiography imaging with ultra-low-dose contrast medium: A preliminary study. Eur. Radiol. 2023, 33, 678–689. [Google Scholar] [CrossRef]
- Hwang, E.J.; Park, C.M. Clinical Implementation of Deep Learning in Thoracic Radiology: Potential Applications and Challenges. Korean J. Radiol. 2020, 21, 511–525. [Google Scholar] [CrossRef]
- Koetzier, L.R.; Mastrodicasa, D.; Szczykutowicz, T.P.; van der Werf, N.R.; Wang, A.S.; Sandfort, V.; van der Molen, A.J.; Fleischmann, D.; Willemink, M.J. Deep Learning Image Reconstruction for CT: Technical Principles and Clinical Prospects. Radiology 2023, 306, e221257. [Google Scholar] [CrossRef]
- Yoon, J.H.; Park, J.Y.; Lee, S.M.; Lee, E.S.; Kim, J.H.; Lee, J.M. Renal protection CT protocol using low-dose and low-concentration iodine contrast medium in at-risk patients of HCC and with chronic kidney disease: A randomized controlled non-inferiority trial. Cancer Imaging 2023, 23, 100. [Google Scholar] [CrossRef]
- Foti, G.; Silva, R.; Faccioli, N.; Fighera, A.; Menghini, R.; Campagnola, A.; Carbognin, G. Identification of pulmonary embolism: Diagnostic accuracy of venous-phase dual-energy CT in comparison to pulmonary arteries CT angiography. Eur. Radiol. 2021, 31, 1923–1931. [Google Scholar] [CrossRef]
- Patel, A.A.; Sutphin, P.D.; Xi, Y.; Abbara, S.; Kalva, S.P. Arterial Phase CTA Replacement by a Virtual Arterial Phase Reconstruction from a Venous Phase CTA: Preliminary Results Using Detector-Based Spectral CT. Cardiovasc. Interv. Radiol. 2019, 42, 250–259. [Google Scholar] [CrossRef]
- Weiss, J.; Notohamiprodjo, M.; Bongers, M.; Schabel, C.; Mangold, S.; Nikolaou, K.; Bamberg, F.; Othman, A.E. Effect of Noise-Optimized Monoenergetic Postprocessing on Diagnostic Accuracy for Detecting Incidental Pulmonary Embolism in Portal-Venous Phase Dual-Energy Computed Tomography. Investig. Radiol. 2017, 52, 142–147. [Google Scholar] [CrossRef]
| Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).