Article

Evaluation of the New Beckmann Coulter Analyzer DxH 900 Compared to Sysmex XN20: Analytical Performance and Flagging Efficiency

by
Maite Serrando Querol
1,2,*,
Javier Nieto-Moragas
1,2,
Anna Marull Arnall
1,2,
Meritxell Deulofeu Figueras
1,2 and
Orlando Jiménez-Romero
1,2
1
Hematology Laboratory, Hospital Universitari de Girona Doctor Josep Trueta, 17190 Girona, Catalonia, Spain
2
Research Group of Clinical Anatomy, Embryology and Neuroscience (NEOMA), Department of Medical Science, Faculty of Medicine, Universitat de Girona, 17003 Girona, Catalonia, Spain
*
Author to whom correspondence should be addressed.
Diagnostics 2021, 11(10), 1756; https://doi.org/10.3390/diagnostics11101756
Submission received: 17 August 2021 / Revised: 20 September 2021 / Accepted: 21 September 2021 / Published: 24 September 2021
(This article belongs to the Special Issue Advances in Hematology Laboratory)

Abstract

Efficiency and accuracy in automated hematology analyzers are very important for clinical laboratories. The purpose was to evaluate the flags and results reported by the newest Beckman Coulter analyzer DxH 900 compared to the Sysmex XN20 system. Samples were analyzed on the XN20 (Sysmex, Kobe, Japan) and on the Beckman Coulter DxH 900 (Beckman Coulter, Miami, Florida, USA). Slide reviews were performed microscopically. Morphologic criteria were used to identify abnormal cells as recommended by International Consensus Group for Hematology (ICSH): blasts, immature granulocytes (IG%), abnormal lymphocytes (ALs) and plasma cells. Results: there was a strong correlation between the analyzers in almost all clinical parameters tested. Both DxH 900 and XN20 showed an excellent degree of association for the leukocyte differential compared to the reference method (manual microscopy). When it comes to IG%, XN20 showed a positive bias for higher results. Related to platelets, there are no differences between the two methods for PLT count. For mean platelet volume (MPV), DxH 900 provided 100% results of the samples analyzed while XN20 while in the XN20 analyzer, 16% of the results were missing. From our results we came to the conclusion that both analyzers, DxH 900 and XN20 were clinically accurate and efficient. Abnormal Lymphocyte detection highlighted the differences between the two technologies as only minimal agreement was obtained. DxH 900 demonstrated higher sensitivity in detecting IG with good correlation with microscopic review. The DxH 900 for platelet clumps identification provides an excellent flag (PLT Clumps) with the highest sensitivity observed in our evaluation.
Keywords: cell blood count (CBC); leukocyte differential; Beckman Coulter DxH 900; Sysmex XN20 cell blood count (CBC); leukocyte differential; Beckman Coulter DxH 900; Sysmex XN20

Share and Cite

MDPI and ACS Style

Serrando Querol, M.; Nieto-Moragas, J.; Marull Arnall, A.; Figueras, M.D.; Jiménez-Romero, O. Evaluation of the New Beckmann Coulter Analyzer DxH 900 Compared to Sysmex XN20: Analytical Performance and Flagging Efficiency. Diagnostics 2021, 11, 1756. https://doi.org/10.3390/diagnostics11101756

AMA Style

Serrando Querol M, Nieto-Moragas J, Marull Arnall A, Figueras MD, Jiménez-Romero O. Evaluation of the New Beckmann Coulter Analyzer DxH 900 Compared to Sysmex XN20: Analytical Performance and Flagging Efficiency. Diagnostics. 2021; 11(10):1756. https://doi.org/10.3390/diagnostics11101756

Chicago/Turabian Style

Serrando Querol, Maite, Javier Nieto-Moragas, Anna Marull Arnall, Meritxell Deulofeu Figueras, and Orlando Jiménez-Romero. 2021. "Evaluation of the New Beckmann Coulter Analyzer DxH 900 Compared to Sysmex XN20: Analytical Performance and Flagging Efficiency" Diagnostics 11, no. 10: 1756. https://doi.org/10.3390/diagnostics11101756

APA Style

Serrando Querol, M., Nieto-Moragas, J., Marull Arnall, A., Figueras, M. D., & Jiménez-Romero, O. (2021). Evaluation of the New Beckmann Coulter Analyzer DxH 900 Compared to Sysmex XN20: Analytical Performance and Flagging Efficiency. Diagnostics, 11(10), 1756. https://doi.org/10.3390/diagnostics11101756

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop