Next Article in Journal
Radioiodine-131 Therapy Used for Differentiated Thyroid Cancer Can Impair Titanium Dental Implants: An In Vitro Analysis
Next Article in Special Issue
Clinical Applications of Magnetic Resonance-Guided Radiotherapy: A Narrative Review
Previous Article in Journal
Molecular In-Depth Characterization of Chondrosarcoma for Current and Future Targeted Therapies
Previous Article in Special Issue
Dosimetric Benefit of Adaptive Magnetic Resonance-Guided Stereotactic Body Radiotherapy of Liver Metastases
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Review

Image-Guided Proton Therapy: A Comprehensive Review

James M. Slater, MD Proton Treatment and Research Center, Loma Linda University, Loma Linda, CA 92354, USA
*
Author to whom correspondence should be addressed.
Cancers 2023, 15(9), 2555; https://doi.org/10.3390/cancers15092555
Submission received: 4 March 2023 / Revised: 24 April 2023 / Accepted: 26 April 2023 / Published: 29 April 2023
(This article belongs to the Special Issue Image-Guided Radiotherapy for Cancers)

Abstract

:

Simple Summary

In proton therapy, there is a sharp peak in the delivered dose followed by a rapid falloff, known as the Bragg peak, which is not present in photons. This allows for treatment plans that deliver lower doses to normal tissue than can be performed with photons. This requires a high degree of accuracy and precision of delivery due to the short distance between an area of high and low doses. Image guidance allows for better visualization of the target and more accurate delivery of proton (and photon) radiation. The equipment used to deliver proton therapy differs in several ways from that of photon radiation, which impacts the methods used for image guidance in proton therapy. This paper aims to summarize the various methods of image guidance in current proton therapy and their relative advantages and disadvantages, as well as areas for future improvements.

Abstract

Image guidance for radiation therapy can improve the accuracy of the delivery of radiation, leading to an improved therapeutic ratio. Proton radiation is able to deliver a highly conformal dose to a target due to its advantageous dosimetric properties, including the Bragg peak. Proton therapy established the standard for daily image guidance as a means of minimizing uncertainties associated with proton treatment. With the increasing adoption of the use of proton therapy over time, image guidance systems for this modality have been changing. The unique properties of proton radiation present a number of differences in image guidance from photon therapy. This paper describes CT and MRI-based simulation and methods of daily image guidance. Developments in dose-guided radiation, upright treatment, and FLASH RT are discussed as well.

1. Introduction

Image guidance for external beam radiation therapy has revolutionized the field of radiation oncology. Image guidance is the use of imaging at the pretreatment or treatment stage that leads to an action that improves or verifies the accuracy of radiotherapy. Due to the Bragg peak, proton therapy (PT) is able to deliver a highly conformal dose to a target; however, the Bragg peak must be correctly placed in the target in order to utilize the potential of proton therapy. Compared to photons, protons have additional uncertainties in the range, or penetration, of the beam in tissue; these uncertainties are predominately influenced by the densities of tissue through which the beam passes. Image guidance (along with proper immobilization) minimizes these uncertainties and illustrates the greater importance it has on proton therapy compared to photon therapy. Further illustrating this point is that when the first hospital-based proton therapy center began operation in 1990, image guidance was used for every fraction of treatment that was delivered [1].
The patient must be positioned properly prior to treatment. This can be achieved by planar X-rays, CT scanners, and visual-based methods. There is also increased interest in using MRIs for positioning, although this has some challenges. The target may also be monitored during the treatment for motion. There are differences in the ways that image guidance is used for protons vs. photons due to the intrinsic properties of protons as well as the differences in hardware between linear accelerators and proton gantries. Dose-guided radiation therapy seeks to spatially measure doses from the byproducts of proton radiation. Finally, FLASH radiotherapy, an ultrahigh dose rate method of external beam radiation, which will involve predominantly protons, has its own set of unique issues for image guidance that will be mentioned.

2. Imaging for Simulation

Different methods of imaging have differing benefits and drawbacks when used for simulation (See Table 1).

2.1. CT-Based Simulation

One key difference between protons and photons when using CT scans for planning is the additional step of deriving proton-stopping power from Hounsfield Units (HU). This process of deriving the actions of the charged proton from the actions of the chargeless, massless X-ray photon leads to greater range uncertainty, roughly 3.5% of the absolute range [2,3]. Additional margins have to be added to target volumes to compensate, decreasing the therapeutic benefit of protons.
Dual-energy CT can be used to better differentiate the composition of materials [4] and further reduce the range uncertainty [5,6]. Another solution is the use of proton CT, which allows for direct reconstruction of proton stopping power, although these are currently not in widespread use due to issues with spatial resolution [7]. One group has compared images from a clinically realistic proton CT scanner to those from photon CT scanners and found proton-stopping power measurement discrepancies of up to 40% in regions with mixed content of air, soft tissue, and bone, such as sinuses [8]. A comparison between a dual-energy CT scanner and a proton CT prototype in calculating relative stopping power (RSP) of phantoms of a known RSP showed that both scanners were accurate to within 1%, that the proton CT was slightly more accurate than the dual-energy CT scanner, despite characteristic artifacts reducing the accuracy of the proton CT [9].

2.2. MRI-Based Simulation

MRI-only simulation for treatment planning has also been explored. One benefit of this approach is the superior contrast with MRI when compared to CT. This also may eliminate the need for registration of an MRI to a CT simulation, which introduces additional uncertainty, especially when an MRI is obtained at a different timepoint or in a different position from the simulation CT. Possible drawbacks include an initial unfamiliarity with performing MRI simulation as well as possible incompatibility and artifacts with metallic hardware such as older pacemakers [10].
Another drawback of MRI-only simulation relevant for both proton-based (as well as photon-based) treatment is a possible geometric distortion of images due to changes in the magnetic field uniformity from both the MR system as well as the patient’s body itself. An excellent review by Schmidt and Payne goes into further detail [11].
The main challenge for implementing MRI-only simulation is that the MRI data needs to be converted into a pseudo-CT dataset for patient setup and dose calculation. One strategy includes assigning electron density values to predefined tissue levels [12,13]. Another involves the creation of an MR-to-CT image atlas, with the registration of the patient’s MRI to the atlas and conversion of the MRI intensity to HU values [14]. Each comes with drawbacks, which makes them not ideal for MRI-only simulation use [15].
More recently, there has been interest in developing methods of automated planning in order to improve the workflow and reliability of MR-based planning [13]. One group has used a machine learning-based method to generate pseudo-CT images from pretreatment MRIs and retrospectively analyzed the dose–volume histogram metrics from these pseudo-CTs compared to their actual plans, which were made for the CT’s acquired pretreatment. They found that this model created accurate pseudo-CT images with good agreement from their simulation CT [16]. The value of pseudo-CT in MRI-PT has yet to be validated prospectively to the authors’ knowledge.
Table 1. Advantages and Disadvantages of Photon CT, MRI, and Proton CT for Simulation.
Table 1. Advantages and Disadvantages of Photon CT, MRI, and Proton CT for Simulation.
MethodAdvantagesDisadvantages
Photon CTWidespread availability and familiarity of CT simulation for therapists and physicians
Dual-energy CT provides additional clarification of soft tissue density
Risk of Contrast-induced Nephropathy
Uncertainty with the conversion from HU to proton-stopping power
MRISuperior soft tissue contrast to Photon CT
Possibility of acquiring sequences that provide additional information about tumor biology
Registration of diagnostic MRI to simulation CT is not needed
Initial unfamiliarity with performing MRI simulation
Older pacemakers may not be MRI-compatible [10]
Requires generation of pseudo-CT for patient setup and dose calculation
Proton CTDirect visualization of proton-stopping powerCurrent lack of proton CT scanners ready for clinical use
Lower-quality anatomic visualization

3. Imaging for Pretreatment Positioning

Various imaging modalities are used for pretreatment positioning (See Table 2), often in complement with one another.

3.1. 2D Kilovoltage Imaging

3.1.1. Anatomical Structure

Due to the increased sensitivity of patient positioning to dosimetry, image guidance was implemented early in the history of proton therapy [5]. Proton treatment systems have kV X-ray tubes mounted in the nozzle for a beams-eye-view setup as well as orthogonal imagers, which can be used prior to volumetric imaging. These can be moved into and out of the beam path prior to treatment. Daily images can be taken during setup compared to the digitally reconstructed radiographs (DRRs) from the time of the simulation, while the corrective moves can be performed based on the bony anatomy [17], often on a robotic 6-degree-of-freedom couch [18,19,20]. One caveat is that these can lengthen the path and increase the spot size for pencil beam scanning (PBS) gantries if downstream of the beamline vacuum window. This has been avoided in some systems by placing the X-ray beams on the robotic C-arms, behind the last gantry bending magnet, or attaching them to rooms instead of gantries [21]. One of the reasons for the early adoption of orthogonal kV radiography was the poor quality of proton radiography at the time [22]. One downside of orthogonal X-ray is that there is not as much information about possible tissue changes along the beam path as volumetric imaging, which could lead to overshoot or undershooting [22]. Moreover, since the virtual proton and XR sources are not perfectly coincident with one another, the XR tube only approximates the beam’s eye view of the proton beam [23].

3.1.2. Fiducial Markers

In addition to using bony anatomy for the daily setup, fiducial markers may also be used for an additional setup. The advantage of using fiducial markers is that they may be placed in soft tissue, in close proximity to the target, and may provide a better surrogate for the location of the target than bony anatomy. Intraprostatic fiducial markers are placed, which allows for daily orthogonal kV imaging for IGRT of the prostate with both passively scattered protons [24] and PBS proton therapy [25]. This allows for a two-step setup, where the couch adjustments are performed based on the bony anatomy first. Then, another adjustment is performed based on the fiducial markers, or if the displacement of the markers is unacceptably high, the patient is not treated but instead undergoes preparation for treatment (bladder filling and rectal emptying), before repeating the setup.
For uveal melanoma treatments, tantalum clips are placed around the tumor behind the globe in the operating room prior to simulation. In the simulation, there was immobilization with a rigid mask and bite block, and the eye is fixed using stimulus-directed light. Pretreatment, orthogonal kV X-rays are taken and aligned with those taken at the time of the simulation [26]. At Loma Linda University, patients are treated with partial breast irradiation with protons, which can present a challenge due to the inter- and intrafraction motion of the target. Patients lie prone supported by vacuum bags with a custom low-density foam around the breast and plastic breast cup, which provides reproducible placement of the breast and limits respiratory motion. The kV X-rays are taken pretreatment and the radio-opaque surgical clips are aligned with the DRR to determine any setup corrections [27].
However, the fiducial markers may not move exactly as the target tissue. Thus, fiducial markers have a significant effect on dosimetry in proton therapy compared to photon therapy. They may cause HU artifacts on the planning CTs, which could lead to an under or overshoot from the proton beams. Moreover, the markers placed near the end of the proton path may strike them and cause an underdosing of the target region. This effect may be greater with gold markers, when markers are placed closer to the end of the beam or when there are fewer treatment beams used [21]. A hydrogel fiducial marker has been developed for use with proton radiation and has shown negligible depth-dose perturbation in a phantom [28].

3.1.3. Fluoroscopy

Fluoroscopy has been used in both pre- and intrafraction to tailor the respiratory gating parameters. Intrafraction fluoroscopy can be used to gate respiratory motion [29]. One downside to fluoroscopy is the significant dose associated with this procedure [21].

3.2. 3D CT Imaging

The next step in image-guided proton therapy arrived with the use of volumetric imaging. CT can provide a 3D definition of the anatomy and greater information about the location of soft tissues than 2D kV radiography. It also allows for the possibility of adaptive planning [30]. This may be especially useful in disease sites that are prone to target shifts due to tumor volume changes and weight loss, such as head and neck tumors [31].

3.2.1. CT on Rails

Patients can be moved on the treatment couch on rails from the gantry to CT scanners in the treatment room. Having the CT scanner several meters away from the proton gantry at an angle reduces the interference with the treatment workflow and neutron exposure; however, the CT and couch are at risk of colliding with one another in this setup. Furthermore, a CT on rails avoids the image quality issues of the cone-beam CT (CBCT) mentioned below. The major drawback to this approach is the lack of imaging capability at this isocenter, with no repeat verification being performed after the setup correction. Further, the movement of the couch and CT will lead to additional time for image guidance [22]. At the Paul Scherrer Institute, patients are positioned and imaged outside the treatment room using the same CT scanner that is used in the simulation. Then, they are moved into the treatment room where verification X-rays are taken prior to treatment. This frees up time in the treatment room to allow for more time on the beam [32].

3.2.2. Cone-Beam CT

In addition to patients or CT scanners being moved for pretreatment imaging, CBCT usage is increasing for proton therapy. There are several options for the placement of the CBCT, which have been used by different facilities. It is typically placed on the gantry, and images are acquired as the gantry rotates, similar to linear accelerators [23]. For some facilities with partial gantries and limited space in the treatment room, the CBCT hardware may be installed onto the nozzle instead of the gantry, which must be retracted while not in use. Some facilities use a CBCT mounted on a robotic C-arm as well, which allows for imaging at the treatment isocenter or off of it. Due to the design, the source-to-imager distance is greater in CBCT for proton gantries vs. photon gantries, which reduces the number of scattered photons reaching the detector [22,23].
Daily pretreatment CT imaging does increase the imaging dose received by patients. Though relatively small in magnitude compared to the treatment dose [33], there is an increasing awareness that the cumulative dose may be significant, particularly in pediatric patients due to the concern for late effects and the distribution of the radiation doses [34]. This has led some to recommend the use of fiducial markers and kV X-rays instead of CBCT for daily soft tissue visualizations in pediatric patients [35]. However, one study found that the CBCT dose could be reduced by 81–98% and remains accurate when used in the setup, based on bony anatomy [36].
There is decreased accuracy with CBCT compared to diagnostic CT due to increased scatter, field-of-view (FOV) limitations, ring artifacts, and patient size. This can lead to dose errors when used for adaptive planning [21,22]; however, with correction methods, these errors can be reduced to around 1% [37]. Synthetic CTs can be created by using a neural network to correct daily CBCT imaging errors. Then, these synthetic CTs can be used for accurate dose calculations in adaptive proton therapy [38].

3.3. Visual Image Guidance

3.3.1. Marker-Based

Visual image guidance can be used in addition to planar or volumetric X-ray-based imaging or can be used alone. External markers can be placed at the relevant locations on the patient’s skin. One advantage of this approach is the high frequency of measurement, which is appealing for tracking mobile targets and gating delivery [21]. As above, the location of external markers does not always reflect the location of an internal target. Moreover, there needs to be consistency in the daily placement of the markers. This approach has already been used in patients for photon therapy, and reflective spheres have been placed on a breathing phantom for a scanning ion beam [39].

3.3.2. Surface-Based

Surface imaging consists of 3D mapping of the patient’s surface generated by a computer. This may be especially useful for targets that have a variable shape and location with respect to bony anatomy. Another advantage to this approach is that there is no need for external marker placement. At Massachusetts General Hospital, patients underwent postmastectomy radiation with spot scanning proton therapy after localization with a surface imaging system [40]. Surface imagers can also validate the operation of couch moves and the position of the nozzle [21,41].

3.4. MRI Guidance

MRI-guided photon therapy has become more popular in recent years. It has superior soft tissue contrast and no moving parts for 3D imaging, which allows for real-time motion monitoring [23]. This raises the possibility of reducing the normal tissue dose with MRI guidance. PTV margin reduction studies have been performed with MRI-guided photon SBRT and have shown a decrease in acute toxicity compared to CT-guided photon SBRT [42]. Due to these advances in photon therapy, there is an interest in the possibility of using MRI guidance for proton therapy.
MRI guidance eliminates the problem of additional ionizing radiation in contrast to CT guidance [43]. The superior soft tissue resolution of MRI over CT and the possibility of real-time imaging make it an attractive option for image guidance.
Pretreatment daily MRI guidance would allow for direct visualization of tumors not easily visualized by a CT or planar X-ray.
One use of MRI for image guidance that comes to mind, given its high resolution and that there is no additional dose for daily imaging, would be for online adaptive planning. A major challenge in the use of MRI guidance for proton therapy, which does not relate to photons, is the interaction between the magnetic fields within the MRI scanner and the charged protons, which causes deflections in the beam path. Due to the number of corrections that need to be made to place the Bragg peak at the same location for adaptive planning based on real-time MRI (changes in scan setting due to the magnetic field, changes in energy due to change in path length), pencil-beam scanning would be the only choice [44].
MRIs do not provide direct information on the stopping power, which can limit the feasibility of the online adaptive MRI-PT. Weekly offline on-treatment MRIs have been shown to improve plan quality for pediatric patients undergoing intensity-modulated proton therapy (IMPT) for CNS malignancies [45]. It has been shown that dose calculations based only on daily MRIs converted to synthetic CTs are feasible [15].
There have been several recent advances in making MRI-guided radiation therapy ready for clinical use. Proton gantries with MRI capabilities are currently under construction. Studies have been conducted that investigate the feasibility of MRI-only workflows [44]. There have been investigations into the hardware required to deliver MRI-guided PT, including magnet design, radiofrequency shielding requirements, rotation of the couch and gantry, measurement of the beam in the setting of strong magnetic fields, and correction of the beam [44]. One group has performed work to account for beam deflections from the fringe and imaging magnetic fields by using gantry angle offset and PBS nozzle skew along with patient-specific optimization within a PBS system [46].

4. Real-Time Imaging and Dose-Guided RT

4.1. PET

One of the limitations in PET-based range verification in protons is the difference between the Bragg peak and the PET signal.
In contrast to photon-based external beam radiation therapy, which can be measured with electronic portal imaging devices (EPIDs) after the beam has passed through the patient, the protons stop in the patient, making direct detection of the dose impossible. However, proton radiation produces byproducts such as positron and gamma radiation as it interacts with matter in a patient. Measurement of these byproducts in space and time allows for the location of the dose and is an area of ongoing research. Online monitoring of the dose may reduce range uncertainty and allow for further reduced margins, decreasing the dose to normal tissues even more. This may also allow for a better ability to determine when a re-simulation scan needs to be performed [47].
The challenges, which include the registration issues with the positioning errors in offline imaging and the cost and time demands for the development of the positron emission tomography (PET) systems, designed specifically for on-beam measurement of dose in vivo, have led to a limited amount of clinical use at this time [48]. A recent study [49] conducted at the National Center of Oncological Hadrontherapy (CNAO, Pavia, Italy) analyzed the in-beam PET data generated from eight patients treated with proton therapy at their center. They found that the standard deviation of interfraction PET activity profiles was 2.3–2.5 mm for patients without anatomical changes during treatment. Furthermore, large variations in PET range data correlated well with areas of anatomical change found on CT scans taken during the treatment course, suggesting that interfraction PET data can be used to detect morphological changes that may have a significant impact on dosimetry.
Positronium atoms consist of a positron bound to an electron. Roughly 40% of the positron annihilations in vivo occur via the formation of positronium atoms [50]. Para-positronium decays into two photons, while ortho-positronium decays into three photons. The rate at which these three-photon decays happen is proportional to the level of oxygen concentration. Therefore, positronium imaging may allow for the identification of hypoxic tissues [48]. A new PET detection system may allow for both range monitoring [51] as well as positronium imaging [52].

4.2. Prompt Gamma

Prompt gamma waves are produced instantaneously and have almost no interaction with the patient as they exit [21]. Detecting these gamma waves may allow for the precise location of the Bragg peak in three dimensions and the determination of the dose online in real time. Several groups are developing systems for prompt gamma detectors but there has not currently been any widespread adoption of prompt gamma detectors in patients [53]. One system is the use of a slit camera detector system, whereby prompt gamma radiation is emitted along the proton radiation path and passes through a tungsten slit collimator below the patient onto a floor-based segmented detector, which allows for spatial resolution of the prompt gamma distribution. The absolute range of protons can be determined from this distribution. This is compared to a reference range prompt gamma distribution generated from the planning CT.
One group used a slit camera detector system to validate their predictions of the proton range with dual-energy CT vs. single-energy CT, taken daily from patients undergoing proton treatment for prostate cancer [54]. They reported that the integration of prompt gamma imaging adds about 1 min per treatment field to the clinical workflow and that dual-energy CT-based predictions of the stopping power ratio agree with prompt gamma imaging more than single-energy CT-based predictions.

4.3. Ionoacoustics

The heat generated from the energy losses from the generated protons causes thermal expansion of the tissues. This leads to thermoacoustic emissions at a practically instantaneous timescale. The measurement of these waves by ultrasound, known as ionoacoustics, to map dose delivery is also an area of active investigation [55].

5. Image Guidance for Upright Treatment

A significant portion of the cost of treatment to construct a proton facility is due to the gantry that allows for different beam angles [56]. The large size of the gantry and the building required to house it makes up the majority of the cost.
One alternative is the use of a fixed horizontal beam line with a rotating chair positioning system, which may reduce the costs as it is less expensive to move the patient in the beam line vs. the beam around the patient. Previously constructed fixed beam lines are being repurposed by removing the old treatment chairs and installing new upright patient positioning systems [57], which allow for more flexibility in the treatment angles. Orthogonal kV X-rays can be received during pretreatment, similarly to receiving the treatment lying down [58].
Although CT simulations may be performed in the lying position and the images converted to a supine position for planning [59], there are significant deformations in the anatomy from a lying position than when treated upright, due to the relative change in direction of gravity. Thus, a dedicated CT scanner for upright daily image guidance is preferred for accurate treatment. Various models exist at certain proton centers and include scanners above the patient that descend for imaging alongside scanners that are mounted on columns and rotate around the patient, allowing for an isometric setup. Proton CT for daily setup is facilitated by the fixed beamline and rotating patient chair. For further reading, Volz et al. provided an excellent review of upright patient setup and guidance [60].

6. FLASH RT

The FLASH effect is defined as a decrease in radiation-induced normal tissue toxicities with dose delivery at ultra-high dose rates, compared to conventional dose rates used clinically [61]. Treatment of the first patient with FLASH-RT was delivered with electrons to a cutaneous T-cell lymphoma with favorable outcomes for tumor control and toxicity [62]. Groups are working on delivering FLASH-rate protons in clinics [63].
A recent trial has shown the feasibility of using FLASH protons in patients [64]. In this trial, the Bragg peak was placed distal to the patient, causing this beam to behave like a photon beam while in the patient, meaning that there was no sharp dose buildup or falloff.
The very short delivery times for FLASH would decrease the need for intrafraction motion management using conventional dose rates. Conversely, a precise pretreatment setup verification is required to ensure that the treatment is delivered to the intended target [65]. The slow acquisition time of CBCT would limit its utility for real-time imaging. Thus, MRI can provide information such as oxygenation and inflammation pre- and post-treatment, which may be useful for biological modeling [65]. One group has developed an inverse planning tool to optimize IMPT to pull the Bragg peak into the target volume and achieve FLASH dose rates in silico [66]. The higher dose-per-pulse rate of FLASH improves the ionoacoustic signal-to-noise ratio, which has led to an increased interest in this technology [65].

7. Conclusions

The importance of image guidance for proton therapy is clear due to the dosimetric properties of protons. Advances in image guidance in this space have generally followed advances in photon image guidance, possibly due to the much greater numbers of linear accelerators vs. proton gantries. However, as the number of proton gantries increases (119 gantries operational as of January 2023 with 37 under construction and another 29 in planning as of March 2023 [67]) and the number of patients treated with protons grows, it is important that there are continued advancements in image guidance for proton therapy so that it can fulfill its potential dosimetric benefits.

Author Contributions

Conceptualization, S.A.L. and G.Y.Y.; writing—original draft preparation, S.A.L.; writing—review and editing, S.A.L., J.M.S. and G.Y.Y.; supervision, G.Y.Y.; project administration, G.Y.Y.; funding acquisition, G.Y.Y. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported by the Del E. Webb Endowed Chair in Proton Research.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Slater, J.M.; Archambeau, J.O.; Miller, D.W.; Notarus, M.I.; Preston, W.; Slater, J.D. The proton treatment center at loma linda university medical center: Rationale for and description of its development. Int. J. Radiat. Oncol. Biol. Phys. 1992, 22, 383–389. [Google Scholar] [CrossRef] [PubMed]
  2. Paganetti, H. Range uncertainties in proton therapy and the role of monte carlo simulations. Phys. Med. Biol. 2012, 57, R99–R117. [Google Scholar] [CrossRef] [PubMed]
  3. Wohlfahrt, P.; Möhler, C.; Hietschold, V.; Menkel, S.; Greilich, S.; Krause, M.; Baumann, M.; Enghardt, W.; Richter, C. Clinical implementation of dual-energy ct for proton treatment planning on pseudo-monoenergetic CT scans. Int. J. Radiat. Oncol. 2017, 97, 427–434. [Google Scholar] [CrossRef] [PubMed]
  4. McCollough, C.H.; Leng, S.; Yu, L.; Fletcher, J.G. Dual- and multi-energy CT: Principles, technical approaches, and clinical applications. Radiology 2015, 276, 637–653. [Google Scholar] [CrossRef] [PubMed]
  5. MacKay, R.I. Image guidance for proton therapy. Clin. Oncol. R. Coll. Radiol. G. B 2018, 30, 293–298. [Google Scholar] [CrossRef]
  6. Algranati, C.; Strigari, L. Imaging strategies in proton therapy for thoracic tumors: A mini review. Front. Oncol. 2022, 12, 833364. [Google Scholar] [CrossRef]
  7. Dedes, G.; Dickmann, J.; Giacometti, V.; Rit, S.; Krah, N.; Meyer, S.; Bashkirov, V.; Schulte, R.; Johnson, R.P.; Parodi, K.; et al. The role of monte Carlo simulation in understanding the performance of proton computed tomography. Z. Für Med. Phys. 2022, 32, 23–38. [Google Scholar] [CrossRef]
  8. DeJongh, D.F.; DeJongh, E.A.; Rykalin, V.; DeFillippo, G.; Pankuch, M.; Best, A.W.; Coutrakon, G.; Duffin, K.L.; Karonis, N.T.; Ordoñez, C.E.; et al. A comparison of proton stopping power measured with proton CT and X-ray CT in fresh post-mortem porcine structures. Med. Phys. 2021, 48, 7998–8009. [Google Scholar] [CrossRef]
  9. Dedes, G.; Dickmann, J.; Niepel, K.; Wesp, P.; Johnson, R.P.; Pankuch, M.; Bashkirov, V.; Rit, S.; Volz, L.; Schulte, R.W.; et al. Experimental comparison of proton ct and dual energy X-ray CT for relative stopping power estimation in proton therapy. Phys. Med. Biol. 2019, 64, 165002. [Google Scholar] [CrossRef]
  10. Yang, B.; Yuan, J.; Cheung, K.Y.; Huang, C.-Y.; Poon, D.M.C.; Yu, S.K. Magnetic resonance-guided radiation therapy of patients with cardiovascular implantable electronic device on a 1.5 T magnetic resonance-linac. Pract. Radiat. Oncol. 2022, 12, e56–e61. [Google Scholar] [CrossRef]
  11. Schmidt, M.A.; Payne, G.S. Radiotherapy planning using MRI. Phys. Med. Biol. 2015, 60, R323. [Google Scholar] [CrossRef]
  12. Stanescu, T.; Jans, H.-S.; Pervez, N.; Stavrev, P.; Fallone, B.G. A study on the magnetic resonance imaging (MRI)-Based radiation treatment planning of intracranial lesions. Phys. Med. Biol. 2008, 53, 3579. [Google Scholar] [CrossRef]
  13. Lenkowicz, J.; Votta, C.; Nardini, M.; Quaranta, F.; Catucci, F.; Boldrini, L.; Vagni, M.; Menna, S.; Placidi, L.; Romano, A.; et al. A deep learning approach to generate synthetic CT in low field MR-guided radiotherapy for lung cases. Radiother. Oncol. 2022, 176, 31–38. [Google Scholar] [CrossRef]
  14. Guerreiro, F.; Burgos, N.; Dunlop, A.; Wong, K.; Petkar, I.; Nutting, C.; Harrington, K.; Bhide, S.; Newbold, K.; Dearnaley, D.; et al. Evaluation of a multi-atlas CT synthesis approach for MRI-Only radiotherapy treatment planning. Phys. Med. 2017, 35, 7–17. [Google Scholar] [CrossRef]
  15. Guerreiro, F.; Koivula, L.; Seravalli, E.; Janssens, G.O.; Maduro, J.H.; Brouwer, C.L.; Korevaar, E.W.; Knopf, A.C.; Korhonen, J.; Raaymakers, B.W. Feasibility of MRI-Only photon and proton dose calculations for pediatric patients with abdominal tumors. Phys. Med. Biol. 2019, 64, 055010. [Google Scholar] [CrossRef]
  16. Wang, T.; Manohar, N.; Lei, Y.; Dhabaan, A.; Shu, H.-K.; Liu, T.; Curran, W.J.; Yang, X. MRI-Based treatment planning for brain stereotactic radiosurgery: Dosimetric validation of a learning-Based pseudo-CT generation method. Med. Dosim. 2019, 44, 199–204. [Google Scholar] [CrossRef]
  17. Takagi, M.; Demizu, Y.; Fujii, O.; Terashima, K.; Niwa, Y.; Daimon, T.; Tokumaru, S.; Fuwa, N.; Hareyama, M.; Okimoto, T. Proton therapy for localized prostate cancer: Long-term results from a single-center experience. Int. J. Radiat. Oncol. Biol. Phys. 2021, 109, 964–974. [Google Scholar] [CrossRef]
  18. Forsthoefel, M.K.; Ballew, E.; Unger, K.R.; Ahn, P.H.; Rudra, S.; Pang, D.; Collins, S.P.; Dritschilo, A.; Harter, W.; Paudel, N.; et al. Early experience of the first single-room gantry mounted active scanning proton therapy system at an integrated cancer center. Front. Oncol. 2020, 10, 861. [Google Scholar] [CrossRef]
  19. Schulte, R.W. Strategies for image-guided proton therapy of cancer. Oncol. Hematol. Rev. US 2007, 1, 75. [Google Scholar] [CrossRef]
  20. Wang, N.; Ghebremedhin, A.; Patyal, B. Commissioning of a proton gantry equipped with dual X-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system. Med. Phys. 2015, 42, 2979–2991. [Google Scholar] [CrossRef]
  21. Paganetti, H. Proton Therapy Physics, Second Edition; Series in Medical Physics and Biomedical Engineering; CRC Press: Boca Raton, FL, USA, 2018; ISBN 978-1-351-85575-4. [Google Scholar]
  22. Landry, G.; Hua, C.-H. Current state and future applications of radiological image guidance for particle therapy. Med. Phys. 2018, 45, e1086–e1095. [Google Scholar] [CrossRef] [PubMed]
  23. Zhang, M.; Zou, W.; Teo, B.-K.K. Image guidance in proton therapy for lung cancer. Transl. Lung Cancer Res. 2018, 7, 160–170. [Google Scholar] [CrossRef] [PubMed]
  24. Mendenhall, N.P.; Hoppe, B.S.; Nichols, R.C.; Mendenhall, W.M.; Morris, C.G.; Li, Z.; Su, Z.; Williams, C.R.; Costa, J.; Henderson, R.H. Five-year outcomes from 3 prospective trials of image-guided proton therapy for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2014, 88, 596–602. [Google Scholar] [CrossRef] [PubMed]
  25. Kubeš, J.; Haas, A.; Vondráček, V.; Andrlík, M.; Navrátil, M.; Sláviková, S.; Vítek, P.; Dědečková, K.; Prausová, J.; Ondrová, B.; et al. Ultrahypofractionated proton radiation therapy in the treatment of low and intermediate-risk prostate cancer-5-year outcomes. Int. J. Radiat. Oncol. Biol. Phys. 2021, 110, 1090–1097. [Google Scholar] [CrossRef]
  26. Fuss, M.; Loredo, L.N.; Blacharski, P.A.; Grove, R.I.; Slater, J.D. Proton radiation therapy for medium and large choroidal melanoma: Preservation of the eye and its functionality. Int. J. Radiat. Oncol. Biol. Phys. 2001, 49, 1053–1059. [Google Scholar] [CrossRef]
  27. Wroe, A.J.; Bush, D.A.; Schulte, R.W.; Slater, J.D. Clinical immobilization techniques for proton therapy. Technol. Cancer Res. Treat. 2015, 14, 71–79. [Google Scholar] [CrossRef]
  28. Zhang, M.; Reyhan, M.; Kim, L.H. Depth dose perturbation by a hydrogel fiducial marker in a proton beam. J. Appl. Clin. Med. Phys. 2015, 16, 5090. [Google Scholar] [CrossRef]
  29. Tanaka, S.; Miyamoto, N.; Matsuo, Y.; Yoshimura, T.; Takao, S.; Matsuura, T. First experimental results of gated proton imaging using X-ray fluoroscopy to detect a fiducial marker. Phys. Med. Biol. 2021, 66, 18NT03. [Google Scholar] [CrossRef]
  30. Li, H.; Dong, L.; Bert, C.; Chang, J.; Flampouri, S.; Jee, K.-W.; Lin, L.; Moyers, M.; Mori, S.; Rottmann, J.; et al. AAPM task group report 290: Respiratory motion management for particle therapy. Med. Phys. 2022, 49, e50–e81. [Google Scholar] [CrossRef]
  31. Beddok, A.; Vela, A.; Calugaru, V.; Tessonnier, T.; Kubes, J.; Dutheil, P.; Gerard, A.; Vidal, M.; Goudjil, F.; Florescu, C.; et al. Proton therapy for head and neck squamous cell carcinomas: A review of the physical and clinical challenges. Radiother. Oncol. 2020, 147, 30–39. [Google Scholar] [CrossRef]
  32. Bolsi, A.; Lomax, A.J.; Pedroni, E.; Goitein, G.; Hug, E. Experiences at the paul scherrer institute with a remote patient positioning procedure for high-throughput proton radiation therapy. Int. J. Radiat. Oncol. 2008, 71, 1581–1590. [Google Scholar] [CrossRef]
  33. Almén, A. ICRP publication 129 radiological protection in cone beam computed tomography (CBCT). Radiat. Prot. Dosim. 2016, 171, 418–420. [Google Scholar] [CrossRef]
  34. Alcorn, S.R.; Zhou, X.C.; Bojechko, C.; Rubo, R.A.; Chen, M.J.; Dieckmann, K.; Ermoian, R.P.; Ford, E.C.; Kobyzeva, D.; MacDonald, S.M.; et al. Low-dose image-guided pediatric CNS radiation therapy: Final analysis from a prospective low-dose cone-beam CT protocol from a multinational pediatrics consortium. Technol. Cancer Res. Treat. 2020, 19, 1533033820920650. [Google Scholar] [CrossRef]
  35. Hess, C.B.; Thompson, H.M.; Benedict, S.H.; Seibert, J.A.; Wong, K.; Vaughan, A.T.; Chen, A.M. Exposure risks among children undergoing radiation therapy: Considerations in the era of image guided radiation therapy. Int. J. Radiat. Oncol. 2016, 94, 978–992. [Google Scholar] [CrossRef]
  36. Olch, A.J.; Alaei, P. How Low Can You Go? A CBCT dose reduction study. J. Appl. Clin. Med. Phys. 2021, 22, 85–89. [Google Scholar] [CrossRef]
  37. Li, H.; Hrinivich, W.T.; Chen, H.; Sheikh, K.; Ho, M.W.; Ger, R.; Liu, D.; Hales, R.K.; Voong, K.R.; Halthore, A.; et al. Evaluating proton dose and associated range uncertainty using daily cone-beam CT. Front. Oncol. 2022, 12, 830981. [Google Scholar] [CrossRef]
  38. Thummerer, A.; Seller Oria, C.; Zaffino, P.; Meijers, A.; Guterres Marmitt, G.; Wijsman, R.; Seco, J.; Langendijk, J.A.; Knopf, A.-C.; Spadea, M.F.; et al. Clinical suitability of deep learning based synthetic CTs for adaptive proton therapy of lung cancer. Med. Phys. 2021, 48, 7673–7684. [Google Scholar] [CrossRef]
  39. Fattori, G.; Saito, N.; Seregni, M.; Kaderka, R.; Pella, A.; Constantinescu, A.; Riboldi, M.; Steidl, P.; Cerveri, P.; Bert, C.; et al. Commissioning of an integrated platform for time-resolved treatment delivery in scanned ion beam therapy by means of optical motion monitoring. Technol. Cancer Res. Treat. 2014, 13, 517–528. [Google Scholar] [CrossRef]
  40. Depauw, N.; Batin, E.; Daartz, J.; Rosenfeld, A.; Adams, J.; Kooy, H.; MacDonald, S.; Lu, H.-M. A novel approach to postmastectomy radiation therapy using scanned proton beams. Int. J. Radiat. Oncol. Biol. Phys. 2015, 91, 427–434. [Google Scholar] [CrossRef]
  41. Freislederer, P.; Kügele, M.; Öllers, M.; Swinnen, A.; Sauer, T.-O.; Bert, C.; Giantsoudi, D.; Corradini, S.; Batista, V. Recent advanced in surface guided radiation Therapy. Radiat. Oncol. Lond. Engl. 2020, 15, 187. [Google Scholar] [CrossRef]
  42. Kishan, A.U.; Ma, T.M.; Lamb, J.M.; Casado, M.; Wilhalme, H.; Low, D.A.; Sheng, K.; Sharma, S.; Nickols, N.G.; Pham, J.; et al. Magnetic resonance imaging–guided vs computed tomography–guided stereotactic body radiotherapy for prostate cancer: The MIRAGE randomized clinical trial. JAMA Oncol. 2023, 9, 365–373. [Google Scholar] [CrossRef] [PubMed]
  43. Nien, H.-H.; Wang, L.-Y.; Liao, L.-J.; Lin, P.-Y.; Wu, C.-Y.; Shueng, P.-W.; Chung, C.-S.; Lo, W.-C.; Lin, S.-C.; Hsieh, C.-H. Advances in image-guided radiotherapy in the treatment of oral cavity cancer. Cancers 2022, 14, 4630. [Google Scholar] [CrossRef] [PubMed]
  44. Oborn, B.M.; Dowdell, S.; Metcalfe, P.E.; Crozier, S.; Mohan, R.; Keall, P.J. Future of medical physics: Real-time MRI-guided proton therapy. Med. Phys. 2017, 44, e77–e90. [Google Scholar] [CrossRef] [PubMed]
  45. Acharya, S.; Wang, C.; Quesada, S.; Gargone, M.A.; Ates, O.; Uh, J.; Krasin, M.J.; Merchant, T.E.; Hua, C. Adaptive proton therapy for pediatric patients: Improving the quality of the delivered plan with on-treatment MRI. Int. J. Radiat. Oncol. 2021, 109, 242–251. [Google Scholar] [CrossRef]
  46. Burigo, L.N.; Oborn, B.M. Integrated MRI-guided proton therapy planning: Accounting for the Full MRI field in a perpendicular system. Med. Phys. 2022, 49, 1853–1873. [Google Scholar] [CrossRef]
  47. Parodi, K.; Polf, J.C. In vivo range verification in particle therapy. Med. Phys. 2018, 45, e1036–e1050. [Google Scholar] [CrossRef]
  48. Parodi, K.; Yamaya, T.; Moskal, P. Experience and new prospects of PET imaging for ion beam therapy monitoring. Z. Med. Phys. 2023, 33, 22–34. [Google Scholar] [CrossRef]
  49. Moglioni, M.; Kraan, A.C.; Baroni, G.; Battistoni, G.; Belcari, N.; Berti, A.; Carra, P.; Cerello, P.; Ciocca, M.; De Gregorio, A.; et al. In-vivo range verification analysis with in-beam PET data for patients treated with proton therapy at CNAO. Front. Oncol. 2022, 12, 929949. [Google Scholar] [CrossRef]
  50. Moskal, P.; Jasińska, B.; Stępień, E.Ł.; Bass, S.D. Positronium in medicine and biology. Nat. Rev. Phys. 2019, 1, 527–529. [Google Scholar] [CrossRef]
  51. Rucinski, A.; Baran, J.; Garbacz, M.; Pawlik-Niedzwiecka, M.; Moskal, P. Plastic Scintillator Based PET Detector Technique for Proton Therapy Range Monitoring: A Monte Carlo Study. In Proceedings of the 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC), Sydney, NSW, Australia, 10–17 November 2018; IEEE: Sydney, NSW, Australia, 2018; pp. 1–4. [Google Scholar]
  52. Moskal, P.; Kisielewska, D.; Curceanu, C.; Czerwiński, E.; Dulski, K.; Gajos, A.; Gorgol, M.; Hiesmayr, B.; Jasińska, B.; Kacprzak, K.; et al. Feasibility study of the positronium imaging with the J-PET tomograph. Phys. Med. Biol. 2019, 64, 055017. [Google Scholar] [CrossRef]
  53. Wrońska, A.; for the SiFi-CC Group. Prompt Gamma Imaging in Proton Therapy—Status, Challenges and Developments. J. Phys. Conf. Ser. 2020, 1561, 012021. [Google Scholar] [CrossRef]
  54. Berthold, J.; Khamfongkhruea, C.; Petzoldt, J.; Thiele, J.; Hölscher, T.; Wohlfahrt, P.; Peters, N.; Jost, A.; Hofmann, C.; Janssens, G.; et al. First-in-human validation of CT-based proton range prediction using prompt gamma imaging in prostate cancer treatments. Int. J. Radiat. Oncol. 2021, 111, 1033–1043. [Google Scholar] [CrossRef]
  55. Mast, T.D.; Johnstone, D.A.; Dumoulin, C.L.; Lamba, M.A.; Patch, S.K. Reconstruction of thermoacoustic emission sources induced by proton irradiation using numerical time reversal. Phys. Med. Biol. 2023, 68, 025003. [Google Scholar] [CrossRef]
  56. Yan, S.; Lu, H.-M.; Flanz, J.; Adams, J.; Trofimov, A.; Bortfeld, T. Reassessment of the necessity of the proton gantry: Analysis of beam orientations from 4332 treatments at the massachusetts general hospital proton center over the past 10 years. Int. J. Radiat. Oncol. Biol. Phys. 2016, 95, 224–233. [Google Scholar] [CrossRef]
  57. Unique Partnership with Hampton University Proton Therapy Institution Is Set to Deliver Advances in Proton Arc Therapy Research. Available online: https://www.leocancercare.com/blog/unique-partnership-with-hampton-university-proton-therapy-institution-is-set-to-deliver-advances-in-proton-arc-therapy-research (accessed on 28 March 2023).
  58. Sun, J.; Kong, L.; Chen, Z.; You, D.; Mao, J.; Guan, X.; Wu, X.; Sheng, Y. Clinical implementation of a 6D treatment chair for fixed ion beam lines. Front. Oncol. 2021, 11, 694749. [Google Scholar] [CrossRef]
  59. Sheng, Y.; Sun, J.; Wang, W.; Stuart, B.; Kong, L.; Gao, J.; You, D.; Wu, X. Performance of a 6D treatment chair for patient positioning in an upright posture for fixed ion beam lines. Front. Oncol. 2020, 10, 122. [Google Scholar] [CrossRef]
  60. Volz, L.; Sheng, Y.; Durante, M.; Graeff, C. Considerations for upright particle therapy patient positioning and associated image guidance. Front. Oncol. 2022, 12, 930850. [Google Scholar] [CrossRef]
  61. Hughes, J.R.; Parsons, J.L. FLASH radiotherapy: Current knowledge and future insights using proton-beam therapy. Int. J. Mol. Sci. 2020, 21, 6492. [Google Scholar] [CrossRef]
  62. Bourhis, J.; Sozzi, W.J.; Jorge, P.G.; Gaide, O.; Bailat, C.; Duclos, F.; Patin, D.; Ozsahin, M.; Bochud, F.; Germond, J.-F.; et al. Treatment of a first patient with FLASH-radiotherapy. Radiother. Oncol. J. Eur. Soc. Ther. Radiol. Oncol. 2019, 139, 18–22. [Google Scholar] [CrossRef]
  63. Van Marlen, P.; Dahele, M.; Folkerts, M.; Abel, E.; Slotman, B.J.; Verbakel, W.F.A.R. Bringing FLASH to the clinic: Treatment planning considerations for ultrahigh dose-rate proton beams. Int. J. Radiat. Oncol. Biol. Phys. 2020, 106, 621–629. [Google Scholar] [CrossRef]
  64. Mascia, A.E.; Daugherty, E.C.; Zhang, Y.; Lee, E.; Xiao, Z.; Sertorio, M.; Woo, J.; Backus, L.R.; McDonald, J.M.; McCann, C.; et al. Proton FLASH radiotherapy for the treatment of symptomatic bone metastases: The FAST-01 nonrandomized trial. JAMA Oncol. 2023, 9, 62–69. [Google Scholar] [CrossRef] [PubMed]
  65. El Naqa, I.; Pogue, B.W.; Zhang, R.; Oraiqat, I.; Parodi, K. Image guidance for FLASH radiotherapy. Med. Phys. 2022, 49, 4109–4122. [Google Scholar] [CrossRef] [PubMed]
  66. Kang, M.; Wei, S.; Choi, J.I.; Lin, H.; Simone, C.B. A universal range shifter and range compensator can enable proton pencil beam scanning single-energy bragg peak FLASH-RT treatment using current commercially available proton systems. Int. J. Radiat. Oncol. 2022, 113, 203–213. [Google Scholar] [CrossRef] [PubMed]
  67. PTCOG—Facilities in Planning Stage. Available online: https://www.ptcog.ch/index.php/facilities-in-planning-stage (accessed on 22 March 2023).
Table 2. Methods of Pretreatment Image Guidance.
Table 2. Methods of Pretreatment Image Guidance.
TypeMethodAdvantagesDisadvantages
kV planar XRkV X-rays are taken from the nozzle and compared to DRRs acquired at time of simulation or fiducial markersSimple
Able to visualize bony anatomy or radio-opaque visual markers
Inferior soft tissue contrast
May not allow for direct visualization of the target
Limited information on changes in tissue proximal to target
CT on RailsPatients are imaged at a separate CT scanner before being moved on rails to the gantry couch, which has been registered to the CT scanner couchAvoidance of imaging artifacts associated with CBCT
Preferred in treatment rooms that cannot fit the CBCT system
Reduces time that gantry is used for imaging
Additional ionizing radiation
Possible additional movement if moving patient from CT scanner to gantry
Additional time to move couch from CT to gantry
No possibility of real-time guidance
CBCTCBCT is placed on a gantry, nozzle, or robotic C-arm which rotates around the patient on the gantry couchAdditional soft tissue visualization compared to planar kV XR
Allows for adaptive planning
Additional ionizing radiation
Imaging artifacts
No possibility of real-time guidance
MRIPatient lies in a specially constructed MRI scanner which allows for proton beam delivery either in-line or perpendicular to magnetic fieldNo additional ionizing radiation
Superior soft tissue contrast
Possibility of real-time guidance
Not currently available clinically
Complex deflections of proton beam and dose due to magnetic field
Older pacemakers may not be MRI-compatible
Surface TrackingSystems of visual cameras are installed in the gantry rooms and monitor moves in markers placed on the patient’s surface or the patient’s skin.No additional ionizing radiation
Real-time monitoring
Pre- and intrafraction patient monitoring
Can be complementary to CBCT
Verification of nozzle setup
Surface movement does not always correspond to target movement
Decreased accuracy in area of flat patient surface
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Lane, S.A.; Slater, J.M.; Yang, G.Y. Image-Guided Proton Therapy: A Comprehensive Review. Cancers 2023, 15, 2555. https://doi.org/10.3390/cancers15092555

AMA Style

Lane SA, Slater JM, Yang GY. Image-Guided Proton Therapy: A Comprehensive Review. Cancers. 2023; 15(9):2555. https://doi.org/10.3390/cancers15092555

Chicago/Turabian Style

Lane, Shelby A., Jason M. Slater, and Gary Y. Yang. 2023. "Image-Guided Proton Therapy: A Comprehensive Review" Cancers 15, no. 9: 2555. https://doi.org/10.3390/cancers15092555

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop