Next Article in Journal
Images from Bits: Non-Iterative Image Reconstruction for Quanta Image Sensors
Next Article in Special Issue
Improvement of Depth Profiling into Biotissues Using Micro Electrical Impedance Spectroscopy on a Needle with Selective Passivation
Previous Article in Journal
Queuing Time Prediction Using WiFi Positioning Data in an Indoor Scenario
Previous Article in Special Issue
Optimized Detection of Plasmodium falciparum Topoisomerase I Enzyme Activity in a Complex Biological Sample by the Use of Molecular Beacons
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Evaluation of Hyaluronic Acid Dilutions at Different Concentrations Using a Quartz Crystal Resonator (QCR) for the Potential Diagnosis of Arthritic Diseases

by
Luis Armando Carvajal Ahumada
1,2,3,4,*,
Marco Xavier Rivera González
1,2,
Oscar Leonardo Herrera Sandoval
3,4 and
José Javier Serrano Olmedo
1,2
1
Center for Biomedical Technology (CTB), Technical University of Madrid (UPM), 28040 Madrid, Spain
2
Networking Research Center of Bioengineering, Biomaterials and Nanomedicine (CIBER-BBN), Technical University of Madrid (UPM), 28040 Madrid, Spain
3
Research and Technological Development Center of Electrical, Electronic and ICT Industry (CIDEI), 111311 Bogotá, Colombia
4
Faculty of Engineering and Basic Sciences, Central University, 111711 Bogotá, Colombia
*
Author to whom correspondence should be addressed.
Sensors 2016, 16(11), 1959; https://doi.org/10.3390/s16111959
Submission received: 10 August 2016 / Revised: 30 September 2016 / Accepted: 20 October 2016 / Published: 22 November 2016
(This article belongs to the Special Issue Point-of-Care Biosensors)

Abstract

:
The main objective of this article is to demonstrate through experimental means the capacity of the quartz crystal resonator (QCR) to characterize biological samples of aqueous dilutions of hyaluronic acid according to their viscosity and how this capacity may be useful in the potential diagnosis of arthritic diseases. The synovial fluid is viscous due to the presence of hyaluronic acid, synthesized by synovial lining cells (type B), and secreted into the synovial fluid thus making the fluid viscous. In consequence, aqueous dilutions of hyaluronic acid may be used as samples to emulate the synovial fluid. Due to the viscoelastic and pseudo-plastic behavior of hyaluronic acid, it is necessary to use the Rouse model in order to obtain viscosity values comparable with viscometer measures. A Fungilab viscometer (rheometer) was used to obtain reference measures of the viscosity in each sample in order to compare them with the QCR prototype measures.

1. Introduction

The synovial fluid is produced by the synovium membrane inside the articular cartilage. It can be considered a dialysate of plasma with a high content of hyaluronic acid [1,2,3,4]. The amount of the synovial fluid in a joint is no more than 3.5 mL (knee) [4,5] where the main functions of the synovial fluid are the lubrication of the joints and the absorption of the mechanical load produced when walking or running (knees) or holding up weight (cubit and shoulder) [1,2,3,4,5]. The viscosity of the synovial fluid depends on the concentration of the hyaluronic acid. There are studies that relate the low viscosity of the synovial fluid (low concentration of hyaluronic acid) with arthrocentesis, osteoarthritis and inflammatory arthropathies [1,2,3,4,6,7]. Due to the concentration of hyaluronic acid in the synovial fluid being ≈ 3.5 mg/mL [6], the sample may be considered a non-Newtonian fluid with viscoelastic properties [5,8]. Due to this lubricant function, the analysis of synovial fluid is important for the detection and diagnosis of arthritic diseases.
At present, the evaluation of the viscosity of synovial fluid samples is done through macroscopic analysis. This analysis evaluates the color, transparency and viscosity of the sample [1,2,3,9,10]. The results of a macroscopic analysis make the classification of the samples possible in healthy, inflammatory or septic, according to Table 1 [5,9,11].
The viscosity test involves extending the sample on a surface and measuring the length before the thread is cut. If the resulting length is greater than 3 cm, the sample is healthy; if not, the sample displays a problem and more tests are necessary [5].
Although the viscosity of the sample is a fundamental parameter in the macroscopic analysis [3], the clinical method to obtain its value is primarily qualitative. The sample of synovial fluid is obtained through a puncture in the joint and it is extended on a surface with the same needle used to extract it. If the final length is greater than 3 cm, the sample is considered healthy; but if it is lower than 3 cm, the sample is considered abnormal [11].
The macroscopic method provides a subjective measure of viscosity that depends, for example, on the skills and strength of the laboratory researcher, the stretching and the surface. Sometimes the viscosity is not even taken into account, only the appearance [1].
Due to the importance of the hyaluronic acid concentration in the synovial fluid, the small amount of sample in the joints and the current subjective method to obtain a measure of the viscosity, it has been proposed to use a piezoelectric biosensor based on a quartz crystal resonator (QCR) to obtain the measure of viscosity of the sample.
There are several studies linking the use of the Quartz Crystal Microbalance (QCM) to the detection of diseases characterizing biological fluids as blood [12,13,14,15], urine [16,17], saliva [17,18,19] and synovial fluid [20,21]. The methods proposed in these studies have proven to be successful according to their results. However, they are also more complex because they involve the functionalization of the electrode in the crystal in order to detect a specific protein related to the disease. There are also studies that relate similar piezoelectric techniques to sense the viscosity of biological samples, such as: the use of cantilevers [22,23,24]. Sensors based on cantilevers are also valid to obtain the viscosity values in biological samples; the physics principle is the same. However, the cantilever transducer is more fragile because it only has one fixed contact point. In consequence, the manufacture and use of these biosensors have a higher level of difficulty.
In contrast to the standard method applied in clinical and others more complex piezoelectric techniques, the method proposed enables us to know the viscosity value for synovial fluid samples in a simple way because the proposed technique does not require the functionalization of the electrode and the manufacture of the biosensor is easy.
In preliminary tests, aqueous dilutions of hyaluronic acid at different concentrations have been used to emulate the synovial fluid (both healthy and abnormal). As reference equipment, the Fungilab viscometer has been used to obtain viscosity measures of each sample in order to compare them with experimental measures.

2. Theory

The present theory is exposed in two parts. The first involves the design and manufacture of the piezoelectric biosensor, and the second contains the mathematical explanation about the viscoelastic behavior of the samples used and the equations needed to make the proposed analysis.

2.1. Quartz Crystal Resonator

The quartz crystal resonators are electro-mechanical transducers with two electrodes which are sensitive to the deposition of mass [25,26,27,28] or to contact with liquid [29,30,31]. The quartz crystal has a resonance frequency determined by its geometry, in particular, by its thickness. The crystal is stimulated by an AC voltage between its electrodes and, as a consequence, the crystal vibrates producing shear waves [32]. When the signal frequency is close to the resonance frequency of the crystal, the intensity of the vibration increases. By performing a frequency sweep close to the resonance, and measuring the voltage and current in the crystal, it is possible to know the behavior of the admittance of the crystal for such frequencies [30,31,33,34,35,36].
When the crystal is in contact with a mass or liquid, its resonance frequency changes according to two equations: the Sauerbrey equation for mass deposition (Equation (1)) [25] and the Kanazawa equation (Equation (2)) for fluids [37]:
Δ f = 2 f 0 2 A ρ q G q Δ m
Δ f = n f 0 3 / 2 ρ L η L π ρ q G q
where ρ q and G q are the specific density and the shear modulus of quartz respectively, f 0 is the fundamental resonance frequency of the quartz, related to its thickness, and Δ m is the thin film of mass deposited, A is the piezoelectrically active crystal area, ρ L and η L are the density and viscosity of the fluid respectively and n is the overtone number [30,37,38].
The Kanazawa equation only applies to Newtonian liquids (the viscosity is always constant) [34]. However, aqueous dilutions of hyaluronic acid and the synovial fluid are Non-Newtonian fluids [5,39,40]. For this reason, it is necessary to use a more general equation to find the viscosity value:
Δ F = 1 + i 2 π ρ q G q ρ L ω ( η L i η L )
η L = G L ω = π ρ q G q ρ L f r Δ F Δ Γ f 0 2
η L = G L ω = 1 2 π ρ q G q ρ L f r ( Δ Γ 2 Δ F 2 ) f 0 2
According to Johannsmann [34], Equations (3)–(5) are more general expressions and these can be used to obtain the values of two components of the viscosity for non-Newtonian fluids. ω is the angular frequency, η L and η L are the real and the imaginary parts of the viscosity, and G’L and G’’L are the real and the imaginary part of the shear modulus for the liquid respectively. Δ F and ΔΓ are the resonance frequency shift and the bandwidth shift respectively (Figure 1).
According to Figure 1, Δf and ΔΓ are obtained by using the following equations:
Δ f = f s f 0
Δ Γ = Γ s Γ 0
For Newtonian liquids ( η L = const, η L = 0), Δf and ΔΓ are equal and opposite. For viscoelastic liquids ( η L = η(ω), η L ≠ 0) [34].
There is another important variable called motional resistance (R). Motional resistance is defined as the impedance value at series resonant frequency [41]. This resistance is related to mechanical energy losses in the crystal during its vibration [31]. In this study, R is calculated as the inverse value of the conductance curve peak. The change in motional resistance (ΔR) due to the deposition sample is calculated based on the following equation:
Δ R = R s ( w i t h   s a m p l e ) R 0 ( u n d a m p e d )

2.2. Homogeneous Viscoelastic Samples with Pseudo-Plastic Behavior

When a viscoelastic sample with pseudo-plastic behavior is deposited and characterized using a QCR, the frequency shift obtained is several times lower than expected [39,40]. In consequence, the Equations (3)–(5) cannot be applied directly because the apparent viscosity obtained will be similar to the viscosity of the sample solvent [39].
Additionally, it is not possible to compare the QCR measures with the viscometer measures directly because the pseudoplastic behavior is related to the shear rate. In other words, the apparent viscosity for pseudoplastic fluids (hyaluronic acid), defined as the slope of the curve in Figure 2, decreases when the shear rate increases [42]. For Newtonian fluids, it is constant. Figure 2 presents this phenomenon.
The shear rate applied to the viscometer is several units of magnitude lower than the shear rate in the QCR because the frequency in the viscometer is less to 1 Hz and the QCR works at 10 MHz. In order to compare the apparent viscosity at a low shear rate (viscometer) with the apparent viscosity at a high shear rate (QCR), an additional factor χ[ωτ0] is necessary, where ω is the angular frequency and τ0 is the relaxation time of the sample [40]. According to [39,40], the Rouse model is appropriate to explain the behavior of polymeric fluids like hyaluronic acid samples at different shear rates. Figure 3 shows this behavior.
In the Rouse model, the relaxation time is defined as [39,40,43]:
τ 0 = 6 η 0 M π 2 ρ R T
where M is the molecular weight of the polymer, R is the gas constant (8.314 J·K−1·mol−1), T is the absolute temperature (288 K for experiments), η0 is the zero shear rate viscosity and ρ is the fluid density. Zero shear rate viscosity was obtained using the Fungilab viscometer at the minimum shear rate (0.1 rpm). These values have been saved into a database in the biosensor’s software.
According to [40], the frequency shift Δ ω obtained can be expressed as:
Δ ω = Δ ω 0 · χ [ ω τ 0 ]
where, Δω0 is only related to the viscous component of the viscoelastic sample (Newtonian behavior). This equation is the same Kanazawa Equation (2):
Δ ω 0   ~   ω 0 2 / 3 ( ρ η 0 ρ q G q )
Using factor χ and the frequency shift Δ ω obtained with the QCR, is possible to obtain a new frequency shift value Δω0 comparable to the viscometer measure. In other words, factor χ allows us to obtain the frequency shift generated for an equivalent Newtonian fluid (viscous component) Δω0 by using the QCR measures.
According to [40], factor χ is defined as:
χ [ ω τ 0 ] = η η 0 ( ( 1 + t a n 2 ( δ ) ) 1 tan ( δ ) )
where the loss tangent tan ( δ ) = η [ ω ] / η [ ω ] .
Using the generalized Maxwell model [40,44] the following equations are obtained (Appendix A):
η η 0 = 1 ζ ( ν ) k = 1 k ν k 2 ν + ( τ 0 ω ) 2
η η 0 = τ 0 ω ζ ( ν ) k = 1 k ν k 2 ν + ( τ 0 ω ) 2
For the Rouse model, ν = 2 and ζ(ν) is the Riemann Zeta Function [40,44].

3. Materials and Methods

3.1. Samples

The concentration of the hyaluronic acid in the healthy synovial fluid is around 3.5 mg/mL [6,7,45]. In traumatic arthritis and osteoarthritis the hyaluronic acid concentration decreases to around 1.3 mg/mL, and for rheumatoid arthritis the concentration is lower, only 0.84 mg/mL [6]. In this study, the hyaluronic acid dilutions were done at the same concentrations in order to emulate the healthy and abnormal synovial fluid. One additional sample at 7 mg/mL was made with the purpose of obtaining an extra point in the analysis of results. The hyaluronic acid (Mw = 1.5 MDa) was supplied by Sigma Aldrich (St. Louis, MO, USA).
Dilutions were prepared by weighing the hyaluronic acid salt (HA) using the analytic balance and mixing the HA with cold water, (water temperature was 4 °C). The amount of hyaluronic acid and the water volume were calculated according to the desired concentration. After mixing HA and cold water, the solution was stirred gently and continuously during 5 min. Finally, the samples were stored in the refrigerator for 8 h. This process is needed so that the mixture does not contain air bubbles (homogeneous solution). With the aim of evaluating the performance of the QCR biosensor and its detection limit, additional glycerol solutions were created at different concentrations in milli-q water (40%, 30%, 20%, 10%, 5%, 2.5%, 1.25%, 0.63%, 0.31%).

3.2. Crystals and Holder Cell

The 10 MHz quartz crystals were purchased from International Crystal Manufacturing (Oklahoma City, OK, USA). The static cell was obtained from Gamry Instruments Inc. (Warminster, PA, USA).

3.3. Viscometer of Reference

The rotational viscometer Alpha series manufactured by Fungilab has been used as reference equipment for the viscosity measures. Its resolution is 0.1 mPa·s according to manufacturer. This equipment is in the Center for Biomedical Technology (CTB).

3.4. Biosensor

The quartz crystal has two electrodes, one on each side. The active electrode receives the sample on its surface and is connected to the electric ground to give more stability to the measures. The proposed circuit gets the voltage and current signal for several frequencies in a configurable bandwidth close to the crystal series resonance frequency (Fs). The circuit is a variation of the Nakamoto and Kobayashi proposal [46]. The electronic system includes a DAC (AD9850 module, Analog Devices, Norwood, MA, USA) to get a high resolution in the measures. The AD9850 module is capable of creating a sinusoidal signal of between 1 and 40 MHz with a frequency step of 1 Hz. The amplitude of the input signal in the crystal is 1 Vp. The sweep frequency is configured close to the resonance frequency of the quartz crystal to identify the expected behavior. The quartz crystal is placed in an array with a toroidal-core transformer used as a current sensor [30,46]. The secondary coil of the transformer gives a voltage proportional to the current in the primary which is the crystal current. The voltage is shifted 90° with respect to the original current signal. The voltage signal (V2) related to the current through the crystal (I1) and the voltage used to stimulate the crystal (V1) are multiplied by using an AD835 mixer (manufacturer) according to Figure 4a.
The output signal in the mixer is filtered using an active filter with an LT1057 amplifier (Linear Technology, Milpitas, CA, USA) to obtain the DC component. According to [30], the DC component is related to the susceptance of the quartz crystal.
The acquisition system uses the ADC port of the Arduino DUE card (https://www.arduino.cc/) to obtain the output signals (voltage, current, susceptance). The digital resolution of the acquisition is 12 bits. With this system, the acquisition time per sweep cycle takes a few seconds. Figure 4b shows the blocks diagram of the prototype.
A Labview code has been developed to control the Arduino card. Through this program the user can choose the initial parameters (range of frequency sweep and amplitude of signal) and obtain the voltage and current signals on the quartz crystal for every frequency from it. The conductance and susceptance graphs are shown in the Labview interface, ΔF, ΔΓ and ΔR are calculated with the same code. Figure 5 displays the QCR prototype system.
Additionally, the Labview software has access to a database created with the external variables needed to obtain the correction factor χ: zero shear viscosity, molecular weight, the gas constant, the density of the fluid, temperature and the Riemann Zeta Function value for the Rouse model. The software uses the database in order to calculate the factor χ. The user can edit the zero shear viscosity, the density or the temperature if the value is known.
In this study, the concentration of the samples is always known. However, for an unknown sample, the biosensor uses the change in the motional resistance (ΔR) in order to select the correct factor χ. In the results section is shown the ΔR values for each sample. These resistance values are also saved in the database for reference.

3.5. Measurement Protocol

(1) Select a frequency range to make the sweep and to obtain the conductance curve in the frequency domain for the bare crystal (without sample). Calculate the serial resonance frequency (F0), the half band half width (Γ0) and the motional resistance (R0) using the developed code in Labview for each cycle. These measures are taken 10 times and the final values are the average of the measures; (2) Have the sample analyzed in the electrode surface of the quartz crystal (50 μL of volume); (3) Adjust the frequency range in Labview software to obtain the conductance curve of the quartz crystal in contact with liquid sample; (4) The measurements with samples (Fs, Γs and Rs) are taken 100 times (approx. 3 min) and the final values are the average among the measures; (5) Obtain ΔF, ΔΓ and ΔR.
The density values were obtained using an analytical balance. The mass of each solution was determined from 1 mL of its volume.

3.6. Protocol to Clean the Crystal

(1) Add abundant water to wash the residues on the crystal; (2) Add alcohol (isopropanol) to clean the crystal surface; (3) Add acetone to complete the electrode cleaning (remove solvents); (4) Dry the electrode surface with a liquid nitrogen pistol.

4. Results

4.1. Biosensor Performance

The sensitivity of the QCR prototype was obtained by following the measurement protocol steps using glycerol aqueous solutions at concentrations commented in Section 3.1.
Table 2 shows the viscosity obtained using the Kanazawa Equation (2). The results were obtained by averaging twenty individual measures for each solution. Variability in measures is in part due to the cleaning process and the crystal reposition into the cell.
The results allow us to see the capacity of the prototype sensor to detect the viscosity samples in up to 6.3 mg/mL of the solution (limit of detection). Figure 6 shows the performance of the QCR biosensor vs. Fungilab Viscometer. According to the Figure 6, the Fungilab viscometer does not differentiate samples with a concentration lower than 5%.
According to the response curve (Figure 7), the sensitivity of the QCR prototype is around 4 Hz per 1 μg/mL of sample concentration change.

4.2. Hyaluronic Acid Samples

According to the methods exposed, the following data were obtained for the hyaluronic acid mixtures:
Table 3 shows the measures obtained using the QCR prototype with different samples deposited on the electrode crystal. F0, ΔF, ΔΓ and ΔR correspond to the average of 20 independent measures for each sample.
Table 4 shows the apparent viscosity obtained using the Equations (4)–(6) with the frequency shift Δf and bandwidth shift ΔΓ measured. It is not possible to compare at a high shear rate (QCR measures) and low shear rate (viscometer measures) for the same sample.
According to the methods section, it is necessary to use factor χ to compare the biosensor measures with the viscometer measures. Using Equations (9)–(14), factor χ[ωτ0] and the frequency shift Δω0 solely related to the viscous component of the viscoelastic sample were calculated for each sample. Table 5 and Table 6 show the results.
Figure 8 shows the response curve of the biosensor versus the square root of the density viscosity product (Figure 8a) and versus the concentration of hyaluronic acid in the samples (Figure 8b). The frequency shift shown corresponds only to the viscous component of the viscoelastic samples.
Figure 9 shows the comparison between the apparent viscosity obtained by Rouse model using the QCR measures and the viscometer measures. The grey data in Figure 9 were obtained with the same Fungilab viscometer at a higher shear rate (1.5–2.5 rpm) than zero shear rate viscosity (0.1 rpm).

5. Discussion

The proposed QCR prototype has a good performance. The results illustrate the capability of the QCR sensor to measure small changes in the composition of glycerol aqueous solutions. Figure 7 shows the relationship between the concentration of the sample and the sensor response ΔF. According to Figure 7, the sensitivity of the sensor is ≈ 4.06 Hz per 1 μg/mL concentration sample. The smallest incremental change of concentration that could be detected in the biosensor (resolution) was 6.25 μg/mL.
According to the results for hyaluronic acids solutions, the presence of hyaluronic acid in an aqueous solution results in a complex rheological behavior; this solution is catalogued into the pseudoplastic fluids. Due to this viscoelastic and pseudoplastic behavior, it is not possible to compare the QCR measures (high shear rate) and the viscometer measures (low shear rate) for the same sample. Table 4 shows these results.
The QCR measures require an additional factor to obtain results comparable with the viscometer results. The Rouse model is a good alternative to obtain the additional factor χ in order to correct the experimental measures.
By applying the Rouse model to the experimental results, the new apparent viscosities were obtained for each sample. These apparent viscosities are only related to the viscous component of hyaluronic acid solutions. In consequence, the new apparent viscosities exhibit the same behavior as the viscometer measures (Figure 9). In other words, the new frequency shifts obtained correspond to an equivalent fluid with the viscous component alone. The new apparent viscosities obtained were calculated using Equation (11). Table 6 shows the results obtained after applying the Rouse model. The relative error increases as the concentration of the sample decreases. However, the viscosities obtained after applying the Rouse model are comparable with the reference viscosities obtained with the viscometer.
In this study, the concentration of the hyaluronic acid is already known for every sample. However, for future clinical samples, the concentration, the zero shear viscosity and the density are unknown variables. In such cases, it is necessary to use the motional resistance value in order to relate the response of the QCR with the correction factor χ. For unknown samples, the ΔR value obtained experimentally is compared with the ΔR values in the database and the system selects the parameters required to calculate the factor χ and adjust the viscosity value. Table 3 shows the ΔR values obtained for each sample.
The obtained results confirm the relationship between hyaluronic acid concentration and the viscosity of the fluid. Considering the concentration samples used in this study are very similar to the ones reported for arthritic diseases [6,7] and the results shown, the QCR biosensor would be capable of identifying synovial fluid samples (healthy or with any medical affectation).

6. Conclusions

Quartz crystal resonators are capable of identifying changes of viscosity in biologic fluid samples due to changes in the concentration of the fluid. The sensitivity of the QCR prototype was around 4 Hz per 1 μg/mL of sample concentration change. In this study, aqueous solutions of hyaluronic acid at different concentrations have been characterized through a QCR biosensor. The results prove that it is possible to relate the concentration of hyaluronic acid with the sample viscosity. Considering that hyaluronic acid is a key component in synovial fluid and taking into account that the hyaluronic acid concentrations used in this work are related to arthritic diseases [6,7], the QCR biosensor may be used in applications related to the diagnosis of such diseases. Additionally, the QCR technique may be beneficial in synovial fluid characterization because the amount of synovial fluid available in clinical samples is too little. For physicians, the fact of knowing the viscosity value of the synovial fluid in their patients will help them to make a timelier diagnosis, and to prepare more specific treatments in accordance with the status of the disease. As future work, testing will be performed with real synovial fluid in small mammals. Also, an artificial neural network will be implemented with a robust database. Such a neural network will provide the sensor a better capacity to select the correct factor χ to adjust the viscosity value.

Acknowledgments

The authors thank the Research and Technological Development Centre of Electrical, Electronical and ICT Industry (CIDEI) for its participation with ideas and developing of the prototype system. We also thank participants from Bioinstrumentation and the Nanomedicine group in the Center for Biomedical Technology (CTB) for its recommendations and loan of equipment. We thank COLCIENCIAS for supporting this study through grant No. 0375-2013 with project number 6570577636375 and Universidad Central (Colombia) for supporting this research through the Research Cluster on Converging Sciences and Technologies (NBIC).

Author Contributions

The prototype and the experiments were conceived and designed by Luis Armando Carvajal Ahumada; The prototype and experiments were performed by Luis Armando Carvajal Ahumada and Marco Xavier Rivera González, the data was analyzed by Luis Armando Carvajal Ahumada and Marco Xavier Rivera González; Reagents/materials/analysis tools were contributed by Oscar Leonardo Herrera Sandoval and Jose Javier Serrano Olmedo; the paper was written by Luis Armando Carvajal Ahumada, Marco Xavier Rivera González.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

The following abbreviations were used in this manuscript:
H.AHyaluronic Acid
QCRQuartz crystal resonator
MMedia
SDStandard deviation

Appendix A

Equations (12) and (13) emerge from the Maxwell constitutive Equation [44]:
τ + λ 1 τ t = η 0 γ ˙
where: γ ˙ is the rate of strain tensor, η0 is the zero shear rate viscosity and λ 1 is the relaxation time.
Solving for τ, integrating:
τ ( t ) = t ( η 0 λ 1 γ ˙ ( t ) ) e t λ 1 d t e t λ 1 + k e t λ 1
where primes have been added to t’s in the integrand in order to avoid confusion. k is the integration constant. If we prescribe that the stress in the fluid is finite at t = − , we must set k on zero.
Applying L’Hopital’s rules in Equation (A2):
lim t τ ( t ) = lim t ( ( η 0 λ 1 γ ˙ ( t ) ) e t λ 1 1 λ 1 e t λ 1 ) = η 0 γ ˙ ( )
Therefore, if γ ˙ ( ) is finite, the stress is also finite at t = − . Consequently, the Maxwell constitutive equation can be written in the form:
τ ( t ) = t ( η 0 λ 1 e ( t t ) λ 1 ) γ ˙ ( t ) d t
The quantity in brackets is called the relaxation modulus for Maxwell fluid. This expression shows that the stress at the present time t depends on the rate of strain at time t as well as on the rate of strain at all past times t’ with a weighting factor (relaxation modulus) that decays exponentially as one goes backwards in time.
Taking into account:
γ y x ( t 0 , t ) = t 0 t γ ˙ y x ( t ) d t
Then:
γ t = γ ˙ ( t )
γ y x ( t , t ) = t t γ ˙ ( t ) d t
Now, integrating by parts τ(t):
τ ( t ) = + t ( η 0 λ 1 2 e ( t t ) λ 1 ) γ ˙ ( t , t ) d t
The quantity in brackets is called memory function for Maxwell fluid. In this form the Maxwell model states that the stress at the present time t depends on the history of the strain for all past times < t < t .
Now, considering the Equation (1) as a sums of partial stresses (Generalized Maxwell Model):
τ ( t ) = k = 1 τ k ( t )
τ k + λ k τ k t = η k γ ˙
The shear stress can be written as:
τ ( t ) = t [ k = 1 η k λ k e ( t t ) λ k ] γ ˙ ( t ) d t
τ ( t ) = + t [ k = 1 η k λ k 2 e ( t t ) λ k ] γ ˙ ( t ) d t
where:
η k = λ k k λ k
λ k = λ k α
λ is a constant time and α is a dimensionless quantity which describes the slope of η versus ω on a log-log plot at high ω.
The Rouse molecular theory for dilute polymer solutions gives α = 2. Additionally, in the general linear viscoelastic model, the Equations (A11) and (A12) can be written as:
τ = t G ( t t ) γ ˙ ( t ) d t
τ = + t M ( t t ) γ ˙ ( t , t ) d t
where the memory function M is the derivative of the relaxation modulus G. Now, supposing a small amplitude oscillatory motion for the fluid:
γ ˙ y x ( t ) = γ ˙ 0 y x cos ( ω t )
γ ˙ 0 y x is real and positive.
Resolving Equation (A15) taking into account Equation (A17):
τ y x = t G ( t t ) γ ˙ 0 y x cos ( ω t ) d t
τ y x = γ ˙ 0 y x 0 G ( s ) cos ( ω ( t s ) ) d s
τ y x = [ 0 G ( s ) cos ( ω s ) d s ] γ ˙ 0 y x cos ( ω t ) [ 0 G ( s ) sin ( ω s ) d s ] γ ˙ 0 y x sin ( ω t )
Considering:
τ y x = η γ ˙ 0 y x cos ( ω t ) η γ ˙ 0 y x sin ( ω t )
Then:
η = 0 G ( s ) cos ( ω s ) d s
η = 0 G ( s ) sin ( ω s ) d s
Now, replacing the Equation (A11) in Equation (A22):
η = 0 [ k = 1 η k λ k e ( s ) λ k ] cos ( ω s ) d s
Solving for integration by parts two times:
0 [ k = 1 η k λ k e ( s ) λ k ] cos ( ω s ) d s   =   k = 1 η k λ k 2 1 ω 2 k = 1 ( 1 λ k 2 ω 2 ) * [ 0 [ k = 1 η k λ k e ( s ) λ k ] cos ( ω s ) d s ]
This equation can be written as:
η = k = 1 η k λ k 2 1 ω 2 k = 1 ( 1 λ k 2 ω 2 ) * η
Solving for η :
η = k = 1 η k 1 + ( λ k ω ) 2
Solving in similar way for η :
η = k = 1 η k λ k ω 1 + ( λ k ω ) 2
In addition, using the expressions (A13) and (A14) in Equations (A27) and (A28), these results become:
η η 0 = 1 ζ ( α ) k = 1 k α k 2 α + ( λ ω ) 2
η η 0 = λ ω ζ ( α ) k = 1 k α k 2 α + ( λ ω ) 2
Changing λ to τ0 and α to ν. Equations (A29) and (A30) are the same Equations (13) and (14).

References

  1. Pascual, E.; Jovani, V. Synovial fluid analysis. Best Pract. Res. Rheumatol. 2005, 19, 371–386. [Google Scholar] [CrossRef] [PubMed]
  2. Brannan, S.R.; Jerrard, D.A. Synovial fluid analysis. J. Emerg. Med. 2006, 30, 331–339. [Google Scholar] [CrossRef] [PubMed]
  3. Denton, J. Synovial fluid analysis in the diagnosis of joint disease. Diagn. Histopathol. 2012, 18, 159–168. [Google Scholar] [CrossRef]
  4. Sterling, W. Rheumatology Secrets, 3rd ed.; Elsevier: Philadelphia, PA, USA, 2015. [Google Scholar]
  5. Araceli, M.-C.; Núñez, C.; Cabiedes, J. Synovial fluid analysis. Reumatol. Clín. 2010, 6, 316–321. [Google Scholar]
  6. Stafford, C.; Niedermeier, W.; Holley, H.; Pigman, W. Studies on the concentration and intrinsic viscosity of hyaluronic acid in synovial fluids of patients with rheumatic diseases. Ann. Rheum. Dis. 1964, 23, 152–157. [Google Scholar] [CrossRef] [PubMed]
  7. Decker, B.; Mcguckin, W.F.; Mckenzie, B.F.; Slocumb, C.H. Concentration of hyaluronic acid in synovial fluid. Clin. Chem. 1959, 5, 465–469. [Google Scholar] [PubMed]
  8. Schurz, J.; Ribitsch, V. Rheology of synovial fluid. Biorheology 1987, 24, 385–399. [Google Scholar] [PubMed]
  9. Gallo Vallejo, F.J.; Giner Ruiz, V. Diagnosis: Synovial fluid analysis. Aten. Primaria 2014, 46, 29–31. [Google Scholar] [CrossRef]
  10. Gatter, R.A.; Andrews, R.P.; Cooley, D.A.; Fiechtner, J.J.; Minna, D.A.; Phelps, P.; Wasner, C. American college of rheumatology guidelines for performing office synovial fluid. J. Clin. Rheumatol. 1995, 1, 194–196. [Google Scholar] [CrossRef] [PubMed]
  11. Rojas, C. Estudio del Liquido Sinovial; Guías de Procedimientos en Reumatología: Bogotá, Colombia, 2011; pp. 41–47. (In Spanish) [Google Scholar]
  12. Muüller, L.; Sinn, S.; Drechsel, H.; Ziegler, C.; Wendel, H.-P.; Northoff, H.; Gehring, F.K. Investigation of Prothrombin Time in Human Whole-Blood Samples with a Quartz Crystal Biosensor. Anal. Chem. 2010, 82, 658–663. [Google Scholar] [CrossRef] [PubMed]
  13. Kömpf, D.; Held, J.; Müller, S.F.; Drechsel, H.; Tschan, S.C.; Northoff, H.; Mordmüller, B.; Gehring, F.K. Real-time measurement of Plasmodium falciparum-infected erythrocyte cytoadhesion with a quartz crystal microbalance. Malar. J. 2016, 15, 317. [Google Scholar] [CrossRef] [PubMed]
  14. Oberfrank, S.; Drechsel, H.; Sinn, S.; Northoff, H.; Gehring, F.K. Utilisation of quartz crystal microbalance sensors with dissipation (QCM-D) for a clauss fibrinogen assay in comparison with common coagulation reference methods. Sensors 2016, 16, 282. [Google Scholar] [CrossRef] [PubMed]
  15. Chen, Z.; Li, Q.; Chen, J.; Luo, R.; Maitz, M.F.; Huang, N. Immobilization of serum albumin and peptide aptamer for EPC on polydopamine coated titanium surface for enhanced in-situ self-endothelialization. Mater. Sci. Eng. C 2016, 60, 219–229. [Google Scholar] [CrossRef] [PubMed]
  16. Yang, Y.; Tu, Y.; Wang, X.; Pan, J.; Ding, Y. A label-free immunosensor for ultrasensitive detection of ketamine based on quartz crystal microbalance. Sensors 2015, 15, 8540–8549. [Google Scholar] [CrossRef] [PubMed]
  17. Alenus, J.; Ethirajan, A.; Horemans, F.; Weustenraed, A.; Csipai, P.; Gruber, J.; Peeters, M.; Cleij, T.J.; Wagner, P. Molecularly imprinted polymers as synthetic receptors for the QCM-D-based detection of l-nicotine in diluted saliva and urine samples. Anal. Bioanal. Chem. 2013, 405, 6479–6487. [Google Scholar] [CrossRef] [PubMed]
  18. Kim, Y.K.; Lim, S.-I.; Cho, Y.-Y.; Choi, S.; Song, J.-Y.; An, D.-J. Detection of H3N2 canine influenza virus using a Quartz Crystal Microbalance. J. Virol. Methods 2014, 208, 16–20. [Google Scholar] [CrossRef] [PubMed]
  19. He, F.; Zhang, L. Rapid Diagnosis of M. tuberculosis Using a Piezoelectric Immunosensor. Anal. Sci. 2002, 18, 397–401. [Google Scholar] [CrossRef] [PubMed]
  20. Ota, S.; Ozeki, T.; Hoshino, H.; Terao, K. AB0547 Rapid and novel immune assay platform of serum tocilizumab concentration applied by quartz-crystal microbalance (QCM) biosensor system. Ann. Rheum. Dis. 2013, 71, 669. [Google Scholar] [CrossRef]
  21. Sakti, S.P.; Wahyuni, F.; Juswono, U.P.; Aulanni, A.M. Development of QCM Biosensor with Specific Cow Milk Protein Antibody for Candidate Milk Adulteration Detection. Sens. Transducers 2013, 149, 143–148. [Google Scholar] [CrossRef]
  22. Agoston, A.; Keplinger, F.; Jakoby, B. Evaluation of a vibrating micromachined cantilever sensor for measuring the viscosity of complex organic liquids. Sens. Actuators A Phys. 2005, 123–124, 82–86. [Google Scholar] [CrossRef]
  23. Cakmak, O.; Elbuken, C.; Ermek, E.; Mostafazadeh, A.; Baris, I.; Erdem Alaca, B.; Kavakli, I.H.; Urey, H. Microcantilever based disposable viscosity sensor for serum and blood plasma measurements. Methods 2013, 63, 225–232. [Google Scholar] [CrossRef] [PubMed]
  24. Eris, G.; Bozkurt, A.A.; Sunol, A.; Jonáš, A.; Kiraz, A.; Alaca, B.E.; Erkey, C. Determination of viscosity and density of fluids using frequency response of microcantilevers. J. Supercrit. Fluids 2015, 105, 179–185. [Google Scholar] [CrossRef]
  25. Sauerbrey, G. Verwendung von Schwingquarzen zur Wägung dünner Schichten und zur Mikrowägung. Zeitschrift für Physik 1959, 155, 206–222. [Google Scholar] [CrossRef]
  26. Höök, F.; Kasemo, B.; Nylander, T.; Fant, C.; Sott, K.; Elwing, H. Variations in coupled water, viscoelastic properties, and film thickness of a Mefp-1 protein film during adsorption and cross-linking: A quartz crystal microbalance with dissipation monitoring, ellipsometry, and surface plasmon resonance study. Anal. Chem. 2001, 73, 5796–5804. [Google Scholar] [CrossRef] [PubMed]
  27. García-Martinez, G.; Bustabad, E.A.; Perrot, H.; Gabrielli, C.; Bucur, B.; Lazerges, M.; Rose, D.; Rodriguez-Pardo, L.; Fariña, J.; Compère, C.; et al. Development of a mass sensitive quartz crystal microbalance (QCM)-based DNA biosensor using a 50 MHz electronic oscillator circuit. Sensors 2011, 11, 7656–7664. [Google Scholar] [CrossRef] [PubMed]
  28. Fredriksson, C.; Kihlman, S.; Rodahl, M.; Kasemo, B. The Piezoelectric Quartz Crystal Mass and Dissipation Sensor: A Means of Studying Cell Adhesion. Langmuir 1998, 14, 248–251. [Google Scholar] [CrossRef]
  29. Rodahl, M.; Höök, F.; Kasemo, B. QCM Operation in Liquids: An Explanation of Measured Variations in Frequency and Q Factor with Liquid Conductivity. Anal. Chem. 1996, 68, 2219–2227. [Google Scholar] [CrossRef] [PubMed]
  30. Carvajal Ahumada, L.A.; Peña Pérez, N.; Herrera Sandoval, O.L.; del Pozo Guerrero, F.; Serrano Olmedo, J.J. A new way to find dielectric properties of liquid sample using the quartz crystal resonator (QCR). Sens. Actuators A Phys. 2016, 239, 153–160. [Google Scholar] [CrossRef]
  31. Auge, J.; Hauptmann, P.; Hartmann, J. New design for QCM sensors in liquids. Sens. Actuators B 1995, 24, 43–48. [Google Scholar] [CrossRef]
  32. Ferreira, G.N.M.; da-Silva, A.-C.; Tomé, B. Acoustic wave biosensors: Physical models and biological applications of quartz crystal microbalance. Trends Biotechnol. 2009, 27, 689–697. [Google Scholar] [CrossRef] [PubMed]
  33. Arnau, A. A Review of Interface Electronic Systems for AT-cut Quartz Crystal Microbalance Applications in Liquids. Sensors 2008, 8, 370–411. [Google Scholar] [CrossRef]
  34. Johannsmann, D. The Quartz Crystal Microbalance in Soft Matter Research, Soft and Biological Matter, 1st ed.; Springer: Gewerbestrasse, Switzerland, 2015; Volume 53. [Google Scholar]
  35. Jakoby, B.; Art, G.; Bastemeijer, J. Novel analog readout electronics for microacoustic thickness shear-mode sensors. IEEE Sens. J. 2005, 5, 1106–1111. [Google Scholar] [CrossRef]
  36. Resa, P.; Castro, P.; Rodríguez-López, J.; Elvira, L. Broadband spike excitation method for in-liquid QCM sensors. Sens. Actuators B Chem. 2012, 166–167, 275–280. [Google Scholar] [CrossRef] [Green Version]
  37. Keiji Kanazawa, K.; Gordon, J.G. The oscillation frequency of a quartz resonator in contact with liquid. Anal. Chim. Acta 1985, 175, 99–105. [Google Scholar] [CrossRef]
  38. Cassiède, M.; Daridon, J.-L.; Paillol, J.H.; Pauly, J. Characterization of the behaviour of a quartz crystal resonator fully immersed in a Newtonian liquid by impedance analysis. Sens. Actuators A Phys. 2011, 167, 317–326. [Google Scholar] [CrossRef]
  39. Fang, J.; Zhu, T.; Sheng, J.; Jiang, Z.; Ma, Y. Thickness Dependent Effective Viscosity of a Polymer Solution near an Interface Probed by a Quartz Crystal Microbalance with Dissipation Method. Sci. Rep. 2015, 5, 8491. [Google Scholar] [CrossRef] [PubMed]
  40. Nwankwo, E.; Durning, C. Mechanical response of thickness-shear mode quartz-crystal resonators to linear viscoelastic fluids. Sens. Actuators A Phys. 1998, 64, 119–124. [Google Scholar] [CrossRef]
  41. Lucklum, R. The Δf–ΔR QCM technique: An approach to an advanced sensor signal interpretation. Electrochim. Acta 2000, 45, 3907–3916. [Google Scholar] [CrossRef]
  42. García-Abuín, A.; Gómez-Díaz, D.; Navaza, J.M.; Regueiro, L.; Vidal-Tato, I. Viscosimetric behaviour of hyaluronic acid in different aqueous solutions. Carbohydr. Polym. 2011, 85, 500–505. [Google Scholar] [CrossRef]
  43. Larson, R.G. The rheology of dilute solutions of flexible polymers: Progress and problems. J. Rheol. 2005, 49, 1–70. [Google Scholar] [CrossRef]
  44. Bird, R.B.; Armstrong, C.R.; Hassager, O. Dynamics of Polymeric Liquids, 2nd ed.; Wiley-Interscience: New York, NY, USA, 1987; Volume 25. [Google Scholar]
  45. Balazs, E.A. The Physical Properties of Synovial Fluid and the Special Role of Hyaluronic Acid. Disord. Knee 1974, 2, 63–75. [Google Scholar]
  46. Nakamoto, T.; Kobayashi, T. Development of circuit for measuring both Q variation and resonant frequency shift of quartz crystal microbalance. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1994, 41, 806–811. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Change in the morphology of the conductance curve for bare crystal and in contact with a liquid sample.
Figure 1. Change in the morphology of the conductance curve for bare crystal and in contact with a liquid sample.
Sensors 16 01959 g001
Figure 2. Shear Stress and Shear Rate for Newtonian (red) and Pseudoplastic (black) fluids.
Figure 2. Shear Stress and Shear Rate for Newtonian (red) and Pseudoplastic (black) fluids.
Sensors 16 01959 g002
Figure 3. Factor χ for polymeric fluids (pseudoplastic behavior).
Figure 3. Factor χ for polymeric fluids (pseudoplastic behavior).
Sensors 16 01959 g003
Figure 4. (a) Simplified Diagram Acquisition System of the Biosensor; (b) Blocks Diagram System of the Biosensor.
Figure 4. (a) Simplified Diagram Acquisition System of the Biosensor; (b) Blocks Diagram System of the Biosensor.
Sensors 16 01959 g004
Figure 5. Gamry Cell (a) and Prototype System (b).
Figure 5. Gamry Cell (a) and Prototype System (b).
Sensors 16 01959 g005
Figure 6. QCR biosensor performance vs. Fungilab viscometer performance.
Figure 6. QCR biosensor performance vs. Fungilab viscometer performance.
Sensors 16 01959 g006
Figure 7. Response curve—Δf vs. Concentration of the glycerol samples (Newtonian behavior).
Figure 7. Response curve—Δf vs. Concentration of the glycerol samples (Newtonian behavior).
Sensors 16 01959 g007
Figure 8. Response curve of the QCR biosensor vs. the square root of the density-viscosity product (a) and Response curve of the QCR biosensor vs. concentration (b).
Figure 8. Response curve of the QCR biosensor vs. the square root of the density-viscosity product (a) and Response curve of the QCR biosensor vs. concentration (b).
Sensors 16 01959 g008
Figure 9. Apparent viscosity obtained by Rouse model vs. viscometer measures.
Figure 9. Apparent viscosity obtained by Rouse model vs. viscometer measures.
Sensors 16 01959 g009
Table 1. Macroscopic classification of synovial samples.
Table 1. Macroscopic classification of synovial samples.
NormalInflammatorySepticHemorrhagic
AppearanceTransparentTurbidTurbidBloody
ColorStraw-coloredYellow-cloudyOpaqueRed-opaque
ViscosityHighLowLowVariable
Table 2. Glycerol solutions results (average ± SD).
Table 2. Glycerol solutions results (average ± SD).
Concentration (μg/mL)ΔFs (Hz)ΔΓ (Hz)Density (mg/mL)Viscosity (mPa·s)
400.0−3660 ± 493697 ± 2311063.1679 ± 0.08
300.0−2968 ± 513042 ± 1210542.1935 ± 0.07
200.0−2617 ± 862619 ± 1410481.6643 ± 0.05
100.0−2281 ± 482287 ± 3310231.2470 ± 0.01
50.0−2064 ± 322076 ± 2110101.0525 ± 0.03
25.0−2011 ± 222032 ± 289991.0097 ± 0.03
12.5−2004 ± 542033 ± 3310070.9544 ± 0.01
6.3−1920 ± 491926 ± 2110060.9020 ± 0.01
3.1−1911 ± 921943 ± 189980.9140 ± 0.06
Table 3. Hyaluronic Acid (H.A) results (Average ± SD).
Table 3. Hyaluronic Acid (H.A) results (Average ± SD).
Concentration H.A (mg/mL)F0 (Hz)ΔF (Hz)ΔΓ (Hz)ΔRDensity (mg/mL)Viscosity (η0) (mPa·s)
7.009983452 ± 45−3411 ± 113615 ± 1524.6 ± 0.421032.23809
3.509983394 ± 54−3381 ± 123563 ± 1321.1 ± 0.481017.89265
1.299983373 ± 33−3315 ± 83463 ± 1518.5 ± 0.521014.3291
0.859983428 ± 44−3269 ± 63341 ± 916.2 ± 0.361019.5368
Table 4. Apparent viscosity obtained with the QCR biosensor and Fungilab viscometer.
Table 4. Apparent viscosity obtained with the QCR biosensor and Fungilab viscometer.
Concentration H.A (mg/mL)QCR-Viscosity (mPa·s)Viscometer (mPa·s)
73.25748.12
3.53.13186.60
1.292.9950.00
0.852.8537.78
Table 5. Factor χ.
Table 5. Factor χ.
Concentration H.A (mg/mL)τ (μs)η′/η0η″/η0tan(δ)χ
7.000.300.152.840.050.06
3.500.100.271.660.160.15
1.290.040.460.990.460.32
0.850.030.540.850.630.39
Table 6. ΔF0 and equivalent viscosity related only to the viscous component of hyaluronic acid mixtures.
Table 6. ΔF0 and equivalent viscosity related only to the viscous component of hyaluronic acid mixtures.
Concentration H.A (mg/mL)ΔF0 (Hz)Equivalent Viscosity QCR (mPa·s)Viscometer (mPa·s)% Error with Viscometer
7.00−53603683.6748.129%
3.50−23175129.6186.6031%
1.29−1047626.850.0046%
0.85−832316.737.7856%

Share and Cite

MDPI and ACS Style

Ahumada, L.A.C.; González, M.X.R.; Sandoval, O.L.H.; Olmedo, J.J.S. Evaluation of Hyaluronic Acid Dilutions at Different Concentrations Using a Quartz Crystal Resonator (QCR) for the Potential Diagnosis of Arthritic Diseases. Sensors 2016, 16, 1959. https://doi.org/10.3390/s16111959

AMA Style

Ahumada LAC, González MXR, Sandoval OLH, Olmedo JJS. Evaluation of Hyaluronic Acid Dilutions at Different Concentrations Using a Quartz Crystal Resonator (QCR) for the Potential Diagnosis of Arthritic Diseases. Sensors. 2016; 16(11):1959. https://doi.org/10.3390/s16111959

Chicago/Turabian Style

Ahumada, Luis Armando Carvajal, Marco Xavier Rivera González, Oscar Leonardo Herrera Sandoval, and José Javier Serrano Olmedo. 2016. "Evaluation of Hyaluronic Acid Dilutions at Different Concentrations Using a Quartz Crystal Resonator (QCR) for the Potential Diagnosis of Arthritic Diseases" Sensors 16, no. 11: 1959. https://doi.org/10.3390/s16111959

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop