Medicina is published by MDPI from Volume 54 Issue 1 (2018). Previous articles were published by another publisher in Open Access under a CC-BY (or CC-BY-NC-ND) licence, and they are hosted by MDPI on mdpi.com as a courtesy and upon agreement with Lithuanian Medical Association, Lithuanian University of Health Sciences, and Vilnius University.
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: Data of one-year prospective clinical study

by
Reda Žemaitienė
1,*,
Martynas Špečkauskas
1,2,
Brigita Glebauskienė
1 and
Vytautas Jašinskas
1
1
Department of Eye Diseases
2
Institute for Biomedical Research, Kaunas University of Medicine, Lithuania
*
Author to whom correspondence should be addressed.
Medicina 2008, 44(12), 936; https://doi.org/10.3390/medicina44120117
Submission received: 19 May 2008 / Accepted: 5 December 2008 / Published: 10 December 2008

Abstract

Objective. To evaluate and compare the impact of two sharp-edge optic foldable intraocular lenses (IOLs) of similar design made from different material (hydrophilic acrylic or hydrophobic acrylic) on visual function, anterior and posterior capsule opacification at one-year follow-up after cataract phacoemulsification.
Material and methods. Seventy-two eyes of 72 patients scheduled for cataract surgery were included in a prospective clinical study. Two foldable sharp-edge optic posterior chamber acrylic IOLs of similar design were used. Thirty-nine eyes of 39 patients received a single-piece hydrophilic acrylic (PC 511, Ophtec) IOL and 33 eyes of 33 patients – single-piece hydrophobic acrylic (AcrySof, SA60AT, Alcon) IOL. Visual acuity, anterior capsule opacification (ACO), capsulorrhexis/optic overlapping, and posterior capsule opacification (PCO) were evaluated. The intensity of ACO was assessed subjectively. PCO values in the entire IOL optic area and in the central 3-mm optic zone were assessed using a photographic image analysis system (EPCO 2000). The patients were examined at one year postoperatively.
Results.
There were no significant differences in best-corrected visual acuity and capsulorrhexis/optic overlapping between IOL types at 1-year follow-up after surgery. In the single-piece hydrophilic acrylic IOL group, the grade of ACO density was significantly higher in capsulorrhexis rim area (1.56±0.71 and 1.00±0.75) and in the capsule/optic area (1.62±0.67 and 1.00±0.75) (P<0.05). PCO values of the entire IOL optic area (0.12±0.13 and 0.024±0.02) as well as in the central 3-mm optic zone (0.06±0.11 and 0.001±0.003) was significantly higher in the single-piece hydrophilic acrylic IOL group one year postoperatively (P<0.05). In 33.3% of cases of the single-piece hydrophilic acrylic IOL group, contraction of haptics to IOL optics was present one year postoperatively, which was not present in any case of the single-piece hydrophobic acrylic IOL group.
Conclusions
. One-year follow-up after cataract surgery has shown a significant difference in ACO and PCO development comparing single-piece hydrophilic acrylic and single-piece hydrophobic acrylic intraocular lenses. The effect of hydrophobic acrylic foldable lenses on preventing anterior and posterior capsule opacification is mainly a result of the acrylic hydrophobic biomaterial.
Keywords: cataract phacoemulsification; anterior capsulorrhexis; anterior capsule opacification; posterior capsule opacification; hydrophobic acrylic intraocular lens; hydrophilic acrylic intraocular lens cataract phacoemulsification; anterior capsulorrhexis; anterior capsule opacification; posterior capsule opacification; hydrophobic acrylic intraocular lens; hydrophilic acrylic intraocular lens

Share and Cite

MDPI and ACS Style

Žemaitienė, R.; Špečkauskas, M.; Glebauskienė, B.; Jašinskas, V. Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: Data of one-year prospective clinical study. Medicina 2008, 44, 936. https://doi.org/10.3390/medicina44120117

AMA Style

Žemaitienė R, Špečkauskas M, Glebauskienė B, Jašinskas V. Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: Data of one-year prospective clinical study. Medicina. 2008; 44(12):936. https://doi.org/10.3390/medicina44120117

Chicago/Turabian Style

Žemaitienė, Reda, Martynas Špečkauskas, Brigita Glebauskienė, and Vytautas Jašinskas. 2008. "Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: Data of one-year prospective clinical study" Medicina 44, no. 12: 936. https://doi.org/10.3390/medicina44120117

Article Metrics

Back to TopTop