Next Article in Journal
DepressionMIGNN: A Multiple-Instance Learning-Based Depression Detection Model with Graph Neural Networks
Previous Article in Journal
Research and Implementation of Travel Aids for Blind and Visually Impaired People
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Current Sensor with Optimized Linearity for Lightning Impulse Current Measurement

1
China Electric Power Research Institute, Beijing 100192, China
2
State Grid Sichuan Electric Power Research Institute, Chengdu 610041, China
*
Author to whom correspondence should be addressed.
Sensors 2025, 25(14), 4516; https://doi.org/10.3390/s25144516
Submission received: 2 April 2025 / Revised: 7 June 2025 / Accepted: 12 June 2025 / Published: 21 July 2025
(This article belongs to the Section Electronic Sensors)

Abstract

Impulse current measurement technology is widely used in various applications, including lightning protection monitoring in power systems, welding current measurement in aircraft and shipbuilding industries, as well as high-current measurement in pulsed power systems. With the advancement of industrial technology, the measurement range of impulse currents has continuously expanded, reaching levels as high as mega-amperes (MA). The calibration of the scale factor for impulse current measurement devices is determined through comparison with standard measurement devices. Developing high-accuracy impulse current measurement devices and accurately judging their characteristics are prerequisites for ensuring the precise calibration of impulse current values. This paper introduces two different types of high-impulse current measurement devices. Experimental studies were conducted on the scale factor and response characteristics of the sensors. The scale factor extension calibration method for sensors under high currents of more than 100 kA has also been introduced. Test results indicate that the developed impulse current measurement devices can serve as standard measurement devices for high impulse current measurement.

1. Introduction

Impulse current measurement technology is widely applied in power systems, meteorological research, industrial manufacturing, and power measurement of pulse power sources. The measurement accuracy of impulse current directly affects the safe and stable operation of power systems, relates to the precise control of industrial production processes, and determines the accurate power calculation of high-power pulse current sources [1,2,3,4,5]. Shunt, Rogowski coils, and magneto-sensitive sensors are commonly used for lightning impulse current measurements. Shunts, designed with a coaxial structure, offer excellent response characteristics and shielding performance, were usually adopted as standard measurement devices. Rogowski coils and magnetic field sensors, operating on non-contact measurement principles, inherently offer wider current measurement ranges and have gained extensive application in field measurement scenarios [6,7,8,9].
With the advancement of various industries, the demand for measuring the amplitude of impulse currents continues to rise. In the pulse power field, impulse current amplitudes can reach several megaamperes (MA), placing higher demands on the response characteristics, stability, and thermal tolerance performance of impulse current measurement devices [10,11]. The calibration of impulse current measuring devices by comparison with national/international standard lightning impulse current measurement systems constitutes a critical approach to ensuring measurement accuracy. However, internationally recognized standard systems have a rated current not exceeding 100 kA [12,13], which cannot meet full-scale calibration requirements. Supplementary linearity calibration tests were needed for the accurate measurement of impulse current value. Consequently, the development of high-accuracy, large-current linearity measurement devices to evaluate the nonlinearity characteristics of high-current impulse measuring devices under large currents has become a critical aspect of high-current impulse measurement technology [14].
Rogowski coils and magnetic field sensors are always used as high impulse current measurement sensors, owing to their advantages of non-contact connection, negligible heating effects, and independence from rated current limitations. Rogowski coils encompass both magnetic-core structures and air-core measurement coils. When measuring high currents exceeding 100 kA, air-core Rogowski coils are generally employed to prevent distortion of measurement results caused by magnetic core saturation [15,16,17,18,19].
Magnetic field sensor based on magneto resistive chips—including TMR (Tunnel Magnetoresistance), GMR (Giant Magnetoresistance), and Hall effect types—are now widely adopted in current measurement technologies. However, such sensors remain unsuitable as standard current measurement devices due to their significant susceptibility to ambient temperature and humidity variations [20,21,22,23,24]. Similarly, fiber-optic magnetic field sensors face stability limitations [25,26,27]. Both of the above magnetic field sensors are currently primarily applied to AC/DC current measurements. When utilized for impulse current measurements, further expansion of their frequency bandwidth characteristics is required. Consequently, this study primarily focuses on magnetic field sensors based on inductive coil structures.
During high-current impulse discharges, intense electromagnetic interference is generated in the surrounding space. This necessitates that current sensors possess exceptional shielding capabilities and interference immunity. In-depth investigation into the sensor’s shielding structure and circuit parameter design is required [28,29] to ensure accurate current measurements across the full-scale range. For the developed current sensors, determining their linearity parameters accurately constitutes another key research focus.
This study develops high-precision air-core Rogowski coils and magnetic field sensors based on Printed Circuit Board coil configurations to achieve accurate measurement of impulse currents at the hundred-kA level. First, the structures and operational principles of both sensors are analyzed, with corresponding structural designs and electrical circuit parameters proposed. Subsequently, key parameters including frequency bandwidth and sensitivity are analyzed and calculated. An impulse high-current characteristic test platform is established to validate the accuracy of the determined linearity parameters, and an experimental methodology for large-current measurements using these sensors is presented.

2. Lightning Impulse Current Measurement Sensor

2.1. Rogowski Coil

The non-intrusive installation of the Rogowski coil determines that it is not directly affected by the thermal effects of high current. The factors affecting its measurement accuracy mainly include bandwidth limitations, eccentricity effects, winding uniformity, leakage flux, and external electromagnetic interference. Under different test currents, the electromagnetic interference caused by transient discharge varies. Therefore, a Rogowski coil with strong shielding capability should be selected for linearity testing.
The structure diagram of the Rogowski coil is shown in Figure 1. Taking the cross-sectional area micro-element dS, the magnetic flux passing through the micro-element of this area is shown in Formula (1).
The magnetic flux through a single-turn current coil is
d ϕ = B · d S = μ r μ 0 i 2 π y d x d y ,
where i is the current to be measured, μ0 is the vacuum permeability (4π × 10−7 H/m), μr is the relative magnetic permeability of magnetic core material, and μr = 1 when the winding skeleton is made of insulating material, y is the coordinate value in the Y-direction on the core interface.
The magnetic flux ϕ linked by a single turn coil is
ϕ = S μ 0 μ r i 2 π y d x d y = μ 0 μ r i 2 π h 2 h 2 a b 1 y d x d y = μ 0 μ r h i 2 π ln b a
The induced electromotive force of the current coil is
e ( t ) = N d ϕ d t = N μ 0 μ r h 2 π ln b a · d i d t d x d y ,
where N is the number of turns of the coil.
In order to enhance the sensitivity of the sensor and ensure its uniform winding, a current coil based on a multi-layer Printed Circuit Board (here inafter referred to as PCB) structure design was chosen [10,11], as shown in Figure 2. The top and bottom layers of the multi-layer PCB form a Faraday cage shield to block interference from external electric fields. The two copper-clad shielding layers are interconnected through ground wire vias, allowing charge flow for the neutralization of reverse electric fields from external sources. The internal structure and appearance of the Rogowski coil are as depicted in the figure. The coil is equipped with an analog integration circuit, with a designed scale factor of 1.25 V/kA.
The characteristic parameters of the designed Rogowski coil were determined through theoretical analysis and simulation calculations, as shown in Table 1. The simulation results demonstrate that the designed Rogowski coil meets the requirements for measuring lightning current waveforms of (4/10) μs and (8/20) μs.

2.2. Magnetic Field Sensor

The magnetic field sensor developed by CEPRI is based on a constant-speed Archimedean spiral coil. Similarly to the Rogowski coil, it measures current based on the magnetic field, and the difference is that the sensing turns of an Archimedean coil are arranged in the same plane, and the conductor under test does not pass through the center of the Archimedean coil. The spiral coil is also printed on a PCB circuit board to improve the distribution uniformity of the coil. The voltage induced by the magnetic sensor is proportional to the change in the current being measured. The total induced electromotive force of the Archimedean spiral coil is the result of the superposition of the induced electromotive forces from multiple arc segments and the coil’s center. As a result, its coil area utilization efficiency is higher.
The positional relationship between the magnetic induction coil and the current-carrying conductor is shown in Figure 3. To maximize the magnetic flux passing through the induction coil, the plane of the coil must be perpendicular to the direction of the magnetic field generated by the current.
The time-domain expression of the impulse current i(t) is set as
i ( t ) = I a ( t ) ,
where I is the peak value of the impulse current, and a(t) is the functional expression of the unit impulse current.
For ease of derivation, assume that the plane of the Archimedean spiral lies in the XOY plane, with its starting point coinciding with point O. The Cartesian coordinate function of the Archimedean spiral is
x = r cos θ , y = r sin θ , r = b θ ,
R = a + bθ.
As shown in Figure 4, r is the distance from a point on the spiral to the fixed point. When θ = 0, a is the distance from the starting point of the spiral to the origin of the polar coordinates. θ is the total rotation angle (in radians) of the spiral.
To further improve the sensitivity of the Archimedean spiral coil, a double-layer series configuration is employed, effectively doubling its sensitivity. Figure 5 illustrates the schematic diagram, and the physical prototype of the magnetic field sensor based on the Archimedean spiral coil.
The direction of the magnetic flux density is perpendicular to the plane of the spiral. Therefore, the expression for the magnetic flux density at any point P1(rcosθ, rsinθ) on the double-layer spiral is
B ( t ) = μ 0 I a ( t ) 2 π ( d r cos θ ) = μ 0 I a ( t ) 2 π ( d r cos θ ) .
Then, the induced electromotive force dE on the infinitesimal arc at point P1 (in Figure 4) is
d E = d B ( t ) d t d S .
Thus, the induced electromotive force E for an Archimedean spiral coil of double layer can be calculated as
E = 2 × d d t S B ( t ) d S = 2 × d d t 0 2 n π 0 b θ B ( t ) r d r d θ = 2 × μ 0 I 2 π d a ( t ) d t 0 2 n π 0 b θ r ( d r cos θ ) d r d θ = μ 0 I π d a ( t ) d t 0 2 n π 0 b θ r ( d r cos θ ) d r d θ .
Simulation calculations of the Archimedean spiral coil were performed based on the designed parameters of the Archimedean coil, and the frequency response characteristics of the Archimedean coil were obtained through analysis, as shown in Figure 6. The frequency range spans from a minimum of 3 Hz to a maximum of 21 MHz. The frequency band characteristics indicate that its high-frequency performance fully meets the requirements for measuring impulse lightning currents and its low-frequency performance is slightly inferior compared to the Rogowski coil.
For a magnetic field sensor, when its measurement range of magnetic field values is fixed, the linearity of its scale factor (the variation in the scale factor under different magnetic field values) is the same. Therefore, when the magnetic field sensor is placed close to the current conductor under test, the linearity of the scale factor over a wide magnetic field range can be determined by comparing it with a standard shunt. For example, test 1: when the magnetic field sensor is placed 10 mm away from the current conductor, a comparison with a 100 kA standard shunt can determine the corresponding scale factor and its variation for each magnetic field value within a test range of up to 89.5 A/m. Test 2: When the current under test increases to 500 kA, the distance between the magnetic field sensor and the current conductor can be increased to 30 mm or more. Under a test current of 100 kA, the measured magnetic field value and the corresponding scale factor of the sensor can be determined. Based on the linearity of the scale factor established in test 1, the scale factor corresponding to each current point from 100 kA to 500 kA can be determined for the magnetic field sensor. Based on the scale factor under 500 kA, the magnetic field sensor can serve as a high-current measurement standard for currents exceeding hundreds of kA for other current measurement devices. Compared with the Rogowski coil, the Archimedean magnetic field sensor theoretically can provide better measurement accuracy since it does not require curve fitting or extrapolation to determine its scale factor.
As a single magnetic field sensor may be susceptible to spatially interfering electromagnetic fields, by designing a ring-shaped insulating support framework and arranging multiple sensor arrays, the effects of spatial co-directional electromagnetic interference can be effectively mitigated, as shown in Figure 7. For the array structure of the Archimedean magnetic field sensor construction, a circular (annular) sensor mounting framework must be machined. This framework should feature uniformly distributed arrayed mounting holes. Each mounting hole should incorporate multiple fixation slots. These slots allow flexible adjustment of the sensor’s radial distance from the central conductor under test. The output voltage signals from all magnetic field sensors must be summed. The aggregated signal is connected to the output terminal via coaxial cable. The output is then fed into a data acquisition system for multi-array magnetic field measurements.

3. 100 kA Lightning Impulse Current Test

3.1. Test of Rigid Rogowski Coil

(1) Scale factor calibration
Figure 8 shows the Schematic Diagram for the scale factor calibration for the Rogowski coil. The impulse current generator employs an energy storage capacitor with C = 6 μF, an inductor with L = 3.6 μH, and a resistor R = 0.73 Ω. Simulation results demonstrate that this configuration can generate an impulse current waveform with time parameters T1(Front time)/T2(time to half value) of 3.97/11.03 μs. The experimental layout diagram is shown in Figure 9.
To ascertain the scale factor and linearity parameters of the developed Rogowski coil, two standard shunts were selected for measuring (4/10) μs and (8/20) μs lightning current waveforms, with rated currents of 40 kA and 100 kA, respectively. The peak current measurement uncertainty is 0.4% (k = 2), while the time parameter measurement uncertainty is 1.0% (k = 2). The calibration test current value is from 5 kA to 40 kA with the 40 kA standard shunt. The calibrated scale factor of the Rogowski coil is reflected by the variation in scale factor under different test current values, as shown in Figure 10. The calibration waveform is as shown in Figure 11. Based on the determined scale factor, the fitting curve formula for calibrating the scale factor y is established as
Y = 9 × 10−5x + 1.2493.
The 100 kA standard shunt was then used to calibrate the scale factor of the Rogowski coil within the 40 kA–100 kA range. The calibration results are shown in Table 2. During the calibration tests, the Rogowski coil’s scale factor was preset to 1.25 V/kA, and each set of experimental data was based on 10 repeated measurements. By comparing the actual calibrated scale factors with the fitted calibration scale factors determined using Equation (4), it was found that the deviation of the calibrated scale factors at each current point within the 40 kA–100 kA range did not exceed 0.1%. This confirms the excellent linearity of the developed Rogowski coil, and the fitted scale factor can be reliably applied to determine the scale factor beyond the 100 kA current range.
The relative error of two measurement devices is as follows: the relative error of front time is within 1%, and that of the half-peak time is within 0.5%, indicating that the developed Rogowski coil exhibits excellent response characteristics.
(2) Step response experiment
Based on the test circuit constructed in Figure 12, a step response test was conducted on the Rogowski coil to evaluate its frequency response characteristics. The step waveform generator can produce a square wave current with a pulse width of 1 μs and an amplitude of 100 A based on the discharge of the voltage on the coaxial cable. Figure 13 shows the measured step response waveform of the Rogowski coil, where the green curve represents the direct output signal from the Rogowski coil, and the blue waveform represents the output signal after integrating the Rogowski coil’s output signal.
The step response data of the Rogowski coil yielded a calculated partial response time of 83.45 ns, with an experimentally measured response time of 78.43 ns and an overshoot of 1.74%. The settling time was 380 ns. According to standard requirements, the coil is capable of measuring impulse current waveforms with rise times shorter than 1 μs.
(3) Short term stability
A 100 kA impulse current waveform was applied to the Rogowski coil test circuit 20 times consecutively with 3 min intervals between measurements. After the test, the scale factor of the Rogowski coil was measured again. The variation between the two scale factors represents the short-term stability of the Rogowski coil. The scale factor of the Rogowski coil changed by 0.11% before and after the test.
(4) Measurement Uncertainty Evaluation
Based on the above test results, the measurement uncertainty of the Rogowski coil was evaluated. Table 3 lists the uncertainty components of the calibration factor measurement results for the Rogowski coil. Within the 100 kA current range, the relative measurement uncertainty of the calibration factor for the Rogowski coil is Urel = 0.52% (k = 2).

3.2. Test of Magnetic Field Sensor

Figure 14 illustrates the schematic diagram for the scale factor calibration test of the magnetic field sensor with a standard shunt. The magnetic field sensor is mounted on a fixture 10 mm away from the current-carrying rod, and the measured signals are transmitted to the digital recorder through coaxial cables. The test is performed at five current at 20 kA, 40 kA, 60 kA, 80 kA, and 100 kA. The error between the shunt and the Rogowski coil is evaluated by the average test results by repeating the test five times. The experimental layout diagram of the magnetic field sensor is as shown in Figure 15.
Fix the magnetic field sensor at a distance of 10 mm from the current-carrying conductor. Apply a current of 100 kA to the conductor, the corresponding magnetic field measurement range for the magnetic field sensor is 89.5 A/mm according to the simulation based on Electromagnetic simulation software Ansoft 16.0.3. The picture of the simulation results is as shown in Figure 16.
Figure 17 displays the calibrated scale factor of the magnetic field sensor under different currents. The scale factor of the magnetic field sensor rises as the test current increases. The scale factor changes by about 0.3% from 20 kA to 100 kA, with similar patterns for both positive and negative polarities.
When the actual current to be measured increases further, the distance between the magnetic field sensor and the current-carrying conductor can be increased to ensure that the measured magnetic field remains within the calibrated scale factor range of the sensor. This enables the measurement of scale factors for currents exceeding 100 kA.

4. Conclusions

To achieve an accurate measurement of impulse currents at the hundred-kA level, this paper presents two distinct structural configurations of impulse high-current measuring devices.
The developed Rogowski coil employs a multi-layer PCB circuit design, enhancing measurement sensitivity while maintaining electromagnetic interference immunity. Optimizing the matching of circuit parameters effectively expands its measurement bandwidth. Experimental results demonstrate a step response rise time of 83 ns and settling time of 380 ns, fully satisfying rapid impulse current measurement requirements. Within the 100 kA range, calibration against a standard shunt determined its scaling factor with linearity better than 0.3% and peak current relative measurement uncertainty of 0.52%. For currents above 100 kA, curve fitting establishes linearity parameters of the scaling factor, with accuracy experimentally verified across the 50 kA to 100 kA range.
The developed magnetic field sensor employs a double-layer Archimedean coil design, offering significantly reduced size and more convenient installation compared to Rogowski coils. Depending on field test conditions, measurements can be performed using either a single sensor or multiple sensors arranged in an array configuration. Calibration against a standard shunt determines the scale factor within a specified magnetic field range and establishes its linearity. When tested at currents between 20 kA and 100 kA, the sensor demonstrates linearity better than 0.2% within the corresponding magnetic field range. For currents exceeding this range, increasing the distance between the magnetic field sensor and the current-carrying conductor under test allows recalibration of the scaling factor within the rated current limits of the standard shunt. This approach, combined with the predetermined linearity parameters, enables impulse current measurements at substantially higher current levels.
Any of the aforementioned scale factor calibration methods can be selected to satisfy the accuracy requirements. In calibration tests, appropriate calibration methods and standard equipment can be selected based on actual needs and existing test conditions.

Author Contributions

Conceptualization, W.L. project administration, Y.D., investigation, Z.L., data curation, S.X.; validation, J.F. and K.H.; software, Z.W., writing—review and editing, W.L., supervision, F.Z. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported by the State Grid science and technology project (5700-202455253A-1-1-ZN).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Dataset available on request from the authors.

Conflicts of Interest

Author Wengting Li, Yinglong Diao, Feng Zhou, Zhaozhi Long, Jiawei Fan, Kangmin Hu and Zhehao Wang were employed by the company China Electric Power Research Institute, and author Shijun Xie was employed by the company State Grid Sichuan Electric Power Research Institute. The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  1. Destefan, D.E. Calibration and testing facility for resistance welding current monitors. IEEE Trans. Instrum. Meas. 1996, 45, 453–456. [Google Scholar] [CrossRef]
  2. Ziegler, S.; Woodward, R.C.; Iu, H.H.; Borle, L.J. Current Sensing Techniques: A Review. IEEE Sens. J. 2009, 9, 354–376. [Google Scholar] [CrossRef]
  3. Yao, X.; Chen, J.; Sun, W. Experimental study on measuring error in impulse current system. High Volt. Eng. 2006, 32, 12–15. [Google Scholar]
  4. Yuan, Y.; Ren, W.; Wang, A.; Jiang, N.; Li, H. Study on calibration of impulse current measuring system. High Volt. Appar. 2015, 51, 102–106. [Google Scholar]
  5. Zou, J.; Duan, X.; Zhang, T. The simulating calculation and experimental research of Rogowski coil for current measurement. Trans. China Electrotech. Soc. 2001, 16, 81–84. [Google Scholar]
  6. Liu, J.; Xu, Q.; Li, S.; Yin, Y.; Feng, J.; Zhou, X. Fold Band Shunt Used in measuring the ns-grade pulse current. High Volt. Eng. 2006, 32, 57–59. [Google Scholar]
  7. Yu, C.; Yu, D.; He, J. Analysis of step response of plate-shunt. J. Xi’an Jiao Tong Univ. 1991, 25, 65–72. [Google Scholar]
  8. Fang, H. Low-inductance shunt for high impulse current measurement. High Volt. Appratus 1982, 18, 18–24. [Google Scholar]
  9. Gerasimov, A. Wide-range inductive sensors of currents with nanosecond rise times for measuring parameters of high-current pulses (Review). Instrum. Exp. Tech. 2002, 45, 147–161. [Google Scholar] [CrossRef]
  10. Jun, Y.; Liu, H.; Yang, G.; Ling, G. Design of a Flexible Rogowski Coil with Active integrator Applied in Lightning Current Collection. In Proceedings of the 2016 33rd International Conference on Lightning Protection (ICLP), Estoril, Portugal, 25–30 September 2016. [Google Scholar]
  11. Metwally, I.A. Self-Integrating Rogowski coil for high-impulse current measurement. IEEE Trans. Instrum. Meas. 2010, 59, 353–360. [Google Scholar] [CrossRef]
  12. IEC Standard 62475; High-Current Test Techniques—Definitions and Requirements for Test Currents and Measuring Systems. IEC: Geneva, Switzerland, 2010.
  13. Istrate, D.; Blanc, I.; Fortuné, D. Development of a Measurement Setup for High Impulse Currents. IEEE Trans. Instrum. Meas. 2013, 62, 1473–1478. [Google Scholar] [CrossRef]
  14. Chiampi, M.; Crotti, G.; Morando, A. Evaluation of flexible Rogowski Coil performances in power frequency applications. IEEE Trans. Instrum. Meas. 2011, 60, 854–862. [Google Scholar] [CrossRef]
  15. Nurmansah, A.P.; Hidayat, S. Design and testing PCB Rogowski-coil current sensor for high current application. In Proceedings of the 2017 International Conference on High Voltage Engineering and Power Systems (ICHVEPS), Bali, Indonesia, 2–5 October 2017. [Google Scholar]
  16. IEEE PSRC Special Report. Practical Aspects of Rogowski Coil Applications to Relaying. September 2010, pp. 1–72. Available online: https://www.pes-psrc.org/kb/report/034.pdf (accessed on 11 June 2025).
  17. Chekurov, Y.; Hällström, J. Influence of busbar geometry on AC current measurement using Rogowski coil. In Proceedings of the 2008 Conference on Precision Electromagnetic Measurements Digest, Broomfield, CO, USA, 8–13 June 2008; pp. 541–543. [Google Scholar]
  18. Sarwade, A.N.; Katti, P.K.; Ghodekar, J.G. Use of accurate measurement of Secondary Current contaminated with CT saturation in Distance Protection Scheme. In Proceedings of the 2016 IEEE 6th International Conference on Power Systems (ICPS), New Delhi, India, 4–6 March 2016. [Google Scholar]
  19. Texas Instrument Design. Active Integrator for Rogowski Coil Reference Design with Improved Accuracy for Relays and Breakers” TIDUBY4A–April 2016–Revised September 2016. Available online: https://www.ti.com/tool/TIDA-00777 (accessed on 11 June 2025).
  20. Sanchez, J.; Ramirez, D.; Ravelo, S.I.; Lopes, A.; Cardoso, S.; Ferreira, R.; Freitas, P.P. Electrical Characterization of a Magnetic Tunnel Junction Current Sensor for Industrial Applications. IEEE Trans. Magn. 2012, 48, 2823–2826. [Google Scholar] [CrossRef]
  21. Marcellis, A.D.; Cubells-Beltran, M.D.; Reig, C.; Madrenas, J.; Zadov, B.; Paperno, E.; Cardoso, S.; Freitas, P.P. Quasi-digital front-ends for current measurement in integrated circuits with giant magnetoresistance technology. IET Circuits Devices Syst. 2014, 8, 291–300. [Google Scholar] [CrossRef]
  22. Poon, T.Y.; Tse, N.C.F.; Lau, R.W.H. Extending the GMR Current Measurement Range with a Counteracting Magnetic Field. Sensors 2013, 13, 8042–8059. [Google Scholar] [CrossRef] [PubMed]
  23. Li, Z.; Zhang, S.; Wu, Z.; Abu-Siada, A.; Tao, Y. Study of Current Measurement Method Based on Circular Magnetic Field Sensing Array. Sensors 2018, 18, 1439. [Google Scholar] [CrossRef] [PubMed]
  24. Ripka, P.; Kejik, P.; Kaspar, P.; Draxler, K. Precise DC current sensors. In Proceedings of the IEEE Instrumentation and Measurement Technology Conference and IMEKO Tec, Brussels, Belgium, 1 February 1996; pp. 1479–1483. [Google Scholar]
  25. Karvan, K.; Bhargab, D. Polymer transduction cavity based optical fiber Fabry—Perot interferometer for VOC sensing. Results Opt. 2023, 12, 100479. [Google Scholar]
  26. Gong, Z.; Xiang, Z.; OuYang, X.; Zhang, J.; Lau, N.; Zhou, J.; Chan, C.C. Wearable Fiber Optic Technology Based on Smart Textile: A Review. Materials 2019, 12, 3311. [Google Scholar] [CrossRef] [PubMed]
  27. Yerin, C.V.; Vivchar, V.I. Ellipsometry of magnetic fluid in a magnetic field. J. Magn. Magn. Mater. 2020, 49, 166144. [Google Scholar] [CrossRef]
  28. Draxler, K.; Styblikova, R. Magnetic shielding of Rogowski coils. IEEE Trans. Instrum. Meas. 2018, 67, 1207–1213. [Google Scholar] [CrossRef]
  29. Kyriazis, G.A. The effectiveness of current-comparator magnetic shields against leakage fields at power frequencies. In Proceedings of the Conference Digest Conference on Precision Electromagnetic Measurements, Ottawa, ON, Canada, 16–21 June 2002; pp. 34–35. [Google Scholar]
Figure 1. Diagram of the Structure of a Roche Coil. a: inner radius of the coil, b: outer radius of the coil, h: height of the coil.
Figure 1. Diagram of the Structure of a Roche Coil. a: inner radius of the coil, b: outer radius of the coil, h: height of the coil.
Sensors 25 04516 g001
Figure 2. Schematic diagram of multi-layer PCB Rogowski coil.
Figure 2. Schematic diagram of multi-layer PCB Rogowski coil.
Sensors 25 04516 g002
Figure 3. Relationship between coil and current coupling.
Figure 3. Relationship between coil and current coupling.
Sensors 25 04516 g003
Figure 4. Receiving area of coil magnetic field.
Figure 4. Receiving area of coil magnetic field.
Sensors 25 04516 g004
Figure 5. Structure and physical diagram of magnetic field sensor.
Figure 5. Structure and physical diagram of magnetic field sensor.
Sensors 25 04516 g005
Figure 6. Frequency characteristic analysis of the Archimedean coil.
Figure 6. Frequency characteristic analysis of the Archimedean coil.
Sensors 25 04516 g006
Figure 7. Schematic diagram of the magnetic field sensor with an array structure.
Figure 7. Schematic diagram of the magnetic field sensor with an array structure.
Sensors 25 04516 g007
Figure 8. Schematic diagram for calibratoin test based on Rogowski coil. T—Transformer, D—High Voltage Silicon Stack, C—Charging Capacitor, G—Discharge Gap, L—Circuit Resistance, R2—Circuit Resistance, Coil—Rogowski Coil, Shunt—Current Divider, M—Digital Recorder.
Figure 8. Schematic diagram for calibratoin test based on Rogowski coil. T—Transformer, D—High Voltage Silicon Stack, C—Charging Capacitor, G—Discharge Gap, L—Circuit Resistance, R2—Circuit Resistance, Coil—Rogowski Coil, Shunt—Current Divider, M—Digital Recorder.
Sensors 25 04516 g008
Figure 9. Experimental layout diagram for the Rigid Rogowski Coil.
Figure 9. Experimental layout diagram for the Rigid Rogowski Coil.
Sensors 25 04516 g009
Figure 10. Scale factor calibration and extrapolation data.
Figure 10. Scale factor calibration and extrapolation data.
Sensors 25 04516 g010
Figure 11. Calibration waveforms.
Figure 11. Calibration waveforms.
Sensors 25 04516 g011
Figure 12. Schematic diagram of the Rogowski coil square wave response test circuit.
Figure 12. Schematic diagram of the Rogowski coil square wave response test circuit.
Sensors 25 04516 g012
Figure 13. Step response waveform of the Rogowski coil.
Figure 13. Step response waveform of the Rogowski coil.
Sensors 25 04516 g013
Figure 14. Schematic diagram for linearity test based on magnetic field sensor. T—Transformer, D—Silicon Stack, C—Charging Capacitor, G—Discharge Spark Gap L—Circuit Inductane, R2—Circuit Resistance, Sensor—Magnetic Field Sensor, Shunt—Shunt Resistor, M—Data Logger.
Figure 14. Schematic diagram for linearity test based on magnetic field sensor. T—Transformer, D—Silicon Stack, C—Charging Capacitor, G—Discharge Spark Gap L—Circuit Inductane, R2—Circuit Resistance, Sensor—Magnetic Field Sensor, Shunt—Shunt Resistor, M—Data Logger.
Sensors 25 04516 g014
Figure 15. Experimental layout diagram.
Figure 15. Experimental layout diagram.
Sensors 25 04516 g015
Figure 16. Simulation distribution map of magnetic field around current conductor.
Figure 16. Simulation distribution map of magnetic field around current conductor.
Sensors 25 04516 g016
Figure 17. Scale Factor calibration results of magnetic field sensor.
Figure 17. Scale Factor calibration results of magnetic field sensor.
Sensors 25 04516 g017
Table 1. Circuit parameters of the designed Rogowski coil.
Table 1. Circuit parameters of the designed Rogowski coil.
Coil ParametersValueCoil ParametersValue
Self-inductance 3.04 μHDistributed capacitance2.32 (nF)
Low-frequency cutoff Frequency80 HzHigh-frequency cutoff Frequency1.8 (MHz)
Sensitivity1.253 (V/kA)Rise time120 (ns)
Table 2. Calibration scale factor of the Rogowski coil within the 100 kA range.
Table 2. Calibration scale factor of the Rogowski coil within the 100 kA range.
Standard Current (kA)Measured Current (kA)Calibrated Scale Factor k1 (V/A)Scale Factor of Fitting Curve
k2 (V/A)
Relative Deviation
(%)
49,624.149,632.01.25321.2529−0.02
59,809.359,852.31.25391.25380.02
69,742.369,825.81.25451.25470.02
79,856.380,034.71.25581.2556−0.02
89,769.289,998.51.25621.2565−0.02
95,280.395,576.91.25691.2574−0.02
49,624.149,818.21.25791.2583−0.04
Table 3. Measurement uncertainty evaluation of the Rogowski coil.
Table 3. Measurement uncertainty evaluation of the Rogowski coil.
Measurement Uncertainty
Components
ParameterMeasurement Uncertainty ComponentsParameter
Introduced by measurement repeatability uA0.0002Introduced by the short-term stability uB30.00063
Introduced by the standard measuring device uB10.002Introduced by long-term stability uB40.001
Introduced by the linearity of the Rogowski coil’s scale factor uB20.001//
Combined standard measurement uncertainty
u C = u A 2 + u B 1 2 + u B 2 2 + u B 3 2 + u B 4 2
2.53 × 10−3Expanded measurement uncertainty (k = 2)Urel = 0.52%
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Li, W.; Diao, Y.; Zhou, F.; Long, Z.; Xie, S.; Fan, J.; Hu, K.; Wang, Z. Current Sensor with Optimized Linearity for Lightning Impulse Current Measurement. Sensors 2025, 25, 4516. https://doi.org/10.3390/s25144516

AMA Style

Li W, Diao Y, Zhou F, Long Z, Xie S, Fan J, Hu K, Wang Z. Current Sensor with Optimized Linearity for Lightning Impulse Current Measurement. Sensors. 2025; 25(14):4516. https://doi.org/10.3390/s25144516

Chicago/Turabian Style

Li, Wenting, Yinglong Diao, Feng Zhou, Zhaozhi Long, Shijun Xie, Jiawei Fan, Kangmin Hu, and Zhehao Wang. 2025. "Current Sensor with Optimized Linearity for Lightning Impulse Current Measurement" Sensors 25, no. 14: 4516. https://doi.org/10.3390/s25144516

APA Style

Li, W., Diao, Y., Zhou, F., Long, Z., Xie, S., Fan, J., Hu, K., & Wang, Z. (2025). Current Sensor with Optimized Linearity for Lightning Impulse Current Measurement. Sensors, 25(14), 4516. https://doi.org/10.3390/s25144516

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop