Next Article in Journal
Electronic and Optical Properties of Transition-Metal-Modified BiFeO3: A First Principles Study
Previous Article in Journal
Impact of Elevated Temperatures on the Integrity of PIR-Core Roof Sandwich Panel Connections
Previous Article in Special Issue
Corticotomy Depth as a Modulator of Orthodontic Tooth Movement and PDL Stress—A Finite Element Study
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
This is an early access version, the complete PDF, HTML, and XML versions will be available soon.
Article

FEA Simulation of Crimping Pressure Distribution in Titanium and Teflon Stapedotomy Prostheses

1
Department of Mechanics, Mathematics and Management, Polytechnic of Bari, 70125 Bari, Italy
2
Department of Translational Biomedicine and Neuroscience, University of Bari Aldo Moro, 70125 Bari, Italy
3
UOC Otolaryngology, University AOUC Policlinico di Bari, 70125 Bari, Italy
*
Author to whom correspondence should be addressed.
Materials 2026, 19(1), 65; https://doi.org/10.3390/ma19010065 (registering DOI)
Submission received: 19 November 2025 / Revised: 11 December 2025 / Accepted: 19 December 2025 / Published: 23 December 2025

Abstract

Stapedotomy is performed to restore ossicular chain sound transmission by inserting a piston prosthesis that couples the long process of the incus to the oval window, thereby addressing conductive hearing loss associated with otosclerosis. This study investigates the effects of crimping force, prosthesis material, and loop geometry on incus to optimize fixation while minimizing complications such as incudal necrosis. Finite element analyses were performed to quantify interface pressures and von Mises stresses for titanium prostheses with loop-band widths of 0.2, 0.3, and 0.5 mm under crimping forces of 300–500 mN and for polytetrafluoroethylene (PTFE) prostheses with loop outer diameters (OD) of 1.2, 1.4, and 1.8 mm. The analysis results showed that PTFE prostheses generated significantly lower interface pressures and stress compared to titanium. For PTFE prostheses, the equivalent von Mises stresses remained well below the critical threshold, with values ranging from 3.5 MPa up to peaks of approximately 43 MPa depending on the loop’s outer diameter. In contrast, titanium prostheses exhibited a marked dependency on crimping force and band width. At a force of 300 mN, stresses were modest (approximately 16–24 MPa). However, when increasing the force to 400 mN, stresses approached the critical threshold (up to approximately 53 MPa). With crimping forces of 500 mN, especially with band widths greater than 0.3 mm, stresses exceeded the cortical bone strength threshold (approximately 61–64 MPa), indicating an increased risk of mechanical overload and potential incudal necrosis. These findings highlight the importance, in a clinical context, of controlling the crimping force and selecting the material and geometry of the prosthesis to achieve secure coupling while preserving the incus’s structural integrity.
Keywords: stapedotomy; crimping force; FEA; Teflon; titanium stapedotomy; crimping force; FEA; Teflon; titanium

Share and Cite

MDPI and ACS Style

Ceddia, M.; Quaranta, N.; Pontillo, V.; Murri, A.; Pantaleo, A.; Trentadue, B. FEA Simulation of Crimping Pressure Distribution in Titanium and Teflon Stapedotomy Prostheses. Materials 2026, 19, 65. https://doi.org/10.3390/ma19010065

AMA Style

Ceddia M, Quaranta N, Pontillo V, Murri A, Pantaleo A, Trentadue B. FEA Simulation of Crimping Pressure Distribution in Titanium and Teflon Stapedotomy Prostheses. Materials. 2026; 19(1):65. https://doi.org/10.3390/ma19010065

Chicago/Turabian Style

Ceddia, Mario, Nicola Quaranta, Vito Pontillo, Alessandra Murri, Alessandra Pantaleo, and Bartolomeo Trentadue. 2026. "FEA Simulation of Crimping Pressure Distribution in Titanium and Teflon Stapedotomy Prostheses" Materials 19, no. 1: 65. https://doi.org/10.3390/ma19010065

APA Style

Ceddia, M., Quaranta, N., Pontillo, V., Murri, A., Pantaleo, A., & Trentadue, B. (2026). FEA Simulation of Crimping Pressure Distribution in Titanium and Teflon Stapedotomy Prostheses. Materials, 19(1), 65. https://doi.org/10.3390/ma19010065

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop