Next Article in Journal
Predicting Continued Participation in Government-Initiated Public Sports Clubs: The Role of Prior Knowledge and Awareness through Health Communications
Next Article in Special Issue
Application of Olfactometry to Assess the Anti-Odor Properties of Filtering Facepiece Respirators Containing Activated Carbon Nonwovens
Previous Article in Journal
“You Need to Get Over the Difficulties and Stand Up Again”—A Qualitative Inquiry into Young Nurses’ Coping with Lateral Violence from the Feminist Perspective
Previous Article in Special Issue
Development of a Dust Respirator by Improving the Half Mask Frame Design
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

In-House Filtration Efficiency Assessment of Vapor Hydrogen Peroxide Decontaminated Filtering Facepiece Respirators (FFRs)

1
Radiation Biology Section, Biomedical Physics Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
2
Emerging Technology Unit, Biomedical Physics Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
3
Perioperative Services Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
4
Protocol Service Nursing Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
5
Quality Management Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
6
Infection Control and Hospital Epidemiology Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
7
General Corporate Consultancy Department, King Faisal Specialist Hospital and Research Centre (KFSH&RC), Riyadh 11211, Saudi Arabia
*
Author to whom correspondence should be addressed.
Int. J. Environ. Res. Public Health 2021, 18(13), 7169; https://doi.org/10.3390/ijerph18137169
Submission received: 25 April 2021 / Revised: 28 June 2021 / Accepted: 1 July 2021 / Published: 4 July 2021
(This article belongs to the Collection Occupational Safety and Personal Protective Equipment)

Abstract

:
To cope with the shortage of filtering facepiece respirators (FFRs) caused by the coronavirus disease (COVID-19), healthcare institutions have been forced to reuse FFRs using different decontamination methods, including vapor hydrogen peroxide (VHP). However, most healthcare institutions still struggle with evaluating the effect of VHP on filtration efficiency (FE) of the decontaminated FFRs. We developed a low-cost in-house FE assessment using a novel 3D-printed air duct. Furthermore, we assessed the FE of seven types of FFRs. Following 10 VHP cycles, we evaluated the FE of KN95 and 3M-N95 masks. The 3M-N95 and Benehal-N95 masks showed significant lower FE (80.4–91.8%) at fine particle sizes (0.3–1 µm) compared to other FFRs (FE ≥ 98.1%, p < 0.05). Following 10 VHP cycles, the FE of KN95 masks was almost stable (FE stability > 99.1%) for all particle sizes, while 3M-N95 masks were stable only at 2 and 5 µm (FE stability > 98.0%). Statistically, FE stability of 3M-N95 masks at 0.3, 0.5, and 0.7 µm was significantly lower (p ≤ 0.006) than 2 and 5 µm. The in-house FE assessment may be used as an emergency procedure to validate the decontaminated FFRs, as well as a screening option for production control of FFRs. Following VHP cycles, both masks showed high stability at 5 µm, the size of the most suspected droplets implicated in COVID-19 transmission.

1. Introduction

The rapid increase in novel coronavirus disease (COVID-19) patients caused a dramatic global shortage of filtering facepiece respirators (FFRs), in particular, N95 masks [1,2]. The N95 mask is a single-use respirator that can, when fitted correctly, block up to 95% of as small as 0.3 µm or larger particles. The N95 masks are approved by the US Food and Drug Administration (FDA) and by the USA National Institute for Occupational Safety and Health (NIOSH) [3,4]. The FFRs, including N95 masks, are considered a critical component of infection prevention and control by decreasing COVID-19 transmission from patients to health care professionals. However, N95 masks are designed for single usage, and in cases of infectious agents, it may also be used for a single patient [5,6]. Therefore, under the massive demand for FFRs during the COVID-19 pandemic, NIOSH issued a recommended guidance for the extended use and limited reuse of N95 respirators in an emergency health care setting [6]. Although the NIOSH guidance could significantly reduce the consumption of N95 masks during the pandemic, two studies were concerned with safety of the extended use of N95 FFRs [7,8]. These concerns were related to the long survival of the SARS-CoV-2 virus on the outer layer of N95 masks, resulting in cross-contamination through direct contact between staff or indirectly by touching contaminated surfaces [7,8]. Therefore, these decontamination methods may be needed for future pandemics.
There are more than five well-characterized N95 decontamination procedures, including vapor hydrogen peroxide (VHP), moist heat incubation, microwave oven, ultraviolet germicidal irradiation (UVGI), gamma radiation, and ethylene oxide [9,10,11,12,13]. Although the USA Centers for Disease Control (CDC) has not approved the routine decontamination of N95 masks, it released emergency guidelines on the N95 decontamination methods, indicating that UVGI, moist heat, and VHP have shown the most promising results [10].
There are advantages and disadvantages to each of the three previous decontamination procedures [14,15,16]. The UVGI and moist heat procedures are ubiquitous, inexpensive, accessible, with high throughput and ease of use and no chemical residues left after the decontamination procedures [15]. However, concerns were raised by a number of researchers regarding the ability of UVGI to sterilize throughout the thickness of the FFRs, and thermal deformation and shadowing was observed on the FFRs following UVGI treatment [15,17]. The moist heat decontamination procedure, either by using moist heat incubation or autoclave, seemed to cause adverse effects on filtration efficiency (FE) and airflow of FFRs [16]. Several studies showed that the VHP decontamination procedure had no adverse effect on either FE, airflow, physical characteristic, or fit test of FFRs, with high germicidal efficacy [14,18]. Therefore, the VHP decontaminant procedure was the most interesting procedure to be locally applied among other decontaminant procedures.
In response to the shortage of FFRs, the FDA has issued three emergency use authorizations (EUA) for the emergency use of VHP to decontaminate FFRs using STERIS, Stryker and STERRAD systems [19,20,21]. The STERIS sterilization systems can be applied up to 10 times for each N95 mask using one of five models of STERIS sterilizers (V-PRO 1 Plus, maX, maX2, 60, and V-PRO s2) [22]. By using the NIOSH test procedure to measure FE [23], the 3M company showed that the VHP-STERIS decontamination method does not affect the FE in seven models of 3M-N95 masks; 1860, 1860S, 1870+, 8110S, 8210, 9205+, and 9210+ [24].
The STERIS sterilization system was selected among the other two systems to be applied in this study for two reasons. The first was the availability of four STERIS sterilizer units (STERIS V-PRO maX model) compared to a single STERRAD sterilizer unit (100NX model) at the perioperative services department in our institution, while the Stryker system was unavailable. The second reason was the limited number of approved-VHP cycles (two cycles maximum) obtained using the STERRAD sterilization system compared to the STERIS sterilization systems (10 cycles maximum) [19,21].
Although the VHP-STERIS decontamination is an FDA–EUA-approved method, in-house FE validation of the decontaminated N95 masks is needed as a quality assurance process. Currently, most healthcare institutions are struggling with how to validate the FE of decontaminated N95, as the NIOSH test procedure is not widely available, in addition to its high cost, time-consumption, and complexity [25]. Therefore, a number of healthcare institutions have developed in-house methods to evaluate the FE of decontaminated masks using the available equipment and techniques as alternatives to the NIOSH method [9,26,27].
In May 2020, during the peak of COVID-19 pandemic in Saudi Arabia, we successfully implemented the STERIS sterilization systems to decontaminate N95 masks as a crisis capacity strategy. The current study aimed to develop a simple, quick, and low-cost in-house FE assessment procedure that can be used as a screening option for production control of FFRs andas a quality assurance process, of decontaminated FFRs. Furthermore, the study aimed to assess the FE of a number of FFRs, some of which were N95 masks, and evaluate the effect of VHP-STERIS decontamination method on FE of N95 and KN95 masks.

2. Materials and Methods

2.1. Filtering Facepiece Respirators (FFRs)

The FFRs used in this study are described in Table 1. Briefly, seven types of FFR masks were used. A set of five each of Gerson 1730, Medline, Benehal, N99/N95 SpectraShield Plus, and KN95 (duck shape) masks were included. In addition, a set of 12 each of 3M-N95 (3M-8210) and KN95 (molded shape) masks were also tested.

2.2. Filtration Efficiency (FE) Measurement

The purpose of using 3D-printing technology was to create a specialized air duct that could be used to assess FE of different mask types with no air leak during the FE assessment. The 3D-printing technology was employed to produce each part of the air duct with a precise shape to serve specific functions. Although such designs can be manufactured in mechanical engineering facilities, 3D-printing was the most accessible and available option during the COVID-19 lockdown.
As shown in Figure 1, a novel specialized air duct was designed using 3D-printing technology to measure the FE of different mask types.
The 3D digital design, in STereoLithography (STL) format, of the air duct is available in Supplementary Materials 1. Briefly, the air duct was made by Prusa 3D-printer (model MK3S, product ID: PRI-MK3S-COM-ORG-PEI) using Polylactic Acid (PLA) filaments (Prusa Research, Prague, Czech Republic, product ID: FLM-PLA-175-YEL) and designed by two softwares; Shapr3D modelling software (https://www.shapr3d.com/ accessed on 3 July 2021) and Slic3R slicing software (https://slic3r.org/ accessed on 3 July 2021), used to generate the G code required to print the model by the Prusa printer with a dimension of 19-cm-long, 14-cm-wide, and 12-cm-high. The air duct consisted of two parts, head and tail, that could be tightly joined together by three mold bolts to squeeze a mask in a sandwich manner. Several air duct designs were fabricated and tested to ensure no air leak was present between both parts. The source of the measured particles was the ambient particulate matter (PM) air. An electrical fan flowed the PM air through the air duct tail. The fan was controlled by a variable transformer (model: 3PF1010, Staco Energy Production Company, Miamisburg, OH, USA) fixed at 47 volts to give a face velocity of 0.4 m/s; measured by Velocicalc Air Velocity Meter 9545 (TSI, Shoreview, MN, USA, product ID# 9545-A). The head of the air duct was designed to fully contain a stainless steel part of an AeroTrak particle counter (TSI, Model 9306, product ID: SKU 9306-03). The AeroTrak particle counter is an optical-based particle counter that measures the particle number concentration at six different particle sizes (0.3, 0.5, 0.7, 1, 2, and 5 µm) using light scattering, a method that is based upon the amount of light deflected by a particle passing through the detection area of the counter [28]. The AeroTrak particle counter was calibrated for each use by Purge (Zero Count) Filter (TSI, product ID# 700005). The air particles entered the air duct from the tail to the head and penetrated the squeezed mask at a flow rate of 2.8 L/min. The particle number concentration of the PM air was assessed at least five times before FFR assessment. For each particle size, FE of each mask was assessed for 1 min and calculated using the following formula:
FE   ( % ) = 100 ( number   of   penetrated   particles   average   number   of   particles   in   air × 100 )
Although the NIOSH’s standard sampling time is 10 min, the 1-min sampling time was selected based on an early pilot experiment performed in our laboratory. The experiment showed no overall significant difference between FE sampling times of 1 and 10 min of either KN95 (n = 5) or N95-3M (n = 5) FFRs with p-values of 0.757 and 0.699, respectively (Supplementary Materials 2). Therefore, the 1-min sampling time was selected since it is more suitable for such an in-house method, especially with a large number of FFRs and VHP cycles, as shown later.

2.3. Filtration Efficiency (FE) Following VHP Cycles

For FE evaluation, 12 masks of each 3M-N95 and KN95 were subjected to 10 VHP cycles using STERIS sterilization systems (Mentor, OH, USA). The full protocol of VHP-STERIS sterilization systems can be found on the STERIS website [29]. Briefly, at the FFRs collection station, the used FFRs were carefully labelled and pouched in a VHP sterilization pouch (Tyvek, Product ID#12340). Before sealing the pouch, a VHP indicator strip (Sterrad, Irvine, CA, USA, product ID#14100) was inserted in the pouch to confirm the masks’ VHP exposure. A maximum of 10 pouches could be processed per each non-lumen cycle (28 min for each cycle) using the STERIS V-PRO maX sterilizers (STERIS).
The FE assessment of decontaminated FFRs was conducted 10 times on 10 different days for 10 VHP cycles; therefore, the PM air particle number concentration was different for each day of the experiment. To minimize these differences, we conducted the 10 FE assessments at the same place using the same particle counter. Overall, these minor differences were relatively small and statistically considered. After each VHP cycle, the stability of FE for each particle size was calculated using the following formula:
S t a b i l i t y   o f   F E   ( % ) = ( F E   a t   c y c l e   X   F E   o f   c o n t r o l   ( c y c l e   0 ) × 100 )

2.4. Statistical Analysis

To examine statistical differences in FE and FE stability of FFRs for different particle sizes and over multiple VHP decontamination cycles, the parametric one-way pepeated measures analysis of variance test was used to detect significant differences in the means values for each set of data. In case the data did not pass the Shapiro–Wilk normality test, the nonparametric Friedman repeated measures analysis of variance on ranks test, which compares medians, was applied. The percent uncertainty of the particle counter at each particle size was calculated by dividing the absolute uncertainty (standard error) by the average of measurements multiplying by 100. All the statistical analyses were performed using SigmaPlot 12.5 for Windows (SPSS Inc., Chicago, IL, USA).

3. Results

3.1. Filtration Efficiency (FE) of FFRs

As shown in Figure 2, KN95 (molded shape), Gerson 1730, Medline N95, N99/N95 SpectraShield, and KN95 (duck shape) masks showed an FE of 98.1% or above at all particle sizes. At 2 and 5 µm particle diameters, all FFRs, except Benehal-N95, showed an FE of 95% or above. The 3M-N95 and Benehal-N95 masks were less efficient (FE = 80.4–91.8%), particularly at low particle diameters (0.3–1 µm), than other FFRs (FE = 98.6–99.9%). Statistically, the overall FE values of FFRs failed to pass the normality test, therefore, the median FE values were used for statistical analysis. The Friedman repeated measures analysis of variance on ranks showed that the overall median FE values for the 3M-N95 and Benehal-N95 masks were significantly different from the N99/N95 SpectraShield mask (p < 0.05, respectively) with no other significant differences observed.

3.2. Filtration Efficiency (FE) Stability of N95 and KN95 Masks Following VHP-STERIS Sterilization Systems

The FE stability of the two FFRs (KN95 and 3M-N95) masks was studied for up to 10 VHP-STERIS sterilization cycles (Figure 3). Results showed that for all sizes of particle diameters, the FE of KN95 masks was almost stable (FE stability > 99.1%) over 10 VHP cycles. However, FE of 3M-N95 masks was stable only at particle diameters of 2 and 5 µm (FE stability > 98.0%), while it showed wide variation in FE stability (73.9–97.0%) at particle diameters between 0.3 and 1 µm. Statistically, the 3M-N95 masks showed an overall significant (p < 0.001) reduction in FE stability with increased VHP cycles using the Friedman repeated measures analysis of variance on ranks test. In particular, FE stability of 3M-N95 masks at 0.3, 0.5, and 0.7 µm over 10 VHP cycles was significantly lower than FE stability at 5 µm with p values of <0.001, <0.001, and 0.006, respectively. Similarly, FE stability of 3M-N95 masks at 0.3 and 0.5 µm over 10 VHP cycles was significantly lower than FE stability at 2 µm with a p-values of 0.002 for both.

4. Discussion

The procedure of decontamination and subsequent reuse of N95 masks is recommended to be applied only as a crisis capacity strategy with the objective of making healthcare institutions self-sufficient for N95 practice [6]. This study provided a simple, quick, and low-cost in-house method for the FE assessment of FFRs. Furthermore, the study evaluated the FE of different FFR types pre- and post-VHP decontamination procedure.
The parameter comparison between the in-house FE assessment method and NIOSH FE assessment method is presented in Table 2 [23].
The in-house FE assessment was developed as an emergency alternative to NIOSH FE method to reduce the SARS-Cov-2 spread. In the in-house FE assessment, the FFRs were exposed to the (PM) air rather than artificial aerosol clouds. Furthermore, the in-house FE assessment does not assess very small particles (<0.3 µm), which are way below the sizes (2–5 µm) suspected of transmitting viruses via expelled droplets during speaking and coughing [30,31]. Although the airflow rate of the in-house FE assessment (2.8 L/min) is lower than NIOSH FE method (85 L/min), it is more compatible with the exhaled-breath flow rate for healthy people (5–10 L/min) than the NIOSH method [32]. The AeroTrack particle counter used in this study (Figure 1) has been used in several studies to assess FE of FFRs as a low-cost alternative to the TSI automated filter tester [9,26,27].
The cost of the in-house FE assessment is insignificant compared to the approved assessments for two reasons. The first reason is the low cost of the recent handheld particle counters such as the AeroTrak (TSI 9306) particle counter or its equivalent compared to the stationary particle counter. Briefly, the high cost of the stationary particle counter has been a barrier to long-term PM air monitoring and studies [33]. However, a recent generation of low-cost handheld particle counters showed high monitoring of ambient air particles in indoor and outdoor environments with high spatiotemporal resolution [34]. The second reason is the availability of handheld particle counters in most hospitals as they are widely used to assess indoor air quality either in the operation theatre or patients’ rooms [35,36]. Therefore, the handled particle counters can be employed for FE assessment of decontaminated FFRS in a pandemic with no extra cost. In terms of 3D-printing cost, the total approximate cost of the filaments used to fabricate the 3D-printed air duct is less than $10, as the 3D-printer (~$750) and the slicing software ($20) are available.
Results showed that five out of seven FFRs retain an FE capacity of >95% at particle sizes of 0.3 µm or larger (Figure 2). Only two types of FFRs (3M-N95 and Benehal) displayed statistically significant (p < 0.05) reduced FE (80.4–91.8%), particularly at fine particle sizes (0.3–1 µm), compared to other FFRs (FE = 98.6–99.9). In agreement with our findings, a recent study showed that 3M-N95 masks displayed an FE of 88.1, 91.4, and 89.2% at particle sizes of 0.3, 0.5, and 1 µm, respectively [27]. Another study also showed that a similar type of 3M-N95 mask (3M-1801) has an FE of ~85% at 0.3 µm [37].
The size of the SARS-CoV-2 virus is between 0.07 µm to 0.09 µm; however, it is transmitted through respiratory droplets (>5 µm) that are released from infected patients during coughing, sneezing, or speaking and then transmitted to another healthy person by fomite transmission [38,39,40]. Although airborne SARS-CoV-2 transmission, caused by the transmission of droplet nuclei particles (<5 µm), was not implicated by the World Health Organization to be the main transmission route of SARS-CoV-2, recent reports indicated that droplet nuclei particles might play a more prominent role in COVID-19 infection, in particular, during medical procedures that generate aerosols [41,42,43,44]. A recent report using data from a number of experimental studies and theoretical models showed that the minimum size of droplets that can contain and then transmit SARS-CoV-2 was calculated to be approximately 4.7 µm [31]. Therefore, the current findings may not, hypothetically, affect the ability of the two FFRs (3M-N95 and Benehal) to prevent SARS-CoV-2 transmission; further research is needed.
While the FE of KN95 mask was stable over 10 VHP cycles, the 3M-N95 was only stable at the highest (2 and 5 µm) particle diameters (Figure 3). The 3M-N95 appeared somewhat unstable for smaller particles (<2 µm) with increasing VHP cycles. A recent study showed that the FE of KN95 and another model of 3M mask (3M-1860) were stable after a single VHP cycle at a particle size of 1 µm, which is compatible with our findings [45]. Another study also observed a stable FE (only 0.4% reduction in FE) of 3M-1860 N95 following a single STERIS-VHP cycle using the NIOSH assessment method to evaluate FE [46]. Furthermore, four reports confirmed the latter studies and showed that 3 to 10 VHP cycles did not affect the FE stability of different types of 3M-N95 masks [47,48,49,50].

4.1. Uncertainties and Repeatability of the Experimental Measurements

The percent uncertainty of the particle counter (AeroTrak TSI 9306), given at one standard deviation, was calculated based on the standard error for 10 measurements of the ambient air assessed for each batch of particle size. These 10 measurements were obtained from the control measurements of the third VHP cycle, and they were selected due to the high number of control measurements assessed at the same time on the same day. As shown in Table 3, the percent uncertainties of our particle counter were acceptable and comparable to the reference uncertainties of the manufacturer calibration at all particle sizes [28,51]. The uncertainties related to other physical parameters, such as pressure, flow rate, and measurement time, could be assumed as negligible as they were included in the particle counter parameters and considered in the overall uncertainties.
The standard error was used to evaluate the repeatability of the experimental measurements in our study. In Figure 2, the highest standard error was 3.6, observed for the average FE assessment (87.1%) of Benhal N95 masks at a particle size of 0.3 µm (Supplementary Materials 3). In Figure 3 (left panel), the highest standard error was 0.4, observed for the average FE assessment (99.6%) of KN95 masks at a particle size of 0.7 µm, while in the right panel, the highest standard error was 4.9, observed for the average FE assessment (87.1%) of N95-1820 masks at a particle size of 0.5 µm (Supplementary Materials 3 and 4).

4.2. Study Limitations

The study has limitations. Only two types of FFRs and one VHP sterilization system have been tested. The in-house FE assessment has less sensitive parameters than the NIOSH FE method regarding airflow, pressure, and sampling time. Furthermore, the NIOSH method used artificial aerosol clouds to assess FE, which seem more sensitive than ambient particulate matter air.

5. Conclusions

Although the in-house FE assessment seems less restrictive than the NIOSH FE method, it may have a potential benefit as an emergency procedure to validate the FE of decontaminated FFRs where the NIOSH method is unavailable. Furthermore, the in-house FE assessment can be used as a screening option for production control of FFRs as it is faster and cheaper than the standardized procedures.The KN95 mask showed higher FE stability than 3M-N95 masks at small particle sizes (0.3–2 µm) following VHP cycles. However, both masks showed high FE stability at 5 µm, the size of the most suspected droplets implicated in COVID-19 transmission; however, further research is needed. For future research, monitoring the potential changes in the physical properties of decontaminated FFRs, such as size, color, solidity, and smell, should be considered [52]. Furthermore, the electrostatic charge analysis in the filtration layer within FFRs following VHP cycles should be considered a new aspect of FFR evaluation in further research [53]. Additionally, the 3D-printed specialized air duct could be used to evaluate other FFR decontamination methods and the efficiency of non-medical masks such as surgical and cloth masks.

Supplementary Materials

The following are available online at https://www.mdpi.com/article/10.3390/ijerph18137169/s1, Supplementary Materials 1: The 3D digital design of the air duct is in STL format (the file is submitted along with the manuscript in a single file). Supplementary Materials 2. Filtration efficiency (FE) comparison between 1 and 10 min sampling times for KN95 (n = 5) and 3M-N95 (n = 5) FFRs. Supplementary Materials 3: The raw data of filtration efficiency (FE) of FFRs. Supplementary Materials 4: The raw data of filtration efficiency (FE) stability of 3M-N95 and KN95 masks following VHP-STERIS sterilization systems. Supplementary Materials 5: The raw data of experimental uncertainties analysis.

Author Contributions

Conceptualization, K.A.-H., G.A., S.A.-G., B.M., and R.A.; methodology, K.A.-H., A.N., J.L., S.F., and N.F.; investigation, K.A.-H., G.A., and B.M.; analysis, K.A.-H. and G.A.; resources and consultation, R.A. and B.M.; writing—original draft preparation, K.A.-H.; writing—review and editing, G.A. and B.M. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by King Abdulaziz City for Science and Technology (KACST) via The Fast Track Funding Path for Coronavirus Program (project number 5-20-01-558-0005; RAC# 2200047).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The raw data supporting the conclusions of this study will be made available by the authors upon request, without undue reservation.

Acknowledgments

We acknowledge the generous funding of King Abdulaziz City for Science and Technology (KACST). We would like to thank Mehenna Arib, Najla Al-Harbi, Sara Bin Judia, Maha Al-Ghamdi, Rachelle Labbao, Amr El-Sayed, Fatima Mohammad, and Ferdinand Comia for their technical assistance and administrative support. We are indebted to Hanan Al-Saif and her team in the Change Management Division for supporting this project. We acknowledge the support of KFSH&RC administration and cooperation of the Clinical Engineering Department, Research Centre Logistics and Supply Chain Management for their logistic support.

Conflicts of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Abbreviations

CDCthe USA Centers for Disease Control
COVID-19the novel coronavirus disease
EUAemergency use authorization
FDAthe US Food and Drug Administration
FFRsfiltering facepiece respirators
FEfiltration efficiency
NIOSHthe USA National Institute for Occupational Safety and Health
PLApolylactic acid
PMambient particulate matter
STLStereoLithography
UVGIultraviolet germicidal irradiation
VHPvapor hydrogen peroxide

References

  1. Feng, S.; Shen, C.; Xia, N.; Song, W.; Fan, M.; Cowling, B.J. Rational Use of Face Masks in the COVID-19 Pandemic. Lancet Respir. Med. 2020, 8, 434–436. [Google Scholar] [CrossRef]
  2. Burki, T. Global Shortage of Personal Protective Equipment. Lancet Infect. Dis. 2020, 20, 785–786. [Google Scholar] [CrossRef]
  3. National Institute for Occupational Safety and Health (NIOSH). Cfr 84 Respiratory Protective Devices: Final Rules and Notice. Federal Register; Federal Register US Centers for Disease Control and Prevention, National Institute for Occu-Pational Safety and Health: Atlanta, GA, USA, 1997.
  4. Centers for Disease Control and Prevention (CDC). Niosh-Approved N95 Particulate Filtering Facepiece Respirators. Available online: https://www.cdc.gov/niosh/npptl/topics/respirators/disp_part/n95list1.Html (accessed on 21 January 2021).
  5. Fisher, E.M.; Shaffer, R.E. Considerations for Recommending Extended Use and Limited Reuse of Filtering Facepiece Respirators in Health Care Settings. J. Occup. Environ. Hyg. 2014, 11, D115–D128. [Google Scholar] [CrossRef]
  6. Centers for Disease Control and Prevention (CDC). Recommended Guidance for Extended Use and Limited Reuse of N95 Filtering Facepiece Respirators in Healthcare Settings. Available online: https://www.cdc.gov/niosh/topics/hcwcontrols/recommendedguidanceextuse.Html#ref19 (accessed on 13 January 2021).
  7. Rebmann, T.; Carrico, R.; Wang, J. Physiologic and Other Effects and Compliance With Long-Term Respirator Use Among Medical Intensive Care Unit Nurses. Am. J. Infect. Control. 2013, 41, 1218–1223. [Google Scholar] [CrossRef]
  8. Chin, A.W.H.; Chu, J.T.S.; Perera, M.R.A. Correspondence. Stability of Sars-Cov-2 in Different Environmental Conditions. Lancet Microbe 2020, 1, E10. [Google Scholar] [CrossRef]
  9. Cramer, A.; Enze, T.; Galanek, M.; Lamere, E.; Li, J.; Gupta, R.; Short, M. Assessment of the Qualitative Fit Test and Quantitative Single-Pass Filtration Efficiency of Disposable N95 Masks Following Gamma Irradiation. JAMA Netw. Open 2020, 3, E209961.e61. [Google Scholar] [CrossRef]
  10. Centers for Disease Control and Prevention (CDC). Implementing Filtering Facepiece Respirator (FFR) Reuse, Including Reuse After Decontamination, When There Are Known Shortages of N95 Respirators. Available online: https://www.cdc.gov/coronavirus/2019-ncov/hcp/ppe-strategy/decontamination-Reuse-respirators.Html (accessed on 13 January 2021).
  11. Rodriguez-Martinez, C.E.; Sossa-Briceño, M.P.; Cortes, J. Decontamination and Reuse of N95 Filtering Facemask Respirators: A Systematic Review of the Literature. Am. J. Infect. Control. 2020, 48, 1520–1532. [Google Scholar] [CrossRef]
  12. Derraik, J.G.B.; Anderson, W.A.; Connelly, E.A.; Anderson, Y.C. Rapid Review of Sars-Cov-1 and Sars-Cov-2 Viability, Susceptibility to Treatment, and the Disinfection and Reuse of Ppe, Particularly Filtering Facepiece Respirators. Int. J. Environ. Res. Public Health 2020, 17, 6117. [Google Scholar] [CrossRef]
  13. Kierat, W.; Augustyn, W.; Koper, P.; Pawlyta, M.; Chrusciel, A.; Wyrwol, B. The Use of UVC Irradiation to Sterilize Filtering Facepiece Masks Limiting Airborne Cross-Infection. Int. J. Environ. Res. Public Health 2020, 17, 7396. [Google Scholar] [CrossRef]
  14. O’Hearn, K.; Gertsman, S.; Webster, R.; Tsampalieros, A.; Ng, R.; Gibson, J.; Sampson, M.; Sikora, L.; McNally, J.D. Efficacy and Safety of Disinfectants for Decontamination of N95 and SN95 Filtering Facepiece Respirators: A Systematic Review. J. Hosp. Infect. 2020, 106, 504–521. [Google Scholar] [CrossRef] [PubMed]
  15. Nicolau, T.; Filho, N.; Zille, A. Ultraviolet-C As a Viable Reprocessing Method for Disposable Masks and Filtering Facepiece Respirators. Polymers 2021, 13, 801. [Google Scholar] [CrossRef] [PubMed]
  16. Gertsman, S.; Agarwal, A.; O’Hearn, K.; Webster, R.; Tsampalieros, A.; Barrowman, N.; Sampson, M.; Sikora, L.; Staykov, E.; Ng, R.; et al. Microwave- and Heat-Based Decontamination of N95 Filtering Facepiece Respirators: A Systematic Review. J. Hosp. Infect. 2020, 106, 536–553. [Google Scholar] [CrossRef] [PubMed]
  17. Fisher, E.M.; Shaffer, R.E. A Method to Determine the Available UV-C Dose for the Decontamination of Filtering Facepiece Respirators. J. Appl. Microbiol. 2010, 110, 287–295. [Google Scholar] [CrossRef] [PubMed]
  18. Kumar, A.; Kasloff, S.B.; Leung, A.; Cutts, T.; Strong, J.E.; Hills, K.; Vazquez-Grande, G.; Rush, B.; Lother, S.; Zarychanski, R. N95 Mask Decontamination Using Standard Hospital Sterilization Technologies. MedRxiv 2020. [Google Scholar] [CrossRef]
  19. The United States Food and Drug Administration (FDA). The U.S. Food and Drug Administration (FDA) Issued an Emergency Use Authorization (EUA) for the Emergency Use of the Steris Sterilization System for Use in Decontaminating Compatible N95 Respirators. Available online: https://www.fda.gov/media/136843/Download (accessed on 21 January 2021).
  20. The United States Food and Drug Administration (FDA). The U.S. Food and Drug Administration (FDA) Issued an Emergency Use Authorization (EUA) for the Emergency Use of the Stryker Sterilization System for Use in Decontaminating Compatible N95 Respirators. Available online: https://www.fda.gov/media/136976/Download (accessed on 6 June 2021).
  21. The United States Food and Drug Administration (FDA). The U.S. Food and Drug Administration (FDA) Issued an Emergency Use Authorization (EUA) for the Emergency Use of the Sterrad Sterilization System for Use in Decontaminating Compatible N95 Respirators. Available online: https://www.fda.gov/media/136884/Download (accessed on 6 June 2021).
  22. STERIS Healthcare. Instructions for Healthcare Facilities: Decontamination of Compatible N95 Respirators Using the Steris Sterilization Systems. Available online: https://www.steris.com/-/media/documents/pdfs/covid19-Landing-page/6-8/steris--Instructions-for-Healthcare-Facilities-6520.Ashx (accessed on 13 January 2021).
  23. Centers for Disease Control and Prevention (CDC). Assessment of Filter Penetration Performance and Fit for Decontam-Inated N95 Respirators. Available online: https://www.cdc.gov/niosh/npptl/respirators/testing/pdfs/NIOSHApproved_Decon_TestPlan10.Pdf (accessed on 12 April 2021).
  24. 3M. Decontamination of 3m Filtering Facepiece Respirators, Such As N95 Respirators, in the United States—Considerations. Available online: https://multimedia.3m.com/mws/media/1824869O/decontamination-Methods-for-3m-Filtering-Facepiece-Respirators-Technical-bulletin.Pdf (accessed on 13 January 2021).
  25. Migliore, M. Testing the Limits of Facemask Filtration." International Filtration News. Available online: https://www.filtnews.com/Testing-the-Limits-of-Facemask-Filtration/ (accessed on 3 July 2021).
  26. Maher, B.; Chavez, R.; Tomaz, G.C.; Nguyen, T.; Hassan, Y. A Fluid Mechanics Explanation of the Effectiveness of Common Materials for Respiratory Masks. Int. J. Infect. Dis. 2020, 99, 505–513. [Google Scholar] [CrossRef] [PubMed]
  27. Cramer, A.K.; Plana, D.; Yang, H.; Carmack, M.M.; Tian, E.; Sinha, M.S.; Krikorian, D.; Turner, D.; Mo, J.; Li, J.; et al. Analysis of SteraMist Ionized Hydrogen Peroxide Technology in the Sterilization of N95 Respirators and Other PPE. Sci. Rep. 2021, 11, 1–10. [Google Scholar] [CrossRef]
  28. Salimifard, P.; Rim, D.; Freihaut, J.D. Evaluation of Low-Cost Optical Particle Counters for Monitoring Individual Indoor Aerosol Sources. Aerosol Sci. Technol. 2019, 54, 217–231. [Google Scholar] [CrossRef]
  29. STERIS Healthcare. Steris Decontamination Solutions for Compatible N95 Respirators for Healthcare Facilities Outside the United States. Available online: https://www.steris.com/healthcare/Steris-Decontamination-Solutions-for-Compatible-n95-Respirators-Non-United-States (accessed on 13 January 2021).
  30. Chao, C.; Wan, M.P.; Morawska, L.; Johnson, G.; Ristovski, Z.; Hargreaves, M.; Mengersen, K.; Corbett, S.; Li, Y.; Xie, X.; et al. Characterization of Expiration Air Jets and Droplet Size Distributions Immediately at the Mouth Opening. J. Aerosol Sci. 2009, 40, 122–133. [Google Scholar] [CrossRef] [Green Version]
  31. Lee, B. Minimum Sizes of Respiratory Particles Carrying Sars-Cov-2 and the Possibility of Aerosol Generation. Int. J. Environ. Res. Public Health 2020, 17, 6960. [Google Scholar] [CrossRef]
  32. Bikov, A.; Paschalaki, K.; Logan-Sinclair, R.; Horváth, I.; Kharitonov, S.A.; Barnes, P.J.; Usmani, O.S.; Paredi, P. Standardised Exhaled Breath Collection for the Measurement of Exhaled Volatile Organic Compounds by Proton Transfer Reaction Mass Spectrometry. BMC Pulm. Med. 2013, 13, 43. [Google Scholar] [CrossRef] [Green Version]
  33. Castell, N.; Dauge, F.R.; Schneider, P.; Vogt, M.; Lerner, U.; Fishbain, B.; Broday, D.; Bartonova, A. Can Commercial Low-Cost Sensor Platforms Contribute to Air Quality Monitoring and Exposure Estimates? Environ. Int. 2017, 99, 293–302. [Google Scholar] [CrossRef] [PubMed]
  34. Rai, A.; Kumar, P.; Pilla, F.; Skouloudis, A.N.; DI Sabatino, S.; Ratti, C.; Yasar, A.-U.-H.; Rickerby, D. End-User Perspective of Low-Cost Sensors for Outdoor Air Pollution Monitoring. Sci. Total Environ. 2017, 607–608, 691–705. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  35. Dlugosz, L. Ambient Airborne Particle Concentrations in Hospital Inpatient Rooms. ASHRAE Trans. 2014, 120. Available online: link.gale.com/apps/doc/A394346839/AONE?u=anon~26a941c3&sid=googleScholar&xid=2e41d8d9 (accessed on 25 April 2021).
  36. Aganovic, A.; Cao, G.; Stenstad, L.-I.; Skogås, J.G. Impact of Surgical Lights on the Velocity Distribution and Airborne Contamination Level in an Operating Room With Laminar Airflow System. Build. Environ. 2017, 126, 42–53. [Google Scholar] [CrossRef]
  37. Bopp, N.E.; Bouyer, D.H.; Gibbs, C.M.; Nichols, J.E.; Ntiforo, C.A.; Grimaldo, M.A. Multicycle Autoclave Decontamination of N95 Filtering Facepiece Respirators. Appl. Biosaf. 2020, 25, 150–156. [Google Scholar] [CrossRef]
  38. Kim, J.-M.; Chung, Y.-S.; Jo, H.J.; Lee, N.-J.; Kim, M.S.; Woo, S.H.; Park, S.; Kim, J.W.; Kim, H.M.; Han, M.-G. Identification of Coronavirus Isolated from a Patient in Korea With COVID-19. Osong Public Health Res. Perspect. 2020, 11, 3–7. [Google Scholar] [CrossRef] [Green Version]
  39. Kwon, N.-J.; Kang, C.-K.; Choe, P.-G.; Yun, J.; Seong, M.-W.; Seo, J.-S.; Oh, M.-D. Virus Isolation from the First Patient With Sars-Cov-2 in Korea. J. Korean Med. Sci. 2020, 24, e84. [Google Scholar]
  40. O’Dowd, K.; Keerthi, M.N.; Parnia, F.; Snehamol, M.; Grant, J.; Moran, R.; Bartlett, J.; Bird, J.; Pillai, S.C. Face Masks and Respirators in the Fight Against the Covid-19 Pandemic: A Review of Current Mate-Rials, Advances and Future Perspectives. Materials 2020, 13, 3363. [Google Scholar] [CrossRef]
  41. World Health Organization (WHO). Modes of Transmission of Virus Causing Covid-19: Implications for Ipc Precaution Recommendations: Scientific Brief, 27 March 2020; World Health Organization: Geneva, Switzerland, 2020. [Google Scholar]
  42. World Health Organization. Transmission of SARSCoV-2: Implications for Infection Prevention Precautions. Scientific Brief. 9 July 2020. Available online: https://www.who.int/news-room/commentaries/detail/transmission-of-sars-cov-2-implications-for-infection-prevention-precautions (accessed on 9 July 2020).
  43. Ong, S.W.X.; Tan, Y.K.; Chia, P.Y.; Lee, T.H.; Ng, O.T.; Wong, M.S.Y.; Marimuthu, K. Air, Surface Environmental, and Personal Protective Equipment Contamination by Severe Acute Respiratory Syndrome Coronavirus 2 (Sars-Cov-2) from a Symptomatic Patient. JAMA 2020, 323, 1610–1612. [Google Scholar] [CrossRef] [Green Version]
  44. Liu, Y.; Ning, Z.; Chen, Y.; Guo, M.; Liu, Y.; Gali, N.G.; Sun, L.; Duan, Y.; Cai, J.; Westerdahl, D. Aerodynamic Characteristics and Rna Concentration of Sars-Cov-2 Aerosol in Wuhan Hospitals During Covid-19 Outbreak. BioRxiv 2020. [Google Scholar] [CrossRef]
  45. Cai, C.; Floyd, E.L. Effects of Sterilization With Hydrogen Peroxide and Chlorine Dioxide Solution on the Filtration Efficiency of N95, KN95, and Surgical Face Masks. JAMA Netw. Open 2020, 3, e2012099. [Google Scholar] [CrossRef]
  46. Oral, E.; Wannomae, K.K.; Connolly, R.L.; Gardecki, J.A.; Leung, H.M.; Muratoglu, O.K.; Griffiths, A.; Honko, A.N.; Avena, L.E.; McKay, L.G.A. Vapor H2o2 Sterilization As a Decontamination Method for the Reuse of N95 Respirators in the Covid-19 Emergency. MedRxiv 2020. [Google Scholar] [CrossRef]
  47. Viscusi, D.J.; King, W.P.; Shaffer, R.E. Effect of Decontamination on the Filtration Efficiency of Two Filtering Facepiece Respirator Models. J. Int. Soc. Respir. Prot. 2007, 24, 93. [Google Scholar]
  48. Bergman, M.S.; Viscusi, D.J.; Heimbuch, B.K.; Wander, J.D.; Sambol, A.R.; Shaffer, R.E. Evaluation of Multiple (3-Cycle) Decontamination Processing for Filtering Facepiece Respirators. J. Eng. Fibers Fabr. 2010, 5. [Google Scholar] [CrossRef] [Green Version]
  49. Kumar, A.; Bhattacharjee, B.; Sangeetha, D.N.; Subramanian, V.; Venkatraman, B. Evaluation of Filtration Efficacy of Various Types of Face-Masks Using Ambient and Pao Aerosols Following With Different Sterilization Methods. MedRxiv 2020. [Google Scholar] [CrossRef]
  50. Wigginton, K.R.; Arts, P.J.; Clack, H.L.; Fitzsimmons, W.J.; Gamba, M.; Harrison, K.R.; LeBar, W.; Lauring, A.S.; Li, L.; Roberts, W.W. Validation of N95 Filtering Facepiece Respirator Decon-Tamination Methods Available at a Large University Hospital. Open Forum Infect. Dis. 2021. [Google Scholar] [CrossRef]
  51. TSI Incorporated. Calibration Certificate Failure Explanation for Aerotrak Particle Counters. Available online: https://tsi.com/getmedia/490d98b8-d8e2-4c2d-9632-2b036ac1657f/Calibration_Cert_Failure_Explanation_AeroTrak_APCs_CC-119-US-web?ext=.Pdf (accessed on 6 June 2020).
  52. Viscusi, D.J.; Bergman, M.S.; Eimer, B.C.; Shaffer, R.E. Evaluation of Five Decontamination Methods for Filtering Facepiece Respirators. Ann. Occup. Hyg. 2009, 53, 815–827. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  53. DeAngelis, H.E.; Grillet, A.M.; Nemer, M.B.; Wasiolek, M.A.; Hanson, D.J.; Omana, M.A.; Sanchez, A.L.; Vehar, D.W.; Thelen, P.M. Gamma Radiation Sterilization of N95 Respirators Leads to Decreased Respirator Performance. PLoS ONE 2021, 16, e0248859. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Illustration of the 3D-printed specialized air duct used to measure the filtration efficiency (FE) of FFRs. (AC) are 3D designs from different angles of the air duct. (D) is the actual air duct connected to AeroTrak particle counter (1), through the head (2) of the air duct showing a tested mask (3), bound to the tail (4) of the air duct and the electric fan (5). The overall dimensions of the air ducts are 19-cm-long, 14-cm-wide, and 12-cm-high; the head dimensions are 10-cm-long, 14-cm-wide, and 12-cm, while the tail dimensions are 9-cm-long, 14-cm-wide, and 12-cm.
Figure 1. Illustration of the 3D-printed specialized air duct used to measure the filtration efficiency (FE) of FFRs. (AC) are 3D designs from different angles of the air duct. (D) is the actual air duct connected to AeroTrak particle counter (1), through the head (2) of the air duct showing a tested mask (3), bound to the tail (4) of the air duct and the electric fan (5). The overall dimensions of the air ducts are 19-cm-long, 14-cm-wide, and 12-cm-high; the head dimensions are 10-cm-long, 14-cm-wide, and 12-cm, while the tail dimensions are 9-cm-long, 14-cm-wide, and 12-cm.
Ijerph 18 07169 g001
Figure 2. Filtration efficiency (FE) of different types of FFRs. Seven types of FFRs were applied to FE evaluation for different particle sizes (0.3–5 µm) using AeroTrak particle counter; KN95 (moulded shape, n = 10), 3M-1820 (n = 10), Gerson-1730 (n = 5), Medline (n = 5), Benehal (n = 5), N99/N95 SpectraShield Plus (n = 5), and KN95 (duck shape, n = 5) masks. Symbols represent the mean, and error bars indicate the standard error. Statistically, an overall significant difference (p = 0.001) was observed between the median FE values between the FFRs, with a p-value of 0.001. Pairwise comparison: 3M-N95 vs. N99/N95 SpectraShield: p < 0.05; Benehal-N95 vs. N99/N95 SpectraShield: p < 0.05; no other pairwise comparison significant differences were observed. Raw data are listed in Supplementary Materials 3.
Figure 2. Filtration efficiency (FE) of different types of FFRs. Seven types of FFRs were applied to FE evaluation for different particle sizes (0.3–5 µm) using AeroTrak particle counter; KN95 (moulded shape, n = 10), 3M-1820 (n = 10), Gerson-1730 (n = 5), Medline (n = 5), Benehal (n = 5), N99/N95 SpectraShield Plus (n = 5), and KN95 (duck shape, n = 5) masks. Symbols represent the mean, and error bars indicate the standard error. Statistically, an overall significant difference (p = 0.001) was observed between the median FE values between the FFRs, with a p-value of 0.001. Pairwise comparison: 3M-N95 vs. N99/N95 SpectraShield: p < 0.05; Benehal-N95 vs. N99/N95 SpectraShield: p < 0.05; no other pairwise comparison significant differences were observed. Raw data are listed in Supplementary Materials 3.
Ijerph 18 07169 g002
Figure 3. Stability of filtration efficiency (FE) of KN95 and 3M-N95 masks over H2O2 decontamination cycles. Twelve of each KN95 (left panel) and 3M-N95 (right panel) masks were cycled in VHP-STERIS sterilization systems 10 times. The masks were applied to FE stability evaluation after each VHP cycle for different particle sizes (0.3–5 µm) using AeroTrak particle counter. Symbols represent the mean, and error bars indicate the standard error. Statistically, the 3M-N95 masks showed an overall significant (p < 0.001) reduction in FE stability with increased VHP cycles. Pairwise comparison: 0.3 µm vs. 5 µm, p < 0.001; 0.5 µm vs. 5 µm, p < 0.001; 0.7 µm vs. 5 µm, p = 0.006; 0.3 µm vs. 2 µm, p = 0.002; 0.5 µm vs. 2 µm, p = 0.002; no other pairwise comparison significant differences were observed. Raw data are listed in Supplementary Materials 4.
Figure 3. Stability of filtration efficiency (FE) of KN95 and 3M-N95 masks over H2O2 decontamination cycles. Twelve of each KN95 (left panel) and 3M-N95 (right panel) masks were cycled in VHP-STERIS sterilization systems 10 times. The masks were applied to FE stability evaluation after each VHP cycle for different particle sizes (0.3–5 µm) using AeroTrak particle counter. Symbols represent the mean, and error bars indicate the standard error. Statistically, the 3M-N95 masks showed an overall significant (p < 0.001) reduction in FE stability with increased VHP cycles. Pairwise comparison: 0.3 µm vs. 5 µm, p < 0.001; 0.5 µm vs. 5 µm, p < 0.001; 0.7 µm vs. 5 µm, p = 0.006; 0.3 µm vs. 2 µm, p = 0.002; 0.5 µm vs. 2 µm, p = 0.002; no other pairwise comparison significant differences were observed. Raw data are listed in Supplementary Materials 4.
Ijerph 18 07169 g003
Table 1. Description of FFRs used in the study.
Table 1. Description of FFRs used in the study.
FFRsTypeValve
Yes/No
CompanyCountryLot NumberNIOSH
Approved
Gerson 1730
Particulate Respirator
N95NoLouis M. GersonUnited StatesTC-84A-0160Yes
Medline-Cone StyleN95NoMedlineUnited StatesTC-84A-5411Yes
Benehal Particulate
respirator face mask
N95YesSuzhou Sanical
Protective Product
China541529Yes
SpectraShield+N99/N95NoNexera MedicalUnited StatesA84740.B1Yes
KN95 (duck shape)KN95NoYuyao Yukang
Medical Equipment
China20200506No 1
3M-8210 2N95No3MUnited States 3A12199Yes
KN 95 (molded shape)KN95NoZhongShan XiaoLan YiShuai Gament
Factory
China2020042701No 1
1 Meets Chinese standards GB2626:2006; 2 FDA-EUA-approved for decontamination and reuse; 3 Multinational conglomerate corporation.
Table 2. Comparison between in-house and NIOSH filtration assessment methods.
Table 2. Comparison between in-house and NIOSH filtration assessment methods.
In-House MethodNIOSH Method
Particle size (µm)≥0.3 *≥0.075
Flow rate (L/min)2.885.0
Aerosol concentration (mg/m3) Ambient particulate matter air (20–50)200
Testing filterTSI AeroTrak particle counterTSI 8310 automated filter tester
Sampling time/FFR (minute)110
* 0.3–5 µm.
Table 3. Comparison between percent uncertainty between the current study and the manufacturer calibration of AeroTrak (TSI 9306) particle counter.
Table 3. Comparison between percent uncertainty between the current study and the manufacturer calibration of AeroTrak (TSI 9306) particle counter.
Particle Size (µm)Percent Uncertainty (%)Manufacturer’s Percent Uncertainty (%)
0.31.43.9
0.53.03.9
0.74.73.9
13.13.9
24.53.8
54.43.8
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Al-Hadyan, K.; Alsbeih, G.; Nobah, A.; Lindstrom, J.; Falatah, S.; Faran, N.; Al-Ghamdi, S.; Moftah, B.; Alhmaid, R. In-House Filtration Efficiency Assessment of Vapor Hydrogen Peroxide Decontaminated Filtering Facepiece Respirators (FFRs). Int. J. Environ. Res. Public Health 2021, 18, 7169. https://doi.org/10.3390/ijerph18137169

AMA Style

Al-Hadyan K, Alsbeih G, Nobah A, Lindstrom J, Falatah S, Faran N, Al-Ghamdi S, Moftah B, Alhmaid R. In-House Filtration Efficiency Assessment of Vapor Hydrogen Peroxide Decontaminated Filtering Facepiece Respirators (FFRs). International Journal of Environmental Research and Public Health. 2021; 18(13):7169. https://doi.org/10.3390/ijerph18137169

Chicago/Turabian Style

Al-Hadyan, Khaled, Ghazi Alsbeih, Ahmad Nobah, Jeffrey Lindstrom, Sawsan Falatah, Nawarh Faran, Salem Al-Ghamdi, Belal Moftah, and Rashed Alhmaid. 2021. "In-House Filtration Efficiency Assessment of Vapor Hydrogen Peroxide Decontaminated Filtering Facepiece Respirators (FFRs)" International Journal of Environmental Research and Public Health 18, no. 13: 7169. https://doi.org/10.3390/ijerph18137169

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop